Spectramag-6 Six-Channel Spectrum Analyser for Magnetic Field and Vibration Surveys. For innovation in magnetic measuring instruments

Similar documents
Data Acquisition & Signal Conditioning Units

Innovation in Magnetic Measuring Instruments. Data Acquisition & Conditioning Units

INNOVATION IN MAGNETICS DS0013/27. Mag-03. Three-Axis Magnetic Field Sensors.

INNOVATION IN MAGNETICS DS-2201/8 DAS1. Data Acquisition System.

Innovation in Magnetic Field Measuring Instruments. Mag648 & Mag649. Low Power Three-Axis Magnetic Field Sensors.

Mag-03. Three-axis Magnetic Field Sensors. For innovation in magnetic measuring instruments

HCS1 Three-Axis Helmholtz Coil System

Innovation in Magnetic Field Measuring Instruments. Grad601. Magnetic Gradiometer System

Mag-01H. Fluxgate Declinometer/Inclinometer. Mag-01H D/I. World leaders in high precision magnetic field measurements

INNOVATION IN MAGNETICS DS0070/12. Mag-01H D/I. Declinometer/Inclinometer System.

Innovation in Magnetic Measuring Instruments. Mag-03. Three-Axis Magnetic Field Sensors

Spacemag & Spacemag-Lite

Innovation in Magnetic Field Measuring Instruments. Mag-03. Three-Axis Magnetic Field Sensors.

Mag-03. Three-Axis Magnetic Field Sensors.

Mag High Temperature Three-Axis Magnetic Field Probes

Operation Manual for Mag612 Three-Axis Magnetic Field Probe

VSEW_mk2-8g. Data Sheet. Dec Bruno Paillard

Vibration Sentry RT. Data Sheet. May Bruno Paillard

Measuring static magnetic fields

Handheld Gaussmeter Gaussmeter

Operation Manual for. Geolabsoft for Windows (PC version)

Practical and Compact!

Acoustic Camera Nor848A

VIBBOX. 32, 48, or 64-Channel Sound & Vibration Solution Expansion to 256 Channels. Key Feature of Each VIBbox: Table 1. Key Features of VIBbox Models

Kistler LabAmp. Electronics & Software. Charge Amplifier and Data Acquisition Unit for Dynamic Measurements. Type 5165A...

OPERATING MANUAL AND TECHNICAL REFERENCE

Product Identification Product name Package Noise* Range in µt Mag-03 MC= Cylindrical B = Basic ±70 MCES = Cylindrical, environmentally sealed to IP64

Portable vibration calibrator Model 28959G

VIBbox 64-Channel Sound & Vibration Solution

DT8824 High Stability, High Accuracy, Ethernet Instrument Module

Vibration Transducer Calibration System

Analog Input Sample Rate

high performance signal conditioner cards

Operation Manual for. Mag-01 & Mag-01H Single Axis Fluxgate Magnetometers with Mag Probes B-G

VIBRATION METERS VX-54WH. Hand-Arm Vibration VX-54WB. Whole-Body Vibration VX-54WS. Habitability on Ships. 3-Axis Vibration Meter VM-54

Advanced Test Equipment Rentals ATEC (2832)

The Concept of Sample Rate. Digitized amplitude and time

Measurement microphone. MI series

DT9837. High Performance USB Module for Sound & Vibration Analysis. DT9837 Features

FDU108 Fluxgate Sensor Data Acquisition Unit

Instruction Manual. G.R.A.S. 12AA 2-Channel Power Module with gain, filters and SysCheck generator. LI April 2015

Instruction Manual. CCP Supply Type 12AL SOUND & VIBRATION. Skovlytoften 33, 2840 Holte, Denmark

SPIDER-81 HARDWARE SPECIFICATIONS (v6.0)

WALKING TEST KIT WT5000

Measurement Equipment for Future Transient Recorder TransCom-CompactX MF Instruments GmbH

Operation Manual for. Mag628/629 and Mag669 Three-Axis Aerospace Magnetometers.

