CH142: Instructions for the Vernier LabPro

Similar documents
Appendix C. Vernier Tutorial

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips

Logger Pro 3. Quick Reference

848/877 Titrino plus. Tutorial EN

HI FORMOL NUMBER MINITITRATOR for Fruit Juices and Wines analysis

Introduction: Features:

and Multiparameter Meters

Aqueous GPC Instructions - Detailed

LAB 02: Graph Matching

LABQUEST 2 DATA LOGGER

Instruction Manual HI Titratable Acidity Mini Titrator & ph Meter for Dairy Products

WK7200. Water Quality Test Kit USER S MANUAL

Capstone Software Download and Installation

Department of Physics & Astronomy Lab Manual Undergraduate Labs. A Guide to Logger Pro

INSTRUCTION MANUAL CONDUCTIVITY-METER MODEL CC-01

HI FREE & TOTAL SULFUR DIOXIDE MINITITRATOR for wine analysis

870 KF Titrino plus. Installation and tutorial EN

Field Equipment for Measuring Electrical Conductivity, Humidity and ph of Soil

916 Ti-Touch. Tutorial EN

Thermo Scientific Orion Star A214 Benchtop ph/ise Meter. Instruction Sheet

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet

Preliminary Lab Exercise Part 1 Calibration of Eppendorf Pipette

Instruction Manual. Laboratory EC Meter NeoMet Series. EC-400L (EC/TDS/ Salinity/Temp)

Version 1.1. COPYRIGHT 1999 Tufts University and Vernier Software. ISBN (Windows) ISBN (Macintosh)

DataHub: FAQs/Troubleshooting

A150 Benchtop Conductivity Meter Instruction Manual

Montgomery County Community College Document Number: DP DeKalb Pike Revision Number: 2. Page 1 of 9 SOP: BioLogic LP Chromatography System

Physics 1050 Experiment 2. Acceleration Due to Gravity

GRAPH MATCHING EQUIPMENT/MATERIALS

USER GUIDE. Salinity Meter Pen Style Water Quality Meter. Model EC170

Conductivity and TDS Meter Pen Style Water Quality Meter

PH450 Analyzer Proper Calibration Steps

ORION Research Inc. Table of Contents

DO Meter Kit. Instruction Manual

Operation Manual. MODEL ph/ec80. Microcomputer Based ph/condctivity/tds/ Sanility & Temperature Pocket Meter. ph/ec80

User's Guide. Online Meters. PH / ORP / TDS / Conductivity.

LabQuest Quick-Start Guide

HI TOTAL ACIDITY MINITITRATOR & ph METER for Water Analysis

Thermo Scientific Orion Star A326 Portable ph/rdo/do Meter. Instruction Sheet

FROG-4000TM Quick Start Guide

Physics 1020 Experiment 3. Acceleration of Falling Objects

915 KF Ti-Touch / 916 Ti-Touch. Compact titrator for routine analysis

EasyData Guidebook. Easy and Simple Data Collection using Vernier Sensors and Texas Instruments graphing calculators

Thermo Scientific Orion Star A325 Portable ph/conductivity Meter. Instruction Sheet

The Olis Titrator Series

Dionex AS-AP Sample Conductivity and ph Accessory Setup and Operation Guide

USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter

STARTER 3000 Bench ph Meter Instruction Manual

Application Bulletin 318 e. Installation Instructions of TitrIC 4

USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements

ph, ORP, and TDS ORP Continuous Monitor OPERATING INSTRUCTIONS 8. Warranty 9. Return of Goods

PerkinElmer Series 200 HPLC: Operating Procedure

Application Bulletin 321 e. Installation Instructions of TitrIC 7

The Economy Conductivity Benchtop Meter

Operations Manual EcoSense ph100a

Using DataQuest on a Handheld

High-Temperature Susceptometer (Kappabridge)

chamber called an Zeta-Meter 4.0 Unit

Manual HPLC-4 Alliance-UV

Mini Mini GlobiLab Software Quick Start Guide

DDS-12DW Benchtop Conductivity Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

