Dual channel temperature logger with two voltage inputs 0-5V Instruction Manual

Similar documents
Dual channel temperature logger with two binary inputs and LCD display Instruction Manual

Dual channel temperature logger with display Instruction Manual

Temperature logger with display Instruction Manual

Temperature logger with internal sensor Instruction manual

Digital thermo-hygro-barometer with external probe Manual for use

Digital thermometer-hygrometer with external temperature probe Instruction manual

Programmable temperature transducer from Pt1000 sensor to current loop 4-20 ma Instruction Manual

Programmable transmitter of temperature, relative humidity and other derived humidity values with 0 to 10 V outputs Instruction Manual

TRANSDUCER T4311, T4411

T2214. Instruction Manual

Programmable transmitter of temperature, relative humidity and other derived humidity values with RS485 serial output Instruction Manual

T2314, T2414. Instruction Manual

Instruction Manual. T3110 T3110L T3110Ex

T2314, T2414. Instruction Manual

RS Stock No Instruction Manual RS Input Data Logging Thermometer

GREISINGER electronic GmbH. D Regenstauf, Hans-Sachs-Straße 26. T-Logg 120 W -... T-Logg 120 K -...

RS Stock No Instruction Manual RS Input Data Logging Thermometer

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26

Version 12/2013 EN. User Manual Radon Scout/Radon Scout Plus Radon logging

CATALOG OF MEASURING INSTRUMENTS

HygroLog HL20 and HL21 Data Logger. User Guide

HygroLog HL20 and HL21 Data Logger. User Guide

531K. Process Display Temperature display for Pt100 and Ni100 RTD's. Installation and Operating Manual

USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER MODEL : ARC2020

WARNING: Do not use the thermometer/data logger before you read the users manual and the following instructions.

T3311, T3313, T3411 T7310, T7410. Instruction Manual

MIRRIN. Multi-Channel Load Logger MODEL: LL400 User Manual 1 GETTING STARTED INTRODUCTION CONNECTIONS

USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER TPLR-96 Series

Instruction manual M1140 M1200 M1220 M1300 M1320 M1321 M1322 M1323 M1440 Multilogger

F1000 User's Manual. (Version: V1.01)

Instruction manual M1140 M1200 M1220 M1300 M1320 M1321 M1322 M1323 M1440 Multilogger

Programmable Relay ZEN V2 Units

OPERATING INSTRUCTION

Intelligent single channel temperature transducer with Ethernet connection P8511

UNIVERSAL MONITORING SYSTEM MS5D, MS5

Anemometer SD Card Datalogger. Instruction Manual

BreezeMAX Wi² and BreezeACCESS Wi² Quick Installation Guide

DM650TM SENSOR / WALL MOUNTED TEMPERATURE DISPLAY

GREISINGER electronic GmbH. D Regenstauf, Hans-Sachs-Straße 26. T-Logg 100. T-Logg 100 E

Hand-held thermometer Model CTH7000

User's Guide. MiniTec TM Series Model MN25 MultiMeter

MP 101A HUMIDITY TEMPERATURE PROBE INSTRUCTION MANUAL

HD 2013 HD 2013-D AA-26 AA-27

Universal monitoring system with data logger MS6D, MS6-Rack, MS6R

Universal monitoring system with data logger MS6D, MS6-Rack, MS6R COMPLETE SOLUTION FOR MONITORING OF TEMPERATURE, HUMIDITY AND OTHER SIGNALS

Energy Management Energy Transducer Type ET112

HOBO State Data Logger (UX90-001x) Manual

Temperature and humidity regulator H3020 H3021 H3023 H3027. Instruction Manual

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

DPL4000 Portable Low Range Dew Point Analyzer

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

Note: the wiring of the dead-man buttons has been changed since October 2001.

