Manual No Revision MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER

Similar documents
Instruction Manual HI Autoranging & Logging Portable Waterproof Microprocessor EC/TDS/NaCl/ C Meter

HI 2300 Auto-ranging Microprocessor EC/TDS/NaCl/ºC Bench Meter

USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements

USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter

Instruction Manual HI 9033 HI Waterproof Multi-Range EC and TDS Meters for Field Applications.

HI HI 9034 HI HI 9130 HI HI 9635 Portable Waterproof Multi-Range Conductivity/TDS/ Fertilizer Meters

Instruction Manual HI 2550 Multiparameter ph/orp/ C EC/TDS/NaCl Bench Meter

HI HI N HI HI

Instruction Manual Mi 170 Bench Meter

USER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe

USER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe

Instruction Manual HI HI Portable Microprocessor Printing and Logging Multi-Range Conductivity Meters

HI HI N HI HI

C100 Handheld Cond./TDS/Sal. Meter. Instruction Manual

Operation Manual. Hand-held ph/mv/temperature Meter. ph6810

WalkLAB HC Professional Conductivity-TDS-Salinity Meter. Operations Manual 15,0,11,4,1,14,5,10,13,2,9,6,3,12,7,8

Operations Manual EcoSense ph100

Instruction Manual HI Two Channels. EC/TDS/NaCl/Resistivity, Temperature Bench Meter.

HI HI HI HI Portable Microprocessor Thermistor Thermometers

HI ma Simulator

Operation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M

Datalogging Conductivity/ TDS Meter. Instruction Manual

TDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT. Pool Professional s Meter

HI ma Simulator

HI 8751 HI 8752 HI 8753 Portable Thermistor Thermometers

Tracer TM Salt/EC/TDS PockeTester Code 1749

Operations Manual EcoSense ph100a

Portable Conductivity Meter M1410. User Manual 1.002

DDS-12DW Benchtop Conductivity Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

Large Display Salinity & Conductivity Pens &

HI HI HI HI HI HI 98185

The Economy Conductivity Benchtop Meter

Operation Manual. MODEL ph6175. Microcomputer Based ph/orp/temperature Benchtop Meter

Operation Manual CX20 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller

User Manual DOCUMENT #601119REF. EcoSense ph1000a. ph, mv (ORP), and Temperature Benchtop Meter USER MANUAL. English

User Guide. ExStik FL700. Fluoride Meter

EC/TDS/SAL TRACERTM POCKETESTER CODE Conductivity/TDS/Salinity/Temperature/Meter

Instruction Manual. Multiparameter Benchtop Meter

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Instruction Manual HI 9835 HI Auto Ranging EC/TDS/NaCl/ Temperature Meters

Platinum RTD Thermometer MODEL NO

Operation Manual. Hand-held ph/orp/temperature Meter 6010M

Operation Manual CX10 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller

DO Meter Kit. Instruction Manual

HI HI HI HI HI HI HI

INSTRUCTION MANUAL. Multiparameter Bench Meter. Mi 180. ph/mv/ec/tds/nacl/temperature.

SoilStik ph Meter PRODUCT MANUAL

User Manual DOCUMENT #606042REF. EcoSense EC300A EcoSense EC300M Portable Conductivity, Salinity and Temperature Instruments. USER MANUAL English

WalkLAB ph mv/orptemperature

User's Guide. ExStik EC400. Conductivity / TDS / Salinity / Temperature Meter MODE HOLD

User Guide. ExStik II EC400. Conductivity / TDS / Salinity / Temperature Meter

Professional Waterproof Meters

Operation Manual. MODEL ph/ec80. Microcomputer Based ph/condctivity/tds/ Sanility & Temperature Pocket Meter. ph/ec80

Global Water Instrumentation, Inc.

ph 300 ph 301 ph 302 Microprocessor GLP Bench-top ph/mv/ise/ C Meters Instruction Manual

Platinum RTD Thermometer MODEL NO

Operation Manual MODEL 3173/3173R. MICROCOMPUTER BASED Conductivity/Salinity/TDS &Temperature BENCH METER 3173/3173R

Dissolved Oxygen Meter Kit DOK. Instruction Manual

Tracer ph/tds/salt PockeTester TM Code 1766

DIGITAL SOUND METER WITH JUMBO DISPLAY

AquaShock ph Meter K

User Guide. ExStik TM Model RE300 Waterproof ORP Meter. Patent Pending RE300 ORP

USER GUIDE. Salinity Meter Pen Style Water Quality Meter. Model EC170

Professional Benchtop Conductivity meter BC3020

HI 9828 Multiparameter Water Quality Meter...

