CARBON NANOTUBE FLAT PLATE BLACKBODY CALIBRATOR. John C. Fleming

Similar documents
Dr. Larry J. Paxton Johns Hopkins University Applied Physics Laboratory Laurel, MD (301) (301) fax

7.1 Advanced Himawari Imager (AHI) Design and Operational Flexibility Dr. Paul C. Griffith

The Council for Optical Radiation Measurements (CORM), NIST July 30 - Aug 1, 2018

TracePro Stray Light Simulation

Mu lt i s p e c t r a l

Philpot & Philipson: Remote Sensing Fundamentals Interactions 3.1 W.D. Philpot, Cornell University, Fall 12

THE COMPARISON OF DIFFERENT THERMAL ANALYSIS SOFTWARE FOR THERMAL SIMULATION OF THE SPACE TELESCOPE

Light Tec. Characterization of ultra-polished surfaces in UV and IR. ICSO October 2016 Biarritz, France

EO-1 Stray Light Analysis Report No. 3

ENHANCEMENT OF DIFFUSERS BRDF ACCURACY

Class 11 Introduction to Surface BRDF and Atmospheric Scattering. Class 12/13 - Measurements of Surface BRDF and Atmospheric Scattering

11.2 ABI s Unique Calibration and Validation Capabilities

CAL/VAL JAXA Agency Report

Mirror Manufacturing Status Ben Gallagher Technology Days in the Government 2007 July 31, 2007

Light Tec Scattering measurements guideline

Light Tec Scattering measurements guideline

Light Tec Scattering measurements guideline

Total and Spectral solar Irradiance Sensor (TSIS-1) Project Overview

CRISM (Compact Reconnaissance Imaging Spectrometer for Mars) on MRO. Calibration Upgrade, version 2 to 3

Daniel A. Lavigne Defence Research and Development Canada Valcartier. Mélanie Breton Aerex Avionics Inc. July 27, 2010

Assembly of thin gratings for soft x-ray telescopes

Mission statement. Infrared Cameras, Detectors & Technology

Infrared Scene Simulation for Chemical Standoff Detection System Evaluation

Spectral imagers for nanosatellites

Component Level Prediction Versus System Level Measurement of SABER Relative Spectral Response

Lecture 13.1: Airborne Lidar Systems

Evaluation of Satellite Ocean Color Data Using SIMBADA Radiometers

NEW OPTICAL MEASUREMENT TECHNIQUE FOR SI WAFER SURFACE DEFECTS USING ANNULAR ILLUMINATION WITH CROSSED NICOLS

High-Performance Thermal Imaging Cameras

Light Tec Scattering measurements guideline

Assessments of MODIS On-orbit Spatial and Spectral Characterization

NEAR-IR BROADBAND POLARIZER DESIGN BASED ON PHOTONIC CRYSTALS

FIXED FOCAL LENGTH LENSES

Improvements to Ozone Mapping Profiler Suite (OMPS) Sensor Data Record (SDR)

Polarization Ray Trace, Radiometric Analysis, and Calibration

Silicon Carbide Pointing Mirror Gimbal System (PMGS) Summary

BROADBAND SPECTRAL-POLARIMETRIC BRDF SCAN SYSTEM AND DATA FOR SPACECRAFT MATERIALS

Laser Beacon Tracking for High-Accuracy Attitude Determination

2017 Summer Course on Optical Oceanography and Ocean Color Remote Sensing. Introduction to Remote Sensing

A multidiscipline solution to support space systems engineering

Reflective Illumination for DMS 803 / 505

NIRvana: 640ST. Applications: Nanotube fluorescence, emission, absorption, non-destructive testing and singlet oxygen detection

Improvement of Infrared Imaging Video Bolometer Systems in LHD )

Light: Geometric Optics

Evaluation of radiative power loading on WEST metallic in-vessel components

dq dt I = Irradiance or Light Intensity is Flux Φ per area A (W/m 2 ) Φ =

High spatial resolution measurement of volume holographic gratings

Diffractive Optics for Moon Topography Mapping

An Optical System to Transform the Output Beam of a Quantum Cascade Laser to be Uniform

INFOGR Computer Graphics. J. Bikker - April-July Lecture 10: Shading Models. Welcome!

