Cary 50. Pre-Installation Manual

Similar documents
Firmware Guide. Keysight PXIe Chassis Family

Agilent OBSAI Protocol Tester

Agilent 70612B K18 Switch Matrix

Keysight E5864A Removable Hard Drive for Series Logic Analyzers. Installation Guide

Getting Started with the Agilent Serial BERT N4906B. You only need a few minutes to get started with the Serial BERT.

Agilent E2943A/E2944A ATCA Probes for Advanced Switching Interconnect

Agilent Gb/s Serial BERT

BIT-3000 Dynamic Sequencing Generator and Analyzer. Datasheet 1.11

Agilent 85130C NMD 3.5 mm to Type-N Adapter Kit

Agilent CSV Export Utility

Keysight 11878A 50 Ohm 3.5 mm Adapter Kit

Agilent N2739A 1000 Series Oscilloscope Rack Mount Kit

Agilent Technologies E5346A 38-Pin Probe and E5351A 38-Pin Adapter Cable

Agilent Technologies E5385A 100-Pin Probe

Keysight 11636B APC-3.5 mm Power Divider DC to 26.5 GHz

Keysight 85130F NMD 2.4 mm to 3.5 mm Adapter Kit

Agilent OpenLAB. Data Store. Backup and Restore Guide

Agilent OpenLAB. Data Store. Maintenance Guide

Agilent Protocol Analyzer and Jammer for USB. Quick Start Guide

Keysight 85130D NMD 3.5 mm to 3.5 mm Adapter Kit

Agilent Series Logic Analysis System

* * Agilent Power Distribution Unit (PDU) Installation Guide

Agilent OpenLAB Chromatography Data System (CDS)

87421/22A Power Supply. Operating and Service Manual

Installation Note. Removable Hard Disk Drive Upgrade Kit. For All PNA Series RF Network Analyzers. Network Analyzer Model Number

Agilent OpenLAB Data Analysis Upload Download Tool. User s Guide

Agilent InfiniiMax II 1168A/1169A Probes

Keysight 54006A 6 GHz Passive Divider Probe Kit. User s Guide

Agilent N2916B 6000 and 5000 Series Oscilloscope Rack Mount Kit

iconverter 2-Module Power Chassis

Agilent 1260 Infinity Purification Solution

DDR/LPDDR Custom Configuration Creator Tool. User s Guide

Agilent OpenLAB. Agilent Instrument Controller Installation and Configuration Guide

BenchCel Workstations Software

Agilent OpenLAB Chromatography Data System

Omnitron Systems Technology, Inc. 1. iconverter. 19-Module Managed Power Chassis User s Manual

Agilent Cary 60 Spectrophotometer. User s Guide

Agilent OpenLAB Chromatography Data System (CDS)

Agilent OpenLAB Chromatography Data System (CDS)

86100 Series Infiniium DCA Oscilloscope

Agilent Dissolution Tester Driver

Dell Precision Workstation T7500/T7500n Setup and Features Information

Agilent E2094M IO Libraries

Agilent OpenLAB Chromatography Data System (CDS)

Quick Start/Installation Guide

Installation Note. Source Attenuators and Bias Tees Upgrade Kit. For E8362B/C PNA Series Microwave Network Analyzers. Network Analyzer Model Number

Agilent N4219B Packet Analysis Probe for Serial ATA. User s Guide

Agilent Technologies E5339A 38-Pin Low-Voltage Probe

Agilent OpenLAB Chromatography Data System (CDS)

Agilent EZChrom SI. Startup Guide

Agilent Softtouch 2.0 Midbus Probe

CrystalView DVI Multi INSTALLATION AND OPERATIONS MANUAL Stancliff Road Phone: (281)

Agilent 34826A BenchLink Data Logger for 34980A. Getting Started Guide. Agilent Technologies

Agilent Flowmeter ADM2000

Agilent ChemStation Plus

Keysight Add Source and Receiver Attenuators Upgrade Kit

Agilent OpenLAB Data Store. Maintenance Guide

Dell MD1280 Storage Enclosure Getting Started Guide

Product Documentation

COOPER POWER SERIES. Input/Output (I/O) module installation instructions. Voltage Regulators MN225067EN