Cole-Parmer Instrument Company

Instruction Manual CON 400 and CON 410 Waterproof Hand-held Conductivity/TDS Meter

Operation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller

MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL

OPERATION MANUAL MODEL TT-LP

INSTRUCTION MANUAL. Model Dual Input RTD Thermometer. Measures two temperatures simultaneously. Dual RTD probe inputs

HI 9828 Multiparameter Water Quality Meter...

HI HI HI HI Portable Microprocessor Printing and Logging PT-100 Thermometers

Oyster Meter ph-conductivity-tds-salinity-orp(mv)

DPH230SD USER S MANUAL

MULTI- PARAMETER TESTER

User Manual. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500. Additional User Manual Translations available at

Instruction Manual RS-1660

Global Water Instrumentation, Inc.

Handheld Water Quality Meters 100 Series

User's Guide. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500

User Guide. Moisture Meter. Model MO250

Instruction Manual Cond 6+ Conductivity/Temp TDS 6+ TDS/Temp Salt 6+ Salinity/Temp

HI98190, HI Calibration Check Waterproof ph/mv/ise/temperature Meters INSTRUCTION MANUAL

DIGITAL POCKET HEAT INDEX MONITORS USER S MANUAL

A150 Benchtop Conductivity Meter Instruction Manual

Introduction: Features:

Resolution selectable from 0.01 and ph. Range ph. Accuracy ± ph

HI 3220 HI 3221 HI 3222

Instruction Manual Temp JKT Temperature Meter Series


PRO 360. Owner's Manual

Conductivity and TDS Meter Pen Style Water Quality Meter

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

WK7200. Water Quality Test Kit USER S MANUAL

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet

Instruction Manual COND 600/610 Waterproof Handheld Conductivity Meter

STATIC METERS Models 211 & 212. SURVEYORSTAT Models 211XL & 212XL

VELOCICALC Air Velocity Meter

OPERATION MANUAL. JENCO MODEL 6312 CTB MICROCOMPUTER BASED Conductivity/TDS & Temperature TRANSMITTER & CONTROLLER

Transcription:

Manual No. 9835 Revision 0-401 MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER

Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. This meter is in compliance with European Community (CE) directives EN 50081-1 and EN 50082-1. TABLE OF CONTENTS Each meter is supplied complete with: GLI-76302W Conductivity/TDS Probe with built-in temperature sensor Instruction Manual AA size Alkaline Batteries (four) Rugged Carrying Case. NOTE: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in their original packaging together with the supplied accessories. PRELIMINARY EXAMINATION... 2 GENERAL DESCRIPTION... 2 FUNCTIONAL DESCRIPTION... 3 SPECIFICATIONS... 3 CONNECTIONS... 4 TAKING MEASUREMENTS... 4 AUTO-RANGING... 5 TEMPERATURE COMPENSATION... 5 CONDUCTIVITY/TDS CALIBRATION...5-6 NaCl CALIBRATION...6-7 TEMPERATURE CALIBRATION... 7 TEMPERATURE ADJUSTMENT...7-8 CONDUCTIVITY vs. TEMPERATURE CHART... 8 SETUP...8-9 BATTERY REPLACEMENT... 9 PROBE MAINTENANCE... 10 WARRANTY... 10 PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, notify the transit carrier immediately. GENERAL DESCRIPTION The GLI-9835 is a waterproof portable microprocessor-based conductivity/tds/nacl meter. The auto-ranging feature of the conductivity and TDS ranges automatically sets the meter to the scale with the highest possible resolution. The measurements are automatically (ATC) or manually (MTC) compensated for temperature. The temperature coefficient value is user selectable. It is possible to disable the temperature compensation and measure the actual conductivity. The meter is equipped with a stability indicator. The Battery Error Preventing System (BEPS) switches the meter off when the batteries are too weak to support proper function. Model GLI-9835 Portable Conductivity/TDS Meter 2 Rev. 0-401

