INSTRUCTION MANUAL. AquiStar GDL. Dissolved Oxygen Datalogger

Similar documents
INSTRUCTION MANUAL. Reading. AquiStar Smart Sensors. With an INW Panel Meter

INSTRUCTION MANUAL. AquiStar Turbo. Turbidity/Temperature Smart Sensor and Datalogger

INSTRUCTION MANUAL. AquiStar Aqua4Plus. Control Software for INW Smart Sensors

INW Multi-Parameter. Smart Sensor and Datalogger ISO USA PROUDLY MADE IN THE

INSTRUCTION MANUAL. AquiStar PT2X-BV. Barometric/Vacuum Sensor and Datalogger

INSTRUCTION MANUAL. AquiStar DO2. Dissolved Oxygen Smart Sensor and Datalogger

Aqua4Plus 2.0. Control Software for LevelSCOUT, BaroSCOUT, PT2X, CT2X PROUDLY MADE IN THE USA

INW Turbo Tubidity Smart Sensor with Data Logging

INSTRUCTION MANUAL. AquiStar T1. Temperature Sensor

Turbo. Tubidity Smart Sensor with Data Logging ISO USA. Precision Environmental Sensors An ONICON Brand PROUDLY MADE IN THE

DO2. Dissolved Oxygen Smart Sensor and Datalogger Instructions ISO USA. Precision Environmental Sensors An ONICON Brand PROUDLY MADE IN THE

PD10. Pulse Divider Instructions PROUDLY MADE IN THE USA

BaroSCOUT. Barometric/Temperature Smart Sensor and Datalogger ISO USA. Precision Environmental Sensors An ONICON Brand PROUDLY MADE IN THE

INW LevelSCOUT Level/Temperature Smart Sensor and Datalogger

Aqua4Plus. Control Software for Seametrics Smart Sensors ISO USA. Precision Environmental Sensors An ONICON Brand PROUDLY MADE IN THE.

Turbo. Tubidity Smart Sensor with Data Logging ISO USA. Precision Environmental Sensors An ONICON Brand PROUDLY MADE IN THE

BaroSCOUT. Barometric/Temperature Smart Sensor and Datalogger ISO USA. Precision Environmental Sensors An ONICON Brand PROUDLY MADE IN THE

INSTRUCTION MANUAL. AquiStar PT2X. Pressure/Temperature Smart Sensor and Datalogger

User s Guide. OM-CP-PRHTEMP2000 Pressure, Humidity and Temperature Data Logger with LCD

Operating Bulletin MODBUS-RTU. The Fastest Flow Controller Company in the World!

PT35. Dual Pulse Timer Instructions. Precision Flow Measurement An ONICON Brand PROUDLY MADE IN THE USA

Global Water Instrumentation, Inc.

INW T1/T1R Submersible Temperature Smart Sensor

ES-600 Ozone Controller Operation Manual

AccuTrac Multi-Point Data Acquisition Software User Guide

SLUG DATA-LOGGER MANUAL. YieldPoint Inc. JULY 2009

PRODUCT CONTENTS... 3 II. REQUIRED MATERIALS... 3 III. OVERVIEW... 3 IV. BENCH LAYOUT...

User Manual TLS HDMI Switch Splitter 3/2

T1/T1r. Submerible Temperature Smart Sensor ISO USA. Precision Environmental Sensors An ONICON Brand PROUDLY MADE IN THE

Eco Sensors OZONE CONTROLLER Model OS-6 Instructions for Use. General and New Features

LevelSCOUT. Level/Temperature Smart Sensor and Datalogger ISO USA. Precision Environmental Sensors An ONICON Brand PROUDLY MADE IN THE

HD-1X4-4K User Manual

HDTVMT User Manual

MODEL 245 / 345 High Range

Model 282 Total Interface Isolation Module Reference Manual Rev. A Print Table of Contents 1.0 General Description 2.

