testo Saveris Adjustment software Instruction manual

Similar documents
testo Saveris Adjustment software Instruction manual

testo Comfort Software Basic 5 Instruction manual

testo Comfort Software Basic 5

testo EasyClimate Software Instruction manual

testo Comfort Software Professional 4 Instruction manual

testo Comfort Software Professional 4 Instruction manual

Measurement data monitoring with testo Saveris PROF. Startup instruction

testo Comfort Software Professional 4

Installing TAM Air Assistant Software TM

QUICKBOOKS PRO 2008 INSTALLING QUICKBOOKS PRO AND STUDENT DATA FILES

testo 184 Data logger Instruction manual

testo 175 Data loggers Instruction manual

Connected to the FP World

QUICKBOOKS PRO 2009 INSTALLING QUICKBOOKS PRO AND STUDENT DATA FILES

Fiery Network Controller for Xerox WorkCentre 7800 Series. Welcome

Firmware, Database, & PC Application Update Installation Instructions

3001D Smart Safe Manual

3001D Smart Safe Manual

Operating manual. GTL - Configuration tool. Please keep the manual for future use.

Software EasyKool. Instruction manual

RICOH Driver Installation Guide

USB-500 Series Getting Started Contents

testo 176 Data loggers

DRAWings 8 and Wings modular 8 Embroidery Software INSTALLATION GUIDE

testo Sound Level Meter Instruction Manual

ModemUSB/E12 User Manual v0.1

PHARO User Configuration Software UCS USER MANUAL Rev.0-15/07/ pharo.reer.it

This is a GENERAL Servant Keeper Network Installation help sheet. If you need further assistance, please contact your network administrator.

USER MANUAL DHP-P500AV/DHP-P501AV VERSION 1.2

OKIFAX 5650 MFP Option Kit. Overview. Setup Guide

IP NETWORK CAMERA CMP-NWIPCAM20 CMP-NWIPCAM30

EX700i Print Server/Integrated Fiery Color Server. Welcome

KV-SS090. Operating Instructions. Instant Scanning Software. Model No.

BAUDTEC TW263R4 ADSL2+ MODEM ROUTER System Requirements

Table of Contents. Page ii

HD Analog Recorder. Quick Start Guide CJ-HDR104. Model No. CJ-HDR108. Version 1.0.2

Instruction manual. testo easyemission Software

Operating Instructions

UNIVERSAL SOFTWARE. Universal Software. Data Sheet

User s Guide. CD Library II Software for Windows XP and Vista. CD Library II 4.0 Software CD Library software for CD Library products

PL-85PE 85Mbps High Speed HomePlug Ethernet Adapter

SCT 3000 Smartline Configuration Toolkit Model SCT 101 Specifications

See Clean. Installation and Set Up Guide V with Confidence. 3M Clean-Trace Data Trending Software. 3M Food Safety

Table of Contents. Contacting Technical Support Warranty Registration Hardware Installation...8. Using the Setup Wizard...

Preface. Manual Revisions. Trademarks

USB-Link Technical Guide

On-Line Order Booking Software Online Printer s s Manual

Firmware, Database, & PC Application Update Installation Instructions

Operating Manual. Windows Operating Software GSOFT 3050 as of version 3.0. for Handheld Instruments GMH 3000 / GMH 5000 / GHMsilverline

INIT USER S MANUAL (1)

GE Sensing. Kaye RF ValProbe. Wireless Thermal Validation System. User s Guide

EM4029 R1 10/100/1000 Mbps PCI-e network adapter

Table of Contents. Configuration...13 Setup Security...17 Application QoS...18 Advanced QoS...19 Reset Troubleshooting...

testo 176 Datenlogger testo 176 Data loggers testo 176 Enregistreur de données Bedienungsanleitung Instruction manual Mode d'emploi

TELTONIKA ModemUSB/G10 User Manual v0.1

USB 3.0 Dual Head Display Adapter

Quick Installation Guide TU-S9

Printing Your First Page. Attaching the Paper Support. Plugging in the Printer. Checking the Printer

EW7033 R2 USB 3.1 / 3.0 Enclosure for 2,5 SATA hard disks

CommLink IV Technical Guide

Version /20/2012. User Manual. PowerLine AV 500 Network Starter Kit DHP 501AV

Operating Instructions Driver Installation Guide

USB 3.0 Docking Station

Bluetooth Micro Dongle User s Guide. Rating: 5V DC 80mA Made in China

USB 3.0 Single/Dual Display Docking Station. User's Manual

Amcrest Eco HDCVI DVR Quick Start Guide

BiPAC 7800VDP(O)X. Dual-band Wireless-N VoIP ADSL2+ (VPN) Router. Quick Start Guide

