Calibration of the depth-measuring device of Rockwell hardness testing machines

Similar documents
Introducing the world most advanced Shore/IRHD durometer DIGI TEST II

HIGH PRECISION SURVEY AND ALIGNMENT OF LARGE LINEAR COLLIDERS - HORIZONTAL ALIGNMENT -

The Virtual CMM a software tool for uncertainty evaluation practical application in an accredited calibration lab

IECEx Certificate of Conformity

Hardness measurement of rubber and elastomers IRHD ISO 48 VLRH DIN ISO 27588

A NON-IMAGING OPTICAL SYSTEM FOR CHARACTERISATION OF BALL- SHAPED MICRO-INDENTERS

D-510 PISeca Single-Electrode Capacitive Sensor Probe


STABILITY OF THE 32 MN FORCE CALIBRATION MACHINE AND INDUSTRIAL APPLICATIONS

Analysis of Combined Probing Measurement Error and Length Measurement Error Test for Acceptance Testing in Dimensional Computed Tomography

Product Information. LS 1679 Incremental Linear Encoder with Integrated Roller Guide

TESA UPD Gauge Block Comparator with a 25 mm Measuring Span

COMPARISON OF ERROR MAPPING TECHNIQUES FOR COORDINATE MEASURING MACHINES USING THE PLATE METHOD AND LASER TRACER TECHNIQUE

Thickness of the standard piece: 10 mm The most important calibration data are engraved in the side face of the specimen.

Stress Analysis of Cross Groove Type Constant Velocity Joint

Orbital forming of SKF's hub bearing units

Calibration Equipment

The high speed profile sensor. 2D/3D profile sensor SENSORS & SYSTEMS SENSORS & SYSTEMS. High speed profile measurement

Operation Manual. Autocollimator Test Wedge for quick testing of Electronic Autocollimators

Autocollimator Calibration for Synchrotron Metrology: Advancing from Plane to Solid Angle

UNITEST - A new device for the static and dynamic testing of unconventional machine structures

A U T O C O L L I M A T O R S

Product Information. ECI 119 Absolute Rotary Encoder


Operating Instructions

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

Coordinate Measuring Machines with Computed Tomography

Height Gauges. Height Gauges K-1

Provisional Product Information. ECN 1123 EQN 1135 Rotary Encoders with EnDat 2.2 for Safety-Related Applications

TRIBOMETERS MICRO-MACRO TRIBOLOGY TESTING

ECI 1319 EQI 1331 EBI 1335 Absolute Rotary Encoders without Integral Bearing. Product Information

Good Practice guide to measure roundness on roller machines and to estimate their uncertainty

Product Information. QUADRA-CHEK 3000 Evaluation Electronics For Metrological Applications

TL Laser Systems for Tool Measurement

DYNAMIC MODELING OF WORKING SECTIONS OF GRASSLAND OVERSOWING MACHINE MSPD-2.5

ND 1202 ND 1203 ND monochrome flat-panel display for position values, dialogs and inputs, graphics functions and soft keys

FlexAirConnecT Dust Insensitive Multi-Fiber Connector with Low Loss and Low Mating Force

Micro Cutting Tool Measurement by Focus-Variation

PRECISION MEASURING INSTRUMENTS

IECEx Certificate of Conformity

CALIBRATION. Sylvac Dial Gauge Calibrator

What makes Bolt Self-loosening Predictable?

Product Information. ECN 1023 EQN 1035 Rotary Encoders with EnDat 2.2 for Safety-Related Applications

QM-HEIGHT SERIES HIGH-PERFORMANCE HEIGHT GAUGE PRE1424(2) SMALL TOOL INSTRUMENTS AND DATA MANAGEMENT

HOW TO PROVE AND ASSESS CONFORMITY OF GUM-SUPPORTING SOFTWARE PRODUCTS

Accuracy for CMMs and Machine Tools

Calibration Procedure For An Industrial Robot

DYNAMIC, FLEXIBLE, PRECISE

IECEx Certificate of Conformity

T EN TROVIS Electropneumatic Positioner

Length Measurement. Measuring lengths up to 30 m

CALCULATION OF 3-D ROUGHNESS MEASUREMENT UNCERTAINTY WITH VIRTUAL SURFACES. Michel Morel and Han Haitjema

Product and Applications Description. Application Programs. Note. instabus EIB Technical Product Information. June 2005

MINISCALE PLUS Product Overview

8661 EN. Precision Torque Sensor Non-contact transmission for rotating applications optional measurement of angle and speed.