Determining satellite rotation rates for unresolved targets using temporal variations in spectral signatures

Supplementary Figure 1 Optimum transmissive mask design for shaping an incident light to a desired

Ghost and Stray Light Analysis using TracePro. February 2012 Webinar

Easy integration into complex experimental setup

TEMPO & GOES-R synergy update and! GEO-TASO aerosol retrieval!

Radiometry (From Intro to Optics, Pedrotti 1-4) Radiometry is measurement of Emag radiation (light) Consider a small spherical source Assume a black

Visible-frequency dielectric metasurfaces for multi-wavelength achromatic and highly-dispersive holograms

Design and Simulation of a NDIR Gas Sensor System with Ray Tracing Using a Monte-Carlo Method

The Evolution of Thermal Imaging Cameras

Optics Vac Work MT 2008

HARP-NEF Front End Assembly

Calibration Techniques for NASA s Remote Sensing Ocean Color Sensors

Working with M 3 Data. Jeff Nettles M 3 Data Tutorial at AGU December 13, 2010

Chapter 32 Light: Reflection and Refraction. Copyright 2009 Pearson Education, Inc.

Module 5: Video Modeling Lecture 28: Illumination model. The Lecture Contains: Diffuse and Specular Reflection. Objectives_template

Defect Repair for EUVL Mask Blanks

Introduction to Diffraction Gratings

Ralf K. Heilmann CAT-GS: Critical-Angle Transmission Grating Spectrometer January 27,

Review of paper Non-image-forming optical components by P. R. Yoder Jr.

Progress of the Thomson Scattering Experiment on HSX

ENMAP RADIOMETRIC INFLIGHT CALIBRATION

LaserDec CL - Product Description -

IR-Signature of the MULDICON Configuration Determined by the IR-Signature Model MIRA

Chapter 13 RADIATION HEAT TRANSFER

Agilent Cary Universal Measurement Spectrophotometer (UMS)

Novel Magnetic Field Mapping Technology for Small and Closed Aperture Undulators

Micro Sun Sensor with CMOS Imager for Small Satellite Attitude Control

dq dt I = Irradiance or Light Intensity is Flux Φ per area A (W/m 2 ) Φ =

Physics 11. Unit 8 Geometric Optics Part 1

A Guide to Reflectance Coatings and Materials

Analysis of Crosstalk in HgCdTe based Vertical Photoconductive LWIR Detector Arrays

On-orbit Calibration of the Geostationary Imaging Fourier Transform Spectrometer (GIFTS)

Continuous Wave 532 nm Diode-Pumped Solid State Laser CDPS532M

IMAGING AND FIELD SPECTROMETER DATA S ROLE IN CALIBRATING CLIMATE-QUALITE SENSORS Kurtis, J. Thome NASA/GSFC Greenbelt, MD

Three-dimensional imaging of 30-nm nanospheres using immersion interferometric lithography

axis, and wavelength tuning is achieved by translating the grating along a scan direction parallel to the x

Optical Scattering. Analysis. Measurement and SPIE PRESS. John C. Stover THIRD EDITION. Bellingham, Washington USA

Polarization Properties of ACS

Dr. Charles Browning Director, Materials & Manufacturing Air Force Research Laboratory

International Conference on Space Optics ICSO 2008 Toulouse, France October 2008

GEOG 4110/5100 Advanced Remote Sensing Lecture 2

Application of a Model to Evaluate Infrared Exposure Limits in Aluminum Foundries Based on Threshold Temperature in the Range of nm

