MULTI meter MU 6100 H

Similar documents
TruLab ph/ise 1320(P)

Operating manual. ph 7310 ph 7310P. Printermodul. ph 7310P. ph ph meter

Operating manual. Oxi 3315 O [Sal] mg/l C. Menu. Digital meter for digital IDS dissolved oxygen sensors

Cond 7310 Cond 7310P

ph/ion meter IS 2000 L

Operating manual. Multi 350i CAL STO ESC. ph / ISE / DO / conductivity measuring instrument. ba75420e04 03/2005

OPERATING MANUAL. ba75863e19 04/2015. Multi 3410 DIGITAL METER FOR DIGITAL IDS SENSORS

Operating manual. Multi Digital ph / D.O. / conductivity meter for 2 digital IDS sensors

OPERATING MANUAL. ba75803e06 06/2013. [Tr25][nLF][ /cm] μs/cm. Menu. Cond 3310 CONDUCTIVITY METER

OPERATING MANUAL. ba77041e02 11/2015. HandyLab 680. HandyLab 680 DIGITAL METER FOR DIGITAL IDS SENSORS

1000 L INSTRUCTION MANUAL

Handheld meter ph 330i/340i

Operating manual M545P M 545P. Pinnacle series. Laboratory ph meter. ba75564e03 02/2007

ProLab 4000 ph/ise/conductivity

Operating manual M530P M 530P. Pinnacle series. Laboratory ph meter. ba75562e02 12/2005

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet

Operating manual. ph/ion/cond 750. Multiparameter laboratory meter for ph, ion concentration and conductivity

Operating manual. VARIO ph. Hand-held ph meter

for measuring ph values and redox potential, conductivity and dissolved oxygen Content

ProLab Operating manual. ph / ISE / conductivity / D.O. meter with automatic sensor recognition and electronic access control

ProLab Operating manual. ph / ISE meter with automatic sensor recognition and electronic access control. DRAFT January 18, :18 pm

Operating Manual. inolab ph Level 2. Laboratory ph meter with integrated printer. ba12219e04 10/

Accuracy when going to press

Thermo Scientific Orion Star A326 Portable ph/rdo/do Meter. Instruction Sheet

Thermo Scientific Orion Star A325 Portable ph/conductivity Meter. Instruction Sheet

ProfiLine Handhelds. For ph, ISE, Dissolved Oxygen and Conductivity

photoflex STD Handheld colorimeter

ProLab 1. ProLab 1000: ph for the professional. ph Meter ProLab 1000: Superior technique in a classy metal body. Precise measurements

HI 9828 Multiparameter Water Quality Meter...

inolab 740 with terminal 740

HI 9828 Multiparameter Water Quality Meter...

Thermo Scientific Orion Star A324 Portable ph/ise Meter. Instruction Sheet

inolab 7000 Series Accurate. Compliant. Sensitive. The inolab family: Advanced measuring technology offers new functions including automatic AutoRead

Thermo Scientific Orion Star A214 Benchtop ph/ise Meter. Instruction Sheet

ProfiLine Single Parameter Instruments Data-logging Communication Ground Water Wastewater Mobile Lab Measurements Surface waters Measurements Communic

Quality Electrochemical Analysis Professional Meters Basic Meters Sensors

Handheld Water Quality Meters 100 Series

User Manual. Thermo Scientific Orion Star A320 Series

Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter

Portable Conductivity Meter M1410. User Manual 1.002

Thermo Scientific Orion Star A223 and Star A323 Portable RDO/DO Meter. Instruction Sheet

Quality Electrochemical Analysis Professional Meters Basic Meters Sensors

Multiparameter Waterproof Meters. Multiparameter measurement in compact, IP67 rated waterproof enclosures

Benchtop Multi-parameter Water Quality Meter

Hanna Instruments HI9819X Multiparameter series

Benchtop and portable meters

Colorado Series. CR-10 Portable ph / mv / Temperature Meter Operation Manual

INSTRUCTION MANUAL. Portable ph and ph/ise Meters. Models 210Aplus 230Aplus 250Aplus 290Aplus

and Multiparameter Meters

Electrochemistry Meters. Series SD (IP 67 waterproof) Electrochemistry Meters

Portavo 904 X Multi. Multiparameter Ex

ProLab Operating manual. ph/ise meter with automatic sensor recognition and electronic access control. DRAFT January 28, :34 pm

Multiparameter Waterproof Meters. Multiparameter measurement in compact, IP67 rated waterproof enclosures

Agilent 3200M Multi-Parameter Analyzer

Dissolved Oxygen Meter Kit DOK. Instruction Manual

Research Grade Meters. ph, ISE, EC and DO. Laboratory bench meters with a color, graphic display and capacitive touch keypad

ProLab Operating manual. ph/ise/conductivity measuring instrument with automatic sensor recognition and electronic access control

MultiLine IDS Portable Meters. PORTABLE MULTIPARAMETER MEASUREMENT OF ph, DISSOLVED OXYGEN, CONDUCTIVITY AND TURBIDITY

User Manual. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500. Additional User Manual Translations available at

Integrated Instruction Manual WP 600 Series Meters

USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements

English. Operating manual. Multiparameter portable instrument HD Keep for future reference. Companies / Brands of GHM

SensoDirect ph200 ph-/redox-/temperature Meter Operating instructions

Technical Information Liquiline To Go CYM290 and Liquiline To Go Ex CYM291

portable waterproof multiparameter meter

ph/orp Meter SD Card real-time Datalogger

Single Parameter Meters Lightweight and versatile ph, EC and DO meters that can be used in portable, wall-mount and benchtop configurations

Instruction Manual HI Multiparameter Meter with GPS

Water Quality Meter. Water Quality Meter Kit. Instruction Manual

Handheld Water Quality Meters 100 Series

Logger WQL-Cond PC program WQL-Log

WK7200. Water Quality Test Kit USER S MANUAL

HI 3220 HI 3221 HI 3222

Instruction Manual HI Multiparameter

Handheld Water Quality Meters 100 Series

703 Laboratory Conductivity Meter

User Manual. Portamess 913 Cond and Portamess 913 X Cond

DO Meter Kit. Instruction Manual

Operating the meter 13. j h g. 4.1 Backside layout Pin assignments RS232 connection

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com DPI 820. Dual Input Thermometer. User manual.

ph 300 ph 301 ph 302 Microprocessor GLP Bench-top ph/mv/ise/ C Meters Instruction Manual

ph and Electrochemistry

Resolution selectable from 0.01 and ph. Range ph. Accuracy ± ph

Instruction Manual HI Multiparameter Meter with GPS

PV Remote Unit A 1378 Instruction manual Version 2.1, Code no

ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:

Introduction: Features:

Submersion/Insertion ph/orp Sensors

Portavo 921 S. Portavo S. The Basic Unit

Operating instructions. Standstill monitor A / / 2011

C100 Handheld Cond./TDS/Sal. Meter. Instruction Manual

WTW Multi 340i/3400i Portable Meters

Operating manual LRD 325. Conductivity measuring cell to be installed in pipelines

Portable ph/orp/conductivity Meter D-74 Instruction Manual CODE:GZ

Instruction Manual HI 2550 Multiparameter ph/orp/ C EC/TDS/NaCl Bench Meter

Field-Lab Analysers for ph, mv, Conductivity, TDS, Salinity, Dissolved Oxygen, Turbidity & Temperature

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Industrial. Laboratory and Field. Accessories. Product Selection Guide. Analyzers 1 Controllers 2 Transmitters 3 Monitors 4

Research Grade Meters

Electrodes ph EC DO /14 dge 2 P/L e

Transcription:

MULTI meter INSTRUCTION MANUAL European Catalogue Number(s): S1: 665-0311 S2: 665-0312 Version: 2 Issued: September 11, 2014 11:39 am ba77077e02 09/2014

Legal address of manufacturer: VWR International bvba Researchpark Haasrode 2020 Geldenaaksebaan 464 B-3001 Leuven +32 16 385011 http://be.vwr.com Manufactured in Germany. 2 ba77077e02 09/2014

Contents - Contents 1 Safety........................................ 7 1.1 Safety information............................... 7 1.1.1 Safety information in the operating manual..... 7 1.1.2 Safety signs on the meter.................. 7 1.1.3 Further documents providing safety information. 7 1.2 Safe operation.................................. 8 1.2.1 Authorized use........................... 8 1.2.2 Requirements for safe operation............. 8 1.2.3 Unauthorized use......................... 8 2 Overview..................................... 9 2.1 meter............................... 9 2.2 Sensors....................................... 9 3 Technical data................................ 10 3.1 Measuring ranges, resolution, accuracy............. 10 3.1.1 ph/orp............................... 10 3.1.2 Oxi................................... 11 3.1.3 Cond.................................. 12 3.2 General data.................................. 13 4 Commissioning............................... 15 4.1 Scope of delivery............................... 15 4.2 Power supply.................................. 15 4.3 Initial commissioning............................ 15 4.3.1 Inserting the batteries..................... 16 5 Operation.................................... 17 5.1 General operating principles...................... 17 5.1.1 Keypad................................ 17 5.1.2 Display................................ 18 5.1.3 Status information (meter)................. 18 5.1.4 Connectors............................. 19 5.2 Switching on the meter.......................... 19 5.3 Switching off.................................. 20 5.4 Navigation.................................... 20 5.4.1 Operating modes........................ 20 5.4.2 Measured value display................... 21 5.4.3 Menus and dialogs....................... 21 ba77077e02 09/2014 3

Contents 5.4.4 Elements in menus and dialogs............. 21 5.4.5 Navigation example 1:Setting the language.... 22 5.4.6 Example 2 on navigation: Setting the date and time............................. 24 5.5 Channel display................................ 26 5.5.1 Display of several sensors in the measuring mode............................. 26 6 ph value..................................... 27 6.1 Measuring.................................... 27 6.1.1 Measuring the ph value................... 27 6.1.2 Measuring the temperature................. 28 6.2 Calibration.................................... 29 6.2.1 Why calibrate?.......................... 29 6.2.2 When do you have to calibrate?............. 29 6.2.3 Automatic calibration (AutoCal)............. 29 6.2.4 Manual calibration (AnyCal)................ 32 6.2.5 Calibration points........................ 35 6.2.6 Calibration data.......................... 36 7 ORP voltage.................................. 38 7.1 Measuring.................................... 38 7.1.1 Measuring the ORP...................... 38 7.1.2 Measuring the temperature................. 39 7.2 ORP calibration................................ 39 8 Dissolved oxygen............................. 40 8.1 Measuring.................................... 40 8.1.1 Measuring D.O........................... 40 8.1.2 Measuring the temperature................. 41 8.2 Calibration.................................... 42 8.2.1 Why calibrate?.......................... 42 8.2.2 When to calibrate?....................... 42 8.2.3 Calibration procedure..................... 42 8.2.4 Calibration in water vapor-saturated air (air calibration vessel)..................... 42 8.2.5 Calibrating with a comparison measurement (OxiComp)............................. 43 8.2.6 Calibration data.......................... 45 9 Conductivity................................. 47 9.1 Measuring.................................... 47 9.1.1 Measuring the conductivity................. 47 9.1.2 Measuring the temperature................. 48 9.2 Calibration.................................... 49 9.2.1 Why calibrate?.......................... 49 4 ba77077e02 09/2014

Contents 9.2.2 When to calibrate?....................... 49 9.2.3 Determining the cell constant (calibration in control standard)...................... 49 9.2.4 Calibration data......................... 50 10 Settings..................................... 52 10.1 Measurement settings........................... 52 10.1.1 Settings for ph measurements.............. 52 10.1.2 Buffer sets for calibration.................. 53 10.1.3 Calibration interval....................... 54 10.1.4 Settings for ORP measurements............ 55 10.1.5 Settings for D.O. sensors.................. 56 10.1.6 Calibration interval....................... 57 10.1.7 Settings for conductivity measuring cells...... 58 10.1.8 Calibration interval....................... 60 10.1.9 Temperature compensation................ 60 10.1.10 Setting the TDS factor.................... 61 10.2 Sensor-independent settings..................... 62 10.2.1 System................................ 62 10.2.2 Data storage............................ 63 10.2.3 Automatic Stability control................. 63 10.3 Reset........................................ 63 10.3.1 Resetting the measurement settings......... 63 10.3.2 Resetting the system settings.............. 65 11 Data memory................................. 66 11.1 Manual storage................................ 66 11.2 Automatic data storage at intervals.............. 67 11.3 Measurement data memory...................... 69 11.3.1 Editing the measurement data memory....... 69 11.3.2 Erasing the measurement data memory...... 71 11.3.3 Measurement dataset.................... 71 11.3.4 Memory locations........................ 71 12 Transmitting data (USB interface)................ 72 12.1 Options for data transmission..................... 72 12.2 Connecting a PC.............................. 73 12.3 MultiLab Importer.............................. 73 13 Maintenance, cleaning, disposal, accessories..... 74 13.1 Maintenance.................................. 74 13.1.1 General maintenance activities............. 74 13.1.2 Replacing the batteries................... 74 13.2 Cleaning..................................... 75 13.3 Packing...................................... 75 13.4 Disposal..................................... 75 ba77077e02 09/2014 5

