Product Data Sheet Technical Publications SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET for ARTIS Q / Q.ZEN / ZEE BIPLANE IECAX480V225A Main Disconnect Panel
Bevco Engineering Company, Inc. CONFIDENTIAL AND PROPRIETARY This Bevco Engineering Company document contains confidential and proprietary information of Bevco Engineering Company. The intellectual properties contained herein may not be used or copied to the detriment of Bevco Engineering Company. The information contained in this document may not be copied without the express written approval of Bevco Engineering Company. CAUTION! Troubleshooting and servicing should be performed by a qualified electrician. WARNING! De-energize and lockout the main disconnect before servicing this panel. Table of Contents 1.0 DESCRIPTION OF OPERATION... 3 SYSTEM CONFIGURATION... 5 2.0 SPECIFICATIONS... 6 Electrical Specifications... 6 Seismic Specifications:... 6 Electrical Wiring Diagram... 6 Environmental Specifications:... 6... 6 4.0 PANEL DIMENSIONS... 7 6.0 DOCUMENT REVISION HISTORY... 8 COMPONENTS SUPPLIED WITH EACH PANEL: The IEC Main Disconnect Panel This Installation, Operations & Service Manual (3) sets of Emergency Power Off pushbuttons with 3NC contacts on each EPO Drawings and Electrical Schematics IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 2 of 8
1.0 DESCRIPTION OF OPERATION INTEGRATED ELECTRICAL CABINET This panel incorporates several features desirable for system installations to minimize down time, protect the generator electronics, and to reduce operational delays after a power outage. This product replaces the power panel, the lighting control relay box, and the Emergency Power Off controls. The panel has a main circuit breaker, Q1 with individual branch breakers for each generator and various cooling options. When the main circuit breaker is turned off, all power circuits within the panel will be de-energized except the power circuits for the Room Light control and the X-Ray ON warning lights. The conductors that are fed from other sources are wired using yellow conductors. The generators power is protected by electronic circuit breakers, Q2 and Q3, and is controlled by contactors, K2 and K6. The contactors will open with any loss of power or by pressing any Emergency Power Off (EPO) pushbutton. The K2 and K6 contactor control circuits are factory configured to automatically re-energize the generators upon restoration of facilities power. The control circuit may be re-configured to require the operator to manually restart the generators once the incoming power has been restored. This protects the sensitive electronic circuits of the generators from sags and surges that immediately follow power loss from blackouts, storms, utility reclosure operations, and out of phase automatic transfer switch operations. The SC1 cabinet power is protected by an electronic circuit breaker, Q4, and is controlled by contactor, K7. The contactor will open with any loss of power or by pressing any Emergency Power Off (EPO) pushbutton. The K7 contactor control circuit are factory configured to automatically re-energize the SC1 cabinet upon restoration of facilities power. The control circuit may be re-configured to require the operator to manually restart the SC1 cabinet once the incoming power has been restored. This protects the sensitive electronic circuits of the generators from sags and surges that immediately follow power loss from blackouts, storms, utility reclosure operations, and out of phase automatic transfer switch operations. The control circuits are low voltage 24 VDC and are fully powered from within the panel. The restart functionality and EPO circuitry is controlled with a safety relay, K10. See page 11 for detail. The white SAFETIES OK indicator light on the front of the panel is illuminated when none of the EPOs are pressed and the safety relay K10 is reset. When the white light is active, pressing the green START pushbutton will cause the Angio system to be energized. The green START button will illuminate, and the white SAFETIES OK light will go off. Pressing the STOP button will de-energize the imaging system. Any EPO pressed while the generator systems are energized will result in the immediate de-energizing of the Angio system. If an EPO is pressed at any time, the EPO must be reset which will cause the SAFETIES OK light to activate. Then the START button will activate the imaging system. IMPORTANT: If building power is removed from the panel while the system is energized, the system will re-energize when building power is restored without any human interaction. Additional provisions are made to integrate the X-Ray ON warning lights and room lighting with the X- Ray Equipment. The facility lighting panel provides 120- or 277-volt power that is controlled by contacts relays of K4 and K5 mounted in the IEC. The signal controlling the relays comes directly from the Siemens Generator/Power Cabinet. The IEC will accept signals from the generator at either 24 Volts AC or DC. IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 3 of 8
IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 4 of 8
UTILITY SUPPLY 480V 3-PHASE 180A MAX Q1 225A WIRESIZE: 4AWG- 350MCM SYSTEM CONFIGURATION KEY TO COMPONENTS: Q1 MAIN CIRCUIT BREAKER Q2 GENERATOR 2 CIRCUIT BREAKER Q3 GENERATOR 1 CIRCUIT BREAKER F2 POWER SUPPLY FUSES F4 DC CIRCUIT BREAKER G01 DC POWER SUPPLY K1 SAFETY CONTACTOR K2 GENERATOR 2 CONTACTOR K4 ROOM LIGHT RELAY K4A ROOM LIGHT RELAY 24VDC K5 X-RAY ON WARNING LIGHT RELAY K5A X-RAY ON WARNING LIGHT RELAY 24VDC K6 - GENERATOR 1 CONTACTOR K7 SYSTEM CABINET RELAY K10 SAFETY RELAY Q2 100A F2 4A Q4 30A Q3 100A G01 DC POWER SUPPLY K7 95-120A WIRE SIZE: 10-2/0AWG F4 5A SC1 DOOR SWITCH ROOM LIGHT X-RAY ON 30A MAX K2 95-120A MAX WIRE SIZE: 10-2/0AWG DC CONTROL CIRCUITS K1 SAFETY CONTACTOR K4 ROOM LIGHT RELAY K4A ROOM LIGHT RELAY 24VDC K5 X-RAY ON WARNING LIGHT RELAY K5A X-RAY ON WARNING LIGHT RELAY 24VDC K10 SAFETY RELAY EMERGENCY POWER OFF CIRCUITS X-RAY ON WARNING LIGHT 20A MAX LIGHTING POWER PANEL 120/277V K6 95-120A WIRE SIZE: 10-2/0AWG GENERATOR 2 80A MAX EPO EPO EPO ROOM LIGHTS 20A MAX GENERATOR 2 80A MAX IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 5 of 8
2.0 SPECIFICATIONS Electrical Specifications UL design and ratings of this panel require the application of wire at the 75 C ampacity ratings of NEC table 310.16. Wire rated at higher temperature ratings, such as 90 C, may be used but ONLY applied at the 75 C ampacity ratings of NEC Table 310.16. Catalog Number Incoming Voltage & Frequency Incoming Power Configuration Main Circuit Breaker Rating Regulatory Markings IECAX480V225A 480VAC, 50/60Hz 3Ø + Ground + Neutral 225 Amps culus Short Circuit Current Rating is 35,000 Amperes RMS Symmetrical @ 480V Seismic Specifications: This Integrated Electrical Cabinet has been certified by an independent California structural engineer in conformance with the shake testing requirements of ICC- AC 156. The California OSHPD number is OSP-0457-10. The seismic performance characteristics are as follows: SDS(g) 2.56; z/h 1.0 ; Ip 1.5 Electrical Wiring Diagram The Electrical Wiring Diagram is adhered to the inside of the enclosure door for convenient reference to the circuitry. A complete set of the panel drawings are also attached to the back of this manual and shipped with the disconnect panel. Environmental Specifications: FOR INDOOR USE ONLY Temperature 59-90 F (15-32 C) Humidity 30-75%, NON-CONDENSING PARTS, TECHNICAL SUPPORT AND SERVICE FOR INQUIRIES DURING INSTALLATION CONTACT BEVCO ENGINEERING, SUSSEX WI (262) 820-2400 FOR ISSUES FOLLOWING INSTALLATION, CONTACT SIEMENS UPTIME CENTER, (800) 888-7436 IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 6 of 8
4.0 PANEL DIMENSIONS Refer to the Table and Illustrations below for the physical dimensions of the panel. The panel may be recessed into the wall for semi-flush installations. See Illustrations below for typical front and side views with the mounting hole pattern. Refer to the drawing set provided with the panel for total weight and center of gravity dimensions. PANEL EXTERNAL PHYSICAL DIMENSIONS Parameter Dimensions Height 36.0 inches 914 mm Width 30.0 inches 762 mm Total Depth 9.1 inches 230 mm Enclosure Depth 8.0 inches 203 mm Mounting Hole Diameter 7/16 inches 11.1 mm Mounting Hole Width 28.5 inches 724 mm Mounting Hole Height 34.5 inches 876 mm Weight 157 pounds 71.2 kg IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 7 of 8
6.0 DOCUMENT REVISION HISTORY REV DATE AUTHOR DESCRIPTION OF CHANGE 1 04/19/2014 RJB Initial Release 2 02/28/2015 RJB Changed Artis name on first page 3 03/09/2015 XYUE Update manual based on current schematic 4 03/19/2015 JRB Added additional Provision on page 3 50 11/01/2016 DRO IEC Redesign 51 07/10/18 JT Updated system configuration data IECAX480V225A Product Data Sheet_R51 Siemens IEC Manual For AX Systems Page 8 of 8