FUNCTIONAL DESCRIPTION (see drawing below) 1) Power adapter socket 2) Probe connector 3) Liquid crystal display (LCD) 4) ON/OFF key, to turn the meter on and off 5) ALT key, to alternate key function 6) RANGE/FIXED key, to select measurement ranges or (with ALT key) freeze the current range on the LCD 7) CAL/CALT key, to enter calibration mode 8) CFM key, to move down or (with ALT key) confirm values 9) ATC/TC key, to select temperature compensation mode or (with ALT key) to view the temperature coefficient value 10) FNC key, to move up or (with ALT key) enter setup mode Range Resolution Accuracy (@20 C/68 F) Typical EMC Deviation Conductivity Calibration Conductivity (auto-ranging) TDS (auto-ranging) NaCl Calibration SPECIFICATIONS NaCl Temp. Cond. TDS NaCl Temp. Cond. TDS NaCl Temp. Cond. TDS NaCl Temp. 0.00 to 29.99 30.0 to 299.9 300 to 2999 3.00 to 29.99 ms/cm 30.0 to 200.0 ms/cm Up to 500.0 ms/cm* 0.00 to 14.99 ppm 15.0 to 149.9 ppm 150 to 1499 ppm 1.50 to 14.99 g/l 15.0 to 100.0 g/l Up to 400.0 g/l* 0.0 to 400.0% 0.0 to 60.0 C *When temperature compensation is disabled, actual conductivity or TDS reading is the non-temperature compensated value of the solution. 0.01 (0.00-29.99 ) 0.1 (30.0-299.9 ) 1 (300-2999 ) 0.01 ms/cm (3.00-29.99 ms/cm) 0.1 ms/cm (over 30.0 ms/cm) 0.01 ppm (0.00-14.99 ppm) 0.1 ppm (15.0-149.9 ppm) 1 ppm (150-1499 ppm) 0.01 g/l (1.50-14.99 g/l) 0.1 g/l (over 15.0 g/l) 0.1% 0.1 C ± 1% of reading (± 0.05 or 1 digit, whichever is greater) ± 1% of reading (± 0.03 ppm or 1 digit, whichever is greater) ± 1% of reading ± 0.4 C ± 1% of reading ± 1% of reading ± 1% of reading ± 0.1 C Automatic 1 point with 6 memorized solutions (84, 1413, 5000, 12880, 80000 or 111800 ) One point with 100%NaCl sea water standard solution Temp. Calibration Two points at 0 and 50 C Temp. Compensation Automatic or manual from 0 to 60 C (can be disabled to measure actual conductivity) Temp. Coefficient 0.00 to 6.00%/ C (conductivity and TDS only); 1.90%/ C default TDS Factor 0.40 to 0.80 (0.50 default) Power Supply Four 1.5 volt AA alkaline batteries/150 hours continuous use (included), or 12 VDC adapter Casing IP 67 Environment 0 to 50 C/100% RH Dimensions 7.7 x 3.1 x 2.41 inches (196 x 80 x 60 mm) Weight 18 oz. (500 g) Rev. 0-401 3 Model GLI-9835 Portable Conductivity/TDS Meter

CONNECTIONS NOTES: I If the meter displays "----", the reading is out of range. Remove the back cover, unwrap the batteries and install them while paying attention to the polarity. Replace the back cover. Alternatively, connect the voltage adapter to the power socket. Connect the conductivity/tds probe to the 7-pin connector located on the top of the instrument. Tighten the threaded ring. Make sure the probe sleeve is properly inserted (see arrow above). TAKING MEASUREMENTS 1. Press the ON/OFF key to turn the meter on. 2. Immerse the probe into the solution to be tested. The sleeve holes must be completely submerged. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve. 3. If needed, press the RANGE key repeatedly until the desired range (conductivity, TDS or NaCl) is selected on the LCD. 4. Allow for the reading to stabilize. The upper LCD displays the measure in the selected range while the temperature is displayed on the lower LCD. II III IV V VI VII VIII If the reading is unstable, the stability indicator "~" blinks. The "gm" indication on the LCD means g/l. Make sure the meter is calibrated before taking measurements. If measurements are taken successively in different samples, to have accurate readings it is recommended to rinse the probe thoroughly with deionized water before immersion in the samples. To maximize battery life, the meter is automatically switched off after 5 minutes of non-use. To reactivate the meter, press the ON/OFF key. This feature can be disabled by entering the setup mode and selecting the "AoF" item (see SETUP for details). TDS reading is obtained multiplying the conductivity reading by the TDS factor, which has a default value of 0.50. It is possible to change the TDS factor in the 0.40 to 0.80 range by entering the setup mode and selecting the "tds" item (see SETUP for details). When the use of an alternate function (FNC, CFM, GLP, FIXED, TC or CALT) is needed, press and hold the ALT key first and then the second key. Model GLI-9835 Portable Conductivity/TDS Meter 4 Rev. 0-401

