MANUAL. Handgelenksband WST 100. Handgelenksband- und Schuhtester. performance test for Electrostatic Wrist Straps and Conductive Shoes

Similar documents
Wrist Strap & Shoe Tester WST 100

PGT 120. Personnel Grounding Tester. Wrist Strap and Footwear Tester. Users's Manual

ELECTROSTATIC FIELDMETER

RangerBOSS Network Ready Constant Monitor Model CM2800

Zero Volt Monitor Installation, Operation and Maintenance

Product Overview. System Overview

TEST AND MEASUREMENT SRM 200 SRM 200/EFM 51 STARTERKIT

Operating Instructions

Electrostatic Field Meter EFM 022

Operating Instructions

MANUAL. Electrostatic Field Meter EFM 022. System Prof. Dr.-Ing. Hans Kleinwächter

IntelliVue MP20/30. Hardware Upgrade Installation Guide. IntelliVue Patient Monitor MP20/30. Patient Monitoring M B *M B*

T R A N S F O R M I N G T E C H N O L O G I E S, L L C O U T S T A N D I N G A L T E R N A T I V E S I N S T A T I C C O N T R O L

Ohm Metrics. Instruction Manual. Dual Footwear Wrist Strap Tester Model PDT800K

All-Spec Industries Ph: Web: Wilmington, NC Fx:

WALKING TEST KIT WT5000

T R A N S F O R M I N G T E C H N O L O G I E S, L L C O U T S T A N D I N G A L T E R N A T I V E S I N S T A T I C C O N T R O L

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

Ground Master Monitor Installation, Operation and Maintenance

Copra Moisture Tester. Operating Manual

User's Guide. MiniTec TM Series Model MN25 MultiMeter

IMPORTANT SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS

Table of Contents. 1.Introduction... P.2. 2.Product Tour... P Operating Instructions... P Important Notes... P.10

INSTALLATION INSTRUCTIONS

Mini Digital Multimeter

Operating Manual Quattrino RE529, two tone

GSM HOME SECURITY SYSTEM

OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS

Model DT-315P DIGITAL STROBOSCOPE INSTRUCTION MANUAL

IDEAL INDUSTRIES INC. TECHNICAL MANUAL MODEL:

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: 57314

ReferenceMate portable reference source REF2500 operation manual

Part No. Z , IA Jul OPERATION MANUAL. High Voltage Digitalmeter A

DPL4000 Portable Low Range Dew Point Analyzer

TG basic 1. Appliance tester for testing protective measures in accordance with DIN VDE (DIN EN 62638)

Autoranging True RMS Multimeter User Manual

Battery Powered Portable Appliance Tester with Display & Memory User Manual Issue 2.1

Operating Instructions

Digital Keypad Introduction

Mini Digital Multimeter

Release Notes. Electrostatic Discharge and Grounding Guide for Cisco NCS 2000 Series. Prerequisites. Preparing Your Location

User's Guide. Digital Multimeter. Model MN42

CANcheck. Installation Tester for CAN/CANopen Networks MANUAL ENGLISH

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

LSL250. Laser light! Do not stare into the beam or view directly with magnifiers. Class 2 laser EN :

Part No. Z , IA Mar OPERATION MANUAL. High Voltage Digitalmeter A

Thank you very much for purchasing LAISAI LS686 line laser, please read this instruction manual before operating it. Content

3M Constant Monitors WSMONITOR1, WSMONITOR 2, WSMONITOR3

Mini Digital Multimeter

POCKET MULTIMETER Model No: MM18

3M Wrist Strap and Ground Monitor 773. User s Guide

2 in1 LAN Tester & Multimeter. Model: PCE-LT 1

Digital Power Probes, ACL120, ACL220, ACL Internal, and DCL5 (#30015, #30023, #30024, #30025)

HI-POT TESTER User Manual

2 in 1 LAN Tester and Multimeter Model:

E102 - Advanced Strain Gauge Transducer Display Interface. Contents

All in One Multi-purpose Cable Tester and Cable Tracer

Micro-Ohmmeters. Digital 10A. Models 6240 & Measure low resistance with high accuracy. Selectable test current and resistance ranges

BC-9000 Field Calibration Procedure CF1_FIELDCAL_BC9000 Revised 03/16/2013

CHAPTER 2: USING MANAGER PREFERENCE. 2.1 Using Manager Mode Using CSAFE Console (back side) Access Layout...

