RSA-PE-20. Introduction Unpacking and Inspection... 2

Similar documents
Cary Fiber Optic Probes

Operating Procedure PerkinElmer LAMBDA 950 (Integrating sphere Mode)

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

FluxGage. FluxGage. LED Luminaire Measurement System User Manual

Contents Safety.. 1 General. 1 Electrical.. 1 Warning 1 Working Principle 2 Unpacking Instructions 2 Specifications 3 Installation.. 3 Introduction..

Thermo Scientific Spectronic 200 Visible Spectrophotometer. The perfect tool. for routine measurements

C101-E137 TALK LETTER. Vol. 14

6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory

Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc.

Philpot & Philipson: Remote Sensing Fundamentals Interactions 3.1 W.D. Philpot, Cornell University, Fall 12

SOLUTIONS RECERTIFICATION CALIBRATION AND UV/VIS REFERENCE MATERIALS. Photometric and Wavelength Accuracy Reference Material Set

Operating Procedure PerkinElmer LAMBDA 750

Light Tec Scattering measurements guideline

NAV-1 Lens Support. NAV-2 Lens Support BEFORE YOU BEGIN

STEP-BY-STEP INSTRUCTIONS FOR BUILDING AN AUTOCOLLIMATOR. TECHSPEC Optical Cage System

Upgrading LVDS Cables Instruction Sheet

Radiance, Irradiance and Reflectance

Light Tec Scattering measurements guideline

D. Instruction Manual and Experiment Guide for the PASCO scientific Model OS-8523 SLIT ACCESSORY FOR THE BASIC OPTICS SYSTEM

Spectro-115E TM and Spectro-115U TM Spectroelectrochemical System. Operator s Manual

ISP Series. Integrating spheres for all applications. We bring quality to light.

E1135C PDU and Pod Upgrade Procedure

Improving Traceability of Fluorescence Calibrations to Practical Colorimetric Applications

Section. Service & Maintenance. - Core & Hard Disk Drive (HDD) - Amplifier - Monitor - UPS - Dollar Bill Acceptor - Fan Filter G - 1

Megatouch FORCE Monitor Chassis Board Replacement

Light Tec Scattering measurements guideline

UV / VIS SPECTROPHOTOMETER

Installation Instructions

Light Tec Scattering measurements guideline

Calibration Procedures for LI-COR Spectroradiometers, Radiation Sensors & Lamps

Installation Instructions

ABM's High Resolution Mask Aligner Features:

Analysis of spectrophotometer specular performance using goniometric information

The Council for Optical Radiation Measurements (CORM), NIST July 30 - Aug 1, 2018

UV-6 Series Double Beam UV/Vis Spectrophotometers

Optical Fiber Assemblies

ATTENTION: OBSERVE PRECAUTIONS FOR HANDLING ESD-SENSITIVE DEVICES

SPECTROMETERS USER MANUAL. Ver. 1.2_09/16. Making spectroscopy brighter

Model : LUV-300 Please read this manual carefully before using the instrument Labnics Equipment

dq dt I = Irradiance or Light Intensity is Flux Φ per area A (W/m 2 ) Φ =

Installation Guide. Retrofit Kit for USB Ready Intraoral Systems

Model 2657A-LIM-3 LO Interconnect Module

Cutter Option Installation Instructions

E3S-A. Built-in Amplifier Photoelectric Sensor (Medium Size) Ordering Information. Built-in Amplifier Photoelectric Sensors. Horizontal. 7 m.

Reflective Illumination for DMS 803 / 505

BPL SERIES INSTALLATION INSTRUCTIONS THIS SHEET CONTAINS IMPORTANT SAFETY INSTRUCTIONS. SAVE THESE INSTRUCTIONS.

Physical Optics. You can observe a lot just by watching. Yogi Berra ( )

Service & Maintenance

Ag Leader Technology. DirectCommand Installation Hardi 20-pin Interface Kit (Sprayer Chassis Mount)

USER S MANUAL

Table of Contents 1. PRINCIPLE, APPLICATION AND FEATURES PRINCIPLE APPLICATION FEATURES SPECIFICATIONS AND ACCESSORIE

D&S Technical Note 09-2 D&S A Proposed Correction to Reflectance Measurements of Profiled Surfaces. Introduction

Two slit interference - Prelab questions

UV WinLab Software The Lambda Series

UNICO 2100 SERIES SPECTROPHOTOMETER Model 2100 Spectrophotometer Model UV 2100 Spectrophotometer

EFFECTS OF MEASUREMENT GEOMETRY ON SPECTRAL REFLECTANCE AND COLOR

INSTRUCTIONS FOR THE INSTALLATION OF THE INFINITY "L" DISPLAY HOOD (INTO PREVIOUSLY INSTALLED INFINITY "L" SYSTEMS)

WAVELENGTH MANAGEMENT

FTIR-7600 FT-IR Spectrometer Accessories Price List 2011

LS-785 High Throughput Lens Spectrograph User Manual

Lab 5: Diffraction and Interference

CDRPan. Installation. Instructions for Siemens OP-10 FOR PANORAMIC SYSTEMS

Model RD-4 Remote Time Display

PS/IO Circuit Board Retrofit

TIVO UPGRADE INSTRUCTIONS (c) , Adberg Consulting LLC. All rights reserved.

Diving into a new dimension: Quality meets efficiency

Application Note. Measuring the Geometric Attributes of Your Products AN

A TCP/IP network CAT 5 cable If the network is faster than 10baseT a switching hub will be needed Static IP address

Validation of the Gonioreflectometer

P160 User s Manual Manuel de l utilisateur Anwenderhandbuch Manuale per l operatore Manual del usuario

MBE Mounts and Adapters

CAMERA ASSEMBLY. Removal/Replacement of the Camera Box Assembly APR-CA. Install Camera Assembly. Remove Camera Assembly

Instructions to Install Retrofit Kit RVMC 4/5000 Machine (MDB Only)

Keysight N9030A Signal Analyzer Option BBA Analog Baseband IQ Inputs Retrofit Kit

Light Sources opmaak.indd :32:50

Integrating Sphere Uniform Light Source Applications

Light Tec. Characterization of ultra-polished surfaces in UV and IR. ICSO October 2016 Biarritz, France

Spectrographs. C. A. Griffith, Class Notes, PTYS 521, 2016 Not for distribution.

