Silicon Avalanche Photodiodes in Dynamic Light Scattering

Similar documents
Single Photon Counting Module COUNT blue -Series

Single Photon Counting Module

Standard-Area Silicon PIN Photodiodes (500µm)

Photodetector unit with built-in power supply and amplifier

TRiCAM APPLICATIONS KEY FEATURES. Time Resolved intensified CAMera. TRiCAM 13001A01 31/10/2013

LASERS FOR ANALYTICAL INSTRUMENTATION

SpectraTec X. Multi- λ Laser Light Source BLUE SKY RESEARCH WAVELENGTHS

XUV pulse FLASH

LASERS FOR ANALYTICAL INSTRUMENTATION

SYNCERITY TM 1024 x 256

specular diffuse reflection.

Progress of the Thomson Scattering Experiment on HSX

PureGain PG1600P. Data Sheet. Fixed Gain, Compact EDFA with Digital Control Electronics. Contents: Features:

Overview of WT&T LED light sources (fiber-coupled and free-space) WT&T Inc. Contact:

A New Model for Optical Crosstalk in SinglePhoton Avalanche Diodes Arrays

LASERS FOR ANALYTICAL INSTRUMENTATION

High-Power Femtosecond Lasers

LIFA KEY FEATURES APPLICATIONS. Fluorescence Lifetime Attachment LIFA 15001A02 16/03/2015

Preconfigured Optical Setups for the General Purpose Optical Bench for Frontier FT-IR Spectrometers. FT-IR Spectroscopy.

FCI-XXXA Large Active Area 970nm Si Monitor Photodiodes

Models 1417 and 1437 User s Manual. High-Speed Photodetector Modules

Master Oscillator Power Amplifier SLD System MOPA-SLD-850

Development of Ultrafast CXRS system in Heliotron J. Graduate School of Energy Science Kyoto University LU XIANGXUN 03/15/2016

RoHS COMPLIANT 2002/95/EC

Introduction to Diffraction Gratings

POWERSIGHT LASER DOPPLER AND PHASE DOPPLER MEASUREMENT SYSTEMS

SPECTRUM. The world s first fully automated Raman AFM. AFM - confocal Raman - SNOM - TERS AFM KPFM. Raman. AFM-Raman characterization of PS-PVAC

DEPARTMENT OF CHEMICAL ENGINEERING LABORATORY FOR THE PREPARATION OF NANO AND MICROMATERIALS

Continuous Wave 532 nm Diode-Pumped Solid State Laser CDPS532M

SUPPLEMENTARY INFORMATION

Lab 5: Diffraction and Interference

High-Accuracy LIBS with Nanosecond and Picosecond Time Resolution Enabled by Ultrasensitive emiccd Technology

405 nm Diode Lasers. 14-pin BF, D-type OEM Module & Turn-key Systems. 405 nm Single-Mode Wavelength Stabilized Diode Lasers

Detectors for Fiber Optics: InGaAs PIN Photodiodes

Development of automated ultraviolet laser beam profiling system using fluorometric technique

LIFA. SPECIFICATIONs. Fluorescence Lifetime Attachment LIFA14001A02 25/02/2014

Realizzazione di un sistema LiDAR basato su fotomoltiplicatori in silicio

WinCamD-LCM 1" CMOS Beam Profiling Camera, SuperSpeed USB 3.0, * nm * model-dependent

FluxGage. FluxGage. LED Luminaire Measurement System User Manual

C101-E137 TALK LETTER. Vol. 14

High spatial resolution measurement of volume holographic gratings

Evaluation of Satellite Ocean Color Data Using SIMBADA Radiometers

Minimizes reflection losses from UV - IR; Optional AR coatings & wedge windows are available.