DST550 Digi-Stem Thermometer

TCNM-ACBB1 Installation Manual

Control Panel CP600 CP635, CP635-B, CP635-WEB

DMT340 Series Dewpoint and Temperature Transmitters for Very Dry Conditions

RH/Temperature SD Card Datalogger. Instruction Manual

DESIGO RX Individual room controllers. for fan-coil systems, chilled ceilings and radiators, with LONMARK-compatible bus communications

Control Panel CP600 CP651, CP651-WEB

600 g x 0.02 g DIGITAL BALANCE

Precision Hand-held Thermometer

Autoranging True RMS Multimeter User Manual

LDCL DIGITAL CONTROLLER OPERATING MANUAL

PCD3.C90. Date Modification Asked by: Version E/A Bus Connector PCD2.K106 change to PCD3.K106 Müller.R 1.1

Programmable indoor transmitter of temperature T0318, T0418

PTU300 Combined Pressure, Humidity and Temperature Transmitter

User's Guide. Digital Multimeter. Model MN42

DATA LOGGER (Version V1.3)

OPERATION MANUAL. auto range + type K Temp. optional Humidity, Light, Anemometer SOUND LEVEL METER Model : SL-4112

Druck DPI 620 advanced modular calibrator

OPERATION MANUAL. separate probe, RS-232 SOUND LEVEL METER. Model : SL-4013

Control Panel CP600-eCo CP604, CP604-B

GB/UK. Operating manual. Data logger EBI USB

HOBO Light On/Off Data Logger (UX90-002x) Manual

ReCon T Energy Logger

Operating Manual Digital Quick Response Thermometer GMH 3210 For Thermocouple Probes Type J, K, N, S, T from Version 1.0

TM221M16R controller M IO relay

2 in 1 LAN Tester and Multimeter Model:

DMT345 and DMT346 Dewpoint Transmitters

Hand-held thermometer Model CTH6200

Main contacts Auxiliary contact Number of input channels Rated voltage Model Category 24 VAC/VDC 3PST-NO SPST-NC 1 channel or 2 channels possible

Advanced Test Equipment Rentals ATEC (2832) HMP45A&HMP45D HUMIDITY AND TEMPERATURE PROBES Operating Manual

METREL test and measurement accessories: 3-phase AktivGT / Machine adapter A1322 Instruction manual Version 1.0, Code no.

MC 11 EB-2 Power supply cabinet with external bus, AC version

Oven Loggermaster BEVS User Manual

DM650VI SENSOR/WALL MOUNTED VOLTAGE/mA DISPLAY

FEMA ELECTRÓNICA. Series BDF. LARGE DISPLAYS for parallel BCD code. USER S MANUAL (HT r190307) BDF-xx-BCD for parallel BCD code

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB

PolluTherm. calculator hous. 2 from hous. 1 until X X X X X* X* M-Bus- M-Bus and 2 2PM pulse inputs X X X X

TM3TI4 module TM3-4 inputs temperature

Absolute Encoder Multiturn

HOBO State Data Logger (UX90-001x) Manual

OWNER S MANUAL 9908-TE. HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER

NOVOHALL Rotary Sensor non-contacting. Series RSC2800 digital SSI, SPI, Incremental

HygroLog NT Humidity Temperature Data Logger. Instruction Manual

E3S-A. Built-in Amplifier Photoelectric Sensor (Medium Size) Ordering Information. Built-in Amplifier Photoelectric Sensors. Horizontal. 7 m.

HOBO Motor On/Off Data Logger (UX90-004x) Manual

HEC. General Operating, Maintenance and Installation Manual

Millenium 3. General Characteristics. General environment characteristics for CB, CD, XD, XR and XE product types

HygroLog NT Humidity Temperature Data Logger Instruction Manual

Transcription:

LOGGER S0541 Dual channel temperature logger with two voltage inputs 0-5V Instruction Manual