TRIBOMETERS MICRO-MACRO TRIBOLOGY TESTING

Inspection Technology Europe BV Allied NDT Engineers

Quick Start. This document describes how to install the Juniper Networks PTX5000 Packet Transport

Spectrophotometric Methods of Refractive Indices Measurement

Mu lt i s p e c t r a l

Instructions to Install Retrofit Kit 6/7000 Machine

Standard Operating Procedure for the Horiba FluroMax-4

Removal and Installation8

dq dt I = Irradiance or Light Intensity is Flux Φ per area A (W/m 2 ) Φ =

TIVO DVR UPGRADE INSTRUCTIONS (#80-HD)

Operating Instructions OEM Spectrometer RU 60

i-series Videoconference Fixture

CDRPan. Installation. Instructions for Siemens / Sirona Orthophos 3 FOR PANORAMIC SYSTEMS

B&W RearView Camera Installation & Operation

PHYSICS 116 POLARIZATION AND LIGHT MEASUREMENTS

Scanner 2000 microefm QuickStart. Installing the Scanner Remote Mount. Direct Mount NUFLO. Part No , Rev. A

ENGR142 PHYS 115 Geometrical Optics and Lenses

SPECIAL INSTRUCTIONS FOR CAPACITORS COMPACT GENERATORS

QCOLVCZ4M/Z6M SV Series Interface Cable Kit for Z4M and Z6M

WAVELENGTH MANAGEMENT

EMC 10T "CE" Mechanical Upgrade Procedure

PHOS mini OPERATING INSTRUCTIONS

Transcription:

RSA-PE-20 Introduction... 1 Unpacking and Inspection... 2 Installation and Assembly... 3 Installing the Accessory PC Board Kit... 3 Installing the Accessory in Your Sample Compartment... 6 Accessory Alignment... 6 Diagnostic Scans... 8 System Description and Operation... 11 Integrating Sphere... 11 Transfer Optics... 12 Loading the Sample Holders... 12 Theory of Operation... 13 Operating Procedures for the RSA-PE-20 Accessory... 21 Maintenance... 26 General Information... 26 Mirror Cleaning Procedure... 26 Spectralon Cleaning Instructions... 26 Appendix A Specifications... 27 Appendix B Typical Spectralon Reflectance Values... 28 Appendix C Substitution Error Sample Scans... 29 Appendix D Reflectance Glossary... 31 Appendix E Recommended Reading... 33 AQ-00073-000, Rev. 7

Introduction The Labsphere RSA-PE-20 is a diffuse reflectance and transmittance accessory designed for a number of PerkinElmer UV-VIS spectrometers. The list of compatible spectrometers includes the Lambda 25/35/45 series as well as the older models Lambda 2/12/14/20/40. The RSA-PE-20 accessory essentially is an optical bench, including transfer optics and an integrating sphere, that fits neatly into the sample compartment of the spectrometer instrument. An illustration of the accessory is provided in Figure 1. Since the reference beam passes through the accessory unimpeded, the RSA-PE-20 accommodates either single or double beam configuration spectrometer. The accessory, however, is a single beam device. The accessory is ideal for measuring diffuse reflectance or transmittance for scientific and industrial applications including color, thermochromism, composition of opaque materials such as biological or geological specimens. The geometric measurement capabilities of the accessory include 8 /hemispherical reflectance, 0 /diffuse reflectance, 0 /hemispherical transmittance and 0 /diffuse transmittance. Sample holders for the test and reference samples are supplied with the accessory. This manual provides the installation and operating procedures for the RSA-PE-20. The operation of your spectrometer system with the integrating sphere accessory installed, however, essentially is unchanged. Since the procedures may vary from one spectrometer to the next, you should consult your PerkinElmer instruction manual when taking reflectance and transmittance measurements. Figure 1. The RSA-PE-20 fits into the sample compartment of your PerkinElmer spectrometer. AQ-00073-000, Rev. 7 1

Unpacking and Inspection The RSA-PE-20 was thoroughly inspected and calibrated before shipping and should be ready to operate after completing the set-up instructions. All Labsphere instrumentation is packaged and shipped in reinforced shipping containers. The RSA-PE-20 is sold in two different configurations, depending on the source of your order and the model number of your PerkinElmer UV/ VIS spectrometer. Carefully check the components after unpacking for any damage that may have occurred during shipping. If there is any such damage, file a claim immediately with the freight carrier and contact the Labsphere Customer Service Department at!(603) 927-4266. Standard components accompanying all orders: RSA-PE-20 Accessory One 8 and two 0 sample holders Accessory cable assembly Cuvette Holder Tool Kit USRS-99-010 Uncalibrated Spectralon Reflectance Standard Instruction Manual If your system was ordered directly from Labsphere, an instrument modification kit will accompany the accessory and the components listed below will be labeled by Labsphere part number. If your RSA-PE-20 was ordered through your PerkinElmer representative, the modification kit will be provided directly from PerkinElmer and installed by their qualified technician. Accessory PC Board Kit part No. AS-00820-000 for use with Lambda 2 and Lambda 12 model spectrometers RSAPE20 wiring harness Part No. AS-01803-000 for use with the Lambda 14 spectrometer Optional Components Calibrated SRS-99-010 Spectralon Reflectance Standard Custom sample holders AQ-00073-000, Rev. 7 2

Installation and Assembly In all cases, the RSA-PE-20 Accessory PC Board Kit installation should be accomplished by a qualified PerkinElmer technician. A technician is available through your local PerkinElmer representative. Once the accessory is installed, you can disconnect and reconnect the accessory as often as desired using the accessory installation procedure below. The RSA-PE-20 is compatible with a number of older Lambda spectrometer models as well as the current Lambda 25/35/45 series. The RSA-PE-20 installation requirements for each spectrometer are different. Installing the Accessory PC Board Kit If your PerkinElmer spectrometer is one of the current Lambda Series, i.e. Lambda 25/35/45, no kit installation is required - the instrument is already configured to accept the RSA-PE-20 accessory. If your spectrometer is an older Lambda series model, an accessory PC board kit will need to be installed into the spectrometer instrument before installation of the RSA-PE-20 accessory. Kit Modifications to the Lambda 2 and Lambda 12 Spectrometers The Accessory PC Board Kit for the Lambda 2 and Lambda 12 spectrometers includes two components: a ribbon cable, Part No. AS-01803-100, and the EC-12201-000 accessory side plate. J18 J8 Figure 2. Part No. EC-12201-000 and AS-01803-100 of the accessory PC board kit used in modifications to the Lambda 2 or Lambda 12 instrument. 1. Disconnect electric power from the spectrometer. Remove the sample compartment AQ-00073-000, Rev. 7 3