Contents 13.5 Accessories................................... 76 13.5.1 General information...................... 76 13.5.2 ph / ORP.............................. 76 13.5.3 Conductivity............................ 77 13.5.4 D.O................................... 77 14 What to do if................................... 79 14.1 ph/orp...................................... 79 14.2 Dissolved oxygen............................... 80 14.3 Conductivity................................... 82 14.4 General information............................. 83 15 Firmware update.............................. 84 16 Glossary..................................... 85 17 Index........................................ 89 18 Technical service............................. 94 19 Warranty..................................... 94 20 Compliance with local laws and regulations....... 94 6 ba77077e02 09/2014

Safety 1 Safety 1.1 Safety information 1.1.1 Safety information in the operating manual This operating manual provides important information on the safe operation of the meter. Read this operating manual thoroughly and make yourself familiar with the meter before putting it into operation or working with it. The operating manual must be kept in the vicinity of the meter so you can always find the information you need. Important safety instructions are highlighted in this operating manual. They are indicated by the warning symbol (triangle) in the left column. The signal word (e.g. "CAUTION") indicates the level of danger: WARNING indicates a possibly dangerous situation that can lead to serious (irreversible) injury or death if the safety instruction is not followed. CAUTION indicates a possibly dangerous situation that can lead to slight (reversible) injury if the safety instruction is not followed. NOTE indicates a possibly dangerous situation where goods might be damaged if the actions mentioned are not taken. 1.1.2 Safety signs on the meter Note all labels, information signs and safety symbols on the meter and in the battery compartment. A warning symbol (triangle) without text refers to safety information in this operating manual. 1.1.3 Further documents providing safety information The following documents provide additional information, which you should observe for your safety when working with the measuring system: Operating manuals of sensors and other accessories Safety datasheets of calibration or maintenance accessories (such as buffer solutions, electrolyte solutions, etc.) ba77077e02 09/2014 7

Safety 1.2 Safe operation 1.2.1 Authorized use This meter is authorized exclusively for ph, ISE, ORP, dissolved oxygen and conductivity measurements in a field and laboratory environment. Only the operation and running of the meter according to the instructions and technical specifications given in this operating manual is authorized (see section 3 TECHNICAL DATA, page 10). Any other use is considered unauthorized. 1.2.2 Requirements for safe operation Note the following points for safe operation: The meter may only be operated according to the authorized use specified above. The meter may only be supplied with power by the energy sources mentioned in this operating manual. The meter may only be operated under the environmental conditions mentioned in this operating manual. The meter may only be opened if this is explicitly described in this operating manual (example: Inserting the batteries). 1.2.3 Unauthorized use The meter must not be put into operation if: it is visibly damaged (e.g. after being transported) it was stored under adverse conditions for a lengthy period of time (storing conditions, see section 3 TECHNICAL DATA, page 10). 8 ba77077e02 09/2014

Overview 2 Overview 2.1 meter The meter enables you to perform measurements (ph, U, conductivity, dissolved oxygen) ) quickly and reliably. The provides the maximum degree of operating comfort, reliability and measuring certainty for all applications. 3 2 1 1 Keypad 2 Display 3 Connectors 2.2 Sensors A measuring system ready to measure consists of the meter and a suitable sensor. Suitable sensors are ph electrodes, ORP electrodes, D.O. sensors and conductivity measuring cells. ba77077e02 09/2014 9

Technical data 3 Technical data 3.1 Measuring ranges, resolution, accuracy 3.1.1 ph/orp Measuring ranges, resolution Manual temperature input Variable Measuring range Resolution ph -2.0... +20.0 0.1-2.00... +20.00 0.01-2.000... + 19.999 0.001 U [mv] -2500... +2500 1-1200.0... +1200. 0.1 T [ C] -5.0... +105.0 0.1 T [ F] 23.0... +221.0 0.1 Variable Range Increment T manual [ C] -25... +130 1 T manual [ F] -13... +266 1 Accuracy (± 1 digit) Variable Accuracy Temperature of the test sample ph / range * -2.0... +20.0 ± 0.1 +15 C... +35 C -2.00... +20.00 ± 0.01 +15 C... +35 C - 2.000... + 19.999 ± 0.005 +15 C... +35 C U [mv] / range -2500... +2500 ± 1 +15 C... +35 C -1200.0... +1200. ± 0.3 +15 C... +35 C T [ C] / temperature sensor NTC 30 ± 0.1 PT 1000 ± 0.1 * when measuring in a range of ± 2 ph around a calibration point The accuracy values specified here apply exclusively to the meter. The accuracy of the electrodes and buffer solutions has to be taken into account additionally. 10 ba77077e02 09/2014

Technical data 3.1.2 Oxi Measuring ranges, resolution (depending on the sensor) Variable Measuring range Resolution Concentration [mg/l] 0... 20.00 (0... 20.0) 20.0... 90.0 (20... 90) Saturation [%] 0... 200.0 (0... 600) 0... 600 D.O. partial pressure [mbar] 0... 200.0 (0... 1250) 0... 1250 0.01 (0.1) 0.1 (1) 0.1 (1) 1 0.1 (1) 1 T [ C] 0... 50,0 0.1 Accuracy (± 1 digit) Variable Accuracy Concentration [mg/l] ± 0.5 % of measured value at ambient temperature of +5 C... +30 C Saturation [%] ± 0.5% of measured value when measuring in the range of ± 10 K around the calibration temperature D.O. partial pressure [mbar] T [ C] / temperature sensor NTC 30 ± 0.1 PT 1000 ± 0.1 ± 0.5 % of measured value at ambient temperature of +5 C... +30 C Correction functions Temperature compensation Salinity correction Air pressure correction Accuracy better than 2 % at 0... +40 C 0... 70.0 SAL Automatic through integrated pressure sensor in the range of 500... 1100 mbar The accuracy values specified here apply exclusively to the meter. The accuracy of the D.O. sensors has also to be taken into account. ba77077e02 09/2014 11