Instruction Manual for use of S0541 logger Instrument is designed for measurement and record of temperature from two external temperature probes and two external voltage signals 0-5Vdc. Temperature probes and voltage signals are connected by means of connectors. Measured value of temperature and actual values of voltage inputs are displayed on dual line LCD display and are stored in adjusted time interval to internal non-volatile memory. All logger settings and control are performed by means of personal computer and password protection is enabled. It is also possible to switch ON and switch OFF the logger by start/stop magnet (this function can be disabled from the PC). Also automatic switch ON at programmed day and time (maximum one month forward) is enabled. Start/stop magnet enables also to clear the minimum and maximum value memory. Minimum and maximum measured values can be displayed (display switches to actual measured values and min/max values automatically). It is also possible to operate logger with switched OFF display. Short display of actual measured values is enabled by means of magnet. Switched ON logger every 10 seconds (independently on logging interval) updates MIN/MAX memory, compares measured values of each channel with two adjustable limits for each channel and exceeding of limits is indicated on the display (alarm function). Also memory alarm mode is selectable, when alarm is indicated permanently till alarm memory reset. Alarm function is enabled or disabled for each channel individually. Logging mode can be adjusted as non-cyclic, when logging stops after filling the memory. In cyclic mode oldest stored values are overwritten by new. In addition logging mode can be selected when logging is active only if measured value is out of adjusted alarm limits. Stored values can be transferred from logger memory to the PC by means of communication adapter. Communication adapter can be connected to the logger permanently data logging is not interrupted even if data download appears. Logger evaluates minimum battery voltage and its drop below allowed limit is indicated on the display. At the same time value of remaining battery capacity is available by means of the PC program and appears on the logger LCD in % (every time after switch ON). Technical parameters: External temperature probes (can be limited by concrete applied probe): Range of measurement: -90 to +260 C (RTD Pt1000/3850ppm probe) -50 to +150 C (RTD Ni1000/6180ppm probe) Resolution: 0.1 C Accuracy temperature inputs without probe (Pt1000 and Ni1000): ±0.2 C from -50 to +100 C ±0.2 % from reading from +100 to +260 C ±0.4 % from reading from -50 to -90 C Accuracy of the instrument with connected external probe is defined by above input accuracy and the accuracy of connected probe. Type of compatible probe: based on RTD Pt1000/3850ppm or Ni1000/6180ppm equipped with ELKA3008 female connector Recommended probe Pt1000TG8/E (IP67, -80 to +200 C) Response time of N1ATG8 probe (air flow approximately 1 m/s): t63 < 1 min, t90 < 3 min (temperature step 20 C) Voltage inputs (IN3, IN4): Range of measured voltage: 0 to 5 V dc Effective resolution: 13 bits Measurement accuracy: ±0,2 % FS Input resistance: approximately 100k Ohms 2 IE-LOG-S0541-03

Connection of measured signals: shielded cable of 10 m maximum length with connector ELKA 3008V Measuring interval, alarm evaluation, memory MIN/MAX update and display refresh: 10 s Logging interval to memory: 10 s to 24 h (20 steps) Memory capacity: for non-cyclic logging mode 8 126 records for each measured channel for cyclic logging mode 7880 records for each measured channel Specified values are maximum possible and can be reached only if record since last memory erasing was not interrupted. Communication with computer: via RS232 (serial port) by means of COM adapter or USB port by means of USB adapter; data transfer from logger via communication adapter is optical Real time clock: adjustable from computer, integrated calendar including leap years Error of internal RTC: < 200 ppm (i.e. 0.02 %, 17.28 s in 24 h) Power: Lithium battery 3.6 V size AA Battery life: typical (data download to PC is carried out about once a week): 5 years in continuous on-line mode with interval 1 minute: approximately 3 years in continuous on-line mode with interval 10 seconds: approximately 1 year Notice: The above battery lives are valid for logger operation in temperature from -5 to +35 C. If often operated outside of the above temperature range battery life can be reduced to 75%. Protection: IP67 Operational conditions: Operational temperature range: -30 to +70 C Operational humidity range: 0 to 100 %RH Specification of outer characteristics accordingly to Czech National Standard 33-2000-3: normal environment accordingly to appendix NM: AE1, AN1, AR1, BE1 Operational position: negligible Logger installation: by self adhesive Dual Lock, applied to clean, flat surface Not allowed manipulation: it is not allowed to operate instrument with not connected temperature probes - if all probes are not connected, unconnected connectors should be protected by included connector cover. Limit condition: temperature -40 to +70 C, humidity 0 to 100 %RH maximum voltage across voltage input (IN3, IN4): -1 to +7 V dc Storing condition: temperature -40 to +85 C, humidity 0 to 100 %RH Dimensions: 93 x 64 x 29 mm Weight including battery: approximately 130 g Material of the case: ABS Logger operation Logger comes complete with installed battery and switched OFF. Before operation it is necessary by means of installed user PC software to set logging parameters and other features. Free user program is available to download at www.cometsystem.cz. For communication with the PC a communication adapter is necessary (not included in delivery). For connection via RS232 serial port it is necessary to use COM ADAPTER, for connection via USB port it is necessary to use USB ADAPTER. Connect adapter connector to proper computer port and plug the adapter to the guide slots on the side of logger. IE-LOG-S0541-03 3

