CT Reconstruction with Good-Orientation and Layer Separation for Multilayer Objects

Similar documents
High-resolution X-ray CT Inspection of Honeycomb Composites Using Planar Computed Tomography Technology

New Features of SIMTech s Unique CT Reconstruction and Visualization Software

Non-Destructive Failure Analysis and Measurement for Molded Devices and Complex Assemblies with X-ray CT and 3D Image Processing Techniques

Case Studies in the Use of Computed Tomography for Non-Destructive Testing, Inspection and Measurement

CIVA Computed Tomography Modeling

Non-destructive, High-resolution Fault Imaging for Package Failure Analysis. with 3D X-ray Microscopy. Application Note

Coordinate Measuring Machines with Computed Tomography

ULTRASONIC TESTING AND FLAW CHARACTERIZATION. Alex KARPELSON Kinectrics Inc., Toronto, Canada

INDUSTRIAL SYSTEM DEVELOPMENT FOR VOLUMETRIC INTEGRITY

Advanced Computed Tomography System for the Inspection of Large Aluminium Car Bodies

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications

Translational Computed Tomography: A New Data Acquisition Scheme

A Large Scale Multiple Source X-ray CT System for Aerospace Applications

3D Computed Tomography (CT) Its Application to Aerospace Industry

MULTI-PURPOSE 3D COMPUTED TOMOGRAPHY SYSTEM

AUTOFOCUS SENSORS & MICROSCOPY AUTOMATION IR LASER SCANNING CONFOCAL MICROSCOPE IRLC DEEP SEE. Now See Deeper than ever before

Void Detection in Large Solder Joints of Integrated Power Electronics. Patrick Schuchardt Goepel electronics LLC

Reconstruction Methods for Coplanar Translational Laminography Applications

3D X-ray Laminography with CMOS Image Sensor Using a Projection Method for Reconstruction of Arbitrary Cross-sectional Images

Digital Volume Correlation for Materials Characterization

Tomographic X-ray data of a walnut

An Acquisition Geometry-Independent Calibration Tool for Industrial Computed Tomography

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Continuous and Discrete Image Reconstruction

UMASIS, an analysis and visualization tool for developing and optimizing ultrasonic inspection techniques

Computer-Tomography I: Principles, History, Technology

Optiv CT160 Accurate measurement for every detail. Computed Tomography System

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

X-rays see all X-ray computed microtomography (µct) a novel technique available at CERN for material and metrological inspection

Characterization of microshells experimented on Laser Megajoule using X-Ray tomography

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY

LABORATORY SYSTEM FOR X-RAY NANOTOMOGRAPHY

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

Vi Technology 3D Solder Paste Inspection Solutions

Using X-Ray Micro Tomography to assess the quality of Packaging Foil

Computed tomography (Item No.: P )

Advanced materials research using the Real-Time 3D Analytical FIB-SEM 'NX9000'

Measurements using three-dimensional product imaging

Improved Inspection of Miniaturized Interconnections by Digital X-ray Inspection and Computed Tomography

Enhanced material contrast by dual-energy microct imaging

Image Acquisition Systems

Scanning Acoustic Microscopy For Metrology of 3D Interconnect Bonded Wafers

UMASIS, AN ANALYSIS AND VISUALIZATION TOOL FOR DEVELOPING AND OPTIMIZING ULTRASONIC INSPECTION TECHNIQUES

John R. Mandeville Senior Consultant NDICS, Norwich, CT Jesse A. Skramstad President - NDT Solutions Inc., New Richmond, WI

Theoretically-exact CT-reconstruction from experimental data

Introduction to Biomedical Imaging

Phase Reversed Image Reconstruction Method of. Industrial Ultrasonic CT

DAMAGE INSPECTION AND EVALUATION IN THE WHOLE VIEW FIELD USING LASER

Optimized Design of 3D Laser Triangulation Systems

Recognition and Measurement of Small Defects in ICT Testing

Automatic Integrated Circuit Die Positioning in the Scanning Electron Microscope

X-ray Tomography. A superficial introduction, but sufficient enough to get us started in surgical navigation.

