BCD Interface for the 520 Indicator. Installation and Configuration Manual

Similar documents
BCD Interface for the 520 Indicator DISCONTINUED. Installation and Configuration Manual Rev A

DMC-688. When Accuracy Counts

R-30 Price Computing Retail Scale. Version 2. Installation and Service Manual

I/O Module. For Intrinsically Safe Systems Version 2.0. Installation Manual

PC 150. Digital Bench Scale. Operation Manual

Programmable HMI Indicator/Controller. In-Motion Checkweighing System Manual

Ethernet. Interface for 520 and 920i HMI Digital Indicators Version 2.0. Installation and Configuration Manual

Operation Instructions

EtherNet/IP. Interface for 520, 720i, 820i and 920i Indicators. Installation and Programming Manual

EtherNet/IP. Interface for 520 and 920i Indicators. Installation and Programming Manual

PC-150. DIGITAL BENCH SCALE Setup / Operation Manual

DISCONTINUED. Profibus DP. Installation and Programming Manual. Version 1.0

Profibus DP. Installation and Programming Manual. Version 1.0. Indicator Interface Indicator Interface for IQ plus 510 and IQ plus 710 Indicators

CS Series. Digital Counting Scale. Operation Manual

P1 PowerStation Backlight Bulb

ControlKeeper 4. General Information. Connecting Relay Loads. Installation Sheet. Getting Started. Power Supply Wiring. Mounting the Cabinet

Secured Series: Hub Plus Kit Single Door Controller Package Installation Manual

EASON TECHNOLOGY. IO8 & IO24 Break-Out Module

MPP200 User s Manual

PV3500. Fuel Site Controller. Service Manual OPW Fuel Management Systems Manual M Rev. 1

COBALT C INSTALLATION GUIDE RFID CONTROLLER ESCORT MEMORY SYSTEMS. High Frequency Passive Radio Frequency Identification Controller

ADAM Fiber Optic to RS-232/422/485 Converter

Crane Scale. Revision 1.4 February 18, Transcell Technology, Inc. Contents subject to change without notice.

CC186 AND CC186/2 STAND-ALONE OR SYSTEM CLOCK

OPT. Outdoor Payment Terminal. Service Guide OPW Fuel Management Systems Manual No. M Rev. 2

Resolver to Digital Expansion Board

16/24 Port 10/100 Auto-Sensing Dual Speed Ethernet Switch

HUBBCOM GSC3000/GSC4000 Flush-Mount Smart Controller Quick Installation Guide

Broadband Automatic Disconnect Switch. User Manual

INSTRUCTION MANUAL. Sensoray Model 720RB/DIN. Relay I/O Board (Rev A) October 12, 2001

Digital Scale. Operation Manual. Revision 1.1 April 16, Transcell Technology, Inc. Contents subject to change without notice.

Pit Transition with 24" Fan Installation and Operators Instruction Manual

PSA200 User s Manual

User Guide. Control Box. RoscoLED TM.

AUTOMATION. Operator s Manual. PG5 Digital Pressure Gauge

Omnitron Systems Technology, Inc. 1. iconverter. 19-Module Managed Power Chassis User s Manual

PowerFlex DC Stand-Alone Regulator (SAR) Power Interface Circuit Board

FA-2448 SIX POSITION Filter Wheel

WARRANTY & LIMITATION OF LIABILITY

MODBUS MANUAL For the AccUView UV Transmission Monitor

SERVICE MANUAL MODEL SSW-521-D2-ADA

DI-148U. User's Manual. The way PC-based instrumentation should be. 8-Channel USB Data Acquisition Starter Kit. Manual Revision B

Pulse LED Instruction Guide

The Butterfly LaserMount also offers all the features you would expect from a modern butterfly laser diode fixture, including:

iconverter 2-Module Power Chassis

MPP1700 User s Manual

WARRANTY & LIMITATION OF LIABILITY

Data Reporter. Installation-Operation E rev.f

The power behind competitiveness. Delta Infrasuite Power Management. Power Distribution Unit. User Manual.

AUTOMATION. Operator s Manual PG-5. Full Access. DOC Rev. A1 3/10

Installing the A4504A PMC Bridge Adapter and A4509A Expansion Adapter. HP Part No. A Edition E1197 Printed in U.S.A.

