ABB MEASUREMENT & ANALYTICS DATA SHEET. Testomat AW101 Water hardness monitor

Similar documents
AW101 Testomat Water hardness analyzer

8241 Silica Monitor. Cost-effective automated monitoring of silica in a wide range of applications

SM2000 Videographic recorder Spare parts and upgrade kits

SM1000 Videographic recorder Spare parts and upgrade kits

ABB MEASUREMENT & ANALYTICS DATA SHEET. Navigator 600 Phosphate Phosphate analyzer

ControlMaster CMF160 Universal process indicator, fieldmount

ControlMaster CMF160 Universal process indicator, fieldmount

ABB MEASUREMENT & ANALYTICS DATA SHEET. Navigator 600 Silica Silica analyzer

Navigator 600 Phosphate Phosphate analyzer. Cost-effective automated monitoring of phosphate for steam raising applications

C160 Wall-/Pipe Mounted Universal Process Indicator

TEPI11 P/I signal converter

Navigator 600 Silica Silica analyzer. Cost-effective automated monitoring of silica for a wide range of applications

ABB MEASUREMENT & ANALYTICS DATA SHEET. C1950 (STLR & HTST) Pasteurizer recorder and recorder / controller

ABB MEASUREMENT & ANALYTICS DATA SHEET. TB84PH ph /ORP / pion transmitter

Digital Indicator DS 48/24 E

C1950 (STLR & HTST) Pasteurizer Recorder and Recorder Controller

FICHA TÉCNICA DE PRODUTO

TB84TE Conductivity transmitter 2-Electrode. Superior accuracy in low conductivity and pure water. applications

Crane-Boy / Crane-BoyP Load limitation electronics

Control IT Booster Pump Controller

Manual Control Unit GFC 32

Silica Analyzer. Data Sheet. Navigator 600 Silica. Cost-effective automated monitoring of silica for a wide range of applications

Manual Simatek Control Unit GFC 32

ABB i-bus KNX HCC/S Heating/cooling circuit controller

ABB MEASUREMENT & ANALYTICS DATA SHEET. Navigator 500 Hydrazine analyzer

INDI BOY DSP Load limitation electronics

100mm Process Indicator Recorder

ControlMaster CM15 Universal process indicator, 1 /8 DIN. Making process control easy

Bus Connection Unit 560BCU04

RTU511. Binary Output 23BO62. Characteristics. Application. Data Sheet Binary Output 23BO62. The binary output is made via relay contacts.

ControlMaster CM10, CM30, CM50 and CM15 Universal process controllers and indicator Spare parts

Navigator 500 Sodium analyzer. Measurement made easy. Accurate and reliable measurement of sodium in high purity water

VeriMaster. Software tools. Measurement made easy. Overview

MagMaster Electromagnetic flowmeter. MagMaster to WaterMaster electronics upgrade. Introduction

/8 DIN Universal Process Indicator

ABB i-bus KNX BCI/S Boiler/Chiller Interface

FICHA TÉCNICA DE PRODUTO

ControlMaster CM10 Universal process controller, 1 /8 DIN. Making process control easy

Manual Control Unit GFC 16

Wall-/Pipe Mounted Universal Process Indicator

CoriolisMaster Accurate flow measurement of liquids and gases

Indicomp 4 Bargraph Indicator

FS-21.doc Status Filter Control FS-21. EasyRange

Sensyflow FMT200-ECO2 Thermal Mass Flowmeter

/8 DIN Motorized Valve Controller

FICHA TÉCNICA DE PRODUTO

CoriolisMaster mass flowmeter Automation solution for mixing applications

Adapter Plates for continuous Part-Turn Actuators (Contrac)

Spare Parts, Upgrade Kits and Service Kits for SM1000 and SM2000 Videographic Recorders

Thermal Mass Flowmeter Sensyflow FMT200-D

Magnetically Actuated Safety Interlock Switches

Introduction. For more information. HART Protocol 7.0 Valid for software levels from 05.00

Inform IT Single Pen Circular Chart Recorder

ABB MEASUREMENT & ANALYTICS DATA SHEET. IT1 Universal weighing terminal

I/O module (Expansion module)