AUTO-RANGING The conductivity and TDS scales are autoranging. The meter automatically sets the scale with the highest possible resolution. By pressing ALT+FIXED, the auto-ranging feature is disabled and the current range is frozen on the LCD. The "F1" symbol blinks on the LCD. To restore the auto-ranging option, press ALT+FIXED again. NOTE: Auto-ranging is automatically restored if the RANGE key is pressed, if the setup or calibration modes are entered, or if the meter is turned off and back on again. To select the desired option, press the ATC key until the option is briefly displayed on the LCD. NOTE: The default compensation mode is ATC. If temperature compensation is active, measurements are compensated using a default temperature coefficient of 1.90%/ C. It is possible to select a different temperature coefficient (TC) in the 0.00 to 6.00%/ C range by entering the setup mode and selecting the "tc" item (see SETUP for details). The current temperature coefficient can be quickly viewed pressing ALT+TC. The value is briefly displayed on the lower LCD. TEMPERATURE COMPENSATION Three options of compensating temperature are available: Automatic (Atc): The probe has a built-in temperature sensor; the value of the temperature is used to automatically compensate the conductivity/tds reading using 25 C as a reference temperature. This is the default option. Manual (Mtc): The temperature value, shown on the lower LCD, can be manually set by the user with the up and down arrow keys. The compensation is referenced at 25 C. The " C" symbol blinks when this option is active. No Compensation (notc): The temperature reading shown on the lower LCD is not taken into account. The reading displayed on the upper LCD is the actual conductivity or TDS value. The " C" and "%TC" symbols blink when this option is active. CONDUCTIVITY CALIBRATION Calibration is a 1-point procedure. Selectable calibration points are: 0.0 12.88 ms/cm 84.0 80.0 ms/cm 1413 111.8 ms/cm 5.00 ms/cm 1. To enter conductivity calibration, select the conductivity range and press the CAL key. NOTE: TDS reading is automatically derived from the conductivity reading and no specific calibration for TDS is needed. Pressing CAL while TDS range is selected has no effect. 2. Rinse the probe with some of the calibration solution or deionized water. Immerse the probe into the solution. The Rev. 0-401 5 Model GLI-9835 Portable Conductivity/TDS Meter

sleeve holes must be completely submerged. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve. For zero calibration, just leave the dry probe in air. The indications "BUF" and "CAL" are displayed. The upper LCD shows the uncalibrated conductivity reading. The lower LCD shows the buffer value. The stability indicator "~" blinks. 3. Select the desired value with the and keys, if necessary. 4. When the "~" symbol stops blinking, the reading is stable. The "CON" indication starts blinking on the LCD asking for confirmation. 5. Press ALT + CFM to confirm the calibration. If everything is satisfactory, the meter displays the "Stor Good" message and returns to measurement mode. NOTES: I If the uncalibrated reading is too far from the expected value, calibration is not recognized. The "CON" indication does not appear; the "~" and "BUF" symbols blink to signal wrong or contaminated calibration solution. II For best results, choose a conductivity standard solution value close to the sample to be measured. III IV V In order to minimize any EMC interference, use plastic beakers. The meter uses a 1.90%/ C compensation factor during calibration. If the setup item "tc" has been set to a different value, when exiting the calibration mode the value displayed on the upper LCD could be different from the nominal standard solution value. It is possible to set the cell constant value directly without following the calibration procedure. To set the cell constant enter the setup mode and select "CEL" (see SETUP for details). NaCl CALIBRATION Calibration is 1-point at 100.0% NaCl. Use a 100% NaCl standard solution. 1. To enter NaCl calibration, select the NaCl range and press the CAL key. 2. Rinse the probe with some of the calibration solution or deionized water. Immerse the probe into the 100% NaCl standard solution. The sleeve holes must be completely submerged. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve. 3. The indications "BUF" and "CAL" are displayed. The upper LCD shows the uncalibrated NaCl reading in percentage. The lower LCD shows "100." Model GLI-9835 Portable Conductivity/TDS Meter 6 Rev. 0-401

