IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

Similar documents
IDEAL INDUSTRIES INC. TECHNICAL MANUAL MODEL:

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS:

Mini Digital Multimeter

Model INSTRUCTION MANUAL DIGITAL MULTIMETER

OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER

OPERATING INSTRUCTION. Pen-Type Digital Multimeter

ETHOS Auto Ranging Digital Multimeter

User s Guide. 600A AC Clamp Meter. Model 38387

Owner's Manual. True RMS Multimeter. Model No Safety Operation Maintenance Español

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

Mini Digital Multimeter

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at

User's Guide. MiniTec TM Series Model MN25 MultiMeter

SBS -600 Graphical Digital Multimeter SBS- 700 Multi-Function Oscilloscope. Quick Start Guide

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

Mini Digital Multimeter

MTP INSTRUCTION MANUAL

Digital Clamp Meter User Manual

CM-220 True RMS AC CLAMP METER INSTRUCTION MANUAL

R5050. Model. Instruction Manual. TRMS AC/DC Clamp Meter. reedinstruments. www. com

Model A Mini AC/DC Clamp Meter. User's Guide

Model R5010. TRMS Digital Multimeter. Instruction Manual

OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER MAS830 MAS830L

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: 57314

User Guide True RMS Multimeter Extech EX205T

USER'S MANUAL DCL-650

User's Guide. Digital Multimeter plus IR Thermometer. Extech 450 Patented

Model R5010. Instruction Manual. TRMS Digital Multimeter. reedinstruments www.

Autoranging Multimeter plus IR Thermometer Extech 450 Patented

USER MANUAL. True RMS Multimeter. Extech EX430A

POCKET MULTIMETER Model No: MM18

Mini Digital Multimeter Model MN15. User's Guide

User's Guide. 800 Amp Clamp Meters. EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter

User's Guide. True RMS Multimeter plus IR Thermometer. Extech 470 Patent Pending

AC/DC TRUE RMS CLAMP METER ACM-2036

True RMS Multimeter with IR Thermometer

Autoranging Mini Multimeter

ATK-2040 AC/DC TRMS Watt Clamp Meter Users Manual

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

OPERATING INSTRUCTION

Autoranging True RMS Multimeter User Manual

2 in1 LAN Tester & Multimeter. Model: PCE-LT 1

Digital Multimeter User's Manual

MARTINDALE INSTRUCTIONS MM35 DIGITAL MULTIMETER ELECTRIC. Trusted by professionals

USER MANUAL. True RMS Multimeter plus IR Thermometer. Extech EX470A

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com. TRIPLETT Model 9055

OWNER S MANUAL 9908-TE. HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER

User Manual Extech EX820 True RMS 1000 Amp Clamp Meter with IR Thermometer

700 Series 200 Amp Clamp Meters

AutoRanging Digital MultiMeter

User's Guide. Digital Multimeter. Model MN42

True RMS Multimeter with IR Thermometer

User's Guide. True RMS Multimeter plus IR Thermometer. Extech 470 Patented

2 in 1 LAN Tester and Multimeter Model:

DC / AC True RMS Watt Clamp DMM

Model 3526-B Digi-Probe Multimeter

User's Guide. Extech AM A AC Analog Clamp Meter

USER MANUAL. Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX310

7032 Digital-Analog Multimeter

Autoranging Mini Multimeter Model MN16A

Autoranging Multimeter Extech EX503

EARTH RESISTANCE & RESISTIVITY TESTER

Industrial Multimeter Extech EX510

Mini Multimeter with Non-Contact Voltage Detector (NCV)

Digital Clamp-on Meter Instruction Manual

Extech EX820 True RMS 1000 Amp Clamp Meter with IR Thermometer. User's Guide

Model: Pro93 TRUE RMS LEAKAGE CURRENT TESTER

True RMS Industrial Multimeter Extech EX530

DC 100 A Current Probe

True RMS Industrial Multimeter Extech EX520

Wireless TRMS Multimeter and Insulation Tester

TENMA INSTRUCTION MANUAL MULTI-TESTER

True RMS Multimeter with IR Thermometer

SBS-S SERIES DC LOAD BANK USER MANUAL Rev

Digital Clamp-on Meter Amp Plus. Instruction Manual

80W (16V/5A) Power Supply with USB Output. User Manual

Variable Temperature Digital Solder and Testing Station

Monarch Instrument 2002 all rights reserved MONARCH INSTRUMENT. True RMS Multimeter. Instruction Manual. Monarch 120

