Calibration of Optical Rotary Tables using Autocollimators and Laser Interferometer Systems

Similar documents
Operation Manual. Autocollimator Test Wedge for quick testing of Electronic Autocollimators

A U T O C O L L I M A T O R S

PROCESSING OF CALIBRATION MEASUREMENTS ON EZB-3

A NEW SYSTEM FOR COMPARISON CALIBRATION OF VIBRATION TRANSDUCERS AT LOW FREQUENCIES

A Procedure for accuracy Investigation of Terrestrial Laser Scanners

Contents 1 Measurement and Machine Tools An Introduction

The Virtual CMM a software tool for uncertainty evaluation practical application in an accredited calibration lab

Guidelines on the Calibration of Autocollimators. EURAMET Calibration Guide No. 22 Version 1.0 (07/2017)

PrismMaster. Ultra Accurate and Fully Automated Prisms and Polygons Measurement

Good Practice guide to measure roundness on roller machines and to estimate their uncertainty

Vertical angle calibration of a theodolite

Guidelines on the Calibration of Angular Encoders. EURAMET Calibration Guide No. 23 Version 1.0 (02/2018)

ACTA TECHNICA CORVINIENSIS Bulletin of Engineering Tome X [2017] Fascicule 2 [April June] ISSN:

Name Section Date. Experiment Reflection and Refraction

3. Automatic control of multireference equipment for calibration of angle measuring instruments

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

TriAngle. Electronic Autocollimators for Precise Angle Measurement

CALCULATION OF 3-D ROUGHNESS MEASUREMENT UNCERTAINTY WITH VIRTUAL SURFACES. Michel Morel and Han Haitjema

Comparison between 3D Digital and Optical Microscopes for the Surface Measurement using Image Processing Techniques

COMPARISON OF ERROR MAPPING TECHNIQUES FOR COORDINATE MEASURING MACHINES USING THE PLATE METHOD AND LASER TRACER TECHNIQUE

Light and refractive index

Measuring object. Fig. 1: Optical scheme of focus-touching (concentricity measurement with measuring ball)

Horizontal scale calibration of theodolites and total station using a gauge index table

Length, Germany, PTB (Physikalisch-Technische Bundesanstalt)

Monte Carlo method to machine tool uncertainty evaluation

A Novel Kinematic Model of Spatial Four-bar Linkage RSPS for Testing Accuracy of Actual R-Pairs with Ball-bar

RELIABILITY OF PARAMETRIC ERROR ON CALIBRATION OF CMM

METHODS FOR PERFORMANCE EVALUATION OF SINGLE AXIS POSITIONING SYSTEMS: POINT REPEATABILITY

MICHELSON S INTERFEROMETER

Mr. Chaobin Li China Lucky Film Corp. No. 6 Lekai South Street 6, Baoding, Hebei Province, PRC Tel.:

Fast scanning method for one-dimensional surface profile measurement by detecting angular deflection of a laser beam

ANTENNA ALIGNMENT IN A NEAR-FIELD FACILITY

Mirror selection for optical periscopes

IND62 TIM CALIBRATION OF FREE-FORM STANDARD AND THEIR APPLICATIONS FOR IN-PROCESS MEASUREMENT ON MACHINE TOOLS

Highly Efficient Assembly of Lenses with. OptiCentric and OptiCentric Cementing

: Imaging Systems Laboratory II. Laboratory 2: Snell s Law, Dispersion and the Prism March 19 & 21, n 1 n 2

Computer aided error analysis for a threedimensional precision surface mapping system

A RADIAL WHITE LIGHT INTERFEROMETER FOR MEASUREMENT OF CYLINDRICAL GEOMETRIES

The development of a laser array measurement system for three dimensional positioning testing in machine tool

good check of volumetric accuracy. However, if the mea error components. However, if the errors measured are large,

How to Measure Wedge. Purpose. Introduction. Tools Needed

SIMULATION AND VISUALIZATION IN THE EDUCATION OF COHERENT OPTICS

Keysight Technologies 5530 Laser Calibration System Components

Determination and compensation of the reference surface from redundant sets of surface measurements

