Novel evaluation method of low contrast resolution performance of dimensional X-ray CT

Similar documents
Performance Evaluation of 3-Axis Scanner Automated For Industrial Gamma- Ray Computed Tomography

Frequency-based method to optimize the number of projections for industrial computed tomography

Efficient 3D Crease Point Extraction from 2D Crease Pixels of Sinogram

Japan Foundry Society, Inc. Application of Recent X-ray CT Technology to Investment Casting field. Kouichi Inagaki ICC / IHI Corporation

Non-Destructive Failure Analysis and Measurement for Molded Devices and Complex Assemblies with X-ray CT and 3D Image Processing Techniques

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

Quality control phantoms and protocol for a tomography system

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

Coordinate Measuring Machines with Computed Tomography

Comparison of Scatter Correction Methods for CBCT. Author(s): Suri, Roland E.; Virshup, Gary; Kaissl, Wolfgang; Zurkirchen, Luis

DUE to beam polychromacity in CT and the energy dependence

Recognition and Measurement of Small Defects in ICT Testing

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

Comparison of Reconstruction Methods for Computed Tomography with Industrial Robots using Automatic Object Position Recognition

Thickness Measurement of Metal Plate Using CT Projection Images and Nominal Shape

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

3D Computed Tomography (CT) Its Application to Aerospace Industry

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS

Development of Reverse Engineering System for Machine Engineering Using 3D Bit-map Data. Tatsuro Yashiki* and Tarou Takagi*

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications

Basics of treatment planning II

CT vs. VolumeScope: image quality and dose comparison

A deformable model driven method for handling clothes

ACTIVE SEPARATION CONTROL WITH LONGITUDINAL VORTICES GENERATED BY THREE TYPES OF JET ORIFICE SHAPE

Development of a multi-axis X-ray CT for highly accurate inspection of electronic devices

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

DEVELOPMENT OF CONE BEAM TOMOGRAPHIC RECONSTRUCTION SOFTWARE MODULE

CIVA Computed Tomography Modeling

INDUSTRIAL SYSTEM DEVELOPMENT FOR VOLUMETRIC INTEGRITY

Three-dimensional nondestructive evaluation of cylindrical objects (pipe) using an infrared camera coupled to a 3D scanner

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Extracting Woven Yarns of Ceramic Matrix Composite Parts With X-ray CT Scanning

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT

Characterization of microshells experimented on Laser Megajoule using X-Ray tomography

18th World Conference on Nondestructive Testing, April 2012, Durban, South Africa

RELIABILITY OF PARAMETRIC ERROR ON CALIBRATION OF CMM

DEVELOPMENT OF POSITION MEASUREMENT SYSTEM FOR CONSTRUCTION PILE USING LASER RANGE FINDER

Arion: a realistic projection simulator for optimizing laboratory and industrial micro-ct

Applying Hounsfield unit density calibration in SkyScan CT-analyser

Self-formation, Development and Reproduction of the Artificial System

Creating a Multi-material Probing Error Test for the Acceptance Testing of Dimensional Computed Tomography Systems

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Application of 450 kv Computed Tomography to Engine Blocks with Steel Liners

Image Acquisition Systems

Translational Computed Tomography: A New Data Acquisition Scheme

Tomographic Reconstruction

DUAL energy X-ray radiography [1] can be used to separate

Separation of CT Nominal-Actual Comparisons for Compensation of Plastic Injection Molds

3D Surface Reconstruction of the Brain based on Level Set Method

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

Discrete Estimation of Data Completeness for 3D Scan Trajectories with Detector Offset

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system

The development of a laser array measurement system for three dimensional positioning testing in machine tool

Temperature Distribution Measurement Based on ML-EM Method Using Enclosed Acoustic CT System

APS Sixth Grade Math District Benchmark Assessment NM Math Standards Alignment

Interactive Math Glossary Terms and Definitions

Adapted acquisition trajectory and iterative reconstruction for few-views CT inspection

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Investigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications

ULTRASONIC TESTING AND FLAW CHARACTERIZATION. Alex KARPELSON Kinectrics Inc., Toronto, Canada

Section 7.2 Volume: The Disk Method

Natural Viewing 3D Display

Enhanced material contrast by dual-energy microct imaging

C101-E137 TALK LETTER. Vol. 14

acute angle An angle with a measure less than that of a right angle. Houghton Mifflin Co. 2 Grade 5 Unit 6