Detectors for Future Light Sources. Gerhard Grübel Deutsches Elektronen Synchrotron (DESY) Notke-Str. 85, Hamburg

Light Sources opmaak.indd :32:50

Versatile laser Raman Spectrometer. RMP-500 series

Chapter 2: Wave Optics

2011 Optical Science & Engineering PhD Qualifying Examination Optical Sciences Track: Advanced Optics Time allowed: 90 minutes

Company Presentation Optical Components

Parameter Minimum Typical Maximum Units

Near Field Observation of a Refractive Index Grating and a Topographical Grating by an Optically Trapped Gold Particle

OTEB-CO-B Series. Booster EDFA OPERATING MANUAL

ISO INTERNATIONAL STANDARD. Particle size analysis Laser diffraction methods Part 1: General principles

Chapter 24. Wave Optics. Wave Optics. The wave nature of light is needed to explain various phenomena

MonoVista CRS+ Raman Microscopes

Genesis MX STM-Series

976nm Laser Module Stabilized with Narrow-Bandwidth Grating

Control of Light. Emmett Ientilucci Digital Imaging and Remote Sensing Laboratory Chester F. Carlson Center for Imaging Science 8 May 2007

ENHANCEMENT OF DIFFUSERS BRDF ACCURACY

IMAGING SPECTROMETER DATA CORRECTION

Chapter 24. Geometric optics. Assignment No. 11, due April 27th before class: Problems 24.4, 24.11, 24.13, 24.15, 24.24

Physics 625 Femtosecond laser Project

1.3 Customized Solutions (OEM)

Genesis CX SLM-Series

Lecture 4 Recap of PHYS110-1 lecture Physical Optics - 4 lectures EM spectrum and colour Light sources Interference and diffraction Polarization

Introduction of the Industry's Broadest Tunable Quantum Cascade Lasers. February Block Engineering 1

Chapter 24. Wave Optics. Wave Optics. The wave nature of light is needed to explain various phenomena

Mu lt i s p e c t r a l

Dual Fiber SWIR Laser Doppler Vibrometer

Philip E. Plantz. Application Note. SL-AN-08 Revision C. Provided By: Microtrac, Inc. Particle Size Measuring Instrumentation

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

SPECIFICATIONS HAND HELD PARTICLE COUNTER KC Higashimotomachi, Kokubunji, Tokyo , Japan

DEMONSTRATION OF THE EVOLUTION OF SPECTRAL RESOLVING POWER AS A SUPERPOSITION OF HIGHER ORDER DELAYED BEAMS

Modular Raman Spectrometers

TABLE OF CONTENTS PRODUCT DESCRIPTION CINCAM CCD TECHNICAL DATA SENSOR RESPONSE DIMENSIONS CINCAM CCD LARGE FORMAT TECHNICAL DATA SENSOR RESPONSE

Lecture 6 Proper,es of laser beams *

Michelson Interferometer

LASER RANGEFINDER (LRF) SYSTEM- INTEGRATOR KIT 1.5-MICRON LASER RANGEFINDER SYSTEM- INTEGRATOR KIT

PyLoN: Applications: Astronomy, Chemiluminescence, Bioluminescence, Phosphor Imaging, Ultra-low light Imaging and Spectroscopy.

Light Tec Scattering measurements guideline

Confocal Microscope Imaging of Single-Emitter Fluorescence and Hanbury Brown & Twiss Setup for Photon Antibunching. Edward Pei

PRODUCT INFORMATION Model 1000XP Wide Range Particle Spectrometer (WPS )

Time-resolved PIV measurements with CAVILUX HF diode laser

Chapter 22. Reflection and Refraction of Light

Interference of Light

Quad-RO: 4320 X-RAY GROUP

APPLICATIONS KEY FEATURES. High-speed intensified Camera Attachment

P R I C E L I S T. Laser Diagnostic Tools

Grid tunable laser source, Optical power meter, Optical attenuator, Optical switch, etc.

