Ultra-Accurate Measurement for Temperature and Voltage

Similar documents
Ultra-Accurate Measurement for Temperature and Voltage

Ultra-Accurate Temperature Measurement Instruments

Precision Voltage Measurement. Instruments. Applications. Voltage Inputs. High-Stability, Low Drift Voltage References. Key Design Features

Temperature Measurement Instruments Ultra-Accurate... by Design

DT8824 High Stability, High Accuracy, Ethernet Instrument Module

DT9828. Low-Cost, Isolated, Thermocouple Measurement Module. Overview. Key Features. Supported Operating Systems

Analog Input Sample Rate

DT Channel Analog Input USB DAQ Module. Key Features:

VIBBOX. 32, 48, or 64-Channel Sound & Vibration Solution Expansion to 256 Channels. Key Feature of Each VIBbox: Table 1. Key Features of VIBbox Models

VIBbox 64-Channel Sound & Vibration Solution

16-Bit A/D Converter (ENOB=14.1) 16-bit D/A. 16-bit D/A. Digital I/O (8/8) Counter/Timer (2)

DT9828. USB Powered Thermocouple Measurement Module. Key Features: Analog Input Channels

±500 V Isolation Protects Your Data. Sample Rate. 16 bits ±1.25 V, ±2.5 V, ±5 V, ±10 V 100 ks/s 16 2

DT MS/s High-Speed, Isolated Simultaneous USB Data Acquisition Module. Overview. Key Features. Bandwidth of the DT9862

Features Summary. Input Ranges

DT3016. High-Speed, Multifunction PCI Data Acquisition Board. Overview. Key Features. Supported Operating Systems

OBSOLETE. Isolated, Linearized, Thermocouple Input 3B47 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

DT9800 Series. USB Function Module for Data Acquisition

Isolated Voltage Input 3B30 / 3B31 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

DT9853/DT9854 Isolated, Low Cost D/A Modules for USB

Isolated, Process Current Output 7B39 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated Wideband Voltage Input 3B40 / 3B41 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated Process Current Output 3B39 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated, Field Configurable Analog Input 6B11 / 6B11HV FEATURES APPLICATIONS PRODUCT OVERVIEW

Isolated Process Current Input 7B32 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated Voltage Input 7B31 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated, Voltage or Current Input 7B30 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Wide Bandwidth Strain Gage Input 3B18 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated Linearized RTD Input 5B34 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated Process Current Input with Loop Power 7B35

WebDAQ 316. Internet Enabled Thermocouple Logger. Remote Configuration and Monitoring. Virtually Unlimited Storage. Integrated HW and SW

Isolated Current Output 5B39 FEATURES APPLICATIONS PRODUCT OVERVIEW

ECONseries Low Cost USB DAQ

A variety of ECONseries modules provide economical yet flexible solutions

Specifications

DT9837. High Performance USB Module for Sound & Vibration Analysis. DT9837 Features

ECONseries Low Cost USB DAQ

Part Numbers. SNAP-AIARMS-i SNAP-AIARMS-i-FM* SNAP-AIVRMS-i SNAP-AIVRMS-i-FM* SNAP-AIMA-i. SNAP-AIMA-iSRC SNAP-AIMA-iSRC-FM* SNAP-AIMA2-i.

The Tracker 220 Series A Complete Range of Universal Input Digital Panel Indicators for Temperature and Process Measurement

Isolated, Voltage or Current Input 7B41 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

SCM5B. SLX200 Data Acquisition System. isolynx. Data Acquisition Systems. Visit our website Features

TRACKER 220 SERIES. Digital Panel Indicators. A Complete Range of Universal Input Digital Panel Indicators for Temperature and Process Measurement

Operating Manual. USB-based High-Precision 8-channel Temperature Measurement Module MODEL NO

Personal Daqs USB Data Acquisition Modules Models /55 & /56

USB-TEMP-AI. Temperature and Voltage Measurement. User's Guide

UM D. SCPI Programmer s Manual for TEMPpoint

Temperature measurement board, optically isolated, 16/8/4 channels for thermocouples, Pt100, RTD, 18-bit

LAUREL. Laureate RTD Temperature Panel Meter / Controller ELECTRONICS, INC. Features. Description. Specifications

LAUREL ELECTRONICS, INC.

6220 Ethernet-Based Voltage Measurement Module

Detailed Specifications 6 1/2 Digits High-Performance DMM

Cryogenic Temperature Controller Model 24

C100 Portable Multifunction

TC Type Range Conformity Error. J -210 C to +760 C (-347 F to F) ±0.09 C (±0.16 F) K -244 C to C (-408 F to F) ±0.1 C (±0.