Indicating Digital Transmitter for Differential Pressure and Flow. Media 6 with LCD Media 6 with LED

Product Information. ECI 4090S Absolute Rotary Encoder with 180 mm Hollow Shaft and DRIVE-CLiQ Interface for Safety-Related Applications

ROUNDTEST RA Roundness/Cylindricity Measurment

TECHNICAL PAPER SP80 & SP80H

ksa MOS Ultra-Scan Performance Test Data

Single Actuator Shaker Design to Generate Infinite Spatial Signatures

Torque Motor Rotary indexing tables TO. Technology that inspires

Rotary Measuring Technology Absolute encoders, Multiturn, DeviceNet

MEASURING MEANS UNDERSTANDING.

Laser profile sensor, laser scanner 2D/3D for beads and grooves

Series 3730 Type Electropneumatic Positioner

OUT OF MACHINE CALIBRATION TECHNIQUE FOR ANALOG PROBES

DESIGN AND VERIFICATION OF THE TRINANO ULTRA PRECISION CMM

Test Report UST - Universal Surface Tester

Entering a Whole New Dimension The ULTRAHEAT XS Heat Meter

PRISMO Vario. Specifications and Performance Features. Industrial Measuring Technology from Carl Zeiss. Wide sensor variety

The development of a laser array measurement system for three dimensional positioning testing in machine tool

IECEx Certificate of Conformity

Product Information. RCN 6000 Absolute Angle Encoder with Integral Bearing and Large Hollow Shaft

The Uncertainty of Parallel Model Coordinate Measuring Machine

NUMERICAL METHOD TO ESTIMATE TOLERANCES COMBINED EFFECTS ON A MECHANICAL SYSTEM

PORTABLE SURFACE ROUGHNESS GAGE. POCKET SURF

INTERAPID Depth Feet. Brown & Sharpe CENTER FINDER. Model with a flat measuring face. Model with a prismatic measuring face H-16

Position Transducers Pivot head mounting potentiometric up to 300 mm, IP54 Series TEX

MEAM 550 Modeling and Design of MEMS Spring Solution to homework #3. In our notation and values, k = = =

Measurements using three-dimensional product imaging

DMT PERFORMING UNDER PRESSURE DMT. Applications. Special features. Description 1/9

TubeInspect and TubeInspect S

Improved Efficiency for Multi-Technology Inspections (UT and ECT)

n Measuring range 0... ± 0,02 Nm to 0... ± 1000 Nm n Low linearity deviation of ± 0.05 % F.S. n Intelligent operating state indicator

Precimar. PLM and CiM Universal Length Measuring Machines LENGTH MEASURING MACHINES FOR HIGH-END CALIBRATION

ODU-MAC in the Aluminium Frame

Physikalisch-Technische Bundesanstalt P1B

Natural Viewing 3D Display

Product Information ECI 1118 EQI 1130 ECI 1119 EQI Absolute Rotary Encoders without Integral Bearing

Product Information ROC 425 ROQ 437. Absolute Rotary Encoders with Solid Shaft for Safety-Related Applications

LENORD. +BAUER... automates motion. Tower oscillation monitor GEL 3011 for safety applications. Technical information Version General.

Experimental evaluation of static stiffness of a spatial translational parallel manipulator.

BESMAK BMT - E SERIES

ROUNDTEST RA-2200 SERIES

MINIMAT-EC-Servo Screwdriver Spindles

Absolute Encoders Multiturn

A Six Degree of Freedom, Piezoelectrically Actuated Translation Stage

Rotary Measuring Technology

Transcription:

Calibration of the depth-measuring device of Rockwell hardness testing machines F. Löffler, A. Sawla, P. Strobel* Physikalisch-Technische Bundesanstalt, Braunschweig * Heinrich Bareiss Prüfgerätebau GmbH, Oberdischingen Abstract In addition to the indirect verification using reference blocks and a reference indenter, it is possible to realise the calibration by direct verification of the test force, the indenter, the depth-measuring device and the testing cycle. Some conditions of the calibration process, e.g. the verification range, and the necessary accuracy of the depth-measuring device are described in standard ISO 6508-2. With reference to these parameters, a new depthmeasuring calibration device was developed. The study describes the function of this device and the calibration process. The advantages of this new process are that the measurement takes place in the axis of force application and that the calibration is independent of the Rockwell hardness scales. During calibration, the force stresses the calibration device but not the sensitive measuring sensor. Results of a real depth-measuring device verification are also presented. 1 Introduction Apart from the indirect calibration of hardness testing machines by means of reference blocks, also a direct method can be applied for calibration. In this case, the test force, the indenter, the depth-measuring device and the testing cycle have to be calibrated. In standard ISO 6508-2, the admissible deviations from the measuring devices or standards are stated, and also from the components of the hardness testing machine to be tested. On the basis of these technical requirements, a device for the calibration of the depth-measuring device was developed. In this paper, the function of the calibration device as well as the calibration process and the results of the device calibration will be described. 2 State of the art The calibration of the depth-measuring device for Rockwell hardness testing machines is, in the case of the indirect measurement method, carried out with reference blocks and a reference indenter. The hardness of a reference block is determined by means of the hardness testing machine to be calibrated. Afterwards, this hardness value is compared to the known value of the reference block. A disadvantage of this method is that for the calibration of a hardness testing machine several reference blocks have to be used so as to cover different hardness ranges and hardness scales. In consequence, the reference blocks wear out. The measurement uncertainties are limited by the calibration methods for the hardness testing machine and the reference block and by material inhomogeneities. Apart from this indirect method, according to standard ISO 6508-2 also a direct method can be