Model HM-535 Power Supply Installation and Service Instructions

PIECAL 520B & 521B Thermocouple Source Operating Instructions

Wireless Digital Video Doorbell

DPM-3221 Digital DC Power Meter with data logging capability.

NMID. Remote control. SAPIM I/O number. SAPIM I/O number NBRN (RS485) ZM100. Relay, 24 VDC - + UO07 UI11 UI10 UI09 UI08 UI07 UI06 UI05 UI04 UI03 UI02

Installing a New Solo 1150 Hard Drive

DPM Digital DC Power Meter with Data logging capability. User Manual

HOBO State Data Logger (UX90-001x) Manual

Replacing the Gateway M405 Keyboard

Junos WebApp Secure 5.0 Hardware Guide

TTC Series Torque Tool Tester Operation Manual

User Guide True RMS Multimeter Extech EX205T

FG-7000L Digital Force Gauge Operation Manual


SHIMPO INSTRUMENTS. FG-7000T Digital Torque Gauge Operation Manual

DPM Bi-directional Digital DC Power Meter with built-in USB data logger & adapter External shunt model. User Manual

FG-7000 Digital Force Gauge Operation Manual

2.4GHz Wireless Handheld Microphone System & 2.4GHz Wireless Belt Pack Microphone System

U P S USER S MANUAL IMPORTANT SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS. Uninterruptible Power System. Line Interactive (Network) UPS

Operating Manual RE429NT+ Quattrino tone, eco, voice, memo

DX4085 Z-Meter Tester

OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER

User's Manual. Autoranging Datalogging Insulation Tester (Megohmmeter) with PC Interface. Model

BENCH-TOP DIGITAL MULTIMETER INSTRUCTION MANUAL

USER MANUAL. Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX310

ESD Monitoring, Ionization & Test. SmartLog V5. ESD Data Acquisition System. Rev Made in United States of America.

genesis TECHNICAL MANUAL Two-Door Controller GEN-045

DLA. DMX512 Analyzer. DLA Users Manual SV2_00 B.lwp copyright ELM Video Technology, Inc.

MCCB-250 MOLDED-CASE CIRCUIT BREAKER TESTER

Part No. Z , IA Nov OPERATION MANUAL. High Voltage Digitalmeter A

OPERATING INSTRUCTION

SOUND LEVEL METER OPERATION MANUAL. Model : SL Your purchase of this

Two Door Controller GEN-045

SoundwebTM. Installation Guide

SpectraPulse Ultra Wideband Wireless Microphone System

BSL Hardware Guide for MP35 and MP30-1 -

Nvidia Quadro FX 5600 graphics card and auxiliary power adapter cable installation

Operating Instructions PRO-AB /6.15. Leakage Current Measuring Adapter

SIMATIC S7-300 Programmable Controller Getting Started Edition 04/2000 First Steps for Installation and Commissioning Totally Integrated Automation

MONARCH INSTRUMENT. TACH-4A / TACH-4AR Programmable Optical Tachometer. Instruction Manual Rev 2.0

Transcription:

MANUAL Handgelenksband- und Schuhtester Handgelenksband WST 100 performance test for Electrostatic Wrist Straps and Conductive Shoes S