A FEW CASE STUDIES IN UNCERTAINTY USING THE NIST M48 CMM

Evaluating the Performance of a Vehicle Pose Measurement System

Grade 6 Curriculum and Instructional Gap Analysis Implementation Year

arxiv: v1 [physics.ins-det] 21 Aug 2009

Calibration of Angle Standards Technical Protocol (Final) Prepared by NML

SWING ARM OPTICAL CMM

Introduction to Metrology

Centration of optical elements

Development of shape measuring system using a line sensor in a lateral shearing interferometer

XL-80 spares list. Application note. Overview

Calibration of Inertial Measurement Units Using Pendulum Motion

Exp No.(9) Polarization by reflection

Reflection and Refraction of Light

Matija Gubec International School Zagreb MYP 0. Mathematics

AN EVALUATION OF NON-CONTACT LASER SCANNER PERFORMANCE USING CALIBRATED ARTIFACTS

Resource Inspections Canada Inc. Certification Scheme Requirements

Accuracy for CMMs and Machine Tools

Study on Gear Chamfering Method based on Vision Measurement

Autocollimator Calibration for Synchrotron Metrology: Advancing from Plane to Solid Angle

Study on Digitized Measuring Technique of Thrust Line for Rocket Nozzle

PROTOTYPE OF LOAD CELL APPLICATION IN TORQE MEASUREMENT

The Uncertainty of Parallel Model Coordinate Measuring Machine

Operation Trajectory Control of Industrial Robots Based on Motion Simulation

Calibration for Dynamic Balancing Test System of Motorcycle Crankshaft

DESIGNER S NOTEBOOK Proximity Detection and Link Budget By Tom Dunn July 2011

Evaluation strategies in CT scanning

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY

GEOMETRICAL SURVEY OF COMPACT ANTENNA TEST RANGES USING LASER TRACKER TECHNOLOGY

CyberGage360 3D SCANNERS. One-Button Automation for 3D Scan Inspection.

ASSESSMENT OF MEASUREMENT UNCERTAINTY CAUSED IN THE PREPARATION OF MEASUREMENTS USING COMPUTED TOMOGRAPHY

Length, Thailand, NIMT (National Institute of Metrology (Thailand))

Length, United States, NIST (National Institute of Standards and Technology)

BenchMike 283 Series. Off-Line Diameter & Ovality Measurement System. for Wire and Cable Samples

7 Fractions. Number Sense and Numeration Measurement Geometry and Spatial Sense Patterning and Algebra Data Management and Probability

Mathematics. Scheme of Work. Year 7. New Mathematics Framework

Laser diagonal testing

Agilent 10706B High Stability Plane Mirror Interferometer

Reflection, Refraction and Polarization of Light Physics 246

UNITEST - A new device for the static and dynamic testing of unconventional machine structures

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

round decimals to the nearest decimal place and order negative numbers in context

Quality assurance and calibration tasks in the scope of multi-sensor systems

Reflection, Refraction and Polarization of Light

Measurement of period difference in grating pair based on analysis of grating phase shift

Grade 7 Mathematics Performance Level Descriptors

Experiment 6. Snell s Law. Use Snell s Law to determine the index of refraction of Lucite.

Calibration of the depth-measuring device of Rockwell hardness testing machines

Industrial Metrology. Multisensor Measuring Machines O-INSPECT 322/442

ABSTRACT 1. INTRODUCTION

Apply multiplication facts to solve division problems.

Thickness Measurement of Metal Plate Using CT Projection Images and Nominal Shape

Integers & Absolute Value Properties of Addition Add Integers Subtract Integers. Add & Subtract Like Fractions Add & Subtract Unlike Fractions

INFLUENCE OF INITIAL SETUP OF PARTS BEFORE ROUGHNESS EVALUATION

PCMM System Specifications Leica Absolute Tracker and Leica T-Products

Part 1: Plane Mirrors!

