CLEVELAND CONTROLS MODEL 6650 DIGITAL MANOMETER INSTRUCTION MANUAL DM

Similar documents
Model A Mini AC/DC Clamp Meter. User's Guide

Owner's Manual. True RMS Multimeter. Model No Safety Operation Maintenance Español

User's Guide. MiniTec TM Series Model MN25 MultiMeter

User Guide True RMS Multimeter Extech EX205T

Mini Digital Multimeter

User's Guide. Digital Multimeter. Model MN42

Mini Digital Multimeter

Model R5010. Instruction Manual. TRMS Digital Multimeter. reedinstruments www.

Manometers 2PSI, 5PSI, 15PSI, 30PSI, 100PSI , , ,

OWNER S MANUAL 9908-TE. HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER

350 SMART MANOMETER OPERATING INSTRUCTIONS

Model R5010. TRMS Digital Multimeter. Instruction Manual

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389

Mini Digital Multimeter Model MN15. User's Guide

User's Guide. Extech AM A AC Analog Clamp Meter

OPERATING INSTRUCTION

Autoranging True RMS Multimeter User Manual

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

Model INSTRUCTION MANUAL DIGITAL MULTIMETER

2 in1 LAN Tester & Multimeter. Model: PCE-LT 1

Mini Digital Multimeter

User s Guide. 600A AC Clamp Meter. Model 38387

CM-220 True RMS AC CLAMP METER INSTRUCTION MANUAL

AutoRanging Digital MultiMeter

MODEL ATV 4/6 BATTERY POWERED DIGITAL VACUUM INSTRUMENT

2 in 1 LAN Tester and Multimeter Model:

Table of Contents. General Information. Document Sure-Aire Flow Monitoring System. User and Service Manual WARNING CAUTION

Autoranging Multimeter plus IR Thermometer Extech 450 Patented

Autoranging Mini Multimeter

PT DIGITAL INDICATOR OPERATING MANUAL REVISION 3.2 3/ Rev /04

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

OPERATING INSTRUCTION. Pen-Type Digital Multimeter

SA SERIES Arlingate Lane, Columbus, Ohio (614)

INSTRUCTION MANUAL. Digivac 760 Series. Models 100Led P / P / 29.9" 100Led 760 / SixVac 801 / 760 / 2c 801 Micro

True RMS Multimeter with IR Thermometer

User's Guide. Heavy Duty Differential Pressure Manometer. Model

Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only.

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: 57314

User Guide. Heavy Duty Differential Pressure Manometer. Model HD750

NT PRESSURE TRANSDUCER MODELS 4100, 4210, 4130, 4230

True RMS Multimeter with IR Thermometer

USER MANUAL. True RMS Multimeter. Extech EX430A

USER MANUAL. Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX310

HAND-HELD THERMOCOUPLE THERMOMETER SERVICE MANUAL. CATALOG NUMBERS , and , and PROPRIETARY

1. Introduction 2. Specifications 3. Symbol Definition and Button Location 4. Operation Instructions 5. Temperature Measurement 6.

2200 Series. Quick Start Guide SAFETY CHECKLIST. Safety Made Simple

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE PTC-20 PELTIER TEMPERATURE CONTROL SYSTEM

Manual for PSDC_OPT_EIGHT_01: Optics Power Supply

AC/DC TRUE RMS CLAMP METER ACM-2036

MicroTrans EQ Signal Processor. Operation & Maintenance Manual

DP2500 DP0250. Differential Pressure Transmitter. Product Bulletin

User's Guide. Digital Multimeter plus IR Thermometer. Extech 450 Patented

User Manual. Heavy Duty Differential Pressure Manometer. Model SDL720. Additional User Manual Translations available at

Digital Clamp Meter User Manual

DFI 7000 DIGITAL FORCE INDICATOR USER S GUIDE 1.0 INTRODUCTION 2.0 DESCRIPTION. 2.1 Configuration. 2.2 Specifications