2 Handbuch WST 100 Contents Description... 2 Specification... 2 Function... 3 Wrist strap test... 4 Shoe test... 4 Adjustable abilities... 5 Statistisical error memory... 5 Limits... 5 Change Battery... 5 Connection... 6 Shoe electrode connections... 6 Connecting a door-opener... 6 Scope of supply WST 100 SET... 7 Optional accessories... 7 Calibration... 8 That offers great advantages:... 8 Adjust the Limits.....9 Error... 10 Short Instruction... 12 Description The WST100 serves the function test of electrostatic wrist straps and conductive footwear. Before you get access to ESD-secured areas the ESD precautions like wrist strap and conductive footwear should be tested for correct functioning. The WST-100 is a reliable and advantageous solution for this. The appliance is portable and because of its battery power employable everywhere. The setting of the limits and the calibration can be done by the customer himself, very simply. A special feature is the appliance possesses a statistical error memory, in which the errors of statistically appraised are stored. Hand appliance in a plastic housing 150mm x 88mm complete µc controlled. The shoe electrode for this unit is 370mm x 440mm of a plastic plate which includes 2 shoe electrodes for the left and right shoe separately. The display includes 7 LEDs a green LED (8mm) for o.k. and 6 red LEDs for error measage For the additional acoustic signal a buzzer is integrated. The appliance possesses an integrated relay 60V / 0,5A (in the shoe electrode). With the optional plug charger and NiMH accumulators, the appliance can be run permanently.

3 Specification Dimension (L x B x H): Hand appliance: 150mm x 88mm x 35mm Shoe Electrode: 440mm x 370mm x 35mm Weight: Hand appliance: appr. 250 g Shoe Electrode: appr. 2 kg Stromversorgung: 4 x Mignon (AA) Alkaline in the hand appliance Optional: plug power adapter 7,5V- DC / 300mA via Shoe Electrode with 4 x Mignon Battery NiMH (rechargeable) Display: Relay contact: (included in the shoe electrode) Principle of Measurement: 1 LED (8mm) green, 6 LEDs (5mm) red, 1 buzzer potential-free contact (60V/0,5A) Current / Voltage Converter Measure Voltage 20V ±5% Connection: :All connections via 9-pin Sub-D Connector Limits The limits are separate adjustable for the wrist strap test and the shoe test Setting range: Lower limit from 0... 2MOhm Upper limit from 5... 200Mohm Special limits are possible if needed by the customer! Function The testing person should stand with both shoes on the contact plates at the shoe electrode. The Wrist strap-cables are plugged into the measuring instrument. Then press and hold the Test -Plate on the appliance with a finger. When starting the measurement, a short acoustic signal sounds. At first, the resistor of the Wrist strap is measured, after it the resistors from the left and from the right shoe. It is tested whether the resistance values are within the lower and upper limits. If all resistance values are between the lower and upper limits, a short acoustic signal sounds and the green LED lights up. If one or several resistance values is outside of the lower or upper limit, no acoustic signal sounds, only the appropriate red < or > LED lights up. Also several mistakes can be represented simultaneously. For example : Wrist strap > and Shoe left < If the Test Plate is free, the displayed LEDs lights approximately 2 Seconds longer, then the appliance switches off.

4 Handbuch WST 100 Wrist strap test If the wrist strap test is selected, put the wrist band on and connect the grounding cord to one of the lower wrist strap sockets. Depress and hold the metal contact plate until a LED appears or the buzzer sounds If none of the both wrist strap LEDs lights up the wrist strap is o.k. If the red < LED is on, the current limiting resistor in the ground cord is out of tolerance and the ground cord has to be replaced. The > LED indicates that the resistor range is exceeded. First check the grounding cord. Leave the plug end attached to the tester and disconnect the snap end from the wrist band. Connect ground cord snap to the stud in the hand symbol and depress the metal contact again to start a new measure. If no red wrist strap LED shines, the ground cord is o.k. In this case you have a problem with the wrist band or the contact between the skin and the wrist band. This resistance may be caused by extremely dry skin or the presence of hair. The use of a skin lotion is recommended to solve this problem. If a red > LED condition still exists, replace the wrist band. Shoe test Stand with both feet on the left and right electrode of the shoe electrode. Depress the metal contact plate and hold until a LED appears. If no one of the shoe right/left LED light up shoe(s) o.k. If the red LED (shoe right/left) > is illuminated, clean the sole of the right/ left shoe or check your right/left shoe grounding system and retest. If the red LED (shoe right/left) > remains on, replace your shoes or grounding system. If the red LED (right/left) < is illuminated, the minimum resistance of your right/left shoe or grounding system is to low. Maybe, that your shoe are wet, then dry them. otherwise replace your shoes or grounding system. The wrist strap test does not influence the shoe tests. After pressing the metal plate the selected tests from the test mode will run step by step.

