General Information for Automated Inova 500 MHz NMR (Indy)

Similar documents
NMR Training VNMRJ 4.2A/VNMRS 400

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry

Agilent VnmrJ 3.2. Quick Start Guide

c. Saturday and Sunday: 10 minute time blocks and unlimited time.

VnmrJ Walkup. Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No , Rev. 0604

VnmrJ Walkup. Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software Pub. No , Rev. A 0808

Varian Solution NMR Automation Procedure

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS

North Carolina State University Department of Chemistry Varian NMR Training Manual

Varian Solution NMR Procedure

ChemPack Queue. Automation Queue. Walkup Queue. Data Review

Varian Solution NMR Automation Procedure (VNMRS Machines running VNMRJ 4.2 under Red Hat Enterprise 5.1)

Bruker Avance 400 MHz Instructions

UC DAVIS NMR FACILITY VNMRJ SHORT MANUAL FOR VARIAN/AGILENT NMR SPECTROMETERS

Basic 1D Processing. Opening Saved Data. Start the VnmrJ software. Click File > Open.

Acquiring Data in VnmrJ 4.2A

NMR Spectrometer Crib-Sheet

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer

1. Log in at the RESLOG terminal by entering your code and select the instrument you want to use.

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751)

Vanderbilt Small Molecule NMR Center

PINMRF. Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy

DCIF NMR Training Guide Varian Mercury 300 MHz Last Update

Gemini-300, Manual Operation (version 1.0, 4Feb95) 1

EQNMR Facility Rules of Use

PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University

Agilent VnmrJ 3.2 Automation

Bruker NMR Topspin Manual

Quick Guide to Topspin

Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300

Troubleshooting Acquisition Related Problems

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800

NMR Users Guide Organic Chemistry Laboratory

Frequently Asked Questions (FAQ)

Bruker AVANCE III HD 400 MHz Spectrometer Operation

MNova Version Installation Instructions.

NMR Experiment Procedure

Varian NMR Instructions - 1D. Department of Chemistry NMR Facility University of Minnesota

Open Access Manual for Superusers

1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR. Spectrometer without the Sample Changer and IconNMR

DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400

NMR Experiment Procedure

Seized Drugs Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division

NMR Course /13/ :31:47 PM Emory University

DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015

Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised

Bruker Topspin NMR Training Manual

GPS Explorer Software For Protein Identification Using the Applied Biosystems 4700 Proteomics Analyzer

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin.

TIDY LABELS. User Guide

NMR Spectroscopy with VNMRJ & EZ NMR

ChromQuest 5.0 Quick Reference Guide

TraceFinder Analysis Quick Reference Guide

WallSign TASKE Call Center Management Tools Version 7.0. Table of Contents TABLE OF CONTENTS TASKE WallSign Startup and Shutdown...

Processing data with Bruker TopSpin

ONE DIMENSIONAL NMR SPECTROSCOPY. DPX 200 and DPX 300

IS2000. Administrative Operator s Guide

ADVANCED DATA PROCESSING

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer

Chemical Shift Perturbation Tutorial

UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X

NMR Data Acquisition and Processing Procedure

Today s Training Agenda. MMS website: FMAX website:

DCIF FAQ: Frequently Asked Questions Varian Last Update (2/21/2005)

Netzsch TGA Training Notebook

Version Beta, pre-release. zspace Studio Users Guide

Introduction to Microsoft Office PowerPoint 2010

DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015

Basic Procedure 1D NMR Data Collection

Training Manual For Chemistry NMR Facility Spectrometers

Multivariate Calibration Quick Guide

RemoteMRC. Introduction Connecting to a Workstation Connecting to a Workstation Terminal Troubleshooting... 9

MestReC Cheat Sheet. by Monika Ivancic, July 1 st 2005

TraceFinder Analysis Quick Reference Guide

Determination of Drag Coefficient

Goldfish 4. Quick Start Tutorial

SchoolDesk University

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection

Vnmrj 3.1 New and improved features

Analyst QS. Administrator s Guide. Part Number: B July 2004

TopPoint User's Guide

The Faces Scheduling System

MyMack Version 9 Upgrade Details

Setup Examples. RTPView Project Program

Table of Contents. 1. Intended Use Overview Materials Required System Requirements User Mode 4

Sciex QTrap Operational Steps for Trained Personnel

Fitting NMR peaks for N,N DMA

SMART Recorder. Record. Pause. Stop

Agilent 6400 Series Triple Quadrupole LC/MS System

Chromatography Software Training Materials. Contents

Shotcut would be suitable for users who are comfortable using MovieMaker.

