ScienceDirect. AACMM Performance Test: Influence of Human Factor and Geometric Features

Similar documents
ScienceDirect. Study of Virtual Features in the Performance of Coordinate Measuring Arms

Sensor Prototype to Evaluate the Contact Force in Measuring with Coordinate Measuring Arms

Application of a Force Sensor to Improve the Reliability of Measurement with Articulated Arm Coordinate Measuring Machines

OUT OF MACHINE CALIBRATION TECHNIQUE FOR ANALOG PROBES

Monte Carlo method to machine tool uncertainty evaluation

Design and Mechanical Evaluation of a Capacitive Sensor-Based Indexed Platform for Verification of Portable Coordinate Measuring Instruments

TESTING THE CAPABILITIES OF THE THREE-SPHERES ALIGNMENT METHOD FOR LASER TRIANGULATION SENSORS

Measurement procedure for application of white light scanner in the automotive sector

ScienceDirect. The use of Optical Methods for Leak Testing Dampers

EVALUATION OF DIFFERENT PROBING SYSTEMS USED IN ARTICULATED ARM COORDINATE MEASURING MACHINES

Available online at ScienceDirect. Marko Švaco*, Bojan Šekoranja, Filip Šuligoj, Bojan Jerbić

AN EVALUATION OF NON-CONTACT LASER SCANNER PERFORMANCE USING CALIBRATED ARTIFACTS

The Uncertainty of Parallel Model Coordinate Measuring Machine

Investigating the volumetric performance of multi-axis measurement arms

Available online at ScienceDirect. Procedia Engineering 161 (2016 ) Bohdan Stawiski a, *, Tomasz Kania a

The development of a laser array measurement system for three dimensional positioning testing in machine tool

THREE DIMENSIONAL INDUSTRIAL METROLOGY USING X-RAY COMPUTED TOMOGRAPHY ON COMPOSITE MATERIALS

METROLOGY SOLUTIONS THE OPTICAL PORTABLE CMM FOR INDUSTRIAL APPLICATIONS

METHODS FOR PERFORMANCE EVALUATION OF SINGLE AXIS POSITIONING SYSTEMS: POINT REPEATABILITY

THE OPTICAL PORTABLE CMM FOR INDUSTRIAL APPLICATIONS

On the impact of probing form error on form measurement in Computed Tomography

Length, Germany, PTB (Physikalisch-Technische Bundesanstalt)

COMPARISON OF ERROR MAPPING TECHNIQUES FOR COORDINATE MEASURING MACHINES USING THE PLATE METHOD AND LASER TRACER TECHNIQUE

Comparision of contact and contactless measuring methods for form evaluation

LOCAL AND GLOBAL CALIBRATION OF COORDINATE MEASURING MACHINES

19th International Congress of Mechanical Engineering HUURUV 7KHVH WHVWV DUH GHVFULEHG E\ VWDQGDUGV OLNH,62 $PHULFDQ $16,$60( % *HUPDQ 9',9'( DQG

Good Practice guide to measure roundness on roller machines and to estimate their uncertainty

Renishaw touch-trigger probing technology. Rugged and flexible solutions for discrete point measurement on CMMs

Comparison of Probing Error in Dimensional Measurement by Means of 3D Computed Tomography with Circular and Helical Sampling

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system

Contents 1 Measurement and Machine Tools An Introduction

PARAMETERS INFLUENCING THE PRECISION OF SLM PRODUCTION

Improving dimensional accuracy by error modelling and its compensation for 3-axis Vertical Machining Centre

IND62 TIM CALIBRATION OF FREE-FORM STANDARD AND THEIR APPLICATIONS FOR IN-PROCESS MEASUREMENT ON MACHINE TOOLS

Evaluation strategies in CT scanning

The Virtual CMM a software tool for uncertainty evaluation practical application in an accredited calibration lab

Impact of 3D Laser Data Resolution and Accuracy on Pipeline Dents Strain Analysis

A FEW CASE STUDIES IN UNCERTAINTY USING THE NIST M48 CMM

Improvements Metrosoft QUARTIS R12

Innovations in touch-trigger probe sensor technology

Design of three-dimensional photoelectric stylus micro-displacement measuring system

Transactions on Engineering Sciences vol 16, 1997 WIT Press, ISSN

A Comparison of Mechanical Artefacts used in Calibration of Coordinate Measuring Machines (CMMs)

QUALITY CONTROL OF PARTS AND ASSEMBLIES USING SOFTWARE GOM INSPECT BASED ON PORTABLE CMM DATA

Investigation on precise measurement of cutting tool edges using coordinate measuring machines

Comparison ISO to ASME B89.4.1

Available online at ScienceDirect. Energy Procedia 69 (2015 )

A NATIONAL MEASUREMENT GOOD PRACTICE GUIDE A NATIONAL MEASUREMENT GOOD PRACTICE GUIDE. No. 42. No. 42. CMM Verification.

Available online at ScienceDirect. Physics Procedia 56 (2014 ) Moscow, , Russia

DATA SHEET DEA GLOBAL SILVER 09.XX XX.10

ScienceDirect. Reverse Engineering of Parts with Optical Scanning and Additive Manufacturing

GET A WHOLE new measure OF things!

