Strain Gauge Measurement by Applying Displacement using Strain Indicator

Similar documents
EDX-10 Series. Compact Recording System

EDX-10 Series. Compact Recording System

Strain and Force Measurement

Using LabVIEW in Instrumentation and Control Course

SKEE 2742 BASIC ELECTRONICS LAB

CLEVELAND CONTROLS MODEL 6650 DIGITAL MANOMETER INSTRUCTION MANUAL DM

System StrainSmart Data Acquisition System. Micro-Measurements FEATURES DESCRIPTION MODEL 5100B SCANNER SPECIFICATIONS

Strain Gage Measuring Amplifier GSV-1

PCD-400A/430A. Sensor Interface. Carrier Wave Type Compact and Moderate Prices. Data Recorders/Analyzers MEASURING INSTRUMENTS

DFI 7000 DIGITAL FORCE INDICATOR USER S GUIDE 1.0 INTRODUCTION 2.0 DESCRIPTION. 2.1 Configuration. 2.2 Specifications

INSTRUCTION MANUAL. Digivac 760 Series. Models 100Led P / P / 29.9" 100Led 760 / SixVac 801 / 760 / 2c 801 Micro

LAUREL. Laureate Digital Panel Meter for Process and Ratiometric Signals ELECTRONICS, INC. Features. Description

The PM1000 series is a universal 4 digit LED plug-on display for transmitters with 4-20mA 2 wire output and fitted with DIN43650 connector.

LAUREL. Laureate Digital Panel Meter for Load Cell & Microvolt Input ELECTRONICS, INC. Features. Description

SA SERIES Arlingate Lane, Columbus, Ohio (614)

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB

Agilent Strain Measurement Application Using Agilent U2300A Series Data Acquisition Devices With Agilent VEE Pro. Application Note

MODEL 9250 Preliminary data sheet

BRC SERIES 20kW High Voltage Power Supply

4000 SERIES REGULATED HIGH VOLTAGE DC TO DC CONVERTERS

DC-204R DC-204Ra SMART DYNAMIC STRAIN RECORDER. 200kHz 2GB UPS. Strain DC voltage Temperature. TML Pam E-442B. capacity. High speed sampling

Model SRV411i Small Animal Floor Scale Service Manual Part No. MANSRV411i_ Page 1 of 16. Small Animal Floor Scale. Operating and Service Manual

SECTION 1 INTRODUCTION. Walt Kester

BE/EE189 Design and Construction of Biodevices Lecture 5. BE/EE189 Design and Construction of Biodevices - Caltech

StrainSmart Data Acquisition System

Transducers and Transducer Calibration GENERAL MEASUREMENT SYSTEM

Model User Manual Revision E 04/29/99. OEM Gaging System. Part Number R01

NIST Traceable Testing at the Press of a Button

ZENSOL CIRCUIT BREAKER PERFORMANCE ANALYZER

Exercise 1: Identify Types of Switches

ED-1100 PORTABLE VARIABLE EDDY CURRENT INSTRUMENT

E101 - Strain Gauge Transducer Display Module. Contents

DFI 9000 PORTABLE DIGITAL INDICATOR USER S GUIDE

GSV-1A4 M12/2 M12/2. Highlights

TM0-1 AMPLIFIER / CONDITIONER MODULE OPERATORS MANUAL. Transducer Techniques R

EZCT-2000C Plus. current transformer test set. Vanguard Instruments Company, Inc.

Portable vibration calibrator Model 28959G

OPERATING AND SERVICE MANUAL SPECIAL SUPPLEMENT

Standard Options. Model 4100 Position Indicating Meter. Three Phase Motor Control. Positran Transmitter

Model SR411i-KG Patient Floor Scale Service Manual Part No. MAN411i-KG_ Page 1 of 16. Patient Floor Scale. Operating and Service Manual

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB

OPERATION MANUAL - DS-MOD-

Table of Contents. Introduction... 2 Description...2 Specifications...3 Installation.. 5 Calibration.. 6 General Maintenance...7

DVC-350C CALIBRATORS, INC HANDHELD BATTERY-POWERED VOLTAGE CALIBRATOR FEATURES INTRODUCTION GENERAL DESCRIPTION

JR3 EXTERNAL SENSOR ELECTRONICS WITH ANALOG OUTPUT. JR3, Inc. 22 Harter Ave. Woodland, CA 95776

CT-7500 S2. Advanced Test Equipment Rentals ATEC (2832) digital circuit breaker analyzer

Strain Gage Measuring Amplifier GSV-3

MAX Modular Analogue Expandable System

Advanced NI-DAQmx Programming Techniques with LabVIEW

MP3V5050, 0 to 50 kpa, Differential, and Gauge Pressure Sensor

DBK48. Multipurpose Isolated Signal-Conditioning Module. Description. Supports up to Sixteen 8B Modules

CONNECTING SENSORS TO DATA ACQUISITION SYSTEM ANALOG INPUTS

CT-6500 S2. digital circuit breaker analyzer. Vanguard Instruments Company, Inc.

6224 Ethernet-Based Strain Measurement Module

Wide Bandwidth Strain Gage Input 3B18 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Laboratory of Sensors Engineering Sciences 9 CFU

MP-2000 Dual Channel LVDT/RVDT Readout/Controller

DIGITMR S2 PC. digital circuit breaker analyzer. Vanguard Instruments Company, Inc.

Model HU-224/225. Technical Information TI.224/ Humidity Transduce r FOR ADDITIONAL INFORMATION SEE HU-224/225 DATA SHEET

PARAMETER MIN TYP MAX UNITS COMMENTS

User Manual for Portable National Instruments Board, Bridge Amplifiers, and VI Logger Software

SCXI OFFSET-NULL AND SHUNT-CALIBRATION HIGH-VOLTAGE TERMINAL BLOCK INSTALLATION GUIDE

AMP20. User Manual.

Determination of Drag Coefficient

Mounting DIGITAL DISPLAY

PSM-2 POWER SUPPLY MODULE INSTRUCTIONS

MICRO SERIES PROCESS DISPLAYS LARGE DIGIT MODELS

800 Series. Strain Gauge / Load Cell Transmitter with Alarm. exc + sense + bridge + bridge sense exc

JR3 EXTERNAL SENSOR ELECTRONICS WITH SERIAL DATA OUTPUT. JR3, Inc. 22 Harter Ave. Woodland, CA 95776

RESISTANCE/CURRENT METER Model 863/6487

Handheld hydraulic readouts & dataloggers

ECE 2010 Laboratory # 2 J.P.O Rourke

OBSOLETE. Isolated, Linearized, Thermocouple Input 3B47 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Sheet Resistance Meter. Model SRM 232 MODE SELECT STORE SURFACE RESISTIVITY METER SRM-232 POWER. Operator s Guide

Isolated Wideband Voltage Input 3B40 / 3B41 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Handheld hydraulic readouts & dataloggers

HI ma Simulator

8700 SERIES. narda OPERATION MANUAL DIGITAL METER CALIBRATION FIXTURE. Safety Test Solutions P/N an L3 communications company

TECHNICAL MANUAL. Man / Auto. Ten Off / On x100 psi Tension lb X1000 lb 0.

7561-PSD Manual Portable Battery Powered Indicator

Determination of Calibration Factors

EZCT-2000C current transformer test set

HIGH ACCURACY PRESSURE TRANSDUCERS

1523/1524. Reference Thermometers. Measure, graph and record three sensor types with one tool

Isolated Voltage Input 3B30 / 3B31 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

TELEDYNE HASTINGS INSTRUMENTS

Advanced Test Equipment Rentals ATEC (2832)

Industrial Wet/Wet Differential Pressure Transducer

OPERATING INSTRUCTIONS

Webtec Products Limited. Handheld hydraulic readouts & dataloggers

AirTest Model CN9000 Series Sensor Controller

GSV-3USBx2 3,5mV/V 4,2V

Project Progress. 1. Overall Design

MODEL 8000MP LEVEL SENSOR

GSV-6BT GSV-6BT. Description

Model SC500. Programmable Single-Channel Transducer Indicator/Conditioner DESCRIPTION FEATURES

high performance signal conditioner cards

K-Module 544 for: AS 43, AS 45, AS 200 IT. Installation Manual

Calibration of molbox1+ flow terminal Fluke Corporation Fluke Calibration Flow Course 1

SensoTrans DMS A 20220

Transcription:

University Diploma Program Electronic Equipment Maintenance Lab Instructor: Muhammad Ajmal Khan EET-027, Experiment # 8 Strain Gauge Measurement by Applying Displacement using Strain Indicator Objectives: Measuring strain when the strip end is displaced in the strain gauge micrometer device. Apparatus: Strain gauge Staring gauge micrometer Different Weights 1 kg, 2k, 5 kg. Strain Indicator Theory: The strain gauge is a transducer employing electrical resistance variation to sense the strain produced by a force or weight. It is a very versatile detector for measuring weight, pressure, mechanical force, or displacement. The Model P-3500 Strain Indicator is a portable, battery-powered instrument with unique features for use in stress analysis testing, and for use with strain gage based transducers. In use, the operator follows a logical sequence of setup steps by activating color-coded pushbutton controls to prepare the instrument for making accurate and reliable measurements. The P-3500 also incorporates a highly stable DC amplifier, precisely regulated bridge excitation supply, and precisely settable gage factor controls. Static measurements are displayed directly on the indicator's readout with 1 micro-strain resolution. The instrument will accept full-, half-, or quarter-bridge strain gage inputs, and all required bridge completion components for 120, 350 and 1000 ohm gages are built in. EET-027 Lab Manual Copyright Electrical Engineering Department, KFUPM. 35

Gage factor is precisely settable (to a resolution of 0.001) by a front-panel 10-turn potentiometer, and is displayed on the digital readout when the gage factor push button is depressed. Strain Indicator P-3500 Front Panel Procedure: 1. Measuring Strain using Strain Indicator: The P-3500 is designed for ease of operation, the push-button switches and front panel controls are arranged such that the proper setup procedure generally follows a straightforward left-to-right sequence. To measure the strain, the steps is outlined below: 1. Select 1/4-1 /2 position of BRIDGE push button. 2. Select Xl position of MULT push button. 3. Connect strain gage to binding posts connector. These binding posts are color-coded in accordance with conventional practice, and are clearly labeled. Input connections are shown on the inside cover of the instrument. 4. Depress AMP ZERO push button. Allow instrument to warm up for two minutes minimum. Set AMP ZERO control for a readout display of ±0000. This adjustment must be made with MULT in Xl position. 5. Depress GAGE FACTOR push button. Set GAGE FACTOR range switch and GAGE FACTOR control for the desired gage factor. 6. Depress the RUN push button. Set the BALANCE switch and the BALANCE control for a reading of ±0000. This setting must be made with the MULT in the Xl position. EET-027 Lab Manual Copyright Electrical Engineering Department, KFUPM. 36

7. Depress the CAL push button and verify calibration of the instrument. 8. Select the Xl or Xl 0 MUL T position as required. 9. Depress the RUN push button. Displace the end of strip and record the reading of strain in the table 1. 10. Plot the graph of Strain versus Resultant Displacement and find the slope of the graph and find strain at displacement of 0.115 inch from graph. Table 1 S. No. Displacement Resultant Displacement (x-a) Strain (micro-strain) 1 a = 0 2 3 4 5 6 7 8 9 10 11 12 13 EET-027 Lab Manual Copyright Electrical Engineering Department, KFUPM. 37

2. Measuring Differential Voltage of Full Bridge Circuit when Strip is displaced: 1. Connect strain gauge with the bridge circuit as shown the following figure. Set the power supply to 10 Volts and all three resistances are 120 ohms. R 1 R 2 ΔV V s =10V a DVM b R 3 Strain Gauge 2. Find the voltage difference ( Δ V ) across nodes a and b using digital volt-meter (DVM) without any displacement and record the value in the table 2. 3. Apply some displacement using micrometer and find the voltage difference ( Δ V digital volt-meter (DVM) and record the values in the following table 2. ) using 4. Plot the graph of differential voltage versus Resultant Displacement and find the slope of the graph and find the differential voltage at displacement of 0.115 inch from graph. S. No. Displacement Table 1 Resultant Displacement (x-a) Differential Voltage (Volts) 1 a = 0 2 3 4 5 6 7 8 EET-027 Lab Manual Copyright Electrical Engineering Department, KFUPM. 38

9 10 Conclusions: Compare the slope of the two graphs? And write your comments. EET-027 Lab Manual Copyright Electrical Engineering Department, KFUPM. 39