USER MANUAL TOUGH DRIVE EXTERNAL MOBILE HARD DRIVE / 2.5" / USB 2.0. Rev. 848

Comfort software x35 Professional. Instruction manual. testo x35 Professional

Daily Monthly Report Software EcoMeasure III. User manual (Detailed version) Model MES3-SW1-DR-EN

Testo USB driver Windows 7, Windows 8.1, Windows 10. Application information

LUDLUM MODEL 2350(-1) Windows Interface Version LMI Part #: October 2015

Quick Start Manual. Color Image Scanner

Operating Instructions

HygroLog HL20 and HL21 Data Logger. User Guide

Network Video Recorder Quick Start Guide

testo easyheat Configuration and Analysis software Instruction manual

TEKLYNX LABELVIEW A D M I N I S T R A T O R S G U I D E

BEC 5200 Series. (802.11g) ADSL2+ Router. Quick Start Guide

NDT Remote Control Library. User s Manual

BiPAC 5200 Series. (802.11g) ADSL2+ Firewall Router. Quick Start Guide

Setup Guide. Before Setting Up. Installation. Start-up/Shutdown

Getting Started. Read Me First. series. Photo Printer

Digital Quick-Response Thermometer

User Manual. HOPPECKE sun powerpack premium. service software

PL200M. 200Mbps powerline adapter. User Manual

ECLIPSE LV Series Support Tools

GMH Operating Manual for Digital Quick-Response Thermometer

HI-SPEED USB 2.0 FILE TRANSFER CABLE USER MANUAL

DataNet Installation Guide + Quick Start Guide Updated: August 2008

Amcrest HDCVI DVR Quick Start Guide

Installing the Printer Software

Network Guide NPD EN

testo Saveris Update Instruction to Version testo Saveris 4.5 Sp3

Table of Contents. Hardware Installation...7 Push Button Security Using the Setup Wizard... 11

Table of Contents. Hardware Installation...7 Push Button Security...8. Using the Setup Wizard...9

F-Secure Client Security. Quick Installation Guide

NT2 U3. 2-Bay RAID Storage Enclosure. User Manual May 18, 2010 v1.1

testo 6651 Humidity transmitters testo 6600 Probes P2A software Parameterizing, adjusting and analyzing software Instruction manual

BIPAC Broadband Firewall Router. Quick Start Guide

Transcription:

testo Saveris Adjustment software Instruction manual

2

1 Contents 1 Contents 1 Contents... 3 2 About this document... 4 3 Specifications... 5 3.1. Use... 5 3.2. Scope of delivery... 5 3.3. System requirements... 5 4 First steps... 7 4.1. Software installation... 7 4.2. Installing the driver... 7 4.3. Supplying the Ethernet probe with power... 8 4.4. Connecting the Ethernet probe to the computer... 8 4.5. Supplying the radio probe with power... 9 4.6. Connecting the radio probe to the computer... 9 4.7. Starting the software... 10 5 Product description... 12 6 Using the product... 13 6.1. Allocation of password... 13 6.2. Preparing for adjustment... 14 6.3. Performing the adjustment:... 16 6.4. Printing the adjustment values... 17 7 Tips and assistance... 18 3

2 About this document 2 About this document Use > Please read this documentation through carefully and familiarize yourself with the product before putting it to use. Pay particular attention to the safety instructions and warning advice in order to prevent injuries and damage to the products. > Keep this document to hand so that you can refer to it when necessary. > Hand this documentation on to any subsequent users of the product. Knowledge of Windows operating systems is required when working with the software. The description in this instruction manual relates to Windows 10, Windows 8, Windows 7 Symbols and writing standards Representation 1.... 2.... Explanation Note: Basic or further information. Action: more steps, the sequence must be followed. >... Action: a step or an optional step. -... Result of an action. Menu Elements of the program interface. [OK] Buttons of the program interface....... Functions/paths within a menu.... Example entries 4

3 Specifications 3 Specifications 3.1. Use The Saveris adjustment software is used to adjust the radio/ethernet probes connected to the Saveris base. Using the Saveris adjustment software and a reference measuring instrument, the temperature and the relative humidity of each individual probe can be adjusted. After successful adjustment, the current adjustment data is stored in the probe. At the same time, the adjustment software accepts this data so that the adjustment histories are available. The testo 400/650 with precision humidity probe (order no. 0636 9741) or the precision temperature probe (order no. 0614 0240) is recommended as the reference measuring instrument for adjusting the probes. 3.2. Scope of delivery The following components are included in the delivery: CD with testo Saveris adjustment software including instruction manual. Adjustment cable for connecting the radio probes to the USB adapter. USB cable including USB adapter for connecting the Ethernet and radio probes to the computer. 3.3. System requirements Computer For smooth work with the software, the following requirement should be met: Pentium processor of at least 1.2 GHz or equivalent 256 MB RAM 50 MB unused hard drive capacity CD-ROM drive USB 2.0 interface Internet Explorer 5.5 Service Pack 1 or higher 5