Study on Tool Interference Checking for Complex Surface Machining

Transcription:

Calibration of Optical Rotary Tables using Autocollimators and Laser Interferometer Systems Prof. S. Zahwi, Prof. M. Amer, National Institute for Standard, Egypt; Prof. M. A. Abdo, Al-Azhar University, Faculty of Engineering, Egypt; Researcher Assistance A. El-Melegy, National Institute for Standard, Egypt Keywords: Angle, Measurement, Rotary table. Abstract: Calibration of rotary tables can be performed using autocollimators or laser interferometer systems. In this investigation a comparison between these two methods of calibration as applied on an optical rotary table has been carried out. The calibrated table is a Carl Zeiss optical rotary table of resolution 30. An Elcomat 3000 autocollimator system and an Agilent 5529A laser interferometer system have been used in this calibration. A precise indexing table of 360 serrations with accuracy of 0.5 manufactured by Davis and a glass polygon of 12 faces with accuracy of 0.3 delivered by MOLLER WEDEL GmbH has been used. The indexing table is used in co-operation with the laser system while both of the polygon and the indexing table are used with autocollimator system. Repeated results from the different methods of calibration were carried out and presented in the paper. Uncertainties associated with the measurement results for each method have been estimated using the procedures mentioned in GUM. The paper discusses the accuracies as well as difficults and ease associated with each method. Introduction: Engineering measuring devices, such as rotary and indexing tables are used for angle measurement carried out in industry [1]. Calibration of rotary tables requires two basic components, a reference standard such as a precise polygon or a precision reference indexing table for generating a reference angle nominally equal to the angle being calibrated and an angle measuring instrument used as an angle comparator such as an autocollimator or laser interferometer system to measure small deviations in the measured angle from the reference angle [2]. Autocollimators have a long history of being used in metrology laboratories for calibration of angle standards, such as polygons, indexing tables, rotary tables and angle gauge blocks [2]. Rotary tables can be calibrated by autocollimator against a precision polygon; or against a precise indexing table using a reflecting mirror on the top of the table. Laser interferometer system can

10th IMEKO TC14 Symposium on also be used to calibrate rotary tables against precise indexing table with retroreflectors optics. In this investigation, a comparison of calibration results on an optical rotary table by the two systems is carried out. Associated uncertainties are estimated according to GUM [4] and [5]. Experimental Setup: An optical rotary table having a measuring range of 360 and a resolution of 30 is calibrated. This table is calibrated using two measuring systems, laser interferometer system Autocollimator head Polygon and autocollimator system. Each of the two systems has a 0.05 resolution. The calirotary table Display unit bration of rotary tables is based on angle standards, two angle standards, namely polygon and Autocollimator head an indexing table has been used. A calibrated precision optical Polygon of 12 faces calibrated Polygon Rotary table Display unit with uncertainty of 0.03 has been used. An indexing table of 360 range, angle step of 1 and calibrated with uncertainty of 0.6. The rotary table is calibrated at steps of 30 throughout its full range 360. The results of calibanti vibration table drawn in each case ration are corrected for the errors of angle standards used. Error charts are showing the error at each calibrated angle (steps). The calibrations are carried under environmental conditions Anti ofvibration 20 table C 1 C. Figure 1 shows the different experimental setups used. a) Autocollimator with the aid of polygon Autocollimator head Polygon Display unit a) Autocollimator with the aid of polygon Reflector Rotary table Reflector Indexing table Indexing table Air operated lifting fixture Anti vibration table Air operated lifting fixture a) Autocollimator with the aid of polygon a) Autocollimator with the aid of polygon b) Autocollimator with the aid of indexing Reflector table Display unit Laser head b) Autocollimator with the aid of indexing table b) Autocollimator with the aid of indexing table Display unit Laser head Interferometer Angular reflector Indexing table Interferometer Angular reflector Air operated lifting fixture b) Autocollimator with the aid of indexing table Display unit Laser head Interferometer Angular reflector Laser interferometer with the aid of indexing table c) Laserc)interferometer with the aid of indexing Figuretable 1: Calibration of rotary tables using different systems c) Laser interferometer with the aid of indexing table Figure 1: Calibration of rotary tables using different systems Fig. 1: Calibration of rotary tables using different systems

