Skyline High Resolution Metabolomics (Draft)

Similar documents
Skyline Targeted MS/MS

Tutorial 7: Automated Peak Picking in Skyline

Skyline Collision Energy Optimization

Skyline irt Retention Time Prediction

Skyline MS1 Full Scan Filtering

MRMPROBS tutorial. Hiroshi Tsugawa RIKEN Center for Sustainable Resource Science

Skyline Targeted Method Editing

TraceFinder Analysis Quick Reference Guide

Pinpoint Quick Start Guide

Tutorial 2: Analysis of DIA/SWATH data in Skyline

Panorama Sharing Skyline Documents

Skyline Targeted Method Refinement

Skyline Targeted Method Refinement

MRMPROBS tutorial. Hiroshi Tsugawa RIKEN Center for Sustainable Resource Science MRMPROBS screenshot

Progenesis CoMet User Guide

Agilent Triple Quadrupole LC/MS Peptide Quantitation with Skyline

TraceFinder Analysis Quick Reference Guide

Retention time prediction with irt for scheduled MRM and SWATH MS

Progenesis CoMet User Guide

MS1Probe Statistical Processing for MS1 full scan data

Statistical Process Control in Proteomics SProCoP

TraceFinder Administrator Quick Reference Guide

TraceFinder Shortcut Menus Quick Reference Guide

Thermo Xcalibur Getting Started (Quantitative Analysis)

Skyline Processing Grouped Study Data

Skyline Processing Grouped Study Data

What s New in Empower 3

Tutorial for the Cross-link Transition Calculator Skyline Plugin

MassHunter Personal Compound Database and Library Manager for Forensic Toxicology

Agilent G6825AA METLIN Personal Metabolite Database for MassHunter Workstation

Procedures for Group Quantitation

SpectroDive 8 - Coelacanth. User Manual

Agilent G6854AA MassHunter Personal Compound Database

Tutorial for the Cross-link Transition Calculator Skyline Plugin

Agilent G6854 MassHunter Personal Pesticide Database


Xcalibur. QuickQuan. User Guide. XCALI Revision C July For Research Use Only Not for use in Diagnostic Procedures

PRM Method Development and Data Analysis with Skyline. With

To get started download the dataset, unzip files, start Maven, and follow steps below.

Agilent 6400 Series Triple Quadrupole LC/MS System

Agilent MassHunter Qualitative Data Analysis

Protein Deconvolution Quick Start Guide

Agilent G2721AA Spectrum Mill MS Proteomics Workbench Quick Start Guide

MS-FINDER tutorial. Last edited in Aug 10, 2018

PROTOCOL: GCP MS Analysis (MSA)

Agilent RapidFire 300 High-throughput Mass Spectrometry System

Tutorial for the PNNL Biodiversity Library Skyline Plugin

MSstats Group Study Statistics

QuiC 1.0 (Owens) User Manual

irc Pro (Not for diagnostic use)

Xcalibur Library Browser

Progenesis QI for proteomics User Guide. Analysis workflow guidelines for DDA data

Using SIMDIS Expert 9 (Separation Systems) with Clarity

Reference Manual MarkerView Software Reference Manual Revision: February, 2010

MassHunter Pesticides PCD or PCDL Quick Start Guide

Chromeleon / MSQ Plus Operator s Guide Document No Revision 03 October 2009

Basic tasks in Excel 2013

Chromeleon software orientation

Note: The MS/MSALL with SWATH Acquisition MicroApp version 2.0 supports PeakView software versions 2.1 and later.

Agilent MicroLab Quant Calibration Software: Measure Oil in Water using Method IP 426

De Novo Peptide Identification

Getting low-resolution GC-MS data on the JEOL GCMate Introduction

This manual describes step-by-step instructions to perform basic operations for data analysis.

PeakSimple Calibration Tutorial August 2008

Chapter 3: Rate Laws Excel Tutorial on Fitting logarithmic data

Agilent G2727AA LC/MS Data Browser Quick Start

Progenesis LC-MS Tutorial Including Data File Import, Alignment, Filtering, Progenesis Stats and Protein ID

MS-FINDER tutorial. Last edited in Sep. 10, 2018

PEAKS Studio 5 User s Manual

Scientific Graphing in Excel 2007

TraceFinder Acquisition Quick Reference Guide

Annotation of LC-MS metabolomics datasets by the metams package

OpenLynx User's Guide

Processing of Data - Basic Quantitative Calculations

Q Exactive HF-X Software Manual

Importing and processing a DGGE gel image

Pathway Analysis of Untargeted Metabolomics Data using the MS Peaks to Pathways Module