3 Specifications Operating system The adjustment software will function on the following 32-bit and 64-bit operating systems: Windows 7 Windows Server 2008 R2 Windows 8 Windows 8 PRO Windows 8 Enterprise Windows 10 6

4 First steps 4 First steps Administrator rights are required for installation. 4.1. Software installation 1. Insert the program CD into the CD-ROM drive of the computer. - Installation wizard starts automatically > If the installation wizard does not start automatically: Open CD drive in the Windows Explorer start Setup.exe (double-click on left mouse button). 2. Follow the directions of the installation wizard. 3. To stop the software installation: Click on [Finish]. - The software was installed successfully on the computer. 4.2. Installing the driver The drivers required for the Saveris adjustment software can be found on the program CD for the Saveris software (SEE/Prof/CFR). 1. Insert program CD for the Saveris software (SEE/Prof/CFR) in the CD drive. 2. If the installation program starts automatically: click on [Cancel]. 3. To start the Driver.exe file: > Select the CD drive Open the USBDriver folder Open the EthernetProbe folder Start TestoSetup.exe. - The installation wizard opens. 4. Follow the directions of the installation wizard. 5. To finish the driver installation: click on [Finish]. - The driver for the Saveris adjustment software was successfully installed on the computer. 7

4 First steps 4.3. Supplying the Ethernet probe with power 1. Open the cover ➊ for the power supply. 2. Connect mains cable ➋ to the Ethernet probe. 3. Insert mains plug into a socket. - The Ethernet probe is supplied with power 4.4. Connecting the Ethernet probe to the computer 8

4 First steps 1. Open the cover ➊ for the service interface. 2. Connect the USB cable ➋ to the testo USB adapter ➌ and insert in the service interface ➍. 3. Connect the USB cable to the computer. - The Ethernet probe is connected to the computer. 4.5. Supplying the radio probe with power The radio probe is supplied with power via the adjustment cable as soon as it is connected to the computer. 4.6. Connecting the radio probe to the computer 1. Loosen screws on the rear of the probe. 2. Remove housing cover of radio probe. 3. Remove batteries. 9

4 First steps 10 4. Connect the USB cable ➊ to the testo USB adapter ➋. 5. Connect the testo USB adapter ➋ to the adjustment cable ➌. 6. Connect the adjustment cable ➌ to the radio probe as shown. 7. Connect the USB cable ➊ to the computer. - The radio probe is connected to the computer. 4.7. Starting the software Please observe the following notes before starting the Saveris adjustment software: One probe (Ethernet or radio) must be connected to the computer, see Connecting the Ethernet probe to the computer page 8 or Connecting the radio probe to the computer page 9. For secure data transmission, the probes must be supplied with power, see Supplying the Ethernet probe with power page 8 or Supplying the radio probe with power page 9. The user interface of the software is opened in the language of the operating system if this is supported. If the operating system language is not supported, the user interface is in English. Windows program menue 1. Windows 7 > Click on [Start] All programs Testo Saveris Justage Software (double-click on left mouse button). Windows 8 > [Start] right mouse button Search Enter the application name in the search field click on Saveris Justage Software (double-click on left mouse button). Windows 10 / Vista > Click on [Start] Alle apps Testo Saveris Justage Software (double-click on left mouse button). 2. When the User Account Control window opens: click on [Continue]. - The screen asking for a user name and password opens. Please observe the following notes with regard to the user name and password: The user name and password are not linked to each other. The user name is used to record the adjustment

4 First steps process. It is stored in the probe together with the adjustment data to record who has performed an adjustment. The password protects the adjustment data in the probe against unauthorized modification. The password is stored in the probe and a separate password is assigned to each probe. 3. Enter the user name. 4. Enter the password for the probe connected. If you do not yet have an individual password for the probe: enter "testo". 5. Click on [OK]. - The adjustment software starts. 11