Technical data 3.1.3 Cond Measuring ranges, resolution (depending on the sensor) Variable Measuring range Resolution ϰ[µs/cm] 0.000... 1.999* 0.00... 19.99** 0.0... 199.9 200... 1999 0.001 0.01 0.1 1 ϰ [ms/cm] 2.00... 19.99 20.0... 199.9 200... 1000 ρ (Resistivity) [Ohm*cm] ρ (Resistivity) [kohm*cm] ρ (Resistivity) [MOhm*cm] 1.000... 1.999 2.00... 19.99 20.0... 199.9 200... 1999 2.00... 19.99 20.0... 199.9 200... 1999 2.00... 19.99** 20.0... 199.9* SAL 0.0... 70.0 according to the IOT table TDS 0... 1999 mg/l 2.00... 19.99 g/l 20.0... 199.9 g/l 0.01 0.1 1 0.001 0.01 0.1 1 0.01 0.1 1 0.01 0.1 0.1 1 0.01 0.1 T [ C] -5.0... +105.0 0.1 T [ F] +23.0... +221.0 0.1 * only possible with cells of the cell constant, 0.010 cm-1 ** only possible with cells of the cell constant, 0.010 cm -1 or 0.090... 0.110 cm -1 Cell constants Cell constant C Values Can be calibrated in the ranges 0.450... 0.500 cm -1 0.800... 0.880 cm -1 Adjustable 0.090... 0.110 cm -1 0.010 cm -1 (fixed) 0,250... 1,000 cm -1 Reference temperature Reference temperature Adjustable Values 20 C (Tref20) 25 C (Tref25) 12 ba77077e02 09/2014

Technical data Accuracy (± 1 digit) Variable Accuracy Temperature of the test sample ϰ and ρ / temperature compensation None (Off) ± 0.5 % Nonlinear (nlf) ± 0.5 % 0 C... +35 C according to EN 27 888 ± 0.5 % +35 C... +50 C enhanced nlf function Linear (lin) ± 0.5 % +10 C... +75 C SAL / range 0.0... 42.0 ± 0.1 +5 C... +25 C TDS [mg/l] ± 0.2 +25 C... +30 C ± 0.5 % T [ C] / temperature sensor NTC 30 ± 0.1 PT 1000 ± 0.1 The accuracy values specified here apply exclusively to the meter. The accuracy of the measuring cell has also to be taken into account. Further data are given in the documentation of your sensor. 3.2 General data Dimensions ca. 180 x 80 x 55 mm Weight Approx. 0.4 kg Mechanical structure Type of protection IP 67 Electrical safety Protective class III Test certificates CE Ambient conditions Storage -25 C... +65 C Operation +5 C... +55 C Admissible relative humidity Yearly mean: < 75 % 30 days/year: 95 % Other days: 85 % ba77077e02 09/2014 13

Technical data Power supply Batteries Rechargeable batteries Operational life 4 x 1.5 V alkali-manganese batteries, type AA 4 x 1.2 V NiMH rechargeable batteries, type AA (no charging function) Up to 1000 h without / 150 h with illumination ph sensor input Input resistance > 5 * 10 12 ohm Input current < 1 * 10-12 A USB interface Type USB 1.1 USB B (device), data output Baud rate Adjustable: 1200, 2400, 4800, 9600, 19200 Baud Data bits 8 Stop bits 2 Parity None Handshake RTS/CTS Cable length Max. 3 m Guidelines and norms used EMC EC directive 2004/108/EC EN 61326-1 EN 61000-3-2 EN 61000-3-3 FCC Class A Meter safety EC directive 2006/95/EC EN 61010-1 UL 61010-1 CAN/CSA-C22.2#61010-1 IP protection class EN 60529 14 ba77077e02 09/2014

Commissioning 4 Commissioning 4.1 Scope of delivery Meter 4 batteries 1.5 V Mignon type AA Short instructions CD-ROM with USB drivers detailed operating manual Software MultiLab Importer 4.2 Power supply The is supplied with power in the following ways: Battery operation (4 x alkaline manganese batteries, type AA) USB operation via a connected USB-B cable 4.3 Initial commissioning Perform the following activities: Insert the supplied batteries Switch on the meter (see section 5.2 SWITCHING ON THE METER, page 19) Set the date and time (see section 5.4.6 EXAMPLE 2 ON NAVIGATION: SETTING THE DATE AND TIME, page 24) ba77077e02 09/2014 15

Commissioning 4.3.1 Inserting the batteries 1 2 1 Screws 2 Battery compartment CAUTION Make sure that the poles of the batteries are positioned correctly. The ± signs on the batteries must correspond to the ± signs in the battery compartment. You can operate the meter either with normal batteries or with rechargeable batteries (Ni-MH). In order to charge the batteries, an external charging device is required. 1. Place four batteries (type Mignon AA) in the battery compartment. 2. Close the battery compartment. 3. Set the date and time (see section 5.4.6 EXAMPLE 2 ON NAVIGATION: SETTING THE DATE AND TIME, page 24). 16 ba77077e02 09/2014

Operation 5 Operation 5.1 General operating principles 5.1.1 Keypad In this operating manual, keys are indicated by brackets <..>. The key symbol (e.g. <OK>) generally indicates a short keystroke (under 2 sec) in this operating manual. A long keystroke (approx. 2 sec) is indicated by the underscore behind the key symbol (e.g. <OK_>). <F1>: <F1_>: <F2>: <F2_>: <On/Off>: <MODE>: Softkeys providing situation dependent functions, e.g.: <F2>/[USB output]: Outputs data to the USB interface <F2_>/[USB output]: Configures the automatic data output to the USB interface Switches the meter on or off Selects the measured parameter / Quits the settings <CAL>: <CAL_>: <STR>: <STR_>: <RCL>: <RCL_>: < >< >: < _>< _>: <OK>: <OK_>: <HOLD> Calls up the calibration procedure Displays the calibration data Saves a measured value manually Opens the menu for the automatic save function Displays the manually stored measured values Displays the automatically stored measured values Menu control, navigation Increments, decrements values Increments, decrements values continuously Opens the menu for measurement settings / confirms entries Opens the menu for system settings Freezes the measured value (HOLD function) ba77077e02 09/2014 17

Operation 5.1.2 Display 3 ph 2 1 AutoCal TEC HOLD AR Menu 6.093 15.03.2014 08:00 25.0 C 4 5 6 7 8 1 Status information (sensor) 2 Measured value (with unit) 3 Measured parameter 4 Channel display: Plug position of the sensor 5 Sensor symbol (calibration evaluation, calibration interval) 6 Measured temperature (with unit) 7 Status information (meter) 8 Softkeys and date + time 5.1.3 Status information (meter) AR HOLD Stability control (AutoRead) is active (calibration) Measured value is frozen (<HOLD> key) Batteries are almost empty Data are automatically output to the USB-B interface at intervals 18 ba77077e02 09/2014

Operation 5.1.4 Connectors 5 1 2 3 4 1 ph electrode 2 Temperature sensor 3 D.O. sensor or conductivity measuring cell 4 Mini USB-B interface 5 Service interface CAUTION Only connect sensors to the meter that cannot return any voltages or currents that are not allowed (> SELV and > current circuit with current limiting). Almost all customary sensors fulfill these conditions. 5.2 Switching on the meter 1. Switch the meter on with <On/Off>. The meter performs a self-test. The display shows the manufacturer's logo while the self-test is being performed. The measured value display appears. ph 6.949 25.0 C Menu 15.02.2014 14:15 ba77077e02 09/2014 19