TABLE OF CONTENTS PRODUCT DESCRIPTION CINCAM CCD TECHNICAL DATA SENSOR RESPONSE DIMENSIONS CINCAM CCD LARGE FORMAT TECHNICAL DATA SENSOR RESPONSE

Glare Spread Function (GSF) - 12 Source Angle

Polarized Downwelling Radiance Distribution Camera System

HIGH SPEED PROJECTILE RECORDING SYSTEM

WHITE PAPER. Application of Imaging Sphere for BSDF Measurements of Arbitrary Materials

SALT TELESCOPE INSTRUMENT STRUCTURE ANALYSIS

Meteosat Third Generation (MTG) Lightning Imager (LI) instrument performance and calibration from user perspective

Lockheed Martin Nanosystems

Transcription:

CARBON NANOTUBE FLAT PLATE BLACKBODY CALIBRATOR John C. Fleming Ball Aerospace, jfleming@ball.com Sandra Collins, Beth Kelsic, Nathan Schwartz, David Osterman, Bevan Staple Ball Aerospace, scollins@ball.com bkelsic@ball.com nschwart@ball.com dosterma@ball.com bstaple@ball.com ABSTRACT Evolving remote sensing missions present a growing need for satellite sensors with significantly enhanced measurement accuracies beyond current capabilities. For example, the on-board calibration needed for very precise (e.g., <1% radiance uncertainty), spectrally resolved IR radiances typically requires high-emissivity, ε ( 0.999) calibration blackbodies (BBs). The challenge is that conventional BBs have coating emissivities that are usually limited to 0.98. Additionally, they generally have complex geometries that create fabrication and coating non-uniformities which leads to radiance, emissivity, and temperature calibration uncertainties. Moreover, conventional BBs are typically large size, weight, and power (SWaP), which is not optimal for on-board calibration and can be very expensive. At Ball, our approach to enhance remote sensing missions beyond current capabilities is to leverage the unique optical properties of Carbon Nanotubes (CNTs) to develop advanced coatings that overcome the limitations of conventional BB coatings. We have developed Flat-Plate, Extreme-ε (0.999), BB Calibrator based on a Vertically Aligned CNT (VACNT) coating. Key BB elements, such as the extreme-ε VACNT, have been verified for Ball by NIST to be 0.999. Extreme-ε provides a significantly reduced BB radiance uncertainty by minimizing the radiance error due to the radiation reflected from the BB s background environment. Additionally, the flat-plate simplifies instrument design, is optimal for on-board calibration due to its low SWaP, and it eliminates the fabrication and coating difficulties associated with the complex geometry of conventional BBs. This paper summarizes the results of our design, characterization and plan for implementing the technology on a small satellite platform with a micro-bolometer sensor. CNT PROPERTIES Our various Internal Research and Development (IRAD) and Contract Research and Development (CRAD) efforts have collected a wide range of reflectance and environmental data on the VACNT surfaces. This data informs our decision to use the surface in a blackbody and provides high confidence that it will survive the space environment. VACNT Structure The VACNT surface is comprised of multi-walled cylinders of sp2 bonded carbon. These VACNTs are on the order of 10-30 nm in diameter and for our surfaces from 100 to 300 µm in length. The grown VACNT surface has a density greater than 10 10 CNTs/cm 2. Figure 1 shows a schematic view of a single walled CNT and a side view of a VACNT surface. A key aspect of our IRAD effort has been selection of the optimum substrate on which to grow the VACNTs and consistent growth over a range of sizes and thicknesses of the substrate. Copyright 2017 by John C. Fleming, Ball Aerospace. All rights reserved. Page 1 of 8

