MONITOR/DEFIBRILLATOR CONTROLS, CONNECTIONS AND INDICATORS

Similar documents
MONITOR/DEFIBRILLATOR CONTROLS, CONNECTIONS AND INDICATORS

Vital Signs Trending. with the HeartStart MRx and XL+ PUBLISH. Application Note. Introduction

Product Specifications

Philips HeartStart MRx Owners Manual

Philips HeartStart XL Defibrillator Manual

R Series Configuration Manual

Hardware and Components: Philips Monitors in Critical Care

Cardiostart. Biphasic Defibrillator Monitor

URGENT Field Safety Notice Philips HeartStart MRx Monitor/Defibrillator

LEGAL NOTICE INVITATION TO SUBMIT BIDS

DEFRIBILLATORS PARAMEDIC CU-ER1

the next generation life energy

STEP 2: VITALSBRIDGE TO PATIENT MONITOR CONNECTION GUIDE VITALSBRIDGE AND 100

SpO 2 Upgrade Instructions

Defibrillation made simple

STEP 6: USING THE VITALSBRIDGE WITH A STANDARDIZED PATIENT VITALSBRIDGE AND 100

Resusci Anne Advanced SkillTrainer

etco 2 SpO 2 IBP medlab 2017/2018 OEM Modules for Patient Monitoring medizinische Diagnosegeräte GmbH

ACCESSORY GUIDE REV D

ALS life support on the run. Philips M3536A HeartStart MRx Monitor/Defi brillator

Defibtech DDU-100 Semi-Automatic External Defibrillator

VitalsBridge. Directions for Use MODEL VB2.0



HeartStart XL+ Service Library. Service Manual. HeartStart XL Edition 3

Argentine Manufacturer of Hospital Equipment. Electrocardiograph Pulse Oximeters Vital Signs Monitors

Patient Simulator Series

ACCESSORIES BROCHURE LIFEPAK 15 MONITOR/DEFIBRILLATOR. Genuine Accessories from Physio-Control

Resusci Anne Simulator

BeneVision DMS Workstation Default Settings

PW 406. Features &Benefits: Multi parameters Patient monitor

ARGUS PRO LifeCare 2. Patient monitor, 12-channel ECG, defibrillator and pacemaker in a compact format F A C T U R M A N U A N D I N E D E U R O I G N

M3/M4 Monitors (M3046A)

X SeriesTM. Xtremely. Small, Light, and Powerful

2012/2/3. imec12/10/8

Version 4 User Guide

Connex Spot Monitor (CSM) 1.x Software Version

*M C* Philips A1 and A3 Patient Monitors M3926A M3923A M3924A M3927A M3928A M3929A. Service Guide. Part Number M C

Propaq LT Monitors. Propaq LT Base Model. Propaq LT with Wireless Option/ Legendary quality and durability Heart rate, 3-or 5-lead ECG, SpO 2

INTRODUCTION The Monitor, interchangeable modules and storage batteries are described below. Refer to the table below for specifications.

M Series AHA/ERC Guidelines 2005 Software Configuration Overview

TRANQUILITY VS VERSATILITY IN VITAL SIGNS A M E R I C A S TRANQUILITY VS A M E R I C A S

LIFEPAK 15 Test Calibration Procedure ( TCP)

ACENDIS EMERGENCY ACENDIS APM-700 / 700 XL MODULAR PATIENT MONITOR

Defibtech DDU-2300 Semi-Automatic External Defibrillator

Cardiac Monitoring Catalog USA

IntelliVue Patient Monitor. MP20/30/40/50/60/70/90 For monitor release B.1 with software revision B.10.xx. IntelliVue MP20/30/40/50/60/70/90

Test equipment, adapters and facilities

BeneVision DMS Server Default Settings

Infinity Delta and Delta XL Patient Monitors

SERVICE MANUAL. Defibrillator TEC-5601/TEC-5602 TEC-5611/TEC-5621 TEC TEC-5600 series B