Replacing the PanelMate Power Pro 1785 Series, PanelMate epro 7585x-8 and 7685x-8 Series Backlight Assembly

Phase Loss Protection Upgrade. Phase Loss Protection Upgrade. In this bulletin:

3710 ACM 3750 PDC 3800 RTU. ISOCOM Communications Card Retrofit Instructions

gbox Ethernet to GPI Interface Users Manual

INSTRUCTION MANUAL. Models: H100 Light Meter

Rev. A. ANC Series RS-485/RS-422 Synchronous Clock Display. Antona Corporation (818) URL:

Integrated Stepper Drive & Motor

USER MANUAL CCHP-6000-XX 6kJ/s Capacitor Charging Power Supply

Reference Manual EPM-PS1. PC/104-Plus Power Supply Module DOC. REV. 4/29/2009

SySTIUM TECHNOLOGIES. Assembly Guide. Model 133i

HVPS-C. High Voltage Power Supply. HVPS-C. Package Contents: Sold Separately. Table of Contents:

Instruction and Operation Manual. CRMS Contact Resistance Measurement System. Model Number Caution!

T R A N S F O R M I N G T E C H N O L O G I E S, L L C O U T S T A N D I N G A L T E R N A T I V E S I N S T A T I C C O N T R O L

Mini-DVI-WP Series WALL PLATE MINIATURE MULTIMODE FIBER OPTIC DVI TRANSMISSION SYSTEM

Series CTA CAP TORQUE TESTERS. User s Guide

Service Manual For. -IP Voice over Internet Protocol -ACH** Handset cord length Inches

SOLVES-IT! 4 CHANNEL TIMER APPLICATION MODULE

Model 815 User s Manual

SATA II HDD Canister KISS DA 435 Quick Reference Guide

ACLD Channel Opto-isolated Digital Input Board

TC1880 Series. 4/5/6/8 Channel RS-232 FIBER OPTIC MICRO MUX User's Manual

Installation Note. Configurable Test Set Upgrade Kit. For PNA Series RF Network Analyzers (E8801A, E8802A, and E8803A) Network Analyzer Model Number

Thank you for choosing the DIL LaserMount from Arroyo Instruments. The DIL LaserMount is designed for high performance and long term use.

Owner s Manual. TSD-DCPDV DC Power Distribution with Fixed & Variable Outputs. TSD-DCPDV DC Power Distribution. AtlasIED.com

FNet Repeater Installation & Operator s Instruction Manual

User Guide. DMXSM-H18 Splitter. DMXSM-H81 Merger. Document Release Aug Revision B

INSTALLATION INSTRUCTIONS

DCS-E 1kW Series, DLM-E 3kW & 4kW Power Supplies

AC300/AC400 SERIES DYNAMIC BRAKING and ADDITIONAL FORM C RELAY. INSTALLATION INSTRUCTIONS Document Number:

EDG Port Industrial 10/100 Mbps Ethernet Switch. User Manual

The HPV Explorer is a serial communication program used to work with the HPV 600/900 s parameters via a computer s serial port

Dear Customer, User Memo: Please visit us on facebook or twitter! Thank you for purchasing this product.

User s Guide. Series CT CLOSURE TORQUE TESTER. User s Guide

Seven Channel Remote Control

LB-ICP LINK BRIDGE TM INLINE CONTROL PROCESSOR

Zonit μats TM Users Guide μats1-lv Version 1.2

CubePro. Main PCB Replacement Guide. Prosumer 3D Printer. Original Instructions

CTI 3570A SERIES INDUSTRIAL ETHERNET SWITCH INSTALLATION AND OPERATION GUIDE Version 1.2

CA-A480-A Elevator Controller. Reference & Installation Manual

Tornado F-5. Motor Backspin Detect Relay And Probe For F-5 Motor Controller

Basketball Shot Clock Set LX2180 Manual

SERVICE MANUAL MODEL SSW-520-F

USER S MANUAL ENTITY. Energy Meter TRINITY. Entity Operational Manual

EPX Touchscreen and Display Replacement Assemblies

OPERATIONS MANUAL PCM-I/O48

Replacing Preamplifier Circuit Boards

Relay Board. User Manual. 1.0, Oct 2013

RESIDENTIAL OPERATOR MOTOR CONTROL BOARD REPLACEMENT INSTRUCTIONS

Model No. ST200-S-RCP

Transcription:

BCD Interface for the 520 Indicator Installation and Configuration Manual 7627

Contents About This Manual....0 Introduction... 2.0 Installation... 2 2. BCD Option Card Installation....................................................... 2 2.2 Dip Switch Configuration.......................................................... 4 3.0 Data Information... 6 3. Operation of Remote Switch Inputs.................................................. 9 3.2 Floating Decimal Point........................................................... 3.3 Polarity Switch................................................................. 4.0 Parts List... 2 BCD Option Limited Warranty... 3 Technical training seminars are available through Rice Lake Weighing Systems. Course descriptions and dates can be viewed at www.ricelake.com or obtained by calling 75-234-97 and asking for the training department. 2008 Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America. Specifications subject to change without notice. January 2008

2 BCD Interface Installation and Configuration Manual

About This Manual This manual is intended for use by service technicians responsible for installing the BCD option in 920i digital weight indicators. This manual applies to indicators using Version.04 of the 920i software. Some procedures described in this manual require work inside the indicator enclosure. These procedures are to be performed by qualified service personnel only..0 Introduction The BCD interface is used with 520 digital weight indicators to send weight and status information to remote displays, printers and other peripheral devices. The data is updated after each display update and continuously available for use. The BCD interface option is CMOS, NMOS, and TTL compatible. BCD outputs are capable of sourcing/sinking 20mA. The BCD option comes with a BCD interface option card and ribbon cable. An external port provides for connection of a DB-37 interface cable (which exits the back of the 520) for communication to a peripheral device. See the 520 Installation Manual (PN 68973) for proper connections on the 520 CPU board. The BCD (binary-coded decimal) option uses a straight assignment of the binary equivalent. The weights of the BCD code are 8, 4, 2, and. See Table - for decimal to binary conversion. BCD Decimal Digit 8 4 2 0 0 0 0 0 0 0 0 2 0 0 0 3 0 0 4 0 0 0 5 0 0 6 0 0 7 0 8 0 0 0 9 0 0 Table -. Decimal to Binary Conversion Introduction

2.0 Installation Figure 2- shows the BCD option card as installed in the 520 indicator. The indicator enclosure must be open to install the BCD card. The 520 has no on/off switch. Before opening the unit, ensure the power cord is disconnected from the power outlet. Use a wrist strap to ground yourself and protect components from electrostatic discharge (ESD) when working inside the indicator enclosure. These units use double pole/neutral fusing which could create an electric shock hazard. Procedures requiring work inside the indicators must be performed by qualified service personnel only. 2. BCD Option Card Installation. Disconnect the 520 indicator from power source. 2. Place the indicator on an antistatic work mat. Remove screws that hold the backplate or cover to the enclosure body, then lift the backplate or cover away from the enclosure and set it aside. 3. Use the 8-32 nuts in hardware kit to secure DB-37 connector plate to backplate of indicator. BCD Option Board Configuration Dip Switch Figure 2-. 520 Indicator with BCD Option Installed 4. Carefully align the large option card connector (J) with connector J2 on the 520 CPU board. Press down to seat the option card in the CPU board connector. 5. Use the 4-40 screws provided in the option kit to secure the other end of the option card to the threaded standoffs on the 520 CPU board. 6. Plug 40-pin ribbon cable to J4 on BCD card. Do not connect communications to J3 if switch SW-4 is in Units mode. 2 BCD Interface Installation and Configuration Manual

7. Plug serial cable to J3 on BCD card. Plug other end of serial cable to J8 on 520 CPU board. Use the nylon tie in the parts kit to secure the serial cable to the mounting pem (see Figure 2-2 below). NOTE: When connecting serial cable to J8 of 520 CPU board, be sure that red wire goes to TX (transmit), black wire to GND (ground), and green wire to RX (receive). Figure 2-2. Serial Cable Tie Snaps onto the 520 CPU board. (J2) J 34 33 2 2 2 3 4 5 6 7 8 OPEN R6 D L D 2003 RICE LAKE WEIGHING SYSTEMS C2 U R7 C25 C C6 C24 R8 C2 U6 C7 R R2 U2 R3 C6 C7 C3 VR J2 ISP 0 C8 C9 VR2 U3 3.3V C3 R R2 R3 C20 C4 R 4 C2 R4 R 5 C4 R5 2 3 4 5 6 7 8 2 3 4 5 6 7 8 OPEN C26 5V R6 C8 C9 C0 C22 U4 SW U5 U7 U8 J3 Tx Gnd Rx C5 C C5 C23 P/N Revision GD GND J4 39 40 R9 R0 R7 R8 R9 Plug in ribbon connector to the back of the 520 indicator. SW Configuration Dip Switch Figure 2-3. BCD Option Card Installation 3