Signal Conditioning Modules and Terminal Boards

Trident and Trident X2 Digital Process and Temperature Panel Meter

Chapter 1 Introducing the OM-NET-TC... 6 Ethernet interface... 6 Functional block diagram... 6

A variety of ECONseries modules provide economical yet flexible solutions. Waveform Generation

CONTROL MICROSYSTEMS Thermocouple Analog Input Module. Hardware Manual

Fluke 8845A/8846A Digital Multimeters. Precision and versatility for bench or systems applications

Inform IT Single Pen Circular Chart Recorder

12-36 VDC/12-24 VAC Power Option 4-Digit Display, 0.56 (14.2 mm) or 1.20 (30.5 mm)

NI2400/NI816/NI4866 DATALOGGER NI2400/NI816/NI /16/24 CHANNEL ETHERNET DATA LOGGER with Embedded Web Server

Model 3724 Dual 1x30 FET Multiplexer Card

Agilent L4421A 40-Channel Armature Multiplexer

WHERE THE FUTURE MEETS THE PRESENT. Datascan. solo Network Based Data Acquisition System

Fluke 8845A/8846A Digital Multimeters. Precision and versatility for bench or systems applications

R1M-GH THERMOCOUPLE & DC INPUT MODULE MODEL. Remote I/O R1M Series. (16 points)

Fluke 8845A/8846A Digital Multimeters. Precision and versatility for bench or systems applications

High-Temp PRT Module 2561

USB Multi-sensor Measurement and Data Logger. User's Guide

100mm Process Indicator Recorder

CALYS 50. Field multifunction calibrator for basic use. Calibration measurement

StrainSmart Data Acquisition System

DI-924MB. Thermocouple/Millivolt Input Module

USB DIO High Voltage and Relay

High-Precision Data Acquisition

AMC 8 & IOC 8T. Analog Monitoring Card Type AMC 8 and Input / Output Card Type IOC 8T FEATURES

Agilent L4433A Dual/Quad 4x8 Reed Matrix

WLS-TC Specifications

AIM & THURLBY THANDAR INSTRUMENTS

Open. Powerful. Application Builder for Test & Measurement Systems.

AIM & THURLBY THANDAR INSTRUMENTS

Sorensen DLM 600 Series W V 2 75 A. Half Rack Programmable DC Power Supply

5B Series 16 Channel Backplane 5B01 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Sorensen XEL & New XEL-P Series

AIM & THURLBY THANDAR INSTRUMENTS

NEW. Programmable Resistance Decade Substituter. PRS-330 Page 1 of 5. IET LABS, INC. in the GenRad Tradition

DATAWATCH IX MONITOR/RECORDER OVERVIEW FEATURES AND BENEFITS. Datasheet

6220 Ethernet-Based Voltage Measurement Module

Safety Barriers. CEAG safety barriers can be used for all kinds of instrumentation applications.

AMC8 and IOC8T Analog monitoring card and input / output card

Measurement Systems Datascan Installation and User Guide

USB 1608G Series USB Multifunction Devices

1/32-DIN TEMPERATURE CONTROLLER INSTALLATION, WIRING AND OPERATION MANUAL FORM 3882

Techniques to Reduce Measurement Errors in Challenging Environments

RSTI-EP Slice I/O. Analog Input Modules `EP-3164, EP-3264, EP-3124, EP-3368, EP-3468 EP-3664, EP-3704, EP GFK-2960F September 2018

- SELF-TUNING PID ALGORITHM - INTUITIVE COLOR DISPLAY WITH TEXT MESSAGING - UNIVERSAL PROCESS AND TC/RTD INPUTS - MULTI- FUNCTION RAMP- DWELL/PROCESS

Sorensen DLM 600 Series W V 2 75 A. Half Rack Programmable DC Power Supply

DT3000 Series. DT3000 Series. Key Features. at full rated throughput.

