For technical support please contact

Similar documents
DaqLink Calibration Procedure

Multi-functional precision thermometer CTR3000

DaqPRO Solution. User Guide INNOVATIVE MONITORING SOLUTIONS ALL IN ONE SOLUTION FOR DATA LOGGING AND ANALYSIS.

TITAN S8. Portable Data Acquisition Logger. Product User Guide

2 Product Overview. 5 User Interface. 14 Logging Data. 17 Viewing Data. 19 Managing Data. 24 Device Settings. 26 Specifications.

Product User Guide. QuadTemp and OctTemp. QuadTemp 4-Channel Thermocouple Based Temperature Data Logger

MC1000 Multi-Function Calibrator

DSTCAL Software User Manual 09/16/09 WD1037 Rev E Revised 09/09/14

L200 User Manual. Operating Instructions for L200-TC & L200-PT Temperature Monitor. RoHS compliant

CONTROL MICROSYSTEMS Thermocouple Analog Input Module. Hardware Manual

DataNet Solution. Wireless intelligent logging network. Sense it. Manage it. H E

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB

IST ULTRASTAB Power Supply for high precision transducers

Analogue Input Module. user manual All Models

Specification for temperature sensor calibration setup

Model 3750 Multifunction I/O Card

IST ULTRASTAB Power Supply for IT/ITN ULTRASTAB transducers

96x48. gal/h. Digital panel meter 4½-digit DV4. Without setpoints. Sensor supply

Thermo Electric Cooling Temperature Controller TEC Controller / Peltier Driver ±10 A / up to ±21 V

Hand-held thermometer Model CTH7000

Thermo Electric Cooling Temperature Controller TEC Controller / Peltier Driver ±4 A / up to ±21 V. Miniature OEM TEC Controller

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

WORLD LEADING PRODUCTS FOR LASER SCIENTISTS AND ENGINEERS

Isolated Wideband Voltage Input 3B40 / 3B41 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Bluetooth/USB Data Logger USER S MANUAL. Hantek 365A/B/C/D/E/F V

5 Performance Verification

AMETEK JOFRA CSC200. Temperature Calibrator. Reference Manual

Cube67 - Modular I/O station

DS5000-IBFS. 5kA rms - 8kA DC Fluxgate current transducer. Features. Description. Application. Precision Innovation /

BC-7000 Field Calibration Procedure 55 AMP

Wiring guide for dual-output vibration and temperature sensors

GSV-1A4 M12/2 M12/2. Highlights

Ultra-Accurate Temperature Measurement Instruments

Portable Multi-Function Calibrator Model CEP6000

CALYS 50. Field multifunction calibrator for basic use. Calibration measurement

Isolated Voltage Input 3B30 / 3B31 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

C100 Portable Multifunction

Wide Bandwidth Strain Gage Input 3B18 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Thermo Electric Cooling Temperature Controller Dual TEC Controller / Peltier Driver 2x (±16 A / up to ±31 V) OEM Dual-Channel TEC Controller

UNIVERSAL TEMPERATURE TRANSMITTER

Thermo Electric Cooling Temperature Controller TEC Controller / Peltier Driver ±10 A / up to ±21 V

DC-204R DC-204Ra SMART DYNAMIC STRAIN RECORDER. 200kHz 2GB UPS. Strain DC voltage Temperature. TML Pam E-442B. capacity. High speed sampling

AiM User Manual. Channel Expansion connection and configuration for AiM loggers. Release 1.01

OBSOLETE. Isolated, Linearized, Thermocouple Input 3B47 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Hand-held thermometer Model CTH7000

-auto Power Off- T1/T2/T1-T2

M3S Series Super-space-saving Signal Conditioners. PC CONFIGURATOR SOFTWARE Model: M3SCFG. Users Manual

Installation Instructions for: Channel Thermocouple Amplifier

WIRING FOR MXL PRO- 05

Industrial. Laboratory and Field. Accessories. Product Selection Guide. Analyzers 1 Controllers 2 Transmitters 3 Monitors 4

Model 3724 Dual 1x30 FET Multiplexer Card

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB

SEM310 SEM310X HART UNIVERSAL TEMPERATURE TRANSMITTER

Model R5005. Instruction Manual. True RMS Industrial Multimeter. reedinstruments. www. com

! Ranges from -45º to 650ºC. ! Rapid heating, cooling and settling. ! Switch test, ramp, step and preset functions

5504 Thermocouple Analog Input Module

IP 20. Rated. with terminals

Analog signal to RS485 or RS232 Converter, A/D Converter

Engineering Note: EN0047 Commissioning a Hydro-Probe with a Command Alkon EZCal Station

