Steps for LCMS-8040 Triple Quadrupole

Similar documents
Agilent 6400 Series Triple Quadrupole LC/MS System

3 Starting the chiral HPLC and logging into LC Real Time Analysis.

Agilent MassHunter Workstation Software 7200 Accurate-Mass Quadrupole Time of Flight GC/MS

Getting low-resolution GC-MS data on the JEOL GCMate Introduction

Agilent GC/MSD Instructions

How to Run the CASPiE GCMS

Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017

Operating Manual. High Performance Liquid Chromatograph. Scientific Equipment Center, Prince Of Songkla University

HITACHI LACHROM ELITE

Chromeleon / MSQ Plus Operator s Guide Document No Revision 03 October 2009

Manual HPLC-4 Alliance-UV

Shimadzu IMD-MS news For Sales Forces of Shimadzu LCMS No. MSnews42 September, 2010

Sciex QTrap Operational Steps for Trained Personnel

Q Exactive HF-X QuickStart Guide

ChromQuest 5.0 Quick Reference Guide

Agilent 6100 Series Quad LC/MS System

TraceFinder Analysis Quick Reference Guide

Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS

Thermo. EASY-nLC User Guide for LC Devices

Customers should provide all necessary operating supplies upon request of the engineer.

Automated Submission for Analysis Processing (ASAP) Operator s Guide Dionex Corporation

MassLynx 4.1 SCN 804. Release Notes. SQ Detector 2

EASY-nLC Virtual Instrument for Xcalibur 2.1.x Quick Reference Guide

ANALYTICAL SYSTEMS. L3320 Autosampler with excellent accuracy and linearity. System configuration examples

Agilent G1978B Multimode Source for 6410 Triple Quad LC/MS

Agilent Triple Quadrupole LC/MS Peptide Quantitation with Skyline

Aqueous GPC Instructions - Detailed

Release Notes. Analyst TF Software

EASY-nLC Series User Guide for the Xcalibur Data System (Xcalibur version 2.1 or later)

Xcalibur. QuickQuan. User Guide. XCALI Revision C July For Research Use Only Not for use in Diagnostic Procedures

Agilent Instrument Control Framework A for Agilent LC/CE and GC Instruments in Non-Agilent Chromatography Data Systems Release Notes

For additional information, see these documents on the Breeze 2 software DVD:

Agilent G1732AA MSD Security ChemStation

Agilent 6300 Ion Trap LC/MS Systems Quick Start Guide

Short instructions. Olympus FV1000

Agilent MassHunter Qualitative Data Analysis

Agilent G3250AA LC/MSD TOF System

Agilent G1978A Multimode Source for 6300 Series Ion Trap LC/MS

PerkinElmer Series 200 HPLC: Operating Procedure

PANalytical Epsilon 5 XRF spectral acquisition, processing and reporting

Data Acquisition with 3400/3600 GC Control

Validation Sample Set Generator

MassLynx 4.2 SCN 989 Release Notes

2-1. the ESI Probe Assembly. the PEEK Safety Sleeve and Sample Tube to the ESI Probe

GC-MS ChemStation Reference Guide. Last Updated July 2014

Release Notes. Analyst Software

Analyst QS. Administrator s Guide. Part Number: B July 2004

Getting Started. Environmental Analysis Software

Netzsch TGA Training Notebook

Using the Chromeleon Summit 2 Tandem Operation Wizard

irc Pro (Not for diagnostic use)

The Agilent 5975 inert MSD: New Tools for the Forensic Analyst

Agilent MassHunter GCMS Data Acquisition for the 5977 Series MSD System

Agilent G2727AA LC/MS Data Browser Quick Start

Agilent MassHunter Easy Access Software

Quick Start Guide. Introduction. Importing Raw Data. Method 1: Use Open Data Panel

Agilent G1978A Multimode Source for G1946/G1956 LC/MSD

ACQUITY UPLC Systems with 2D Technology Capabilities Guide

PLATINblue MSQ Plus. Connection Guide V6951A UHPLC/HPLC. Mass spectrometry

DESCRIPTION. DataApex Clarity 32bit multi instrument chromatography station for Windows

Detailed procedures, operation instructions, maintenance, and emergency contact information list is attached.