O-ARM IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE

Development of an automated bond verification system for advanced electronic packages

Wieblinger Weg 92a, Heidelberg, Germany, Phone: , Fax: ;

LJMU Research Online

Industrial Computed Tomography

Dr. Javier Santillan, San Carlos, CA

Using Superpowers for Hardware Reverse Engineering. Joe Grand Grand Idea Studio, Inc.

Multi-spectral (W, Mo, Cu, and Ag) XCT measurements

TFT-LCD Technology Introduction

The new generation of industrial computed tomography. The Desktop CT exact S

[HALL PROBE GRADIOMETRY ]

Rule-based inspection of Wafer surface

MCT225 HA Metrology CT

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

Adapted acquisition trajectory and iterative reconstruction for few-views CT inspection

MEDICAL IMAGING 2nd Part Computed Tomography

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT

A new Concept for High-Speed atline and inlinect for up to 100% Mass Production Process Control

Plane Wave Imaging Using Phased Array Arno Volker 1

Optical Topography Measurement of Patterned Wafers

The Electrochemical Innovation Lab X-ray Suite: from macro- to nano-ct

Alignment and Other Challenges in Reconstructing Cryotomograms with IMOD

Traceable 3D X-Ray Measurements ZEISS METROTOM

Applications of Temporal Phase Shift Shearography

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units

X-ray Industrial Computed Laminography (ICL) Simulation Study of Planar Objects: Optimization of Laminographic Angle

Fast Reconstruction of CFRP X-ray Images based on a Neural Network Filtered Backprojection Approach

Digital Image Processing

4. Recommended alignment procedure:

Virtual Interaction System Based on Optical Capture

VJ Technologies Inspection Services Division. By Bob Maziuk, February 2015

How does the ROI affect the thresholding?

VALIDATION OF THE SIMULATION SOFTWARE CIVA UT IN SEPARATED TRANSMIT/RECEIVE CONFIGURATIONS

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

Central Slice Theorem

Available online at ScienceDirect

Medical Image Processing: Image Reconstruction and 3D Renderings

3D X-Ray Microscopy: A Non Destructive High Resolution Imaging Technology That Replaces Physical Cross-Sectioning For 3DIC Packaging

Advanced Reconstruction Techniques Applied to an On-Site CT System

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset

ADVANCED ULTRASOUND WAVEFORM ANALYSIS PACKAGE FOR MANUFACTURING AND IN-SERVICE USE R. A Smith, QinetiQ Ltd, Farnborough, GU14 0LX, UK.

Computed tomography of simple objects. Related topics. Principle. Equipment TEP Beam hardening, artefacts, and algorithms

Advanced Non-Destructive Testing by High Resolution Computed Tomography for 3D analysis of Automotive Components

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Advances in Flexible Hybrid Electronics Reliability

Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing

Transcription:

17th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China CT Reconstruction with Good-Orientation and Layer Separation for Multilayer Objects Tong LIU 1, Brian Stephan WONG 2, Tai Chong CHAI 3, and Andrew. A. MALCOLM 1 1 Singapore Institute of Manufacturing Technology 71 Nanyang Drive, Singapore 638075 Phone: +65 67938326; Fax: +65 67916377; E-mail: tliu@simtech.a-star.edu.sg 2 School of MAE, Nanyang Technological Univ., Singapore; E_mail: 3 Institute of Microelectronics, Singapore; E-mail: Abstract Nowadays, multilayer designs such as temperature co-fired ceramic (LTCC) and stacked IC packaging are increasingly popular in electronics and semiconductor industries. For these multilayer structures, X-ray CT is thought as one of the essential tools for inspecting their internal defects. A common requirement for X-ray CT inspection of multilayer objects is to visualize the reconstructed CT results in separated layers so that defect analysis can be conducted layer-wise and conveniently corresponded to manufacturing process. One characteristic of multilayer objects is that the layers are usually thin and have a large area-to-thickness aspect ratio. This property sometimes may lead to failure in inspecting thin defects. For example, thin and large delamination-like defects in a multilayer packaging are in many cases invisible to X-ray 2D imaging. If the user has no prior knowledge about their existence, they are also quit easy to be missed with CT images if the reconstructed object has a tilted orientation. In this paper, we demonstrate that a good-oriented CT reconstruction for planar object can make layer separation much more simple and effective. Through the study of a 5 layer packaging sample, we also show that more detailed information of the delamination-like defects can be observed. Subject terms: Computed tomography, multilayer object, layer separation, NDT, CT. 1. Background Traditional micro-ct usually reconstructs an object without considering its actual scanning orientation with respect to the detector plane and the rotary unit. As a result, the object is usually reconstructed with a tilted orientation (Fig. 1a). Although this has no any impact on the reconstruction quality, it indeed causes some trouble in the processes of separating layers in a reconstructed multilayer component such as stack IC chip or low temperature co-fired ceramic (LTCC) device [1]. Several powerful software packages are available in the market for CT visualization, such as VGStudio Max 2 and I-view 3. With these softwares, the way to separate a multilayer component with a tilted reconstruction orientation is to cut it with a clip-plane defined along the normal of the layer plane. This process is generally very time-consuming even for experienced users. It is also quite subjective because it relies purely on the personal judgment of the user for the definition of the clip-plane. An imperfect definition would lead to an unacceptable separation of layers (Fig. 1b).

The tilted orientation of a reconstructed planar object comes from two sources. Firstly, almost all CT software performs the reconstruction with a default scanning start angle, for example zero. However, in reality, an object may be scanned with any initial orientation, as illustrated in Fig.2 (a). A difference between the default start angle and the actual start angle would result in the titling of the object CT slice with respect to the reconstruction matrix. Secondly, when mounting a planar object, its primary plane may form a small angle with the rotation axis (Fig.2). This situation would lead the reconstructed object to be misaligned in the third dimension in a volume reconstruction process. (a) Fig. 1 (a) An object is reconstructed with a tilted orientation with respect to the dimensions of reconstruction box; A poor cutting with a inaccurate clip-plane definition. In principle, this problem can be solved physically by carefully positioning the planar object on the rotator before the scanning. However, in many cases this has proven to be impractical or at least very difficult to achieve because of the lack of accurate referencing and the variation of sample shape and size. Recently we have developed an automated algorithm that enables users to reconstruct planar objects with a wanted orientation [4,5]. That is, to make the object s primary plane aligned with one plane of the reconstruction volume so that layer separation can be carried easily along one dimension of the reconstruction volume. This technology first uses one sinogram of the object to automatically determine the actual start angle of a scan, with which the CT reconstruction is conducted to achieve an aligned orientation between the object slice and the reconstruction matrix. After all slices reconstructed, the height position of the object cross-section on each slice is calculated and an image shifting operation to all slices is applied to make the object is also well aligned in the third dimension of the reconstruction volume. As a result, the primary plane of the reconstructed object will be parallel to one of the planes of the reconstruction volume. With such an orientation, displaying and extracting layers becomes a trivial operation. In this paper, through investigating the defects inside a five-layer packaging sample, we will demonstrate the benefit of the new technology over the traditional method. To support our statements and conclusions drawn from the study, some C-mode Scanning Acoustic Microscopy (CSAM) images are also provided as reference [6]. 2. Methodology The sample was investigated with C-mode Scanning Acoustic Microscopy (CSAM) first and then with a micro-focus X-ray inspection machine (XIM). The comparison of the CSAM image and 2D X-ray image showed some inconsistent results, which made it hard to draw a conclusion. Then an X-ray CT scan was used to create the 3D model of the sample. First, we reconstructed the object by using commercial software with a default scanning 2