2.2 Dip Switch Configuration The BCD option card contains dip switches (see Figure 2-3 on page 3). Move the switches to configure BCD interface parameters (see Table 2-). SW- SW-2 SW-3 SW-4 SW-5 SW-6 SW-7 SW-8 DEC_ DEC_2 POL_SEL Units/Com Select Units Input Select Tri-State Select Reserved Reserved Table 2-. Dip Switch Configuration Modes Unit Status Bit/Communications Output Selection See Table 2-2 for unit status bit/communications output select configuration. SW-4 Mode 0 Unit Status Bit mode Indicates primary or secondary Communications Select mode RS-232 Table 2-2. Units Status Bit/Communications Select Mode Configuration By connecting J3 on the BCD board to J8 on the 520 board, communications mode allows the throughput of connector J3 to be used as a wiring path for communications from the 520 indicator to an external device through the DB-37 connector. It is not a communications port, rather an alternative route for wiring a 520 communication port to an external device. Do not connect communications to J3 if switch SW-4 is in Units mode. BCD Option Card 520 Indicator J3- TX J8- TX J3-2 Gnd J8-2 Gnd J3-3 RX J8-3 RX Table 2-3. J3/J8 Connections Units/Tri-State Input See Table 2-4 below, for information on configuring units or tri-state input modes. SW-5 SW-6 Mode 0 0 Not valid 0 Units enable 0 Tri-State enable Off Table 2-4. Units/Tri-State Input Configuration 4 BCD Interface Installation and Configuration Manual

Decimal Rotation Configuration See Table 2-5 below for information on configuring dip switch for decimal rotation. See Section 3.2 on page for more information on decimal rotation. SW- SW-2 Parallel Output Format 0 0 No rotation 0 Rotate left enable 0 Rotate right enable (primary only) Reserved Polarity Bit Select Table 2-5. Decimal Rotation Configuration See Table 2-6 below, for information on configuring polarity bit select mode. See Section 3.3 on page for more information on polarity switch. SW-3 State + Polarity (logic = +5V) ; Polarity (logic 0 = common) 0 + Polarity (logic 0 = common); Polarity (logic = +5V) Table 2-6. Polarity Bit Select Configuration Installation 5

3.0 Data Information The following sections provide information on BCD pin-outs, data information, and remote switches. DB-37 Leading MSD 2 (Most Significant Digit) 4 8 2 4 8 2 4 8 2 4 8 LSD 2 (Least Significant Digit) 4 8 Polarity Bit Zero Bit Print Enable Bit Data Valid Bit Overload Bit Motion Bit Units Bit/TX Serial Output Gross/Net Bit Gross/Net Switch Input Zero Switch Switch Input Hold Switch Switch Input Auto Tare Switch Input Unit Select (or Tri-State Enable) Switch Input Serial Input Logic Common Logic Common 4th SD 3rd SD 2nd SD 0 2 4 6 8 8 7 6 20 24 23 22 2 32 27 26 25 3 30 29 28 9 3 5 7 9 5 35 34 37 4 33 3 36 2 Logic 0 = See Section 3.3 on page 0 Logic = Center of Zero Logic = Valid Print Logic 0 = Data Registers Changing Logic 0 = Overload/Underload Logic 0 = Scale in Motion Logic = Secondary or TX Serial Output: if dip switch configuration Logic = Net Mode is pr operly selected See Figure 3-2 RX Serial Input: if dip switch configuration is properly selected. No connect otherwise Digital Common (Not Chassis or Earth Ground) See note below Figure 3-. BCD External Connection NOTE: The 520 parallel BCD option does not have a floating ground. For proper operation, it is necessary to connect Logic Common of the BCD peripheral device to Logic Common pins and 2 of the BCD option. 6 BCD Interface Installation and Configuration Manual

