Calibrating FieldPoint Analog Input Modules with Calibration Executive

Similar documents
NI 9213 Calibration Procedure

FieldPoint. FP-1000/1001 FieldPoint Quick Start Guide. Install the Network Module Mount the module on a DIN rail. What You Need to Get Set Up

INSTALLATION INSTRUCTIONS LFH160 Cable

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9208E 16-Channel, ±20 ma, 24-Bit Analog Input Module

NI CB-37F-LP. Introduction INSTALLATION GUIDE. Connector Block for 37-Pin D-Sub Devices

LabVIEW DSP Module. Contents. Introduction RELEASE NOTES. Version 1.0

NI 653x Cable Adapter

NI SMB-2145/2146/2147/2148

OPERATING INSTRUCTIONS AND SPECIFICATIONS NI 9871E

VME-GPIB Installation and User Guide

You need the following to set up and use your terminal block:

FP-PS-4. Features. Installation

OPERATING INSTRUCTIONS AND SPECIFICATIONS NI 9870E

QUICK START GUIDE. What You Need to Get Set Up

Contents. Software. CALIBRATION PROCEDURE NI 9216/ Channel, 24-Bit, 100 Ω/1000 Ω RTD Analog Input Module

NI 6521 Specifications

FP-AO-200 AND CFP-AO-200

FP-AO-210 AND CFP-AO-210

FP-RTD-122. Introduction CALIBRATION PROCEDURE. Version 2.0

Getting Started with R Series Intelligent DAQ

CALIBRATION PROCEDURE NI PXIe Channel, Isolated Voltage/Current Analog Output Module

Contents. Software Requirements. CALIBRATION PROCEDURE NI PXIe-4357

Getting Started with R Series Multifunction RIO

LabVIEW DSP Module Release and Upgrade Notes

FP-2000/2010 FieldPoint Quick Start Guide

CALIBRATION PROCEDURE CAL Contents. Software

CALIBRATION PROCEDURE NI channel, ±2- ma/±10 V, 24-bit Analog Input Module. ni.com/manuals

NI PCIe-784xR, NI PXI-784xR, NI PXIe-784xR, NI USB-784xR, NI PCIe-785xR, NI PXI-785xR, NI PXIe-785xR, NI USB-785xR, and NI PXIe-786xR

Sample. LabWindows TM /CVI TM Core 1 Exercises Manual

cfp-rly-421 Features Installing the cfp-rly-421 FieldPoint Operating Instructions Eight-Channel SPST Relay Module

Using Measurement & Automation Explorer (MAX) with VXI

Getting Results with the NI PXI-6585R

NI R Series. Contents. Software CALIBRATION PROCEDURE

CALIBRATION PROCEDURE PXIe channel, 16-bit, 400 ks/s/ch, Ch-Ch Isolated Analog Input Module. ni.com/manuals

FP-1600 FieldPoint Quick Start Guide

TBX-96 TERMINAL BLOCK

Caution Electrostatic discharge can damage your controller. To avoid such damage, handle the controller only in a proper ESD-controlled environment.

Sample. LabWindows TM /CVI TM Core 1 Course Manual

Contents. Software. CALIBRATION PROCEDURE NI Channel, ±5 V, 24-Bit Software-Selectable IEPE and AC/DC Analog Input Module

NI SPEEDY-33. '7 NATIONAL INSTALLATIONGUIDE. P~ClCl J ~ Ig~~!1 "INSTRUMENTSN. 1~~~~~~g~~lm.OOOQ{JOOO,

Power Analyzer Firmware Update Utility Version Software Release Notes

Getting Results with the NI 5752R

FieldPoint cfp-20xx. Quick Start Guide. What You Need to Get Set Up

NI SD Card Memory Module

Getting Started with NI-Motion for NI 73xx Motion Controllers

NI USB Introduction USER GUIDE AND SPECIFICATIONS. 4-Channel, ±10 V, 16-Bit Analog Voltage Output Module

Contents. Software. CALIBRATION PROCEDURE NI Channel, 300 Vrms, 24-Bit, Simultaneous, Channel-to-Channel Isolated Analog Input Module

Sample. LabVIEW TM Core 1 Exercises. Course Software Version 2010 August 2010 Edition Part Number B-01. LabVIEW Core 1 Exercises

NI CardBus-8310-to-PXI Expansion Kit

NI Channel, 0 20 ma, 16-Bit Analog Current Output Module

USER GUIDE Universal Motion Interface (UMI)-7764

Upgrading Luminex IS 2.3 to Bio-Plex Manager 6.1 Software. For technical support, call your local Bio-Rad office, or in the US, call