LightCycler 480. Instrument and Software Training. Relative Quantification Module. (including Melting Curve)

Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017

Importing data in a database with levels

QPCR Tutorial. Version 3.0. Create Date Sep 10, 2008 Last modified May 25, 2009 by Stephan Pabinger

Skyline Command-Line Interface

Scientific Graphing in Excel 2013

Installation Instructions

Galaxie Photodiode Array Software

PEAKS Studio 5.1 User s Manual

Spectronaut Pulsar X

Agilent 6300 Ion Trap LC/MS Systems Quick Start Guide

Near real-time processing of proteomics data using Hadoop. Hillman, Chris; Ahmad, Yasmeen; Whitehorn, Mark; Cobley, Andrew

StarFinder. Operation Manual. Star Chromatography Workstation Version 6

BIOINF 4399B Computational Proteomics and Metabolomics

TraceFinder Acquisition Quick Reference Guide

UPDATING PESTICIDE RETENTION TIME LIBRARIES FOR THE AGILENT INTUVO 9000 GC

Overview. Experiment Specifications. This tutorial will enable you to

MilkyWay Proteomics Documentation

Release Notes. Analyst Software with Hotfixes to February 2011

MALDI Software Users Guide (simplified version)

Tutorial for Windows and Macintosh SNP Hunting

QuiC 2.1 (Senna) User Manual

v MODFLOW Transient Calibration Calibrating transient MODFLOW models GMS Tutorials Time minutes

Transcription:

Skyline High Resolution Metabolomics (Draft) The Skyline Targeted Proteomics Environment provides informative visual displays of the raw mass spectrometer data you import into your Skyline documents. Originally developed for proteomics use, Skyline has been extended to work with generalized small molecules. This tutorial explores using Skyline for targeted quantification of small molecules using high resolution data. In this tutorial, you will learn about Quantification of Polyunsaturated Fatty Acids in Plasma using NIST SRM-1950 as Single-Point External Calibrant, used along with stable-isotope internal standards. You will explore: Use of molecular formula and adducts to assign precursor accurate mass Analysis of high resolution accurate mass precursor quantification data in Skyline Quantification using Single-Point External Calibration with SIL internal standards and surrogate standards. Use of Concentration Multiplier to adjust calibration range for each analyte. Skyline aims to provide a vendor-neutral platform for targeted quantitative mass spec research. It can import raw data from the instrument vendors Agilent, SCIEX, Bruker, Shimadzu, Thermo-Scientific and Waters. The ability to import data across various instrument platforms greatly facilitates crossinstrument comparisons and large multi-site studies. This remains equally true in using it to target small molecules, as it has been for years in the field of proteomics. If you have not already looked at the Skyline Small Molecule Targets tutorial you should do so now, in order to pick up a few basics about how Skyline works with small molecule descriptions including chemical formulas and adducts. Getting Started To start this tutorial, download the following ZIP file: https://skyline.gs.washington.edu/tutorials/hiresmetabolomics.zip Extract the files in it to a folder on your computer, like: C:\Users\bspratt\Documents This will create a new folder: 1

C:\Users\bspratt\Documents\HiResMetabolomics It will contain all the files necessary for this tutorial. Now start Skyline, and create a new empty document. Importing a Small Molecule Transition List into a Skyline Document The easiest way to get a small molecule transition list into a Skyline document is to start with an empty document and use the Edit > Insert > Transition List menu item. Transition list insert To begin, do the following: If you are an Excel user, open the provided PUFA_TransitionList.xlsx file. Otherwise, you can open the CSV version PUFA_TransitionList.csv in any text editor. On the Skyline Edit menu, choose Insert and click Transition List. Skyline will show the Insert form, which may start out looking like this: If it does, you can change it to accept small molecule fields by doing the following: Click the Small molecules option at the bottom of the form. The form should now look something like this (you may have a different column selection and order from previous uses of Skyline): 2

You can see that there are some extra column headers in the Insert form, and the column order is not the same in the form as in the spreadsheet. Both issues are easy to correct: Click the Columns button and uncheck the columns that do not appear in the spreadsheet. This should result in a column picking menu like the one shown below: Next do the following to reorder the columns in the Insert form: Click and drag each column header you want to move to the order matching the spreadsheet. Once you have selected and arranged your columns, the insert form should now appear as shown below: 3

Now: Copy the transition list from Excel (or your text editor, for CSV) and paste into the Insert form. Make sure to omit the header row. There is an intentional error in the transition list: charge is given as 1, but the adduct is [M-H]. If you click the Check for Errors button you will see this: Change the charge value to -1 in each row and try the Check for Errors button again. You will see this: 4

