Nano DSC and MCDSC. Site Preparation Guide

Similar documents
TAM Air Isothermal Calorimeter. Site Preparation Guide

Discovery SDT650. Site Preparation Guide. Revision A Issued March Page 1

TA Instruments Installation Requirements for Microcalorimeter Systems

TA Instruments Installation Requirements for Microcalorimeter Systems

Discovery Hybrid Rheometer

FOX-50 Heat Flow Meter

Thermal Diffusivity Instruments DLF-2 with EM-1600

TA Instruments Installation Requirements for Microcalorimeter Systems

Agilent 2100 Bioanalyzer System & Applications Site Preparation Checklist

Customer Responsibilities

SpiritStation by Napa Technology

Customer Responsibilities

WineStation by Napa Technology

Pre-installation guide for analytical systems User manual HPLC

Customer Responsibilities

NOTE: All facility requirements must be met before 3D Systems Customer Support can schedule a trip to your facility to install the printer.

Avanti HP Series Centrifuges Preinstallation

DIMENSION/WEIGHT kpa. 0 to 20 m / 0 to 60 ft 0 to 2 bar / 0 to 29 psi. 0 to 50 m / 0 to 160 ft 0 to 5 bar / 0 to 72.5 psi

Facility Requirements Guide. ProJet X60 Series. INCLUDES: ProJet 160 ProJet 260C ProJet 360 ProJet 460Plus ProJet 660Pro ProJet 860Pro

Model KB 400 Cooling incubators with compressor technology

Cary 50. Pre-Installation Manual

E-Tool Mobile Manager 6 Drawer Rugged Managed (Rev. 0)

Haze 800 DMX. User Guide. English

Customer Responsibilities

Dell MD1280 Storage Enclosure Getting Started Guide

NOTE: Use the System Setup program to view microprocessor information. For more information, see "Using the System Setup Program."

Mechanically Activated Device for Dry Run Protection in Applications with LORENTZ Solar Pump Systems

OmniSmart 230V 350VA 225W CE/IEC Medical Grade Line-Interactive UPS with Built-In Isolation Transformer

HMIPPH7A0701 Panel PC Performance - Hard Disk - 15'' - AC - 0 slot - fan

4600 Series RAID-Enabled

4700 Series RAID-Enabled Industrial PCs DS _RAID(C)

StorNext M440 Site Planning Guide

Cobalt 8 Pre-Installation & Site Preparation Guide

Junos WebApp Secure 5.0 Hardware Guide

Eco-Surge 8-Outlet Surge Protector, 8-ft. Cord, 2880 Joules, Tel/Modem/Coax/Ethernet Protection

E-Series Site Preparation Guide

Obtaining Documentation and Submitting a Service Request, page xvii Safety Warnings, page xvii Safety Guidelines, page xx

Micro GC Fusion Quick Start Guide

AVR Series 120V 550VA 300W Ultra-Compact Line-Interactive UPS with USB port

UPGRADE PROGRAM. V250 Series Pumps vs V301 Series Pumps. Technical Memo

AVR Series 120V 700VA 350W Ultra-Compact Line-Interactive UPS with USB port

Dell Inspiron 1300 Basic Features, Affordable Widescreen DELL INSPIRON 1300 REVIEWERS GUIDE. Basic Features, Affordable Widescreen

TA Instruments New Features in TAM Air Assistant TM and TAM Assistant TM Software

FT-20 Fire Training Fog Machine User Manual English

EMC SYMMETRIX VMAX 10K

Isotemp. Bath Circulators. Improve the productivity. of your lab with solutions. from Fisher Scientific

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Luminex Prima

Breeze 2 Installation and Configuration Guide

Installation Job Aid for Ethernet Routing Switch 3600 Series

Owner s Manual. Isolate. Restore. Inspire! Power Conditioners Audio / Video Power Isolation Units Rack Mount / Consumer Series

AVR Series 120V 750VA 450W Ultra-Compact Line-Interactive UPS with USB port

SAGE SCIENCE INC. BluePippin Installation Guidelines Rev. A Originated March 9, 2015 Revised: Page 1 of 7. Prepared by: Alex Vira SUMMARY