access door and open the top cover to gain access to the instrument. 2. Loosen the two side plate mounting fasteners on the top, inside end of the plate. One of these screws fastens the cover leash - its removal will permit the spectrometer cover to fall back. 3. Remove the large ribbon cable connector from the header on the existing side plate circuit board and lift the plate out of the instrument. 4. Install the new side plate, Part No. EC-12201-000, into the instrument in the same orientation as the original plate. 5. Reconnect the large ribbon cable into the same header connector located on the newly installed side plate. 6. Unplug the detector cable that runs from the detector module to the motherboard at connector P8. Route the same end of this cable to header P8 on the newly installed side plate. Note that the header is keyed. 7. Plug the gray connector on Part No. AS-01803-100 into the side plate header labeled P18. Note that this connector is keyed. 8. Route the black connector of Part No. AS-01803-100 to the instrument motherboard at P8. This connector is keyed. 9. The individual wires attached to Part No. AS-01803-100, labeled +5V and GND, should be plugged into the motherboard electrical studs labeled 5V/REL1 and GND respectively. 10. Tighten the two mounting screws that secure the side plate to the instrument enclosure. 11. Proceed to the accessory installation procedure. Kit Modifications to the Lambda 14 Spectrometer The Accessory PC Board Kit for the Lambda 14 spectrometer includes three components: a ribbon cable, Part No. AS-01803-000, and the two components listed above, Part No. AS-01803-100 and EC-12201-000. P1 P3 Figure 3. Part No. AS-01803-000 used in modification of the Lambda 14 spectrometer. 1. Disconnect electric power from the spectrometer. Remove the sample compartment access door and open the top cover to gain access to the instrument. 2. Loosen the two side plate mounting fasteners on the top, inside end of the plate. One of these screws fastens the cover leash - its removal will permit the spectrometer cover to fall back. 3. Remove the large ribbon cable connector from the header on the existing side plate AQ-00073-000, Rev. 7 4

circuit board and lift the plate out of the instrument. 4. Install the new side plate, Part No. EC-12201-000, into the instrument in the same orientation as the original plate. 5. Reconnect the large ribbon cable into the same header connector located on the newly installed side plate. 6. Remove the cable running from the detector module to the motherboard header at P9. This cable will not be used. 7. Plug the single connector end of Part No. AS-01803-000 into the header at the detector module. 8. Connect the gray 16-pin connector on the opposite end of Part No. AS-01803-000 into the P9 header on the motherboard. 9. Connect the remaining black connector end of Part No. AS-01803-000 into the newly installed accessory side plate header labeled P8 as shown in Figure 4. Figure 4. Making the detector cable hookup to the side plate. 10. Plug the gray connector on Part No. AS-01803-100 into the side plate header labeled P18. Note that this connector is keyed. 11. Route the black connector of Part No. AS-01803-100 to the instrument motherboard at P8. This connector is keyed. 12. The individual wires attached to Part No. AS-01803-100, labeled +5V and GND, should be plugged into the motherboard electrical studs labeled +5V and GNDA respectively. 13. Tighten the two mounting screws that secure the side plate to the instrument enclosure. 14. Proceed to the accessory installation procedure. Kit Modifications to the Lambda 20 and Lambda 40 Spectrometers The accessory kit and associated cables for modification to the Lambda 20 and Lambda 40 spectrometers are provided by PerkinElmer. Consult your PerkinElmer technician for installing the accessory kit for these instruments. AQ-00073-000, Rev. 7 5

Installing the Accessory in Your Sample Compartment 1. Unplug the instrument and remove the sample compartment cover and any accessories within the sample compartment. 2. Mount the 8 sample holder against the reflectance port of the integrating sphere, tightening the two set screws using a 0.035" hex wrench. 3. Mount a 0 sample holder against the transmittance port of the integrating sphere, tightening the two set screws with the 0.035" hex wrench. 4. String the accessory connector cable through the sample compartment wall cable run leading to the side plate. Attach the 9-pin connector to the 9-pin port marked INTEGRATING SPHERE (J88) on the side of the instrument. Accessory Cable Accessory Preamp Board Figure 5. The accessory cable connects the sphere detector to the instrument electronics. 5. Plug the remaining end of the detector cable into the mating plug located in the preamp board enclosure underneath the base of the accessory. 6. Carefully lower the accessory into the sample compartment and slide it horizontally until the four self-retaining mounting screws on the accessory base mate with the sample compartment base. Tighten the screws with a 2.5 mm hex wrench. 7. Lower the instrument cover, replace the sample compartment door and reconnect the power plug to the instrument. Accessory Alignment The accessory was aligned precisely at the Labsphere factory before shipping and should be ready for use upon installation. Check the alignment of the accessory before using and, if necessary follow this procedure to adjust the optical alignment. 1. Load the Spectralon reflectance standard at the reflectance port. 2. Close the sample compartment, turn on the instrument and boot your UV Winlab software application. AQ-00073-000, Rev. 7 6

WARNING: Do not touch the mirror surfaces with your bare fingers. 3. Configure the instrument for white light operation by entering 0 nm in the Instrument section of the Lambda Manual Control Display. 4. Dim the room lights. M3 M2 Reference Beam Transmittance Port Sample Beam Integrating Sphere M1 Reflectance Port Figure 6. Optical setup of the RSA-PE-20 accessory. Make sure your UV Winlab software is configured for the sample beam at the front port. 5. Place a piece of white translucent tissue paper in front of mirror M2. Adjust the thumbscrews on M1 so the beam is centered. 6. Place the white paper in front of mirror M3. Adjust the thumbscrews of M2 so that the beam is centered on mirror M3. 7. Place the white paper in front of the transmittance port of the sphere. Adjust the thumbscrews of M3 so that the beam enters the transmittance port without clipping the edges of the port. The transmittance port can be viewed by looking into the sphere through the front reflectance port. 8. Place a piece of lens paper at the reflectance port. Adjust the thumbscrews of mirror M3 so that the beam is centered on the reflectance port. If the beam can not be centered, repeat the previous steps making minor adjustments in either direction so the beam is centered on the reflectance port and is not clipped at the transmittance port. 9. The accessory is now properly aligned. You can now proceed to the next section on diagnostic scans. AQ-00073-000, Rev. 7 7

Diagnostic Scans This section of the instruction manual contains the procedures for a series of diagnostic scans on your accessory. These scans were performed on your accessory and our instrument before we shipped the RSA-PE-20 to you.you should perform these scans again on your instrument after installing the accessory to validate proper operation of the accessory. The scan results should be retained for future use. If problems develop with your accessory in the future, repeat the diagnostic scans and send us copies for analysis. The quality assurance documentation includes a copy of the diagnostic scans we performed at Labsphere. The instrument parameters used for each diagnostic scan are listed under the test results. You should use these same parameters for your diagnostic scans. The user should note that a background correction is executed as part of the 100% Baseline Scan and used for all subsequent diagnostic scans. Sample Beam Energy Scan The energy scan is an indicator of accessory throughput. 1. Set up instrument parameters to conform to those listed in Table 1. These parameters settings should be identical to those listed in the quality scans accompanying your shipment. Instrument Parameter Setting Lamps UV/Vis ON Beam Single Beam Method Scan Data Interval 1 nm Abscissa Start 250 Abscissa End 1100 Slit Mode Fix Slit 2 nm Integration Time 0.08 s Ordinate Mode E1 Sample Beam Front Attenuators 100% Table 1. Instrument setup for sample beam energy scan. 2. Load the reflectance standard at the sample reflectance port. The 8 wedge should be installed. 3. Close the sample compartment door. Execute a scan from your UV Winlab software and save the results. AQ-00073-000, Rev. 7 8