Transcription:

Ultra-Accurate Measurement for Temperature and Voltage MEASURpoint is an ultra-accurate instrument for measuring any combination of thermocouple, RTD, and voltage inputs. MEASURpoint is available as a USB or Ethernet (LXI Class C compliant) instrument, and provides ±500V, ±1400V, or ±3500V isolation channel-tochannel and to earth ground. MEASURpoint includes the QuickDAQ application to configure and acquire temperature, resistance, and voltage channels, and to display, log, analyze, and export data to other formats including Excel. A ready-to-measure program for immediate productivity. DT8875 DT8874 DT9874 DT8876 MEASURpoint provides isolation to earth ground of up to ±3500V continuously or 5000V for transients, making it ideal for grid and alternative energy applications including: Gas turbines Wind turbines Battery storage Power load distribution and management Figure 1. MEASURpoint is an ultra-accurate instrument for measuring any combination of thermocouple, RTD, or voltage inputs with the QuickDAQ application. Available for USB or Ethernet, it provides up to 5000V isolation. MEASURpoint incorporates proprietary ISO-Channel technology that makes measurements almost indestructible and eliminates any common mode noise and ground loop problems under all environmental conditions. In addition, up to forty-eight configurable input channels (depending on the model) offer ultimate flexibility to the user. Isolation All systems have 10Hz/channel throughput rate and use a high stability 24-bit delta-sigma A/D per channel. Features Summary DT9874/DT8874 DT8875 DT8876 Working ±500V ±1400V ±3500V Transient (peak) 500V 2500V 5000V Ch-to-Ch 1000V 2800V 7000V Channels # of Channels (max) 48 40 20 Sensor Thermocouple* RTD** Voltage Accuracy ±75mV, ±1.25V, ±10V, ±100V, ±400V ±75mV, ±1.25V, ±10V, ±100V, ±600V ±75mV, ±100mV, ±1.0V, ±1.25V, ±10V Thermocouple ±0.16 C ±0.24 C ±0.32 C RTD (PT1000) ±0.003 C ±0.003 C ±0.003 C Voltage (10V range) ±0.3mV ±0.3mV ±0.3mV CJC Configuration CJC per ch CJC per ch *Type J, K, T, B, E, N, R, S, for all models. **Type PT100, PT500, PT1000. 2, 3, or 4-wire. CJC per 4 ch (embedded in block)

DT8874 DT8875 DT8876 Figure 2. MEASURpoint offers 3 different levels of isolation. The DT8874 input channel board offers the smallest size and easily accommodates ±500V isolation. The DT8875 offers wide spacing between channel elements to accommodate ±1400V isolation. The DT8876 offers the widest spacing to accommodate ±3500V isolation. Key Features: ISO-Channel technology provides galvanic isolation channel-to-channel and to earth ground: ±500V for the DT8874 ±1400V for the DT8875 ±3500V for the DT8876 Dedicated 24-bit: Delta-Sigma ADC/ch 8DI to 48DI simultaneous analog inputs Sampling rates: up to 10Hz/ch Software selectable ranges on voltage boards: DT8874: ±10V, ±100V, ±400V DT8875: ±10V, ±100V, ±600V DT8876: ±0.1V, ±1V, and ±10V Auto-calibrating front-end: resets the zero point on each power-up Measurement Instrument Calibration Utility: allows in-field calibration Digital I/O galvanically isolated to 250V: 8 in/8 out Packaging options: compact, rugged 2U, half-rack enclosure (DT8874) and industrial 5U, 19-inch rack enclosure (DT8875/DT8876) USB version of DT8874: DT9874 Thermocouple Input channel Features: Voltage input range: ±75mV Dedicated CJC (cold junction compensation) input for each thermocouple channel Supports B, E, J, K, N, R, S, and T thermocouple types Break-detection circuitry: detect open thermocouple inputs RTD Input channel Features: Voltage input range: ±1.25V 4-wire, 3-wire, or 2-wire PT100, PT500, and PT1000 RTDs supported MEASURpoint Software Features: QuickDAQ Acquire and display from all Data Translation USB and Ethernet devices that support analog input streaming. Acquire data, record data to disk, display the results in both a plot and digital display, and read a recorded data file. Two additional options can be purchased to add FFT analysis capabilities to the base package. Also, data can be exported to other applications like Microsoft Excel and MATLAB for more advanced analysis IVI-COM driver works in any development environment that supports COM programming, including Measure Foundry, Visual Basic, Visual C#.NET, MATLAB, VEE Pro, LabVIEW or LabWindow, and others SCPI support: SCPI (Standard Commands for Programmable Instruments) is a universal programming language for electronic test and measurement instruments, based on the IEEE 488.1 and IEEE 488.2 standards Analog Input Flexibility The standard MEASURpoint instrument provides up to 48 configurable channels (depending on the model), allowing for ultimate flexibility with thermocouple, RTD, and voltage inputs. Because MEASURpoint architecture uses an A/D per channel, sampling rates of up to 10Hz per channel simultaneously can be achieved. The RTD input channels provide a 4-wire RTD input with Kelvin sensing for maximum accuracy by eliminating errors due to wire resistance. You can attach a voltage input or any of the following RTD types to these channels in a mix and match fashion: Platinum 100 Ω (Pt100), Platinum 500 Ω (Pt500), or Platinum 1000 Ω (Pt1000) RTD using an European alpha curve of 0.00385 or an American alpha curve of 0.00392. The supported temperature measurement range for these RTD types is 200 C ( 328 F) to 850 C (1562 F). You can also measure a resistance value, in Ohms, if desired. The analog input range is ±1.25V. CJC Circuit Thermocouples are relative not absolute temperature measuring devices that generate voltage as a function of the temperature difference between both ends. The DT9874, DT8874, and DT8875 incorporate an independent CJC circuit for every channel. The DT8876 incorporates a CJC for each bank of 4 channels. The board-level CJC is embedded in a metal bar for excellent thermal conductivity and optimum consistency. 2

