From lab to production, providing a window into the process. Universal Input Indicator Start-up Guide. Operating Manual.

Similar documents
Model No Universal Input Indicator Start-up Guide

Operating Manual. Dynisco /8 DIN Indicator Concise Product Manual

Instruction Manual. Instrument Amplifier HE 017

From lab to production, providing a window into the process. Dynisco's ATC990 and UPR900 Applications and Setup

IPM500 Series QuickStart Manual

1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882

Veeder-Root brand Series S628 Strain Gauge (S628-6XXX)

-DIN & DIN RaPID TEMPERATURE CONTROLLERS. Product Manual (June 1999)

PX Series Balances. Quick Start Guide. Please download the user manual from

Series S628 AC Volts/Amps (S628-3XXX) Veeder-Root brand. Technical Manual

Truckweigh. operators sheet STGGLOBAL.NET

Omni. User Manual. Economic Self-Tune PID Temperature Controller. Omni 96. Omni 72 72X72. Omni ~ 265 V AC SUPPLY

7561-PSD Manual Portable Battery Powered Indicator

DUAL COLOUR DISPLAY DC PROCESS INDICATOR. Product Manual

IPM650 Intelligent Panel-Mount Display

5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL Fruitland Ave Los Angeles, CA UM Channel Monitor.

DS 96/48 PK. Digital indicator. Manual 42/33-90 EN Rev. 00 DIM

Analyzer/Sensor to measure UV Spectral Absorbance

masibus : UT-94 Page 1 of 6 INTRODUCTION SPECIFICATIONS Dimensions Document Ref No. m98a/om/101 Issue No: 08

Mercury 11-10PV Pack / Condenser Controller Installation & User Guide V6 or Higher

M260 Load Cell Amplifiers

PROCESS ANALYSERS. SERVOTOUGH FluegasExact Gas Analyser. QuickStart Manual. Part Number: D Revision: 1 Language: UK English

6-Channel Monitor. Installation and Operation Manual

Digital Keypad Introduction

LAUREL. Laureate Digital Panel Meter for Load Cell & Microvolt Input ELECTRONICS, INC. Features. Description

LAUREL. Laureate Digital Panel Meter for Process and Ratiometric Signals ELECTRONICS, INC. Features. Description

Operating Manual. Dynisco ATC990 Graphical 1/4 DIN Process Controller Concise Product Manual

MYRIAD QLC 4-CHANNEL MONITOR/CONTROLLER INSTRUCTION MANUAL

Instruction Manual. ROYTRONIC EXCEL Series AD9 Programming Guide Electronic Metering Pumps

The PM1000 series is a universal 4 digit LED plug-on display for transmitters with 4-20mA 2 wire output and fitted with DIN43650 connector.

User Manual Digi-Sense 12-Channel Benchtop Data Logging Thermocouple Thermometer

AS Keypad User Manual

PROCESS ANALYSERS. SERVOTOUGH OxyExact (2200) Operator Manual. Part Number: 02210/001A Revision: 4 Language: UK English

Trident and Trident X2 Digital Process and Temperature Panel Meter

Security System User Guide

LAUREL ELECTRONICS, INC.

T80 Universal Transmitter S80 Intelligent Sensors

Contents. Temperature & Process Measurement Indicators Setup Guide

Model SC500. Programmable Single-Channel Transducer Indicator/Conditioner DESCRIPTION FEATURES

Using a Memory Card. Chapter 4. Chapter Objectives. Supported Memory Cards

Instruction Manual July sitrans RD200

SINGLE-CHANNEL PROCESS CONTROLLERS & TEMP. DISPLAYS D3820/D3830 Series

SERIES C 1ZC 16C 18C 19C 25C

TRACKER 240 SERIES. Load Cell and Weighing Indicators. A Precision Measurement Instrument with Outstanding Features

Multi Units Humidity/Temperature Transmitter. HygroViewer Model JW300 User Manual

MICRO SERIES PROCESS DISPLAYS LARGE DIGIT MODELS

5450 NW 33rd Ave, Suite 104 Fort Lauderdale, FL Fruitland Ave Los Angeles, CA SST7000 SST7100. Speed Switch / Transmitter

10 Thomas, Irvine, CA USA Tel: (949) Fax: (949) IPM650 Product Manual

Omni + User Manual. Dual Setpoint Temperature Controller with Programmable Input & TIMER. Omni 96+ Omni 72+ Omni 48+ TMR TMR TMR

D45 System. Digital colour entrance panel with addresses list

FG-3000R Digital Force Gauge Operation Manual

TC Type Range Conformity Error. J -210 C to +760 C (-347 F to F) ±0.09 C (±0.16 F) K -244 C to C (-408 F to F) ±0.1 C (±0.