OZONE SWITCH Model OS-6. OS-6 Features

Eco Sensors OZONE CONTROLLER Model OS-6 Instructions for Use. General and New Features

OPERATING AND SERVICE MANUAL. Universal Interface Device 47

ACLD Channel Opto-isolated Digital Input Board

DTG for Retorts. Style FH3 or FH4

User s Guide. OM-CP-HITEMP140-M12 High Temperature Data Logger with an M12 Probe Connector

MODEL 245 / 345 DigiComp

NOTES TABLE OF CONTENTS

Model 8732 IAQ-CALC TM Indoor Air Quality Meter

OPERATING AND SERVICE MANUAL. Universal Interface Device 47

VP Process Inc. Model: VP-EC-RDU Modbus RTU LCD Display

It s Under Control. Installation and Operation Guide CPB-1. Control Port Connecting Block V 1.1

Smart-Box Series. Electronic Lockbox User Manual. Model: 7000,

V-9908 MESSAGE/PAGE PANEL

PIM-Mini Pulsed Current Source Operation Manual

FA-2448 SIX POSITION Filter Wheel

Woolich Racing. USB ECU Interface User Guide

Mobile: Fax:

MTX-A Temperature Gauge User Manual

Model 206 Operating Instructions. Setra Systems, Inc. 159 Swanson Road, Boxborough, MA

Portable Wireless Infrared Printer (w/ Disposable Batteries) Part No

SC532 9-Pin Peripheral to RS232 Interface

6001TA Terminal Adapter Installation Instructions

MSD IGNITION FOR THE BOMBARDIER DS 650 QUAD PN /PN 4205

REMOTE FRONT PANEL EMULATION

SD1306. Speed Dome IP Camera. Quick User Guide

Xxxxx Xxxxxx OM-CP-QUADVOLT. OM-CP-OCTVOLT 8-Channel Low Level DC Voltage Data Logger. 4-Channel Low Level DC Voltage Data Logger INSTRUCTION SHEET

User Guide. DMXSM-H18 Splitter. DMXSM-H81 Merger. Document Release Aug Revision B

DICKSON FH320/325 & FT300/325 DICKSON. Temperature/Humidity Recorder. Specifications. Applications & Product. Product. Contents: Getting Started

DI-148U. User's Manual. The way PC-based instrumentation should be. 8-Channel USB Data Acquisition Starter Kit. Manual Revision B

LAUNCH TECH CO., LTD.

DICKSON MP100 DICKSON. Operation Manual. Applications. Product. Specifications. Product. Contents:

AE14-503J SERIES PRODUCTION DISPLAY SYSTEM

IAQ-CALC TM Indoor Air Quality Meter

MPP200 User s Manual

SP-400 Portable Water Analyzer Operation Manual. Rev. A Firmware version 1.02

DICKSON TC200/TH300 DICKSON. Handheld Temperature and Temperature / RH Indicators. Useful Features. Applications & Product. Specifications.

INSTALLATION INSTRUCTIONS 5" SINGLE CHANNEL ULTIMATE TACH

VELOCICALC Air Velocity Meter

MULTIPLEXER MX4. Operation and Installation Manual. Pacific Micro Systems

RTD SAMPLER OPERATOR S MANUAL

LevelSCOUT. Level/Temperature Smart Sensor and Datalogger ISO USA. Precision Environmental Sensors An ONICON Brand PROUDLY MADE IN THE

Data Reporter. Installation-Operation E rev.f

Exposure Monitoring. Model 8560 INSPECTAIR CO 2. Meter. Operation and Service Manual , Revision B June 2006

MODBUS MANUAL For the AccUView UV Transmission Monitor

MEC-USB-M002. User s Manual

TB-24P/24R/16P8R. Relay Outputs or Isolated Inputs Terminal Board Series

MEC-LAN-M002. User s Manual

THECHARGEHUB.COM. User Manual. For Square & Round Models

ACT-IR100M+/100M IrDA IR Printer Adapter

BT-HP1 OWNERS MANUAL

USER MANUAL. MODEL 2014 Interface Powered RS-530 to V.35 Converter. SALES OFFICE (301) TECHNICAL SUPPORT (301)

CONTROLLER (Dimensions are approximate) CTS HAND HELD II

½ MicroJet HRP Printhead: (part# -MJHRP) $1, Printhead, Mounting Bracketry, Software, Instruction, Cables.