SensyMaster High quality and cost effective solution for precise and dynamic direct mass flow measurement for gases

Process Photometer for total chlorine ppm with cleaning function

InCos-P Pressure sensor 20 Pa Pa

Messaging module CP-C MM

Voltage monitoring relays CM-ESS.1 For single-phase AC/DC voltages

RTU500 series Data Sheet Power Supply CP-E 24/2.5

LineMaster 100 Continuous-line Recorder

WaterMaster Select Electromagnetic flowmeter. The perfect fit for water industry applications

Sensyflow FMT200-ECO2 Thermal Mass Flowmeter

Automatic Insulating In-Line Oil Tester DTA 100 E-AD

FICHA TÉCNICA DE PRODUTO

Aztec 600 aluminium Aluminium analyzer. Reliable on-line monitoring of aluminium for potable water applications

Navigator 540. Measurement made easy. Introduction. For more information

Safety Light Grid Orion2 Extended

DIMENSION/WEIGHT kpa. 0 to 20 m / 0 to 60 ft 0 to 2 bar / 0 to 29 psi. 0 to 50 m / 0 to 160 ft 0 to 5 bar / 0 to 72.5 psi

The PM1000 series is a universal 4 digit LED plug-on display for transmitters with 4-20mA 2 wire output and fitted with DIN43650 connector.

RTU560. CT/VT Interface 560CVT10. Meter Operation. Application. Characteristics. Data Sheet CT/VT Interface 560CVT10. Instantaneous Meter Values

Operating Instructions VEGAMET 381

ABB MEASUREMENT & ANALYTICS DATA SHEET. ControlMaster CM10 Universal process controller, 1/8 DIN

FICHA TÉCNICA DE PRODUTO

Electronic timer CT-ERS.12 ON-delayed with 1 c/o (SPDT) contact

Manual Control Unit GFCD 16

SPARE PARTS LIST CP COMPRESSOR. Model CPD CPD 75

Gas Controller System DGC-06

ControlMaster CMF160 Universal process indicator, fieldmount

Control Panel CP600-eCo CP604, CP604-B

RTU560 Connections and Settings DIN Rail RTU 560CIG10

Immersion Temperature Sensors

7110 MTF-FG. Transmitter for Turbidity and Solids Content ISO Technical Information TI 232e

ATS-16 HV USER MANUAL. Automatic Transfer Switch 16A / 230Vac V090318

Pluggable interface relays CR-M Miniature relays

Pluggable interface relays CR-P Pcb relays

Instruction Manual. M Pump Motor Controller. For file reference, please record the following data:

Using the WirelessHART transmitter TTF300-W as repeater to extend the availability of a WirelessHART network

Operating Manual -ENGLISH- Page 1 KWG-ISO

Table of Contents. Introduction... 2 Description...2 Specifications...3 Installation.. 5 Calibration.. 6 General Maintenance...7

LME620-AI/-AN (Contrac) Electrical Linear Actuator

Temperature Switches

Supplementary operating instructions for your air curtain system Controller TMC 500

Aztec 600 iron Iron analyzer. Reliable on-line monitoring of iron for potable water applications

Switching relay CT-IRE with 1 c/o (SPDT) contact

RMZ782B RMZ783B. Extension Modules. Building Technologies HVAC Products. Synco 700. for heating controller RMH760B

Low Range Manganese Analyzer

AB Axis Industries. Use

DEEP SEA ELECTRONICS PLC DSE103 MK II Speed Switch Operators Manual

Transcription:

ABB MEASUREMENT & ANALYTICS DATA SHEET Testomat AW101 Water hardness monitor

TE STO MAT AW 101 WATER H ARDNE SS MONITOR DS/AW100-E N RE V. E Measurement made easy A unique water hardness monitoring system with superior performance Automatic on-line monitoring of residual / total water hardness ideal for water softening and potable water plants Menu-driven, programmable functions with clear LCD text display simple operation Programmable hardness units dh, f, ppm CaCO3 and mmol l-1 Configurable initiation of analysis automatic intervals (programmable 0 to 99 minutes) from external flow signal (turbine meter) external stop/start Remote diagnostic alarm dirty measuring chamber low reagent level 500 ml (1 pint) reagent storage bottle extended operation period Analog output 0 to 0mA and 4 to 0mA enables the use of process recorders for data recording Two fully adjustable limit contacts enables separate ranges of hardness levels to be monitored

