Contents. Software. CALIBRATION PROCEDURE NI Channel, ±5 V, 24-Bit Software-Selectable IEPE and AC/DC Analog Input Module

Similar documents
Contents. Software. CALIBRATION PROCEDURE NI Channel, 300 Vrms, 24-Bit, Simultaneous, Channel-to-Channel Isolated Analog Input Module

Contents. Software. CALIBRATION PROCEDURE NI 9216/ Channel, 24-Bit, 100 Ω/1000 Ω RTD Analog Input Module

CALIBRATION PROCEDURE NI channel, ±2- ma/±10 V, 24-bit Analog Input Module. ni.com/manuals

±10 V, 20 MS/s/ch, 14-Bit, 4-Channel C Series Digitizer Module

CALIBRATION PROCEDURE PXIe channel, 16-bit, 400 ks/s/ch, Ch-Ch Isolated Analog Input Module. ni.com/manuals

NI PCIe-784xR, NI PXI-784xR, NI PXIe-784xR, NI USB-784xR, NI PCIe-785xR, NI PXI-785xR, NI PXIe-785xR, NI USB-785xR, and NI PXIe-786xR

NI R Series. Contents. Software CALIBRATION PROCEDURE

CALIBRATION PROCEDURE NI PXIe Channel, Isolated Voltage/Current Analog Output Module

Contents. Software Requirements. CALIBRATION PROCEDURE NI PXIe-4357

CALIBRATION PROCEDURE CAL Contents. Software

GETTING STARTED GUIDE NI Channel, 50 Arms Continuous, 24-Bit, Analog Input Module

SCXI -1102/B/C. Contents CALIBRATION PROCEDURE. For NI-DAQmx

SCXI -1104/C. Contents CALIBRATION PROCEDURE. For NI-DAQmx

CALIBRATION PROCEDURE SCXI Contents. Software Requirements. ni.com/manuals

12-Channel, ±24 V Precision PXI Source Measure Unit

Contents. Software CALIBRATION PROCEDURE NI TC-4353

24-Channel, ±24 V Precision PXI Source Measure Unit

Contents. CALIBRATION PROCEDURE NI PXIe-5624R 2 GS/s, 12-Bit IF Digitizer

GETTING STARTED GUIDE NI AO, 0 ma to 20 ma, 16 Bit Simultaneous

GETTING STARTED GUIDE NI Channel Solid-State Relay (SSR) Digital Output Module

GETTING STARTED GUIDE NI V, 32-Channel (Sinking Input), 7 µs C Series Digital Module

NI USRP-2900/2901. Contents. Electromagnetic Compatibility Guidelines GETTING STARTED GUIDE. Universal Software Radio Peripheral

R Series Digital I/O Module for PCI Express, 128 DIO, 512 MB DRAM, Kintex-7 160T FPGA. This document describes how to begin using the PCIe-7821R.

Electromagnetic Compatibility Guidelines

R Series Reconfigurable I/O Module (AI, AO, DIO) for PXI Express, 6 AI, 18 AO, 48 DIO, 1 MS/s AIO, 512 MB DRAM, Kintex-7 160T FPGA

OPERATING INSTRUCTIONS AND SPECIFICATIONS NI 9476E. 32-Channel, 24 V Sourcing Digital Output Module

Reconfigurable Embedded Chassis with Integrated MXI-Express (x1) MXI-Express Switch

Contents. Software Requirements CALIBRATION PROCEDURE NI TB-4353

To calibrate the NI 5105, install the following versions of NI-SCOPE. NI-SCOPE Version

GETTING STARTED GUIDE NI Channel, 60 V, High-Speed, Sourcing Digital Output Module

NI ELVIS RIO Control Module

R Series for USB Multifunction RIO with Kintex-7 70T FPGA. ni.com/manuals

GETTING STARTED GUIDE NI AI, 100 S/s/ch Simultaneous, Universal Measurements

R Series for USB Multifunction RIO with Kintex-7 160T FPGA. ni.com/manuals

Bracket and Mounting Dimensions

NI PXIe-4138/4139. Contents. Electromagnetic Compatibility Guidelines GETTING STARTED GUIDE. Single-Channel System SMU

R Series for USB Multifunction RIO with Kintex-7 70T FPGA. ni.com/manuals. Electromagnetic Compatibility Guidelines