Operations Manual EcoSense ph100

MEC-COM-M114. User s Manual

The HPV Explorer is a serial communication program used to work with the HPV 600/900 s parameters via a computer s serial port

SP-C1 Mobile Docking Station Installation Guide

MSD Three Stage Delay Timer PN 7760

PCM-7140 Pulsed Current Source Operation Manual

TM0-1 AMPLIFIER / CONDITIONER MODULE OPERATORS MANUAL. Transducer Techniques R

User Manual for XL-J1939

Model 8702 DP-CALC Micromanometer

MEC-SAT-M002. User s Manual

Troubleshooting Tips & Procedures. Model 8800 series. GeoNet Wireless

Transcription:

INSTRUCTION MANUAL AquiStar GDL Dissolved Oxygen Datalogger

Table of Contents 1 Introduction... 3 What is a GDL?... 3 Initial Inspection and Handling... 3 Do s and Don ts... 3 Getting Started... 4 GDL Connectors... 4 Connecting Your DO Sensor to the GDL... 4 Connecting GDL to Computer... 4 Aqua4Plus Software... 5 Warm Up Time... 5 Output... 5 Calibration... 5 Dissolved Oxygen... 6 Temperature... 7 Maintenance... 8 Changing Batteries... 8 Appendix A: Using USB to Serial Cables... 8 Reordering Information... 9 Limited Warranty/Disclaimer - AquiStar GDL/DO Datalogger... 10

2 Information in this document is subject to change without notice and does not represent a commitment on the part of the manufacturer. No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopying and recording, for any purpose without the express written permission of the manufacturer. 1997-2009 Instrumentation Northwest, Inc. Registered trademarks and trademarks belong to their respective owners.

Introduction 3 What is a GDL? INW s AquiStar GDL is a datalogger to which various sensors can be attached and then read from INW s Aqua4Plus control software. The GDL can read the sensors in real time or record sessions of data for later upload. This industry standard digital RS485 interface device records thousands of records, operates on low power, and features easy-to-use software with powerful features. The GDL Dissolved Oxygen version connects to our Modbus dissolved oxygen sensor and returns a dissolved oxygen reading in ppm and a temperature reading in either degreees Celsius, Fahrenheit, or Kelvin. The GDL is powered by either an internal 12 VDC supply (eight AA batteries) or an external 12 VDC supply. This supply also powers the dissolved oxygen sensor - supplying power to it only as needed to conserve power. The unit is programmed using a laptop or desktop Windows based computer via its RS485/RS232 connector and easy to use Aqua4Plus software. Once programmed the unit will measure and collect data on a variety of time intervals. Initial Inspection and Handling Upon receipt of your datalogger, inspect the shipping package for damage. If any damage is apparent, note the signs of damage on the appropriate shipping form. After opening the carton, look for concealed damage. If concealed damage is found, immediately file a claim with the carrier. Do s and Don ts Do handle the device with care. Don t install the device such that the box or any of its connectors are submerged. Don t support the device with the connector or with the connectors of an attached cable. Don t bang or drop the device on hard objects.

4 Getting Started GDL Connectors This GDL comes with two connectors - one for connecting the DO sensor and one communication connector for connecting to the computer. Communication Connector Figure 1: Connectors Sensor Connector Connecting Your DO Sensor to the GDL Connect the sensor to the female 5 pin connector. The connector is wired as follows: Connector Sensor wire Pin Color Code Function 1 Red Power 2 Green D- 3 White D+ 4 n/c 5 Black & Shield Grd Connecting GDL to Computer Connect the GDL to your computer s serial port, as shown below. (For USB connections, see Appendix A.) Interface RS485/RS232 Cable Adapter GDL Serial Port Figure 2: Connecting to computer PC or Laptop Computer

Aqua4Plus Software 5 The GDL comes with the Aqua4Plus host software that is installed on your PC or laptop. This software is used to program the datalogger, to retrieve data from the logger, to view collected data, and to export data to external files for use with spreadsheets or databases. Refer to the Aqua4Plus software manual for details on installing and using Aqua4Plus. Warm Up Time The DO sensor requires a warm-up time of five seconds before readings are stable. The GDL board has a programmable field for warm-up time. Set this warm up time by clicking on Warm Up Time... on the Configure menu. Settings are in milli-seconds so set to 5000 ms. Output The dissolved oxygen channel returns readings in ppm. All readings should be positive. Readings below zero indicate an error code. If you get a reading of -90, double check your connection to the sensor. This error indicates that the GDL cannot read the sensor. If you get any other negative readings, contact INW technical support. The temperature channel returns readings in degrees Celsius, Fahrenheit, or Kelvin. Set your desired units from the Display Units selection on the Options menu. Calibration The dissolved oxygen sensor is calibrated at the factory and should rarely need recalibration. If you want to make slight adjustments to the calibration on either the dissolved oxygen channel or the temperature channel, do so from the Field Calibration option on the Configure Menu. (Note that these adjustments are saved on the GDL unit, NOT on the sensor itself. If you remove the sensor and replace it with a different sensor, you should re-do the field calibration or return them to the default values of m=1 and b=0.) If you want to do a field calibration, follow the instructions below. You can select to do either a one- or a two-point calibration.