Operation 5.3 Switching off 1. Switch the printer off with <On/Off>. Automatic shut-off function The instrument has an automatic shut-off function in order to save the batteries (see section 10.2.1 SYSTEM, page 62). The automatic shut-off function switches off the meter if no key is pressed for an adjustable period. The automatic shut-off function is not active if the communication cable is connected if the Automatic data storage function is active, or with automatic data transmission Display illumination The meter automatically switches off the display illumination if no key is pressed for 30 seconds. The illumination is switched on with the next keystroke again. You can also generally switch the display illumination on or off (see section 10.2.1 SYSTEM, page 62). 5.4 Navigation The principles of navigation in menus and dialogs are explained in the following sections. 5.4.1 Operating modes The instrument has the following operating modes: Operating mode Measuring Calibration Storing in memory Transmitting data Setting Explanation The measurement data of the connected sensor are shown in the measured value display The course of a calibration with calibration information, functions and settings is displayed The meter stores measuring data automatically or manually The meter transmits measuring data and calibration records to a USB-B interface automatically or manually. The system menu or a sensor menu with submenus, settings and functions is displayed Only those displays and functions are available in the active operating mode that are currently being required. 20 ba77077e02 09/2014

Operation 5.4.2 Measured value display In the measured value display, open the setting menus with <OK>. The current functions of the softkeys are shown on the display. Use <OK> (short pressure) to open the menu for calibration and measurement settings for the displayed measured parameter. Use <OK_> (long keystroke (approx. 2 s) to open the Storage & config menu with the sensor-independent settings. Use the keys of the keypad to carry out further functions such as storage or calibration (see section 5.1.1 KEYPAD, page 17). These functions are not available in other operating situations. 5.4.3 Menus and dialogs The menus for settings and dialogs in procedures contain further subelements. To select a subelement, use the < >< > keys. The current selection is displayed with a frame. To make further settings, switch to the next higher menu level with <F1>[Back]. Use <MODE> to return to the measured value display. 5.4.4 Elements in menus and dialogs Submenus The name of the submenu is displayed at the upper edge of the frame. Submenus are opened by confirming with <OK>. Example: System General Interface Clock Service information Reset Back 15.02.2014 14:15 Settings Settings are indicated by a colon. The current setting is displayed on the right-hand side. The setting mode is opened with <OK>. Subsequently, the setting can be changed with < >< > and <OK>. Example: ba77077e02 09/2014 21

Operation General Language: Deutsch Beep: Off Illumination: On Contrast: 50 % Switchoff time: 1 h Temperature unit: C Stability control: On Back 15.02.2014 14:15 Functions Functions are designated by the name of the function. They are immediately carried out by confirming with <OK>. Example: Display the Calibration record function. ph Calibration record Calibration data storage Buffer: One point calibration: Calibration interval: Unit for slope: TEC Yes 7 d mv/ph i 4.00 7.00 10.00 (20 C) 15.02.2014 Back 14:15 Messages Information is marked by the i symbol. It cannot be selected. Example: ph Calibration record Calibration data storage Buffer: One point calibration: Calibration interval: Unit for slope: TEC Yes 7 d mv/ph i 4.00 7.00 10.00 (20 C) 15.02.2014 Back 14:15 5.4.5 Navigation example 1:Setting the language 1. Press the <On/Off> key. The measured value display appears. The instrument is in the measuring mode. 22 ba77077e02 09/2014

Operation ph 6.949 25.0 C Menu 15.02.2014 14:15 2. Open the Storage & config menu with <OK_>. The instrument is in the setting mode. Storage & config System Data storage Back 15.02.2014 14:15 3. Select the System submenu with < >< >. The current selection is displayed with a frame. 4. Open the System submenu with <OK>. System General Interface Clock Service information Reset Back 15.02.2014 14:15 5. Select the General submenu with < >< >. The current selection is displayed with a frame. 6. Open the General submenu with <OK>. ba77077e02 09/2014 23

Operation General Language: Deutsch Beep: Off Illumination: On Contrast: 50 % Switchoff time: 1 h Temperature unit: C Stability control: On Back 15.02.2014 14:15 7. Open the setting mode for the Language with <OK>. General Language: Deutsch Beep: Off Illumination: On Contrast: 50 % Switchoff time: 1 h Temperature unit: C Stability control: On Back 15.02.2014 14:15 8. Select the required language with < >< >. 9. Confirm the setting with <OK>. The meter switches to the measuring mode. The selected language is active. 5.4.6 Example 2 on navigation: Setting the date and time The meter has a clock with a date function. The date and time are shown in the measured value display. When storing measured values and calibrating, the current date and time are automatically stored as well. The correct setting of the date and time and date format is important for the following functions and displays: Current date and time Calibration date Identification of stored measured values. Therefore, check the time at regular intervals. After a fall of the supply voltage (empty batteries), the date and time are reset. 24 ba77077e02 09/2014

Operation The date format can be switched from the display of day, month, year (dd.mm.yyyy) to the display of month, day, year (mm/dd/yyyy or mm.dd.yyyy). 1. In the measured value display: Open the Storage & config menu with <OK_>. The instrument is in the setting mode. 2. Select and confirm the System / Clock menu with < >< > and <OK>. The setting menu for the date and time opens up. 3. Select and confirm the Time menu with < >< > and <OK>. The hours are highlighted. Clock Date format: dd.mm.yyyy Date: 15.02.2014 Time: 14:15:25 Back 15.02.2014 14:15 4. Change and confirm the setting with < >< > and <OK>. The minutes are highlighted. 5. Change and confirm the setting with < >< > and <OK>. The seconds are highlighted. 6. Change and confirm the setting with < >< > and <OK>. The time is set. 7. If necessary, set the Date and Date format. The setting is made similarly to that of the time. 8. To make further settings, switch to the next higher menu level with [Back]<F1>. or Switch to the measured value display with <MODE>. The instrument is in the measuring mode. ba77077e02 09/2014 25