INSTRUCTION MANUAL ORP-METER MODEL COR-01

Instruction Manual HI 9033 HI Waterproof Multi-Range EC and TDS Meters for Field Applications.

ph/orp Meter SD Card real-time Datalogger

Thermo Scientific Orion Star A324 Portable ph/ise Meter. Instruction Sheet

Installation Instructions of TitrIC 1

Checklist for the SMPS & CPC (5.400 & 5.500) PART 1 Setup of the external Plumbing (for the DMA and the CPC):

CyAn ADP Guide. Starting Up

C100 Handheld Cond./TDS/Sal. Meter. Instruction Manual

Operation Manual. Hand-held ph/mv/temperature Meter. ph6810

Large Display Salinity & Conductivity Pens &

815 Robotic USB Sample Processor XL

Equipment Overview: Safety:

INSTRUCTION MANUAL. Portable ph and ph/ise Meters. Models 210Aplus 230Aplus 250Aplus 290Aplus

Physics 1020 Experiment 2. Motion in One Dimension

Logger Pro 3 Release Notes

Bante 221 Portable ph/orp Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

GeneMate Benchtop ph Meters Operation Manual

HI 2300 Auto-ranging Microprocessor EC/TDS/NaCl/ºC Bench Meter

ph 8+ DHS ph- mv - Temperature GLP - USB Data Logger - RS232 printer

Portable Conductivity Meter M1410. User Manual 1.002

Bante901 Benchtop ph/conductivity Meter Instruction Manual

Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter

Using I9131 Sodium Combination ISE on 3200I/M

Thermo Scientific Orion Star A223 and Star A323 Portable RDO/DO Meter. Instruction Sheet

BL121 Pool & Spa Controller. ph/orp Swimming Pool Controller with Built-In Chemical Feed Pumps

Handheld Water Quality Meters 100 Series

915 KF Ti-Touch. Tutorial EN

Operation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M

Manual No Revision MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER

HI HI 9034 HI HI 9130 HI HI 9635 Portable Waterproof Multi-Range Conductivity/TDS/ Fertilizer Meters

Instruction manual of the controller unit. PCS ph-redox-chlorine-temp PCS ph-oxygen-temp PCS ph-redox-temp

Operating Instructions. Titrator DL15

Visual Physics - Introductory Lab Lab 0

MultiTherm Shaker. Operations Manual Item: H5000-H and H5000-HC

CHLORINE DIOXIDE (0 to 5.00 mg/l) Method 10126

Tracer ph/tds/salt PockeTester TM Code 1766

4801P ph/orp controller

User Manual. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500. Additional User Manual Translations available at

Transcription:

1 CH142: Instructions for the Vernier LabPro INTRODUCTION: The Vernier LabPro is a versatile data collection interface that can be used in combination with more than 45 sensors, and with a computer, TI graphing calculator, Palm OS handheld or battery powered as a stand-alone data logger. Because of this versatility, the Vernier LabPro can be used in many different ways in the classroom or in the field. The Vernier LabPro can be easily connected to a computer and used to titrate a strong and/or weak acid with a strong base (0.1M NaOH). The titration can be studied by monitoring changes in ph (using a ph electrode) or conductivity (using a conductivity probe). The interpretation of the differences in the shape of a titration curve and the information that can be determined from the equivalence point of a titration can be valuable information in the identification and concentration of an acid. Conductivity curves can be used to describe the behavior of these acids in an aqueous environment. INSTRUCTIONS FOR USE: A. Turn on the power to the power strip that is connected to the Vernier LabPro, the stir plate and the power supply for the DCU pump unit. A blinking green light occurs when the interface receives power. The Vernier system will complete series of (7) beeps as the system completes a self check. If no other beeping sounds are given by the interface, then the system is ready for use. The pump should be connected to the DIG/Sonic1 port and an electrode amplifier with an attached ph electrode should be connected to CH1. The conductivity probe will be connected later in the procedure. B. Calibration of the pump (finding the volume of one drop delivered by the pump): Open the desktop folder labeled CH142VernierLabPro. Double click on the icon one minute pump cal.cmbl. The pump will click and a blinking green light(s) will be visible on the interface when recognized by the software. (The toggle switches on the pump should always be set on the computer and DCU settings for the software to communicate to the pump.) NOTE: If the equipment is not set up properly, a Sensor Confirmation window will be displayed. Be sure the sensors are securely attached to the interface and in the correct ports. If the instrument amplifier is shown to be in use in the top portion of the window, turn off this amplifier by clicking on the checkmark in the use box. Check to see that the electrode amplifier and the digital control unit (pump) are listed in the second portion of the box and given a green checkmark to the left of each listing. Click OK. If the window is not displayed, continue on.

