Future Topics. Projection Imaging Dose Reporting, XA 3D Volume Objects. for DICOM WG-02. Presented by Heinz Blendinger, Siemens Medical Solutions

Similar documents
DICOM Enhanced XA/XRF Object New Dimensions for X-Ray Projection Imaging. (DICOM Supplement 83)

Digital Imaging and Communications in Medicine (DICOM) Supplement 167: X-Ray 3D Angiographic IOD Informative Annex

Kevin O Donnell Senior R&D Manager Toshiba Medical Research Institute Co-Chair, DICOM Standards Committee Past Chair, IHE Radiology Planning

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL ADR-1000A /R2 (MIIXR0003EAE)

Digital Imaging and Communications in Medicine (DICOM) Supplement 116: 3D X-Ray Storage SOP Class

Sep, th Edition 897N101668H

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL

DICOM Conformance Statement

DICOM CONFORMANCE STATEMENT MEDIA STORAGE FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A. with XIDF-037A and XIDF-038A or SRS-1000A

Whole Body X-ray CT System Supria. DICOM Conformance Statement

Copyright FUJIFILM Corporation, Japan. August, th Edition 897N100760F

INTERNATIONAL STANDARD

Technical Publications

Technical Publications

Technical Publications

Technical Publications

xorantech.com Suite of DR Products

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

DICOM Conformance Statement for Selenia Dimensions Acquisition Workstation Software Version 1.8 Part Number MAN Revision 002

Technical Publications

Technical Publications

Technical Publications

Fluoroscan InSight V5.0.6 DICOM Conformance Statement

Technical Publications

DICOM CONFORMANCE STATEMENT

Technical Publications

DICOM CONFORMANCE STATEMENT STORAGE SCU FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A. with XIDF-053A or XIDF-056A (MIIXR0001EAB)

DICOM Overview: Stability and Evolution

DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300

DICOM Conformance Statement for FLEXAVISION HB/FD

Interoperability Issues in Image Registration and ROI Generation

DX-D 300. Flexible Direct Radiography System

PLANMECA PROMAX 3D MID CBCT UNIT

Digital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities

IHE Radiation Oncology Technical Framework Supplement. Treatment Delivery Plan Content (TDPC) Rev. 1.1 Trial Implementation

Technical Publications

Radiation Oncology treatment facility design. Simulators, CT scanners, HDR Brachytherapy

Carestream OnSight 3D Extremity System ImageView Software Version 1.0. DICOM Conformance Statement

DX-D 800 DX-D 800 DYNAMIC, 3-IN-1 DIRECT RADIOGRAPHY SYSTEM OFFERING REAL TIME IMAGES FOR FLUOROSCOPY, GENERAL RADIOGRAPHY AND DIRECT EXPOSURES.

Whole Body X-ray CT System Supria. DICOM Conformance Statement

DICOM Conformance Statement

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

Technical Publications

VIVIX Software.

DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL HDR-08A

DAR-8000 Series DICOM Conformance Statement

No. MIIMS0002EAG TOSHIBA MEDICAL SYSTEMS CORPORATION ALL RIGHTS RESERVED

FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300

DX-D 300. Flexible Direct Radiography System

DICOM Conformance Statement for DR-200/DRU-200

DICOM Conformance Statement

EOS. EOS imaging system. DICOM Conformance Statement [1]

Technical Publications

NMKOMMT. pfaf`lj =tip. af`lj=`çåñçêã~ååé=pí~íéãéåí= båöäáëü. spkn. oéäé~ëéç=nrlnolomnr. Ñçê=éêçÇìÅí=îÉêëáçå=Ñêçã=PKN=çåï~êÇë

DAR-9400f/DAR-9500f with Gateway Option DICOM Conformance Statement. Revision D

Caseaccess 1.0 DICOM Conformance Statement

C-View 1.0. DICOM Conformance Statement. MAN Rev 001

O-ARM IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE

Digital Innovation for Medical Equipment

DICOM Conformance Statement

Planmeca ProMax 3D s CBCT unit

Technical Publications

REGIUS CONSOLE CS-2. Revision History. Date Version Description 02/08/2004 Ver This Conformance Statement applies to DICOM3.0 (2000 version).

PrecisePLAN 2.15 Export

DIRECT RADIOGRAPHY. flexible. Delivering. imaging excellence. DX-D Direct Radiography from Agfa HealthCare

Preparing for Part 2 of the ABR Diagnostic Physics Exam

CMT MEDICAL TECHNOLOGIES

Technical Publications

DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 DR SYSTEM

DigiRoebs 3.1. DICOM Conformance Statement V1.0. medigration GmbH. DICOM Conformace Statement Seite 1 / 38.