Getting Started. Read Me First. series. Canon Inkjet Premium Photo Printer. 1 Preparation...P.1. Install the Print Head... P.3

New IdeaCentre tower systems with Intel processor technology deliver high-performance computing at a competitive price

Model KB 400 Cooling incubators with compressor technology

Internet Office 120V 350VA 180W Standby UPS, Ultra-Compact Desktop, USB

Pre-installation Manual

Protect It! 7-Outlet Surge Protector, 6-ft. Cord, 1080 Joules, Fax/Modem Protection, RJ11

ATS22D88S6U. ATS22D88S6U soft starter-ats22-control110v-power 208V(25hp)/230V(30hp)/460V(60hp)/575V(75hp) Product data sheet Characteristics.

Model #: ECO350UPS. 350VA Eco Energy-Saving UPS System. Highlights. Description. Features

Industrial PC - VL BPC

TOT Series Manual. North American CE and UK Models. CE Models. UK Models. North American Models

Omni LCD 650VA Tower Line-Interactive 120V UPS with LCD display and USB port

LABORATORY GRADE REMOTE PROGRAMMING SWITCHING MODE DC REGULATED POWER SUPPLIES SDP / 3636 / User Manual

Table of Contents. 1 Welcome. 2 Install your Access Point-I. 3 Using the Access Point-I

TD2100 Electric Gummed Tape Dispenser Owner s Manual

SmartPro LCD 120V 1500VA 900W Line- Interactive UPS, AVR, 2U Rack/Tower, LCD, USB, DB9 Serial, 8 Outlets

Model VDL 115 Vacuum drying chambers for flammable solvents

3M Pad Staff Workstation Model 895. Site Planning Guide

Installation and Operation Back-UPS Pro 900

SAGE SCIENCE INC. PippinHT Installation Guidelines Rev. B Originated November 21, 2014 Revised December 2, 2014.

Blue e+ chillers with the principle

ATS22D75S6 soft starter-ats22-control 220V-power 230V(18.5kW)/ V(37kW)/500V(45kW)

Dell Vostro 320 Setup and Features Information Tech Sheet

Protect It! 7-Outlet Surge Protector, 7-ft. Cord, 2520 Joules, Fax/Modem Protection, RJ11

Getting Started. Read Me First. series. Photo Printer

AVR Series 120V 700VA 350W Ultra-Compact Line-Interactive UPS with USB port

Model KMF 720 Constant climate chambers with expanded temperature / humidity range

UPGRADE PROGRAM. V70 Series Pumps vs V81T Series Pumps

MOBILE CONNECTOR - GEN 2 OWNER'S MANUAL

Internet Office 120V 750VA 450W Standby UPS, Ultra-Compact Desktop, USB

Model #: INTERNET750U

Sorensen DLM 600 Series W V 2 75 A. Half Rack Programmable DC Power Supply

USTER ZWEIGLE TWIST TESTER 5-SA

Dell Inspiron 300/400 Comprehensive Specifications

PDS100 Programmable Dispensing System SAFETY INSTRUCTIONS

Artico NAS Storage Appliance Site Planning Guide

Operating and monitoring systems Panel PC SERIES 400 Device platform EAGLE PM MT-436-TX

Datasheet.

AVR Series 230V 550VA 300W Ultra-Compact Line-Interactive UPS with USB port, C13 Outlets

APC Matrix-UPS 3kVA 208V/240V In 120/208/240V Out

TOT Series Manual. North American and International Models - without optional faceplate. Receptacle Panel of International Models, CE model shown

APC Smart-UPS C. Efficient, true sine-wave power protection for entry-level servers and network equipment. >

SmartPro LCD 120V 1500VA 900W Line-Interactive UPS, Extended Run, 2U Rack/Tower, LCD Display, USB, DB9 Serial

TA Instruments New Features in TAM Air Assistant TM and TAM Assistant TM Software

Installation Manual of DS-7600NI-H2/P Series. DS-7600NI-H2/P Series. Installation Manual UD.7L0202D1761B01

ATS22D62Q soft starter-ats22-control 220V-power 230V(15kW)/ V(30kW)

NEW APC Smart-UPS. Advanced line interactive power protection for servers and network equipment. The world s most popular network and server UPS.