applied to calibrate the hardness testing machine. According to that standard, the test force, the indenter, the depth-measuring device and the testing cycle have to be calibrated. Whilst the calibration of test force, indenter and testing cycle can be carried out by methods which are already known, only a few methods have been developed so far for the calibration of the depth-measuring device. The reason for this is that the calibration of the depth-measuring device has to be carried out at the preliminary test force for the hardness testing machine which, according to ISO 6508-1, is 98,07 N or 29,42 N, and that, at the same time, the requirements of ISO 6508-2 have to be complied with. To this effect, paragraph 4.4.2 of the standard says: "The instrument used to verify the depth-measuring device shall have an accuracy of 0,0002 mm." This means that the limiting deviation of the calibration device is 0,2 µm. Furthermore, this standard requires that calibration be carried out in different subranges and that, for this purpose, "the indenter shall be displaced by known values in the direction of increasing hardness values." In [1], a device for the calibration of the depth-measuring device is described. With this device, length measurement takes place outside the axis in which the indenter is moved. For the length measurement, a Heidenhain electronic length measuring system - type MT 12 - is used [2]. 3 Design example The above-mentioned requirements for a calibration device are complied with in the following design examples. The schematic diagram A in Fig. 1 shows the main elements of the device. The adapter is fastened in the hardness testing machine, and the baseplate is placed on the sample plate of the hardness testing machine. When the preliminary force is applied, the adapter is pressed against the frame on which the force is transmitted via two legs onto a spring and onto the baseplate. The adjustment unit displaces the legs which are carried in a guide at one end while, at the other end, they displace the frame and also the adapter via joints. The displacement takes place by specified intervals within the depth range of approx. 200 µm with the preliminary force being maintained. After each adjustment, the indication of the depth-measuring device of the hardness testing machine is compared with that of the measurement sensor (the calibration process). The length measurement takes place in the force application axis. The limiting deviation required in the standard for the calibration device becomes possible because the necessary adjustment unit and the preliminary test force are decoupled from the point of measurement and thus do not contribute to the measurement uncertainty budget. When the calibration is finished, the calibration device is dismounted and the indenter remounted into the holding fixture of the hardness testing machine. Another design example is shown in the schematic diagram B in Fig. 1. In this case, the same main elements are used, but this time in a cylindrical set-up. With such a compact setup, the dimensions of the device can be reduced considerably. Moreover, due to the cylindrical arrangement, a cost-advantageous design is achieved as the components are mainly manufactured by turning. As a result, an economical prototype series was realised without altering the measurement principle and the quality of the device, thus creating the basis for widespread use of the device.

This method avoids the above-mentioned disadvantages of indirect calibration and offers the following additional advantages: measurement of length variation in the load axis measurement under a preliminary test force measurement without the preliminary test force being applied to the measurement system calibration of the length measurement unit with only one device for all Rockwell scales and for the entire hardness range decoupling of the acting preliminary test force from the length measurement and at the same time measurement of the depth adjustment in the load axis and decoupling of the adjustment unit from the measurement unit of the calibration device. Version A: holding fixture of indenter of HTM joint adapter frame guidance measuring sensor spring leg baseplate adjustment unit sample plate of HTM Version B: holding fixture of indenter of HTM measuring sensor adapter frame baseplate spring adjustment unit sample plate of HTM Fig. 1: Design example of the calibration device

4 Calibration process For the following description of the calibration process, the calibration device according to schematic diagram A was used: (1) Replacement of the indenter by an adapter and setting-up of the calibration device on the sample plate (2) By triggering off the preliminary force setting on the hardness testing machine, the adapter is pressed onto the frame and stretches the spring via the legs. (4) Positioning of the measuring sensor on the baseplate and folding-in of the adjustment unit. (5) Adjustment of the adapter level by setting the adjustment unit, and step-by-step comparison of the indication of the measuring sensor with the indication of the depthmeasuring device (the calibration process). (6) Removal of the preliminary force. (7) Dismounting of the calibration device and remounting of the indenter. 5 Measurement results The comparison measurements were carried out on a conventional mechanical hardness testing machine according to the calibration process described above. After application of the preliminary force, the adjustment unit of the calibration device is coupled in. Now the adapter which is fixed in the machine holder is levelled down to the position which corresponds to the smallest hardness value of the hardness scale to be calibrated. From here, the adapter is adjusted step by step, each time by one hardness value, in the direction of increasing hardness. For each indication value of the depth-measuring device of the hardness testing machine, the associated value indicated by the calibration device is recorded. At the end, the intervals corresponding to one hardness value are calculated. For the hardness scales HRA to HRK, one hardness value corresponds to a variation of the depth by 2 µm (1 µm for the HRN and HRT scales). As during a hardness test the difference between the depth after application of the preliminary test force and the remaining depth after removal of the additional test force is always determined, this was also the case during this measurement. The results are shown in Fig. 2. Fig. 2: Measuring results by comparing the values of the sensor and the hardness machine