EPWR Series. RoHS. Product Description. Available Products. Installation Instructions 01

E101 - Strain Gauge Transducer Display Module. Contents

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TC-20 TEMPERATURE CONTROL SYSTEM

ScopeMeter. Accessories. PC Software. Protective Cases. Powering and Charging

ETHOS Auto Ranging Digital Multimeter

ATE QUICK START GUIDE. KEPCO An ISO 9001 Company. VOLTAGE/CURRENT STABILIZED POWER SUPPLY I INTRODUCTION

Strain Gauge Measurement by Applying Displacement using Strain Indicator

PP-8516 & PP Electronic Static Pressure Controllers and Transmitters for Cental Air Supply Systems General Instructions

User's Guide. 800 Amp Clamp Meters. EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter

700 Series 200 Amp Clamp Meters

larger, brighter display for ease of use

USER'S MANUAL DCL-650

OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER

Differential Liquid/Gas Pressure Transmitter

DP DP2500 DP0250

Autoranging Multimeter Extech EX503

14 Pressure Measurement. Measuring Devices Description Pressure range Connection Device Page

2016 SIGLENT TECHNOLOGIES CO.,LTD

USER GUIDE. Heavy Duty Differential Pressure Manometer. Model HD700

OWNER'S MANUAL. Model:PSV Amp SWITCHING DC POWER SUPPLY.

BOP QUICK START GUIDE 100W, 200W, 400W. KEPCO An ISO 9001 Company. BIPOLAR OPERATIONAL POWER SUPPLY WARNINGS I INTRODUCTION II SAFETY.

INSTALLATION INSTRUCTIONS

Wilcoxon Research P702B/P702BF-1 operating guide Power unit/amplifier for piezoelectric transducers with internal amplifiers

Series F75 Fail-Safe Module Installation, Operation and Maintenance Instructions

Model 120D and Model 120CEX Wireline Data Recorder. Operation and Maintenance Manual

HP UPS R/T3000 G2. Overview. Precautions. Kit contents. Installation Instructions

Actuator Commissioning Tool

ED-1100 PORTABLE VARIABLE EDDY CURRENT INSTRUMENT

MD9 MULTIDROP INTERFACE INSTRUCTION MANUAL

REMOTE DISCHARGE PROBE

ADC7520 SERIES. 1600W Battery Chargers and Power Supplies

BC-9000 Field Calibration Procedure CF1_FIELDCAL_BC9000 Revised 03/16/2013

IVC-222HPII Precision Voltage/Current Calibrator

This 4200-RM Rack Mount Kit is for installation in 4200-CAB series cabinets only.

Industrial Multimeter Extech EX510

DFI 9000 PORTABLE DIGITAL INDICATOR USER S GUIDE

Model 2657A-LIM-3 LO Interconnect Module

MiG2 CONTROLLERS. 2 & 4 Stage General Purpose Controllers, with Air-conditioning Facilities

AirTest Model CN9000 Series Sensor Controller

Heavy Duty Differential Pressure Manometer. Model User's Guide

AC TO DC ADAPTERS

EPS Power Supply

PanelView Plus/VersaView CE Terminals and Display Modules

1-36V, 0-3A DC Power Supply

Transcription:

CLEVELAND CONTROLS MODEL 6650 DIGITAL MANOMETER INSTRUCTION MANUAL DM-6650.02 1 Instruction Manual DM-6650.02

TABLE OF CONTENTS I.0 INTRODUCTION 3 1.1 General Description... 3 1.2 Specifications... 3 1.3 Unpacking... 4 1.4 Operator Controls... 5 1.5. Connections... 5 1.6 Initial Checkout... 6 2.0 OPERATION 7 2.1 Continuity Testing... 7 2.2 Pressure Measurement... 8 2.3. Analog Output Circuit... 8 2.4 Nickel-cadmium Battery Pack... 8 2.5 Field Maintenance... 9 FIGURES Figure 1: Model 6650 Digital Manometer... 10 Figure 2: Integrated Pressure Sensing System... 10 Figure 3: Continuity Testing & Calibration of a Sensing Switch... 12 Instruction Manual DM-6650.02 2