TA Instruments New Features in TAM Air Assistant TM and TAM Assistant TM Software

ECO Series 120V 900VA 480W Energy-Saving Standby UPS with USB, Muted Alarm and 12 Outlets

Transcription:

Nano DSC and MCDSC Site Preparation Guide Revision A Issued May 2017

Table of Contents Table of Contents... 1 Ideal Setup... 2-3 MCDSC... 2 Nano DSC... 3 System Components... 4 Instrument Measurements... 5 Utility Requirements... 6 Computer Requirements... 7 Accessories... 8-9 MCDSC Water Circulator... 8 Nano DSC Autosampler... 9 Site Preparation Checklist... 10 TA Instrument Offices... 11 Revision A Issued May 2017 Page 1

Ideal Setup MCDSC: IDEAL PLACEMENT AND BENCH MEASUREMENTS Select a location with adequate floor space and a rigid laboratory bench that is level and is in a vibration-free environment. Bench length: 2.4 m (8 ft) Bench depth: 76 cm (30 in) min. Distance from the wall: 20 cm (8 in) min. Allow 38 cm (15 in) floor space and 20 cm (8 in) clearance at the rear of the water circulator for air circulation. A continuous low flow rate dry gas purge is advised when operating the MCDSC below the ambient temperature. Revision A Issued May 2017 Page 2

Ideal Setup NANO DSC: IDEAL PLACEMENT AND BENCH MEASUREMENTS Select a location with adequate floor space and a rigid laboratory bench that is level and is in a vibration-free environment. Bench length: 2.4 m (8 ft) Bench depth: 76 cm (30 in) min. Distance from the wall: 20 cm (8 in) min. Pressure-regulated gas supplied from a tank is required with Autosampler systems. An Autosampler requires an additional 65 cm (26 in) bench space to accommodate up to 6 bottles (0.5 L-4 L) and 1-2 waste outlet collection bottles (1 L-4 L). Additional floor space beneath the Autosampler is required for 4 L spent rinse fluid collection bottle. An additional 30 cm (1 ft) bench space is required to accommodate the Degassing Station. Revision A Issued May 2017 Page 3

System Components MAIN SYSTEM COMPONENTS A B A D B E F C A. Computer (Controller) B. Instrument C. External Water Circulator/Controller (MCDSC only) D. Degassing System (required for all Nano DSC systems) E. Optional Autosampler (Nano DSC only) F. Gas Tank Revision A Issued May 2017 Page 4

Instrument Measurements NANO DSC Height: 28 cm (11 in) Width: 35 cm (14 in) Depth: 53 cm (21 in) Weight: 17 kg (37 lbs) MCDSC Height: 31 cm (12 in) Width: 35 cm (14 in) Depth: 53 cm (21 in) Weight: 21 kg (46 lbs) Revision A Issued May 2017 Page 5

Utility Requirements POWER Item Nano DSC MCDSC Power line Electrical power cord Customer-supplied Requirement 100 240 VAC, 3A, 50/60 Hz 100 240 VAC, 4A, 50/60 Hz Grounded, single-phase line for instrument and computer, not shared with motors, heaters, or compressors 15 A for voltages near 120 VAC 10 A for voltages near 230 VAC The plug of the cord must be rated to carry at least 125% of the product current rating. The cord length must be less than 4.5 meters and must be UL or CSA approved. Surge suppressor power strip GAS Instrument Requirement Nano DSC Autosampler Filtered, compressed nitrogen, regulated to 45 psi (3 bar), with 1/16-inch outlet line diameter. Customer-supplied. Note: Adapters for 1/4- and 1/8 - inch lines are included with the instrument. 2 to 5 customer-supplied bottles for rinsing/cleaning solutions (> 1 L each) MCDSC 2 customer-supplied waste collection bottles (> 2 L each) A dry purge gas source (ex. air, nitrogen, argon) at low pressure (5 10 psig) is required for operation of the instrument with sample temperatures below the dew point conditions in the laboratory. WATER Instrument Requirement MCDSC Requires an external water circulator Revision A Issued May 2017 Page 6