BC-9000 Field Calibration Procedure CF1_FIELDCAL_BC9000 Revised 03/16/2013

Page 2 TECPak User s Manual

R1M-GH THERMOCOUPLE & DC INPUT MODULE MODEL. Remote I/O R1M Series. (16 points)

AS-i Safety Analog Input Module, IP20

RSTI-EP Slice I/O. Analog Input Modules `EP-3164, EP-3264, EP-3124, EP-3368, EP-3468 EP-3704, EP GFK-2960B November 2017

UNIVERSAL CURRENT CONVERTER

User s Guide DP20. Economical Multisignal Meter for Process, Temperatures and Electrical signals. Shop online at omega.com

Autoranging True RMS Multimeter User Manual

Jandel RM3 operating instructions

DM650TM SENSOR / WALL MOUNTED TEMPERATURE DISPLAY

DataNet Solution Wireless intelligent logging network

CAN EGT-8 Amplifier. About:

SINGLE-CHANNEL PROCESS CONTROLLERS & TEMP. DISPLAYS D3820/D3830 Series

Expanding the Basic GL7000 Configuration

Isolated Process Current Input 7B32 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Digital Clamp Meter User Manual

TC 6621 / TC Temperature test tools CALIBRATION. Measure and generation Protected for on-site use Perfect for control and tests

Signal Isolators & Conditioners

Digital Thermometer MEM. PC-Link READ

Operation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller

Ultra-Accurate Measurement for Temperature and Voltage

Stand-Alone, Low-Cost Data Loggers and Accessories. USB-500/600 Series Comparison Chart

LogBox-BLE. INSTRUCTION MANUAL V1.0x NOVUS AUTOMATION 1/51

DBC 150/650 Series. Druck Dry Block Temperature Calibrators. GE Sensing. Features

PCAN-TJA1054 Bus Converter High-Speed CAN to Low-Speed CAN. User Manual. Document version ( )

PROGRAMMABLE NUMERIC CONVERTER CNL 45 CNL 45L CONFIGURATION AND UTILISATION HANDBOOK

USER'S MANUAL DCL-650

Digital Indicating Controllers. JC Series. High performance controllers...at the lowest prices anywhere! Toll Free

High Voltage Power Supplies

RPS-401A, RPS-426A Self-Contained Ultrasonic Sensor

Piccolo2-OEM Fiber-optic Version (Item nr. PICO2-OEM)

Technical Data. C-DIAS-MULTI l/o MODULE CIO 013

LED55WPG Programmable LED Driver

SY021 Portable Load/Force Meter User instructions

DC resistance thermometry bridge Model CTR6000

INSTALLATION INSTRUCTIONS

midi LOGGER 820 GL820-UM-251

AMC 8 & IOC 8T. Analog Monitoring Card Type AMC 8 and Input / Output Card Type IOC 8T FEATURES

Ultra-Accurate Measurement for Temperature and Voltage

TC-4 4 Channel Thermocouple Interface User Manual

RESISTANCE/CURRENT METER Model 863/6487

Transcription:

DaqPRO Calibration Procedure This document outlines calibration procedure for fourtec Fourier Technologies DaqPRO 5300. Calibration is performed on each of the DaqPRO s inputs. Use the calibration sheet at the end of the document to record the calibrated values, and also as a reference for the acceptable tolerance for each input. Please refer to the Calibration chapter in the DaqPRO User Guide before proceeding with the Calibration procedure as outlined below. For technical support please contact support@fourtec.com Note: When calibrating DaqPRO of firmware version 2.0 the logger samples every 10 seconds so the calibration process will be slower. On DaqPRO firmware version 3.0a to 3.0h the logger will sample every second. A - Required Equipment B - Pulse Input Test C - Calibration Procedure C1 - NTC-10K Calibration C2 - NTC-100K Calibration C3-0-10 V Sensor QA & Calibration C4 - Current Sensor QA & Calibration C5-50 mv Sensor QA & Calibration C6 - PT 100 2-wire Sensor Calibration C7 - PT 100 3-wire Sensor Calibration C8 - TCJ, K, T Sensor Calibration Appendix A: DaqPRO Calibration Sheet A - Required Equipment External power supply for DaqPRO Martel MC-1000 calibrator (or suitable alternative calibration device) Martel model 2001 calibrator with plug-in for TC-calibration (or suitable alternative calibration device) PC with CalibratorProject.exe and DaqLab software installed Two green terminal blocks connected to a black and red banana cable (+ is the red cable and - is the black cable). Seven terminal blocks with 680 ᾨ resistors. Seven terminal blocks) with short TCJ sensors. Seven terminal blocks with ±349.5 ᾨ resistors.