start angle. However, due to the existence of a small angle between the primary plane of the sample and the rotation axis and the difference between the actual scanning start angle and the default one, we obtained a reconstructed object with a titled orientation. By carefully defining a clip-plane we were able to extract the images of the two interface layers that are to an extent matching the images of the CSAM images, however, without any prior knowledge about the defects, we were unable to judge how good is the layer images obtained and whether we could observe more details from the reconstructed results. Therefore, we re-did the volume reconstruction with the actual scanning start angle and then perform an axial titling correction to the reconstructed data. This process gave us a reconstructed object with an expected orientation. With this good orientation, we were able to easily extract the individual layers with much more details through an automated layer separation approach. Fig.2. The scanning start orientation of a planar object. (a) An arbitrary start angle; An angle exists between the primary plane of the object and the rotation axis. 3. Experiment and Discussion 3.1 Sample Description The sample to be investigated in this paper is a five-layer packaging sample as illustrated in Fig.3. The three silicon layers are assembled with adhesive materials that form the two interface layers. It has a size of 10mm 10mm 2mm. Each silicon layer is 600µm and each interface layer is 100µm. This sample was specially prepared for evaluating the assembly processing. The inspection objective was to check what would have happened on the two adhesive layers. Interesting defects include voids and delamination-like gaps. Fig. 3. The five-layer package sample 3.2 CSAM and X-ray 2D Images The sample was investigated with C-mode Scanning Acoustic Microscopy (CSAM) first and then with a micro-focus X-ray inspection machine (XIM). For CSAM inspection, one of the two interfaces was scanned first (Fig.4 (a), labelled as the first interface). The sample was then inverted and the second interface was scanned (Fig. 4). For comparison purpose, one X-ray 2D image was also captured with same position as the CSAM images (Fig. 4(c)). 3

The CSAM system used in the experiment was a Sonoscan D300. The sample was scanned at a frequency of 230 MHz, with a scanning area of 20mm-by-20mm and a resolution of 256-by-240 pixels. 4

17th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China The X-ray inspection machine was a Comet/Feinfocus Fox 160.25. It has an open tube with spot size as small as 700nm, and a 200mm x 197mm direct digital detector (Varian Paxscan 2520). The detector has 1408-by-1888 pixels and each pixel is 127 µm in size. The detector can work in either full frame or binning mode in order to generate images of different size to optimise image acquisition time and data storage. (a) (c) Fig. 4 CSAM and X-ray Images. (a) CSAM image of the first interface; CSAM image of the second interface; (c) X-ray 2D image. In principle, voids, cracks, and delamination, if detectable, will show as white areas in both the CSAM and X-ray 2D images. For CSAM this is because acoustic waves cannot pass through an air gap, and for X-ray, because these kinds of defects lead to less absorption of the X-ray beam. For X-ray inspection, it generally makes no difference to have the sample facing up or down. For CSAM, by setting an appropriate time gate, the two interfaces can be scanned and inspected separately. It is easy to become confused when comparing the X-ray images and the CSAM images. While most of the defects on the CSAM image of the first interface can also be found on the X-ray image, the CSAM image of the second interface shows little similarity to the X-ray image. Obviously, there were some thin and large delamination-like defects existing on this interface, however, because they are too thin, they are not visible on the X-ray 2D image. 1 st Interface 2 nd Interface (a) Fig. 5. One CT slice (a) and the 3D model of the reconstructed part of the object with the new method. 3.3 CT reconstruction with the actual start angle and axial tilting correction As 2D X-ray inspection was unable to detect this kind of delamination-like defects, we then resorted to the CT inspection. The sample was scanned with a source-to-image distance of 680mm and a source-to-object distance of 61mm. The source was set at an accelerating voltage of 100kV and a tube current of 37µA using a