Agilent N5410A Fibre Channel Automated Test Application

TA Instruments Installation Requirements for Microcalorimeter Systems

Cary 50 Spectrophotometer. Hardware Manual

EVO-TM2A EVO-TM2B Touch Screen Monitor

TA Instruments Installation Requirements for Microcalorimeter Systems

RZ2 BioAmp Processor. Operator s Manual

Agilent OpenLAB ECM Intelligent Reporter

Keysight E4990A Impedance Analyzer

XBDM. 1015LV, 1020LV, 1030LV, 1020HV Models USER & INSTALLATION MANUAL BYPASS DISTRIBUTION MODULE

Microscan Barcode Reader

Errata. Agilent References in this manual. About this manual. Support for your product

TAM Air Isothermal Calorimeter. Site Preparation Guide

Keysight N7015A/16A Type-C Test Kit. User Guide

Installation and Operations Manual

Agilent U2941A Parametric Test Fixture

Nimsoft Monitor. proxy Guide. v3.1 series

The power behind competitiveness. Delta Infrasuite Power Management. Power Distribution Unit. User Manual.

Keysight Second Source, Combiner, and Mechanical Switches Upgrade Kit

Instruction Manual Standard Multi-Position Stirrers Advanced Multi-Position Stirrers

Setting Up Probes. Online Help

Agilent ChemStation Plus

Model 4001A Modular System Bin Operating and Service Manual

INSTRUCTION MANUAL RA 19-7B RACK ADAPTER HOT SWAP RACK ADAPTER FOR HSF SERIES POWER SUPPLIES

Installation and Quick Start Guide

Model VT120 Fast Timing Preamplifier Operating and Service Manual

Manual Version: V1.00. Video Decoder Quick Guide

RZ5/RZ5D/RZ6 Processor. Operator s Manual

A33606-AIC-02 User's Guide

Version 1.0 March 2014 OLC 10.1

user manual AMOS-2000 Universal Compact, Chassis System, Supporting Nano-ITX Embedded Boards Revision

E-Series Site Preparation Guide

Agilent E5072A Network Analyzer

Automated Tuner System Power Distribution Hub

USER GUIDE. AXIS T8120 Midspan 15 W 1-port ENGLISH

Agilent Lab Advisor. IT Administrator's Guide. Agilent Technologies

User's Guide. Phase Sequence and Motor Rotation Tester Model

Power Supply (48Vdc, 2.5A)

Agilent OpenLAB Chromatography Data System

Wa r n i n g-wa r n i n g-wa r n i n g

1993 Specifications CSJ SPECIAL SPECIFICATION ITEM Transportation Management Console

Transcription:

Cary 50 Pre-Installation Manual

Notices Agilent Technologies, Inc. 1999, 2006 and 2011 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number 8510161800 Edition Third edition, March 2011 Printed in Australia Agilent Technologies, Inc. Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license. Restricted Rights Legend If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as Commercial computer software as defined in DFAR 252.227-7014 (June 1995), or as a commercial item as defined in FAR 2.101(a) or as Restricted computer software as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms, and non- DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data. Safety Notices CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. 2 Cary 50 Pre-Installation Manual

Contents Contents 1. Introduction 5 Pre-installation checklist 6 2. Work Area 7 Suitability 7 Environmental conditions 7 Workbench 8 Electrical specifications 9 3. Equipment On Site 11 Insurance 11 In-house transit routes 11 Inspecting for transit damage 12 Unpacking 13 4. Computer System Requirements 15 Minimum and recommended configurations 15 Setting up your computer 17 Instrument connection 18 Other connections 19 Cary 50 Pre-Installation Manual 3

Contents This page is intentionally left blank. 4 Cary 50 Pre-Installation Manual

Introduction 1. Introduction Pre-installation checklist 6 The Agilent Cary 50 Series spectrophotometers are complete analysis systems that have been fully tested and proven to specification before dispatch from the manufacturing plant. This manual contains general information about preparing the installation site and details of the facilities that must be provided for the installation. The Cary 50 has been designed to be installed by the customer. You may be charged if installation is performed by Agilent service personnel. Cary 50 Pre-Installation Manual 5