Model M3LLC STRAIN GAUGE TRANSMITTER. PC CONFIGURATOR SOFTWARE Model: M3LLCCFG. Users Manual M3LLCCFG USERS MANUAL EM-9197-T

CII: Current Input, Isolated

Model T28 Transmitter

FG-3000 Digital Force Gauge Operation Manual

AUTOMATION. Operator s Manual PG-7. Full Access. Doc Part Rev B, 07/18. Automation Products Group, Inc.

Calibration & Connectivity

Installing Keypad and Backplate

Intrinsically Safe Temperature Concentrator System

Operating Instructions tico 735 AC Volts/Amps Indicator

Elevation Window Control System

1. Installation. 2. Configuration - Operation. 3. Specifications. Safety

USER INSTRUCTION MANUAL FOR LOADCELL TRANSMITTER MODEL TDC/I/0550 (SOFTWARE: VER2A) INDEX

LAUREL. Laureate RTD Temperature Panel Meter / Controller ELECTRONICS, INC. Features. Description. Specifications

Introduction. Features. Index

Digital Bench Scale. Revision 8.93 August 12, 1993

DUAL COLOUR DISPLAY TEMPERATURE INDICATOR. Product Manual

Installation and User Manual

RITZ / Turbine controls DISPLAY

SHIMPO INSTRUMENTS DIGITAL PANEL METER OWNERS MANUAL

TRACKER 220 SERIES. Digital Panel Indicators. A Complete Range of Universal Input Digital Panel Indicators for Temperature and Process Measurement

SERIES CMT CARBON MONOXIDE GAS TRANSMITTER

The Tracker 220 Series A Complete Range of Universal Input Digital Panel Indicators for Temperature and Process Measurement

HL100 Controller Operator's Manual

Master-Touch INSTRUCTION MANUAL. Series 9000MP Multipoint System Series 9600MP System Control Panel Thermal Gas Mass Flowmeters. version 7.

12-36 VDC/12-24 VAC Power Option 4-Digit Display, 0.56 (14.2 mm) or 1.20 (30.5 mm)

HANDHELD CALIBRATOR-MULTIMETER OC502-t

Trident & Trident X2 PD765

GS100P. Weighing Controller. Contents. Order Codes

- SELF-TUNING PID ALGORITHM - INTUITIVE COLOR DISPLAY WITH TEXT MESSAGING - UNIVERSAL PROCESS AND TC/RTD INPUTS - MULTI- FUNCTION RAMP- DWELL/PROCESS

PD765 TRIDENT. Trident Series Process & Temperature Meters.

M A C 3 Wind Speed Alarm & Controller

Quick Start Guide. Cyl-Tel & Tank-Tel Liquid Level Gauges. Designed and Built by: Chart Inc th Street NW New Prague, MN USA (800)

3690 N.W. 53rd Street Fort Lauderdale, FL SC-2124 SC-2124M. 24 Channel Universal Scanner. Installation and Operation Manual P/N 145F-11902

Communications and Displays. SITRANS RD300 - Dual-Input Process Meter Operating Instructions 02/2014

OLS VT10 SERIES STANDARD DUAL DISPLAY PID AUTOTUNE RAMP-TO-SETPOINT ULC, UL, CE (4810) AUTO / MANUAL 15 SECURITY OPTIONS UNIVERSAL INPUT

MTII4200 Level Transmitter Installation, Operation & Maintenance Instructions

+GF+ SIGNET ph/orp Transmitter Instructions

Operation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller

PD765 TRIDENT. Trident Series Process & Temperature Meters.