NO COMPROMISE IN AN EMERGENCY CASE

Faros Explorer Manual

Defibtech DDU-100 Semi-Automatic External Defibrillator User Manual

DPM 4 Patient Monitor. Service Manual

Original Manufacturer

Infinity Delta and Delta XL Infinity Bedside Solutions

StressLocator & Sleep as Android

User Manual. MAXISIM 2000 Patient Simulator

corpuls 3 Any Emergency under control

This document contains important information for the continued safe and proper use of your equipment

Propaq Vital Signs Monitors

Defibtech Lifeline AED Fully Automatic Manual

Defibtech DDU-100 Series Automated External Defibrillator DDU-100 DDU-120

Simulator Maintenance: SimMan3G & SimMan Essential

Defibtech DDU-2400 and DDU-2450 Automated External Defibrillator

i 4 imm Compact Critical Care Monitor GE Healthcare Modular monitoring for critical care

Defibtech Automated External Defibrillator Lifeline/ReviveR DDU-100 Lifeline/ReviveR AUTO DDU-120

SmartMonitor 2 Self Study Program

Defibtech DDU-2000 Series Automated External Defibrillator DDU-2300 DDU-2400 DDU-2450

QUICK REFERENCE GUIDE

Defibtech DDU-2300 Semi-Automatic External Defibrillator

Nursing Solutions with SimPad.

Professional Health Solutions

Service Manual. M4735A HeartStart XL Defibrillator/Monitor

Using the Universal Pendant with Patient Cable

ECG100S. General Information. Product Code

WVSM wireless vital signs monitor

Patient Monitoring Systems. Patient Monitor Vital Sign Monitor Pulse Oximeter

Draft. IntelliVue Cableless Measurements. IntelliVue CL SpO2 Pod and CL NBP Pod Release A.0. Patient Monitoring. Instructions for Use

All Products of Mediana

INSTRUCTION MANUAL. Automatic Blood Pressure Monitor with Arm Cuff. Model HEM-711AC

CU-SP1 INTELLIGENT PUBLIC ACCESS DEFIBRILLATOR. Medical and Surgical Requisites Pty Ltd

FOCUS - ON - PATIENTS. corpuls aed

ipm-9800 Patient Monitor Service Manual

SimPad SkillReporter. User Guide.

Operating Instructions. Panorama. Web Viewer

FREQUENTLY ASKED QUESTIONS

Student Quick Reference Guide

STANDARD OPERATING PROCEDURE TENSOVAL COMFORT

User s Manual. Model May 1, Catalog No

EC-12RT Resting ECG Machine. Technical Specification

OPERATING INSTRUCTIONS. LIFEPAK 12 3D BIPHASIC defibrillator/monitor

Statewide Contract Addendum

This document contains important information for the continued safe and proper use of your equipment

BeneVision Default Settings

ipm 12/iPM 10/iPM 8 ipm 7/iPM 6/iPM 5 Patient Monitor Service Manual

Ansur ProSim 6/8. Users Manual. Software Plug-In

Addendum No. 2 Reference: ITB Procurement of Medical Equipment for the Intensive Care Unit at Alia Hospital Hebron

Bluetooth/RS-232 Enhanced Data Communication

Transcription:

MONITOR/DEFIBRILLATOR CONTROLS, CONNECTIONS AND INDICATORS

Adult Dose 100 120 70 170 50 200 30 20 15 1-10 CONTROLS, CONNECTIONS AND INDICATORS External Power SYNC Mark Event Lead Select Alarm Pause Event Summary Soft Keys Soft Keys Navigation Sync Manual Defib Pacer M onitor 150 On Off On AED Select Energy 1 Charge Shock Menu Select 2 3 Ready For Use (RFU) Therapy Knob Charge Data Card Shock Printer Printer Door Latch Print Speaker Therapy Connector CO 2 Inlet CO 2 Outlet Temp ECG Out (Sync) Jack CO 2 2 1 ECG ECG Invasive Pressure 1 Invasive Pressure 2 NBP ECG SpO 2 Battery/AC Compartment B LAN Connection RS 232 Serial Port AC Power Module DC Power Input Bed Rail Hook Mount Battery Compartment A Battery 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR 12-LEAD ECG