SMMS i Selection Tool

Isolera. User Manual

Jasco 4700 FTIR. Laser Spectroscopy Labs. Jasco 4700 FTIR. Operation instructions. Laser Spectroscopy Labs, UCI

NOTICE: This document contains references to Varian. Please note that Varian, Inc. is now part of Agilent Technologies. For more information, go to

Dionex SRD-10 Suppressor Regenerant Detector Installation Instructions

TraceFinder Analysis Quick Reference Guide

Tutorials. Saturn GC/MS Workstation Version 5.4. Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA /usa

Thermo Xcalibur Getting Started (Quantitative Analysis)

Release Note for Agilent LC and CE drivers Revision A.02.14

Agilent G1701DA MSD ChemStation

ZEPHYR Compact Haze Machine

Agilent 7890 Series Gas Chromatograph

BioTOF. Series. Getting Started

Chromatography Software Training Materials. Contents

Equipment Overview: Safety:

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes. Agilent 6230B Series TOF LC-MS System Installation Checklist

Agilent G6854AA MassHunter Personal Pesticide Database Kit Quick Start Guide

Agilent G6855AA MassHunter Personal

Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start

Agilent G1978A Multimode for 6210/6220 TOF

TraceFinder Acquisition Quick Reference Guide

TraceFinder Acquisition Quick Reference Guide

AGILENT 1100/1200. Clarity Control Module. Code/Rev.: M060/70D Date: 10/5/2017

Configuring and optimizing an IC-MS system using a compact IC and a single quadrupole mass spectrometer

QUANTAX EDS SYSTEM SOP

Operating Manual. GC-APPI Inferface for:

MassHunter File Reader

SIM Direct Inlet Probe. Agilent 5975 MSD. Agilent 7000 Triple Quadrupole

Release Note for Agilent LC and CE Drivers Revision A.02.17

Chromeleon software orientation

Appendix C. Vernier Tutorial

Hardware Guide. AB SCIEX QTRAP 5500 System

Data Acquisition with 3800 GC Control

FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up

Agilent G2070BA GC ChemStation

PurityChrom Bio 5 Software Manual V2650

Agilent GC, MS, and ALS

Soprane II Documentation

Q Exactive HF-X Software Manual

Transcription:

Steps for LCMS-8040 Triple Quadrupole June 2017 Version 2 A. Data Acquisition 1. Check lights on MS. There should be a total of three lights lit by LED. One of them is blinking. Power Blinking Gas Status 2. Check level of solvents A and B on top of instrument. Be sure there is sufficient solvent volume for your run. Bottle A is aqueous and Bottle B is organic. 3. Check the waste can. 4. If changing solvents: a. Hit Purge button on Autosampler. This is set to purge for 25 minutes. b. Turn black knob on LC-20AD Pump A 90 counterclockwise towards open. Press purge on LC-20AD Pump A. 90 counterclockwise 1

c. Turn black knob on LC-20AD Pump B 90 counterclockwise towards open. Press purge on LC-20AD Pump B. d. Turn black knob back to closed on LC-20AD Pump A 90 clockwise when purge is complete (5 minutes). Turn clockwise until you can t turn anymore. The A sticker should be upright. e. Turn black knob back to closed on LC-20AD Pump B 90 clockwise when purge is complete (5 minutes). The B sticker should be upright. 5. Double click on Lab Solutions icon on computer Desktop. 6. Logon as Admin. No password needed. Just hit OK. 7. The Lab solutions windows will open. Click on the Instruments icon 8. Select and Double-click on the LCMS icon 1. Click on Instruments 2. Click on LCMS 2

9. Click on Main Tab on far left. Main 10. Click the Data Acquisition icon in the Main assistant toolbar. Data Acquisition 11. Under File, open Method File. Or double-click on your method to load. 12. Click on Download when complete.. This will start Mass Spect. It will not start heating column, if needed, or flow rate. a. Oven Temperature. Want slightly above room temperature. Usually set at 40 C. Hit icon to start warming column. b. Pump ON/Off icon on the top toolbar. Pump Solvent delivery is started. 3