I.0 INTRODUCTION 1.1 GENERAL DESCRIPTION The Cleveland Controls Model 6650 Digital Manometer provides accurate measurement of differential pressure up to 10" wc. Pressure readings are digitally displayed on the 3-½ digit LCD display, in two common engineering units (Inches H 2 O and kpa). High and low air sample ports located on the front of the instrument accept ¼" ID flexible tubing. Included with the manometer is a 10-foot (10') length of flexible vinyl tubing, to be cut following guidelines in this manual. The manometer is compatible with air and other dry, clean, non-corrosive gases that will not attack polyester, vinyl, silicone, or silicone-based adhesives. In addition, an electrical continuity test circuit is provided for convenient testing and calibration of pressure sensing switches. NO, NC, and COM continuity test connections, plus red and green LED indicators, are located on the front panel. A 3-prong continuity test cord, six-feet (6') long is included with the manometer. An analog output signal (0-1 VDC) is provided to permit remote pressure recording or monitoring. Analog output terminals are located on the rear panel. The manometer is powered by a internal rechargeable battery pack. It can also be connected to an AC line via the power receptacle located on the rear panel and the provided AC cord. The Model 6650 Digital Manometer is factory calibrated and requires no routine maintenance. 1.2 SPECIFICATIONS Range: 0 10.00" wc Units of Measurement: 0.0 10.00" wc; 0.0 2.49 kpa; 0.0 1.999" wc; 0.0 0.497 kpa. Maximum Pressure Input: 5 psi. Over Range Indication: Digit blanking. Accuracy: ±1.5% full scale, ±1 digit, typical at 77 ºF (25 ºC). Temperature Error: ±0 0.014%/ºF, typical; ±0 0.025%/ºC, typical. Ambient Temperature Limits: 40 122 ºF (5 50 ºC). 3 Instruction Manual DM-6650.02

Analog Output: 0 1 VDC. Operator Controls: Power On/Off, Zero Adjustment; Pressure-units Selector; Display Hold. Display: 3-½ digit LCD with low battery indication. Settling Time: 5 seconds. Pressure Input Connection: ¼" ID flexible tubing. Continuity Connection: ⅛" miniature banana jacks with test cable. Analog Output Connection: 5-way miniature binding post (⅛"). Power Input Connection: IEC universal connector with cord. Power Requirements: 120V, 50/60 Hz., 4W. Battery Pack: Rechargeable nickel-cadmium. Operating Time: 15 hours maximum from full charge. Recharging Time: 16 hours maximum. Media: Dry, clean gases that will not attack polyester, vinyl, silicone, or silicone-based adhesives. Case: Black, high-impact plastic with tilt handle. 1.3 UNPACKING The shipping carton contains the following items: Model 6650 Digital Manometer AC cord set (P/N 27560) Transparent flexible vinyl tubing, ¼" ID, ⅜" OD, 10' length (P/N 25781) 3-prong continuity test cord, 6' long (P/N 27636) Aspiration probe,12" (P/N 31309) Impact probe, 12" (P/N 31310) Instruction Manual DM-6650.xx Confirm that each of these items is in the carton. If any item is missing or damaged, please call Customer Service at (216) 398-0330. Instruction Manual DM-6650.02 4