Vibration Analyzer, Data Collector & Dynamic Balancer

THREE-AXIS HALL MAGNETOMETER THM1176 TECHNICAL SPECIFICATIONS

Features. RoHS COMPLIANT 2002/95/EC

RogaDAQ2. Portable Data Acquisition and Analysis

VC431. Triaxial Vibrometer. The VC431 is a high performance vibrometer; the ideal instrument for measuring vibration according to:

IP 20. Rated. with terminals

GE Measurement & Control. from Bently Nevada * Asset Condition Monitoring

Kistler LabAmp. Electronics & Software. Charge Amplifier and Data Acquisition Unit for Dynamic Measurements. Type 5165A...

A variety of ECONseries modules provide economical yet flexible solutions. Waveform Generation

6220 Ethernet-Based Voltage Measurement Module

A variety of ECONseries modules provide economical yet flexible solutions

CEL 240 Series Digital Sound Level Meters Frequently Asked Questions

DT Channel Analog Input USB DAQ Module. Key Features:

RogaDAQ2 Portable Data Acquisition and Analysis

PCE-VD 3 VIBRATION DATALOGGER Instruction Manual

PRODUCT DATA. Charge to CCLD Converter Types 2647 A, B, C, D, D 001, D 002, D 003, D 004, and E. Features and Benefits. Features

OTS-400 Series Optical Sources, Meters, Testers and Kits with Data Storage Capabilities A LANscape Solutions Product

16-Bit A/D Converter (ENOB=14.1) 16-bit D/A. 16-bit D/A. Digital I/O (8/8) Counter/Timer (2)

MB Millenium Single- and Multi-Axis Vibration and Motion Control System

INTRODUCING. vb and SCOUT Series. from Bently Nevada * Asset Condition Monitoring

WebDAQ 504. Internet Enabled Vibration/Acoustic Logger. Remote Configuration and Monitoring. Virtually Unlimited Storage. Integrated HW and SW

RogaDAQ2 NVH Analyzer & Recorder

Let`s get SIRIUS! SIRIUS Overview. SIRIUS from Dewesoft. SIRIUS Overview. The new hardware generation makes your measurement more precise!

Features Summary. Input Ranges

Magnetic Field Mapping System MMS-1-RS KEY FEATURES:

Measurement has never been this easy. 8 high voltage (250 Vrms) analog inputs. 8 TTL event inputs

Coliy Gaussmeter. Smart Guassmeter Probe GSP G100 Handheld Gaussmeter Distribuito in Italia da :

SCP for Plastics Processing

ECONseries Low Cost USB DAQ

6224 Ethernet-Based Strain Measurement Module

m+p SO Analyzer Dynamic Signal Acquisition and Analysis

Instruction Manual. 44AA Mouth Simulator according to ITU-T Rec. P51 with built-in power amplifier 44AB Mouth Simulator according to ITU-T Rec.

brands you trust. VOTES Infinity - Valve Diagnostic System A Breakthrough in Valve Testing Technology

Integrating Sound Level Meters LA-5570/5560/2560

AGATE TECHNOLOGY AT-2040

TES-1353S INSTRUCTION MANUAL

E101 - Strain Gauge Transducer Display Module. Contents

Really Performs! Here is a Hand-Held Data Logger that Royalton Rd. Cleveland, OH USA Tel: (440) Fax: (440)

Velleman LEGEN HEIRWEG GAVERE BELGIUM. PCS64i PC Scope

GFT6084 Quad 2 / 4GS Digitizer

Portable Data Recorder HMG 1000

Optical wireless gaming headset instruction manual. Item name : 2.4G optical wireless gaming headset Model : HW-933M A For PS3 console

DM3100 Solutions. Application-specific Portable Data Capture Systems built on the DM3100 Data Platform. PC-based Data Logger. Digital Oscilloscope

DataFlow. DataFlow. Electronics & Software. Quality Monitoring System. DataFlow Statistics. Type 2805A...

OTS-400 Series Optical Meters, Sources and Testers A LANscape Solutions Product

UltraTEV Plus 2 Make Partial Discharge discovery simple.