12-LEAD ECG Electrode Placement for 12-Lead ECG Acquiring the 12-Lead ECG 1. Turn the Therapy Knob to Monitor. 2. Press the 12-Lead soft key. 3. Check for good signal quality. 4. Press the Start Acquire soft key. If prompted, enter patient age and sex. 5. Keep the patient still during the 10-second acquisition period. Accessing Stored 12-Lead Reports 1. Press the Menu Select button. 2. Use the Navigation buttons to select Reports. Reports for the current patient are listed by date and time. 3. Use the Navigation buttons to select a report and press the Menu Select button. 4. Select Print, Copy, or Delete, and press the Menu Select button. 5. To select another report, repeat steps 1 through 4. 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR ALARMS

USING ALARMS Responding to Alarms 1. Check the patient. 2. Identify the alarm(s) indicated. 3. Silence the alarm(s) by pressing the Menu Select button or one of the Navigation buttons. 4. Address the alarm condition by selecting one of the following options: Alarms Off - Turn the monitoring parameter's alarms off, displaying next to the parameter value. Acknowledge - Recognize that the alarm condition is present, without further action. As the default option, Acknowledge always appears when an alarm occurs. New Limits - Adjust parameter alarm limits accordingly. Note: For only Heart Rate/Arrhythmia alarms, your options are either Acknowledge or Acknowledge and New Limits, depending on the alarm. Checking/Adjusting Alarm Limits Alarm limits are preset for your device based on its configuration and the patient type. When alarms are on, alarm limits are visible next to the measurement's numeric value. To adjust an alarm limit setting: 1. Press the Menu Select button. 2. Select Measurements/Alarms and press the Menu Select button. 3. Select the desired measurement from the menu and press the Menu Select button. 4. Select (measurement) Limits from the menu and press the Menu Select button. 5. Adjust the limit and press the Menu Select button. 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR MONITORING CARBON DIOXIDE

MONITORING CO 2 Using the Nasal FilterLine 1. Attach the FilterLine tubing to the CO 2 Inlet port. 2. Check that both nostrils are clear. 3. Position the nasal FilterLine on the face by inserting the tips into the nostrils. 4. Pass the FilterLine tubing over the ears, then slide the sleeve up the tubing toward the neck to a comfortable fit under the chin. 5. If using dual purpose FilterLine tubing, connect the green tubing to the oxygen source. 6. Check the positioning of the FilterLine regularly to ensure proper function. Using the FilterLine and Airway Adapter 1. Attach the FilterLine tubing to the CO 2 Inlet port. 2. Connect the wide end of the airway adapter to the endotracheal tube. 3. Connect the narrow end of the airway adapter to the ventilator tubing or manual resuscitator. Measuring EtCO 2 and AwRR The CO 2 waveform is automatically displayed in the configured Wave Sector when the FilterLine is connected to the CO 2 Inlet port. The measurement values for End-Tidal CO 2 (EtCO 2 ) and Airway Respiration Rate (AwRR) are displayed in Parameter Block 2 as shown below. SpO2 100 100 90 EtCO2 AwRR 34 50 30 18 30 8 % mmhg rpm 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR MONITORING ECG

MONITORING ECG Selecting the Lead for ECG Monitoring It is important to select a suitable lead for monitoring so that a QRS complex can be accurately detected. Guidelines: - QRS tall and narrow - R-wave above or below baseline (but not bi-phasic) - T-wave smaller than 1/3 R-wave height - P-wave smaller than 1/5 R-wave height - For non-transcutaneously paced patients, the pace pulse should be large enough to be detected with minimal repolarization and not wider than the normal QRS complex. To select a lead for the primary wave sector: 1. Press the Lead Select button. 2. Cycle through lead choices. To select a lead for an additional wave sector: 1. Press the Menu Select button. 2. Select Waves and press the Menu Select button. 3. Select the wave sector and press the Menu Select button. 4. Select the desired lead source and press the Menu Select button. 5. If needed, adjust the ECG wave size (gain) and press the Menu Select button. Note: Arrhythmia analysis occurs at x1 gain (no matter what the display gain is). Changing Pacing Status When monitoring patients with non-transcutaneous pacemakers, it is important to set the paced status correctly. (See MRx Instructions for Use for additional information.) To change pacing status: 1. Press the Menu Select button. 2. Select Patient Info and press the Menu Select button. 3. Select Paced and press the Menu Select button. 4. Select Yes/No and press the Menu Select button. II I + III + + 3-Lead Placement avr II I avl avf + + + 5-Lead Placement III 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR MONITORING INVASIVE PRESSURES