Notice that Skyline has automatically filled in the Label Type column, having determined that the first two entries are a heavy/light labeled pair based on having the same name and formulas that differ only in isotopic labeling (four of the hydrogens are replaced by Deuterium in the second formula). Click on the Insert button. On the Skyline Edit menu, click on Expand All then click on Molecules. 5

Your Skyline window should now look like this, displaying a tree of polyunsaturated fatty acids which we hope to quantify using high resolution extraction, along with their stable-isotope internal standards (as applicable): Note that for heavy labeled pairs, the label is expressed as part of the adduct description for the heavy variant. The [M4H2-H] adduct tells us that four of the molecule s hydrogens are replaced by H2 ( M4H2 ) and that it is ionized by deprotonation ( -H ). You will notice that ALA (alpha-linoleic acid) does not have a heavy-labeled variant. Instead, we will assign a different stable-isotope labeled molecule as its surrogate standard. We will use d5-dha in this case because it is the closest in retention time: right-click on the DHA target and choose the Set Standard Type submenu Click on Surrogate Standard 6

Transition settings Next we have to make sure Skyline s Transition Settings are correctly set for importing the experimental mass spectrometer results. To do this, perform the following steps: On the Settings menu, click Transition Settings. Change settings as necessary to match the following: 7

Note: the f,p setting in Ion Types on the Filter tab means we are interested both fragment and precursor ion transitions. This Skyline document contains only precursor transitions, but the f is harmless. Now we are ready to import the experimental mass spectrometer results. Importing mass spectrometer runs Perform the following steps. On the File menu, click Save. (Ctrl-S) Save this document as SM_HiRes_v1.sky On the File menu, select Import and click on Results. In the Import Results form, choose to import single-injection replicates. For best performance, be sure to select Many in the Files to import simultaneously control at the bottom of the form. Now click the OK button. Select all 16 raw data folders in the tutorial folder by clicking the first listed and then holding down the Shift key and clicking the last. The Import Results Files form should look like: Click the Open button. 8

The files should import within 30 seconds or so, leaving your Skyline window looking something like this: To take advantage of the Skyline summary graphs for viewing individual targets, do the following: On the View menu, choose Peak Areas and click Replicate Comparison. On the View menu, choose Retention Times and click Replicate Comparison. Click and drag these views to dock them above the chromatogram graphs. On the View menu, choose Document Grid. In the Document Grid view, click on the Views control and select Peptide Quantification. Click and drag the Document Grid view and dock it next to the chromatogram graphs. 9

The Skyline window should now look something like this: Checking Peak Integration Looking at the Retention Times Replicate Comparison window we can see by the lack of outliers that Skyline did not have any problems with peak integration. Preparing for Single Point Quantification On the Settings Menu, click on Peptide Settings. Click on the Quantification tab. Modify the settings as needed to look like this: Click on OK. 10

Now, returning to the Document Grid view: Click on the Views control and select Replicates. This step establishes the role of each of the samples ( replicates ) in the study, as standards, unknowns, or quality control samples. Edit the sample types and analyte concentrations as necessary so the view looks like this: Note that in this situation, the concentration given for the NIST sample (NIST-SRM-1950) is the published consensus concentration for FA 18:2 in this reference material, which incidentally is the highest concentration of any analyte in the document. Therefore, in the next step, the concentration multipliers will each be <1 in order to adjust the target concentration of those analytes to their respective NIST SRM-1950 reference values. In the Document Grid, click on the Views control again and select Peptide Quantification. Set the concentration multiplier and normalization method for each target to match this: 11

The default normalization method is ratio to heavy when using stable-isotope internal standards, but we need to set the FA 18:3 analyte to use the previously-defined surrogate standard as the denominator in the light/heavy ratio. 12

Inspect the Calibration Curves Each entry in the Calibration Curve column is a clickable link that brings up the calibration curve view for the molecule in that row. Click on the calibration curve link for FA 18:2 and the Calibration Curve view will appear: Note that the calibration curve by default has a slope =1 and an intercept = 0, as defined by the method of quantification (Settings menu, Peptide Settings, Quantification tab) and the use of a single point of external calibration. The Document Grid or Report Grid can be used to export quantitative data (in micromolar) for each sample analyzed. Conclusion In this tutorial, you have learned how to create a Skyline document that targets small molecules specified as precursor ion chemical formulas and adducts. You imported a multi-replicate data set collected on a Q Exactive orbitrap mass spectrometer for a set of plasma samples,, and saw how many existing Skyline features created initially for targeted proteomics use can now be applied to small molecule data. Small molecule support is still a relatively new feature area for Skyline. As such, you can expect it to continue improving rapidly. 13