User's Guide. Phase Sequence and Motor Rotation Tester Model

Similar documents
INSTRUCTION MANUAL. Model True RMS AC/DC 30A Mini Clamp-on Meter. Introduction. True RMS AC Current and Voltage

User's Guide. Extech AM A AC Analog Clamp Meter

User s Guide. 600A AC Clamp Meter. Model 38387

True RMS AC Voltage/Current Datalogger

User's Guide. Model Laser Photo Tachometer

9040/9040UK. Users Manual. Phase Rotation Indicator

User's Guide. Model High Precision Quad Output DC Power Supply

User's Guide. MiniTec TM Series Model MN25 MultiMeter

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389

Mini Digital Multimeter

User s Manual. Model and Sound Level Calibrators

User's Guide. Video Borescope. Model BR100

Mini Photo Tachometer Model Mini Contact Tachometer Model

User Guide. Moisture Meter. Model MO250

User's Guide. Datalogging Light Meter with PC Interface. Model

User's Guide. Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer. Patented

User's Guide. Cup Thermo-Anemometer. Model AN400

Laser Photo/ Contact Tachometer. m/min. ft/min

Digital Vane Thermo-Anemometer Model

User's Guide. Power Analyzer Model Power Analyzer Datalogger Model Introduction

3-Axis G-Force Datalogger

Heavy Duty Psychrometer + IR Thermometer Model HD500

User Guide. Heavy Duty Differential Pressure Manometer. Model HD750

Integrating Sound Level Datalogger Model

Heavy Duty Datalogging Light Meter

User's Guide. 1/16 th DIN ph Controller. Model 48PH2. Introduction

Heavy Duty Vibration Meter

User's Guide. Hot Wire Thermo-Anemometer Model

User's Guide. Digital Sound Level Meter. Model

User's Guide Video Borescope Model BR200

User Guide. ExStik TM Model RE300 Waterproof ORP Meter. Patent Pending RE300 ORP

INSTRUCTION MANUAL. Model Dual Input RTD Thermometer. Measures two temperatures simultaneously. Dual RTD probe inputs

Digital Sound Level Meter

User's Guide. Vibration Meter and Laser Combination Tachometer Model

User's Manual Sound Level Alert Model SL130

User's Manual. Model Heavy Duty Datalogging Module with Windows Software

SuperHeat Psychrometer Models RH350 and RH355 (kit)

Heavy Duty Datalogger Module

Heavy Duty Mini Vane CFM Thermo Anemometer

User's Guide. Heavy Duty Differential Pressure Manometer. Model

User Guide. ExStik FL700. Fluoride Meter

User Guide. Model Temperature Datalogger Kit Model Temperature and Humidity Datalogger Kit Model SW276 Datalogging Software SW276

User's Guide. Hot Wire Thermo-Anemometer Model Introduction

Heavy Duty CFM Thermo-Anemometer Model

User's Guide. Programmable DC Power Supply 200 Watt (40 Volts / 5 Amps) Model Introduction

User's Guide. Heavy Duty Vane Thermo-Anemometer. Model

Autoranging Multimeter plus IR Thermometer Extech 450 Patented

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at

Heavy Duty Datalogger Module

User's Guide. Temperature / Humidity Datalogger. Model 42270

Power Analyzer Datalogger Model

Autoranging Mini Multimeter Model MN16A

User's Guide. Mini Microscope. Model MC108

MTP INSTRUCTION MANUAL

Mini Digital Multimeter

User's Guide. Model Noise Dosimeter with PC Interface. Introduction

User's Guide. Vane Thermo-Anemometer Datalogger. Model Introduction

User's Guide. True RMS Multimeter plus IR Thermometer. Extech 470 Patent Pending

User's Manual. Digital Sound Level Meter. Model

LINE VOLTAGE TESTER CT101 USER S MANUAL. Please read this manual carefully and thoroughly before using this product.