1.4 OPERATOR CONTROLS All operator controls are located on the front panel. The Zero Control is a single-turn potentiometer used to set the display to 000 when no differential pressure is present at the inputs. The Power Control is a double-action pushbutton switch that controls the AC line and battery power. Power is ON when the switch is pushed in. Power is OFF when the switch is depressed again and returns to the out position. The Hold Control is a double-action pushbutton switch. The pressure reading freezes at the moment the switch is pushed to the in position. Depressing the switch a second time releases the hold. The Unit Selection Controls are four pushbutton switches used to select the unit of measurement displayed. They are interlocked so that only one can be in the in position at a time ensuring that only one unit of measurement can be selected for display. 1.5. CONNECTIONS Connections are located on the front and back panels. Pressure Input Connections are made to the HI and LO pressure ports located on the front panel. The ports are designed to accommodate ¼" ID flexible vinyl tubing (supplied with the Digital Manometer). Electrical Continuity Test Circuit Connections are the ⅛" miniature banana jacks located on the front panel. Connections are made with the test cable included with the Digital Manometer. Analog Output Terminals are the ⅛" binding posts located on the back panel. They accept ⅛" miniature banana plugs, tip jacks, or wire leads to provide output for a remote recording or monitoring device (not included). AC Power Receptacle is located on the rear panel and accepts the AC cord included with the Digital Manometer. 5 Instruction Manual DM-6650.02

1.6 INITIAL CHECKOUT 1.6.1 Connect the power cord. 1.6.2 Press the ON-OFF switch to the ON position and push the desired unit of measurement switch. 1.6.3 When the display appears, adjust the Zero Control until it reads 0.00. 1.6.4 Apply a known pressure to the HI port and confirm that a pressure reading appears. An alternative method is to press a finger tightly over the HI port. The air in the port will be compressed sufficiently to produce a reading. 1.6.5 Plug the continuity test cable into the three test jacks. Connect each color-coded wire to the matching, labeled jack. 1.6. 6 Touch the Normally Open (NO) wire to the Common (C) wire, and confirm that the NO LED (red) lights up. 1.6.7 Touch the Normally Closed (NC) wire to the Common (C) wire, and confirm that the NC LED (green) lights up. Instruction Manual DM-6650.02 6

2.0 OPERATION 2.1 CONTINUITY TESTING 2.1.1 Prove circuit integrity, check and calibrate sensing switches using the continuity testing capability of the Digital Manometer. Note: When testing an installed switch, first disconnect the power source from the switch to avoid shock hazard and damage to the continuity tester. 2.1.2 The continuity circuit consists of two LED s and a series resistor in open circuit condition. Continuity between COMMON (C) and either Normally Open (NO) or Normally Closed (NC) illuminates the red (NO) or green (NC) LED. 2.1.3 A non-resistive ⅛ amp fuse is located in the power supply connection to protect the battery and manometer circuits if the continuity circuit is inadvertently connected to AC power or any other high potential source. 2.1.4 To determine the pressure at which a sensing switch opens or closes, connect the continuity test cable to the three continuity jacks on the front panel of the manometer (see Section 1.6, above). 2.1.5 Connect the color-coded wires of the test cable to the sensing switch contacts. Be sure to connect the common wire (C) to the common contact, the normally open (NO) wire to the normally open contact, and the normally closed (NC) wire to the normally closed contact. 2.1.6 Connect the manometer s pressure inputs in parallel with the switch. Press the manometer s ON/OFF switch to the ON position, and push the 0-10 In. H2O unit of measurement switch. When the display appears, adjust the ZERO control until it reads 0.00. 2.1.7 At pressures below the sensing switch set point, the green LED (NC) will be on, indicating that this circuit is closed. The red LED (NO) will remain off. 2.1.8 At the switch s set point, as indicated on the manometer, the green NC LED will go out and the red NO LED will come on. 7 Instruction Manual DM-6650.02