IVN Safety Guidelines GETTING STARTED GUIDE. 100BASE-T1 Media Converter

GETTING STARTED GUIDE NI S/s/channel, 4-Channel C Series Universal Analog Input Module

NI CMS Contents MAINTAINING HARDWARE. NI Condition Monitoring System for NI InsightCM Enterprise

SAFETY, ENVIRONMENTAL, AND REGULATORY INFORMATION. NI CompactDAQ Four-Slot Controller with Quad-Core Processor

R Series for PXI Express Digital RIO with Kintex-7 325T FPGA. ni.com/manuals. This document explains how to install and configure the NI PXIe-7822R.

NI PXI Contents GETTING STARTED GUIDE. 8 GHz RF Amplifier

NI 9213 Calibration Procedure

Electrical. SPECIFICATIONS SH Cable

NI sbrio Safety Guidelines GETTING STARTED GUIDE. Single-Board RIO OEM Devices. This document describes how to begin using the NI sbrio-9627.

NI CMS Contents MAINTAINING HARDWARE. NI Condition Monitoring System for NI InsightCM Enterprise

NI PXIe Contents GETTING STARTED GUIDE. RF Preselector Module

NOTE TO USERS NI 61xx Pinout Labels for the SCB-68 Multifunction I/O S Series Modules/Devices Using the 68-Pin Shielded Connector Block

USER GUIDE USB-634x/635x/636x OEM X Series USB-6341/6343/6356/6361/6363/6366 OEM Devices

NI USB Contents. Electromagnetic Compatibility Guidelines GETTING STARTED GUIDE. 10 MHz to 18 GHz RF Power Meter

Related Information Refer to the NI Switches Help for detailed information about features and programming options for your NI switch product.

To access NI 5622 documentation, navigate to Start»All Programs»National Instruments» NI-SCOPE»Documentation.

R Series for USB Multifunction RIO with Kintex-7 160T FPGA. ni.com/manuals. Electromagnetic Compatibility Guidelines

SH Cable Specifications

SPECIFICATIONS AND FEATURES GUIDE NI Serial Legacy Hardware

INSTALLATION INSTRUCTIONS Data Management Software Suite in the Evaluation Mode

NI PXIe-5160/5162. Contents. Electromagnetic Compatibility Guidelines GETTING STARTED GUIDE. 10-Bit Oscilloscope

NI PXIe-4080/4081/4082

Foundation Fieldbus Hardware and NI-FBUS Software

Thermal Kit for NI sbrio-9607/9627/9637

NI CMS-9068 and NI MMS-9068

PXIe Contents. Verifying the System Requirements GETTING STARTED GUIDE. 32-Channel Digital Pattern Instrument

NI SMD-7610 Stepper Drives and NI 9512 C Series Modules

Introduction... 1 System Diagram... 2 Hardware... 2 Powering the Hardware... 3 Wiring Harness... 5

NI CVS Required Components. Optional Equipment GETTING STARTED GUIDE. Compact Vision System with USB3 Vision and Reconfigurable I/O

Electromagnetic Compatibility Guidelines

GETTING STARTED GUIDE

NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

Calibrating FieldPoint Analog Input Modules with Calibration Executive

GETTING STARTED GUIDE NI Ch 24 V Sinking DI, 8-Ch 24 V Sourcing DO SIL3 Capable

Single-Channel Precision System Source Measure Unit (SMU)

GETTING STARTED NI 9501 C Series Stepper Drive Modules and NI SoftMotion Module

Embedded Direct Injector Control and Measurement (DCM) Device with Real-Time Processor and Reconfigurable FPGA

To access NI 5105 documentation, navigate to Start»All Programs»National Instruments» NI-SCOPE»Documentation.

SCXI Contents CALIBRATION PROCEDURE

PXI/PCI Contents GETTING STARTED GUIDE. 2 GS/s Oscilloscope

NI WSN-9791 Ethernet Gateway

PXIe Contents GETTING STARTED GUIDE. 32-Channel Digital Pattern Instrument

To access NI 5114 documentation, navigate to Start»All Programs»National Instruments» NI-SCOPE»Documentation.