5 Adjustable abilities In the Set-Up Mode you can choose different functions Test - Mode: Only wrist strap test without Shoe - Test Only Shoe - Test with double shoe electrode Wrist strap and Shoe - Test with double shoe electrode Wrist strap and Shoe - Test with single shoe electrode Relay - Mode: Alarm function (Relay switches on in case of error) Door opener function (Relay switches on if o.k.) Statistisical error memory The appliance possesses 5 statistical memories for Measurement Number, wrist strap <, wrist strap >, shoes <, shoes > This works as follows: After every measurement the measurement number counter increases. If an error < or > is recognized the error counter < or > increases. Therefore the following to each memory applies: Number of the measurements minus errors < minus errors > = number of the O.K. measurements. So you will get a statistics statemant about the quality of the wrist straps and the shoes! Limits The lower and upper limits are fixed by lower and upper calibrated resistors. Factory default is: wrist strap test lower limit 750kΩ upper limit 35 MΩ Shoe test lower limit 100kΩ upper limit 35 MΩ Other limits are possible without extra charge. The limits can be changed by the customer anytime see adjustment of the limits. Evaluation lower limit Calibrated Value (LCV) LED < Calibrated Value (LCV) + 10% LED o.k. upper limit Calibrated Value (UCL) 10% LED o.k. Calibrated Value (UCL) + 10% LED > Change Battery The appliance possesses battery supervision. If the battery voltage is below the limit, an acoustic signal sounds with approximately 1Hz. In this case the batteries must be change. The battery compartment is on the underside of the appliance. When installing in new batteries look for the polarity!! Important! Use alkaline batteries only!

6 Handbuch WST 100 Connection The connection with the appliance takes place at a 9pol.Sub D pin jack. Connecting to this is the shoe electrode, the optional Power supply, the relay output and the serial PC Interface. Only one single cable is needed to connect to the hand appliance pin assignment 1 = U+ from power supply 2 = PC Out 3 = PC In 4 = shoe electrode left shoe 5 = GND (U- power supply) 6 = + measuring voltage 20V= 7 = Relay contact 8 = wrist strap 9 = shoe electrode right shoe Shoe electrode connections Cable to the hand unit At the front-side of the shoe electrode are the plugs for the power supply (right) and for the relay output (left) ( Relay contact 60V / 0,5A potential-free) Connecting a door-opener

7 PC Interface To using the PC interface unplug the plug to the shoe electrode and contact the PC Cable. Warning! Use the special WST-100 PC - cable only! Scope of supply WST 100 SET Hand instrument WST 100 Shoe electrode WST 100 SE 4 x Batteries AA Mignon Alkaline Manual Calibration Certification Optional accessories WST-100 RE Calibrated Resistor for the lower and upper limit with Certification of Calibration. wrist strap Standard - lower limit 750kΩ upper limit 35MΩ Shoe Standard - lower limit 100kΩ upper limit 35MΩ We can deliver each pair of limits into the limit range without extra charge! WST-100 NT The charger set includes 4 x NiMH Mignon batteries and a plug power supply which is connected to the shoe electrode. Important! Before using the power supply replace the alkaline -batteries with rechargeable NiMH types. WST-100 WB Wall connecting plate with testing instruction. Front side German / back side English. The fixation from the WST 100 on the board proceeds by Veltro strips (Dual Lock).