Pin Description Logic Common 2 Logic Common 3 Units (primary/secondary) Switch or Tri-State Enable Input. Serves as remote input switch for the Units or Tri-State function. Change of state toggles units between primary or secondary; or Logic places output in Tri-State mode. See Section 2.2 on page 4 for dip switch configuration 4 Hold Switch. Input freezes the BCD registers 5 Units Bit. Specifies the status of the units; Logic = secondary units, or if enabled by dip switch configuration, accesses TX serial output 6 4th SD -4 7 4th SD -2 8 4th SD - 9 Polarity Bit. Specifies the status of the polarity. See Section 3.3 on page 0 Leading One. Logic if the sixth significant digit is odd; Logic 0 if it is even. See Figure 3- on page 6 Zero Bit. Specifies the status of zero. Logic = center of zero 2 MSD 3 Print Enable Bit. Specifies data valid, no motion, no overload/underload and positive gross weight. Hold switch must be at logic (OFF). Logic = print enabled 4 MSD 2 5 Data Valid Bit. Informs the external device that the BCD output registers have been updated and the data is true. Data is not valid if the display shows other than numeric data, if the display is held (in regulatory modes), or if data integrity has been lost. This bit will toggle momentarily to Logic 0 each time the registers are updated. Logic 0 = data not valid 6 MSD 4 7 Overload Bit. Specifies the status of positive or negative scale overload. Logic 0 = overload/underload 8 MSD 8 9 Motion Bit. Specifies the status of motion on the scale. Logic 0 = scale in motion 20 4th SD 8 2 3rd SD 8 22 3rd SD 4 23 3rd SD 2 24 3rd SD 25 2nd SD 8 26 2nd SD 4 27 2nd SD 2 28 LSD 8 29 LSD 4 30 LSD 2 Data Information 7

Pin Description 3 LSD 32 2nd SD 33 Auto Tare Switch Input. A remote switch for the auto tare function 34 Gross/Net Switch Input. A remote switch for the gross/net function. Signals the indicator to switch between gross and net modes 35 Gross/Net Bit. Specifies the status of the mode of the scale. Logic = Net mode 36 RX serial output, if enabled by dip switch configuration for communication mode 37 Zero Switch Input. A remote switch for the zero function 8 BCD Interface Installation and Configuration Manual

3. Operation of Remote Switch Inputs DB-37 Zero Switch 37 ZERO Gross/Net 34 GROSS/NET Auto Tare 33 AUTO TARE Hold Units Select (primary/secondary) or Tri-State Enable Logic Common 4 3 OFF ON HOLD (If on, this freezes the BCD registers) OFF ON UNITS SELECT (toggles between primary and secondary) or TRI-STATE Output Enable (if Tri-State configuration switch is enabled, On position enables Tri-State mode). Logic Common 2 Figure 3-2. BCD Remote Switches Three of the front panel switches (zero switch, gross/net switch, and auto tare switch), can be operated from a remote location using momentary-action pushbuttons. Note that each push of a remote switch activates the associated function once. The function of the gross/net switch is to change the state of output between gross and net mode. An alternate action switch is suggested to operate the Hold and Units Select/Tri-State Enable. When using the BCD option in Units mode, the switch state is independent of the front panel and toggles between primary and secondary units if any change of state is detected. Tri-State output enable is used to allow other peripheral devices to use the same BCD output. To place the unit in Tri-State mode, set dip switches according to Table 2-4 on page 4. All switch inputs are tied to 5V through pull-up resistors on the BCD option card to ensure reliable operation. NOTE: Connect the pins/switches to Logic Common only, not case ground or other grounds. Data Information 9

Good engineering practices suggest: Switches should be placed no more than six feet (two meters) from the indicator. Cables for low level digital signals should not be placed near high voltage sources or equipment that emits high levels of EMI/RFI. 9 8 7 6 5 4 3 2 0 9 8 7 6 5 4 3 2 37 36 35 34 33 32 3 30 29 28 27 26 25 24 23 22 2 20 2 3 4 5 6 7 8 9 0 2 3 4 5 6 7 8 9 20 2 22 23 24 25 26 27 28 29 30 3 32 33 34 35 36 37 520 ribbon cable header on BCD option 20 2 2 3 22 4 23 5 24 6 25 7 26 8 27 9 28 0 29 30 2 3 3 32 4 33 5 34 6 35 7 36 8 37 9 DB-37 connector connects to the backplate of the 520 indicator DGND (Logic Common) 4SD MSB (8) DGND (Logic Common) 3SD MSB (8) Units/Tri-State enable 3SD (4) Hold bit 3SD (2) Units bit/tx serial output 3SD LSD () 4SD (4) 2SD MSB (8) 4SD (2) 2SD (4) 4SD () 2SD (2) Polarity LSD MSB (8) LSD () LSD (4) Zero bit LSD (2) MSD LSB () LSD LSB () Print enable 2SD LSB () MSD (2) Auto tare Valid data Gross switch MSD (4) Net Overload No connect/rx serial MSD MSB (8) Zero switch Motion bit Figure 3-3. BCD Header to DB-37 Connector Pin-outs Option 0 BCD Interface Installation and Configuration Manual