Section 2 PROGRESS IN LASER FUSION

Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing

Experimental Observation of Invariance of Spectral Degree of Coherence. with Change in Bandwidth of Light

Light Tec. Characterization of ultra-polished surfaces in UV and IR. ICSO October 2016 Biarritz, France

Prizmatix. Optical Specifications. Mic-LED-635 spectrum. BLCC-02 Benchtop LED Current Controller Specifications

Light Tec Scattering measurements guideline

WAVELENGTH MANAGEMENT

DIGITAL DIRECTIONAL INLINE OPTICAL POWER MONITOR/METER

Chapter 25. Wave Optics

Tutorial Solutions. 10 Holographic Applications Holographic Zone-Plate

Continuous Measurement up to 43 MS/s

Transcription:

Silicon Avalanche Photodiodes in Dynamic Light Scattering August 2016

Introduction This application note describes the use of the ID100 single photon counting detector for the measurement of light scattered by small particles. By analyzing the fluctuation of the scattered light intensity it is possible to evaluate the particle size. ID Quantique offers compact and affordable stateof-the-art single photon counting detectors based on silicon avalanche photodiodes with best-in-class timing resolution and a short dead time, these are especially for dynamic light scattering applications. Dynamic Light Scattering (DLS), also known as Photon Correlation Spectroscopy (PCS) or Quasi-Elastic Light Scattering (QELS) is a technique commonly used to determine the size of small particles in a solution [1, 2]. Shining a monochromatic light beam onto a solution with particles causes light to scatter. When the particles are much smaller than the incoming wavelength, this is predominantly Rayleigh scattering. This scattered intensity fluctuates with time. The time-dependent fluctuation occurs because the particles undergo Brownian motion and the distance between them is constantly varying [3]. From the time dependence of the scattering intensity fluctuation and knowing the temperature and viscosity of the medium it is possible to calculate the diffusion coefficient of the particles from which the hydrodynamic diameter can be derived. The hydrodynamic diameter is that of a sphere that has the same translational diffusion coefficient as the particle being measured [3]. If a laser beam is applied to a solution with particles, the scattered light intensity can be measured at any time with the ID100 single photon counting detector and the size of the particles can be evaluated, as described in the following sections. Dynamic Light Scattering and its Advantages DLS is a non invasive technique and it is well established for measuring the size of molecules and particles in the submicron region. This method offers several advantages: Measures particle sizes of 1nm to 1/10 of the wavelength of the light Typically ± 1% precision Short experiment duration (1-2min) Reliable and repeatable analysis Straightforward sample preparation Low sample volumes (as low as 2 ul) Modest development costs

ID100 Single Photon Counter The ID100 single photon counters are based on a reliable silicon avalanche photodiode (APD) sensitive in the visible spectrum. The modules are able to detect weak optical signals down to the single photon level. With their very low dark count rates, timing resolution of 40ps and dead time of 45ns, the modules have an unmatched dynamic range.. The ID100 is easy-to-use, self-contained and can easily be integrated in every optical set-up. Besides a very narrow Instrument Response Function (IRF) it has an excellent timing stability up to count rates of at least 10MHz. The id100 is available in the following versions: ID100-20: free space coupling with 20mm active area ID100-50: free space coupling with 50mm active area ID100-SMF20: Single Mode Fiber coupling with 20mm active area ID100-MMF50: with 50/125µm FC/PC Multi Mode Fiber coupling ID100-MMF50: with 100/140µm FC/PC Multi Mode Fiber coupling The ID100 modules exist in two grades, depending on the dark count rate specifications. For the Ultra- Low Noise grade, the dark count rate is less than 5Hz for the ID100-20 and less than 20Hz for the ID100-50 and ID100-MMF50. The key features are: Broad spectral range: 350 to 900 nm Best in class timing resolution of 40ps Low dark count rate of less than 5Hz Low dead time (45ns) Photon detection probability up to 35% Standard 50 Ω output with BNC connector Fast active quenching circuit Low bias voltage +5V Not damaged by strong illumination For the measurements mentioned below, the ID100-MMF50 detector was used to compute the scattered light intensity.