100% Baseline Scan This scan provides the direct comparison of a sample scan to the baseline at the same sphere configuration. The 100% baseline scan is used for general accessory troubleshooting purposes. 1. Set up instrument parameters as listed in Table 2. Instrument Parameter Setting Lamps UV/Vis ON Beam Double Beam Method Scan Data Interval 1 nm Abscissa Start 250 Abscissa End 1100 Slit Mode Fix Slit 2 nm Integration Time 0.08 s Ordinate Mode %R Sample Beam Front Attenuators 100%/100% Table 2. Instrument setup for the 100% Baseline Scan. 2. Load the reflectance standard at the sample reflectance port. 3. Close the sample door and perform a background correction scan. 4. Perform a sample scan with the same sphere configuration and save the results. Blocked Beam Scan The blocked beam or zeroline scan records the detector noise level from the accessory. You should use the baseline recorded during the 100% baseline scan in the previous procedure, or execute a new baseline if it is not available. 1. Set up instrument parameters as listed in Table 2. 2. Load a metal plate or other non-transmitting sample into the transmittance port. The sample should be absorbent such that reflected light from the sample beam does is not sensed by the reference beam detector. Leave the reflectance standard at the sample reflectance port. 3. Close the sample compartment door. Execute a sample scan and save the results. Light Trap Zeroline The light trap scan determines if the sample beam is correctly centered on the sample reflectance port. You should use the baseline recorded during the 100% baseline scan, or execute a new baseline if it is not available. AQ-00073-000, Rev. 7 9

1. Set up instrument parameters as listed in Table 2. 2. Leave the transmittance port empty and install a light trap over the sample reflectance port. 3. Close the sample compartment door. Execute a sample scan and save the results. Time Drive Scan The time drive scan indicates the stability of the light source and the sphere detector. 1. Set up instrument parameters as listed in Table 3. 2. Load a reflectance standard at the sample reflectance port and close the sample compartment door. Instrument Parameter Setting Lamps UV/Vis ON Beam Double Beam Method Time Drive Total Time 180s Wavelength 500 nm Slit Mode Fix Slit 2 nm Integration Time 0.24 s Ordinate Mode %R Sample Beam Front Attenuators 100%/100% Table 3. Instrument setup for the Time Drive Scan. 3. Perform an Autozero scan at 500 nm. 4. Run a time drive at 500 nm and save the results. AQ-00073-000, Rev. 7 10

System Description and Operation The RSA-PE-20 accessory is specifically designed to measure the reflectance or transmittance of solids, liquids, powders, or other small objects that can fit in the transmittance or reflectance ports. Basic components of the accessory include the integrating sphere, transfer optics and detector preamplification module. Although the RSA-PE-20 is a single beam construction, the accessory is compatible with the double beam configuration of the Lambda Series spectrometers and your UV Winlab software. Integrating Sphere The integrating sphere assembly, shown in Figure 7, is 50 mm in diameter and constructed of Spectralon. Spectralon is the same material used on the reflectance standard supplied with the accessory. Typical reflectance values for Spectralon are provided in Appendix B. Spectralon Sphere Port Frame Port Frame Sample Reflectance Port Transmittance Port Detector Baffles Detector Preamp Board Figure 7. Construction of the integrating sphere. The RSA-PE-20 integrating sphere features two ports: a transmittance port where the beam enters the sphere, and a sample reflectance port. The illustration in Figure 7 shows the accessory integrating sphere along with the transfer optics and preamplification board. The silicon detector is mounted directly to the preamp board underneath. Two Spectralon baffles inside the sphere shield the detector sensor from direct illumination by the ports. The sample holders, mounted either at the transmittance or sample reflectance port, clamp the AQ-00073-000, Rev. 7 11

reflectance standard or sample against the respective port. Three sample holders are provided with the accessory for mounting a sample either at 0 or 8 angles of incidence. The sample holders mount directly to the port frame at each end of the sphere. The sphere detector is a silicon photodiode installed at the detector port at the bottom of the integrating sphere. The photodiode operates in photovoltaic mode; the detector output is conditioned by the preamplifier board for compatibility with the instrument electronics. Detailed information on the photodiode is provided in Appendix A. Transfer Optics The transfer optics of the RSA-PE-20 accessory direct the spectrometer sample beam to the transmittance port on the integrating sphere. The sample beam is the front beam in the accessory. As shown in Figure 6, mirrors labeled M1, M2 and M3 guide the sample beam through the sample transmittance port and onto the sample reflectance port at normal incidence. All mirrors are AlMgF 2 coated. Mirrors M1 and M2 are flat. Mirror M3 is a spherical mirror that condenses the sample beam onto the target sample. The reference beam passes through the accessory unimpeded. Loading the Sample Holders Both transmittance and sample reflectance ports are 5/8-inch holes at either end of the accessory sphere. The transmittance port is closest to mirror M3. A sample holder clamp fits over a dovetail extension located underneath each port, securing the reflectance standard or sample tightly against the port opening. The sample holder with the 8 wedge should be installed at the sample reflectance port when taking 8 hemispherical measurements. The 0 sample holder should be installed at the sample reflectance port when taking diffuse measurements with the specular component excluded. Figure 8 describes the proper way to load the sample holder. When mounting a sample or reflectance standard at any port, you should make sure the reflecting surface lies flat against the wedge and completely fills the port surface area. To remove the sample holder at the sample reflectance port, loosen the set screws along the side of the respective wedge and remove the entire sample holder assembly. AQ-00073-000, Rev. 7 12

8 Wedge Sample or Standard Clamp Figure 8. Loading a standard or sample into the sample holder. In some instances, you may need to load a light trap at the sample reflectance port. The light trap can be loaded using either wedge and fits into the sample holder in the same manner as a reflectance standard. Theory of Operation During the traditional measurement of sample absorption by a spectrometer, the relationship between absorptance and transmittance of the sample beam is described by the Kirchoff equation: A + T = 1, where: A = absorptance and T = transmittance. When measuring the absorptance of a sample in the sample compartment, the detector signal of the spectrometer represents the portion of the sample beam that is not absorbed or scattered by the sample. When using the RSA-PE-20 accessory, it is convenient to use the Kirchoff relationship: AQ-00073-000, Rev. 7 13

A+ T+ R = 1, where R = reflectance of the sample beam at the sample surface. During reflectance measurements, the reflected component of the sample beam is collected by the integrating sphere and detected by the sphere detector. The detector signal represents the part of the sample beam that is not transmitted and not absorbed by the sample substance. Double Beam Spectroscopy In a double-beam, ratio-recording spectrometer, the measurement of reflectance factor and transmittance involves the calculation of a baseline based upon values recorded in a background correction scan. Background correction is used to compensate for changes in sphere efficiency due to the introduction of the sample into the system and for any imbalance in the energy of the sample and reference beams. This correction is performed automatically by the instrument and is the same routine used during spectroscopy evolutions without the accessory. The theoretical basis for the background correction, as it applies to simple reflectance and transmittance measurements, is explained briefly in this section. Radiation from the instrument illumination sources is split into two different beams: the sample beam and the reference beam. Each beam is interrupted periodically by means of an optical chopper such that the integrating sphere is illuminated alternately by the two beams. At any given wavelength, the instrument records the ratio of the signal produced by the detector when the sphere is illuminated by the sample beam to that when the reference beam is measured. Therefore, when a background correction scan is performed, for either reflectance or transmittance measurement, the background correction value B recorded by the instrument may be expressed as: B E s ρ s κ = ---------------- s Eq. 2 E r ρ r κ r where: E s is the energy of the sample beam, E r is the energy of the reference beam, ρ s is the reflectance factor of the sample at the sample reflectance port ρ r is the reflectance factor of the reference port, and κ s and κ r are the efficiency with which the energy reflected from the sample and reference beams is captured by the sphere and converted into a signal by the detector. AQ-00073-000, Rev. 7 14