tungsten target. The object was rotated through 360º to complete the scan with one projection image being taken at each degree. One sinogram was then created, from which the actual scanning start angle can be determined [4]. Then the cone-beam volume reconstruction algorithm was used to reconstruct the object with the actual scanning start angle obtained. One slice image was shown in Fig. 5a, in which the object dimensions were well aligned with the dimensions of the reconstruction matrix. Totally 500 axial slices were reconstructed in this study for the area shown in Fig. 4. Then, the tilting angle of the object s primary plane with respect to the rotation axis was determined from the projection that has the narrowest shadow on detector after performing a fan-beam to parallel-beam conversion. This tilting angle was then used to perform a image shifting operation to all slices to make the reconstructed object also aligned with the reconstruction box in the third dimension, as shown in Fig. 5b [5]. As a consequence, the primary plane of the reconstructed object was parallel to one plane of the reconstruction box and therefore. If a layer-separation operation was necessary, it could be accomplished by cutting the object digitally along one dimension of the reconstruction box. (a) (c) Fig.6 Automated layer separation. (a) The concept; the first interface; (c) the second interface 3.4 Automated layer separation With a good oriented reconstruction, it is also able to perform an automated layer separation. As illustrated in Fig. 6, to obtain the individual layers in a sample, we performed a sum operation over each row and generated a integrated intensity distribution over the row number as shown in Fig. 6a. With the positions of the two layers determined, the images of the two individual layers were obtained automatically (Figs. 6b and 6c). (a) (c) (d) (e) (f) (g) Fig.7. The cross-sectional CT images of the second interface with the new method. Another advantage of the reconstruction with a good orientation lies on the availability to easily show the variation within the interface layer in this study. From the distribution illustrated in Fig. 6a, one could have an 6

estimate about the width of the interfaces that might be several pixels, depending on the reconstruction resolution in the thickness dimension. Therefore, to display the slice with a one-pixel increment, we were able to view the variation of the defects in the interface, as shown in the Fig. 7. This detailed information in Fig. 7 showed rather different scenario to that shown in the CSAM image. Fig.7(a)-(g) showed the cross-sectional images of the second interface from bottom to top. One could immediately notice that Fig.7(c) was very similar to the defects shown in Fig. 4. However, for other positions, the defect was quite different. In fact, the several defects recognized in Fig.4 were actually located at different heights within the second interface. Some of them were located between the second silicon layer and the adhesive material layer; some of them were between the adhesive layer and the third silicon layer. For example, the delamination-like areas shown in Fig. 7(c) as lower gray level became adhesive material areas in Fig. 7(e). What this means to the assembling process might be a question we need to answer in the future. 4. Conclusion In this paper, we demonstrate that delamination-like defects can be more effectively inspected and visualised with an automated good-oriented reconstruction for multilayer planar objects. Due to the small thickness of this kind of defects, 2D X-ray inspection is usually unable to detect them reliably. They are also generally not guaranteed to be detected with traditional CT reconstruction and visualisation approaches due to the tilted orientation of the reconstructed object and the subjective definition of a clip-plane. Our new method allows a good-oriented reconstruction of the planar object and an automatic layer separation of internal layers. Compared to traditional approach, the present method is automated effective and subjective. In many cases, it also allows the user to extract more feature details within a thin layer. This characteristic is proven to be useful in analysing the forming mechanism of the voids or delamination in packaging process. References [1] M. Pacheco, Z. Wang, L. Skoglund, Y. Liu, A. Madina, A. Raman, R. Dias, D. Goyal and S. Ramanathan, Advanced fault isolation and failure analysis techniques for future package technologies, Intel Technology Journal: vol. 9(4), 2005. [2] User s Manual, VGStudio Max 1.2, first printing, Mar, 2004. [3] Users Manual, i-view Workstation Version 1.0, 2003. [4] T. Liu, Fast automated center-of-rotation determination in computed tomography technology, SIMTech Invention Disclosure, ID-P-07-013, 2007. [5] T. Liu, CT reconstruction for planar object and automated layer separation, SIMTech Invention Disclosure, ID-P-07-017, 2007. [6] T. Adams, Acoustic Micro Imaging of Stacked Die, Semiconductor International, http://www.semiconductor.net/article/ca604506.html, 2005. 7