Introduction Pre-installation checklist Use the following checklist to make sure you have the work area ready to install the Cary 50. Preparation requirement Complete The work area meets the environmental requirements (see Page 7). A suitable workbench is available (see Page 8). Suitable electrical power supplies are available (see Page 9). A telephone line is available.* All equipment is on site and has been checked for damage (see Page 11). The Cary 50 has been unpacked (see Page 12). A computer that meets requirements is available (see Page 15). A working Microsoft Windows operating system is installed on the computer (see Page 19). * A telephone outlet near the instrument is recommended. This will allow you to seek assistance when developing methods, or for use by Agilent authorized staff to remotely certify your instrument via modem using Telediagnostics. 6 Cary 50 Pre-Installation Manual

Work Area 2. Work Area Suitability 7 Environmental conditions 7 Workbench 8 Electrical specifications 9 Suitability Your Cary instrument is designed for indoor use. It is suitable for these categories: Installation category I Pollution degree 2 Safety class 3 (EN 61010-1) Environmental conditions Sample preparation areas and materials storage facilities should be located in a separate room. For optimum analytical performance it is recommended that the ambient temperature of the work area be between 20 and 25 C and be held constant to within ±2 C throughout the entire working day. Air conditioning is recommended. The room should be temperaturecontrolled if your analyses are particularly sensitive. The work area should have a dust-free, low humidity atmosphere. Cary 50 Pre-Installation Manual 7

Work Area Table 1 Suitable conditions during instrument transportation, non-operation and operation Condition Altitude (m, ft) Temperature ( C, F) Non-operating (transport) 0 2133, 0 7000 5 45, 41 113 20 80 Operating but not necessarily meeting performance specifications Operating within performance specifications 0 2000, 0 6562 0 853, 0 2800 853 2133, 2800 7000 5 31, 41 88 31 40, 88 104 10 35, 50 95 10 25, 50 77 Relative humidity, non-condensing (%) 80 [80-3.33(t-31)] 8 80 CAUTION Operating specifications for the computer, monitor and printer/plotter may differ from those for the Cary spectrophotometer. You must check in the literature provided with these units and arrange the operating environment to suit the complete system. Workbench The workbench must be stable and strong enough to support the total weight of equipment to be used. Table 2 Cary 50 weights and dimensions Weight (kg, lb) Width (mm, in.) Depth (mm, in.) Height (mm, in.) 21, 46 590, 23 500, 20 200, 8 The bench tops should be large enough to permit a free circulation of air around each unit in the system. Remember to provide space for the computer, monitor and printer/plotter. The Cary 50 has been designed to withstand the weight of a standard 15 inch monitor. The recommended workbench height is approximately 90 cm (36 in.). To avoid damage from spillage of the samples being analyzed, the bench tops should be covered with a material that is corrosionresistant and impervious to liquids. 8 Cary 50 Pre-Installation Manual

Work Area Electrical specifications The Cary 50 uses the power supply in the computer to derive the necessary voltages required to operate the instrument. It is therefore necessary to ensure that electrical power supplies for the computer comply with the rules and/or regulations imposed by local authorities responsible for the supply of electrical energy to the workplace. WARNING Electrical Shock Hazard Good electrical grounding is essential to avoid potentially serious shock hazards. Ensure that power outlets are earth grounded at the grounding pin. All power supplies must be single phase AC (alternating current) voltage, three wire system (active, neutral, earth) and should be terminated at an appropriate power outlet receptacle that is within reach of the computer power cord assembly. For safety reasons, a separate power outlet receptacle should be provided for each unit in the system. Do not plan to use extension cords or outlet adapters. Avoid using power supplies from a source that may be subject to electrical or RF (radio frequency) interference from other services; for example, large electric motors, elevators or welders. Table 3. Mains voltage requirements System unit Required supply voltage Power rating (typical) Cary 50 Power supplied directly from the computer Computer 100, 120, 220, 240 VAC, 50/60 Hz 300 VA Printer 100, 120, 220, 240 VAC, 50/60 ±1 Hz 100 VA +5 V DC <1 A +12 V DC <1.5 A -12 V DC <0.25 A Cary 50 Pre-Installation Manual 9

Work Area NOTE This table is indicative only. Refer to the literature provided with the computer and printer for details of individual power requirements. 10 Cary 50 Pre-Installation Manual