8 Handbuch WST 100 WST-100 PC Including the PC cable with 3,5 Disc. Retrieval and erasure of the statistical EEPROM-Memory Excel Software on Disk/CD to read out the statistical error memory. Calibration By using calibrated resistors by doing a new calibration and the result is saved in the EEPROM, this is also a calibration. With this calibrated resistors you can calibrate all your units at the user s site... recommended test cycle time: annually That offers great advantages: ESD Standard changing! By using the new resistor values, you can change the limits at the user s site! Calibration! By using calibrated resistors, you have only to recalibrate this. Than calibrate your instruments by yourself. No shutdown period and very cheap!!

9 Adjust the limits/calibrate the instrument Is required according to function: Calibrated resistor pairs for wrist strap and/or shoe test Sequence Press and hold button H (hidden), with a small screw driver Press the test plate shortly buzzer sounds! Release the button H Release the test plate Set-Up Mode operating mode puts in: By pressing the test - plate you now can change the operation mode: 1. All red LED s Wrist strap + shoe right + left test 2. Only wrist strap LED s (< >) Wrist strap test only 3. Wrist strap+shoe right LED s (< >) Wrist strap + shoe right test ( single shoe Electrode ) 4. Shoe right + left LED s (< >) Only shoe right / left test ( without wrist strap ) Through short pressing the button H your selected operating mode is saved. By pressing the test plate you can select now the relay mode 1. Only green LED Relay is in door open mode (switch on by o.k.) 2. All red LED s Relay is in Alarm mode (switch on by error) Through short pressing the button H your selected relay mode is saved. The buzzer sounds for approximately 1 second, than all LED's lights. Through pressing the Test Plate (> 1sec), the Set-Up can now be left. Through short pressing the button H you can continue with the adjustment of the limits.

10 Handbuch WST 100 Set-Up Mode Read Calibration Resistors : The LED wrist strap < lights up. (only if wrist strap is tested). Connect calibration resistor module for wrist strap limits, Press briefly button H. If value is O.K., the green LED lights up (+ buzzer for 1 seconds), afterwards the LED wrist strap > lights up. Press briefly button H. If the value is O.K., the green LED lights up (+ buzzer) for 1 second afterwards LEDs shoe right < and shoe left < lights up. (only if Mode Only Wrist strap is not selected) Connect the calibration resistor - module for shoe limits Ppress briefly button H. If value is O.K., the green LED lights (+ buzzers) for 1 seconds, afterwards the LED`s shoe right/left > lights. Press briefly button H. If value is O.K., the green LED lights. The new limits are saved By pressing the test - plate the WST-100 switchs off Limit ranges : lower limit : 0..... 2 MΩ upper limit : 5..... 200 MΩ Special ranges are available! Error If a connected calibration resistor is out of these ranges, all 6 red LEDs flash.. 5 x Error Calibration Value After that the unit switches off. Important! In this case you must repeat the complete Set Up!!!

11 Calibration Since the values are written directly with the adjustment into the EEPROM, the adjustment of the limits with calibrated resistors is also simultaneously the calibration of the appliance. In order to return to the normes, the calibrated resistors can be calibrated by the manufacturer annually. With these calibrated resistors, all appliances of the customer can then be calibrated again. Suggested calibration cycles: annual Calibrated Resistor Module RE-100 A calibrated resistor modulo consists of a 9-pin Sub-D Plug with the calibrated resistors for the lower and upper limits the following standard - modules are offered: A lower limit 100kΩ upper limit 35MΩ B lower limit 750kΩ upper limit 35MΩ Every desired resistance combination within the limit ranges can be delivered without extra charge!

12 Handbuch WST 100 Short Instruction Stand with both feet solidly on the electrodes! Connect the wrist strap! Press and hold the test plate! Wait until the test - result is shown! Green LED o.k. Wrist strap and Shoes o.k. buzzer = 1 x long ESD-secured area can be entered! One or more red LEDs > wrist strap or Shoe to high impedance One or more red LEDs < wrist strap or Shoe to low impedance ESD Entry!!