Emissivity 33 rd Space Symposium, Technical Track, Colorado Springs, Colorado, United States of America Figure 1. Schematic View of a single CNT and SEM image of a VACNT surface VACNT Surface Reflectance and BRDF The reflectance of the VACNT surface is extremely low at all wavelengths we have measured. NIST measured the spectral emissivity of two coupons in 2012. This data in Figure 2 show very high emissivity from the midwavelength infrared (MWIR) through the long-wave infrared (LWIR). Figure 3 shows the bi-directional reflectance distribution function (BRDF) of a typical sample measured at 3.39 µm. The BRDF is very flat up to a 30 angle of incidence (AOI). At a 60 AOI, the specular reflectance increases but is still very low. Also shown on the plot is the typical field angles (±10 ) where a blackbody is observed. At these view angles and even out to a 60 AOI the BRDF is very low. This low BRDF over a wide range of illumination angles will be highly effective in limiting stray light from other sources that are adjacent to the blackbody. 1 0.9995 0.999 0.9985 Sample GN00012-1 Sample GN00012-2 0.998 0.9975 0.997 0.9965 0.996 3 4 5 6 7 8 9 10 11 12 13 14 Wavelength (µm) Page 2 of 8

Total Reflectance (%) BRDF (/sr) 33 rd Space Symposium, Technical Track, Colorado Springs, Colorado, United States of America Figure 2. Spectral Emissivity as Measured by NIST BRDF at 3.39 um 1.E-02 GN12_1_2 5AOI GN12_1_2 30AOI 1.E-03 GN12_1_2 60AOI 1.E-04 Typical Instrument FOV 1.E-05-90 -60-30 0 30 60 90 Scatter Angle Relative to Surface Normal (degrees) Figure 3. BRDF measured at 3.39 µm NIST documented a technique to reduce the reflectance of CNT samples 1. This process was repeated on Ball samples and the reduction was significant. The data below shows the change in reflectance over the 250-2500 nm range. The data in the SWIR range is noisy, but the processing reduced the reflectance by approximately a factor of 3x. We plan to employ this post-processing on the blackbodies we build and test this year. 0.45 VACNT Surface - Change in Reflectance After Post Processing 0.4 0.35 0.3 0.25 Before (dashed) After (solid) SN-T51 After SN-T81 After SN-T87 After SN-T51 Before SN-T81 Before SN-T87 Before 0.2 0.15 0.1 0.05 0 250 500 750 1000 1250 1500 1750 2000 2250 2500 Wavelength (nm) Figure 4. Change in Visible to NIR Reflectance after Post-Processing Page 3 of 8

Environmental Testing We have tested the survivability of the VACNT surface through temperature cycling, vibration, radiation exposure and atomic oxygen (AO) exposure. The only testing that revealed some negative effects was the atomic oxygen exposure. Figure 5 compares the BRDF measured before and after eight temperature cycles from -30 to +50C. Later testing has exposed the VACNT surface to temperatures up to 450K with no evidence of any change. Since the VACNT growth is at temperature well above 450K, this is not a surprising result. Vibration testing of VACNT coupons was performed in 2 axes parallel and perpendicular to the CNT surface at 14.1 Grms (GEVS) and 43.8 Grms. Results from all CNT vibration tests showed contamination levels below CL300. For these results all particulates found were counted even though the majority were not CNTs. This same result was seen on VACNT surfaces provided by multiple suppliers. Figure 6 shows the total reflectance of VACNT coupon before and after exposure to radiation. The exposure was equivalent to 5 years in a 700 km sun-synchronous orbit with a 2x radiation design margin. There is no evidence of any change in reflectance. Figure 7 shows the change in reflectance due to AO exposure. The first round of exposure simulating multi-year exposure in low-earth orbit in the Ram direction completely removed the coating. For the data in the figure the AO exposure was 3E21 atoms/cm2 at 0.03eV. The AO exposure slightly reduced the total reflectance. BRDF at 3.39 µm Before BRDF at 3.39 µm After Figure 5. Comparison of VACNT Surface BRDF Before and After Thermal Cycling Page 4 of 8