MONITORING INVASIVE PRESSURES Changing a Pressure 1. Press the Menu Select button. 2. Using the Navigation buttons, select the Measurement/Alarms menu and press the Menu Select button. 3. Select Press 1 (or Press 2) and press the Menu Select button. 4. Select Label and press the Menu Select button. 5. Select the appropriate label from the list and press the Menu Select button. 6. Select the size of the scale, the Alarm Source type and high/low Alarm Limits. Pressure Label Options ABP - Arterial Blood Pressure ART - Arterial Blood Pressure Ao - Aortic Pressure CVP - Central Venous Pressure ICP - Intracranial Pressure* LAP - Left Atrial Pressure PAP - Pulmonary Artery Pressure RAP - Right Atrial Pressure P1 - Non-specific Pressure P2 - Non-specific Pressure Label (Channel 1) Label (Channel 2) * Cerebral Perfusion Pressure (CPP) is automatically displayed with ICP when one pressure is already set to ABP, ART or Ao and the other pressure is set to ICP. Zeroing 1. Place the transducer at the appropriate level for the measurement site. 2. Close the transducer stopcock to the patient and vent the transducer to atmospheric pressure. Then pick one of the following methods: Using Menu Select button Using a soft key in Monitor Mode 3. Press the Menu Select 3. Press the button under [Zero button. Pressure] on the display. 4. Using the Navigation button, 4. Using the Navigation buttons, select Measurement/Alarms, select the pressure(s) to zero press the Menu select button 5. Select the pressure label to the Menu Select button. zero. 6-Highlight Zero and press the Menu Select button. from the Zero menu and press 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR MONITORING NONINVASIVE BLOOD PRESSURE

MONITORING NBP Preparing 1. Select the appropriately sized cuff. 2. Attach the cuff to the NBP tubing. 3. Insert the NBP tubing into the NBP port. 4. Apply the blood pressure cuff to the patient's arm or leg. a. Ensure the cuff is completely deflated. b. Wrap the cuff around the limb. c. Check that the edge of the cuff falls within the size range markings. 5. Place the limb at the same level as the patient's heart. Changing the NBP Schedule 1. Press the Menu Select button. 2. Using the Navigation buttons, select the Measurement/Alarms menu and press the Menu Select button. 3. Select NBP and press the Menu Select button. 4. Select NBP Schedule and press the Menu Select button. 5. Select the desired interval or Manual, and press the Menu Select button. Interval Options: 1, 2.5, 5, 10, 15, 30, 60 and 120 minutes. A Manual NBP measurement may be taken at any time by pressing the Start NBP soft key. Measuring NBP To measure NBP, press the Start NBP soft key. Cuff pressure is displayed as the cuff inflates and deflates. When the measurement completes, the NBP values are displayed. HR 82 120 120 60 NBP mmhg Manual 120/80 (100) SYS 10:40 160 90 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR OPERATIONAL CHECK

OPERATIONAL CHECK Running the Operational Check 1. Insert a charged battery into the HeartStart MRx. 2. Turn the Therapy Knob to 150J. 3. Press the Menu Select button. 4. Using the Navigation buttons, select Other and press the Menu Select button. 5. Select Operational Check and press the Menu Select button. 6. Select Run Op Check and press the Menu Select button. 7. Press the Menu Select button to acknowledge the message Leaving Normal Operating Mode. 8. Carefully read and respond to the prompts. At the completion of the Operational Check, a report is printed that lists the test results and prompts you to visually inspect the device and cables, and to inventory all accessories and supplies. Note: For device versions greater than or equal to B.05 with Pacing, run Op Check with a pads cable and test external paddles using the Weekly Shock Test. 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR NONINVASIVE PACING (TRANSCUTANEOUS PACING)