Results: An example of rotary table calibration result and associated error charts is shown in figure 2. The total error An Texample of the rotary of rotary table is table determined calibration in each result case. and The calibration associated is repeated error charts 10 times is by shown each system in figure on the 2. same The rotary total table. error The T θ standard of the rotary deviations table of the is determined repeated results in at each each case. measured The angle calibration are shown is repeated in figure 3. Table 1 shows the average total error T and the standard deviation of the scatter of results S(T ) as obtained from 10 the times repeated by each calibrations system using on each the same method. rotary Figure table. 4 represents The standard such results deviations represents such of the results. repeated results at each measured angle are shown in figure 3. Table 1 shows the average total error T θ and the standard deviation of the scatter of results S(T θ) as obtained from the repeated calibrations using each 8 method. Figure 4 represents such results represents such results. errors, errors in arcseco, in arcseco Maximum 6 8 4 Average Maximum 6 2 4 0 Average 2 0 30 60 90 120 150 180 210 240 270 300 330 360-2 0 T rotation angles, in degree -4 0 30 60 90 120 150 180 210 240 270 300 330 360-2 -6 T rotation angles, in degree -4-8 Minimum -6-8 Minimum Figure 2: An example of an error chart. Fig. Figure 2: An 2: An example of an error chart. Normalized standard deviation, 1S 1.8 1.4 1 0.6 0.2-0.2 0 30 60 90 120 150 180 210 240 270 300 330 360-0.6-1 -1.4 Angle, degree -1.8 Fig. Figure 3: Associated 3: Associated standard standard deviation deviation at at the the calibrated calibrated angles. angles. Table 1: Calibration results of the rotary table Table 1: Calibration results of the rotary table Table 1: Calibration results of the rotary table Calibration of of rotary table by by Average total error of table, T,, Standard deviation, deviation, S(T S(T ) ) Autocollimator Autocollimator with with polygon polygon 11.97 11.97 0.73 0.73 Autocollimator with indexing table 12.13 0.96 Autocollimator with indexing table 12.13 0.96 Laser interferometer system 11.98 0.98 Laser interferometer system 11.98 0.98

T T = 11.97 S(T ) = 0.73 Total angular error, T, 10 12 14 a) T T = 12.13 S(T ) = 0.96 Total angular error, T, 10 12 14 b) T T = 11.98 S(T ) = 0.98 Fig. 4: Results of rotary table as calibrated by; 10 12 14 Total angular error, T, a) Autocollimator system with polygon b) Autocollimator system with indexing table c) Laser interferometer c) system with indexing table Figure 4: Results of rotary table as calibrated by;

Uncertainty Estimation: The main factors affecting the calibration results and contributing to the uncertainty are mainly; uncertainty in the reference standard angle, accuracy of the angle comparators, uncertainty due to instabilities in environmental conditions, uncertainty due to eccentricity; uncertainty due to flatness of reflectors and uncertainty due to pyramidal effect. Mathematical Model; The measurand is the angles on the table under test (rotary table): = + s + m + i + e + + f + p ------- (1) Where; is angle on the rotary table (measurand). s is angle on the reference standard. is the difference between the measurand and the angle standard as measured by the angle comparator. m is the correction due to inaccuracy of the angle comparator i is the correction due to instabilities due to environmental conditions on the measurements e is the correction due to eccentricity in fixing the angle standard to the table under test. f is the correction due to flatness deviation of reflectors and polygon faces. p is the correction due to pyramidal error. The uncertainty at any angle based on GUM [4] and according to the mathematical model of equation 1, is: (by differentiation) u 2 ( ) = [ ( )/ ( )] 2 u 2 ( ) + [ ( )/ ( s )] 2 u 2 ( s ) + [ ( )/ ( m )] 2 u 2 ( m ) + [ ( )/ ( i )] 2 u 2 ( i ) + [ ( )/ ( e )] 2 u 2 ( e ) + [ ( )/ ( f )] 2 u 2 ( f ) + [ ( )/ ( p )] 2 u 2 ( p ) (2) u 2 ( ) = c 1 2 u 2 ( s ) + c 2 2 u 2 ( ) + c 3 2 u 2 ( m ) + c 4 2 u 2 ( i )+ c 5 2 u 2 ( e ) + c 6 2 u 2 ( f ) + c 7 2 u 2 ( p ) (3) Where, C i = 1 so; u 2 ( ) = u 2 ( s ) + u 2 ( ) + u 2 ( m ) + u 2 ( i ) + u 2 ( e )+ u 2 ( f ) + u 2 ( p )...(4) The total error of the rotary table; Where; T is the total error of the rotary table. max is the maximum error during calibration of the rotary table. min is the minimum error during calibration of the rotary table. The uncertainty in the total error of the rotary table; T = max - min. (5) u 2 (T ) = 2 u 2 ( ) (6) Substituting from equations 4 & 6, the uncertainty u(t ): u 2 (T ) = 2 [u 2 ( ) + u 2 ( s ) + u 2 ( m ) + u 2 ( i ) + u 2 ( e )+ u 2 ( f ) + u 2 ( p )]...(7) This equation is used in calculating the combined uncertainty of the total error in calibration of the rotary table. u( θ) can be obtained from the repeatability observed in the calibration at each angle. The maximum standard deviation during the 10 repeated calibrations is taken to represent s( θ) of individual measurement results. The standard deviation of the mean [s( θ)/ 10] is taken to represent repeatability in angle measurements. A summary of uncertainty budget in each case are shown in tables 2; 3 and 4