Experimental Setup Figure 1: Dynamic Light Scattering experimental setup. Auto-Correlation Measurements The experimental setup is illustrated in figure 1. A 632nm He Ne laser is used for this measurement. The beam passes through collimation lenses, allowing the light to be focused onto the solution. The scattered light is then detected by the ID100-MMF50 single photon detector. The output signal from ID100-MMF50 is sent to a correlator that computes the number of single photons detected as a function of time. The auto-correlation function is then calculated from the detected photon statistics. Figure 2 shows the normalized intensity correlation function (ICF) fluctuation with time. For a detector with zero after pulsing probability, the normalized ICF curve should go from 1 to zero with the increase of lag time. The deviation from 1 observed at short lag times in Figure 2 is due to to the after pulsing rate of the ID100-MMF50 that is typically 0.5% (you can find all details about its measurements in the annex document).

Figure 2. Intensity correlation function (ICF) dependence with time, in auto-correlation measurement. Courtesy of LS Instruments. Cross-Correlation Measurements The cross correlation measurement is performed with 2 single photon detectors (ID100-MMF50), as shown in figure 3. This measurement consists in illuminating the sample with a laser beam at 632nm and measuring the coincidence of the scattered light at a fixed angle. By performing a cross-correlation on the detected signal it is possible to eliminate the artefact created by the detectors after pulse rate. The specific cross-correlation setup of LS Instruments (figure 3) also allows suppression of photons that are scattered multiple times, which drastically interfere with measurements in turbid samples.

Figure 3: Cross-correlation experimental setup. Courtesy of LS Instruments. Figure 4 shows a comparison of the normalized intensity correlation function between the autocorrelation and cross-correlation measurements. Since the influence of the after pulses is eliminated, the cross-correlation curve has a flat behavior at short lag times. For longer lag times, both curves are similar. As the after-pulsing effect is suppressed reliable particle sizing is now feasible Figure 4: Comparison between cross-correlation and auto-correlation measurements. Courtesy of LS Instruments.

Once the intensity correlation function is measured, the particle size can be calculated by fitting the data with mathematical models based on assumed particle size distributions [3]. With appropriate data analysis DLS is also a sensitive method to detect particle aggregation. Industrial development In the setups described above, the ID100 single photon detector module is used. ID Quantique also provides an OEM (original equipment manufacturer) version of ID100. The ID101 is the smallest, most reliable single photon detector on the market. It is low cost, and easy to integrate mechanically and electrically Thus, the ID101 can be easily mounted on a printed circuit board and integrated in equipment such as spectrometers or microscopes. Conclusion ID Quantique s ID100-MMF50 single photon detector is able to detect weak signals in the visible spectrum with a time resolution of 40ps and a 45ns dead time. The small influence of the after pulse rate is eliminated when performing cross-correlation dynamic light scattering (DLS) measurements. The ID100-MMF50 is an easy-to-use and self-contained detector that can be used for Dynamic Light Scattering DLS) experiments and be implemented in commercial setups thanks to the OEM version ID101. References [1] P. Schurtenberger, and M. Newman, Characterization of Biological and Environmental Particles using Static and Dynamic Light Scattering, In Environmental Particles, J. Buffle, H. van Leeuwen, Eds.; Lewis Publishers: Boca Raton, 1993, 37-115. [2] T. G. Mason and D. A. Weitz, Optical Measurements of Frequency-Dependent Linear Viscoelastic Moduli of Complex Fluids, Phys. Rev. Lett.74 (1995), pp.1250-1253. [3] Michael J Saxton, Ken Jacobson, SINGLE-PARTICLE TRACKING : Applications to Membrane Dynamics Annual Review of Biophysics and Biomolecular Structure Vol. 26: p 373-399 (1997) or Brownian dynamics with hydrodynamic interactions in Article Reference [4] F. Scheffold, Particle Sizing with Diffusing Wave Spectroscopy, Journal of Dispersion Science and Technology, 23 (5) (2002), pp. 591 599. Information in this document is subject to change without notice. Copyright 2017 ID Quantique SA. Printed in Switzerland. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means electronic, mechanical, photocopying, recording or otherwise without the permission of ID Quantique. Trademarks and trade names may be used in this document to refer to either the entities claiming the marks and names or their products. ID Quantique SA disclaims any proprietary interest in the trademarks and trade names other than its own.