Project Goals: Particle size distribution with shape information (roundness, aspect ratio) with highest precision and accuracy.

Similar documents
The Canty SolidSizer TS system was selected for the analysis of the Sugar crystals. Below are the details regarding the testing.

Linescan System Design for Robust Web Inspection

JM Canty International. Sugar Industry

Analysis of Proppants

TechSpec Large Format. # mm FL f/2. Techspec Large Format Fixed Focal Length Lens.

INFINITY-CORRECTED TUBE LENSES

Washability Monitor for Coal Utilizing Optical and X-Ray Analysis Techniques

Particle size range (0.8 microns 1000 microns)

Magnifiers. Available from : SSPA

Rodenstock Products Photo Optics / Digital Imaging

AP Physics Problems -- Waves and Light

TECHSPEC COMPACT FIXED FOCAL LENGTH LENS

Measurements using three-dimensional product imaging

Particle Size Analyzers

Color and Shading. Color. Shapiro and Stockman, Chapter 6. Color and Machine Vision. Color and Perception

FIXED FOCAL LENGTH LENSES

Supporting Information. High-Throughput, Algorithmic Determination of Nanoparticle Structure From Electron Microscopy Images

TELECENTRIC LENSES INNOVATION STARTS HERE... Global Design & Support Rapid Prototyping Volume Manufacturing & Pricing

TELECENTRIC LENSES INNOVATION STARTS HERE... Global Design & Support Rapid Prototyping Volume Manufacturing & Pricing

Rodenstock Products Photo Optics / Digital Imaging

Weatherproof IR Color Day/Night Cameras

Development of Image Processing and Analyzing Software for Nuclear Fuel Pellet End Faces Mandar Joshi a, Praveen Deshpande b, and C. P.

CALIBRATION ASPECTS OF MULTIPLE LINE-SCAN VISION SYSTEM APPLICATION FOR PLANAR OBJECTS INSPECTION

An Intuitive Explanation of Fourier Theory

Miniaturized Camera Systems for Microfactories

MeVis-C, MeVis-Cm and MeVis-CF. High-Resolution Inspection Lenses

Discovery C15 Color Camera

Automated Particle Size & Shape Analysis System

LN-TCL1 & LN-TCL1-LRF

User Manual. Binocular Zoom Stereo Microscope with Boom Stand. MicroscopeNet.com

CAMSIZER XT - the latest in Dynamic Image Analysis

501, , 1052, , 1602, 1604 EXCEL EXCEL 1602UC EXCEL 1052UC EXCEL 501HC. Micro-Vu Corporation. Precision Measurement Systems

Optilia Instruments. Empowering Your Vision!

Ch 22 Inspection Technologies

GRAPHIC-VISION RX300 Digital Defect Detection System

UNIT 102-9: INTERFERENCE AND DIFFRACTION

Powerful & homogeneous projector

High definition digital microscope. visioneng.us/lynxevo

Micro Tech Lab. Operating Instruction LM DSLR Photomicroscope. Depending on the microscope configuration

Particle size and shape characterization by Dynamic Image Analysis

Image Analysis: Evaluating Particle Shape

AV300 VIDEO MEASURING SYSTEM.

P recise Eye. High resolution, diffraction-limited f/4.5 optical quality for high precision measurement and inspection.

Multisensor Coordinate Measuring Machines ZEISS O-INSPECT

Grade 8. Unit 3 Light and Optical Systems

SES123 Computer Methods Lab Procedures

Particle Shape Determination in a Sedimaging Device

Full-HD (1080p) digital microscope. Stunning full-hd 1080p/60fps live video image quality

DeltaPix Modus M12Z Digital microscope

Traffic Sign Recognition Senior Project Proposal

Home Lab 7 Refraction, Ray Tracing, and Snell s Law

Variable Zoom Lenses* USB High-Resolution Camera

I/ Video Capture 1. Remove the len s cover 2. Turn on Computer 3. Open the XCAP For Window icon OK

ZOOM STEREOMICROSCOPE SZ61/SZ51. For Biological Use. The Comfort Zone

Multisensor Coordinate Measuring Machines ZEISS O-INSPECT

EyeTech. Particle Size Particle Shape Particle concentration Analyzer ANKERSMID

Line Scan Lens. XENON-DIAMOND 2.7/111, beta = -2.6x

All Series Video Measuring Machine

Design and production of stereoscopic instruments. User manual for the 3-D macro lens model General. Exposure

Introduction.