2 The main calibration window should be activated and the Collect button found in the top menu bar will be colored green. The table to the left will record collected data in the appropriate columns. The smaller box below the table will count the pulses delivered by the pump. The graph will display a green signal for each pulse. To prime the pump with deionized water, find the inlet tubing (marked as IN on the pump) and submerge the tubing into a clean 100 ml beaker of fresh dh 2 O. To prevent air bubbles from entering the line, make sure that the line is secure and that the opening of the tubing is at the bottom of the beaker. Put the outlet tubing into a waste beaker. Click on the green Collect button. A window will appear that will ask what you would like to do: *erase and continue: will erase the last set of pulses and start at zero *cancel: cancel or exit *append to latest: will add another series of pulses to the last one *store latest run: save data in the table before starting another run Each run is a series of 30 pulses. Click on erase and continue. The pump will run for 30 pulses (refer to the small box under the table). Click on Collect again and then append to latest to add another series of 30 pulses. It will take about 90 pulses to prime the pump with water. Check for bubbles in the tubing. If you see bubbles, check your set up and add another set of pulses. If no bubbles are apparent in the lines, then continue. To calibrate the pump (volume in ul delivered by each pulse) find the mass of a clean dry weigh bottle (no top) to 0.1 mg. Use a clean, dry Kim Wipe to handle the bottle because fingerprints and dirt will add to the mass and result in calibration error. Set the outlet tubing into the weigh bottle and deliver 60 pulses of deionized water to the bottle. Mass the bottle again. Using the density of water at room temperature (0.9982 g/ml at 20C), convert the mass of the water delivered to the corresponding volume of water delivered. Calculate the volume of water delivered in a single pulse by the pump (in ul). Remove the inlet tubing from the beaker of water. Empty the line by delivering 30-60 pulses (of air). Place the outlet tubing into a waste beaker and the inlet tubing into a 100 ml beaker of your standardized NaOH (titrant). Prime the pump with the strong base. Be sure that there are no bubbles in the line. Exit the calibration program by using the LoggerPro menu in the main menu bar Quit LoggerPro.. Save.