Computer Requirements HARDWARE REQUIREMENTS Description Processor Memory Hard drive DVD Screen resolution Graphic card Screen (LCD) size USB Ports (1 for instrument and 1 for Autosampler) Requirement Intel Core 2 Duo or better 2.93 GHz with 3 MB L2 cache Required: 4 GB RAM Recommended: 8 GB RAM 300 GB free space 16x CD-ROM optional. Obtain the latest software updates at http://www.tainstruments.com/support/software-downloadssupport/downloads Minimum: 1280 x 1024 with 24-bit colors Recommended: 1920 x 1080 with 24-bit colors 128 MB DirectX10 or higher recommended Recommended: 24 wide screen 1.1 or 2.0 Item Operating System Browser Other Other Network Conflicts SOFTWARE REQUIREMENTS Requirement Windows 7, 8, 10 Ultimate, Enterprise & Professional Home version not supported Required: 32-bit or 64-bit version Recommended: 64-bit version Internet Explorer Microsoft Operating System Service Pack Windows Operating System and associated Microsoft updates must be upto-date TA Instruments is not responsible for resolving issues associated with connections to your corporate network. TA Instruments is not responsible for resolving hardware/software conflicts created by the addition of third party hardware or software to the computer. Revision A Issued May 2017 Page 7

Accessories WATER CIRCULATOR FOR MCDSC Height: 68 cm (27 in) Width: 37 cm (14.5 in) Depth: 57 cm (22 in) Weight: 55 kg (120 lbs) Requirements Must be placed 20 cm (8 in) from the rear wall. Circulator Power Cooling Gas LN 2 Fluid Light Hardware Software Temp Lab Customer Revision A Issued May 2017 Page 8

Accessories NANO DSC AUTOSAMPLER Height: 35 cm (14 in) Width: 30 cm (12 in) Depth: 56 cm (22 in) Weight: 21 kg (46 lbs) Revision A Issued May 2017 Page 9

Site Preparation Checklist Nano DSC and MCDSC if applicable if applicable Sufficient bench space for instrument, computer, sample prep space and tools, and Degassing System or Autosampler (if needed): Depth: 76 cm (30 in) minimum Distance from the wall: 20 cm (8 in) minimum Nano DSC: Length: 2.8 m (9 ft) for instrument, computer, sample prep and tools, and Degassing Station Nano DSC with Autosampler: Length: 2.8 m (9 ft) for instrument, Autosampler, computer, and sample prep and tools Additional floor space beneath the Autosampler for 4 L collection bottle MCDSC: 2.4 m (8 ft) for instrument, computer, and sample prep space and tools Instrument power: Nano DSC: 100 240 VAC, 3A, 50/60 Hz MCDSC: 100 240 VAC, 4A, 50/60 Hz Water Circulator: 120 VAC, 60 Hz, 13A or 240 VAC 50 Hz, 13A Nano DSC Autosampler: Filtered, compressed nitrogen regulated to 45 psig (3 bar) with 1/16-inch outlet line (or 1/4- or 1/8-inch to be used with included adapter) diameter. Customer-supplied. 2 5 customer-supplied bottles for rinsing/cleaning solutions (>1 L each) 2 customer-supplied waste collection bottles (> 2 L each) MCDSC: A dry purge gas source (ex. Air, nitrogen, argon) at 5 10 psig MCDSC: External water circulator I hereby acknowledge that all utility requirements have been met per the checklist above and that they will be ready at the agreed time of installation. If all utility requirements are not met at the agreed time of installation, additional charges may be incurred for a return Service trip. Customer DD MM YYYY Company City State Country Please send a signed copy of the completed checklist to your local Service representative. Revision A Issued May 2017 Page 10

TA Instruments Offices For information on our latest products, contact information, and more, see our website at: http://www.tainstruments.com. To find your local TA Instruments office and contact information, visit http://www.tainstruments.com/contact/ta-directory/ TA Instruments Waters LLC Corporate Headquarters 159 Lukens Drive New Castle, DE 19720 USA Telephone: 302-427-4000 Fax: 302-427-4001 Email: info@tainstruments.com Revision A Issued May 2017 Page 11