The sphere efficiency factors (κ) for a given sample are a function of many variables including the spatial distribution of the energy reflected from the sample, the reflectance of the sphere wall, the geometry of the sphere (location of ports, baffles, etc.) and the efficiency of the detector itself. Proper measurement procedures make it possible to greatly reduce or eliminate the effect of such factors on transmittance and reflectance factor measurements. During the sample scan of a transmittance measurement, the transmittance (T s ) of a non-scattering sample placed in the path of the sample beam affects the amount of energy entering the sphere and reaching the sample reflectance port. The sample scan value S T of the ratio recorded by the instrument is: S T T s E s ρ s κ s = ---------------------- Eq. 3a E r ρ r κ r The value D T displayed by the instrument, however, is S T divided by the value B recorded in the background correction curve: D T S T T ----- s E s ρ s κ s E r ρ r κ = = -------------------------------------- r = T B E r ρ r κ r E s ρ s κ s Eq. 3b s Therefore, in a transmittance measurement of a non-scattering sample, the value displayed by the instrument is simply equal to the transmittance of the sample. In reflectance measurement for a double beam spectrometer, the standard positioned at the sample reflectance port during the uncorrected background measurement is replaced by another sample of unknown reflectance ρ u. The value S R of the ratio recorded by the instrument during the sample scan is: S R E s ρ u κ = ----------------- u Eq. 4a E r ρ r κ r The spectral data displayed by the instrument is: AQ-00073-000, Rev. 7 15

D R = S ----- R B E s ρ u κ u E r ρ r κ r ρ --------------------------------- u κ = = ----------- u Eq. 4b E r ρ r κ r E s ρ s κ s ρ s κ s Assuming that the spatial distribution of energy reflected from this unknown sample is the same as that of the original standard, the efficiency factor (κ) will be the same for both materials: κ u = Eq. 5 κ s Therefore, the value displayed by the instrument will simply be equal to the ratio of the reflectance factor of the unknown sample to that of the original standard: D R ρ u = ---- Eq. 6a ρ s If the reflectance factor of this original standard is known, it can be used as a reference standard and the reflectance factor of the unknown sample can be derived as follows: ρ u = D R ρ s Eq. 6b This operation is known as the reference correction to distinguish it from background correction described in this manual. The foregoing account involves two simplifying assumptions that bear further discussion. First, the efficiencies of the integrating sphere, represented by κ r,κ s and κ u, are not necessarily constant under changes in the system, as suggested by Equations 3a and 4a. The introduction of a reflectance or transmittance sample into the system may change the sphere efficiency. It is commonly assumed, however, that the sphere efficiencies will change by the same factor. This factor is eliminated when the instrument records its ratio, so that the foregoing analysis remains valid. The second assumption applies to reflectance factor measurements and is represented by Equation 5. Here it is assumed that the geometric scattering properties of the sample and the reference AQ-00073-000, Rev. 7 16

standard are identical. Therefore, the efficiency factor for radiation reflected from each sample will be the same. If the geometric scattering properties of the sample differ greatly from those of the reference, significant systematic errors may be introduced into the measurements. Therefore, the geometric scattering properties of the sample and the reference should be matched as closely as possible. Although your PerkinElmer spectrometer may be a double beam instrument, the RSA-PE-20 is a single beam accessory such that combined operation is in dual beam spectrometer mode. The value of κ r for your system does not change from background correction to sample scan. Single Beam Substitution Error Correction The single beam substitution error, sometimes known as single beam sample absorption error, is the systematic, predictable, and non-random error inherent in single beam integrating spheres measuring reflectance or transmittance. The error is caused by the difference in the throughput of the sphere when the reference makes up a portion of the sphere wall and when the sample is substituted for the reference. In integrating spheres which are designed for dual beam spectrometers, it is not possible to place both the reference beam and the sample beam into the sphere at the same time. These dual beam spectrometers employ a beamsplitter to divide the instrument s beam between the sample and reference channels. Reference Detector (optional) Reference Detector (optional) Sample Beam Physical Reference Sample Beam Sample Reference Beam Reference Beam Background Scan Sample Scan Figure 9. Operation of a dual beam spectrometer. Thus, even though the dual beam spectrometer has two beams, it is still only possible to utilize a single beam integrating sphere accessory. Unfortunately, single beam integrating spheres are subject to substitution errors during sample measurements which must be corrected to obtain accurate results. In reflectance measurements the throughput and corresponding measured reflectance is usually lower when the sample is present since a reference material of high reflectance is used. In transmittance measurements the throughput and the measured transmittance is usually higher when the sample is present since an open port is typically used as a reference. With spectrometers that AQ-00073-000, Rev. 7 17

use a chopped signal between sample and reference, this error does not occur. Double beam spectrometers are able to accept double beam integrating spheres in which both the sample and reference beams are placed in the sphere. Double beam spectrometers employ a chopper to divide the instrument beam between the sample, reference, and dark channels. Although your spectrometer is a double instrument, the RSA-PE-20 is designed as a single beam accessory to accommodate the size of the instrument sample compartment. Since only the sample beam is subject to sphere configuration, your RSA-PE-20 measurements are subject to substitution error. When a sample and a reference are of similar reflectance, the substitution correction is very small; at worst it may reach as much as 4-5%. In quality control applications where a threshold value is used, this may not be a concern, as the error can simply be built into the threshold. This is also true if only peak position information is required, as single beam correction only concerns the photometric scale. There are both physical and mathematical solutions to the problem of single beam substitution error. If the sphere is made sufficiently large and the area of the sample is minimized, the error increasingly diminishes. The use of a large sphere is usually not practical in low cost spectrophotometer systems due to signal-to-noise concerns as well a price considerations. The integrating sphere on your RSA-PE-20 accessory is too small to accommodate the modifications necessary for a physical solution. The mathematical solution is simple in theory, but complex in practice. If a reference is used that is very close in reflectance to the reflectance of the sample, the substitution error becomes negligible. If the user can match the reference reflectance to the reflectance of the sample, or has a large database of different level photometric reflectance standards, one can set up look-up or correction tables to correct for the substitution error. Tables 4 and 5 show reflectance values for twelve levels of photometric gray scale. Table 4 provides data for gray Spectralon exhibiting diffuse reflectance characteristics under 50%. Table 5 provides the data for Spectralon reflectances greater than 50%. The italicized values are the actual reflectance of samples of Labsphere SRS-XX gray scale reflectance material. These reflectance values were determined by measuring the corrected 8 / hemispherical reflectance factors using a Labsphere double beam spectrometer and a double beam accessory with a 150 mm diameter integrating sphere. The second set of reflectance values in regular print were measured on a spectrometer equipped with a Labsphere 50 mm single beam integrating sphere accessory. AQ-00073-000, Rev. 7 18