3 C. To set up a titration using Vernier LabPro: Double click on the CH142Titrate.cmbl icon (found in the desktop CH142VernierLabPro folder). The CH142Titrate.cmbl window will be displayed. As with the calibration software, the table on the left will record collected data in columns labeled with color coded headings. A small window under the table will display ph (live reading). Temperature will be displayed if the temperature probe is connected to the interface. The larger graphing window will display your data as it is being recorded by the interface. The X axis is the volume (ul) delivered by the pump and is set from 0 to 20,000 ul. The Y axis is the sensor output and is set to record ph 0 to12. (NOTE: If necessary, the axes may be rescaled at anytime by clicking on the minimum/maximum value and typing in the preferred value.) To calibrate the volume ( X ) axis, use the scroll bar at the bottom of the table, to find the volume column and double click on the heading. A Calculated Column window will appear. Find the equation box and change ONLY the last number in the equation to the calibrated pulse volume in ul that your pump delivers. Click Done. To calibrate the ph electrode chose the Experiment menu from the LoggerPro menu bar and choose Calibrate. Sliding to the right choose LabPro: 1CH1: Electrode Amplifier. The sensor settings window will appear. Make sure that the current calibration is set for Electrode Amp ph<sensor Page 1>. Click on the calibrate now button. This is a two buffer point calibration. Remove the ph electrode from the electrode storage solution, make sure that the hole in the blue collar is open, rinse well with deionized water and place in the ph 4 (pink) buffer solution. Always make sure the ground glass frit is covered by solution and no bubbles are caught in the protective cage. Once the mv reading has stabilized, type 4.00 into the enter value box and press the Keep button. Repeat to calibrate in ph 7 buffer solution and press the Keep button. Press Done. Add 50 ml of deionized water to a clean, dry 100 ml beaker (lines on beaker are OK). Using a 10 ml volumetric pipet, add 10 ml of acid to the water, place the beaker on the stir plate and add a stir bar. Rinse and dry the ph electrode and carefully place it in the beaker. Rest the electrode in the extended portion of the lip of the beaker. Carefully turn the stir plate on and arrange the beaker so that the stir bar does not hit the electrode cage. Record the starting ph. Using the provided clip, attach the pump outlet tube to the beaker so that the end of the tubing is right at the level of the solution being tested. Press the green Collect button in the main window menu bar to start the titration. For the first titration, click erase and continue. The data will be recorded in the table and the titration and derivative plots will automatically be displayed on the graph.

4 (NOTE: If this is not true, double click on the Sensor Output Y axis label and choose More. Then check the ph and dph/dv boxes for the appropriate run. Remove the check for any unwanted data being displayed on the graph.) Press Stop when the ph has reached 11 to 12 to end the titration ( Stop is where the Collect button was at the start of the run). The color of the curves will match the color of the headings to the columns in the table. To help with the interpretation of the data, the color of the titration and derivative plots should be the same for each titration experiment. If they are different, double click on the ph column heading and choose Options from the Column Options box. Choose a color from the color bar and press Done. Repeat this to make the derivative plot (dph/dv) the same color for the run. Be careful to choose colors that will not to be too dark to cover other runs that will be displayed on the chart. Choose File from the Logger Pro Main window menu bar and Save to assure that the data is not lost while preparing the next titration or analyzing the data. For additional titration experiments, press Collect and Store latest run. The data will be collected in the table and the titration and derivative plots will be overlayed on the graph. When all of the titration experiments are completed, and have been saved, the graph can be printed by choosing File from the LoggerPro main window menu bar and Print Graph. Entering the information in the Printing Options window will provide a label to the hardcopy of your graph. Press OK. Setting the printer to HP4560 will produce a color copy of the data. Select the number of copies and press Print. The copies will be printed in the Schupf Lab. To analyze the data points of interest on the ph and derivative plots, press the X= (also called the Examine ) button in the CH142Titrate.cmbl window menu bar. The color of the data values will match the color of the plotted data. Any point along the curves can be evaluated (also recorded in the table). Record the data points that are important to your experiment. Click on the X= button again to exit data analysis. You can also press the Control-Option keys on the keyboard while holding down the mouse with the cursor placed on the graph. Choosing copy from the menu box will allow the graph to be saved as a MicroSoft Word document. Open Word and use Paste under the Edit menu to copy the graph. The graph can be further edited in Word. If you use the Save as option to save the data, you would need the LoggerPro Software on your computer to reopen the file. To exit the program, choose Quit LoggerPro from the LoggerPro menu bar. Press Save.

5 NOTE: IF YOU ARE FINISHED WITH THE VERNIER SYSTEM, THE TITRANT MUST BE FLUSHED FROM THE PUMP. Reopen the one minute pump calibration.cmbl and place the inlet tubing into a beaker of deionized water. Place the outlet tubing in a waste beaker. Be sure that the line is secure and that the open end of the tubing is placed at the bottom of the beaker to reduce bubbles in the line. Use at least 90 pulses to run deionized water through the pump and tubing. Then place the inlet hose on the benchtop and use 60 pulses to dry the lines. Please rinse the ph electrode and place it in the electrode storage solution (the hole in the blue collar should be open). Please turn off the power strip.