In this guide you will learn how to:

REACH SCREEN SYSTEMS. System Support Manual. User manual for operating the REACH Announcement Tool, Scheduling Tool, and Touch Screen Systems.

PHYSICIAN S OFFICE STAFF Instructions for Paragon s WebStation for Physicians

CHEMISTRY DEPARTMENT Category Work Instruction

Transcription:

General Information for Automated Inova 500 MHz NMR (Indy) Primary Contacts: NMR Facility Manager: David VanderVelde, x3004, davidv@caltech.edu GLA: Pamela Tadross, x6131, 646-220-5789 (cell), 204 Church, ptadross@caltech.edu Additional contact: NMR Facility Spectroscopist: Scott Ross, scottr@caltech.edu The machine is currently capable of running 1 H, 13 C, 19 F, and 31 P samples. Do NOT run experiments on any other nuclei until you have been informed by the facility managers that the probe has been properly calibrated. Inova 500 MHz NMR Computer Terminal: Dell Precision, Intel Pentium 4 Operating System: Linux NMR interface: VnmrJ 2.1B Sample Preparation: All NMR tubes must be 8 inches in length, 5 mm in diameter. The autosampler can break tubes that are larger or smaller. Samples should contain 5 cm of solvent to ensure proper autolocking and autoshimming. Sign Up Rules: Webcal sign-up for the automated 500 MHz NMR is located at: http://chemex.che.caltech.edu/cgi-bin/webcal-3/webcal.cgi?function=webday&cal=indy and can be found by navigating from http://chemex.che.caltech.edu/cgi-bin/webcal-3/webcal.cgi. The time limits are as follows: Every day 8 am 8 pm 1 hr Every evening 8 pm 12 am 2 hr Every night 12 am 8 am 8 hr Please sign up for all experiments run on this machine, both daily and nightly, as you would when using Hg1 or Hg2.

Start Up: Click on the VnmrJ icon on the toolbar along the bottom of the desktop, or type vnmrj in the terminal window. Login: On the login screen, select your research group from the Operator drop down menu. None of the research group operators require a password, so leave this space blank. Click OK. VnmrJ Interface: Sample Loading: Before loading your NMR tube into the rack, check the sample tray menu in VnmrJ to see which slots are available. They are color coded: Gray Green Blue Yellow Purple Red Faded / Greyed out Available Completed Active Submitted Day Queue Submitted Night Queue Error Sample submitted by an operator other than the one currenly logged in

Load your sample into a spinner, adjust its height using the depth gauge, wipe the bottom with a kimwipe, and load it into an open slot. Please leave the spinners in the large test tube rack when not in use. If there are no available spinners, remove an NMR tube from a slot that has already been run, place it in the small test tube rack next to the autosampler tray, and use that spinner. Do NOT submit to slot #1. This is reserved for the standard. If the autosampler malfunctions, it can be turned off by pressing the RED button. Sample Submission: Once your sample is in the rack, click on the corresponding slot number on the sample tray in VnmrJ. An animated circle should appear around that number. Select your experiment from the list on the lefthand side of the screen. There are 4 tabs (Std 1D, Homo 2D, Hetero 2D, Sel 1D), each of which contains a list of experiments. Click on the experiment you want to run and it will be added to the StudyQ in the white space below. The title will read new sample until the sample has been submitted to the queue, at which point the title will become the location in which the FID will be autosaved. Click all the experiments you want to run on the single sample you have selected. They will be listed underneath the title new sample. In order to modify the acquisition parameters of an individual experiment, you must first double click on the experiment name (e.g. [1 min] Proton). If this is done correctly, the font for that experiment should change to bold and italics (e.g. [1 min] Proton [1 min] Proton). You may now make changes to the information under the Start and Acquire tabs directly to the right of the StudyQ. Start tab: Solvent: select the deuterated solvent for your sample from the drop down menu Sample: text entered in the fourth line of the parameter array on the printout. Comment: same as Text command, add text to top left corner of spectrum Acquire tab: Spectral width (ppm) Relaxation delay (sec) Number of scans (nt) Click Show Time to see how your changes affect the acquisition time. To remove an experiment from the StudyQ, click and drag it to the trashcan in the lower lefthand corner. Clicking on the blue experiment title (e.g. Proton) will remove that experiment for the chosen sample. Clicking on the new sample title will remove the entire list of experiments for that sample, effectively removing the sample itself from the queue.