The Geometry of Solids

Multi-spectral (W, Mo, Cu, and Ag) XCT measurements

Chapter 24. Wave Optics. Wave Optics. The wave nature of light is needed to explain various phenomena

Application of MCNP Code in Shielding Design for Radioactive Sources

Chapter 3. Physical phenomena: plane parallel plate. This chapter provides an explanation about how rays of light physically behave when

NDT OF SPECIMEN OF COMPLEX GEOMETRY USING ULTRASONIC ADAPTIVE

A method and algorithm for Tomographic Imaging of highly porous specimen using Low Frequency Acoustic/Ultrasonic signals

Investigation on reconstruction methods applied to 3D terahertz computed Tomography

f. (5.3.1) So, the higher frequency means the lower wavelength. Visible part of light spectrum covers the range of wavelengths from

CLEAR EDDY CURRENT FlAW IMAGE USING COMPUTERIZED TOMOGRAPHY. Hiroshi Hoshikawa and Kiyoshi Koyama

RECONSTRUCTION OF REGISTERED RANGE DATA USING GEODESIC DOME TYPE DATA STRUCTURE

DEVELOPMENT OF PREVENTIVE MAINTENANCE SYSTEM ARISING IN ADVANCED EDDY CURRENT TESTING USING NETWORK TOMOGRAPHY

MEDICAL IMAGING 2nd Part Computed Tomography

SIMULATION OF ARTIFICIAL SYSTEMS BEHAVIOR IN PARAMETRIC EIGHT-DIMENSIONAL SPACE

doi: /

Dimensional Measurement of Nuclear Fuel Pellets using High Energy X-ray CT

DEVELOPMENT OF 3D BIT-MAP-BASED CAD AND ITS APPLICATION TO HYDRAULIC PUMP MODEL FABRICATION

Optiv CT160 Accurate measurement for every detail. Computed Tomography System


AFMG. EASE Seminar September 17 th to 21 st 2018, Berlin, Germany. Agenda. Software-Engineering Research Development

Detection of Narrow Gaps Using Hessian Eigenvalues for Shape Segmentation of a CT Volume of Assembled Parts

Metal Artifact Reduction for Multi-Material Objects

Chapter 24. Wave Optics. Wave Optics. The wave nature of light is needed to explain various phenomena

Evaluation strategies in CT scanning

S. Guru Prasad, Ph.D., DABR

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

Similar Pulley Wheel Description J.E. Akin, Rice University

How does the ROI affect the thresholding?

SUPPLEMENTARY FILE S1: 3D AIRWAY TUBE RECONSTRUCTION AND CELL-BASED MECHANICAL MODEL. RELATED TO FIGURE 1, FIGURE 7, AND STAR METHODS.

6. Find the equation of the plane that passes through the point (-1,2,1) and contains the line x = y = z.

Transcription:

More Info at Open Access Database www.ndt.net/?id=18754 Novel evaluation method of low contrast resolution performance of dimensional X-ray CT Makoto Abe 1, Hiroyuki Fujimoto 1, Osamu Sato 1, Kazuya Matsuzaki 1, Toshiyuki Takatsuji 1 1 National Metrology Institute of Japan (NMIJ/AIST), Umezono 1-1-1, Tsukuba Central-3, Tsukuba, Ibaraki, 305-8563, JAPAN, e-mail: Makoto.abe@aist.go.jp Abstract X-ray CT used for dimensional measurement has been attractive due to its nature of visualizing not only external features of an object to be measured but also internal features. Various influencing characteristics have been studied to establish the metrological traceability and to improve the practical performance. These include e.g. spatial resolution, especially under a condition with lower image contrast. A novel design of a material standard which provides quantitative evaluation possibility of resolution performance under low contrast condition of X-ray CT is proposed. The material standard consists of a tab to be filled with liquid continuously variable in its density, and a number of solid cylinders finished for the surface and the form to be submerged into the tab, which have both a number of discretely variable dimensions and discretely variable X-ray absorption. As the result of the combination, desired low contrast density boundary around the cylinders can be realized. This paper presents the design features of the material standard and the experimental results obtained on a dimensional X-ray CT. The preliminary test results demonstrate possibility of the material standard for testing some of performance of dimensional X- ray CT. Keywords: dimensional metrology, X-ray computed tomography, contrast, resolution, 1 Introduction Use of industrial X-ray CT has been expanded due to advancement of performance of the instruments and spread of digital manufacturing across the globe. Particularly dimensional X-ray CT is further expected to become not only an alternative but also a breakthrough to realize non-destructive three dimensional geometric measurements for both external and internal features of industrial products and the parts [1]. Within various performance indices of dimensional X-ray CT, spatial resolution or the detectability of X-ray CT can be recognized as a conventional subject [2]. Different from X-ray CT for medical usage, typical applications of industrial X-ray CT usually show relatively higher contrast in the reconstructed volume image. Therefore extremely low contrast range, which is concern on medical and biological usage of X-ray CT, doesn t have to be considered. However, existing material standard designs for testing spatial resolution usually focus in relatively high contrast range. It recalls motivation of authors to propose a novel material standard design for testing spatial resolution performance of industrial X-ray CT in relatively lower contrast range with ease to use. Authors propose a material standard design composed of cylinder features and liquid medium continuously adjustable for its density. The cylinders are prepared with several different materials and several different sizes in combination. The liquid medium having density range from 1 g/cm 3 up to 3.1 g/cm 3 is chosen. As the consequence, contrast noise ratio (CNR) arbitrary between +40 down to -10 can be realized in case the material standard is applied to high energy X-ray CT. The proposed material standard is prototyped and experimentally scanned by using a conventional 6 MV X-ray CT with fan beam configuration to observe its basic function. Although difficulty in evaluating detectability in quantitative sense under a very low contrast condition can be pointed out, benefit resulted by the proposed material standard for it to be able to realize testing capability of spatial resolution or detectability of dimensional X-ray CT under relatively lower contrast range is believed to be demonstrated. 2 Design and realization of material standard Several different material standards designed for testing spatial resolution of X-ray CT have commercially been available. The typical structure may be composed of a bead, a wire, or a line pair held in a medium having relatively low density. In case the structure is once molded in the medium, it is difficult for users of the material standard to vary dimension of the feature and difference in attenuation of X-ray in the respective material which build contrast with in reconstructed volume image. Major issues in designing a material standard to be used for testing spatial resolution performance of X-ray CT can be: Selection of feature and structure suitable for spatial resolution and detectability, and Intended contrast range and the adjustment possibility to vary contrast in reconstructed volume image. Contrast in X-ray CT image under a specific measurement condition, including the X-ray energy, is attributed to difference of linear attenuation coefficient between two materials [2]. A possibility to realize the intended contrast is to put two solid www.3dct.at 1

materials each having different X-ray attenuation coefficients side by side. Some existing designs have adopted it, e.g. by molding small feature made of dense material into resin or similar [3, 4]. The other possibility may be to replace one of the materials by liquid [5] although preceding study focuses on medical and clinical usage. It is advantageous to realize a material standard which may provide variety of different contrast values by varying density of the liquid. Several candidates for heavy liquid may be found if toxicity was not questioned, e.g. Bromoform (density up to 2.9), Methylene Iodide (up to 3.3), Tetrabromoethane (up to 3.9), or Clerici solution (up to 5). Certainly it is desirable to have atoxic heavy liquid for safe and ease to use, and Sodium Poly Tungstate (SPT) has been chosen. Its density can easily be adjusted in the range from 1 up to 3.1 by varying the proportion in water solution. It is noted that preparation of SPT water solution is relatively easy and the succeeding adjustment of the density can also easily be performed by using a commercially available density measuring instrument. Materials of the cylindrical features are determined in accordance with the possible density range realized by PST water solution, i.e. Polyethylene (PE) (density 0.98), Polyphenylenesulfide (PPS) (density 1.35), Boron Nitride (BN) (density 2.1), Macor (density 2.52), Aluminum (Al) (density 2.8), and Titanium (Ti) (density 4.5). The representative specification of the prototyped material standard is summarized in Table 1. Table 1 Representative specification of prototyped material standard. Size of cylindrical features to be detected by X-ray CT is determined. Typical resolution performance of MV class X-ray CT may reach around 0.5 or up to 1 line pair by the modulation transfer function with 50% threshold. Range in size of cylinders to be prepared for the material standard is determined to cover from 10 mm down to 0.5 mm in diameter. The cylindrical features are allocated on the base plate of the material standard having the maximum pitch circle diameter of 100 mm in rotation symmetric manner as depicted in Fig. 1. The base plate is so designed as the cylindrical features to be submerged into the tub which can be filled with SPT water solution or with the air. Fig. 1 Overview of cylindrical features allocated on base plate of prototyped material standard. Figure 2 shows photographs of the assembled material standard with: 1) the air, 2) water, and 3) SPT water solution having density of 1.08 respectively. It is noted that submerging operation of cylindrical features allocated on the base plate of the material standard into SPT water solution may not cause problem of air bubble potentially adhered around the cylinder feature even in case of SPT water solution density to be exceeding 3.0. www.3dct.at 2