Mini Digital Multimeter Model MN15. User's Guide

POCKET MULTIMETER Model No: MM18

User's Guide. Digital Multimeter. Model MN42

USER MANUAL. Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX310

Hygro Thermo-Anemometer Model

User Guide True RMS Multimeter Extech EX205T

USER GUIDE Current Calibrator

Model A Mini AC/DC Clamp Meter. User's Guide

Oyster Meter ph-conductivity-tds-salinity-orp(mv)

Mini Digital Multimeter

User's Guide Extech EX840 True RMS 1000 Amp Clamp Meter with IR Thermometer

APO F T1 F T1-T2 ESC - 0 ENTER. [Limits] CLR? SET[ ] LOG READ SAVE READ SHIFT [OFS] [APO] MAX/MIN REL HOLD [TIME] T1-T2 TYPE T1/T2 TYPE T1/T2

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

1000A 3-Phase Power Analyzer/Datalogger

User's Guide. 800 Amp Clamp Meters. EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter

CM-220 True RMS AC CLAMP METER INSTRUCTION MANUAL

High Precision Quad Output DC Power Supply

Earth Resistance and Resistivity Tester

Owner's Manual. True RMS Multimeter. Model No Safety Operation Maintenance Español

USER'S MANUAL DCL-650

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide

USER GUIDE. Dual Input True RMS AC Voltage/Current Datalogger. Model DL160

Variable Temperature Digital Solder and Testing Station

User's Guide. Model (60Hz) Model (50Hz) 1000A 3-Phase Power Analyzer/Datalogger NO.

Dual Input, 3-Display Thermometer/Datalogger Model

Visit us at

Autoranging True RMS Multimeter User Manual

OPERATING INSTRUCTION

v3001 FC Wireless DC Voltage

Autoranging Mini Multimeter

OX 5022-CK OX 5042-CK

OPERATING INSTRUCTION. Pen-Type Digital Multimeter

AC/DC Current Probe Model SL201

Automated Tuner System Power Distribution Hub

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

USER MANUAL. True RMS Multimeter. Extech EX430A

Thermocouple Calibrator

374 FC/375 FC/376 FC. Clamp Meter. Safety Information

700 Series 200 Amp Clamp Meters

Solar Installation PV100 Operating Instructions

Transcription:

User's Guide Phase Sequence and Motor Rotation Tester Model 480403

Introduction Congratulations on your purchase of the Extech Model 408403 Motor and Phase Rotation Indicator. This handheld instrument detects the rotational field of three-phase systems and determines motor-rotation direction. Color-coded test leads are provided for connecting to the system or motor under test. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Safety International Safety Symbols! Caution! Refer to the explanation in this Manual Caution! Risk of electric shock Earth (Ground) Double Insulation or Reinforced insulation AC, Alternating Current or Voltage DC, Direct Current or Voltage Safety Procedures To avoid possible electric shock or fire, observe the following: Read the following information carefully before using or servicing the instrument. Adhere to local and national safety codes. Individual protective equipment must be used to prevent shock and injury. Use of instrument in a manner not specified by the manufacturer may impair safety features/protection provided by the equipment. Avoid working alone. 2

Inspect the test leads for damaged insulation or exposed metal. Check test lead continuity. Damaged leads must be replaced. Do not use the meter if it appears damaged. Use care when working above 30V ac rms, 42V ac peak and 60V dc. Such voltages pose a shock hazard. When using the probes, keep fingers away from probe contacts. Keep fingers behind the finger guards on the probes. Measurements can be adversely affected by impedances of additional operating circuits connected in parallel or by transient currents. Verify operation prior to measuring hazardous voltages (voltages above 30V ac rms, 42V ac peak and 60V dc). Do not use the meter with any of the parts removed. Do not use the meter near explosive gas, vapor, or dust. Do not use the meter in a wet environment. Warranty EXTECH INSTRUMENTS CORPORATION (a FLIR company) warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a six month limited warranty applies to sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department at (781) 890-7440 ext. 210 for authorization or visit our website www.extech.com for contact information. A Return Authorization (RA) number must be issued before any product is returned to Extech. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied. 3