Wavelength Gray0/L2 Gray0/L19 Gray4/L2 Gray4/L19 Gray5/L2 Gray5/L19 Gray7/L2 Gray7/L19 Gray10/L2 Gray10/L19 Gray11/L2 Gray11/L19 300 0.009 0.011 0.053 0.065 0.099 0.123 0.192 0.231 0.337 0.394 0.378 0.436 325 0.009 0.011 0.050 0.064 0.095 0.121 0.188 0.230 0.329 0.392 0.371 0.434 350 0.009 0.012 0.050 0.064 0.093 0.121 0.184 0.230 0.323 0.388 0.366 0.432 375 0.009 0.011 0.049 0.064 0.092 0.122 0.182 0.229 0.319 0.387 0.362 0.431 400 0.010 0.011 0.048 0.064 0.091 0.122 0.181 0.230 0.316 0.386 0.360 0.431 425 0.010 0.011 0.048 0.064 0.091 0.122 0.181 0.231 0.314 0.385 0.359 0.431 450 0.010 0.011 0.048 0.064 0.091 0.122 0.182 0.232 0.313 0.385 0.358 0.432 475 0.010 0.011 0.049 0.064 0.091 0.122 0.183 0.233 0.313 0.385 0.359 0.433 500 0.010 0.011 0.049 0.064 0.092 0.123 0.184 0.235 0.313 0.385 0.359 0.433 525 0.010 0.011 0.050 0.065 0.092 0.123 0.186 0.236 0.313 0.385 0.360 0.434 550 0.010 0.011 0.050 0.065 0.093 0.123 0.187 0.238 0.313 0.385 0.361 0.435 575 0.010 0.011 0.050 0.066 0.093 0.124 0.188 0.239 0.314 0.386 0.362 0.436 600 0.010 0.011 0.051 0.066 0.094 0.124 0.190 0.240 0.315 0.386 0.364 0.436 625 0.010 0.011 0.051 0.066 0.095 0.125 0.191 0.241 0.316 0.386 0.365 0.436 650 0.011 0.011 0.052 0.067 0.095 0.125 0.193 0.243 0.317 0.386 0.366 0.437 675 0.011 0.011 0.052 0.067 0.096 0.126 0.194 0.243 0.317 0.386 0.367 0.437 700 0.011 0.011 0.053 0.068 0.097 0.126 0.195 0.245 0.318 0.387 0.368 0.438 725 0.011 0.011 0.054 0.068 0.098 0.127 0.196 0.245 0.319 0.388 0.369 0.439 750 0.011 0.011 0.054 0.068 0.099 0.127 0.198 0.246 0.320 0.388 0.370 0.440 775 0.012 0.011 0.055 0.069 0.100 0.128 0.199 0.247 0.321 0.388 0.371 0.440 800 0.012 0.011 0.056 0.069 0.101 0.129 0.200 0.249 0.323 0.391 0.372 0.440 825 0.012 0.011 0.057 0.070 0.102 0.130 0.202 0.249 0.324 0.391 0.374 0.442 850 0.013 0.011 0.057 0.071 0.102 0.131 0.203 0.251 0.325 0.391 0.375 0.442 875 0.013 0.012 0.058 0.072 0.104 0.132 0.204 0.251 0.326 0.391 0.376 0.442 900 0.013 0.012 0.059 0.072 0.104 0.133 0.206 0.252 0.327 0.391 0.377 0.441 925 0.014 0.012 0.059 0.072 0.105 0.133 0.207 0.252 0.328 0.391 0.378 0.441 950 0.014 0.012 0.060 0.072 0.106 0.134 0.208 0.252 0.329 0.391 0.379 0.442 975 0.014 0.013 0.061 0.073 0.107 0.134 0.209 0.254 0.331 0.392 0.380 0.443 1000 0.014 0.012 0.061 0.073 0.108 0.135 0.210 0.255 0.332 0.392 0.381 0.443 1025 0.014 0.012 0.062 0.075 0.108 0.135 0.210 0.256 0.333 0.393 0.382 0.443 1050 0.014 0.012 0.062 0.075 0.109 0.135 0.211 0.256 0.334 0.392 0.383 0.444 1075 0.014 0.012 0.062 0.075 0.11 0 0.136 0.212 0.257 0.334 0.394 0.383 0.444 1100 0.015 0.012 0.063 0.076 0.11 0 0.138 0.213 0.257 0.336 0.396 0.385 0.445 Table 4. Reflectance values for gray Spectralon under 50%R. To use the tables, look at the wavelength of interest and find the value under the column marked L2 corresponding as close as possible to sample reflectance you measured with your single beam accessory. These values were determined with a PerkinElmer Lambda 2 double beam spectrometer and an RSA-PE-20 single beam accessory. Compare the charted L2 value and your measured reflectance to the reflectance value under the column marked L19 immediately to its left. The value under L19 is the actual value of reflectance of the sample at that wavelength - you may need to interpolate. These values were determined with a Lambda 19 spectrometer and RSA-PE-19 double beam accessory. The standards used in generating these tables were spectrally flat over a wide wavelength range. Supplemental information is available for some materials along with the slope in their reflectance curves. This technique is the same used for measurements of standards at the National Institute of Standards and Technology (NIST), the National Physical Laboratory (NPL) and the National Research Council Canada (NRCC). The data is available at Labsphere on disk. Be advised that these tables should used for diffusely reflecting materials only. AQ-00073-000, Rev. 7 19