3.2 Floating Decimal Point A BCD receiving device must assume the decimal point in a single location for either units (no decimal point is actually sent on the BCD lines). The BCD offers several configurations for aligning the decimal point. Refer to Table 2-5 on page 5 for dip switch settings. With switch SW- closed (logic 0), the decimal point is rotated left. The BCD option automatically adjusts for the floating decimal point to accommodate units-switching on fixed decimal point devices. It adjusts the decimal point of the units with fewer decimal digits, to match that of the units with the most decimal digits. Example: 272 lb transmits as 2720 is interpreted as 272.0 lb 234.5 kg transmits as 2345 is interpreted as 234.5 kg The receiving device interprets the decimal point at one place. Switch SW-2, when closed (logic 0), shifts the weight data of the primary units to the right one place (if it is configured with a fixed zero), dropping the least significant digit. This function is primarily used to condense the weight data to five and one-half digits by eliminating the fixed zero from the parallel output. Example: Displayed Weight Parallel Output 5th SD 4th SD 3rd SD 2nd SD LSD SD Fixed Zero on Receiving Device 252350 lb 2 5 2 3 5 0 NOTE: With both switches closed, no rotation is performed. Table 3-. Floating Decimal Point Example 3.3 Polarity Switch The BCD option returns polarity on pin 9. The polarity logic may be reversed by setting switch SW-3. See Table 2-6 on page 5 below for information on polarity switch configuration. Data Information

4.0 Parts List See Table 4- below for a list of parts for the BCD option for the 520 indicator. Part Number Part 7548 BCD plate 7655 Cable, 40 pin assembly 55947 37-pin mating connector 5497 Cable, 25-conductor 24AWG Table 4-. BCD Option Parts List 2 BCD Interface Installation and Configuration Manual

BCD Option Limited Warranty Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in materials and workmanship for two years. RLWS warrants that the equipment sold hereunder will conform to the current written specifications authorized by RLWS. RLWS warrants the equipment against faulty workmanship and defective materials. If any equipment fails to conform to these warranties, RLWS will, at its option, repair or replace such goods returned within the warranty period subject to the following conditions: Upon discovery by Buyer of such nonconformity, RLWS will be given prompt written notice with a detailed explanation of the alleged deficiencies. Individual electronic components returned to RLWS for warranty purposes must be packaged to prevent electrostatic discharge (ESD) damage in shipment. Packaging requirements are listed in a publication, Protecting Your Components From Static Damage in Shipment, available from RLWS Equipment Return Department. Examination of such equipment by RLWS confirms that the nonconformity actually exists, and was not caused by accident, misuse, neglect, alteration, improper installation, improper repair or improper testing; RLWS shall be the sole judge of all alleged non-conformities. Such equipment has not been modified, altered, or changed by any person other than RLWS or its duly authorized repair agents. RLWS will have a reasonable time to repair or replace the defective equipment. Buyer is responsible for shipping charges both ways. In no event will RLWS be responsible for travel time or on-location repairs, including assembly or disassembly of equipment, nor will RLWS be liable for the cost of any repairs made by others. THESE WARRANTIES EXCLUDE ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WITHOUT LIMITATION WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NEITHER RLWS NOR DISTRIBUTOR WILL, IN ANY EVENT, BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES. RLWS AND BUYER AGREE THAT RLWS S SOLE AND EXCLUSIVE LIABILITY HEREUNDER IS LIMITED TO REPAIR OR REPLACEMENT OF SUCH GOODS. IN ACCEPTING THIS WARRANTY, THE BUYER WAIVES ANY AND ALL OTHER CLAIMS TO WARRANTY. SHOULD THE SELLER BE OTHER THAN RLWS, THE BUYER AGREES TO LOOK ONLY TO THE SELLER FOR WARRANTY CLAIMS. NO TERMS, CONDITIONS, UNDERSTANDING, OR AGREEMENTS PURPORTING TO MODIFY THE TERMS OF THIS WARRANTY SHALL HAVE ANY LEGAL EFFECT UNLESS MADE IN WRITING AND SIGNED BY A CORPORATE OFFICER OF RLWS AND THE BUYER. 2008 Rice Lake Weighing Systems, Inc. Rice Lake, WI USA. All Rights Reserved. RICE LAKE WEIGHING SYSTEMS 230 WEST COLEMAN STREET RICE LAKE, WISCONSIN 54868 USA Parts List 3