NI PXI/PXIe/PCI Contents GETTING STARTED GUIDE. 14-Bit, 100 MS/s PXI Oscilloscope

NI REM Remote I/O Overview DATASHEET. Bus Coupler for Remote I/O

NI 6703/6704 Calibration Procedure for NI-DAQ mx

NI PXI-5650/5651/5652

NI REM Remote I/O Overview DATASHEET. Analog Output Module for Remote I/O

NI REM Remote I/O Overview DATASHEET. Digital Input Module for Remote I/O

The following conventions appear in this document:

NOTE TO USERS NI 65xx Pinout Labels for the SCB-100 Digital I/O Modules/Devices Using the 100-Pin Shielded Connector Block

NI PXI Contents CALIBRATION PROCEDURE

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9208E 16-Channel, ±20 ma, 24-Bit Analog Input Module

INSTALLATION INSTRUCTIONS Data Management Software Suite

NI Safety Guidelines. Safety Guidelines for Hazardous Locations GETTING STARTED GUIDE. Ethernet Expansion Chassis for C Series Modules

NI crio-9101/9102/9103/9104

GETTING STARTED NI SMD-7611/7612 Stepper Drives and NI 9512 C Series Modules

NI PXIe-6555/6556. Contents GETTING STARTED GUIDE. 200 MHz Digital Waveform Generator/Analyzer with PPMU

INSTALLATION GUIDE AND TERMINAL BLOCK SPECIFICATIONS NI PXIe-4322 and TB Channel, Isolated Voltage/Current Analog Output Module

USER MANUAL NI 9755 NI Powertrain Controls CompactRIO NOx Sensor Module Kit

LabVIEW DataFinder Toolkit

INSTALLATION INSTRUCTIONS SH160F-160M-NI SwitchBlock Cable

NI REM Remote I/O Overview DATASHEET. Digital Output Module for Remote I/O

Transcription:

CALIBRATION PROCEDURE NI 9234 4-Channel, ±5 V, 24-Bit Software-Selectable IEPE and AC/DC Analog Input Module This document contains the verification and adjustment procedures for the National Instruments 9234. For more information on calibration, visit ni.com/calibration. Contents Software... 1 Documentation... 2 Test Equipment... 3 Test Conditions... 4 Initial Setup... 4 Verification... 5 Accuracy Verification... 5 Gain Matching Verification... 6 Phase Matching Verification... 8 Common-Mode Rejection Ratio Verification... 9 IEPE Current Verification... 11 IEPE Compliance Voltage Verification... 12 Adjustment... 14 Accuracy Adjustment... 14 EEPROM Update... 14 Reverification... 15 Accuracy Under Calibration Conditions... 15 Worldwide Support and Services... 16 Software Calibrating the NI 9234 requires the installation of NI-DAQmx 9.2 or later on the calibration system. You can download NI-DAQmx from ni.com/downloads. NI-DAQmx supports LabVIEW, LabWindows TM /CVI, C/C, C#, and Visual Basic.NET. When you install NI-DAQmx you only need to install support for the ADE that you intend to use.

Documentation Consult the following documents for information about the NI 9234, NI-DAQmx, and your application software. All documents are available on ni.com and help files install with the software. NI cdaq-9174/9178 USB Chassis Quick Start NI-DAQmx installation and hardware setup NI 9234 Operating Instructions and Specifications NI 9234 specific information, specifications, and calibration interval NI-DAQmx Readme Operating system and application software support in NI-DAQmx LabVIEW Help LabVIEW programming concepts and reference information about NI-DAQmx VIs and functions NI-DAQmx C Reference Help Reference information for NI-DAQmx C functions and NI-DAQmx C properties NI-DAQmx.NET Help Support for Visual Studio Reference information for NI-DAQmx.NET methods and NI-DAQmx.NET properties, key concepts, and a C enum to.net enum mapping table 2 ni.com NI 9234 Calibration Procedure