Wavelength Gray14/L2 Gray14/L19 Gray16/L2 Gray16/L19 Gray18/L2 Gray18/L19 Gray20/L2 Gray20/L19 Gray22/L2 Gray22/L19 Gray23/L2 Gray23/L19 300 0.512 0.571 0.636 0.691 0.730 0.773 0.791 0.825 0.895 0.912 0.956 0.964 325 0.505 0.571 0.631 0.691 0.725 0.774 0.792 0.827 0.898 0.915 0.965 0.971 350 0.500 0.569 0.627 0.692 0.722 0.775 0.790 0.829 0.899 0.917 0.970 0.975 375 0.495 0.568 0.624 0.691 0.718 0.771 0.789 0.827 0.898 0.917 0.974 0.978 400 0.492 0.568 0.623 0.692 0.716 0.774 0.789 0.832 0.899 0.918 0.978 0.983 425 0.491 0.567 0.622 0.693 0.715 0.775 0.790 0.833 0.899 0.918 0.980 0.985 450 0.490 0.568 0.623 0.693 0.714 0.774 0.791 0.833 0.900 0.920 0.981 0.986 475 0.491 0.568 0.624 0.695 0.715 0.775 0.793 0.835 0.902 0.921 0.983 0.987 500 0.491 0.569 0.624 0.696 0.715 0.776 0.794 0.836 0.902 0.921 0.983 0.987 525 0.492 0.570 0.626 0.696 0.716 0.775 0.795 0.837 0.903 0.922 0.983 0.988 550 0.493 0.570 0.627 0.698 0.716 0.777 0.797 0.839 0.904 0.921 0.983 0.988 575 0.495 0.570 0.628 0.698 0.717 0.776 0.798 0.840 0.905 0.924 0.983 0.987 600 0.496 0.571 0.629 0.698 0.718 0.776 0.800 0.840 0.906 0.923 0.983 0.988 625 0.497 0.572 0.631 0.699 0.719 0.777 0.802 0.841 0.907 0.924 0.983 0.987 650 0.498 0.572 0.632 0.699 0.719 0.776 0.803 0.841 0.907 0.926 0.983 0.988 675 0.499 0.572 0.633 0.701 0.720 0.777 0.805 0.843 0.908 0.924 0.983 0.988 700 0.500 0.574 0.634 0.702 0.721 0.778 0.806 0.843 0.909 0.925 0.983 0.988 725 0.501 0.575 0.635 0.702 0.721 0.776 0.807 0.844 0.910 0.926 0.983 0.988 750 0.503 0.576 0.637 0.703 0.722 0.778 0.809 0.845 0.911 0.926 0.983 0.988 775 0.504 0.575 0.638 0.704 0.723 0.778 0.810 0.846 0.912 0.926 0.984 0.987 800 0.506 0.576 0.640 0.704 0.724 0.777 0.812 0.846 0.913 0.931 0.984 0.990 825 0.507 0.578 0.641 0.704 0.725 0.777 0.813 0.848 0.913 0.930 0.983 0.988 850 0.508 0.578 0.642 0.704 0.725 0.777 0.813 0.848 0.913 0.930 0.982 0.990 875 0.509 0.578 0.644 0.705 0.726 0.778 0.814 0.848 0.912 0.929 0.981 0.988 900 0.511 0.578 0.645 0.706 0.726 0.779 0.814 0.851 0.912 0.930 0.980 0.988 925 0.512 0.579 0.646 0.707 0.727 0.781 0.815 0.851 0.913 0.930 0.980 0.987 950 0.513 0.579 0.647 0.709 0.728 0.781 0.817 0.851 0.914 0.930 0.980 0.988 975 0.515 0.579 0.648 0.710 0.728 0.781 0.818 0.851 0.914 0.931 0.981 0.989 1000 0.516 0.579 0.650 0.710 0.730 0.782 0.820 0.854 0.916 0.931 0.981 0.988 1025 0.517 0.580 0.651 0.711 0.730 0.781 0.821 0.854 0.917 0.932 0.982 0.986 1050 0.517 0.580 0.652 0.712 0.729 0.781 0.818 0.855 0.912 0.932 0.976 0.988 1075 0.518 0.582 0.652 0.713 0.728 0.783 0.816 0.854 0.909 0.931 0.971 0.986 1100 0.519 0.581 0.654 0.714 0.729 0.783 0.817 0.856 0.910 0.932 0.972 0.988 Table 5. Reflectance values for gray Spectralon over 50%R. An example of the use of these tables is shown in Appendix C. In these measurements, Labsphere SCS Spectralon color standards were measured using a dual beam spectrometer with a single beam sphere accessory and a double beam spectrometer with a double beam sphere accessory. The samples were chosen based on their wide reflectance variation across the wavelength spectrum. In each case, the bottom curve represents the measured spectrum generated on the single beam sphere. The boxed curve represents the corrected values using the lookup tables. The light curve is the actual reflectance of the standard as measured on the Lambda 19. The corrected values, using the Substitution Error Correction Tables, give an excellent correlation with actual double beam measurements. Zeroline Correction An integrating sphere is sensitive to small-angle scatter from the sample beam coupling optics. The scattered radiation strikes the wall of the integrating sphere near the sample reflectance port, creating a "halo" surrounding the port. This "halo-effect" gives rise to a small error in measurement of reflectance factor which is most significant when measuring samples of very low reflectance. Fortunately, this error is relatively easy to characterize and correct. The following procedure may be used to correct reflectance factor measurements of diffuse or specular samples, in either the 8 /Hemispherical or 0 /Diffuse geometries: AQ-00073-000, Rev. 7 20

1. Perform a reflectance factor measurement of a given sample as described in the operating procedures. 2. Replace the sample with a light trap. The reflectance of the light trap is approximately 0%. 3. Execute a sample scan to record the measurement data on the light trap. 4. Compute the corrected reflectance factor for the sample, ρ s, as follows: ( R Z)ρ r ρ s = ----------------------- ( 100 Z) where: ρ r = the reflectance value of the reference standard, R = the sample data displayed on the instrument expressed in percent, and Z = the data from the light trap measurement, expressed in percent. Note: This zero-correction procedure can also be used in conjunction with a mask at the sample reflectance port to accurately measure the reflectance factor of samples smaller than the sample port dimensions. Operating Procedures for the RSA-PE-20 Accessory The RSA-PE-20 accessory can measure reflectance factors at either 0 or 8 angles of incidence. The angle of incidence is controlled by the sample holder configuration at the sample reflectance port. You should keep in mind, however, that 8 hemispherical measurements will include the specular component reflecting off the sample surface. Although calibrations using the 0 geometry are available on a custom basis, the standard calibrations at Labsphere are performed at 8. If your application requires NIST traceability, you should use a calibrated reflectance standard and the 8 reflectance geometry. Since any integrating sphere accessory acts as a signal attenuator, you should run scans at a slower scan speed than in ordinary transmittance work without the accessory, or increase the integration time if collecting data at a fixed wavelength. 8 /Hemispherical Reflectance Factor Measurements Hemispherical reflectance measurements can be made on diffuse or specular samples - the procedure is the same. If you are measuring a specular sample, a specular reference standard, such as a calibrated first-surface aluminum mirror, should be substituted for the diffuse reference standard. 1. Secure the 8 reflectance sample holder over the reflectance sample port. 2. Load a calibrated or uncalibrated reflectance standard at the sample reflectance port. AQ-00073-000, Rev. 7 21