TESTOMAT AW101 WATER HARDNESS MONITOR DS/AW100-EN REV. E 3 General The AW100 has been developed as a simple device to monitor the quality of water from softeners. Applications include laundries, utility boiler plants, soft drinks factories, brewing, food processing and potable water plants. It provides an alarm for high hardness that starts an automatic regeneration of the ion exchange beds used most commonly in these applications. The AW100 uses one of four reagents that changes from green to red at a predetermined water hardness level. The reagent is added to a known volume of sample via a small pump until a color change takes place. The volume of reagent required to bring about this reaction (monitored photoelectrically) indicates the level of hardness of the water sample. The AW100 also provides an analog signal, the range of which is determined by the choice of one of the four reagents available. Secure operation without supervision low-level reagent alarm In certain steam boiler installations, the possibility of an unsupervised boiler operating with a depleted supply of reagent would have safety implications; allowing untested sample to flow through the boiler system. To overcome this problem, the AW101 employs a programmable low-level reagent alarm operation function. This monitors the reagent constantly and an alarm is triggered if the available reagent quantity falls below the quantity required for 7 hours usage. Stop function The active analysis cycle can be interrupted by pressing the STOP / Standby key on the display pad. The cycle can be interrupted remotely / automatically using the stop relay. The digital display on the front of the monitor and the analog output show the value of the last analysis cycle. Three alarm outputs are provided two concentration alarms for control purposes and one alarm to drive an audible signal device. There are also inputs for a flow device and suspension of the flow cycle.

4 TESTOMAT AW101 WATER HARDNESS MONITOR DS/AW100-EN REV. E Mode of operation Analysis cycle Pause Flush 1 1 The flow cell is flushed thoroughly to ensure that an uncontaminated sample is analyzed. The flush time is programmable to suit the process sample and conditions. Fill chamber Drain chamber Fill chamber The measuring chamber is filled with a known sample volume. 3 The sample is checked optically to ensure it is clean. 4 A quantity of reagent is titrated and the end-point (color change) is reached. Analysis Drain chamber Pause 3 4 5 6 7 5 The result is evaluated and displayed. If the hardness value is outside the range determined by the reagent used, an alarm state is triggered. 3, 4 and 5 are known as the analysis period. 6 The chamber is drained. 7 There is a pause period (programmable) until the next analysis cycle is started.

TESTOMAT AW101 WATER HARDNESS MONITOR DS/AW100-EN REV. E 5 Reagents available for AW101 Parameter / type of reagent Units Water hardness AW101901 Water hardness AW10190 Water hardness AW101903 Water hardness AW101904 dh German (10 mg / CaO per 1000 ml water) 0.05 to 0.5 (0.1) 0.5 to.5 (0.05) 1.0 to 10.0 (0.).5 to 5.0 (0.) Range f French (10 mg / CaCO3 per 1000 ml water) ppm CaCO3 North America and UK (1 mg CaCO3 per 1000 ml water) 0.09 to 0.89 (0.0) 0.89 to 8.93 (0.) 0.45 to 4.48 (0.1) 4.47 to 44.7 (0.9) 1.79 to 17.9 (0.4) 17.9 to 179 (3.8) 4.48 to 44.8 (0.4) 44.7 to 447 (3.8) mmol/l Internationally recommended units (100 mg CaCO3 per 1000 ml water) 0.01 to 0.09 (0.01) 0.04 to 0.45 (0.01) 0.18 to 1.79 (0.04) 0.45 1 4.48 (0.04) Note. Figures in brackets show measurement resolution Display B C A D 1 0,1 o dh ^ 1:0,30 ^ :0,40 ANALYSE E F G H I testomat ECO ENTER M STANDBY I J L K A Status of limit value displays Displays the status of the limit values LV1 and LV. B Text display Displays the current analysis, all important status results and programming data in a -line LCD. C Alarm Displays a function fault. D Power switch The on / off switch is located on the right-hand side panel. E Unit fuse (inside the unit) Protects outputs against overload and short circuit. G Programming keys (cursorblock with ENTER) These keys are used to enter all values and programming data. Function keys H Manual manual start of an analysis. I STANDBY manual analysis stop / standby. J Alarm cancels alarm message. K I-Key Access all unit information. L M-key Access the programming menu. F Analysis message Displays current analysis.