User Manual VX4234 Digital Multimeter Module

INSTRUCTION MANUAL. Model True RMS AC/DC 30A Mini Clamp-on Meter. Introduction. True RMS AC Current and Voltage

400Amp True RMS AC/DC Clamp Meter with IR Thermometer Model EX623

Wireless TRMS Multimeter

OPERATING MANUAL INDUSTRIAL MULTIMETER CMM-40

Digital Clamp-on Meter Amp Plus. Instruction Manual

Wireless TRMS Multimeter and Insulation Tester

True RMS Digital Bench Multimeter Model:

User's Guide Extech EX840 True RMS 1000 Amp Clamp Meter with IR Thermometer

W Warning To avoid electrical shock, remove test leads and any input signals from the Model 707 Loop Calibrator before opening the case.

User's Guide. Handheld MultiScope. Model

User s Guide. MultiPro Digital MultiMeter Series With PC Interface. Models: MP510 MP520 MP530

High Precision Quad Output DC Power Supply

TT-SI 50 50MHz DIFFERENTIAL PROBE USER S MANUAL

Thermocouple Calibrator

INSTRUCTION MANUAL. Triple Output DC POWER SUPPLY MODELS 1651A & 1652 MODELOS 1651A & El Manual de la Instrucción

Wireless TRMS Multimeter

Digital Clamp-on Meter Instruction Manual

WARNING CAUTION 1. SAFETY

Mid-span 802.3af PoE Injector. Installation Guide KPOE-100

Digital Multimeter Instruction Manual

Transcription:

99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-609 The Service Information provides the following information: Precautions and safety information Specifications Performance test procedure Calibration and calibration adjustment procedure Basic maintenance (replacing the battery and fuse) Form number: TM61609 Revision: 2. Date: June 2002

TABLE OF CONTENTS Title Page Introduction 1 Precautions and Safety Information 1 Safety Information 1 Specifications 2 General Specification 2 Voltage Specifications 2 Current Specifications 2 Resistance Specifications 2 Diode Check Specifications 2 Performance Verification 3 Calibration 4 Replacing the Battery 5 Replacing Fuses 5

Page 1 Introduction Warning To avoid shock or injury, do not perform the verification tests or calibration procedures described in this manual unless you are qualified to do so. The information provided in this document is for the use of qualified personnel only. Caution Model 61-609 contains parts that can be damaged by static discharge. Follow the standard practices for handling static sensitive devices. SAFETY Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use the product only as specified. CAUTION. These statements identify conditions or practices that could result in damage to the equipment or other property. WARNING. These statements identify conditions or practices that could result in personal injury or loss of life. Specific precautions Use proper Fuse. To avoid fire hazard, use only the fuse type and rating specified for this product. Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the product without the covers in place. Electric overload. Never apply a voltage to a connector on the product that is outside the range specified for that connector. Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads while they are connected to a voltage source. Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or damp conditions. Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 FAX 781.665.0780 - TestEquipmentDepot.com

Page 2 General specifications Characteristics Description Display 3-3/4 Digits LCD display Display Count 4000 count, maximum reading 3999 Overrange Indication LCD shows OL Sampling Rate 2 times/second Operating Environment: Relative Humidity 0 C to 40 C (32 F to 104 F) 80% for temperatures up to 31 C Decreasing linearly to 50% at 40 C Storage Environment: -20 C to 60 C (-4 F to 140 F) at <80% relative humidity Power source: One 9V Battery (NEDA 1604) Battery Live: 200 hours typical (alkaline) Low Battery Indicator: symbol indicates low battery voltage ma protection Fuse 500mA, 250V fast acting fuse Dimensions 5.0 H X 2.8 W X 1.4 D 128mmH X 72mmW X 36mmD Weight: Approximately 7.0 oz. or 200g including battery RANGES and ACCURACY SPECIFICATION Function Setting Ranges Accuracy AC Voltage 400 mv ± (1.8% + 4 digits) @45-100 Hz 4 V, 40 V ± (1.3% + 4 digits) @45-500Hz 400V, 750 V ± (1.3% + 4 digits) @50-60Hz DC Voltage 400 m / 4 / 40 / 400 / 1000 V ± (1.0% + 1 digits) AC Current 40 / 400 ma ± (1.5% + 4 digits) 10 A ± (3.5% + 4 digits) DC Current 40 / 400 ma ± (1.0% + 1 digit) 10 A ± (3.0% + 1 digit) Resistance 400 / 4K / 40K / 400K / 4M Ω 40 MΩ ± (0.8% + 3 digits) ± (3.5% + 5 digits) Diode Check DCV 2V @ 0.3 ma approx. ± (3.0% + 3digits) AC Converter: Average responding, RMS Calibrated to Sine Wave Overload Protection: AC and DC Voltage: 1000 VDC or 750VAC rms for no more than 60 seconds. Resistance: 250VDC or peak AC 10A input, un-fused for no more than 60 seconds. ma input, 500mA at 250V fast acting fuse.