CALIBRATION PROCEDURE SCXI Contents. Software Requirements. ni.com/manuals

±10 V, 20 MS/s/ch, 14-Bit, 4-Channel C Series Digitizer Module

VI Logger. Getting Started with VI Logger. Getting Started with VI Logger. May 2003 Edition Part Number B-01

Contents. Software Requirements CALIBRATION PROCEDURE NI TB-4353

NI CAN Demo Box. NI CAN Demo Box User Guide. Power supply (+9 VDC to +12 VDC) 68-pin Multifunction Data Acquisition board

Compact FieldPoint cfp-cal

FP-AIO-600 and cfp-aio-600

NI PXI Contents CALIBRATION PROCEDURE

SCXI HIGH-VOLTAGE 8 4 MATRIX TERMINAL BLOCK

Instruction Manual. VacTest Accessory. Interface Converter (Bluetooth Module) RS485 - Bluetooth. Vacuum Measurement Equipment

NI 6584R User Guide and Specifications

Dataliner Message Displays Using DL50 Slaves with a DL40 Master

SCXI -1102/B/C. Contents CALIBRATION PROCEDURE. For NI-DAQmx

Getting Started with the NI PCI-1405

Getting Started with the NI 8234

NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

SCXI -1104/C. Contents CALIBRATION PROCEDURE. For NI-DAQmx

4 GHz Active Probe P7240 Datasheet

NI USB Introduction USER GUIDE AND SPECIFICATIONS. 4-Channel, 0 20 ma, 16-Bit Analog Current Output Module

Differential Probes TDP1500 TDP3500 Data Sheet

NI 446X Calibration Procedure

Safety. Introduction

Getting Started with the NI PCIe-1433

SCXI Contents CALIBRATION PROCEDURE

AC/DC (120V) Input Module Cat. No IAD Series C

Contents. Software CALIBRATION PROCEDURE NI TC-4353

Sample. LabWindows TM /CVI TM Basics I Course Manual

4 GHz Active Probe P7240 Datasheet

CB-27T, TBX-68T, AND CB-68T ISOTHERMAL TERMINAL BLOCKS INSTALLATION GUIDE

The following conventions appear in this document:

For NI PXI and NI SCXI Switch Modules

NI USB Introduction USER GUIDE AND SPECIFICATIONS. 8-Channel, 24 V Logic, Sinking Digital Input Device

FieldPoint. Note to Users

NI 6703/6704 Calibration Procedure for NI-DAQ mx

USER GUIDE SC-2075 SIGNAL CONDITIONING ACCESSORY

RTPA2A. TekConnect probe adapter for real-time spectrum analyzers. Tektronix high-performance probing solutions. Applications. Notice to EU customers

Purchasing. Operations 3% Marketing 3% HR. Production 1%

Compact FieldPoint cfp-1808

ChemiDoc and ChemiDoc MP Imaging Systems with Image Lab Touch Software Installation Qualification Protocol (IQ)

Tektronix Logic Analyzer Probes P5900 Series Datasheet

END-OF-SALE AND END-OF-LIFE ANNOUNCEMENT FOR THE CISCO MEDIA CONVERGENCE SERVER 7845H-2400

Step 1: New Portal User User ID Created Using IdentityIQ (IIQ)

Troubleshooting Ethernet Problems with Your Oscilloscope APPLICATION NOTE

SCXI Equipment and Other Test Requirements CALIBRATION PROCEDURE. Test Equipment. Software and Documentation

Getting Started with the NI LabVIEW Embedded Module for ADI Blackfin Processors Version 8.6

NI USB Introduction USER GUIDE AND SPECIFICATIONS. 8-Channel, 24 V Logic, Sourcing Digital Output Device

Getting Started with the NI PCIe-8255R

NI PCIe-8255R Quick Start Guide

Transcription:

USER GUIDE Calibrating FieldPoint Analog Input Modules with Calibration Executive Contents This document contains step-by-step instructions for loading and running National Instruments Calibration Executive to calibrate FieldPoint analog input (AI) modules, including the FP-AI-100, FP-AI-110, FP-AI-111, FP-RTD-122, and FP-TC-120. Equipment and Other Test Requirements... 2 Test Equipment... 2 Documentation... 3 Software... 3 Test Conditions... 3 Calibration Procedures... 4 Configuring the FieldPoint AI Module for Calibration... 5 Connecting the Calibrator to an FP-TC-120 Module... 10 Connecting the Calibrator to an FP-AI-1xx Module... 11 Using Precision Resistors to Calibrate an FP-RTD-122 Module... 12 Running the Calibration Executive Procedure... 13 Troubleshooting and Common Questions... 14 Technical Support Resources... 14 NI Web Support... 14 Worldwide Support... 15