6 TESTOMAT AW101 WATER HARDNESS MONITOR DS/AW100-EN REV. E Analog output Another possibility for monitoring the analysis is the connection to a process recorder or supervisory system. For this purpose the unit is equipped with a programmable current output. Output values of 0 to 0mA and 4 to 0mA can be selected for retransmission of the measured value. Example ma CaCO 3 0 179 Reagent AW101903, 0 to 0 ma 1 107.4 8 71.6 4 0 0 Typical recorder output

TESTOMAT AW101 WATER HARDNESS MONITOR DS/AW100-EN REV. E 7 Specification Display Ranges Determined by reagent (see table on page 5) Alarms Three relay outputs limit values (LV1, LV) and one fault alarm Resistive load 4 A Alarm displays The following faults are shown on the display and also activate the alarm output: Low water pressure Function fault optics Measuring fault analysis Function fault dosing pump Function fault outlet to drain Reagent low level Measuring fault dirtiness Measuring fault turbine meter Measuring range exceeded Front panel indicators Programme in operation Analysis stopped Lack of reagent Satisfactory result Unsatisfactory result Viewing window Outputs Current output 0 to 0 ma or 4 to 0 ma Max. load 500 Ω EMC Conformity EN50081-1, EN5008-, EN61010-1 Power supply Voltage 115, 30 or 4 V AC ±10 %, 50 / 60 Hz Power consumption 30 VA Unit protection 115 V, 30 V: T 0.1 A 4 V: T 1.0 A Environmental data Sample pressure 0.1 to 3 bar (1.5 to 45.5 psi) Sample temperature 10 to 40 C (41 to 104 F) max. Ambient temperature 10 to 45 C (41 to 14 F) Mechanical data Ingress protection IP65 Dimensions 380 x 459 x 80 mm (15 x 18.8 x 11 in.) Weight 9 kg (19.8 lbs.) Consumables Reagent consumption 0.07 ml (0.00013 pint) per test

8 TESTOMAT AW101 WATER HARDNESS MONITOR DS/AW100-EN REV. E Overall dimensions Dimensions in mm (in.) 380 (14.96) 80 (11.0) 66 (10.47) 459 (18.07) 44 (16.7) Typical installation diagram Control unit LV1 OUT 0 to 0 ma 4 to 0 ma PLC LV IN Fault alarm Flow meter F1 Raw water inlet Soft water outlet Testomat AW101 Softener Brine tank Monitor drain Treated only drain

TESTOMAT AW101 WATER HARDNESS MONITOR DS/AW100-EN REV. E 9 Terminal block identification No. Terminal Type Function Note PE IN Mains protective earth (5x) Earth / Ground 1 3 to 5 6 to 8 9 10 11 1 13 14 15 16 17 18 19 0 1 3 L N n l IN OUT Mains, L = Live Mains, N = Neutral Neutral, switched (8x) Live, switched (8x) LV1 OUT Limit value output 1 normally closed Limit value output 1 common Limit value output 1 normally open LV OUT Limit value output normally closed Limit value output common Limit value output normally open Alarm OUT Fault message normally closed Fault message common Fault message normally open Stop IN OUT + OUT IN IN External analysis stop Common earth for inputs Water meter input Common earth for inputs Mains input 115 V, 30 V or 4 V AC Mains voltage, max. 4 A Volt-free relay output, max. load 40 V AC, 4 A Volt-free relay output, max. load 40 V AC, 4 A Volt-free relay output, max. load 40 V AC, 4 A Only for volt-free normally open/normally closed contact Only for volt-free normally open/normally closed contact Note technical data of turbine OUT 0 or 4 to 0 ma Current output + (0 or 4 to 0 ma) 3 4 + OUT +1 V for hall-sensor (turbine) Note technical data of turbine Max. power input of sensor must not exceed 0 ma Terminal block for inputs Stop, IN and output OUT Terminal block for mains connection and relay outputs Stop IN OUT +1 18 19 0 1 3 4 n n n l l l PE PE PE L N + 1 9 10 11 1 13 14 15 16 17 K1 K K3 Terminal block labels