PCD-400A/430A. Sensor Interface. Carrier Wave Type Compact and Moderate Prices. Data Recorders/Analyzers MEASURING INSTRUMENTS

DIAGNOSTICS PN# / / FSP

OTS-600 Series Optical Sources, Meters, Testers and Kits with Data Storage Capabilities

Wireless High-Speed Sensing Platform

Sensors and Systems. Short Form. that protect your plant...

Wilcoxon Research P702B/P702BF-1 operating guide Power unit/amplifier for piezoelectric transducers with internal amplifiers

Wireless High-Speed Sensing Platform

Installation and Operating Instructions for Differential Pressure Sensor Water Tank Level Long Range Telemetry System.

Transcription:

Spectramag-6 Six-Channel Spectrum Analyser for Magnetic Field and Vibration Surveys For innovation in magnetic measuring instruments

Spectramag-6 Six-Channel Spectrum Analyser for Magnetic Field and Vibration Surveys Spectramag-6 is a six-channel, 24-bit data acquisition and spectrum analysis system, designed for use with the Bartington Instruments Mag-03 range of 3-axis fluxgate magnetometers. In addition to magnetometers, the system also has an ICP interface, allowing the connection of a range of accelerometers and microphones. All six-channels are simultaneously sampled, making the Spectramag-6 ideally suited for recording and analysis of magnetic field and/or vibration data in three axes. Typical applications include magnetic and vibration measurements for pre-installation surveys for MRI systems, electron microscopes and similar sensitive equipment, general magnetic measurements, dual magnetometer differential measurements, site surveys and recording magnetic fields due to 50/60Hz mains supplies. The system consists of an interface unit and Windows based PC software. The interface unit is linked to the host PC via a USB2 connection. The software-based nature of the instrument allows for easy upgrading, simply by downloading the latest software version from the internet. Main Features 6-channel, simultaneously sampled, 24 bit data acquisition Magnetic Field and Vibration measurement inputs Time domain and Frequency domain display, with zoom facility 100μs to 10s sample intervals Fixed scan length or continuous acquisition mode (sample rate dependent) Compatible with all Mag-03 and Mag-01MS magnetometers Direct connection of ICP accelerometers or microphones Programmable pass/fail test profiles for time & frequency domains Software based instrumentation permits easy upgrades Operates under Windows 98, 2000 or XP Operates from mains power or internal, rechargeable battery use in the field, with a laptop PC Additional features Averaging for Frequency domain plots, Total field magnitude (xyz vector sum) Selectable front-end gain amplifier Choice of various FFT windowing functions Display cursors Data can be exported as graphics in bitmap or JPG format, or as time-stamped data values Results scaled in engineering units for standard range of sensors. Optional accessories Tripod and adaptor for Mag-03 magnetometers Rugged carrying case

Typical Spectramag-6 Display Cursor and Markers Pass/Fail limits shown on charts Pass/Fail result or{ Modes of use The six input channels are arranged in 2 groups of 3 inputs, which are independently selected for magnetic or vibration measurements. This allows connection of : Two 3-axis magnetometers One 3-axis magnetometer and up to three single-axis accelerometers Up to six single-axis accelerometers For MRI pre-installation surveys the Mag-03MS1000 three-axis magnetic field sensor, with a full-scale range of ±1000μT (±10Gauss) and a resolution down to a few nt, is recommended. The ICP interface provides a 4mA constant current source via a BNC connector and a gain between 1 and 1000 can be selected for vibration measurements down to a few μg. A minimum system for magnetic field measurement comprises: Spectramag-6 unit + Mag-03 magnetic field sensor + Mag-03 cable + tripod + Mag-03 tripod adaptor. A Windows PC with USB2.0 is also required.