6 Dissolved Oxygen One-Point Calibration: --Computing Calibration Value-- Place sensor in water. Allow time for sensor to reach equilibrium. Using an accurate alternative measuring device, measure the dissolved oxygen content of the water. In the Ref deg box for the first point, enter this value. Click first Measure button. When readings have stabilized to your satisfaction, click the OK button in the pop-up box. --Applying Calibration Value-- Click the Apply button to apply calibration value. The computed b value will be transferred to the calibration field. Click OK to save the value to the sensor. Two-Point Calibration: --First Calibration Point-- Place sensor in water with first concentration. Allow time for sensor to reach equilibrium. Using an accurate alternate measuring device, measure the dissolved oxygen content of the water. In the Ref deg box for the first point, enter this value. Click first Measure button. When readings have stabilized to your satisfaction, click the OK button in the pop-up box. --Second Calibration Point-- Place sensor in water with second concentration. Allow time for sensor to reach equilibrium. Using an accurate alternate measuring device, measure the dissolved oxygen content of the water. In the Ref deg box for the second point, enter this value. Click second Measure button. When readings have stabilized to your satisfaction, click the OK button in the pop-up box. --Applying Calibration Values-- Click the Apply button to apply calibration values. The computed m and b values will be transferred to the calibration fields. Click OK to save the values to the sensor.

Temperature 7 One-Point Calibration: --Computing Calibration Value-- Place sensor in water. Allow time for sensor to reach thermal equilibrium. Using an accurate alternate measuring device, measure the temperature of the water. In the Ref deg box for the first point, enter this temperature. Click first Measure button. When readings have stabilized to your satisfaction, click the OK button in the pop-up box. --Applying Calibration Value-- Click the Apply button to apply calibration value. The computed b value will be transferred to the calibration field. Click OK to save the value to the sensor. Two-Point Calibration: --First Calibration Point-- Place sensor in water at first temperature. Allow time for sensor to reach thermal equilibrium. Using an accurate alternate measuring device, measure the temperature of the water. In the Ref deg box for the first point, enter this temperature. Click first Measure button. When readings have stabilized to your satisfaction, click the OK button in the pop-up box. --Second Calibration Point-- Place sensor in water at second temperature. Allow time for sensor to reach thermal equilibrium. Using an accurate alternate measuring device, measure the temperature of the water. In the Ref deg box for the second point, enter this temperature. Click second Measure button. When readings have stabilized to your satisfaction, click the OK button in the pop-up box. --Applying Calibration Values-- Click the Apply button to apply calibration values. The computed m and b values will be transferred to the calibration fields. Click OK to save the values to the sensor.

8 Maintenance Changing Batteries Because changing the batteries involves opening the weather-tight seal, this must be done in a clean, dry environment to avoid contamination or moisture damage to the circuitry. The GDL runs on eight AA alkaline batteries. To replace the batteries, remove the four corner screws and lift off the lid. Gently lift out the battery pack. Replace the batteries, and then place the lid on the box, tightening the screws securely. NOTE: The box is directionally keyed. Failure to replace the lid correctly will prevent a tight seal and will result in water leakage. No slot Slot Figure 3: Note keyed slot in box before replacing lid. Appendix A: Using USB to Serial Cables The standard communication cable/rs485-232 adapter that comes with the GDL plugs into a 9-pin serial port on the PC or laptop. Many new computers, especially laptops, do not come with 9-pin serial ports. If you have one of these computers, or if all of your serial ports are in use, you can connect to a GDL using a USB to Serial cable, as shown in figure 4. Interface Cable RS485/RS232 Adapter USB to Serial Cable GDL USB Port PC or Laptop Computer Figure 4: Connection using a USB to Serial Cable