Meter Description 1. L1 test Lead input to meter 2. L3 test lead input to meter 3. L2 test lead input to meter 4. L1 status LED 5. Orientation symbol for noncontact testing 6. L3 status LED 7. L2 status LED 8. Power ON status LED 9. ON button 10. Left/Right Clockwise/Counter- Clockwise status LEDs 11. Test lead alligator clips Operation Determine Phase Rotation Direction 1. Connect the supplied color-coded test leads to the meter s test lead input jacks at the top of the meter. 2. Connect the test probes to the three mains phases for the system under test. 3. Press the ON/OFF button to turn the meter on. 4. L1, L2, and L3 indicators will illuminate for each phase connected, respectively. 5. The clockwise and counter-clockwise arrows with the left/right L or R icons display the phase rotation direction (phase sequence) of mains as connected. 6. To understand the status of a test, based on the condition of the meter s front panel LEDs, refer to the table printed on the meter: Note: The rotational indicator LEDs illuminate even if one of the test probes is connected to a neutral or ground conductor instead of one of the mains phases. 4

Check Motor Turn Direction (contact method) 1. Connect the supplied color-coded test leads to the meter s test lead input jacks at the top of the meter. 2. Connect the test (alligator) probes to the three motor connections (L1 to U, L2 to V, and L3 to W). 3. Press the ON/OFF button. The green ON indicator shows that the instrument is ready for testing. 4. Turn the motor shaft half of a revolution towards the right. 5. The clockwise and counter-clockwise arrows with the left/right L or R icons display the orientation of the motor. Check Motor Turn Direction (non-contact method) 1. Disconnect the test leads from the meter and from any other devices under test. 2. Hold the meter close (one inch or closer) to the motor, parallel to the length of the motor shaft. 3. The bottom of the meter should face the drive shaft and the back of the meter should be flush with the motor (refer to diagram at right). Note: There is an orientation symbol on the meter to assist. 4. Press the ON/OFF button. The green ON indicator shows that the instrument is ready for testing. 5. The orientation of the motor is represented by the clockwise and counter-clockwise arrows with the left/right L or R icons display. Face the rear of the motor (meter will appear upside down) when reading the status of the display LEDs. Note: The meter will not operate with motors controlled by frequency converters. 5

Magnetic Field Detection 1. Hold the meter in the area under test. 2. A magnetic field is present if either the clockwise or the counterclockwise LED illuminates. Battery replacement The meter uses a 9V battery (supplied) as a power source. If the green LED does not light when the ON/OFF button is pressed, replace the battery. 1. Place the meter on a clean, flat, non-abrasive surface. 2. Loosen the rear bottom Phillips head screw to access the battery compartment. 3. Lift the battery access door and replace the 9V battery observing polarity. 4. Place the battery carefully in the compartment without crimping the black and red battery connecting wire. 5. Close the access door and secure the screw. Calibration and Repair Services Extech offers repair and calibration services for the products we sell. Extech also provides NIST certification for most products. Call the Customer Care Department for information on calibration services available for this product. Extech recommends that annual calibrations be performed to verify meter performance and accuracy. Support line (781) 890-7440 Technical Support: Extension 200; E-mail: support@extech.com Repair & Returns: Extension 210; E-mail: repair@extech.com Product specifications subject to change without notice For the latest version of this User Guide, Software updates, and other up-to-the-minute product information, visit our website: www.extech.com Extech Instruments Corporation, 285 Bear Hill Road, Waltham, MA 02451 Copyright 2008 Extech Instruments Corporation (a FLIR company) All rights reserved including the right of reproduction in whole or in part in any form. 6

Specifications Environmental Specifications Operating Temperature 32 to 104 o F (0 to 40 o C) Operating Humidity 15 to 85% relative humidity Operating altitude 2000m Pollution Degree 2 Approvals DIN / VDE 0411 IEC 61010 DIN / VDE 0413-7 Mechanical Specifications Dimensions (H x W x D): 5.1 x 2.7 x 1.3 (130 x 69 x 32mm) Weight 4.6 oz. (130g) Safety Specifications Electrical Safety Approvals IEC 61010/EN61010, IEC 61557-7/EN 61557-7 Maximum Operating Voltage (Ume) 600 V Protection Levels CAT lll 600V Electrical Specifications Battery type 9V battery Current Consumption 20mA max. Battery life One (1) year minimum Nominal Voltage 40 to 600 VAC Frequency Range (fn) 2 to 400Hz Nominal Test current (in per phase) < 3.5mA 7