Specification Resolution 24 bit A-D Converter Input channels 6 selectable in groups of 3 for magnetometer or accelerometer Input coupling Magnetometers DC or AC selectable with 0.01Hz (-3dB) high-pass filter Accelerometers AC with 0.1Hz (-3dB) high-pass filter Sampling Interval 100μs (min) to 10s (max) Up to 100,000 samples (PC dependent) Continuous sampling mode (slower sample rates only) Frequency range 0-3.5kHz (-3dB point), reduced to 1kHz for gain of 1000 Input impedance (magnetometer inputs) 1MΩ ICP constant current 4mA ± 20% for cables up to 1km in length Gain control software selected x1/x10/x100/x1000 Spectrum range software selected as sample rate or maximum frequency Output interface USB2 Software Windows 98/2000/XP compatible Controls power on/off switch Connectors 2 x Hirose RM15TPD10P fixed plug to magnetic field sensors 6 x BNC sockets for ICP piezoelectric vibration sensors/microphone preamplifiers 1 x usb to PC 1 x 2.1mm socket for 12V input from mains adaptor for recharging Frequency domain display options Amplitude spectrum (RMS) Amplitude spectral density (RMS/ Hz) Power supply Internal rechargeable battery with universal mains adaptor for charging Battery charging time 10 hours for full charge Battery life (typical) 8 hours Enclosure Aluminium Dimensions (mm) 210 x 170 x 112 Weight (kg) 2.85 Operating temperature -10ºC to +50ºC Storage temperature -10ºC to +70ºC Suitable ICP vibration sensor PCB Piezoelectronics type 393A03 (1V/g) low-noise rugged PCB Piezoelectronics type 393B31 (10V/g) low noise rugged Carrying case dimensions (mm) 610 x 230 x 200 Total weight with carrying case 12kg with Spectramag-6, Mag-03 magnetometer, 5m cable and tripod. Specifications of the products described in this brochure are subject to change without prior notice. Windows and Microsoft are registered trademarks of Microsoft Corporation. Bartington is a registered trademark of Bartington Instruments Ltd. Bartington Instruments 5 & 10 Thorney Leys Business Park Witney, Oxford, OX28 4GE, England T +44 1993 706565 F +44 1993 774813 E sales@bartington.com www.bartington.com DS2021/7

SPECTRAMAG-6 VERSION 6 UPGRADE NEW FEATURES TO EXTEND THE CAPABILITIES OF THE SPECTRAMAG-6 SYSTEM. Features Benefits V4.2 V6.1 Increased no. of samples Multiple-scan logging Record seven times more data in a single scan Permits many days of automatic data recording 100,000 samples 700,000 samples Up to 10000 * back-to-back scans Digital prefiltering Post process filtering Anti-alias oversampling Allows filtering of sampled data, displayed in near realtime Digital filtering can be applied to saved scan data Cumulative frequency spectrum can be generated from multiple saved scans Reduce aliasing effects at low data sampling rates Low-pass filter with variable cut-off frequency Low pass filter Cumulative FFT Included * Number of scans is dependant on PC performance and other scan settings With V6.1 software, users can perform operations that are difficult to achieve with earlier versions: Oversampling allows data to collected at slow sampling rates, but without the danger of aliasing of high frequency signals Digital filtering permits discrimination of different frequency components in both time & frequency domains (in V4.2, this can only be done effectively in the frequency domain). If your signals include two or more distinct frequency components (say 50Hz and 1kHz), you can now filter the time domain display to view just the lower frequency. Digital filtering can be applied after data has been acquired and recorded. You would record data at high speed to prevent aliasing, then apply digital filters afterwards (post process) in a controlled fashion. Multiple-scan logging extends the maximum length of data recording to many days, compared to a few hours achievable with V4.2 For the fastest 10kHz sampling rate (used in single-scan mode), the maximum number of samples is now 700,000 per channel, increasing the scan period to 70 seconds. All new Spectramag-6 systems will include V6.1 software. Existing customers can download V6.1 from the www.bartington.com Bartington Instruments Limited 5 & 10 Thorney Leys Business Park Witney Oxfordshire OX28 4GE ENGLAND Telephone: +44 1993 706565 Facsimile: +44 1993 774813 E-mail: sales@bartington.com Internet: www.bartington.com Feb 2008