The reference value for all measurements is the position of the adapter after application of the preliminary test force. In this measurement example, it is at the indication 100. For the indication 20, for example, which would correspond to a hardness of 20, a theoretical difference of (100-20) 2 µm = 160 µm derives. In this case, however, a depth difference of 160,8 µm was measured from which the deviation from the theoretical value of 0,8 µm follows, which on the left in the diagram is scaled. The total measurement deviations range from - 0,1 µm to + 1,3 µm. After adjustment of the depth-measuring device around a value of + 0,6 µm, the limiting deviation is at ± 0,7 µm. The limiting deviation of ± 1,0 µm required by the standard is thus complied with. 6 Contributions to measurement uncertainty The factors influencing the measurement uncertainty of the calibration device and of the calibration method are in particular the following: limits of permissible errors of the measuring sensor axial misalignment between the adapter and the calibration device tilting of the adapter in relation to the measuring axis force variation during measurement resolution of the indication shocks According to information from the manufacturer, the limit of permissible errors of the measuring sensor is 0,15 µm. At the Geometrical Standards Section of the PTB, the measuring sensor was calibrated additionally in the measurement range from -100 µm to +100 µm. For this purpose, in each subrange (0 to +100 µm and 0 to -100 µm), 10 comparison measurements were carried out starting with 0 up to higher absolute values and back again to 0. The result of these measurements in the subrange from 0 to 100 µm is shown in Fig. 3. The expanded measurement uncertainty is U = 0,5 µm and derives from the standard deviation by multiplication by the coverage factor k = 2. The calculation basis is the "Guide to the Expression of Uncertainty in Measurement." Due to the manufacturing tolerances, the possible axial misalignment is 0,01 mm for the position adapter and 0,2 mm for other adapters which, however, were not used in the measurement described above. In connection with tilting, the axial misalignment can lead to a deviation. The deviation resulting from the tilting of the adapter can be determined by means of the geometry relations of the calibration device. This way, tilting between the measurement axis of the hardness testing device to the measurement axis of the calibration device of 0,1 leads to a deviation of 2 µm. The deviation of the preliminary test force of ± 2 % which is admissible according to standard DIN EN ISO 6508-2 can bring about a change in length which is also dependent on the design and the materials used for the calibration device. For the calibration device shown in schematic diagram A, this means a deviation of 13 nm. The real variation of the preliminary test force, however, was considerably lower for the hardness testing machine used.

correction value 1,40 1,20 1,00 0,80 0,60 0,40 0,20 up down 0,00 0-10 -20-30 -40-50 -60-70 -80-90 -100 indicated value Fig. 3: Result of the comparison measurement in the range from 0 to -100 µm. The correction value derives from the indicated value of the measuring sensor minus the value of the standard measuring device. The resolution of the indication depends strongly on the type of indication of the respective hardness testing machine. For a dial gauge, a resolution of ± 0,1 scale divisions can be assumed. In the case of perceptible vibrations, the measurement has to be rejected because these vibrations can have a considerable influence on the measurement. 7 Summary On the basis of standard DIN EN ISO 6508-1 and -2, a calibration device and an appropriate calibration method were developed. By means of this calibration method, the measurement values of the depth-measuring device of a hardness testing machine are compared to those of the calibration device. The depth-measuring device is thereby subjected to a load (preliminary test force). The function of the calibration device was tested in a comparison measurement. The measurement on a conventional, mechanically working hardness testing machine showed that the limiting deviation of ± 1,0 µm required by the standard is complied with. 9 Literature [1] Pereira, M. F. F.: Calibration of hardness machines and machine parts; in VDI REPORT No. 1194, pp. 287 to 292, 1995 [2] N. N.: Operating Instructions of Heidenhain "Heidenhain-Metro MT 12 / MT 25 / MT 12B / MT 25B;" No. 227 366 94 5/96 F. Löffler, A. Sawla: Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig P. Strobel: Heinrich Bareiss Prüfgerätebau GmbH, Breiteweg 1, 89610 Oberdischingen