Connecting logger to the computer by means of communication adapter Notice: USB connector can be located also at the computer front side After connecting the logger to the computer reading of logger info is enabled by means of the PC software and also setting of the instrument accordingly to the user needs (menu Configuration / Setting of instrument parameters). Before logging start it is necessary: check or optionally set the logger real time clock select suitable logging interval select logging mode (cyclic or non cyclic) set for each temperature channel type of temperature probe (Pt1000 or Ni1000) disable channels for logging with not connected temperature probe or voltage signals switch ON the logger (or switch OFF, if it is about to be switched ON by magnet or automatically with delayed start) enable or disable the option to switch ON the logger by magnet enable or disable the option to switch OFF the logger by magnet enable or disable the option to clear minimum and maximum value memory by magnet set date and time of logger automatic switching ON logger or disable this option select if record runs permanently or only if alarm is active If alarms are about to be applied, set both limits for each measured quantity and enable alarm optionally enable permanent alarm indication (alarm with memory) switch ON or switch OFF the display of logger optionally switch ON displaying of MIN/MAX values on the LCD reset memory of MIN/MAX values (if needed) check free space in data memory, optionally erase data memory of the logger enter password if protection against unauthorized manipulation with the logger is necessary Logging interval between to subsequent measurements is specified by the user. Memorizing of the first value is synchronized with the internal real time clock, so the logging is performed at sharp multiples of minutes, hours and days. E.g. after starting the logging with the 15 minute interval the first value is not stored immediately, but after the internal clock gets the status of a quarter, a half or a whole hour. After starting the logging with the 6 hours interval the first value is stored at that whole hour to perform the storing also at 00.00, i.e. at the beginning of the day. First storing is performed at 6.00,12.00, 18.00 or 00.00hour - at the hour from the above nearest to the logging start. After communication with computer or after start by magnet logger automatically waits for the 4 IE-LOG-S0541-03

nearest whole multiple of time and then first measurement is performed. This is also necessary to take into account when setting time of automatic logger switch ON. Notice: If logger operates as permanently connected to the computer, using of magnet start/stop is disabled. To enable the logger control by magnet is suitable only in cases, when possibility of unauthorized manipulation to the logger operation is eliminated. Data logger connection to voltage signal Connect data logger to the voltage signal by shielded cable and three pin socket ELKA. Connector connection to the cable is described at the end of this manual (page 9). Connector pins are marked with colors. Connect signals in accordance with the figure below. Both voltage inputs are galvanically connected and must be supplied from one source. Short change of polarity does not cause the damage, but data logger does not measure correctly (records zero voltage). Permanent incorrect polarity is not allowed (increased consumption from the battery). Note: all GND terminals are galvanically connected in data logger. 1 Legend of connection of connector ELKA: INx brown (input) GND blue (ground) GND black (shielding) Reading on the display in usual operation (logger switched ON) After switching ON the logger all LCD symbols are displayed for checking the display. Then actual date and time in logger is displayed for approximately 4 s. Consequently reading of estimated remaining battery capacity for approximately 2 s is displayed (values 0 to 100%). It is valid if logger is operated at temperature from -5 to +35 C. If logger is operated often outside of the above temperature range battery life can be reduced to 75%, i.e. if displayed remaining battery capacity drops below 25%, it is recommended to replace the battery. If display is switched ON, actual reading of measured values is displayed - channel 1 temperature ( C 1) at the LCD upper line, channel 2 temperature ( C 2) at the LCD lower line. Symbol LOG indicates data logging in progress if it blinks, data memory is filled up to more than 90%. Other measured or calculated values are displayed always after 5 seconds. IE-LOG-S0541-03 5