Equipment and Other Test Requirements This section describes the equipment, documentation, software, and test conditions you need to calibrate a FieldPoint module using Calibration Executive. Test Equipment Calibration requires using a high-precision calibrator with the voltage accuracy in parts per million (ppm) listed in Table 1. NI recommends that you use the following instruments for calibration: Table 1. Test Equipment Specifications and Recommended Models FieldPoint Module Required Accuracy Recommended Instrument FP-AI-100 500 ppm Fluke 5500A FP-AI-110 100 ppm Fluke 5500A FP-AI-111 100 ppm Fluke 5500A FP-RTD-122 6.5 digits NI-4070 FP-TC-120 30 ppm Fluke 5520A FieldPoint FP-1000 serial network module FP-TB-CAL FieldPoint Calibration Terminal Base (part number 777519-90), a modified FieldPoint base that enables you to write new calibration constants to the module EEPROM 11 24 V power supply NI PS-2 power supply RS-232 serial cable If the recommended instrument is not available, use the accuracy specifications to choose a substitute calibration standard. Calibrating FieldPoint Analog Input Modules 2 ni.com

Documentation This section describes the documentation you need to calibrate a FieldPoint AI module. In addition to this calibration procedure, you may need to refer to the following documentation: Calibration Executive Software User Manual FP-1000/1001 User Manual FP-1000/1001 Quick Start Guide Operating instructions for the AI module You can download these documents from the NI Web site at ni.com/ manuals. Software The Calibration Executive setup program installs Calibration Executive, the FieldPoint software, and the associated files listed in Table 2. Refer to the Calibration Executive Software User Manual for more information about installing and uninstalling Calibration Executive software and procedures. Table 2. Calibration Executive Directories and Supporting Files Directory Name Calibration Executive\Procedures\ FP-XX-XXX\Limits Calibration Executive\Procedures\ FP-XX-XXX Calibration Executive\Procedures\ FP-XX-XXX\Support Files Contents Microsoft Access database that stores the calibration limits for FieldPoint AI modules (excluding FP-TC-120) Directory structure that contains the calibration procedures Required support files Test Conditions Follow these guidelines to optimize connections and test conditions during calibration: Keep connections to the module short. Long cables and wires act as antennae, picking up extra noise that can affect measurements. Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets. Maintain a temperature between 18 and 28 C. If you intend to operate the module at a specific temperature outside this range, calibrate the module at that temperature. National Instruments Corporation 3 Calibrating FieldPoint Analog Input Modules

Calibration Procedures Keep relative humidity below 80%. Allow a warm-up time of at least 15 minutes to ensure that the measurement circuitry is at a stable operating temperature. This section explains how to configure and set up the module for calibration, and how to run the Calibration Executive procedure. Refer to Table 3 for the approximate times to complete the procedure in both automated and manual modes. Table 3. Approximate Time to Complete Calibration Executive Procedure Module Automated Mode Manual Mode FP-TC-120 20 minutes 45 minutes FP-AI-100 120 minutes 180 minutes FP-AI-110 45 minutes 90 minutes FP-AI-111 60 minutes 90 minutes FP-RTD-122 10 minutes 10 minutes Note For an explanation of automated versus manual calibration, refer to the Automated Versus Manual Calibration section of the Calibration Executive Software User Manual. Calibrating FieldPoint Analog Input Modules 4 ni.com

Configuring the FieldPoint AI Module for Calibration Complete the following steps to build the FieldPoint bank: 1. Attach the network module to the FP-TB-CAL as shown in Figure 1. 1 2 3 1 Local Bus Connectors Firmly Mated 2 Protective Cover 3 Baud Rate Switches Figure 1. Connecting FieldPoint Terminal Bases 2. Install the FieldPoint AI module on the calibration terminal base as described in the module operating instructions. 3. Connect the bank to the host computer using an RS-232 cable. 4. Use the DIP switches on the network interface module to select a baud rate that your computer supports. Table 4 shows the available baud rates. The default baud rate for the FP-1000 is 115,200, which provides the fastest performance. You do not need to change this setting unless you experience communication problems. National Instruments Corporation 5 Calibrating FieldPoint Analog Input Modules