4. When the "~" symbol stops blinking, the reading is stable. The "CON" indication starts blinking on the LCD asking for confirmation. 5. Press ALT + CFM to confirm the calibration. If everything is satisfactory, the meter displays the "Stor Good" message and returns to measurement mode. NOTES: I If the uncalibrated reading is too far from the expected value, the calibration is not recognized. The "CON" indication does not appear; the "~" and "BUF" symbols blink to signal wrong or contaminated calibration solution. II The meter uses a 1.90%/ C compensation factor during calibration. If the setup item "tc" has been set to a different value, when exiting the calibration mode the value displayed on the upper LCD could be different from the nominal calibration value. TEMPERATURE CALIBRATION (for technical personnel only) The calibration is 2 points at 0.0 C and 50.0 C: 3. The lower LCD displays "0.0 C"; the "BUF" and "CAL" symbols appear. 4. When the reading is stable, the "CON" symbol starts to blink. 5. Press ALT+CFM to confirm. The lower LCD displays 50.0 C. 6. Immerse the probe in a 50 C temperature bath. 7. When the reading is stable, the "CON" symbol starts to blink. 8. Press ALT+CFM to confirm and return to normal operation. TEMPERATURE ADJUSTMENT The temperature reading can be manually fine-tuned by following this procedure: 1. Press ALT+CALT to enter the temperature calibration mode. 2. Press CAL to enter the temperature adjustment mode. The upper and lower LCD will display the current temperature reading. 3. Adjust the temperature reading on the upper LCD using the arrow keys. The maximum adjustment is ±1 C around current reading. 1. Immerse the probe in a 0 C temperature bath. 2. Press ALT+CALT to enter temperature calibration mode. Rev. 0-401 7 Model GLI-9835 Portable Conductivity/TDS Meter

4. Press ALT+CFM to confirm. The meter returns to measurement mode and displays the new temperature. C F 27 28 29 30 31 80.6 82.4 84.2 86 87.8 84 87 89 90 92 94 1413 1467 1494 1521 1548 1575 5000 5190 5286 5383 5479 5575 12880 13370 13620 13870 14120 14370 80000 111880 83000 115700 84900 117700 86300 119700 88200 121800 90000 123900 NOTES: I Press ALT+CALT to escape without any changes. SETUP II It is possible to enter the temperature adjustment mode only if the probe is connected. CONDUCTIVITY VERSUS TEMPERATURE CHART Setup is used to view or change the instrument parameters. 1. To enter setup, press ALT+FNC when the meter is in measurement mode. The conductivity of an aqueous solution is the measure of its ability to carry an electrical current by means of ionic motion. The conductivity invariably increases with increasing temperature. It is affected by the type and number of ions in the solution and by the viscosity of the solution itself. Both parameters are temperature dependent. The dependency of conductivity on temperature is expressed as a relative change per degree Celsius at a particular temperature, commonly as percent per C. The following table lists the temperature dependence of the conductivity standard solutions used for calibration: 2. "Set" is displayed on the upper LCD. The lower LCD displays the blinking code of the current setup item. 3. Select the desired setup item using the or key. 4. Press ALT+CFM to confirm. C F 0 5 10 15 16 17 18 19 20 21 22 23 24 25 26 32 41 50 59 60.8 62.6 64.4 66.2 68 69.8 71.6 73.4 75.2 77 78.8 84 64 65 67 68 70 71 73 74 76 78 79 81 82 84 86 1413 776 896 1020 1147 1173 1199 1225 1251 1278 1305 1332 1359 1386 1413 1440 5000 2760 3180 3615 4063 4155 4245 4337 4429 4523 4617 4711 4805 4902 5000 5096 12880 7150 8220 9330 10480 10720 10950 11190 11430 11670 11910 12150 12390 12640 12880 13130 80000 111880 48300 65400 53500 74100 59600 83200 65400 92500 67200 94400 68500 96300 69800 98200 71300 100200 72400 102100 74000 104000 75200 105900 76500 107900 78300 109800 80000 111800 81300 113800 NOTE: If ALT+FNC are pressed before item confirmation, the meter will escape and return to measurement mode. 5. Once the desired setup item has been selected, its current value blinks (if it is a changeable parameter). Model GLI-9835 Portable Conductivity/TDS Meter 8 Rev. 0-401