Operation 5.5 Channel display The manages the connected sensors and displays which sensor is plugged to which connection. ph 6.949 24.8 C Channel display Display of the plug position for the respective parameter The filled bar indicates for each connected sensor to which plug position of the meter it is connected. 15.02.2014 14:15 USB output 5.5.1 Display of several sensors in the measuring mode The measured values of the connected sensors can be displayed in the following ways: Clear display of all connected sensors Detailed display of one sensor With the softkey you can very easily switch between the two display types. The suitable softkey is displayed depending on the operating situation. ph 6.966 25.0 C ϰ 0.840 1/cm 25.0 C ph 6.966 25.0 C 15.02.2014 14:15 USB output 15.02.2014 14:15 USB output 26 ba77077e02 09/2014

ph value 6 ph value 6.1 Measuring 6.1.1 Measuring the ph value NOTE When connecting an grounded PC/printer, measurements cannot be performed in grounded media as the values would be incorrect. The USB interface is not galvanically isolated. To ensure the high measurement accuracy of the measuring system, always measure with a calibrated electrode (see section 6.2 CALIBRATION, page 29). 1. Connect the ph electrode to the meter. 2. If necessary, select the measured parameter with <MODE>. 3. When measuring without temperature sensor: Temper the test sample, or measure the current temperature. Enter the temperature value in the menu. 4. Immerse the ph electrode in the test sample. The measured value is checked for stability (automatic stability control). The display of the measured parameter flashes. 5. Wait for a stable measured value. The display of the measured parameter no longer flashes. ph 6.949 24.8 C Menu 15.02.2014 14:15 Stability control (AutoRead ) Criteria for a stable measured value The stability control function (AutoRead) continually checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of measured values (see section 10.2.3 AUTOMATIC STABILITY CONTROL, page 63). The Stability control function checks whether the measured values are stable ba77077e02 09/2014 27

ph value within the monitored time interval. Measured parameter Time interval Stability in the time interval ph value 15 seconds Δ : better than 0.01 ph Temperature 15 seconds Δ : better than 0.5 C The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer. Freezes the measured value (HOLD function) With the HOLD function, you can freeze the current measured value. The displayed measured value stops changing until you switch the HOLD function off. 1. Freeze the measured value with <HOLD>. The [HOLD] status indicator is displayed. 2. Release the frozen measured value again with <HOLD>. The HOLD function is switched off. The [HOLD] status display disappears. 6.1.2 Measuring the temperature For reproducible ph measurements, it is essential to measure the temperature of the test sample. You have the following options to measure the temperature: Automatic measurement of the temperature with the temperature sensor (NTC30 or Pt1000) integrated in the sensor. Measurement by an external temperature sensor. Manual determination and input of the temperature. The measuring instrument recognizes whether a suitable sensor is connected and automatically switches on the temperature measurement. The display of the temperature indicates the active temperature measuring mode: Temperature sensor Resolution of the temp. display Temp. measurement yes 0.1 C Automatic with temperature sensor - 1 C Manual If you wish to measure (or calibrate) without temperature sensor, proceed as follows: 1. Measure the current temperature of the test sample. 2. In the <OK>/pH/Man. temperature menu, set the temperature value with < >< >. 28 ba77077e02 09/2014

ph value 6.2 Calibration 6.2.1 Why calibrate? ph electrodes age. This changes the zero point (asymmetry) and slope of the ph electrode. As a result, an inexact measured value is displayed. Calibration determines and stores the current values of the zero point and slope of the electrode. Thus, you should calibrate at regular intervals. 6.2.2 When do you have to calibrate? After connecting a sensor Routinely within the framework of the company quality assurance When the calibration interval has expired 6.2.3 Automatic calibration (AutoCal) Make sure that in the sensor menu, Buffer menu, the buffer set is correctly selected (see 10.1.1 SETTINGS FOR PH MEASUREMENTS, PAGE 52). Use any one to five buffer solutions of the selected buffer set in ascending or descending order. Below, calibration with Technical buffers (TEC) is described. When other buffer sets are used, other nominal buffer values are displayed. Apart from that, the procedure is identical. If single-point calibration was set in the menu, the calibration procedure is automatically finished with the measurement of buffer solution 1 and the calibration record is displayed. 1. Connect the ph electrode to the meter. 2. When measuring without temperature sensor: Temper the buffer solution, or measure the current temperature. Enter the temperature value in the menu. 3. In the measured value display, select the measured parameter ph or mv with <MODE>. 4. Start the calibration with <CAL>. The calibration display for the first buffer appears (voltage display). ba77077e02 09/2014 29

ph value ph Buffer 1 AutoCal TEC -180.0 24.8 C mv 15.02.2014 14:15 5. Thoroughly rinse the electrode with deionized water. 6. Immerse the electrode in the first buffer solution. 7. When measuring without temperature sensor: Temper the buffer solution, or measure the current temperature. Enter the temperature value in the menu. 8. Start the measurement with <OK>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes. ph AutoCal TEC AR Buffer -180.0 24.8 C 15.02.2014 14:15 10.011 mv 9. Wait for the end of the measurement with stability control or accept the calibration value with <OK>. The calibration display for the next buffer appears (voltage display). 10. If necessary, finish the calibration procedure as a single-point calibration with <MODE>. The calibration record is displayed. For single-point calibration, the instrument uses the Nernst slope (-59.2 mv/ph at 25 C) and determines the zero point of the electrode. Continuing with twopoint calibration 11. Thoroughly rinse the electrode with deionized water. 12. Immerse the electrode in the second buffer solution. 30 ba77077e02 09/2014

ph value 13. When measuring without temperature sensor: Temper the buffer solution, or measure the current temperature. Enter the temperature value in the menu. 14. Start the measurement with <OK>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes. ph Buffer 7.000 0.0 mv 24.8 C AutoCal TEC AR 15.02.2014 14:15 15. Wait for the measurement with stability control to be completed or terminate the stability control and take over the calibration value with <OK>. The calibration display for the next buffer appears (voltage display). 16. If necessary, finish the calibration procedure as a two-point calibration with <MODE>. The calibration record is displayed. Continuing with three- to five-point calibration 17. Thoroughly rinse the electrode with deionized water. 18. Immerse the electrode in the next buffer solution. 19. When measuring without temperature sensor: Temper the buffer solution, or measure the current temperature. Enter the temperature value in the menu. 20. Start the measurement with <OK>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes. ph AutoCal TEC AR Buffer 180.0 24.8 C 4.010 mv 15.02.2014 14:15 ba77077e02 09/2014 31

ph value 21. Wait for the measurement with stability control to be completed or terminate the stability control and take over the calibration value with <OK>. The calibration display for the next buffer appears (voltage display). 22. If necessary, use <MODE> to finish calibration or Continue calibrating using the next buffer with <OK>. Calibration is automatically completed after the last buffer of a buffer set has been measured. Then the calibration record is displayed. The calibration line is determined by linear regression. 6.2.4 Manual calibration (AnyCal) Make sure that in the sensor menu, Buffer menu, the AnyCal buffer set is correctly selected (see section 10.1.1 SETTINGS FOR PH MEASUREMENTS, page 52). Use any one to five buffer solutions in ascending or descending order. If single-point calibration was set in the menu, the calibration procedure is automatically finished with the measurement of buffer solution 1 and the calibration record is displayed. 1. Connect the ph electrode to the meter. The ph measuring window is displayed. 2. When measuring without temperature sensor: Temper the buffer solution, or measure the current temperature. Enter the temperature value in the menu. 3. In the measured value display, select the measured parameter ph or mv with <MODE>. 4. Start the calibration with <CAL>. The calibration display appears (voltage display). ph AnyCal Buffer 0.0 24.8 C mv 1 15.02.2014 14:15 5. Thoroughly rinse the electrode with deionized water. 6. Immerse the electrode in the first buffer solution. 32 ba77077e02 09/2014