Table 4. Baud Rate Switch Settings for the FP-1000/1001 Switch Positions 6 8 Network Module Baud Rate 300 1,200 2,400 9,600 19,200 38,400 57,600 115,200 5. Connect the positive wire of the power supply to the first V terminal of the network module. Connect the negative wire of the power supply to the first C terminal of the FP-1000 network module as shown in Figure 2. Calibrating FieldPoint Analog Input Modules 6 ni.com

To 11 24 VDC Power Supply + CC V V Figure 2. FP-1000/1001 Power Connector Pinout 6. Launch Measurement & Automation Explorer (MAX). 7. In the MAX configuration tree, right-click Devices and Interfaces and select Create New. National Instruments Corporation 7 Calibrating FieldPoint Analog Input Modules

8. Select FieldPoint Serial, then click Next. 9. Change the serial configuration if you need to, then click Finish. Calibrating FieldPoint Analog Input Modules 8 ni.com

10. Under My System»Data Neighborhood, expand the configuration tree until the I/O module you want to calibrate appears. If you do not see the module, the network module is not communicating properly with either the host or the I/O module. Check the connections and try again. 11. Close MAX before you go on to the next section. You do not need to save the.iak file. Caution You must close MAX and keep it closed during the remainder of this procedure to prevent a system crash. National Instruments Corporation 9 Calibrating FieldPoint Analog Input Modules

Connecting the Calibrator to an FP-TC-120 Module If you are calibrating the FP-TC-120, make the following connections as shown in Figure 3: 1. Connect the positive output (voltage) of the calibrator to terminal 1 of the calibration terminal base. 2. Connect the negative output (voltage) of the calibrator to terminal 2 of the calibration terminal base. 1 7 V V C C V 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 V C 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 C 2 6 5 3 4 1 FP-1000 Network Module 2 TB-Cal Calibration Terminal Base 3 To Positive Output (Voltage) of Calibrator 4 To Negative Output (Voltage) of Calibrator 5 To 11 24 VDC Power Supply 6 RS-232 Serial Cable to COM Port 7 Baud Rate Switches Figure 3. FP-TC-120 Connections to Calibrator Calibrating FieldPoint Analog Input Modules 10 ni.com

Connecting the Calibrator to an FP-AI-1xx Module If you are calibrating an FP-AI-100, FP-AI-110, or FP-AI-111 module, make the connections shown in Figure 4. 1. Connect the positive output (current) of the calibrator to terminal 2 of the calibration terminal base. 2. Connect the negative output (current) of the calibrator to terminal 18 of the calibration terminal base. While the Calibration Executive procedure runs, you are prompted to make additional connections between the calibrator and the module you are calibrating. Follow the onscreen instructions at each prompt. 1 7 V V C C V 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 V C 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 C 2 6 5 3 4 1 FP-1000 Network Module 2 FP-TB-CAL Calibration Terminal Base 3 To Positive Output (Current) of Calibrator 4 Negative Output (Current) of Calibrator 5 To 11 24 VDC Power Supply 6 RS-232 Serial Cable to COM Port 7 Baud Rate Switches Figure 4. FP-AI-1xx Connections to Calibrator for Current Offset Testing National Instruments Corporation 11 Calibrating FieldPoint Analog Input Modules

Using Precision Resistors to Calibrate an FP-RTD-122 Module Calibrating the FP-RTD-122 requires four precision resistors with less than 25 ppm/ C temperature drift. The resistors must be of the following values: 100 Ω, 350 Ω, 1 kω, and 3.5 kω. Use the DMM to determine the real value of each reference resistor. Take four-wire measurements, connecting the DMM to each resistor as shown in Figure 5. HI SENSE HI DMM SENSE LO LO Figure 5. Connecting the DMM to a Resistor When you run the calibration procedure, Calibration Executive prompts you to connect each resistor to channel 0 of the FP-RTD-122 and enter the actual resistance value of each resistor. Connect each resistor as shown in Figure 6. EXCITE FP-RTD-122 SENSE COM Figure 6. Connecting the FP-RTD-122 to a Resistor Calibrating FieldPoint Analog Input Modules 12 ni.com