Digital Imaging and Communications in Medicine (DICOM) Supplement 176: Second Generation Radiotherapy. Additional RT Treatment Modalities

Planmeca ProMax 3D Max CBCT unit

Beschreibung Gesamt / Total Description

DIRECT RADIOGRAPHY. automated. Delivering the fully. digital X-ray room. DX-D 600: Direct Radiography from Agfa HealthCare

Gradient-Based Differential Approach for Patient Motion Compensation in 2D/3D Overlay

The Role of DICOM for Modular Perioperative Surgical Assist Systems Design

The smaller, simpler, safer alternative to radioisotope irradiators. Precision s MultiRad160 is ideal for irradiating larger cell cultures and/or

Measurement of Skin Dose

Dx Server for Windows NT DICOM 3.0 Conformance Statement

Introduction to Biomedical Imaging

Technical Publications

Dx Server for Windows NT DICOM 3.0 Conformance Statement

Senographe Essential Acquisition System

Carestream DRX-Revolution Software Version 5.6. DICOM Conformance Statement

Villa Sistemi Medicali DRF 4343

Technical Publications

DAR-9400f / DAR-9500f DICOM Conformance Statement Revision J

The digital EVOlution in cassette size format

SonicDICOM PACS. Install Manual. December 14, JIUN Corporation. All rights reserved.

Radiation Dose Structured Reports

1.1. FireCR Calibration

DICOM CONFORMANCE STATEMENT. FOR CTport FOR TOSHIBA WHOLE-BODY X-RAY CT SCANNER A-SERIES MODEL CGPS-007A, CGPS-008A CGPM-009A, CGPM-010A CGPM-012A

DICOM CONFORMANCE STATEMENT. BrainLAB PatientBrowser 1.0

IHE Radiation Oncology Technical Framework Supplement. Treatment Delivery Record Content (TDRC) Revision 1.0 Draft for Public Comment

Raffaello Conformance Statement for DICOM V3.0

AGFA HEALTHCARE DICOM Conformance Statement

IHE Endoscopy Technical Framework Supplement. Endoscopy Image Archiving (EIA) Rev. 1.1 Trial Implementation

Version 7 DICOM Conformance Statement. Document Version 3.02, June 2009

IHE Radiation Oncology Technical Framework Supplement. Treatment Delivery Workflow (TDW) Draft for Public Comment

Technical Publications

Transcription:

Future Topics for Projection Imaging Dose Reporting, XA 3D Volume Objects DICOM WG-02 Presented by Heinz Blendinger, Siemens Medical Solutions 1

Presentation outline Dose Reporting Why Dose Reporting? User Scenarios (as of today) Solution outline Next steps XA 3D Volume Object Issues to be solved Solution outline Scope of Work Invitation 2

Why Dose Reporting? MPPS solution delivers Dose Data into an environment where further evaluation is not typically expected. Initiated from IEC Flexible and equipment-dependent structure for dose information needed (Projection X-Ray, CT, Mammography). Close relationship to images generated by irradiation shall be kept, but application of dose shall be recorded independently from image storage. Retrieval and Data Mining of dose information shall be possible in future. Support new applications related to dose evaluation and reporting. Transposing of dose information into other presentations (graphical, new algorithms) shall be possible. Establish a clear responsibility of and capability for X-Ray equipment to store dose-related information. 3

User scenarios (as of Supplement 94) The solution for Dose Reporting shall be scalable from simple Radiography to complex interventional X-Ray procedures. Dose Reports shall be interchangeable between different devices Acquisition Systems, Display Stations, Dose Report Stations, Radiation Safety Station, PACS. The possibility to create secondary reports to summarize dose with extended scope (e.g. visit, certain time period) shall be provided. 4

Scenario Basic Dose Reporting User Acquisition Modality Long-term Storage Display Station Set up acquisition Perform acquisition Create Dose Report Verify, add comments to Dose Report Send images and Dose Report Retrieve Dose Report 5

Scenario - Dose Reporting for Non-Digital Imaging User Acquisition Modality Dose Report Station Long-term Storage Set up acquisition Perform acquisition Enter Dose Information Create Dose Report Send Dose Report 6