Fig. 2 Prototyped material standard assembled by cylindrical features submerged in medium with different density. 3 Preliminary experiment A preliminary test measurement of the prototyped material standard is performed by using a conventional high energy X-ray CT with beam energy of 6 MV. The representative specification of the X-ray CT system is summarized in Table 2. Table 2 Representative specification of X-ray CT used for preliminary experiment. To perform a series of preliminary experiment, SPT water solution density is determined as depicted in Table 3. The representative operational condition for the experiment is as shown in Table 4. Table 3 SPT water solution density for adopted for experiment Table 4 Representative operational condition of experiment. 3.1 Reconstructed image of prototyped material standard The experimentally scanned and reconstructed images are shown in Fig. 3. Respective figures depict the volume image sliced along a plane perpendicular to the nominal axis of the cylinder features installed in the prototyped material standard. The plane is nominally perpendicular to the rotation axis of the X-ray CT as well. It is seen that background brightness in the material standard increases as density of liquid medium around the cylinder features increases. Most bright circles around the top of the image correspond to the cylindrical features made of Ti, and counter clockwise to Al, Macor, BN, PPS, and PE respectively. www.3dct.at 3

Fig. 3 Experimentally obtained scan image of prototyped material standard with different medium densities. 3.2 Evaluation by signed contrast noise ratio Signed contrast noise ratio defined by equation (1) is evaluated on respective reconstructed image of the prototyped material standard with different SPT water solution density. (1) Where is averaged CT value collected in the area corresponding to the cylinder features with the size of 10 mm in diameter. The bounding circular area for the averaging operation is 7 mm in the diameter to exclude fluctuating portion. is averaged CT values collected in the area corresponding the background medium close to the cylinder in interest and having square area of 12.8 mm by 12.8 mm. and are standard deviation of the CT values. Figure 4 summarizes the experimental result of signed contrast noise ratio in respect of medium density of SPT water solution. It is seen that the SCNR values in the range from +43 down to -12 is realized by using the prototyped material standard. Although plots are sparse in the figure, it is noticed that arbitrary contrast value can easily be adjusted by controlling density of the medium. Fig. 4 Summarized result of signed contrast noise ratio plotted with respect to medium density. www.3dct.at 4

4 Conclusion A design of material standard for testing low contrast resolution and detectability suitable for MV class high energy X-ray CT is proposed. The prototyped material standard is realized and scanned by using a conventional high energy X-ray CT. Benefit resulted by the proposed material standard for it to be able to realize testing capability of spatial resolution of high energy X- ray CT under relatively lower contrast range is believed to be demonstrated. Succeeding study will be continued for evaluating spatial and contrast detectability based on analysis of modulation transfer function although difficulty in evaluating detectability stably under a very low contrast condition can be pointed out. Acknowledgements Authors appreciate effort paid for a series of experiments and data processing performed by Ms. M. Ro, and cooperation made by Mr. K. Sato, Hitachi Corporation. The study is partly supported by METI (Ministry of Economy, Trade and Industry in Japan) project, Development of High energy, high resolution X-ray CT. References [1] T. Takatsuji, M. Abe, and H. Fujimoto, Dimensional X-ray CT in Japan, development, application and standardization, Proc. ict2014, pp. 169-180, 2013. [2] ISO 15708-1:2002, Non-destructive testing - Radiation methods - Computed tomography Part 1: Principles. [3] http://www.material standardlab.com/catphan-material standards/ [4] https://www.kyotokagaku.com/jp/educational/products/detail03/ph-9.html [5] N. Kasraie et. al., Characterization of atherosclerotic plaque: a contrast-detail study using multidetector and cone-beam computed tomography, Journal of applied clinical physics, Vol. 15, No. 1, p. 290, 2014 www.3dct.at 5