B. Create or Edit a Method 1. If you need to create a New Method file, start with MS Tab. 1) Set the MS Instrument Parameters. Click on MS tab. Each individual MS condition is called an event and polarity can be set for each event. The system can be scanned in positive and negative modes simultaneously. This can be accomplished by setting events. Each event has its own parameters MS tab a) Select polarity of the event to set: Positive or Negative. b) Click the button for the data s acquisition mode; MRM(±), Product Ion Scan(±), Precursor Ion Scan (±), or Neutral Loss Scan (±), SIM (±) or scan (±). c) The event is now added to the Event Table. 4

] b.) If choosing SIM, be sure to put in M/Z and change dwell time to 20msec. 5 Add M/Z and change Dwell time to 20msec for SIM.

c) Set Same measurement time for each event by right clicking and choosing Set Same Measurement time followed by OK. Set Same Measurement Time d.) Type OK 6

2. Set Interface parameters. a. ESI and/or APCI mode. b. Nebulizing Gas flow 2 L/min. (3 L/min for higher flow rates) c. DL temp =250 (300 for higher flow rates) d. Heat Block temp =400C. e. Check off drying gas flow 15 L/min. 3. Set Data Acquisition parameters. A. Enter the LC Stop Time as time from one sample to the next. B. Click Apply to All acquisition time to set end time for each detector to the same time as the LC detector. Apply to All Acquisition Time 7

4. Under Pump, keep as binary gradient, even if running isocratic. a. Set Pump Flow rate and initial concentration for gradient systems, Pump B conc (never 100% H2O for C18 RP column). b. Set Pressure Max for your column ( i.e 6000)(Look up on internet). 5. Set Oven Temperature slightly above room temperature. Usually set at 40 C. Hit icon to start warming column 6. Click on the LC Time Prog. tab to enter gradient conditions. A. Set Time, Module, Command, and Value. 8

B. Click on Draw Curve to visualize your method. C. Click on the MS Instrument On/Off icon on the top toolbar to start DL heater and heat block temperature control. 7. Click on Download when complete. Download 8. Click on the Pump ON/Off icon on the top toolbar. Pump Solvent delivery is started. 9

9. Under Autosampler, change Purge Time to 10 min. Change Purge time to 10.00 min 10. Under File, choose Save Method File As.. a) Enter the File name and click Save. 10

C. Click the Realtime Batch icon 1. A realtime batch table will open. 2. Under File, choose New Batch File. a. Enter Vial #, b. Tray name =1 c. Sample Name d. Sample ID e. Sample Type-unknown or standard f. Method File set up previously g. For Data File write unique name of sample(can t use name already assigned to previous sample) h. Injection volume from 0.1ul to 100µl. 3. Right click on screen and choose Settings or click on Settings icon 4. Go to Shutdown tab under Settings. Enter conditions for automatic shutdown with a cooling time of longer than the running time. Be sure all instruments are checked off to shutdown after run. Shutdown is checked Cool down time longer than run time All instruments checked off 11

5. Under File, choose Save Batch method as... 6. Select the rows and numbers to be analyzed. 7. Click on Start Realtime Batch icon. 8. 35. Select All Rows or Selected Rows. 9. 36. Hit Stop button if you need to stop the run. 12

D. Data Analysis 1. Click on the Postrun icon 2. Double-click on Postrun.(Browser is used to look at more than one sample at a time) 13

3. Drag and drop a file into the window. Or Double-click on file. 4. Average MS Spectrum- a. Click on Average Spectrum on toolbar. A stable spectrum can be displayed by totaling and averaging the spectra within a certain time range. 14

5. Subtract MS Spectrum a. Click Subtract Spectrum on MS Data Analysis toolbar. A cleaner-looking spectrum can be displayed by subtracting the background MS spectrum from the averaged spectrum. 15

6. Add spectra to Spectrum Process Table a. The spectrum displayed can be entered in the Spectrum Process Table. b. Click on Register to Spectrum Process table in MS Analysis Toolbar. c. Click OK. d. The Spectrum is registered to the Spectrum Processed Table. 16

E. Report 1. Save screen as seen a. Under File, choose Print Graph Image, then PDF Output. b. Save on USB drive. F. Shutdown 1. Close Lab Solutions by going under File and choosing Exit. 17

2. A shutdown sub-window opens. Be sure everything is checked off. 3. Under File, choose Exit Lab Solutions. 18

19