Eye-Pal Vision Quick Start Rev. A

Algorithm User Guide:

12X Zoom. Incredible 12X (0.58-7X) magnification for inspection of a wider range of parts.

Always a new perspective. IEEE1394 DIgITAL CCD AND CMOS CAMERA LINE

Swift SM100 Series Stereo Microscope

Autofocus Systems in Machine Vision. Thomas Schäffler

Quick Scope. CNC/Manually-operated Vision Measuring Microscope

e-bridge Color Profile Tool Quick Start Guide

Optilia Instruments. Empowering Your Vision! Soporte y Garantía Local en México

BOLERO. Hair volume/surface measurement fly-away/frizz analysis system. Sebastien BREUGNOT & Robert GEORGE Bossa Nova Technologies

Independent Resolution Test of

TECHSPEC UC SERIES FIXED FOCAL LENGTH LENSES

Sharp Eyes for Large Format Photography

Precise Dynamic Measurement of Structures Automatically Utilizing Adaptive Targeting

Measurements of Flexible Displays - Challenges and Solutions -

Infrared Dome CCD Camera User Manual

Lecture 4 Image Enhancement in Spatial Domain

System for assisted inspection of stents

Pixel Size Calibration of Video Probe Measuring Machines James G. Salsbury Center for Precision Metrology University of North Carolina at Charlotte

POME A mobile camera system for accurate indoor pose

Model CI-P170. Citadel Panoramic Situational awareness camera for armoured vehicles Operation Manual - Preliminary. PAL and NTSC Versions.

Counting Particles or Cells Using IMAQ Vision

Optimized Design of 3D Laser Triangulation Systems

4Kx4K CCD Imager for the 3.6m DOT

Low Mag Video Lenses. Low Mag Video Lenses. Video lenses for machine vision inspection. Macro zoom lenses for close-up inspection.

E3S-A. Built-in Amplifier Photoelectric Sensor (Medium Size) Ordering Information. Built-in Amplifier Photoelectric Sensors. Horizontal. 7 m.

M40 Microscope User s Manual

scs1 highlights Phone: Fax: Web:

Product information. Hi-Tech Electronics Pte Ltd

PATVIS APA. Process analytical technology visual inspection system for automated particle analysis. Computer Vision Systems

Textbook Reference: Physics (Wilson, Buffa, Lou): Chapter 24

Using the Large Linear Verification Kit

Module 7 - Housing Bubble Dashboard - Part 1-1

Intui ve easy image capture and documenta on. Stand alone, wireless or PC connec vity. High quality stands for precision and stability

Digital Microscope LDM-B1 SERIES

Using Edge Detection in Machine Vision Gauging Applications

Original grey level r Fig.1

OIW-EX Series of Oil In Water Analyzers MiView Handbook Document code: OIW-HBO-0013 Version: EX July 2013

FCM-series Lenses for Luminus Devices SST-50 and SST-90 LEDs

All forms of EM waves travel at the speed of light in a vacuum = 3.00 x 10 8 m/s This speed is constant in air as well

Transcription:

CANTY PROCESS TECHNOLOGY 60 Donner Road Lockport, NY 14094 Phone: (716) 62-4227 Fax: (716) 62-4228 e-mail: sales@jmcanty.com Lab Test Report Sample Identity- Granular material described as: Purpose: SAM1 Pellet Sample 1 SAM2 Pellet Sample 2 SAM3 Pellet Sample 3 The dry granular particles were sized on a glass plate using a CantyVision Particle Size Analysis software. This report describes the measurement set up and presents measurement results. Project Goals: Particle size distribution with shape information (roundness, aspect ratio) with highest precision and accuracy. Lab Setup: The samples provided were analyzed for size and shape using a Canty SolidSizer sensor, shown in Figure 1. The granular material enters the sensor at the flange fitting seen at the top left in the illustration. An adjustable down tube deposits a thin layer of granules on a vibrating feeder tray. The granules travel along the tray and drop in a thin curtain at the discharge end of the tray. A camera observes the falling curtain of particles. The camera output for the free falling particles is a silhouette image of each particle in front of a lighted panel. The Canty Vector running the CantyVision Software processes these images. The Color measurements were made using the Color Speck Analysis System shown in Figure 2. The granular material enters the sensor at the hopper-funnel at the top right. An adjustable down tube deposits a thin layer of granules on a vibrating feeder tray. The granules travel along the tray and are observed by a color camera from above. Uniform white light is provided to illuminate the material traveling on a white tray. 1

Figure 1, SOLIDSIZER Industrial Solid Particle Analysis System sensor assembly includes light source and camera-lens Calibration: The SolidSizer sensor optical magnification can be adjusted over a wide range. For this application, magnification was adjusted to a magnification to permit detection of the 2 mm granular material. A pixel scale factor of 0.0273414 mm per pixel was used for all sample material. The resulting process image field of view (FOV) is 17.0 mm horizontal by 13.12 mm vertical and the solid particles appear dark with a bright background. Results: Figure 3 below shows a typical process image captured for SAM-1 in the SolidSizer. The test scan processing for the Figure 3 image in seen in Figure 4 illustrating how the Canty Software separates the particles from the background and measures the Area, Perimeter, Major Axis, and Minor Axis for each detected particle. The table included in this screen image lists the dimensions, in mm, of the particles imaged in the figure. The computer screen image in Figure also contains a selected individual particle, which is identified by a yellow box surrounding it. The dimensions of this particular particle are shown on the first row of the table. This data (area, perimeter, major axis and minor axis) is written to a text file during each run time and evaluated using a Microsoft Excel template. From these data measurements, plots can be created using either major or minor axis as a size basis. For comparison to sieve screen data, minor axis has been selected for reporting here. 2

Figure 3, Typical Process Image of SAM1 Granules 3

Figure 4, Test Scan of Figure 3 by Canty Vision Software detects three particles, report size in mm The center particle was selected by clicking on the particle. The selected particle is moved to top of Results listing in Figure, and is highlighted blue to allow careful identification of each particle. When these visual verify functions seen in Figure 4 and are satisfactory, a measurement data run of the sample is started to gather a full data set for about 2,000 particles. 4

Figure 7, Test Scan of 13 particles, report size in mm Figure 8, Scan Results Table of Figure 7 shows size and average color for detected particles

Percent 0 9 90 8 20 particles 80 7 D0 1.974 mm 70 6 SGN 197 60 0 4 40 3 30 2 20 1 0 1 1.2 1. 1.7 2 2.2 2. 2.7 3 3.2 3. Minor Axis, mm Figure 9, Volume Based Size distribution Results, SAM-1 The Percent Passing plot is computed from the volume of each particle and the D0 minor axis size is where 0 percent of particles are larger and 0 percent are smaller. 6

Percent less than 0 9 90 8 80 7 70 6 60 0 4 40 3 30 2 20 1 0 1 1.1 1.2 1.3 1.4 1. 1.6 1.7 1.8 1.9 2 Aspect Ratio Aspect Ratio by Volume AR 1.03 AR 0 1.19 AR 9 1.46 Average of all 20 particles AR 1.266 Figure, Aspect Ratio distribution Results, SAM-1 The Percent Passing plot is computed from the volume of each particle and the AR0 Aspect Ratio is where 0 percent of particles are larger and 0 percent are smaller. The AR, AR 0, and AR 9 are presented to characterize the shape or sharpness of the distribution curve. The volume based AR0 value of 1.19 is compared to the count based average Aspect Ratio of 1.266 and implies that the smaller particles have a higher aspect ratio. 7