10 TESTOMAT AW101 WATER HARDNESS MONITOR DS/AW100-EN REV. E Ordering information Testomat water hardness monitor AW101/ X X X Version AW101 Water hardness 1 Mains supply 115 V AC 50 / 60Hz 30 V AC 50 / 60Hz 4 V AC 50 / 60Hz Menu language English German French 1 3 1 3 Spares Pressure regulator AW101601 AW10160 AW101603 AW101604 AW101605 AW101606 AW101607 AW101608 Regulator / filter housing Regulator plug T000, kpl Flow regulator valve Retaining pin for regulator plug Inlet filter Spring for inlet filter Inlet connector Plug-in connector Measuring chamber AW101611 Sight-glass window 30 x 3 with seal AW10161 Sight-glass window 30 x 3 AW101613 Sight-glass retaining disc AW101614 Screw spindle M3 x 40 AW101615 Latch fastener TL 800-7-1 AW101616 Plastic plug AW101617 Measuring chamber T000 Holding block for measuring chamber AW1016 Magnetic stirrer AW10163 Plug-in connector G3/8 in. AW10164 Solenoid valve / way AW10165 Rear guide bar for measuring chamber Dosing Pump DOSIClip AW101631 AW10163 AW101633 AW101634 AW101635 Jet pump complete Suction capillary Pressure capillary complete Base circuit board T1 complete Magnet (4 V DC) Bottle connection / suction tube AW101641 Screwed cap with bottle insert T000 AW10164 Screwed cap GL3 only AW101643 Bottle insert for screwed cap with push fit suction Electrical components AW101651 Fuse M4A AW101656 Cable sleeve 7 10 AW101657 Mains on / off switch AW101658 Cover for mains on / off switch AW101659 Multi-pin strap cable 10 pole with EMI filter clamp AW101660 Multi-pin strap cable 6 pole with EMI filter clamp AW101661 Cable loom V complete (for valves) AW101663 Cable loom for mains on / off switch complete AW101664 Fuse T0.16A AW101665 Fuse T1.0A Spare parts for -3 years operation AW101611 x sight-glass window 30 x 3 with seal AW101605 1 x Inlet filter (optional) AW101701 Gasket set T000 Number required subject to maintenance regime (see manual) AW101664 1 x fuse T0.16A AW101665 1 x fuse T1.0A Reagents AW101901 AW10190 AW101903 AW101904 TH005 water hardness 0.89 to 8.93 ppm CaCO3 TH05 water hardness 4.47 to 44.7 ppm CaCO3 TH100 water hardness 17.9 to 179 ppm CaCO3 TH50 water hardness 44.7 to 447 ppm CaCO3 Service

TE STO MAT AW 101 WAT ER H ARDNE SS MONITOR DS/AW100-E N RE V. E 11

ABB Limited Measurement & Analytics Oldends Lane, Stonehouse Gloucestershire, GL10 3TA UK Tel: +44 (0)1453 86 661 Fax: +44 (0)1453 89 671 ABB Inc. Measurement & Analytics 15 E. County Line Road Warminster, PA 18974 USA Tel: +1 15 674 6000 Fax: +1 15 674 7183 abb.com/measurement We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document. We reserve all rights in this document and in the subject matter and illustrations contained therein. Any reproduction, disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior written consent of ABB. Copyright 018 ABB. All rights reserved. 3KXA87510R1001 DS/AW100-EN Rev. E 03.018