When you are finished modifying the parameters of all the experiments for a given sample, you are ready to submit that sample to the DayQ or NightQ. Click Submit DayQ / Submit NightQ. The new sample title will change to the name of the directory to which your data will be saved and a padlock icon will appear next to each experiment listed under that title signifying that no further changes may be made. NOTE: The StudyQ only shows experiments submitted under the current operator. To view a complete list of all experiments submitted in chronological order, type status in the command line. The progress of the current acquisition may be monitored from the bar at the bottom center of the main page. All entries must be recorded in the logbook. Autosaved Data: As with Hg3, when each acquisition is completed the spectrum will be plotted automatically. In the parameters listed at the top lefthand corner of the plot, you will see three lines that look similar to this: Automation directory: /home/autouser/vnmrsys/data/auto_2006.08.18_01 File: Proton01 Sample id: /home/operator/vnmrsys/data/autosave_files/2006aug1801 The automation directory is where ALL data files are autosaved. However, it will be easier to find your data in your specific operator autosave directory, where only spectra acquired by your operator (e.g., research group) will be saved. This location is shown under the sample ID. So in the example above, the proton FID acquired on August 18, 2006 is saved under the file name Proton01 in the directories listed on the first and third lines. Data Processing: The easiest way to load your FID for processing is to click and drag it from the StudyQ. First click the X at the top lefthand corner of the sample tray, which should reveal a black screen (or possibly the pulse sequence of the most recently run experiment). Click and drag the line that has the experiment title and spectrum icon into the black window to load your spectrum. You may also need to use the wft command to bring your spectrum on screen. To get back to the sample tray, click the same top lefthand button (which will now be a O). Under the Process tab (next to Start and Acquire), you may perform drift and baseline corrections, autofind integrals, adjust the display, etc. All the commands that worked in Vnmr are also available in VnmrJ using the command line (e.g. sp=##p wp=##p, vsadj, isadj, aph, dpf, dpir, dscale, setint(#,#), etc.)

Data Processing Toolbar: The plot commands for VnmrJ are also the same as Vnmr. The following command will plot partial parameters, peak frequencies, integrals, and the scale for the portion of the spectrum currently on screen: Saving Data: pl ppa ppf pir pscale page You may save your processed FIDs as you would on Hg3. Directories have been made for each of the research group operators. Files should be saved in the directory /home/operator/vnmrsys/data. An example of this would be: /home/stoltz/vnmrsys/data/ You may make your own directory within the data/ directory for your research group. To make the directory gradstudent within the NMRfacility group data directory, select File Save and navigate to the data/ directory. Then type in the full path name for the new directory: /home/nmrfacility/vnmrsys/data/gradstudent and press New Dir. You can also do this from the command line by typing: mkdir( /home/nmrfacility/vnmrsys/data/gradstudent/ ) In order to save your processed data to this new directory, select File Save and navigate to the directory. Enter a file name and press Save.

Retrieving Your Sample: If there are no experiments submitted after your own, the machine will idle with your sample in the magnet. If you would like to retrieve it, do so by replacing it with the D 2 O standard sample. This is accomplished by clicking the Standard button on the toolbar on the top of the main screen. NOTE: You may receive the error message seqfil info cannot be found when attempting to load the standard sample using the Standard button. If this happens, type s2pul in the command line and click the Standard button again. Switch Operator (Logout): When not in use, VnmrJ should idle on the login screen. This can be done by selecting File Switch Operator from the main page. Do NOT exit VnmrJ. Problems / Emergencies: Please contact David VanderVelde (x3004) or Pamela Tadross (x6131 or 204 Church) if there are any problems with the NMR.