Galvanic Isolation with ISO-Channel ISO-Channel uses galvanic isolation methods to guarantee up to ±3500V isolation (depending on the model) between any input channel to any other input channel and earth ground. Common-mode noise and ground loop problems are eliminated with ISO-Channel since sensors that are at different ground reference levels are easily accommodated, even if they are at widely differing voltages of hundreds of volts or transients to thousands of volts. The result is that accuracy is preserved for all sensor inputs. This is especially useful when conditions change in the electrical environmental due to motor current surges, electromagnetic radiation, or noisy industrial equipment turning on/off. A vast majority of thermocouple applications reside in industrial environments. ISO- Channel technology makes measurements almost indestructible. Triggers A trigger is an event that occurs based on a specified set of conditions. Acquisition starts when the instrument detects the initial trigger event and stops when the buffer has been filled or you stop the operation. MEASURpoint instruments support a software trigger and an external trigger on digital input line 0. Digital Input/Output Lines MEASURpoint instruments feature eight, isolated, digital input lines. The digital input lines operate from +3 to +28V DC, with a switching time of 2ms maximum. MEASURpoint instruments are perfect for driving relays directly, featuring eight, isolated, digital output lines. The outputs are solid-state relays that operate at ±30V and 400mA peak (AC or DC) with a switching time of 2ms maximum. + - CJC* +10 na Break Detection +425 µa Current Source 2.5 Hz Filter*** Sense Return x20 24-Bit A/D Thermocouple Channel 24-Bit A/D RTD Channel Up to ±3500V Isolation Barrier Control FPGA 64 kb SRAM Embedded Controller 1 of 8 Digital Input Isolators 1 of 8 Digital Output Isolators Ethernet 37-Pin D-Shell Connector LAN ENet Link ENet Activity LAN Power USB ** Limit + 2.5 Hz Filter*** Open TC ** 1Ω 1MΩ 24-Bit A/D ID ROM - Sense Return Voltage Channel Calibration ROM *CJC per channel on DT9874 and DT8874. CJC every 8 channels on DT8875 and every 4 channels for DT8876. **Not used on LXI instruments. ***DT8875 and DT8876 use 4.5 Hz filter. Figure 3. Block diagram. 3

Channel 1 Floating Front End: ISO-Channel CH 1 1MΩ Filter Diff Amp 24-bit Sigma- Delta A/D Channel Ground for Sensor 1 Channel 48 Floating Front End: ISO-Channel Up to ±3500V Isolation ROM SRAM Embedded Controller 8 Digital In 8 Digital Out Ethernet (LXI) Up to CH 48 1MΩ Filter Diff Amp 24-bit Sigma- Delta A/D USB Channel Ground for Sensor 48 Figure 4. An A/D per channel and a converter for each A/D provides channel-to-channel isolation, where each signal can float to its own ground reference. High-Stability, Low Drift Voltage References MEASURpoint uses high-precision, high-stability, low-drift voltage references rated at 4 PPM per degree and 100 PPM drift per year. This means MEASURpoint is accurate now and will remain that way over time. Custom Designed Converters Our custom converters circuits have a unique power distribution system that supplies power to only 2 of the 6 boards at any one time. Cycling non-adjacent boards in this manner creates less power surges, reduces noise, and improves the overall system performance. Field Calibration Users can calibrate any MEASURpoint instrument in the field using precise calibration equipment and the Measurement Instrument Calibration Utility. This utility allows you to revert to the factory calibration for any or all channels, or revert back to the last user calibration values, if desired. In addition, this utility generates a report that lists the starting and ending calibration values for each channel, allowing traceability. Remote Measurements The network-ready versions of MEASURpoint are LXI Class C compliant and provide a standard Ethernet connection to support remote monitoring and control from the field or on the factory floor. Channels can be expanded by simply adding more instruments to the network. 4