ph value 7. When measuring without temperature sensor: Temper the buffer solution, or measure the current temperature. Enter the temperature value in the menu. 8. Start the measurement with <OK>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes. ph Buffer 0.0 24.8 C 7.000 mv AnyCal AR 15.02.2014 14:15 9. Wait for the end of the measurement with stability control or accept the calibration value with <OK>. The calibration display for the setting of the nominal buffer value appears. ph AnyCal Buffer 7.000 24.8 C 1 15.02.2014 14:15 10. Set the nominal buffer value for the measured temperature with < >< >. 11. Accept the set calibration value with <OK>. The calibration display for the next buffer appears (voltage display). 12. If necessary, finish the calibration procedure as a single-point calibration with <MODE>. The calibration record is displayed. For single-point calibration, the instrument uses the Nernst slope (-59.2 mv/ph at 25 C) and determines the zero point of the electrode. Continuing with twopoint calibration 13. Thoroughly rinse the electrode with deionized water. ba77077e02 09/2014 33

ph value 14. Immerse the electrode in the second buffer solution. 15. When measuring without temperature sensor: Temper the buffer solution, or measure the current temperature. Enter the temperature value in the menu. 16. Start the measurement with <OK>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes. 17. Wait for the measurement with stability control to be completed or terminate the stability control and take over the calibration value with <OK>. The calibration display for the setting of the nominal buffer value appears. ph AnyCal Buffer 4.035 24.8 C 2 15.02.2014 14:15 18. Set the nominal buffer value for the measured temperature with < >< >. 19. Accept the set calibration value with <OK>. The calibration display for the next buffer appears (voltage display). 20. Finish the calibration procedure as a two-point calibration with <MODE>. The calibration record is displayed. Continuing with three- to five-point calibration 21. Thoroughly rinse the electrode with deionized water. 22. Immerse the electrode in the next buffer solution. 23. When measuring without temperature sensor: Temper the buffer solution, or measure the current temperature. Enter the temperature value in the menu. 24. Start the measurement with <OK>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes. 25. Wait for the measurement with stability control to be completed or terminate the stability control and take over the calibration value with <OK>. The calibration display for the setting of the nominal buffer value appears. 34 ba77077e02 09/2014

ph value ph AnyCal Buffer 9.958 24.8 C 3 15.02.2014 14:15 26. Set the nominal buffer value for the measured temperature with < >< >. 27. Accept the set calibration value with <OK>. The calibration display for the next buffer appears (voltage display). 28. Use <MODE> to finish calibration or Continue calibrating using the next buffer with <OK>. After the fifth buffer has been measured the calibration is automatically finished. Then the calibration record is displayed. The calibration line is determined by linear regression. 6.2.5 Calibration points Calibration can be performed using one to five buffer solutions in any order (single-point to five-point calibration). The meter determines the following values and calculates the calibration line as follows: Determined values Displayed calibration data 1-point Asy Zero point = Asy Slope = Nernst slope (-59.2 mv/ph at 25 C) 2-point 3-point to 5-point Asy Slp. Asy Slp. Zero point = Asy Slope = Slp. Zero point = Asy Slope = Slp. The calibration line is calculated by linear regression. You can display the slope in the units, mv/ph or % (see section 10.1.1 SETTINGS FOR PH MEASUREMENTS, page 52). ba77077e02 09/2014 35

ph value 6.2.6 Calibration data The calibration record is automatically transmitted to the interface after calibrating. Displays the calibration data Displaying the calibration data memory The calibration record of the last calibration is to be found under the menu item, <OK> / Calibration / Calibration record. To open it, press the <CAL_> key in the measured value display. Subsequently, you can transmit the displayed calibration data to the interface, e.g. to a PC, with the <F2>/[USB output] key. The calibration records of the last calibrations are to be found in the menu, <OK> / Calibration / Calibration data storage. Menu item Calibration / Calibration data storage / Display Calibration / Calibration data storage / Output to USB Setting/ function Explanation - Displays the calibration record. Further options: Scroll through the calibration records with < >< >. Output the displayed calibration record to the interface with <F2>/[USB output]. Quit the display with <F1>/ [Back] or <OK>. Switch directly to the measured value display with <MODE>. - Outputs the calibration records to the interface. Calibration evaluation After calibrating, the meter automatically evaluates the calibration. The zero point and slope are evaluated separately. The worse evaluation of both is taken into account. The evaluation appears on the display and in the calibration 36 ba77077e02 09/2014

ph value record. Display Calibration record Zero point [mv] Slope [mv/ ph] +++ -15... +15-60.5... -58.0 ++ -20... <-15 or >+15... +20 + -25... <-20 or >+20... +25 - -30... <-25 or >+25... +30 >-58.0... -57.0-61.0... <-60.5 or >-57.0... -56.0-62.0... <-61.0 or >-56.0... -50.0 Clean the electrode according to the electrode operating manual Error Error <-30 oder >+30 <-62.0 oder >-50.0 Error elimination (see section 14 WHAT TO DO IF..., page 79) Calibration record (example) Ser. no. 11292113 CALIBRATIONpH 15.02.2014 15:55 AutoCal TEC Buffer 1 4.01 Buffer 2 7.00 Buffer 3 10.01 Voltage 1 184.0 mv Voltage 2 3.0 mv Voltage 3-177.0 mv Temperature 1 24.0 C Temperature 2 24.0 C Temperature 3 24.0 C Slope -60.2 mv/ph Asymmetry 4.0 mv Sensor +++ etc... ba77077e02 09/2014 37