One terminal block with a resistor of precisely 6.8k ᾨ. USB cable for DaqPRO B - Pulse Input Test: 1. Turn on the Martel MC-1000 calibrator. 2. Using the V/mA/Hz button choose Out Hz. Choose any frequency between 1 and 10 and press Enter. a. Rate = every second. b. Samples = 500. 3. Open the Setup dialogue and select Pulse Meter in the first I/O entry. Click Finish. Input a frequency signal. 4. Input a Pulse signal using the green connector and the black and red banana cables. As mentioned before, the red cable goes to the + and the black to the. 5. Run the logger using Logger>Run and check whether the difference between the reading corresponds to the frequency displayed on the Martel MC-100 calibrator. 6. Stop logging using Logger>Stop and disconnect the DaqPRO using Logger>COM Setup>Work offline 7. Turn off the device using the Off button situated on the front panel. C - Calibration Procedure For all tests and calibrations use the following Setup: 1. Sample rate = Every second 2. Samples = 500 3. Average = 4 samples C1 - NTC-10K Calibration 1. Open Setup and select NTC-10K in the first I/O entry. Click Finish. 2. Short circuit the + and of IN 1. 3. Open the Calibrate Sensors dialogue and select the NTC-10K sensor for calibration. 4. In the Calibration dialog, click Restore default settings. 5. Click Start. 6. Wait while the readings stabilize. The peak point is the chosen one. 7. Click Get Zero > OK. 8. Select the Martel MC-1000 calibrator. There is an option, which allows outputting the resistance signal in V/Om/Hz using the 680 ᾨ resistance. 9. Set the DaqPRO IN 1 to 680 ᾨ resistance using the connector with the banana cables connected so as to check the calibration results. 10. To run the test, click Run and make sure that all temperature readings are within the 1 ºC range of 100 ºC. 11. Stop logging.

NTC-10K Input QA 1. Open Setup and select the NTC-10K sensor in all I/O entries. 2. Insert terminal blocks with 680 ᾨ resistors into INs 2 8 and press Run. 3. Check to see that all readings resemble those of IN 1. 4. Stop logging. C2 - NTC-100K Calibration 1. Open Setup and select NTC-100K in the first I/O entry. 2. Short circuit the + and of IN 1. 3. Open the Calibrate Sensors dialogue and select NTC-100K for calibration. 4. In the calibration dialog, click Restore default settings. 5. Click Start. 6. Wait while the readings stabilize. The peak point is the chosen one. 7. Click Get Zero > OK. 8. Input 6.8Kᾨ into IN 1 of the DaqPRO using the calibrator. 9. To run the test, click Run and make sure that all temperature readings are within the 1 ºC range of 100 ºC. 10. Stop logging. C3-0 to 10 Volt Input QA & Calibration 1. Open Setup and select 10 Volt in all of the I/O entries. Click Finish. 2. In the Martel MC-1000 calibrator select a 5V signal in the V/Om/Hz output. 3. Click Start. 4. Input in sequential order a 5V signal into all IN entries. There is an option to do this using two DVM probes that are mounted on the Martel MC-1000 calibrator outputs. 5. Check to see that all INs produce readings of 5V or within the immediate range of 5V. 6. In order to run the voltage test, select the 10V sensor in the first IN in the Setup dialogue and press Finish. 7. Click Start. 8. In the Martel MC-1000 calibrator select a 0V signal in the V/Om/Hz output and record the reading at IN 1. 9. In the Martel MC-1000 calibrator select a 10V signal in the V/Om/Hz output and record the reading at IN 1. 10. Calibrate the input according to these results. 11. In order to calibrate this sensor perform the following: a. Go to Logger>Calibrate Sensors>Sensor 10V b. Enter the values for 0V and 10V in the suitable places and press OK. 12. Check to see that calibration was successful by running the logger and verifying the values. 13. 0V and 10V must appear in the V/Om/Hz output.