QuickDAQ QuickDAQ allows you to acquire and display from all Data Translation USB and Ethernet data acquisition devices that support analog input streaming. Combine QuickDAQ with Data Translation hardware to acquire data, record data to disk, display the results in both a plot and digital display, and read a recorded data file. Be productive right out of the box with this powerful data logging software. Data can be exported to other applications like Microsoft Excel and The Mathworks MATLAB for more advanced analysis. Two additional options can be purchased to add FFT analysis capabilities to the base package. Key Features: QuickDAQ Base Package (Free) Ready-to-measure application software Configure, acquire, log, display, and analyze your data Customize many aspects of the acquisition, display, and recording functions to suit your needs FFT Analysis Option (License Required) Includes all the features of the QuickDAQ Base Package Perform single-channel FFT operations Configure and view dynamic performance statistics Supports Hanning, Hamming, Bartlett, Blackman, Blackman Harris, and Flat Top response windows Advanced FFT Analysis Option (License Required) Includes all the features of the QuickDAQ Base Package and FFT Analysis Package Perform 2-channel FFT operations Supports real, imaginary, and Nyquist display functions Additional FFT analysis functions supported: Exponential, Force, Cosiner Taper Save data to.uff file format Figure 6. The main web page displays information about your instrument on the network. Figure 5. QuickDAQ ships free-of-charge and allows you to get up and running quickly. Additional Software Support The following software support is available for all MEASURpoint instruments: Measurement Instrument Calibration Utility The calibration utility allows you to calibrate a VOLTpoint, TEMPpoint, or MEASURpoint instrument in the field. Using this utility, you can revert to the factory calibration for any or all channels, or revert back to the last user calibration values, if desired. In addition, this utility generates a report that lists the starting and ending calibration values for each channel, allowing traceability. Eureka Discovery Utility This utility helps you locate or discover all LXI (Ethernet) instruments that are connected to your system and provides the following information about your instrument: the IP address, manufacturer, model number, serial number, and version of the firmware that is running on your instrument. In addition, you can use this utility to configure Windows firewall settings and update the firmware for your Data Translation LXI instrument. Instrument Web Interface This built-in interface allows you to verify the operation of your instrument and perform basic functions with Internet Explorer and no additional software. Using it, you can configure your instrument, control output signals, measure input signals, and save results to disk. IVI-COM Driver This driver is provided to write application programs for MEASURpoint using an IVI-COM instrument interface. It can be used with programs written in Visual C#, Visual Basic for.net, or C++ under Visual Studio 2003 to 2012. You can also use the IVI-COM driver with LabVIEW from National Instruments or MATLAB and the Instrument Control Toolbox from the MathWorks to program MEASURpoint instruments. SCPI Commands Use VISA or network sockets to program and control MEASURpoint LXI instruments by sending SCPI commands. Comprehensive user manual and example programs provided. 5

Ordering Summary MEASURpoint Instruments DTx87x xx -xx -xx 4 = ±500V (DT9874, DT8874) 5 = ±1400V (DT8875) 6 = ±3500V (DT8876) 8 = Ethernet 9 = USB (available with ±500V only) = Thermocouple Channels = RTD Channels = Voltage Channels DT9874, DT8874, DT8875 00 = No Channels 08 = 8 Channels 16 = 16 Channels 24 = 24 Channels 32 = 32 Channels 40 = 40 Channels 48 = 48 Channels (DT9874 and DT8874 only) DT8876 00 = No Channels 04 = 4 Channels 08 = 8 Channels 12 = 12 Channels 16 = 16 Channels 20 = 20 Channels Ordering Example DT9874-16T-16R-16V MEASURpoint USB instrument configured with 16 thermocouple channels, 16 RTD channels, and 16 voltage channels. For additional channel configurations, please call 1-800-525-8528 or email info@datatranslation.com to discuss your requirements. Options STP37 Digital I/O screw terminal panel EP333 Cable for attaching the STP37 to the MEASURpoint instrument EP373 Single Rack-Mount Kit for DT9874/DT8874 EP374 Dual Rack-Mount Kit for DT9874/DT8874 EP395 Surface mount Kit for DT8875/DT8876 (mounts DT8875/DT8876 to a surface, such as a table). EP396 Rack-Mount Kit for DT8875/DT8876 Enclosure Options MEASURpoint Instruments with Rack Mount Kit Rugged NEMA enclosures. Copyright 2013 Data Translation, Inc. All rights reserved. All trademarks are the property of their respective holders. Prices, availability, and specifications are subject to change without notice.