ORP voltage 7 ORP voltage 7.1 Measuring 7.1.1 Measuring the ORP NOTE When connecting an grounded PC/printer, measurements cannot be performed in grounded media as the values would be incorrect. The USB interface is not galvanically isolated. 1. Connect the ORP electrode to the meter. 2. If necessary, select the U (mv) display with <MODE>. 3. When measuring without temperature sensor: Temper the test sample, or measure the current temperature. Enter the temperature value in the menu. 4. Rinse the ORP electrode and immerse it in the test sample. The measured value is checked for stability (automatic stability control). The display of the measured parameter flashes. 5. Wait for a stable measured value. The display of the measured parameter no longer flashes. U 157.0 mv 24.8 C Menu 15.02.2014 14:15 Stability control (AutoRead ) Criteria for a stable measured value The stability control function (AutoRead) continually checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of measured values (see section 10.2.3 AUTOMATIC STABILITY CONTROL, page 63). The Stability control function checks whether the measured values are stable within the monitored time interval. Measured parameter Time interval Stability in the time interval ORP 15 seconds Δ : better than 0.3 mv Temperature 15 seconds Δ : better than 0.5 C The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer. 38 ba77077e02 09/2014

ORP voltage Freezes the measured value (HOLD function) With the HOLD function, you can freeze the current measured value. The displayed measured value stops changing until you switch the HOLD function off. 1. Freeze the measured value with <HOLD>. The [HOLD] status indicator is displayed. 2. Release the frozen measured value again with <HOLD> or <MODE>. The HOLD function is switched off. The [HOLD] status display disappears. 7.1.2 Measuring the temperature For reproducible ORP measurements, it is essential to measure the temperature of the test sample. You have the following options to measure the temperature: Automatic measurement of the temperature by the temperature sensor (NTC30 or Pt1000) integrated in electrode. Measurement by an external temperature sensor. Manual determination and input of the temperature. The measuring instrument recognizes whether a suitable sensor is connected and automatically switches on the temperature measurement. The display of the temperature indicates the active temperature measuring mode: Temperature sensor Resolution of the temp. display Temp. measurement yes 0.1 C Automatic with temperature sensor - 1 C Manual If you wish to measure without temperature sensor, proceed as follows: 1. Measure the current temperature of the test sample. 2. Set the temperature value with < >< >. or In the <OK>/U/Man. temperature menu, set the temperature value with < >< >. 7.2 ORP calibration ORP electrodes are not calibrated. You can, however, check ORP electrodes by measuring the ORP of a test solution and comparing the value with the nominal value. ba77077e02 09/2014 39

Dissolved oxygen 8 Dissolved oxygen 8.1 Measuring 8.1.1 Measuring D.O. NOTE When connecting an grounded PC/printer, measurements cannot be performed in grounded media as the values would be incorrect. The USB interface is not galvanically isolated. 1. Connect the D.O. sensor to the meter. The D.O. measuring screen is displayed. 2. If necessary, select the measured parameter with <MODE>. 3. Immerse the D.O. sensor in the test sample. O 2 [OXY 11] 7.9 mg/l 25.1 C 15.02.2014 08:00 USB output 4. If necessary, select the measured parameter with <MODE>. D.O. concentration [mg/l] D.O. saturation [%] D.O. partial pressure [mbar]. Salinity correction When measuring the concentration of solutions with a salt content of more than 1 g/l, a salinity correction is required. For this, you have to measure and input the salinity of the measured medium first. When the salinity correction is switched on, the [SAL] indicator is displayed in the measuring window. You can switch the salinity correction on or off and enter the salinity in the menu for calibration and measurement settings (see section 10.1.5 SETTINGS FOR D.O. SENSORS, page 56). Stability control (AutoRead ) The stability control function (AutoRead) continually checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of measured values (see section 10.2.3 AUTOMATIC STABILITY CONTROL, 40 ba77077e02 09/2014

Dissolved oxygen Criteria for a stable measured value page 63). The Stability control function checks whether the measured values are stable within the monitored time interval. Measured parameter Time interval Stability in the time interval D.O. concentration 20 seconds Δ : better than 0.05 mg/l D.O. saturation 20 seconds Δ : better than 0.6 % D.O. partial pressure 20 seconds Δ : Better than 1.2 mbar Temperature 15 seconds Δ : better than 0.5 C The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer. Freezes the measured value (HOLD function) With the HOLD function, you can freeze the current measured value. The displayed measured value stops changing until you switch the HOLD function off. 1. Freeze the measured value with <HOLD>. The [HOLD] status indicator is displayed. 2. Release the frozen measured value again with <HOLD>. The HOLD function is switched off. The [HOLD] status display disappears. 8.1.2 Measuring the temperature For reproducible D.O. measurements, it is essential to measure the temperature of the test sample. The temperature is automatically measured by the temperature sensor (NTC30 or Pt1000) integrated in the sensor. The D.O. sensor OXY 11 has an integrated temperature sensor. ba77077e02 09/2014 41

Dissolved oxygen 8.2 Calibration 8.2.1 Why calibrate? D.O. sensors age. This changes the slope of the D.O. sensor. Calibration determines the current slope of the sensor and stores this value in the instrument. 8.2.2 When to calibrate? After connecting a sensor Routinely within the framework of the company quality assurance When the calibration interval has expired 8.2.3 Calibration procedure The provides 2 calibration procedures: Calibration in water vapor-saturated air. Use an OxiCal air calibration vessel for the calibration. Calibration via a comparison measurement (e.g. Winkler titration according to DIN EN 25813 or ISO 5813). At the same time, the relative slope is adapted to the comparison measurement by a correction factor. When the correction multiplier is active, the [Factor] indicator appears in the measuring window. 8.2.4 Calibration in water vapor-saturated air (air calibration vessel) For this calibration procedure, the Comparison meas. setting must be set to Off in the Calibration menu. Proceed as follows to calibrate the instrument: 1. Connect the D.O. sensor to the meter. 2. Put the D.O. sensor into the air calibration vessel. The sponge in the air calibration vessel must be moist (not wet). Leave the sensor in the air calibration vessel for a time long enough to adjust. 3. Start the calibration with <CAL>. The last calibration data (relative slope) is displayed. 42 ba77077e02 09/2014

Dissolved oxygen O 2 0.88 25.1 C OxiCal 15.02.2014 08:00 4. Start the measurement with <OK>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. A progress bar is displayed and the display of the measured parameter flashes. 5. Wait for the end of the measurement with stability control or accept the calibration value with <OK>. The calibration record is displayed and output to the interface. 6. Using <F1>/[Continue] or <OK>, switch to the measured value display. The meter displays a D.O. saturation of approx. 101.7 % in the air calibration vessel. 8.2.5 Calibrating with a comparison measurement (OxiComp) For this calibration procedure, the Comparison meas. setting must be set to On in the Calibration menu. Before calibrating via a comparison measurement, the sensor should be calibrated in the air calibration vessel. 1. Connect the D.O. sensor to the meter. 2. Immerse the D.O. sensor in the reference solution. 3. Start the calibration with <CAL>. O 2 7.9 mg/l 25.1 C OxiComp 15.02.2014 08:00 ba77077e02 09/2014 43