6. To change the value, use the or key. 7. Press ALT+CFM to confirm. NOTE: Press ALT+FNC before confirmation to escape without changing the previously set value. The following table lists the setup items, their valid range of values, and the factory default settings: Item Description Valid values Default tc Temp. comp. coeff. 0.00-6.00%/ C 1.90 tce Temp. comp. mode Atc, Mtc, notc Atc tds TDS factor 0.40 to 0.80 0.50 CEL Cell constant (K) 0.500 to 1.700 1.000 Aof Auto-Off enable 1 ON, OFF ON ver Firmware release Chr Battery level test 2 1 Once enabled, the Auto-Off time is fixed at 5 minutes. 2 When the battery level test is selected (Chr), the LCD will display the remaining percentage of battery charge. 100% means fully charged battery and 0% corresponds to the minimum battery level that allows the meter to operate. enable it until new batteries are inserted or an external power adapter is used. When "LOBAT" appears, batteries have typically 10% of their life left and the meter is still measuring properly. Very Weak Batteries -- The meter shuts off to avoid erroneous operations. NOTE: If the meter is not powered for several minutes (e.g. in dead battery condition), the current date and time are lost. Battery replacement must only take place in a non-hazardous area using 1.5V alkaline AA type batteries. To replace run-down batteries, simply remove the two screws on the rear cover of the instrument and replace the four 1.5V AA batteries with new ones, observing correct polarity. New batteries allow approximately 150 hours of continuous use (with 2700 ma/h batteries and without backlight). A 12 VDC power adapter can also be used to power the unit. NOTE: The instrument uses the following configuration. The battery charge level calculation is based on a typical alkaline battery discharge curve. If the meter is connected to an external power adapter and "Chr" is selected, the LCD will display "LINE." BATTERY REPLACEMENT When the batteries are inserted and no power adapter is connected, the meter can recognize different battery charge levels. It is recommendable to purchase the GLI voltage adapters that use the proper polarity configuration. However, the meter can be used with other adapters. In this case, remember to check the correct polarity of your adapter before connecting it to the meter. Low Batteries -- "LOBAT" indication is displayed on the LCD. Backlight is automatically disabled and it is not possible to Rev. 0-401 9 Model GLI-9835 Portable Conductivity/TDS Meter

PROBE MAINTENANCE Rinse the probe with clean water after measurements. If a more thorough cleaning is required, remove the probe sleeve and clean the probe with a cloth or a nonabrasive detergent. Make sure to properly reinsert the sleeve in the right direction onto the probe. After cleaning the probe, always recalibrate the instrument. WARRANTY GLI International, Inc. warrants the Model GLI-9835 Hand-held/Portable Conductivity/ TDS Meter to be free from defects in material or workmanship for a period of one year (12 months) from the date of shipment of this product from our facility. The GLI-76302W Conductivity Sensor is warranted for a period of 6 months. A warranty claim will not be honored if defects are not reported within the warranty period, or if GLI International determines that defects or damages are due to normal wear, misapplication, lack of maintenance, abuse, improper installation, alteration or abnormal conditions. GLI International s obligation under this warranty shall be limited to, at its option, replacement or repair of this product. The product must be returned to GLI International, freight prepaid, for examination. The product must be thoroughly cleaned and any process chemicals removed before it will be accepted for replacement or repair. GLI International s liability shall not exceed the cost of the product. Under no circumstances will GLI International be liable for any incidental or consequential damages, whether to person or property. GLI International will not be liable for any other loss, damage or expense of any kind, including loss of profits, resulting from the installation, use, or inability to use this product. Model GLI-9835 Portable Conductivity/TDS Meter 10 Rev. 0-401