Table 2: Uncertainty budget for rotary table calibration by autocollimator system with the aid of polygon Type U(x i) Distribution. Sensitivity Coefficient Quantity x i Standard Uncertainty u(x i) c i [c iu(x i)] 2 Degree of freedom i Repeatability of measurements; A 0.82 Normal 0.41 1 (0.41) 2 9 Uncertainty in polygon; s B 0.30 Normal 0.15 1 (0.15) 2 Autocollimator accuracy; m B 0.05 Rectangular 0.03 1 (0.03) 2 Environmental instability; i B 0.10 Normal 0.05 1 (0.05) 2 59 Flatness of polygon s mirrors; f B 0.05 Rectangular 0.03 1 (0.03) 2 Pyramidal error; p B 0.31 Normal 0.16 1 (0.16) 2 27 Combined Standard Uncertainty (u c) (0.47) 2 eff = 16 Expanded Uncertainty (U): U = k u c = 2 0.47 = 0.94 Where, k is a coverage factor providing a level of confidence of approximately 95 %. For a normal distribution, k=2 Uncertainty in the total error of the rotary table (U T ): According to equation (7); U T = ( 2) U( ) = ( 2) (0.94) = 1.33 Table 3: Uncertainty budget for rotary table calibration by autocollimator system with the aid of the indexing table Type U(x i) Distribution Sensitivity [c iu(x i)] 2 Coefficient Quantity x i Standard Uncertainty u(x i) c i Degree of freedom i Repeatability of measurements; A 1.02 Normal 0.51 1 (0.51) 2 9 Uncertainty in Indexing table; s B 0.60 Normal 0.30 1 (0.30) 2 Autocollimator accuracy; m B 0.05 Rectangular 0.03 1 (0.03) 2 Environmental instability; i B 0.10 Normal 0.05 1 (0.05) 2 59 Flatness of reflecting mirror; f B 0.05 Normal 0.025 1 (0.025) 2 Eccentricity; e B 0.23 Normal 0.12 1 (0.12) 2 19 Pyramidal error; p B 0.31 Normal 0.16 1 (0.16) 2 27 Combined Standard Uncertainty (u c) (0.63) 2 eff = 21 Expanded Uncertainty (U): U = k u c = 2 0.63 = 1.26 Where, k is a coverage factor providing a level of confidence of approximately 95 %. For a normal distribution, k=2 Uncertainty in the total error of the rotary table (U T ): According to equation (7); U T = ( 2) U( ) = ( 2) (1.26) = 1.78 Table 4: Uncertainty budget for rotary table calibration by laser system with the aid of indexing table. Quantity Type U(x i) Distribution. Standard Sensitivity [c iu(x i)] 2 x i Uncertainty Coefficient u(x i) c i Repeatability of measurements; A 0.76 Normal 0.38 1 (0.38) 2 Uncertainty in Indexing table; s B 0.60 Normal 0.30 1 (0.30) 2 Laser system accuracy; m B 0.05 Rectangular 0.03 1 (0.03) 2 Environmental instability; i B 0.10 Normal 0.05 1 (0.05) 2 Eccentricity; e B 0.23 Normal 0.12 1 (0.12) 2 Combined Standard Uncertainty (u c) (0.51) 2 Expanded Uncertainty (U): U = k u c = 2 0.51 = 1.02 Where, k is a coverage factor providing a level of confidence of approximately 95 %. For a normal distribution, k=2 Degree of freedom i 9 59 13 eff = 29 Uncertainty in the total error of the rotary table (U T ): According to equation (7); U T = ( 2) U( ) = ( 2) (1.02) = 1.44 Discussion: A summary of the average total error in the rotary table T Ө and the uncertainty associated in each case are shown in figures 5.