9 USB-to-Serial cables are readily available from many electronics and computer stores, as well as numerous sites on the Internet. INW has tested and recommends the Keyspan USA-19HS. It is available from INW as well as from many sites on the Internet. Install as follows: Plug into USB port. Install the drivers provided with the particular unit. Determine the port number to which the adapter is assigned. Right-click on My Computer. From the popup menu, select Manage to open the Computer Management window. On left panel, click on Device Manager. On right panel, double-click on Ports. A list of active COM ports will be displayed. Note the COM number assigned to the adapter you just installed. For example: Close Manager. Connect to the GDL (figure 4 above). On the Aqua4Plus software, select the COM port noted above. (If you do not see your new COM port in the drop-down box, open the Communications dialog box from the Options menu. Increase the Highest COM port number, up to a maximum of 15.) Reordering Information For sales & service offices, please contact: Instrumentation Northwest, Inc. www.inwusa.com 800-776-9355

10 LIMITED WARRANTY/DISCLAIMER - AquiStar GDL/DO DATALOGGER A. Seller warrants that products manufactured by Seller when properly installed, used, and maintained, shall be free from defects in material and workmanship. Seller s obligation under this warranty shall be limited to replacing or repairing the part or parts or, at Seller s option, the products which prove defective in material or workmanship within ONE (1) year from the date of delivery, provided that Buyer gives Seller prompt notice of any defect or failure and satisfactory proof thereof. Any defective part or parts must be returned to Seller s factory or to an authorized service center for inspection. Buyer will prepay all freight charges to return any products to Seller s factory, or any other repair facility designated by Seller. Seller will deliver replacements for defective products to Buyer (ground freight prepaid) to the destination provided in the original order. Products returned to Seller for which Seller provides replacement under this warranty shall become the property of Seller. This limited warranty does not apply to lack of performance caused by abrasive materials, corrosion due to aggressive fluids, mishandling or misapplication. Seller s obligations under this warranty shall not apply to any product which (a) is normally consumed in operation, or (b) has a normal life inherently shorter than the warranty period stated herein. In the event that equipment is altered or repaired by the Buyer without prior written approval by the Seller, all warranties are void. Equipment and accessories not manufactured by the Seller are warranted only to the extent of and by the original manufacturer s warranty. THE FOREGOING WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES, WHETHER ORAL, WRITTEN, EXPRESSED, IMPLIED OR STATUTORY. IMPLIED WARRANTIES OF FITNESS AND MERCHANTABILITY SHALL NOT APPLY. SELLER S WARRANTY OBLIGATIONS AND BUYER S REMEDIES THEREUNDER (EXCEPT AS TO TITLE) ARE SOLELY AND EXCLUSIVELY AS STATED HEREIN. IN NO CASE WILL SELLER BE LIABLE FOR CONSEQUENTIAL DAMAGES, LABOR PERFORMED IN CONNECTION WITH REMOVAL AND REPLACEMENT OF THE SENSOR SYSTEM, LOSS OF PRODUCTION OR ANY OTHER LOSS INCURRED BECAUSE OF INTERRUP- TION OF SERVICE. A NEW WARRANTY PERIOD SHALL NOT BE ESTABLISHED FOR REPAIRED OR REPLACED MATERIAL, PRODUCTS OR SUPPLIES. SUCH ITEMS SHALL REMAIN UNDER WARRANTY ONLY FOR THE REMAINDER OF THE WAR- RANTY PERIOD ON THE ORIGINAL MATERIALS, PRODUCTS OR SUPPLIES. B. With respect to products purchased by consumers in the United States for personal use, the implied warranties including but not limited to the warranties of merchantability and fitness for a particular purpose, are limited to twelve (12) months from the date of delivery. Some states do not allow limitations on the duration of an implied warranty, so the above limitation may not apply to you. Similarly, some states do not allow the exclusion or limitation of consequential damages, so the above limitation or exclusion may not apply to you. This limited warranty gives you specific legal rights; however, you may also have other rights which may vary from state to state.

1997-2009 by Instrumentation Northwest, Inc. All rights reserved. Instrumentation Northwest and INW are trademarks registered with the U.S. Patent & Trademark Offi ce. Doc# 9B0843r1 05/2009 / PN 6D285-NI INW 8902 122nd Avenue NE Kirkland, WA 98033 USA 425-822-4434 FAX 425-822-8384 / info@inwusa.com