Scenario Dose Reporting Post-Processing Acquisition Modality or Dose Report Station Long-term Storage Radiation Safety Station Send Dose Report Retrieve Dose Reports Send Regulatory Dose Report Store Cumulative Dose Report Retrieve Referenced Images Store Dose Map Create Dose Map 7

Print a dose report Patient Position: HFS 02-Mar-05 13:34:47 1 CARD FIXED Auto 2s 60F/s 02-Mar-05 13:36:08 A 62kV 99mA 3.6ms 0.0CL small 0.9Cu 23cm 0.7µGym² 0.1mGy 28LAO 12CRA 127F 2 CARD FIXED Auto 3s 60F/s 02-Mar-05 13:36:18 A 62kV 99mA 3.6ms 0.0CL small 0.9Cu 23cm 0.8µGym² 0.1mGy 0LAO 0CRA 157F Patient Position: HFS 02-Mar-05 13:36:48 ***Accumulated exposure data*** 02-Mar-05 13:36:49 Phys: TBD Exposures: 2 Fluoro: 0.3min Total: 2.5µGym² 0.3mGy A Fluoro: 0.3min 1.0µGym² 0.1mGy Total: 2.5µGym² 0.3mGy B Fluoro: 0.0min 0.0µGym² 0.0mGy Total: 0.0µGym² 0.0mGy ======================================================================================= Example with irradiation events and accumulated values section 8

Solution outline Projection X-Ray Structure to encode accumulated values as well as individual irradiation event data in one object is needed. Relationship to patient, study, procedure shall be provided. Possibility to create an additional dose report summary derived from basic reports shall be available Supplement 94 defines open solution to be extended for equipment dose reports not yet specified Existing DICOM services for workflow management and information retrieval shall be applicable to the solution. 9

Solution outline X-Ray accumulated data Equipment identification and calibration information Scope of report and reference information Accumulated values, one set for each plane (X-Ray source) Dose Area Product (grand total, total for fluoroscopy and acquisition modes) Dose at Reference Point (RP) (grand total, total for fluoroscopy and acquisition modes). Grand total of radiographic frames in report scope Total time of fluoroscopy 10

Solution outline X-Ray irradiation event data General Event information (method, plane, type) Link to acquired image(s), if any Dose at RP, Dose Area Product Acquisition parameters (e.g. kv, ma, filtration), some with per frame encoding option. Geometry information (e.g. C-arm/Table Angulations, Source to Distances, Collimated Area, Patient Orientation/Position) Timing of X-Ray application Event Unique Identifier 11

Next Steps Proceed Supplement 94 Diagnostic X-Ray Radiation Dose Reporting to Final Text. Extent the Standard with CT and Mammography equipment by defining the related Dose Report Templates. Foster implementations to gain experience in features connected with Dose Reporting (e.g. Features of Radiation Safety Station) Apply results from practical experience to extent existing Standard according to further needs. 12

XA 3D Volume Object Problems to be solved Solution outline Scope of Work Invitation 13

XA 3D Volume Objects Problems to be solved Volume Objects encoding Make 3D reconstruction from 2D projection imaging an integral part of XA procedure, no extra CT modality. Use new enhanced encoding schemes where applicable Provide concept for 3D (color) overlay. Volume content Define relation between volume reference system and C-arm isocenter. Preserve geometrical data and other specifics of the 2D acquisition context Volume application Align with future 3D Presentation State, Segmentation IOD (Supplement 111), multi-dimensional image registration. Provide concept for 4D viewing 3D multi-volume sets. 14

From 2D rotational angiograms 15

.. to XA 3D volumes 16

XA 3D Solution outline Think of one Storage Class to store Pixel data volume ( XA volume ) another Storage Class to store specific XA 3D results (e.g. segmentations, marks, meshes, surfaces 3D annotations/measurements ) that can be combined with a future 3D Presentation State solution 17

XA 3D Scope of work Enhanced XA Volume Pixel Object (NEW, derived from enhanced XA SOP Class) Scope of work Future links Future: 3D Pres State The objects will relate to each other. XA 3D object is entry point for XA 3D viewing. NEW XA 3D Object (Segmentation, 3D graphics, surfaces,...) 18

Invitation to WG-02 In case you are interested in XA 3D work-item, please contact WG-02 chairman or visit our next meetings. Contact WG-02 chairman Rainer.Thieme@siemens.com Next meetings: October 27 th and 28 th, 2005 in Washington December 12 th to 14 th, 2005 at DIN (Berlin) 19