Equipment On Site 3. Equipment On Site Insurance 11 In-house transit routes 11 Inspecting for transit damage 12 Unpacking 13 Insurance As the carrier s liability ceases when the equipment is delivered, Agilent recommends that the instrument owner arranges separate insurance to cover transportation from the delivery point to the installation site. The delivery point will vary according to the carrier, the shipping method and in some cases the terms of sale. Some carriers will deliver only to their own distribution center while others may deliver to your off-loading bay. Very few carriers will deliver to the actual installation site. In-house transit routes In-house transit routes must be carefully considered. Vertical, horizontal and turning clearances should be calculated from the shipping carton dimensions of the spectrophotometer. Table 4 Shipping weights and dimensions Weight (kg, lb) Width (mm, in) Depth (mm, in) Height (mm, in) 23, 50.7 615, 24 710, 28 360, 14 Cary 50 Pre-Installation Manual 11

Equipment On Site Inspecting for transit damage Transit damage can be obvious or concealed and in either case will be admitted by the carrier only if it is reported within the terms of the carrier s agreement. For any claims against damage in transit, the following general rules apply: Before accepting delivery, you must inspect the packages for signs of obvious damage. The nature of any obvious damage must be noted on the carrier s waybill, which then must be countersigned by a representative of the carrier. Within the time limit stated in the terms and conditions of carriage, a further inspection must be made for concealed damage. If any damage is found at this stage, the carrier must be notified in writing and all packaging material must be retained for subsequent inspection by a representative of the carrier. A copy of any damage report must be forwarded to the Agilent sales office dealing with the supply of your equipment. WARNING Heavy Weight Hazard Many of the packages are large and heavy. To avoid the chance of injury to personnel or accidental damage to the equipment, always use two or more people when handling the packages or lifting equipment into position. NEVER attempt to lift the packages alone. 12 Cary 50 Pre-Installation Manual

Equipment On Site Unpacking After accepting delivery, take the equipment to the installation site. Agilent instruments are inherently robust, and the packaging is designed to prevent internal damage. However, the contents form part of a precision measuring system and all packages should be handled with care. In transit, sharp jolts must be avoided and the packages should not be inverted or tilted unnecessarily. Markings on the shipping cartons generally indicate which side of the package should be kept on top. Unpacking the equipment is your responsibility. As the packages are opened, ensure you received everything you ordered. If there are any discrepancies, notify the supplier. If any items are found to be damaged, immediately notify the carrier and supplier. Check the contents against the enclosed packing lists. Any differences from the original order should be referred immediately to your Agilent sales office. All contents of the shipping packages should be assembled together when installation is to be carried out by Agilent service personnel. Do not discard any packaging components or filler materials To unpack the Cary 50 spectrophotometer: 1 Open the packing box. 2 Lift the Cary 50 from the packaging as depicted in Figure 1. Cary 50 Pre-Installation Manual 13

Equipment On Site Figure 1. Lifting the Cary 50 from the packaging WARNING Heavy Weight Hazard The instrument weighs about 21 kg. To avoid injury to personnel, damage to the instrument, or damage to other equipment, ALWAYS use at least two people to lift it into position. Consult the manuals supplied with your computer, monitor and printer/plotter for their unpacking instructions. 14 Cary 50 Pre-Installation Manual

Computer System Requirements 4. Computer System Requirements Minimum and recommended configurations 15 Setting up your computer 17 Instrument connection 18 Other connections 19 The computer must be certified to at least one of the following standards to ensure that the integrity of the electrical safety of the Cary 50 is not compromised: UL 1950 CSA C22.2 No 950-95 IEC 950 or IEC 60950 EN 60950 WARNING Electrical Shock Hazard To ensure safe operation of the instrument, the switching power supply in the computer must comply with standard IEC 60950. Minimum and recommended configurations The minimum configuration (see Table 5) represents the absolute minimum on which you can run the software. This computer configuration may be out of manufacture, but you may want to use a computer you already have. The recommended configuration is that which you would buy new. Agilent can supply a computer for the Cary 50 instrument in the recommended configuration (see Table 5) as part number 7910026300. The operating system will be pre-loaded. All software disks and manuals will be supplied. Cary 50 Pre-Installation Manual 15