Percent Less Than 0 9 90 8 80 7 70 6 60 0 4 40 3 30 2 20 1 Circularity = 4 Area / (Perimeter) 2 C0 0.779 by Volume Average Circularity 20 particles 0.7440 0 0.4 0.4 0. 0. 0.6 0.6 0.7 0.7 0.8 0.8 0.9 Circularity Figure 11, Circularity distribution Results, SAM-1 The Circularity by volume plot is computed from the volume of each particle and the C0 Circularity is where 0 percent of particles have higher circularity and 0 percent have smaller. The volume based C0 value of 0.779 is compared to the count based average Circularity of 0.7440 and implies that the smaller particles have a lower Circularity. This is expected from the Aspect Ratio characterization for the smaller particles. 8

Percent 0 9 90 8 80 7 4743 particles 70 6 D0 2.34 mm 60 SGN 23 0 4 40 3 30 2 20 1 0 1 1.2 1. 1.7 2 2.2 2. 2.7 3 3.2 3. 3.7 4 4.2 4. 4.7 Minor Axis, mm Figure 12, Volume Based Size distribution Results, SAM-2 The Percent Passing plot is computed from the volume of each particle and the D0 minor axis size is where 0 percent of particles are larger and 0 percent are smaller. 9

Percent less than 0 9 90 8 80 7 70 6 60 0 4 40 3 30 2 20 1 0 1 1.1 1.2 1.3 1.4 1. 1.6 1.7 1.8 1.9 2 Aspect Ratio Aspect Ratio by Volume AR 1.037 AR 0 1.9 AR 9 1.362 Average of all 4743 particles AR 1.16 Figure 13, Aspect Ratio distribution Results, SAM-2 The Percent Passing plot is computed from the volume of each particle and the AR0 Aspect Ratio is where 0 percent of particles are larger and 0 percent are smaller. The AR, AR 0, and AR 9 are presented to characterize the shape or sharpness of the distribution curve. The volume based AR0 value of 1.9 is compared to the count based average Aspect Ratio of 1.16 and implies that the smaller particles have a higher aspect ratio.

Percent Less Than 0 9 90 8 80 7 70 6 60 0 4 40 3 30 2 20 1 C0 0.780 by Volume Average Circularity 4743 particles 0.6372 0 0.4 0.4 0. 0. 0.6 0.6 0.7 0.7 0.8 0.8 0.9 Figure 14, Circularity distribution Results, SAM-2 Circularity The Circularity by volume plot is computed from the volume of each particle and the C0 Circularity is where 0 percent of particles have higher circularity and 0 percent have smaller. The volume based C0 value of 0.780 is compared to the count based average Circularity of 0.6372 and implies that the smaller particles have a lower Circularity. This is expected from the Aspect Ratio characterization for the smaller particles. Discussion: SHAPE CHARACTERIZATION Some particle processors and manufacturers need to characterize the particle shape after dividing the whole sample into a number of bins by size. There are varying specification limits for shape of particles in each size range. This more complex characterization output can be developed from the CantyVision measurement data and simply needs a Sort by Size function to group the particles to the desired bins before the shape calculation is done for each size bin. An Excel template could be developed if this level of shape characterization is needed. Conclusions: The Canty Vision Technology using the SolidSizer sensor and the Vector vision processor provides a Particle Size Distribution that characterizes the size granular materials by minor axis. This volume based size characterization can be directly compared to weight based size measurement methods such as sieve screening. The particle shape characterization available with this vision-based system provides information that is not available from sieve methods. The CANTY equipment provides excellent images for particle observation and shape analysis and can be used to characterize particles equally well in either online or lab environment. 11

12