Instrument now displays value from channel IN3 (LCD upper line) and channel IN4 (LCD lower line). Switched ON logger permanently (with 10 s interval) updates memory of minimum and maximum values of measured values. If displaying of MIN/MAX values is selected, minimum measured value is displayed at the LCD lower line (indicated by symbol MIN) and then similarly maximum measured value (indicated by symbol MAX). Then MIN/MAX values from IN3, IN4 voltage inputs follow. All display cycle repeats periodically, i.e. actual temperature readings follow. Notice: if alarm is active, no MIN/MAX reading is displayed. If display is switched OFF, all above readings are displayed up to estimated remaining battery capacity and then display goes out. If logger is switched ON symbol LOG is displayed (it blinks if memory occupation is higher than 90%). If display is OFF and logger is in mode when record runs only when alarm is active, the LOG symbol is replaced by adjacent symbol " " (hyphen). It appears in case, all measured values are inside of adjusted alarm limits and data logging therefore does not run. Displayed symbol indicates logger is ON. If information on actual measured values is needed, it is possible anytime to display reading by means of the magnet (only if communication adapter is not connected permanently). Plug magnet into guide slots from logger front side for approximately 4 s and wait till reading on the display appears. f logger has enabled the function switch OFF by magnet, resp. MIN/MAX memory clear by magnet, do not remove magnet from guide slots before decimal point symbol goes out logger would be switched OFF, resp. MIN/MAX memory would be cleared! Display reading initialized by magnet goes out automatically after 30 s. Remove the magnet from slots anytime during actual reading is ON or later. Temporary displaying of actual reading by magnet 6 IE-LOG-S0541-03

Alarm indication on the display It is necessary to enable alarm function from the PC and set for each quantity lower and upper limit. If measured value is inside of set limits, alarm of proper quantity is not active. If value of measured quantity gets outside of set limits, alarm of proper quantity is active and it is indicated on the display. It is possible to select "memory alarm mode" when alarm is indicated permanently up to reset from the PC. Active alarm is indicated (if display is ON) by blinking of the value of proper quantity on the display and arrow symbol appears at the LCD upper side at the same time. Arrow 1 indicates active alarm for channel 1 temperature, arrow 2 for channel 2 temperature. Notice: if logger is operated at low temperature (below approximately -5 C), alarm indication by blinking can be indistinct. Indication by arrows works correctly. Messages displayed on the LCD beyond usual operation If measured value is out of the measurable or displayable range numeric reading is replaced by hyphens. If memory is totally full filled in non-cyclic logging mode, logger is switched OFF and message MEMO FULL appears on the LCD. It appears also if logger is operated with switched OFF display. New initialization of logger can occur in switching ON the logger (immediately after displaying of all LCD segments for checking) e.g. after replacement of totally discharged battery for new one. State is indicated by INIT reading. It can be displayed for about 12 s. If battery voltage drop occurred since last internal clock setting below critical limit or battery disconnection for longer period than approximately 30 s, after display switch ON (during date and time display) all four arrows appears as warning to check or set it again from the computer. However all logger functions work without limitation. If reading BAT is displayed periodically on LCD upper line (for 1 s with 10 s interval), the end of estimated battery life is coming however logger functions are not limited. Replace battery as soon as possible! If reading BAT is displayed permanently, battery voltage is low and logger is not possible to switch ON. If logger was switched ON before it, data logging is stopped and logger is switched OFF. Communication with computer can temporarily work. Replace battery as soon as possible! Start / stop by magnet The function must be enabled from the PC before. If only switching OFF by magnet is enabled, it is of course necessary to switch ON the logger from the computer. Notice: It is not possible to combine function switching OFF by magnet and MIN/MAX memory clear by magnet! User software enables only one of them to select. IE-LOG-S0541-03 7