Computer System Requirements NOTE Computers supplied with Letter of Credit orders will be an international brand and will be the recommended configuration or better. Better computer components can be substituted for those listed. For example, processor type, amount of memory, screen size and resolution. Table 5. Minimum and recommended computer requirements Minimum IBM compatible Intel Pentium II processor Recommended IBM compatible 16 MB RAM 64 MB RAM Intel Pentium II processor 150 MB free space on hard disk 500 MB free space on hard disk Video card supporting 800 x 600 resolution, high color (16 bit) mode Super VGA screen Video card supporting 800 x 600 resolution, high color (16 bit) mode Super VGA screen 4 x CD-ROM drive 10 x CD-ROM drive 16 bit sound card 16 bit sound card Windows 95 101 key keyboard Microsoft or compatible mouse One spare AT bus 16 bit ISA expansion slot Microsoft Windows 95, 98 or Windows NT (including Service Pack 4 or later) Microsoft Internet Explorer 4 or later if using Windows 95 of Windows NT Windows 95 101 key keyboard Microsoft or compatible mouse One spare AT bus 16 bit ISA expansion slot Microsoft Windows 95, 98 or Windows NT (including Service Pack 4 or later) Microsoft Internet Explorer 4 or later if using Windows 95 of Windows NT 1 spare computer power supply connector 1 spare computer power supply connector Power supply rating 220 W Power supply rating 220 W 1 spare slot for the accessory cable connector 1 spare slow for the accessory cable connector 16 Cary 50 Pre-Installation Manual

Computer System Requirements Setting up your computer The Cary WinUV software requires Microsoft Windows 95, Windows 98 or Windows NT to be installed on your computer. For instructions on installing this, refer to the documentation supplied with the operating system. NOTE If Agilent is installing the Cary 50 for you, installation of a Windows operating system or Internet Explorer 4 is not included as part of the standard instrument installation. Ensure that you have your display desktop area resolution set to 800 by 600 pixels and that the color palette is set to High Color (16 bit). Check that your sound card and CD-ROM drive are working. You also need to install Microsoft Internet Explorer 4 or later. This does NOT mean that you need to use Internet Explorer 4 as your Web browser the Cary WinUV software uses some of the components that are supplied as part of Internet Explorer 4, but you can continue to use whichever browser you like. You will probably need to purchase Internet Explorer 4 or download it for free from the Microsoft Web site (www.microsoft.com), as typically only earlier versions of Internet Explorer are shipped with Windows. Cary 50 Pre-Installation Manual 17

Computer System Requirements Instrument connection The spectrophotometer is connected to the computer via a shielded interconnection cable provided with the instrument. Connection to the Cary 50 Figure 2. Spectrophotometer to computer connections Connection to the PC Cary 50 to PC cable PC mains connection Cary 50 instrument Figure 3. Cary 50 system interconnection The Cary 50 computer card requires reserved access to: Address space 210-21F Interrupt requests IRQ 3,5,7,10,11,12 (one off) This computer card plugs into a computer I/O connector and it controls the instrument functions. It drives the two motors, lamp module and two detectors. Signals are converted into digital form to be processed by the computer. 18 Cary 50 Pre-Installation Manual

Computer System Requirements Connection of the Cary 50 to the computer card is by a 37-way D-range connector. This connector carries motor drive supplies for the Cary 50 instrument and motor-driven accessories, detector analog signals and digital inputs and outputs. The following supplies are used by the Cary 50. They are taken directly from the I/O bus connector: -12 V DC <0.25 A (maximum) +5 V DC <1 A (maximum) The following power is supplied directly from the computer power supply (standard HDD connector +12 V DC only): +12 V DC <1.5 A (maximum) Typical power consumption (measured) is as follows: When idle (with motors powered) 400 ma Operating (motors and lamp module powered) 1.125 A Power dissipation (typical) instrument only 26 W Operation of motor-driven accessories may increase the +12 volt current by a further 2 amperes (24 watts maximum). Note that these requirements are in addition to that required by the computer itself. Other connections The monitor, keyboard, mouse and printer/plotter are connected to the computer via cables that plug into the back of the computer. Consult your monitor and printer manuals for details of their individual cabling requirements. The sample compartment contains these connections: 3.5 mm phono jack socket in the left wall for accessories. 8-pin DIN connector on the left side for the diode detector. 25-pin D-range connector on the right side for accessories (optional). Cary 50 Pre-Installation Manual 19

Computer System Requirements This page is intentionally left blank. 20 Cary 50 Pre-Installation Manual