C4 - Current Sensor QA & Calibration 1. Open Setup and select Current 4-20 ma in the first I/O entry. 2. Using the V/Om/Hz button, select current signal output to the ma IN of the Martel MC-1000 calibrator. Connect the DaqPRO to the calibrator. 3. Run logger and select a current of 15.38, record the logger value and then enter 5.13 ma and record the logger value. Click Stop. 4. If results are not within the acceptable tolerance, calibrate the sensor. 5. In order to calibrate this sensor go to Logger>Calibrate Sensors>Current 4-20 ma. 6. Enter the values 15.38 and 5.13 ma as Reference Values and the actual results as Logger Value. Click OK. 7. Run logger and check to see that the readings correspond to 15.38 and 5.13 ma. 8. Stop logging. C5-50mV Sensor QA & Calibration THIS SENSOR MUST BE CALIBRATED IN ALL CASES 1. Open Setup and select 50mV in the first I/O entry. 2. Send a signal of up to 50mV to the Martel MC-100 calibrator output. 3. Run logger and feed 0 and 50 mv signals to IN 1. 4. Record the resulting values. 5. In order to calibrate this sensor go to Logger>Calibrate Sensors>Sensor 50 mv. 6. Before calibrating sensor click Restore Defaults and wait until a Ready message appears on the screen. 7. Enter the resulting values in Logger Value and press OK. 8. Run logger and feed 0 and 50 mv signals to IN 1. Stop logging. 9. Check to see that the resulting readings after calibration correspond to 0 and 50 mv. If that fails, recalibrate sensor. C6 - PT 100 2-wire Sensor Calibration 1. Open Setup and select PT-100 2-wire in the first I/O entry. 2. Select the PT-100 sensor in the Martel MC-1000 calibrator. To accomplish that use the TC/RTD and Range keys. 3. Run logger and feed 0 and 350 C signals to IN 1. 4. Record the resulting readings. If these correspond to 0 and 350 C, proceed to the calibration of PT-100 3-wire. Otherwise proceed with the calibration process. 5. In order to calibrate this sensor go to Logger>Calibrate Sensors>Sensor PT 100 2-Wire. 6. Enter the resulting readings (0 & 350 C) in Logger Value and press OK. 7. Check to see that the resulting readings after calibration correspond to 0 and 350 C. If that fails, recalibrate sensor.

C7 - PT 100 3 Wire Sensor Calibration 1. Short circuit the -s of IN1 and IN5. The + insert as in the rest of the sensors. 2. Open Setup and select PT-100 3-wire in the first I/O entry. 3. Run logger and feed 0 and 350 C signals to IN 1. 4. Record the resulting values. If these do not correspond to 0 and 350 C proceed with the following calibration process. 5. In order to calibrate this sensor go to Logger>Calibrate Sensors>Sensor PT 100 3-Wire. 6. As Offset Value enter the difference between the 0 in reference value and 0 in logger value. Click OK. 7. Run the logger and check to see that the results correspond to 0 and 350 C. C8 TCJ, K and T Sensor Calibration 1. Turn on the Martel model 2001 calibrator and connect it to the PC using a 9 pin male-female cable (DT135). 2. Run the CalibratorProject.exe on the PC and select External Off to disable external temperature compensation. Note: If using an alternative calibrator, ensure external temperature compensation is disabled. 3. Close CalibratorProject.exe. 4. Go to Logger>Thermocouple compensation>disable to disable compensation on the DaqPRO. 5. Open Setup and select TCJ sensor in the first I/O entry. Click Finish. 6. Run logger and feed 0 and 1000 C signals using the plug connected to the Martel model 2001 calibrator TC output. 7. Record the resulting readings and in case they do not correspond to the input values 0 and 1000 C, calibrate the thermocouple. 8. In order to calibrate this sensor go to Logger>Calibrate Sensors>Sensor Temperature TCJ. 9. As Offset Value enter the difference between the 0 in reference value and 0 in logger value. Click OK. 10. Run the logger and check to see that the results correspond to 0 and 1000 C (within the acceptable tolerance). 11. To calibrate TCK sensor, repeat the procedure in steps 1-10 above but select the TCK input during logger Setup, and on the calibrator. 12. To calibrate TCT sensor, repeat the procedure in steps 1-10 above but select the TCT input during logger Setup, and on the calibrator. In addition, the reference point for TCT calibration are 0 and 350 C. Input Test with TCJ Plugs 1. Open Setup and select Internal Temperature sensor in the first I/O entry and TCJ in INs 2 7. 2. Enable surrounding temperature compensation using Logger>Thermocouple compensation>enable.

3. Run logger and check to see that the readings in all INs entries do not deviate substantially from those in IN1. 4. Turn off device and disconnect from the power supply. Turn back on and check to see that the date and time were saved correctly. 5. Fill in the calibration form corresponding to the version of DaqPRO.

S/N: Appendix A: DaqPro Calibration Sheet Voltage 0 V ±50 mv 10 V ±50 mv 0-50 mv 0 mv ±250 uv 50 mv ±250 uv PT100 2-wire 350 ºC ±1.5 ºC PT100 3-wire 350 ºC ±1.5 ºC TCJ 5 1000 ºC ±5 ºC TCK 5 1000 ºC ±5 ºC TCT 5 350 º C ±2 ºC NTC-10K (Test after Get Zero process) Calibrator DaqPro Pass/Fail 100 ºC (680Ω) ±1 ºC NTC-100K (Test after Get Zero process) Calibrator DaqPro Pass/Fail 100 ºC (6.8KΩ) ±1 ºC Current 4-20 ma 5.13 ma ±100 ua 15.38 ma ±100 ua All inputs 0-10 V Pulse Counter NTC-10K (All inputs) PASS/FAIL PASS/FAIL PASS/FAIL Tester Name: Date: