UNC Flow Cytometry Core Facility LSR II Operation 2018

Similar documents
FACSLSRFortessa SORP QUICK REFERENCE GUIDE

FACSAria Fusion SORP QUICK REFERENCE GUIDE

LSRFortessa Guide. Cell Evaluation and Therapy Core HO324

BD FACSDiva Software Quick Reference Guide for BD FACSCanto Systems with Loader Option

Unscrew and remove the cap from the waste tank

Log into the computer using your CU Denver username and password. Log into the DIVA software using your CU and Username and the password flow

Canto II Operation Protocol

Flow Cytometry & FACS Core Facility

FACSLyric TM General User instructions

BD FACSMelody Cell Sorter Quick Reference Guide

Setup protocol for Firefly particles. On BD LSR II and LSRFortessa cytometers with single tube loader function

Setup protocol for Firefly particles. On BD FACSCanto cytometers with single tube loader function

Standard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance:

CyAn ADP Guide. Starting Up

Standard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance:

USER GUIDE TO BECKMAN COULTER FC500

FACCalibur Users Guide

The Danish Stem Cell Center (DanStem) Danish Stem Cell Center Flow Cytometry Core Facility. LSR Fortessa Guide

STANDARD OPERATING PROCEDURE FOR THE LSR II

BD Influx Standard Operation Protocol - Basic Operation

Gallios TM Quick Reference

NovoExpress Software Guide. Software Version 1.2.4

BD CellQuest Pro Acquisition Tutorial

BD CellQuest Pro Software Acquisition Tutorial

Customer ( you ) must review these notes prior to installing or operating the Attune NxT Software version ( Attune Software v1.1.0 ).

JMJ-Group Microbiology BMB SDU

Science is hard. Flow cytometry should be easy.

FACScan 4 and 5 Color Upgrade Rainbow Users Guide

Applied Biosystems Attune Cytometric Software version 1.2 Release Notes

SPA III Quick Reference Guide for Custom Panels

FloMax Software for Cytometry. Operating Manual Instrument Control and Acquisition. FloMax Operating Manual Instrument Control and Acquisition

Release Notes Life Technologies Attune NxT Software v2.2

Attune NxT Acoustic Focusing Cytometer

Release Notes Life Technologies Attune NxT Cytometric Software v 2.0.0

Tutorial: FCAP Array Software with BD FACSArray Bioanalyzer

Release Notes Thermo Fisher Scientific Attune NxT Software v2.4

BD High Throughput Sampler User s Guide

Gallios Flow Cytometer

LEGENDplex Data Analysis Software Version 8 User Guide

A Bright Cytometer Easy to use

Release Notes Life Technologies Attune NxT Software v2.3

BD Lyoplate Human Screen Analysis Instructions For analysis using FCS Express or FlowJo and heatmap representation in Excel 2007

Comparing technical specification documents

CELLQuest Acquisition Tutorial

BD Multiwell AutoSampler Additional Features Tutorial

BD CellQuest Pro Analysis Tutorial

Technical Note. Display of BD Accuri C6 Software- Generated FCS 3.0 Files Using FlowJo for PC and Mac v7.6. Introduction

North America/International Japan Europe

Cytomics FC 500 Quick Reference

Data acquisition and analysis of MACSPlex Exosome Kit using the MACSQuant Analyzer Express Mode in MACSQuantify Software version 2.

Spherotech, Inc Irma Lee Circle, Lake Forest, IL

Quick Start Guide: COLLECT

Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.

Instructions for Use. COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software. PN CA (September 2010)

Agilent Feature Extraction Software (v10.5)

FlowJo Software Lecture Outline:

FX500 Exchangeable Fluidics Cell Sorter

CyFlow ML. Healthcare Immunology Pathology Microbiology Cell Biology. Multilaser 16 Parameter Desktop FCM System

MA900 Multi-Application Cell Sorter

LSM510 Confocal Microscope Standard Operation Protocol Basic Operation

Hacking FlowJo VX. 42 Time-Saving FlowJo Shortcuts To Help You Get Your Data Published No Matter What Flow Cytometer It Came From

Flow Cytometry/FACS Bio-Rad S3e Cell Sorter

Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.

MACSQuantify Software 2.11

Aqueous GPC Instructions - Detailed

Import Files from Other Applications

Basic Tutorial. Data Analysis Software for Flow Cytometry Windows/Mac user documentation

CyFlow Cube 6 Instrument Operating Manual

Offline Flow Cytometry Analysis Software: Importance of Scaling and Data Visualization

BD Accuri C6 with single tube loader protocol. For multiplex assays

Seized Drugs Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division

NovoSampler Pro Operator s Guide

Agilent MassHunter Workstation Software Report Designer Add-in

Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017

DIAplex. DIAplex Pro 1.0 Software. Instructions for use. Fast Track Your Research. For research use only

Nanosight NS300 NBTC User Instructions

Indiana Center for Biological Microscopy. BioRad MRC 1024 MP Confocal & Multi-Photon Microscope

Setup and analysis Mic qpcr Cycler. (magnetic induction cycler; Bio Molecular Systems)

Standard Operating Procedure for the Horiba FluroMax-4

Working with the Production dashboard

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Luminex Prima

User Operation Manual

Xfmea Version 10 First Steps Example

Reviewing Hidden Content during Native Review

STAINING PROTOCOL FOR CYTOMETRY (Réf LabEx #1)

BioRobot 9604 Operating Procedure

Leica TCS SP8 X Confocal Microscope and Leica Application Suite X software DRAFT VERSION Room B123b

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted)

Co-Lase Users Guide. DM00001 Rev A (DCN 031) These instructions are protected by copyright. All rights reserved.

LEXT 3D Measuring LASER Microscope

Flow Cytometry Analysis Software. Developed by scientists, for scientists. User Manual. Version Introduction:

1. Getting Started: Brief Step-By-Step Guide (PDF File)

ClonePix TM 2 QUICK SET-UP INTRUCTIONS MANUAL SOFTWARE RELEASE

OpenForms360 Validation User Guide Notable Solutions Inc.

Tutorial: Create an Excel Dashboard. Overview. **** Download the Example Dashboard ****

QAM-I-102 Operation and Calibration of the Autoanalyzer

TraceFinder Analysis Quick Reference Guide

Agilent ChemStation Plus

Getting Started. COULTER EPICS XL TM Flow Cytometer COULTER EPICS XL-MCL TM Flow Cytometer SYSTEM II TM Software. PN BA (October 2010)

Transcription:

Reference: Startup BD FACSDiva 6 Online Course BD Training & elearning 1. Check Sheath and Waste containers: Empty Waste container in the sink Add bleach to the black mark ( ~10% final concentration). Fill Sheath container with Sheath Solution (1X PBS) Do not let fluid level go below the black mark. 2. Turn on FACSFlow Supply System (Fluidics Cart) Green button on the right-hand side. 3. Turn on Cytometer Green button on the right-hand side (watch for position of the HTS/tube switch). 4. Check Sheath filter for air, bleed if necessary. Bleed side tubing. 5. Log in to Windows (Onyen + password). 6. Turn on necessary lasers: GUI-VFL (green arrow icon) for 592nm laser Select laser power, set to 100mW, turn On and Activate. BD Coherent Connection 4 for 355, 405, 488, 532, 561, and 633nm lasers Click Laser On. 7. Launch BD FACSDiva Software (Username + password). 8. Check Flow Cell for bubbles. Prime if necessary. Wait for the computer to establish a connection with the cytometer If the connection fails, restart the cytometer (green button) or alternatively power down the cytometer and the computer and then restart them both. 1

BD FACSDiva Software Menus Overview 2

Quality Control Cytometer Setup & Tracking (CST) CST beads are used to optimize each PMT. This ensures that the ratio of signal to noise is the highest, and the CVs of the positive peak are the smallest. Launching the CST Workspace Select Cytometer Menu > CST CST Workspace Overview 3

Instrument Configuration Select the Configuration that matches the optical bench layout (filters & mirror installed on the cytometer). Within the Cytometer Configuration windows, select a configuration from the left-hand side menu and click Set Configuration. Click OK to close the Window. Close CST Workspace. Note: If necessary change filters & mirror to accommodate your experimental staining panel. Running CST (Performance Check) Prepare the CST beads (1 drop in 350µL of PBS) Verify the bead Lot ID is correct Select Run mode and Low on the fluidics control panel of the cytometer. Click Run in CST Workspace 4

CST Results Under System Summary, check the system s status. If the status is Requires Attention!, you must run a Performance Check for that configuration. If the status is OK, proceed with setting up your experiment. Click the Report and troubleshoot any warnings or failures. 5

Working with BD FACSDiva Software Worksheets are where you create plots, gates, population hierarchies, statistic views, and custom text. There are 2 types of worksheets between which you can toggle (Worksheet View button): Global: Displays data from one tube at a time Uses the same set of plots/gates to analyze any tube in an experiment Used to record data using the Loader or HTS options Used to perform batch data analysis Normal: Used to perform compensation Displays data from several tubes simultaneously Uses different gating strategies (on the same worksheet) for tubes within an experiment Worksheet View button Data displayed in a Global Worksheet changes when the current tube pointer is moved to another tube. 6

Data displayed in a Normal Worksheet is not dependent on where the current tube pointer is set. For example, the plots in the graphics below display data from tubes 001 and 002. Additionally, you can use different gates and gating strategies for each tubes. Here, the P1 gates in the plots are different. 7

Setting Up Your Experiment Adjusting Baseline Voltages Click New Experiment (2 nd icon in the Browser toolbar). OR Recall a saved Experiment Template (Experiment menu>new Experiment ). Rename the Experiment (R-click>Rename). Click on Cytometer Settings and delete unused parameters in the Inspector window. Check the Height (H), Width (W), or both for FSC, SSC, or both to perform doublet discrimination. Check the Log box and the Area (A) box for all fluorescent parameters if running a phenotypic experiment. Right-click Cytometer Settings>Application Settings>Create Worksheet. A sheet titled Global Sheet2 opens. Click on the Experiment (in Browser) and create a New Specimen (3 rd icon in the Browser toolbar) then place the tube pointer on Tube_001. The Acquisition Dashboard becomes available. Filter your samples prior to running on the instrument Load your unstained control on the cytometer. Click Acquire on the Acquisition Dashboard. 8

Adjust FSC-A and SSC-A PMT voltages, as well as the threshold if needed. Adjust the voltages of all your fluorescent parameters in order to place the negative population in the grey box, below the white cross hair. Run a fully-stained sample to make sure all fluorescent parameters are on scale. R-click Cytometer Settings>Application Settings>Save To recall Application Settings: R-click Cytometer Settings>Application Settings>Apply and then R-click Application Settings>Apply Current CST Settings. Note: Using Application Settings, synchronizes the voltages of a specific experiment with the voltage fluctuations recorded by CST during the daily Performance Check. Your voltages will be adjusted to reflect day to day changes in the cytometer signal detection. Note: Alternatively, baseline voltages can be set using an unstained control, adjusting PMT voltage so the population is approximately at 2.5x rsd EN (rsd EN is calculated during CST Baseline determination. See Technical Bulletin Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva V6 March 2012). Compensation Go to Experiment menu>compensation Setup>Create Compensation Controls. Change the Generic Label to the antibody used in your single-stained controls. If using an unstained control, check the Include separate unstained control tube/well box. If not, you will have to manually create a gate around the negative population of all your single-stained controls. Click OK. A Specimen called Compensation Controls is created. Open it by clicking on the +. Place your tube pointer on your first single-stained control or unstained control. Load the corresponding sample on the cytometer. Click Acquire (adjust P1 if necessary you can R-click on the P1 to Apply to All Compensation Controls ). Click Record (5000 events are automatically recorded). Click Next Tube. Load then Acquire/Record your next single-stained control. 9

Once all have been recorded, review/adjust the placement of the snap-to gate P2 for each control. Go to Experiment menu> Compensation Setup>Calculate Compensation Select Apply Only and click OK. Toggle back to the Global Worksheet (Worksheet View button). Review and Enable Compensation Running Samples Place the tube pointer on Tube_001. While Tube_001 is selected, go to the Inspector window>label tab and input the names of your antibodies. Note: If naming multiple tubes simultaneously, use the Experiment Layout (Experiment menu> Experiment Layout ). Rename Tube_001. Create Plots & Gates in the Global Worksheet. Note: Apply a gate to a plot by R-clicking a plot and selecting Show Populations> Name of the Gate or by using the Inspector window (plot tab). Set the Events To Record and the Stopping Gate on the Acquisition Dashboard or in the Experiment Layout (Acquisition tab). Note: To display the name of the applied gate on the header of a plot: click on a plot, go to the Inspector window (title tab) and check the Populations box. R-click on any of your plot and select Show Population hierarchy to visualize simple statistics. Note: To display data using Biexponential display: click on a plot, go to the Inspector window (plot tab) and check the Biexponential X and Y boxes. Load then Acquire/Record your samples (FMOs, biological controls, and experimental samples). 10

Exporting Data Close your experiment (double-click on header) R-click Experiment>Export>FCS files [FCS 3.0] Select Location (path) R-click Experiment>Export>Experiment Template (if desired add information pertaining to your experiment) R-click Experiment>Delete Shutdown Be considerate of others, start cleaning ~9 minutes before the end of your scheduled time 1. Install a tube with 3mL of 1% Contrad with support arm to the side for 1 minute 2. Put support arm under the tube and RUN on HI for 2 minutes 3. Install a tube with 3mL of FACSClean with support arm to the side for 1 minute 4. Put support arm under the tube and RUN on HI for 2 minutes 5. Install a tube with 3mL of dih 2O with support arm to the side for 1 minute 6. Put support arm under the tube and RUN on HI for 2 minutes 7. Leave a tube with ~1ml of dih 2O with support arm under 8. Switch cytometer to STANDBY mode 9. Log off FACSDiva then Quit 10. Turn off lasers 11. Log out of Windows Last user turn off cytometer & FACSFlow Supply System Troubleshooting Coherent Connection not visible Alt + Space, Maximize Coherent Connection not displaying all lasers Tools > Serial Settings > Scan Ports or turn the cytometer off then on Coherent Connection not displaying the 488 laser Open the LSR II lid: if there is blue light, the laser is on. UV laser (355nm) not firing up Set the laser power to 5mW and turn On, then increase to 100mW Cytometer not connected Turn the cytometer off then on Waste Alarm Take the waste lid off and wipe the silver probe dry Sample Not Running due to an empty reservoir Fill the sheath tank, restart, fill the reservoir (halfway), and bleed the sheath filter & side tubing 11