2.2 PRESSURE MEASUREMENT 2.2.1 The measuring element in the Model 6650 Digital Manometer is a piezoresistive pressure transducer. The transducer has a full scale input range of 0-10 wc and a linear output of 1-6 VDC.The Digital Manometer can be used to measure gauge or differential pressures or vacuums. 2.2.2 To measure gauge pressure, connect the pressure source to the HI input. Leave the LOW input open to atmosphere. 2.2.3 To measure differential pressure, connect the higher pressure source to the HI input and the lower pressure source to the LOW input. If a negative reading is displayed, reverse the high and low pressure sources. 2.2.4 To measure gauge vacuum, connect the vacuum source to the LO input. Leave the HI input open to atmosphere. 2.2.5 To measure differential vacuum, connect the lower vacuum source to the LOW input and the higher vacuum source to the HI input.if a positive reading is displayed, reverse the high and low pressure sources. 2.3. ANALOG OUTPUT CIRCUIT 2.3.1 The analog output circuit is a differential amplifier with a gain of x 1n. 2.3.2 The differential 0-5 volt signal on the main PC assembly is connected to an adjustable attenuator, which is set to provide a 0-1 volt signal to the differential amplifier. 2.3.3 This provides a buffered 0-1 volt signal for remote monitoring or recording. 2.4 NICKEL-CADMIUM BATTERY PACK 2.4.1 The internal rechargeable battery pack provides for cordless operation. When fully charged, it will operate for 15 hours before recharging is required. 2.4.2 To charge the battery, connect the manometer to the AC line with the manometer turned OFF. Charging time is 16 hours from fully discharged to fully charged. Additional charging is not harmful. Instruction Manual DM-6650.02 8

2.4.3 For maximum battery life, the battery should be fully discharged and recharged every 3 months. 2.4.4 When the battery voltage drops below 10V, the LO BAT symbol lights up on in the display. 2.5 FIELD MAINTENANCE 2.5.1 The Model 6650 Digital Manometer is a solid state device with no moving parts and no scheduled maintenance requirements other than the battery charging procedure described in Section 2.4. 2.5.2 The manometer is factory calibrated and does not require additional field calibration. 2.5.3 Do not exceed the maximum pressure rating (5 psi). 2.5.4 Use only with compatible media: dry, clean gases that will not attack polyester, vinyl, silicone, or siliconebased adhesives. 9 Instruction Manual DM-6650.02

Figure 1: Model 6650 Digital Manometer A. Digital Readout: 3-½ digit liquid crystal display. B. Field calibration to null readout. C. Continuity tester to verify circuit integrity. D. Air sample ports accept ¼" ID flexible tubing. E. Operation selector allows various conversion factor readings. HOLD freezes the readout for value comparison. F. Analog output terminals for remote recording or monitoring. G. 110 V AC power line receptacle. H. Rechargeable nickel cadmium battery pack. Figure 2 (facing page): Digital Manometer/Integrated Pressure-Sensing System Connect the NO, NC & COM terminals on the air switch to the NO, NC & COM terminals on the manometer. 2. Tubing: Connect the HI side of the air switch to the HI side of the manometer. Connect the LOW side of the air switch to the LOW side of the manometer. 3. Tubing from manometer to T fitting and tubing from air switch to T fitting must be the same length. Also, keep these pieces as short as possible. 4. The tubing connecting the probe(s) to the tee(s) can be a different length. The manometer ships with a 10 ft. piece of tubing to be cut as needed. Illustration is not to scale! Instruction Manual DM-6650.02 10

A. Duct B. Fan/blower. C. Filter/damper. D. Pressure Impact Probe (+). E. Pressure Aspiration Probe (-). F. Air sensing switch. Note: use probes (D + E) together for differential pressure measurement. Important! This illustration is not drawn to scale. It is intended to show the electrical and pressure connections as clearly as possible.note that the pieces of tubing between the manometer and the tees must be the same length as the pieces running from the air switch connectors to the tees. Also, keep them as short as possible.the tubing connecting the probe(s) to the tee(s)can be a different length. The manometer ships with a 10 piece of tubing to be cut as needed. 11 Instruction Manual DM-6650.02

Figure 3: Typical application of Series 6650 Digital Manometer for continuity testing and calibration of an air pressure sensing switch. Cleveland Controls DIVISION OF UNICONTROL INC. Tel: 216-398-0330 Fax: 216-398-8558 1111 Brookpark Rd Cleveland OH 44109 Email:saleshvac@unicontrolinc.com Web page: http://www.clevelandcontrols.com Instruction Manual DM-6650.02 12