Total Reflectance (%) 33 rd Space Symposium, Technical Track, Colorado Springs, Colorado, United States of America Figure 6. Total Reflectance Before and After Radiation Exposure 0.45 0.4 0.35 Sample exposed to thermal AO flux (0.03eV) no VUV, total fluence 3E21 atoms/cm2 Dashed lines are before exposure Solid lines are after exposure 0.3 0.25 0.2 SN160315SIC_T2 Measurement-1 SN160315SIC_T2 Measurement-2 T2 Measurement-1 T2 Measurement-2 0.15 0.1 0.05 0 250 500 750 1000 1250 1500 1750 2000 2250 2500 Wavelength (nm) Figure 7. Change in Reflectance after AO Exposure CNT BLACKBODY DEVELOPMENT Ball s first flight of a system employing VACNT blackbodies is the Compact Infrared Radiometer in Space (CIRiS) expected to launch in 2018. The CIRiS calibration approach uses a scene select (fold) mirror to direct three calibration views to the focal plane array and to transfer the resulting calibrated response to earth images. The views to deep space and two blackbody sources, including one at a selectable temperature, provide multiple options for calibration optimization. 2 Both blackbodies utilize the VACNT technology developed on Ball s various IRAD efforts. One blackbody is thermally connected to the spacecraft structure while the other can be controlled up to 20K above ambient. The scene select mirror is in the path and at the same angle for both imaging and Page 5 of 8

calibration so its transmission and polarization effects are eliminated. Figure 8 shows two views of the CIRiS optical system and spacecraft. Fold mirror CNT blackbody Telescope Space Nadir CNT blackbody Figure 8. Views of CiRIS Design with CNT Blackbodies Ball has been provided funding the last several years from a government customer to further develop VACNT technology for use in future space missions. This funding enabled much of the environmental testing documented above and also blackbody and system level testing of the technology. In 2016 we assembled and tested a tiled blackbody calibrator consisting of four 3 x3 VACNT coated coupons bonded into a 6 x6 blackbody. The setup of the blackbody inside a vacuum chamber is shown in Figure 9. Results of this testing are shown in Figure 10. The uniformity was not as good as we had hoped, but the test results still provided highly valuable lessons learned for assembling large blackbodies from smaller coupons. Figure 9. CRAD Blackbody in Vacuum Chamber Page 6 of 8

250K ΔT=0.25K 350K ΔT=0.35K 450K ΔT=1.1K Figure 10. Test Results from 2016 CRAD Test of Tiled VACNT Blackbody In 2017, our CRAD effort is focused on building and testing a flight-like blackbody. We will build and test two VACNT blackbodies and image them with a state-of-the-art digital infrared focal plane. At each temperature setpoint we will also image a GSE blackbody that provides a highly uniform reference calibration. This test design is shown in Figure 11. Figure 12 shows raytraces to the two VACNT blackbodies. Testing will compare the radiance and uniformity of the two CNT blackbodies to the reference blackbody from 250 to 450K. Testing will also demonstrate the temperature rate of change that we expect to be significantly faster than conventional blackbodies. CNT BB1 CNT BB2 SSM GSE BB Figure 11. Test Design of 2017 VACNT Blackbody Test System Page 7 of 8

Figure 12. Raytrace Paths to Two VACNT Blackbodies SUMMARY Through Ball investment, a NASA program and CRAD funded efforts from a government customer we have made significant strides toward building flight blackbodies incorporating VACNT surfaces. These blackbodies show excellent potential to be used in future NASA and DoD missions requiring exquisite infrared calibration. Key attributes are the high emissivity, survival in challenging environments including thermal, vibration, radiation, and atomic oxygen. This technology will be further demonstrated in a flight-like test this year and the CIRiS launch in 2018. 1 Tomlin, N. A, et. al, Decrease in reflectance of vertically-aligned carbon nanotubes after oxygen plasma treatment, Carbon Journal, 2014, pp. 329-332. 2 Osterman, D. P., et. al., CIRiS: Compact Infrared Radiometer in Space, SPIE 9978, September 2016 Page 8 of 8