6 D. Using the Vernier System to measure Conductivity: Carefully detach the ph electrode from CH1 and place the conductivity probe into CH1. Please check to see that the toggle found on the box connected to the conductivity probe is set to the 0-2000 us/cm range. Double click on the CH142Conductivity.cmbl icon in the CH142Vernier LabPro folder on the desktop. The CH142Conductivity.cmbl window will look and function like the CH142Titration.cmbl window. The software will automatically recognize the conductivity probe and the smaller box under the table will now display Conductivity in us/cm (rather than ph). To calibrate the X axis to the calibrated pump volume used in the titration experiments, double click on the volume column heading in the table and enter the calibrated pulse volume (ul) at the end of the equation in the Calculated Column window. Press Done. To calibrate the conductivity probe, choose Experiment from the LoggerPro main menu bar and choose Calibrate. Slide right to choose LabPro:1CH1:Conductivity. In the Sensor Settings box, check to see that the current calibration reads Con2000uS/cm <Sensor Page 0>. Using Live Calibration, press the calibrate now button. The units should be displayed as us/cm. With the dry probe set on the desktop, enter 0 in the Enter value box and press Keep. The place the probe into the Sodium Chloride Standard (eq. 1000 us/cm) solution. When the mv value is stabilized, type 1000 into the Enter value box and press Keep. Then press Done. Prepare a 10 ml sample of your acid as described in the titration instructions, and carefully rest the conductivity probe on the lip of the beaker. Be sure that the electrode surfaces in the elongated cell are completely submerged in the liquid and that no bubbles are trapped in the cell. Record the starting conductivity measurement. The stir bar should not be in contact with the probe. Attach the pump outlet tube to the beaker so that the end of the tubing is right at the level of the solution being tested. Press the green Collect button in the CH142Conductivity.cmbl menu bar to titrate the acid. Select erase and continue to start the experiment. The conductivity measurements will be recorded in the table and the conductivity plot will be displayed in the graph. (If this is not true, double click on the sensor output Y axis label and choose Conductivity. (Remove the check for any unwanted data being displayed on the graph.). Press Stop when volume of approx. 15000 ul is reached to end the titration ( Stop is where the collect button was at the start of the run). Choose File from the Logger Pro Main window menu bar and Save to assure that the data is not lost while preparing the next titration or analyzing the data.

7 The graph can be printed by choosing File from the LoggerPro main window menu bar and Print Graph. Including the information in the Printing Options window will provide a label to the hardcopy of your graph. Press OK. Setting the printer to HP4560 will produce a color copy of the data. Select the number of copies and press Print. The copies will be printed in the Schupf Lab. To analyze the data points of interest on the conductivity plot, press the X= (also called the Examine ) button in the CH142Titrate.cmbl window menu bar. Any point along the curves can be evaluated (also recorded in the table). Record the data points that are important to your experiment. Click on the X= button again to exit data analysis. You can also press the Control-Option keys on the keyboard while holding down the mouse with the cursor placed on the graph. Choosing copy from the menu box will allow the graph to be saved as a MicroSoft Word document. Open Word and use Paste under the Edit menu to copy the graph. The graph can be further edited in Word. If you use the Save as option to save the data, you would need the LoggerPro Software on your computer to reopen the file. To exit the program, choose Quit LoggerPro from the LoggerPro menu bar. Press Save. Rinse the end of the probe with deionized H 2 O between measurements. If you are finished using the Conductivity Probe, rinse the probe with deionized water and use a Kim Wipe to dry. The probe can be stored safely dry on the benchtop. NOTE: IF YOU ARE FINISHED WITH THE VERNIER SYSTEM, THE TITRANT MUST BE FLUSHED FROM THE PUMP. (REFER TO PAGE 5 FOR THE INSTRUCTIONS FOR CLEAN UP AND SHUT DOWN.)