Running the Calibration Executive Procedure To run the Calibration Executive procedure, complete the following steps: 1. Launch Calibration Executive and follow the instructions in the Calibration Configuration Wizard to load the appropriate FieldPoint calibration procedure for the module. Refer to the Calibration Executive Software User Manual if you need more information on configuring and loading a calibration procedure. 2. Enter the following information about the installed hardware when prompted by Calibration Executive: Selected Calibrator Information The calibrator you are using for this calibration procedure (for example, the Fluke 5700A). Available Address Select None if you are using the manual mode; otherwise, select the GPIB address of the calibrator. 3. When the procedure is loaded, click Run Procedure to begin. 4. Enter the following information in the first dialog box, which is titled FieldPoint Bank Configuration: Baud Rate The correct rate from Table 4 to communicate with the network module. COM Port Number The serial port used to connect to the FP-1000 network module, for example, COM1, COM2, and so on. Device Number The address of the I/O module you want to verify or calibrate. The first module to the right of a network module has an address of 1. If you enter a value incorrectly, Calibration Executive prompts you to re-enter the information. 5. Follow any instructions you receive from Calibration Executive. After the procedure finishes, complete the following steps to see the calibration report: 1. Click View»Reports. 2. Select View to launch your browser and view the report. The calibration report appears as a printable HTML file. You have completed calibrating your FieldPoint AI module with Calibration Executive. Note If the FieldPoint AI module fails after calibration, return it to NI for repair or replacement. National Instruments Corporation 13 Calibrating FieldPoint Analog Input Modules

Troubleshooting and Common Questions When I change the baud rate of the FP-1000 and run the Calibration Executive procedure, a dialog box prompts me to re-enter the correct baud rate. How do I do that? Power down the FP-1000, then power it up again to enable the new settings that you entered. Some test points fail even after I run the Calibration Executive procedure. What can I do to correct this problem? Before running the Calibration Executive procedure, allow the FieldPoint bank to warm up at least 15 minutes to ensure accurate readings. Also, make sure you are using shielded copper wire for all cable connections to the module. In the FP-AI-100 current offset test, my test points are significantly out of specification. What causes this problem? The voltage terminals (1, 3, 5, 7, 9, 11, 13, and 15) on the calibration base may still be connected together from the voltage offset test. Disconnect all the wires that connect these terminals before attempting the current offset test. Communication with the FieldPoint AI module times out. What causes this to happen? The watchdog timer on the FP-1000 may be enabled. Refer to the FP-1000/1001 User Manual to find out how to disable the watchdog timer. Technical Support Resources NI Web Support National Instruments Web support is your first stop for help in solving installation, configuration, and application problems and questions. Online problem-solving and diagnostic resources include frequently asked questions, knowledge bases, product-specific troubleshooting wizards, manuals, drivers, software updates, and more. Web support is available through the Technical Support section of ni.com. Calibrating FieldPoint Analog Input Modules 14 ni.com

Worldwide Support National Instruments corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. National Instruments also has offices located around the world to help address your support needs. You can access our branch office Web sites from the Worldwide Offices section of ni.com. Branch office Web sites provide up-to-date contact information, support phone numbers, email addresses, and current events. If you have searched the technical support resources on our Web site and still cannot find the answers you need, contact your local office or National Instruments corporate. For telephone support in the United States, dial 512 795 8248. For telephone support outside the United States, contact your local branch office: Australia 1800 300 800, Austria 43 0 662 45 79 90 0, Belgium32027570020, Brazil551132623599, Canada (Calgary) 403 274 9391, Canada (Ottawa) 613 233 5949, Canada (Québec) 450 510 3055, Canada (Toronto) 905 785 0085, Canada (Vancouver) 604 685 7530, China 86 21 6555 7838, Czech Republic 420 224 235 774, Denmark 45 45 76 26 00, Finland3850972572511, France330148142424, Germany 49 0 89 741 31 30, India 91 80 51190000, Israel 972 0 3 6393737, Italy 39 02 413091, Japan 81 3 5472 2970, Korea 82 02 3451 3400, Malaysia 603 9131 0918, Mexico 01 800 010 0793, Netherlands 31 0 348 433 466, NewZealand0800553322, Norway47066907660, Poland 48 22 3390150, Portugal 351 210 311 210, Russia 7 095 783 68 51, Singapore 65 6226 5886, Slovenia 386 3 425 4200, South Africa 27 0 11 805 8197, Spain 34 91 640 0085, Sweden 46 0 8 587 895 00, Switzerland 41 56 200 51 51, Taiwan 886 2 2528 7227, Thailand 662 992 7519, United Kingdom 44 0 1635 523545 FieldPoint, National Instruments, NI, and ni.com are trademarks of National Instruments Corporation. Product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your CD, or ni.com/patents. 2003 2004 National Instruments Corp. All rights reserved. *370962B-01* 370962B-01 Sep04