5 Product description 5 Product description User interface 1 Tab: The view of the work space changes according to the tab selected ([New] or [Current]). [New] tab: Editing is performed and the probes readjusted under the [New] tab. [Current] tab: Shows the current adjustment values of the probe Channel 1 = information for adjusting the temperature Channel 2 = information for adjusting the relative humidity 2 Work space: Information is shown and editing performed on the work space. 12

6 Using the product 6 Using the product 6.1. Allocation of password Allocate a probe -specific password to protect the adjustment data against unauthorized access. Please observe the following notes regarding the allocation of passwords: The password protects the adjustment data in the probe against unauthorized modification. The password is stored in the probe and a separate password is assigned to each probe. Once a password has been allocated, the adjustment software can only be opened if the password for the respective probe connected is entered. The password must not have more than 16 Latin characters. 1. Click on the [New] tab. 2. Under the list of functions, click on Password allocation. > If no password has yet been allocated: enter a new password in the text field. > If a password has already been allocated: Click on [Reset password] and enter a new password in the text field. 3. Click on [Accept new password]. - The new password is accepted. 13

6 Using the product 6.2. Preparing for adjustment Please observe the following notes before performing any adjustment: Adjustment data must always be reset. Current measured values of probe must be determined. Resetting the adjustment data 1. Click on the [New] tab. 2. In the list of functions, click on Adjustment data. 3. Click on [Reset adjustment data]. - The confirmation request is opened. > If you wish to reset the adjustment data: click on [Yes] (required for a new adjustment of the probe). > If you wish to cancel the process: click on [No]. 14

6 Using the product Determining the current measured values 1. Expose reference measuring instrument and the probes to be adjusted to the same constant ambient conditions. 2. Click on the [New] tab. 3. In the list of functions, click on Meas. values. 4. Click on [Read in meas. values]. 5. Wait until the current measured values of the probe are displayed in the table. - Channel 1 shows the measured temperature values ( C). - Channel 2 shows the measured relative humidity values (RH%). 6. Wait for equalization period to elapse, until the reference instrument and the probe supply stable measured values. Alternatively, you can also read the measured values from the display on the probe (if present). 15

6 Using the product 6.3. Performing the adjustment: Requirement: The adjustment data have been reset, see Resetting the adjustment data page 14. Current measured values of probe have been determined, see Determining the current measured values page 15. 1. Click on the [New] tab. > If you wish to adjust the temperature ( C) of the probe: click on Channel 1 in the list of functions or > If you wish to adjust the relative humidity (RH%) of the probe: click on Channel 2 in the list of functions. 2. Enter the data sheet number. Assign a unique data sheet number (e.g. probe number and date stamp, as otherwise there is a risk of using a wrong name causing the data to become useless). 3. Enter date of adjustment. 4. Enter text (e.g. product no. of reference measuring instrument or similar) 16

6 Using the product 5. Specify the number of adjustment points. 6. Enter the actual value from the data calculated (see list of functions under Meas. values) in the table (ACTUAL column). The current measured values of the probe must be recalculated for each adjustment point, see Determining the current measured values page 15. 7. Enter the nominal value for the reference measuring instrument in the table (NOMINAL column). - The adjustment software calculates the offset (difference between NOMINAL and ACTUAL values) > Enter date of next adjustment (reminder function). 8. To finish the adjustment: > Transfer adjustment data to the probe and store in the software, do not close the software: click on [Apply]. > Transfer adjustment data to the probe and store in the software, close the software: click on [OK]. The adjustment data are stored under the following path: Windows XP: C:\Documents and Settings\All Users\Documents\Saveris adjustment software. Windows Vista: C:\users\public\documents\saveris adjustment software. Windows 7: C:\users\public\public documents\saveris adjustment software. The adjustment data can be imported into the Saveris software (see Instruction Manual Saveris software). > Cancel the adjustment without transferring the data to the probe and storing it in the software, close the software: click on [Cancel]. 6.4. Printing the adjustment values 1. Click on the [Current] tab. - The current adjustment values for the probe are displayed. 2. To change between the temperature and relative humidity values: Click on Channel 1 or Channel 2 in the list of functions. - The adjustment values for the temperature (channel 1) or relative humidity (channel 2) are displayed. 3. Click on [Print]. - The adjustment values for the temperature ( C) or relative humidity (RH%) are printed out. 17

7 Tips and assistance Performing print settings 1. Click on the [Current] tab. 2. Click on [Set up printer...]. 3. Select the printer and perform the settings. 4. Click on [OK]. - The settings are accepted. - The current adjustment values are shown. 7 Tips and assistance If you have any questions, please contact your dealer or Testo Customer Service. The contact details can be found on the back of this document or on the Internet at www.testo.com/servicecontact.. 18

0970 0460 en 04