NONINVASIVE PACING (TRANSCUTANEOUS PACING) Demand Mode Pacing (Requires ECG leads and pads) 1. Turn the Therapy Knob to Pacer. 2. Press the Lead Select button to select the desired lead. 3. Verify white R-wave markers appear above or on the ECG waveform. 4. Press the Pacer Rate soft key, select the desired number of paced pulses per minute, and press the Menu Select button. 5. If needed, press the Pacer Output soft key, use the Navigation buttons to adjust the initial pacer output, and press the Menu Select button. 6. Press the Start Pacing soft key. 7. Verify white pacing markers appear in front of every QRS. 8. If cardiac capture is not obtained, press the Pacer Output soft key, increase the output until capture occurs, decrease the output to the lowest level that still maintains capture, and press the Menu Select button. 9. Verify the presence of a peripheral pulse. II Paced Rhythm Fixed Mode Pacing (Can be done with pads only) 1. Turn the Therapy Knob to Pacer. 2. Set the pacer mode to Fixed. - Select Pacer Mode from the Main Menu. - Select Fixed and press the Menu Select button. 3. If using leads, press the Lead Select button to select the desired lead. 4. Press the Pacer Rate soft key, select the desired number of paced pulses per minute, and press the Menu Select button. 5. If needed, press the Pacer Output soft key, use the Navigation buttons to adjust the initial pacer output, and press the Menu Select button. 6. Press the Start Pacing soft key. 7. Verify cardiac capture. If it is not obtained, press the Pacer Output soft key, increase the output until capture occurs, decrease the output to the lowest level that still maintains capture, and press the Menu Select button. 8. Verify the presence of a peripheral pulse. NOTE: If monitoring SpO 2 when pacing, activate the Pulse alarms to assess for peripheral perfusion through the SpO 2 transducer. Demand Mode Capture 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR Q-CPR TM

Q-CPR Prior to Use 0-8 1. Verify the Pads/CPR cable is connected to the MRx Therapy port. 2. Verify the Compression Sensor is connected to the Pads/CPR cable. 3. Verify a new Compression Sensor Adhesive Pad is applied to the Compression Sensor. Preparing for Q-CPR 1. Connect the multifunction electrode pads to the Pads/CPR cable. 2. Apply the multifunction pads to the patient in the anterior/anterior position. 3. Position the Compression Sensor on the patient as shown on the green liner (i.e., on the lower half of the sternum). Q-CPR in Manual Defib Mode - Q-CPR starts automatically when you turn the Therapy Knob to 150J or perform a chest compression with the Compression Sensor. - To manually activate Q-CPR, press the Start CPR soft key. Q-CPR in AED Mode - Q-CPR starts automatically when you perform a chest compression with the Compression Sensor during the CPR Pause period. If necessary, press the Pause for CPR soft key to activate the Pause period. - AED Mode provides only corrective text and voice prompt feedback. Q-CPR is a trademark of Laerdal Medical. 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR READY FOR USE INDICATOR

READY FOR USE INDICATOR The Ready For Use (RFU) indicator is located on the upper, right corner of the device. It indicates the status of the therapy delivery functions of the monitor/ defibrillator using the following definitions: A blinking black hourglass symbol indicates the shock, pacing, and ECG functions of the device are ready for use. Sufficient battery power is available for device operation. A blinking red X and a periodic audio chirp indicate no battery is present or a low battery condition. If a battery is inserted and charging, the audio chirp is not present. A solid red X and a periodic audio chirp indicate a failure has been detected that may prevent the delivery of a shock, pacing, or ECG acquisition. An error message will display at power-on. A solid red X without periodic audio chirps indicates either there is no power available, or a device failure has occurred. If, after power is supplied, the indicator reverts to the blinking black hourglass symbol, the device is once again ready for use. 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR MONITORING SpO2