User Manual M1. Thermocouple type K, B, S, N, E, T, R, L, J. Technical features:

UDC 1000 and UDC 1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS

QUICKSTART GUIDE. Infrared Gas Analyzer. Teledyne Analytical Instruments

Brooks Service Tool TM for QUANTIM Precision Mass Measurement and Control

Displays Transducers. Displays and transducers

Iso DIN. User Manual. 1 channel Earth Fault Monitoring

installation quick guide integrated access control & environmental monitoring

TRASK SERIES 2 DIGITAL PANEL METER OWNERS MANUAL

Operating Instructions VEGAMET 381

Transcription:

Universal Input Indicator Start-up Guide Operating Manual -1-

Contents 1. Setting up a unit straight out of the box 1.1. Entry into Configuration mode 1.2. Scrolling through Parameters and Values 1.3. Changing Parameter Values 2 Calibration Mode 2.1 Entry to Calibration Mode 3 Setup Mode 3.1 Entry into the Setup Mode 3.2 Scrolling through Parameters and Values 3.3 Changing Parameter Values 4 Operator Mode 4.1 Entry into Operator Mode 4.2 Scrolling through Parameters and Values 4.3 Changing Parameter Values 1/8 Din Indicator Units Display 5 Alarm Indications 5.1 Resetting Latched Alarm Outputs 5.2 Resetting Alarm 1 Active Time, Minimum PV or Maximum PV 6 Tare Feature 3 3 3 3 7 7 8 8 9 9 10 11 11 11 12 12 12 13-2-

1. Setting up a unit straight out of the box 1.1. Entry into Configuration mode When the unit is first powered on, the message goto Conf, will appear on the screen. This is the first step to set up the unit for the functionality required by the user. To enter configuration mode press the key, this will then prompt you to enter an unlock code. ULoc will appear followed by 0. To enter into the configuration mode the user must enter the correct unlock code using the and keys. The default unlock code is 20, if you do not enter the correct code the unit will revert back to the previous screen asking you to enter the code again. If you forget any of the unlock code there is a hidden read only menu for them. To enter this mode you must power the unit down, whilst powered down you must press the and, keeping them pressed whilst repowering the unit for 10-15 seconds. You will then enter a read only loc code view. If not from first power up Configuration is entered from Select Mode Hold down and press to force the controller into the Select Mode. The SLCT legend is shown for 1 second, followed by the legend for the current mode. Press or to navigate to the Configuration Mode option, then press. Set LED. This flashes in Configuration Mode. 1.2. Scrolling through Parameters and Values Press to scroll through the parameters. While this key is pressed, and up to 1 second after, the parameter legend is shown, followed by the current parameter value. Only parameters that are applicable to the hardware options chosen will be displayed. 1.3. Changing Parameter Values Press to navigate to the required parameter, then press or to set the value as required. Once the desired value is set, press to display YES?, press within 10 seconds, accept the change, otherwise parameter will revert to previous value. Or Press to reject the change and to move onto the next parameter. Hold down and press to return to Select Mode. If there is no key activity for 2 minutes the instrument returns to the operator mode. -3-

1. 1480 Configuration Mode Parameters -4-

-5-

-6-

*Linear Outputs can be configured to provide an adjustable 0.0 to 10.0VDC transmitter power supply for external devices. 2. Calibration Mode 2.1. Entry to Configuration mode Configuration mode must be completed before adjusting Calibration parameters. First select Calibration mode from Select mode. Hold down and press to force the controller into the Select Mode. The SLCt legend is shown for 1 second, followed by the legend for the current mode. Press or to navigate to the Calibration Mode option, then press. You then need to enter the unlock code using the or keys, then press to enter the mode. Press to scroll through the parameters (while this key is pressed, and for 1 sec after, the parameter legend is shown, then the current value). Press or to change the value. To exit from Calibration mode, hold down and press to return to Select mode. Entry into Calibration Mode is security-protected by the Calibration Mode lock code. Default value is IO. Calibration mode will only be displayed if input type is set to ST_G -7-

When the calibration procedure begins ---- appears on the screen. Once Calibration is complete done appears on screen. If there are any Faults with the calibration an error message will appear either Er_r or Er_C. Er_C means the low calibration will fail if the offset is less than -10mV or greater than +10mV. This signifies potential faulty sensors or the high calibration will fail if the count value is less than +20mV or greater than +50mV. This signifies potential faulty sensors Er_r means the high calibration will fail if the mv value is within 10mV of the low calibration value. This is a potential RCAL failure. Setup Mode This mode is normally selected only after Configuration Mode has been completed, or is used when a change to the process set up is required. These parameters must be set as required before attempting to use the indicator in an application. 3 Setup Mode 3.1 Entry into the Setup Mode Setup Mode is entered from Select Mode Hold down and press to force the controller into the Select Mode. The SLCt legend is shown for 1 second, followed by the legend for the current mode. Press or to navigate to the Setup Mode option, then press. Entry into Setup Mode is security-protected by the Setup Mode lock code. Default value is IO. Set LED. This is on in Setup Mode. 3.2 Scrolling through Parameters and Values Press to scroll through the parameters. While this key is pressed, and up to 1 second after, the parameter legend is shown, followed by the current parameter value. -8-