3. Choose your reflectance sample method from the UV Winlab software or manually set all instrument parameters as desired. 4. Perform a background correction scan. 5. Replace the reflectance standard at the sample reflectance port with your sample. 6. Execute a sample scan and save the scan data. 7. Multiply the values recorded in the resulting scan by the known spectral reflectance factor values of the reflectance standard to determine the actual reflectance of the sample. If your reflectance standard is an uncalibrated USRS-99-010 from Labsphere, you can use the reflectance data in Appendix B. 0 /Hemispherical Reflectance Factor Measurements The integrating sphere of the accessory has two sample holders for the sample reflectance port. The standard sample holder is designed to position the sample for illumination by the sample beam at an 8 incident angle. The second sample holder is designed for normal, 0 incidence. With the 0 sample holder in place, any specular component of reflection from the sample is excluded from measurement, since this component is directed out of the sphere through the transmittance sample entrance port. Therefore, the 0 sample holder may be used to make 0 /d reflectance factor measurements. The procedure for such measurement is as follows: 1. Remove the 8 sample holder from the reflectance sample port and install the 0 sample holder. 2. Load a diffuse reflectance standard at the sample port. 3. Choose your reflectance sample method from the UV Winlab software or manually set all instrument parameters as desired. 4. Perform a background correction scan. 5. Replace the reflectance standard at the sample reflectance port with your sample. 6. Execute a sample scan and save the scan data. 7. Multiply the values recorded in the resulting scan by the known spectral reflectance factor values of the reflectance standard to determine the actual reflectance of the sample. If your reflectance standard is an uncalibrated USRS-99-010 from Labsphere, you can use the reflectance data in Appendix B. Note: Measurements performed using the specular excluded geometry should always be interpreted with caution. The transmittance port may not be completely effective in removing the specular reflection component. Reflectance Measurement of Non-Ideal Samples Two simple ideal geometric reflectance distributions may be described: perfect specular reflection, and perfect diffuse reflection. Perfect specular reflection is simply reflection according to Fresnel's law, without scattering. Perfect diffuse reflection is reflection according to Lambert's law. For the purposes of integrating-sphere-based reflectometry, the Spectralon standards provided with this accessory serve as an excellent approximation to the ideal of perfect Lambertian reflection, while a clean, flat, first-surface mirror provides an excellent approximation to the AQ-00073-000, Rev. 7 22

ideal of perfect specular reflection. If your sample approaches perfection, in either diffuse or specular reflectance distribution, your scan data using the procedure above will be accurate. If your sample is not perfectly specular or diffuse, you may decide simply to measure the sample as if it were either nearly specular or Lambertian, accepting the fact that such measurements involve a higher-than-ordinary degree of systematic uncertainty. Such an approach is more than adequate for many applications, especially those in which the emphasis is on repeatability rather than absolute accuracy of measurement. Another alternative might be to scan the sample at the 8 geometry using the diffuse and specular standards, and take the average reflectance of the two. A more ambitious approach involves modeling the geometric reflectance distribution of the sample in question as an ideal "mixed" distribution, composed of two components, one of which is perfectly specular, and the other, perfectly Lambertian. The sample is measured in both total hemispherical (specular included) and diffuse (specular excluded) geometries, and the relative magnitude of the two components is estimated from the results. For the RSA-PE-20 accessory, the specular excluded geometry is at normal incidence where the specular component is rejected from the sphere back through the transmittance port. Systematic errors involved in measuring a partly-specular sample against a diffuse reference, or a partly-diffuse sample against a specular reference, can then be corrected in accordance with this model. An example of such an approach is presented below. 1. A first-surface mirror is measured with the instrument in the standard hemispherical configuration, first using a specular reference standard (M), then using a diffuse reference standard (M'). The geometric correction factor γ is calculated as follows: γ = M M' Eq. 7 2. Next, the test sample is measured, using a diffuse reflectance standard, in both the standard hemispherical, specular included, and diffuse, specular excluded, configurations. These test sample results will be identified as R SIN and R SEX, respectively. 3. Given the assumption that the geometric reflectance distribution of the test sample can be separated into two ideal components, one specular (S) and one perfectly Lambertian (D), and further assuming that the diffuse component of the test samples above approximates a perfect Lambertian distribution, so that its reflectance factor (ρ o ) is the same for both specular-included and specular-excluded geometries, the following relations hold: R SEX = D ρ 0 and R SIN = ( S' + D) ρ 0 Eq. 8 4. Rearranging the relation above, we get the following definitions for the specular and diffuse components: AQ-00073-000, Rev. 7 23

D = R SEX ρ 0 and S' = ( R SIN R SEX ) ρ 0 Eq. 9 where S' represents the contribution of the specular component of the test sample to the apparent hemispherical reflectance factor of the sample. This quantity is analogous to the value M' given in Eq. 7 and its relation to the actual specular component of hemispherical reflectance (S) is described in Eq. 10: S = γs' = γ( R SIN R SEX )ρ 0 Eq. 10 5. On this analysis, the best estimate for the 8 /hemispherical reflectance factor of the test sample (R T ), is given as follows: R T = S+ D = ( γr SIN + ( 1 γ)r SEX )ρ 0 Eq. 11 Transmittance Measurements ASTM E 179-91a describes three categories of transmittance measurement: regular transmittance, diffuse transmittance, and total transmittance. These three terms relate to each other in the same fashion as regular or specular, diffuse and total reflectance. Regular transmittance is the ratio of the undiffused transmitted flux to the incident flux. Diffuse transmittance is the ratio of transmitted flux measured at all forward angles except the regular transmittance angle, to the incident flux. Total transmittance is the ratio of flux transmittance at all forward angles to the incident flux. An illustrative description from ASTM E 179-91a is shown in Figure 10. Impinging Light Beam Retroreflection Diffuse Transmittance Diffuse Reflectance Regular Transmittance Specular Reflection Figure 10. Components of reflectance and transmittance phenomena. AQ-00073-000, Rev. 7 24

The following procedure can be used to measure total transmittance of a sample. 1. Load a diffuse reflectance standard at the sample reflectance port. 2. If you are scanning a liquid sample, load an empty cuvette into the cuvette holder, and load the entire cuvette holder assembly at the transmittance port. Otherwise, leave the transmittance port empty. 3. Choose your transmittance sample method from the UV Winlab software or manually set all instrument parameters as desired. 4. Perform a background correction scan. 5. Load your sample at the transmittance port. If you are measuring a liquid, replace the empty cuvette with your sample cuvette. 6. Execute a sample scan and save the scan data. 7. The value displayed by the instrument is the sample transmittance. Note: When measuring the transmittance of large samples, make sure that the reference or sample beam is not obscured by any part of the sample, and that the sample compartment lid closes completely. The procedure here for total transmittance measurement will produce accurate results for nonscattering samples. For transmittance samples exhibiting high degree of scattering, the scan results will be instrument-specific. You can measure the diffuse transmittance of a test sample using the following procedure: 1. Perform a background correction scan on the accessory with a diffuse reflectance standard loaded at the sample reflectance port, and an empty cuvette, if used, fitted in front of the transmittance port. 2. Replace the reflectance standard at the sample reflectance port with a light-trap. This light trap will capture and exclude from measurement the undiffused portion of the flux transmitted by the sample. 3. Load the test sample or sample cuvette at the transmittance port and execute a sample scan. An approximate regular transmittance of a sample can be determined by measuring the total and diffuse transmittance, as described above, and subtracting the diffuse component from the total transmittance. AQ-00073-000, Rev. 7 25