MONITORING SpO 2 Applying the Sensor Follow the manufacturer's directions for applying and using the sensor, making sure to observe any warnings and cautions. For best results: Make sure the sensor is dry. If the patient is moving, secure the sensor cable loosely to the patient. Make sure the transducer is not too tight. Keep power cables away from the sensor cable and connections. Avoid placing the sensor in an environment with bright lights. If necessary, cover the sensor with opaque material. Avoid placing the sensor on an extremity with poor perfusion. Monitoring SpO 2 1. Connect the appropriate sensor cable to the HeartStart MRx. 2. Apply the sensor to the patient. 3. If the HeartStart MRx is not turned on, turn the Therapy Knob to Monitor. The pleth wave appears after a few seconds, followed by the oxygen saturation value. As the patient's oxygen saturation changes, the SpO 2 value is continuously updated and the associated waveform is displayed. See the Alarms Quick Card for information on adjusting alarms. 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR MONITORING TEMPERATURE

MONITORING TEMPERATURE To Monitor a Temperature 1. Connect the temperature cable to the HeartStart MRx. 2. Select the correct temperature label for your measurement. 3. Check that the current device settings (including alarm settings) are appropriate for the patient. 4. Apply the temperature probe to the patient. Selecting a Temperature Label 1. Press the Menu Select button. 2. Using the Navigation buttons, select the Measurement/Alarms menu and press the Menu Select button. 3. Select the temperature label option currently assigned to your measurement (default is Temp) and press the Menu Select button. 4. Select Label and press the Menu Select button. 5. Select the appropriate label from the list provided and press the Menu Select button. Temperature Label Options Tesoph - esophageal temperature Trect - rectal temperature Tskin - skin temperature Temp - non-specific temperature label Tcore - core temperature Tnaso - nasopharyngeal temperature Tvesic - urinary bladder Tart - arterial temperature Tven - venous temperature 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR VITAL SIGNS TRENDING

VITAL SIGNS TRENDING Viewing Trending Data 1. Place the HeartStart MRx into Monitor Mode. 2. Press the Menu Select button. 3. Using the Navigation buttons, select Trends and press the Menu Select button. The Vital Signs Trending Report appears in the bottom two wave sectors. Scrolling LEFT/RIGHT: Use the << or >> soft keys to scroll left and right (backward/ forward) in the report screen. The soft key will be inactive if there is no more data to be viewed in that direction. UP/DOWN: If there are more vital signs than can be shown on the screen, use the Navigation buttons next to the Menu Select button to scroll up and down with the vertical scroll bar on the display. (Make sure there are no menu items active at the time.) Reading Trending Data - When trending is initially displayed, the latest (most recent) trending data will appear in the far right column; older data flows to the left. - The display will auto-update as new data becomes available as long as the latest data is displayed on the screen. - A -?- indicates that the parameter has invalid information. - A? just before a numeric value indicates questionable data. - An empty space indicates unavailable data. Printing the Vital Signs Trending Report You can print the Vital Signs Trending Report two ways: 1. Press the soft key under the Print Trends label. A report for the displayed period and interval is printed. 2. Press the Summary button and select Trends and then Trends Interval. A report for the entire incident period is printed. 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR

MONITOR/DEFIBRILLATOR WEEKLY SHOCK TEST

WEEKLY SHOCK TEST How to perform the Weekly Shock Test Test each type of patient therapy cable used (multifunction electrode pads or paddles). If you are using paddles, make sure the paddles are secure in their pockets and that the Patient Contact Indicator (PCI) LEDs located on the sternum paddles are not lit. If the LEDs light, adjust the paddles in their pockets. If the LEDs continue to light, clean both adult and pediatric paddle electrode surfaces. If you are using multifunction electrode pads, attach a test load to the end of the patient Therapy Cable. 1. Turn the Therapy Knob to 150J 2. Press the Charge Button. NOTE: If it becomes necessary to disarm the defibrillator, press [DISARM] 3. If using: - Pads, press the Shock button on the HeartStart MRx to deliver a shock into the test load. - External paddles, simultaneously press the shock buttons located on the paddles to deliver a shock into the test load. 4. The strip prints immediately if configured to do so. If the strip does not print immediately, press the Print button. 5. Confirm on the printed strip that the energy delivered to the test load is 150J +/- 23J (127J to 173J). If not, take the device out of use and call for service. 453564042471 Edition 1 September 2006 *453564042471* *1* MONITOR/DEFIBRILLATOR