3.3 Changing Parameter Values Press to select the required parameter, then press or to set the value as required. Once the displayed value is changed, it is effective is immediately. No confirmation of the change is required. Press to move onto the next parameter. Hold down and press to return to Select Mode. If there is no key activity for two minutes the instrument returns to the operator mode. -9-

Alarm parameters marked * are repeated in Configuration Mode. **Once the complete list of Set Up Mode parameters has been displayed, the Operator Mode displays are shown without exiting from Set Up Mode. 4 Operator Mode This is the mode used during normal operation of the instrument. It can be accessed from Select Mode, and is the usual mode entered at power-up. The available displays are dependent upon the setting of the Display Strategy parameter in Configuration Mode. WARNING: IN NORMAL OPERATION, THE OPERATOR MUST NOT REMOVE THE INSTRUMENT FROM ITS HOUSING OR HAVE UNRESTRICTED ACCESS TO THE REAR TERMINALS, AS THIS WOULD PROVIDE POTENTIAL CONTACT WITH HAZARDOUS LIVE PARTS. -10-

CAUTION: Set all Configuration Mode parameters and Set Up Mode parameters as required before starting normal operations. 4.1 Entry into Operator Mode This is the normal operating mode of the instrument from power-up. It can also be accessed from any other mode via Select Mode as follows: Hold down and press to force the controller into the Select Mode. The SLCt legend is shown for 1 second, followed by the legend for the current mode. Press or to navigate to the Operator Mode option, then press. 4.2 Scrolling through Parameters and Values Press to scroll through the parameters. While this key is pressed, and up to 1 second after, the parameter legend is shown, followed by the current parameter value. 4.3 Changing Parameter Values Press to select the required parameter, then press or to set the value as required. Once the displayed value is changed, it is effective is immediately. No confirmation of the change is required. Press to move onto the next parameter. The operator can freely view the parameters in this mode, but alteration depends on the Display strategy setting in Configuration Mode. All parameters in Display strategy 6 are read only, and can only be adjusted via Setup mode. -11-

1/8 Din Indicator Units Display The 1480 1/8 Din indicator has an additional Units Display. In Operator Mode, this display shows C or F when a temperature input range is displayed, and is blank for strain gauge or linear inputs. The units display is also used in other modes as a confirmation of the parameter type currently shown in the main display. 5 Alarm Indications The alarm status screen indicates any active alarms, in addition the associated Alarm LED flashes. For latching alarm outputs, the LED FLASHES when the alarm condition exists, and goes to ON when the alarm condition is no longer present if the output has not yet been reset, to indicate that the relay is in the Latched on condition. 5.1 Resetting Latched Alarm Outputs Latched outputs can be reset whilst the Process variable or Alarm Status screens are displayed, via the Digital Input (if fitted), from the front keypad as follows: Press either or to reset the latched relay(s). Outputs will only reset if their alarm condition is no longer present. CAUTION: A reset will affect ALL latched outputs. 5.2 Resetting Alarm 1 Active Time, Minimum PV or Maximum PV The stored Maximum PV value, Minimum PV value or Alarm 1 active Elapsed Time value can be reset via the Digital Input (if fitted), with a communications command via the RS485 module (if fitted) or from the front keypad as follows: Press to select the parameter to be reset. Press either or for three seconds. The display briefly shows ---- when the value is reset before the unit reverts to the requested display. -12-

Multi-Point Scaling When Multi-Point Scaling is enabled (MPS = EnAb in Configuration Mode), up to 9 breakpoints can be set to linearize the input signal. This only applies to ma, mv or Voltage input types. For each breakpoint the input scale value (ScAn) is entered in % of input span, followed by the value to be shown (disn) in display units. Each breakpoint s input scale value must be higher than the previous value, but the display values can be either higher or lower. Any scale value set to 100% becomes the last in the series. 6 Tare Feature When Tare is enabled (tare = EnAb in Configuration Mode), it can be used to set the displayed value to zero automatically, by making the PV Offset parameter equal, but opposite to, the current process variable value. Tare can be initiated via the Digital Input (if fitted), or by using the following key press sequence: Press until the process variable is displayed. Hold down and together for three seconds until the display shows YES? Release both keys and press within 3 seconds to confirm the request. The Tare request is aborted if this sequence is not followed exactly. -13-