Switching the logger ON by magnet Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point appears right on LCD upper line. After appearance it is necessary immediately (till indication point is displayed) to remove magnet from guide slots and logger switches ON. Switching the logger OFF by magnet The procedure is identical with the above procedure for switching ON. If decimal point does not appear after 1 s, it is necessary remove the magnet and repeat the procedure. Reset of MIN/MAX values by magnet Function enables to clear MIN/MAX values by magnet without using computer. It is necessary to enable the function from the PC software before. Notice: It is not possible to combine this function with function of switching logger OFF by magnet! User software enables to select only one of them (or none). Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point appears right on LCD upper line. After decimal point appearance it is necessary immediately (till indication point is displayed) to remove magnet from guide slots. Reading CLR MIN MAX appears for several seconds and MIN/MAX values will be cleared. Battery replacement Low battery is indicated on the display by blinking of reading "BAT". It can be displayed permanently, if battery voltage is too low. Replace the battery for new one. If logger is operated often in temperature below -5 C or over +35 C and PC program indicates remaining battery capacity below 25% it is also recommended to replace the battery. Applied is Lithium battery 3.6 V, size AA. Battery is located under logger lid. Warning: near battery fragile glass reed contact is located be careful not to damage it. Be careful in battery replacement! 8 IE-LOG-S0541-03

Replacement procedure: switch off the logger by the PC program or by magnet (if low battery allows) unscrew four corner screws and remove the lid remove old battery by pulling the glued tape insert new battery respecting the correct polarity (see symbols + and near battery holder). If you connect new battery up to 30 s, all logger settings remain unchanged. In the opposite case check by means of PC program all settings, especially the real time clock in logger. Attention, inserted battery with incorrect polarity causes logger damage! put the lid back again and screw four screws. Be careful to have the rubber sealing properly in the slot and tighten the screws properly to ensure the water resistance of the instrument. connect logger to the computer and write to it the information on battery replacement (menu Configuration/Battery replacement). This step is necessary to evaluate properly the remaining battery capacity. Old battery or logger itself (after its life) is necessary to liquidate ecologically! Connection of female ELKA 3008V connector for temperature probe and voltage contact It is possible to order as an optional accessory connector for connection of temperature probe. Signal cable should be shielded, round cross section of minimum diameter of 3.5 mm and maximum diameter of 5 mm, maximum length of 10 m and as large as possible cross section of cords (stranded wires). Outer diameter must be from 1 to 1.55 mm including insulation. Connect connector to temperature signal accordingly to the figure by means of the manual included in delivery. Choose the most suitable plastic insert accordingly to the diameter of stranded wires of the cable. The IP67 protection is guaranteed only if connector is assembled correctly! It is necessary to select and connect cable for voltage inputs with respecting the same rules as above. Connect the voltage signals instead of the RTD probe. Connect positive pole to the brown terminal and signal ground to the blue terminal. Due to high impedance of voltage inputs it is necessary to avoid places with higher electromagnetic interference when installing cables from contacts (switchboards, motors, contactors, frequency converters, etc.). Instrument passed through tests of electromagnetic compatibility (EMC): Device conforms in accordance with EN 61326-1 these norms: radiation: EN 55011 Class B immunity: EN 61000-4-2 (levels 4/8 kv, Class A) EN 61000-4-3 (intensity of electric field 3 V/m, Class A) EN 61000-4-4 (levels 1/0.5 kv, Class A) EN 61000-4-6 (intensity of electric field 3 V/m, Class A) IE-LOG-S0541-03 9