Page 3 PERFORMANCE VERIFICATIONS Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning correctly. If the meter does not conform to any of the listed limits the calibration procedure must be performed. Performance Verification Preparation 1. Turn on the calibrator, allow calibrator to warm up. Temperature stabilization should be reached after 30 minutes. 2. Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5V DC. If the battery measures under 7.5V DC, replace the battery before beginning the performance test. 3. Input the values listed in Table 1. Table 1 Performance Verification Function Setting Input Low Limit High Limit ACV 400 mv 390.0 mv AC@ 60Hz 382.6 397.4 ACV 4 V 3.900 V AC@ 60Hz 3.845 3.955 ACV 40 V 39.00 V AC@ 60Hz 38.45 39.55 ACV 400 V 390.0 V AC@ 60Hz 384.5 395.5 ACV 750 V 700 V AC @60Hz 687 713 DCV 400 mv 390.0 mv DC 385.9 394.1 DCV 4 V 3.900 V DC 3.859 3.941 DCV 40 V 39.00 V DC 38.59 39.41 DCV 400 V 390.0 V DC 385.9 394.1 DCV 1000 V 900 V DC 890 910 ACmA 40 ma 39.00 ma AC @ 60Hz 38.37 39.63 ACmA 400 ma 390.0 ma AC @ 60Hz 383.7 396.3 AC A 10 9.00 A AC @ 60Hz 8.64 9.36 DCmA 40 ma 39.00 ma DC 38.60 39.40 DCmA 400 ma 390.0 ma DC 386.0 394.0 DC A 10 9.00 A DC 8.72 9.28 Ω 400 100.0 Ω 98.9 101.1 Ω 4K 1.000K Ω 0.989 1.011 Ω 40K 10.00K Ω 9.89 10.11 Ω 400K 100.0K Ω 98.9 101.1 Ω 4M 1.000M Ω 0.989 1.011 Ω 40M 10.00M Ω 9.60 10.40 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 FAX 781.665.0780 - TestEquipmentDepot.com

Page 4 CALIBRATION Calibration Preparation 1. Turn on the calibrator, allow calibrator to warm up. Temperature stabilization should be reached after 30 minutes. 2. Disconnect the test leads and turn the range switch to OFF. 3. Remove the screw from the back of the meter and uncover the back case. (refer to Figure 1) 4. The case bottom is secured to the case top by two internal snaps (at the LCD end). lift up on the battery end until the case un-snaps. 5. Using a calibrated meter ensure the battery measures a minimum of 7.5V DC. If the battery measures under 7.5V DC, replace the battery. Calibration Procedure It is recommended that all IDEAL meters under go the following calibration procedure on an annual basis. Volts DC Calibration 1. Set the function / range to DC V 2. Connect the calibrator to the V and COM inputs on the meter. 3. Output 3.900V DC. Adjust R42 (refer to Figure 2) until the display reads 3.900 +/-0.001 V Note: This is the only adjustment required for the 61-609. Calibration is complete. R42 500mA/250V Fuse screw 9V Battery Figure 1 Figure 2

Page 5 Battery Replacement 1. Disconnect the test leads from any circuit under test and turn off meter. 2. Remove the screw from the back of the meter and uncover the back case. (refer to Figure 1) 3. Install new 9V battery (NEDA #1604). An alkaline type is recommended. 4. Install new battery into compartment using care not to pinch or bind battery. 5. Replace bottom cover insuring that case is secured at the top by the internal snaps and replace screw. Replacing Fuse 1. Disconnect the test leads and turn the range switch to OFF. 2. Remove the screw from the back of the meter and uncover the back case. (refer to Figure 1) 3. The case bottom is secured to the case top by two internal snaps (at the LCD end). lift up on the battery end until the case un-snaps. 4. Remove the defective fuse (refer to Figure 2 for location of fuse) and replace with a 500mA, 250V fast acting fuse with Part #F-17 specifications. Always use fast acting, high interrupting type fuse. 5. Replace bottom cover insuring that case is secured at the top by the internal snaps and replace screw. Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 FAX 781.665.0780 - TestEquipmentDepot.com