Test Equipment Table 1 lists the equipment recommended for the performance verification and adjustment procedures. If the recommended equipment is not available, select a substitute using the requirements listed in Table 1. Table 1. Recommended Equipment Equipment Recommended Model Where Used Requirements Calibrator Fluke 5700A Accuracy, Adjustment Use a high-precision voltage source with an accuracy 100 ppm and an output that is able to be not grounded (floating). Function Generator NI 4461 Gain Matching Use a function generator capable of sourcing 1 khz and 10 khz ±4.5 V pk AC sinusoidal signals with amplitude uncertainty of ±10% or less. Phase Matching Use a function generator capable of sourcing both 1 khz and 10 khz ±4.5 V pk AC sinusoidal signals with amplitude uncertainty of ±10% or less. CMRR Use a function generator capable of sourcing 1 khz ±4.5 V pk AC sinusoidal signals with amplitude uncertainty of ±0.04dB or less. DMM NI 4070 IEPE Current Use DMM in current mode in the smallest range to measure 2 ma DC with a measurement uncertainty 0.5%. Source Measure Unit (SMU) Keithley Model 2400 IEPE Compliance Voltage Use an SMU capable of sourcing 20 V DC with an accuracy 0.1% while sinking >2 ma DC. BNC Cable All Use a length appropriate for your application. Chassis NI cdaq-9178 All NI 9234 Calibration Procedure National Instruments 3

Test Conditions The following setup and environmental conditions are required to ensure the NI 9234 meets published specifications. Keep connections to the device as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements. Verify that all connections to the device are secure. Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire to eliminate noise and thermal offsets. Maintain an ambient temperature of 23 ±5 C. The device temperature will be greater than the ambient temperature. Keep relative humidity below 80%. Allow a warm-up time of at least 10 minutes to ensure that the NI 9234 measurement circuitry is at a stable operating temperature. Initial Setup Complete the following steps to set up the NI 9234. 1. Install NI-DAQmx. 2. Make sure the NI cdaq-9178 power source is not connected. 3. Connect the NI cdaq-9178 to the system safety ground. a. Attach a ring lug to a 14 AWG (1.6 mm) wire. b. Connect the ring lug to the ground terminal on the side of the NI cdaq-9178 using the ground screw. c. Attach the other end of the wire to the system safety ground. 4. Install the module in slot 8 of the NI cdaq-9178 chassis. Leave slots 1 through 7 of the NI cdaq-9178 chassis empty. 5. Connect the NI cdaq-9178 chassis to your host computer. 6. Connect the power source to the NI cdaq-9178 chassis. 7. Launch Measurement & Automation Explorer (MAX). 8. Right-click the device name and select Self-Test to ensure that the module is working properly. 4 ni.com NI 9234 Calibration Procedure

Verification The following performance verification procedures describe the sequence of operation and test points required to verify the NI 9234 and assumes that adequate traceable uncertainties are available for the calibration references. Accuracy Verification Complete the following procedure to determine the As-Found status of the NI 9234. 1. Set the calibrator to Standby mode (STBY). 2. Connect the NI 9234 to the calibrator as shown in Figure 1. Figure 1. Accuracy Verification Connections to the NI 9234 HI LO AIx Calibrator NI 9234 3. Verify that the calibrator output is not grounded (floating). Notes The analog inputs on the NI 9234 are grounded, so the calibrator must be not grounded (floating) to prevent ground loop measurement errors. To float the output of the Fluke 5700A, enable External Guard mode (EX GRD). 4. Set the calibrator voltage to a Test Point value indicated in Table 4. 5. Set the calibrator to Operate mode (OPR). 6. Acquire and average samples. a. Create an AI voltage task on the NI 9234 according to Table 2. Table 2. NI 9234 Configuration for Accuracy Verification Min (V) Max (V) Scaled Units Channel Configuration -5 5 Volts DC Coupled b. Configure the AI voltage task timing according to Table 3. Table 3. NI 9234 Timing Configuration Sample Mode Samples Per Channel Rate (khz) Finite Samples 10240 10.24 c. Start the task. NI 9234 Calibration Procedure National Instruments 5

d. Average the readings and record the average. e. Compare the average to the limits in Table 4. f. Clear the task. 7. Set the calibrator to Standby mode (STBY). 8. Repeat steps 3 through 7 for each test point in Table 4. Table 4. NI 9234 Verification Test Limits for Accuracy Test Point 1-Year Limits Location Value (V) Lower Limit (V) Upper Limit (V) Max 4.0 3.9952 4.0048 Mid 0.0-0.0012 0.0012 Min -4.0-4.0048-3.9952 Note The test limits listed in Table 4 are derived using the values in Table 21. 9. Disconnect the calibrator from the device. 10. Repeat steps 1 through 9 for each channel on the NI 9234. Gain Matching Verification Complete the following procedure to determine the As-Found status of the NI 9234. 1. Set the calibrator to Standby mode (STBY). 2. Connect the NI 9234 to the calibrator. Refer to Figure 2 for a connection diagram. Figure 2. Gain Matching Verification Connections to the NI 9234 HI LO AIx Calibrator NI 9234 3. Verify that the calibrator output is not grounded (floating). Notes The analog inputs on the NI 9234 are grounded, so the calibrator must be not grounded (floating) to prevent ground loop measurement errors. To float the output of the Fluke 5700A, enable External Guard mode (EX GRD). 4. Set the calibrator voltage to 4.0 V. 5. Set the calibrator to Operate mode (OPR). 6 ni.com NI 9234 Calibration Procedure

6. Acquire and average samples. a. Create an AI voltage task on the NI 9234 according to Table 5. Table 5. NI 9234 Configuration for Gain Matching Verification Min (V) Max (V) Scaled Units Channel Configuration -5 5 Volts DC Coupled b. Configure the AI voltage task timing according to Table 6. Table 6. Timing Configuration Sample Mode Samples Per Channel Rate (khz) Finite Samples 10240 10.24 c. Start the task. d. Average the readings and record the average as V 1. e. Clear the task. f. Set the calibrator to Standby mode (STBY). g. Set the calibrator voltage to -4.0 V. h. Set the calibrator to Operate mode (OPR). i. Start the task. j. Average the readings and record the average as V 2. k. Clear the task. 7. Set the calibrator to Standby mode (STBY). 8. Disconnect the calibrator from the device. 9. Perform the following calculation using the recorded V 1 and V 2 values. V 1 V 2 Gain( db) = 20 log 10 ----------------- 8 10. Repeat steps 1 through 9 for each channel. 11. Compare the channel-to-channel gain difference to the maximum limit of ±40 mdb. If the Gain Matching verification procedure determines that the NI 9234 is outside of the limits, refer to Worldwide Support and Services for assistance in returning the device to NI. NI 9234 Calibration Procedure National Instruments 7

Phase Matching Verification Complete the following procedure to determine the As-Found status of the NI 9234. 1. Connect the NI 9234 to the function generator as shown in Figure 3. Figure 3. Phase Matching Verification Connections to the NI 9234 HI AI0 Function Generator LO AI1 AI2 AI3 NI 9234 2. Configure function generator according to Table 7. Table 7. Function Generator Configuration for Phase Matching Verification Signal Type Amplitude (V pk ) Sinewave ±4.5 3. Set the function generator frequency to a value indicated in Table 10. 4. Acquire and average samples. a. Create an AI voltage task on the NI 9234 according to Table 8. Table 8. NI 9234 Configuration for Phase Matching Verification Min (V) Max (V) Scaled Units Channel Configuration -5 5 Volts AC Coupled Note Wait 20 s for the AC coupling filter on the NI 9234 to settle. b. Configure the AI voltage task timing according to Table 9. Table 9. NI 9234 Timing Configuration Sample Mode Samples Per Channel Rate (khz) Finite Samples 16384 51.2 c. Enable the function generator. d. Start the task. 8 ni.com NI 9234 Calibration Procedure

e. Disable the function generator. f. Clear the task. 5. Calculate the phase for each channel on the NI 9234. NI recommends using the Extract Single Tone Information VI to calculate phase. 6. Compare the channel-to-channel phase difference to the limits in Table 10. 7. Repeat steps 2 through 6 for each test point in Table 10. Table 10. NI 9234 Verification Test Limits for Phase Matching Generator Frequency (khz) Phase Matching Channel-to-Channel Limit 1 ±0.085 10 ±0.49 If the Phase Matching verification procedure determines that the NI 9234 is outside of the limits, refer to Worldwide Support and Services for assistance in returning the device to NI. Common-Mode Rejection Ratio Verification Complete the following procedure to determine the As-Found status of the NI 9234. 1. Connect the NI 9234 to the function generator and short the positive pin and negative shell of the NI 9234 channel. 2. Connect the function generator to the chassis ground as shown in Figure 4. Figure 4. CMRR Verification Connections to the NI 9234 HI AIx LO Function Generator NI 9234 3. Configure the function generator according to Table 11. Table 11. Function Generator Configuration for CMRR Verification Signal Type Amplitude (V pk ) Frequency (khz) Sinewave ±0.5 1 NI 9234 Calibration Procedure National Instruments 9

4. Acquire a reference amplitude. a. Create and configure an AI voltage task on the NI 9234 according to Table 12. Table 12. NI 9234 Configuration for CMRR Verification Min (V) Max (V) Scaled Units Channel Configuration -5 5 Volts DC Coupled b. Configure the AI voltage task timing according to Table 13. Table 13. NI 9234 Timing Configuration Sample Mode Samples Per Channel Rate (khz) Finite Samples 16384 51.2 c. Enable the function generator. d. Start the task. e. Record the measured amplitude as Amplitude Acq. f. Disable the function generator. g. Clear the task. 5. Perform the following calculation using the recorded Amplitude Acq. Amplitude Acq CMRR = 20 log 10 ---------------------------------- 0.5 h. Compare the calculation result to the minimum CMRR limit of 40 db. 6. Repeat steps 1 through h for each channel. If the Common-Mode Rejection Ratio verification procedure determines that the NI 9234 is outside of the limits, refer to Worldwide Support and Services for assistance in returning the device to NI. 10 ni.com NI 9234 Calibration Procedure

IEPE Current Verification Complete the following procedure to determine the As-Found status of the NI 9234. 1. Connect the NI 9234 to the DMM as shown in Figure 5. Figure 5. IEPE Current Verification Connections to the NI 9234 HI LO AIx DMM NI 9234 2. Configure the DMM for a DC current measurement in the lowest appropriate range to measure 2 ma. 3. Measure IEPE Source short circuit current. a. Create an AI voltage task on the NI 9234 according to Table 14. Table 14. NI 9234 Configuration for IEPE Current Verification Min (V) Max (V) Scaled Units Channel Configuration IEPE Excitation -5 5 Volts AC Coupled Enabled b. Configure the AI voltage task timing according to Table 15. Table 15. NI 9234 Timing Configuration Sample Mode Rate (khz) Continuous Samples 51.2 c. Start the task. d. Record the DMM measurement. e. Clear the task. 4. Compare the DMM measurement to the minimum current limit of 2.0 ma. 5. Repeat steps 1 through 4 for each channel. If the IEPE Current verification procedure determines that the NI 9234 is outside of the limits, refer to Worldwide Support and Services for assistance in returning the device to NI. NI 9234 Calibration Procedure National Instruments 11

IEPE Compliance Voltage Verification Complete the following procedure to determine the As-Found status of the NI 9234. 1. Connect the NI 9234 to the DMM as shown in Figure 6. Figure 6. IEPE Compliance Voltage Verification Connections to the NI 9234 (IEPE Source Short Circuit Current Measurement) HI LO AIx DMM NI 9234 2. Configure the DMM for a DC current measurement in the lowest appropriate range to measure 2 ma. 3. Measure IEPE Source short circuit current. a. Create an AI voltage task on the NI 9234 according to Table 16. Table 16. NI 9234 Configuration for IEPE Voltage Verification Channel Configuration AC Coupled with IEPE IEPE Excitation Enabled b. Configure the AI voltage task timing according to Table 17. Table 17. NI 9234 Timing Configuration Sample Mode Rate (khz) Continuous Samples 51.2 c. Start the task. d. Record the DMM measurement DMM Current. e. Perform the following calculation using the recorded DMM Current value. SMU SourceCurrent = -(0.95 DMM Current ) 4. Record the calculation result as SMU SourceCurrent. 12 ni.com NI 9234 Calibration Procedure

5. Connect the SMU to the channel you want to verify on the NI 9234. Refer to Figure 7 for a connection diagram. Figure 7. IEPE Compliance Voltage Verification Connections to the NI 9234 (IEPE Source Compliance Voltage Measurement) HI LO AIx SMU NI 9234 6. Configure the SMU as a current source and set the voltage limit to the lowest appropriate range to measure 24 V. 7. Measure IEPE Source compliance voltage. a. Create and configure an AI voltage task on the NI 9234 according to Table 18. Table 18. IEPE Compliance Verification Equipment Configuration Channel Configuration AC Coupled with IEPE IEPE Excitation Enabled b. Configure the AI voltage task timing according to Table 19. Table 19. NI 9234 Timing Configuration Sample Mode Rate (khz) Continuous Samples 51.2 c. Set the source current on the SMU to the recorded SMU SourceCurrent value. d. Start the task. e. Enable the SMU. f. Record the SMU measurement SMU Voltage. g. Disable the SMU. h. Clear the task. 8. Compare the voltage measurement from the SMU to the minimum voltage limit of 19 V. 9. Repeat steps 1 through 8 for each channel. If the IEPE Compliance Voltage verification procedure determines that the NI 9234 is outside of the limits, refer to Worldwide Support and Services for assistance in returning the device to NI. NI 9234 Calibration Procedure National Instruments 13