Errors in arcsecon 16 14 12 10 8 Autocollimator, Polygon, 11.97, 1.33 Autocollimator, Indexing table, 12.13, 1.78 Laser, Indexing table, 11.98, 1.44 Fig. Figure 5: 5: Comparison comparison of the average results results of of the the rotary rotary table table error error and associated and associated uncertainties. uncertainties. The normalizing error E n (number of consistency) compares between measurement average values [6]. E n values for the results of the rotary table calibration by the different methods investigated are shown in table 5. It is found that E n < 1 in all cases. It could be said that according to the obtained results no significant differences were noticed between the average total errors of rotary table calibration by the two methods investigated.by the two methods investigated. Table 5: E n values for the rotary table Table 5: E n values for the rotary table Comparison between results of E n values Laser with indexing against Autocollimator with indexing 0.07 Laser with indexing against Autocollimator with polygon 0.01 Autocollimator with indexing against Autocollimator with polygon 0.07 The uncertainty contributors are divided into two types of uncertainties, type A (repeatability) and type B (other contributors) as shown in figure 6. Type A (repeatability) is estimated to be (58% 76%) of U(T Ө ). The scale division of the rotary table under test is 30 and by estimation one can estimate up to 6. High repeatability error could be due to this low resolution. The state of art of making rotary tables has improved a lot due to the use of rotary encoders and improving manufacturing techniques. It is now available in the market to get rotary tables used in workshop with resolution of about or less than 1. Analysis of the relative influences of the factors affecting type B evaluation of uncertainty is shown in figure 7. The uncertainty due to accuracy of reference standard accuracy affects type B in the range (49% 79%). The effect due to uncertainty of pyramidal error ranges up to 42%.

The effect due to eccentricity error ranges up to 17%. Other uncertainties (include instabilities, accuracy of instrument used and flatness deviation of reference angle) are relatively small. Uncertainty Types Uncertainty 0.1 m sec2 10000 1000 100 10 Type A Type B 1 Autocollimator & polygon Autocollimator & indexing Laser & indexing Type A 1681 2601 1444 Type B 465 1332 1140 Fig. 6: Types of uncertainties components associated by each method. Eccentricity error 0% Autocollimator & polygon Others 9% Pyramidal error 42% Reference accuracy 49% Autocollimator & indexing table Eccentricity error 15% Others 3% Pyramidal error 15% Reference accuracy 67% Laser & indexing table Eccentricity error 17% Others 4% Pyramidal error 0% Reference accuracy 79% Fig. 7: Pie charts for different contributors of type B evaluation of uncertainty in calibration.

All of the investigated methods (systems) are portable. The calibration of rotary tables by autocollimator system with the aid of polygon comes faster and less time consuming than the other two methods. On the other hand, the angle step of calibration of the rotary table by autocollimator system with the aid of polygon is limited by the number of polygon faces. Conclusion: 1. The averages of the results of the rotary table as obtained by different methods investigated are found to be statistically insignificance. 2. Either of laser interferometer system or autocollimator system is portable. So calibration of rotary table can be performed in calibration site. 3. The calibration of rotary tables by autocollimator system with the aid of polygon comes the easiest; require less time and less experience in application. However the angle step of calibration of the rotary table by autocollimator system with the aid of polygon is limited by the number of polygon faces.

References: [1] J. F. W. Galyer, C. R. Shotbolt, Metrology for Engineers, Cassell, London, third edition, ISBN 0 304 93434 8,1974. [2] J. C. Evans and C. O. Taylerson, Measurement of angle in engineering, HMSO, second edition,1964 [3] Jack A. Stone, Mohamed Amer, Bryon Faust, and Jay Zimmerman, Uncertainties in Small-Angle Measurement Systems Used to Calibrate Angle Artifacts, Journal of Research of the National Institute of Standards and Technology, vol. 109 (3), pp. 319-333, 2004. [4] ISO/IEC Guide 98-3, Uncertainty of Measurement Part 3: Guide to the Expression of Uncertainty in Measurement (GUM: 1995), 2008. [5] Ted Doiron and John Stoup, Uncertainty and Dimensional Calibrations, Journal of Research of the National Institute of Standards and Technology, vol. 102 (6), pp. 647-675, 1997. [6] ISO 13528 Statistical Methods for use in Proficiency Testing by Interlaboratory Comparisons.