Adjustment The following performance adjustment procedure describes the sequence of operations required to adjust the NI 9234. Accuracy Adjustment Complete the following procedure to adjust the accuracy performance of the NI 9234. 1. Set the calibrator to Standby mode (STBY). 2. Connect the NI 9234 to the calibrator as shown in Figure 1. 3. Adjust the NI 9234. a. Initialize a calibration session on the NI 9234. The default password is NI. b. Input the external temperature in degrees Celsius. c. Call the NI 9234 get C Series adjustment points function to obtain an array of recommended calibration voltages for the NI 9234. d. Set the calibrator to a reference value determined by the array of recommended calibration voltages. e. Set the calibrator to Operate mode (OPR). f. Call and configure the NI 9234 gain instance of the DAQmx adjustment function according to Table 20. Table 20. Adjustment Configuration Physical Channel cdaqmod8/aix Reference Value The reference value from the array of adjustment points g. Set the calibrator to Standby mode (STBY). h. Repeat steps d through g for each calibration voltage in the array. i. Disconnect the NI 9234 from the calibrator. j. Short AI and AI- on the BNC connector together for the channel under adjustment on the NI 9234. k. Take a measurement using the NI 9234 offset instance of the DACQmx adjust function for the channel under adjustment. l. Close the calibration session. 4. Repeat steps 1 through 3 for each channel on the NI 9234. EEPROM Update When an adjustment procedure is completed, the NI 9234 internal calibration memory (EEPROM) is immediately updated. If you do not want to perform an adjustment, you can update the calibration date and onboard calibration temperature without making any adjustments by initializing an external calibration, setting the C Series calibration temperature, and closing the external calibration. 14 ni.com NI 9234 Calibration Procedure

Reverification Repeat the Verification section to determine the As-Left status of the device. Note If any test fails Reverification after performing an adjustment, verify that you have met the Test Conditions before returning your device to NI. Refer to Worldwide Support and Services for assistance in returning the device to NI. Accuracy Under Calibration Conditions The following calibration specifications are valid under the following conditions: Ambient temperature 23 ±5 C NI 9234 installed in slot 8 of an NI cdaq-9178 chassis Slots 1 through 7 of the NI cdaq-9178 chassis are empty Note The test limits listed in Table 4 are derived using the values in Table 21. Table 21. NI 9234 Accuracy Under Calibration Conditions Percent of Reading (Gain Error) Percent of Range * (Offset Error) * Range equals 5.1 V ±0.09% ±0.024% (±1.2 mv) Note For operational specifications, refer to the most recent NI 9234 Operating Instructions and Specifications online at ni.com/manuals. NI 9234 Calibration Procedure National Instruments 15

Worldwide Support and Services The National Instruments website is your complete resource for technical support. At ni.com/ support you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers. Visit ni.com/services for NI Factory Installation Services, repairs, extended warranty, and other services. Visit ni.com/register to register your National Instruments product. Product registration facilitates technical support and ensures that you receive important information updates from NI. A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European Communities using the manufacturer s declaration of conformity. This system affords the user protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC for your product by visiting ni.com/certification. If your product supports calibration, you can obtain the calibration certificate for your product at ni.com/calibration. National Instruments corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. National Instruments also has offices located around the world. For telephone support in the United States, create your service request at ni.com/support or dial 1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the Worldwide Offices section of ni.com/niglobal to access the branch office websites, which provide up-to-date contact information, support phone numbers, email addresses, and current events. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patents Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14s, DFAR 252.227-7014, and DFAR 252.227-7015. 20092014 National Instruments. All rights reserved. 372678G-01 Dec14