EMRP JRP IND62 TIM: Use of on-board metrology systems for area-scanning on machine tools

RELIABILITY OF PARAMETRIC ERROR ON CALIBRATION OF CMM

EXPLORER CLASSIC BRIDGE COORDINATE MEASURING MACHINE PRODUCT BROCHURE

Horizontal scale calibration of theodolites and total station using a gauge index table

Linear, and segmented over up to 1000 points Squareness calibration Matrix compensation over up to 30 x 30 points

Proposed new tests for evaluating CMM performance

AN UNIFIED APPROACH TO EVALUATE AND COMPENSATE THE TOOL POSITIONING ERROR IN PARALLEL KINEMATIC MACHINES

Fast checking of CMM geometry with a patented tool

A Novel Kinematic Model of Spatial Four-bar Linkage RSPS for Testing Accuracy of Actual R-Pairs with Ball-bar

A Procedure for accuracy Investigation of Terrestrial Laser Scanners

Simulation Supported POD Methodology and Validation for Automated Eddy Current Procedures

good check of volumetric accuracy. However, if the mea error components. However, if the errors measured are large,

Boosting the Efficiency of Wind Power Plants. Stringent Demands for Boosting the Efficiency of Wind Power Plants

Research on error detection technology of numerical control machine tool. Cao YongJie

PROBE RADIUS COMPENSATION AND FITTING ERRORS IN CAD-BASED MEASUREMENTS OF FREE-FORM SURFACE: A CASE STUDY

Precimar. PLM and CiM Universal Length Measuring Machines

arxiv: v1 [physics.ins-det] 21 Aug 2009

MACRO-SCALE PRECISION ALIGNMENT. 3.1 Precision Machine Design Alignment Principles

Precimar. PLM and CiM Universal Length Measuring Machines LENGTH MEASURING MACHINES FOR HIGH-END CALIBRATION

NEW MONITORING TECHNIQUES ON THE DETERMINATION OF STRUCTURE DEFORMATIONS

TESTING VISION CMMS AND THE INTERNATIONAL STANDARD, ISO

PCMM System Specifications Leica Absolute Tracker and Leica T-Products

PCMM System Specifications Leica Absolute Tracker and Leica T-Products

A Precision Grinding Technique for Radome Inner Surfaces. Z.Y. Jia, T. Ji, D.M. Guo and G.H. Bian

COMPACT RANGE CALIBRATION AND ALIGNMENT

Tele-operation Construction Robot Control System with Virtual Reality Technology

Industrial Measuring Technology from Carl Zeiss CONTURA CONTURA G2 RDS G2 AKTIV

A universal method to optimise measurement uncertainty, time and cost for CMM scanning technology

Calibration of Optical Rotary Tables using Autocollimators and Laser Interferometer Systems

12 m. 30 m. The Volume of a sphere is 36 cubic units. Find the length of the radius.

Available online at ScienceDirect. Procedia Manufacturing 6 (2016 ) 33 38

Experimental accuracy assessment of different measuring sensors on workpieces with varying properties

GEOMETRIC CALIBRATION OF CMMs USING 3D LENGTH MEASUREMENTS

ASSESSMENT OF MEASUREMENT UNCERTAINTY CAUSED IN THE PREPARATION OF MEASUREMENTS USING COMPUTED TOMOGRAPHY

CyberGage360 3D SCANNERS. One-Button Automation for 3D Scan Inspection.

Ch 22 Inspection Technologies

Geometrical assessment of solar concentrators using close-range photogrammetry

DIMENSIONAL INSPECTION IN QUALITY CONTROL

CNC Milling Machines Advanced Cutting Strategies for Forging Die Manufacturing

Introduction to Metrology. ME 338: Manufacturing Processes II Instructor: Ramesh Singh; Notes: Profs. Singh/Melkote/Colton

Incorporating Thermal Expansion into CAD-Based 3-Dimensional Assembly Variation Analysis

ScienceDirect. Evaluation of Measuring Capability of the Optical 3D Scanner

Available online at ScienceDirect. Procedia CIRP 24 (2014 ) 1 6

Experimental study on performance verification tests for coordinate measuring systems with optical distance sensors

INFLUENCE OF INITIAL SETUP OF PARTS BEFORE ROUGHNESS EVALUATION

Industrial Measuring Technology from Carl Zeiss. `lkqro^ `lkqro^ do=oap do=^hqfs

Computer aided error analysis for a threedimensional precision surface mapping system

Transcription:

Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 69 ( 2014 ) 442 448 24th DAAAM International Symposium on Intelligent Manufacturing and Automation, 2013 AACMM Performance Test: Influence of Human Factor and Geometric Features Daniel González-Madruga a, Joaquin Barreiro a, Eduardo Cuesta b *, Beatriz González a, Susana Martínez-Pellitero a a Department of Mechanical, Informatics and Aeroespatiale Engineering, University of León, Campús of Vegazana, León 24071. b Department of Manufacturing Engineering, University of Oviedo, Campus of Gijón, Asturias 33203. Abstract Articulated Arms Coordinate Measuring Machines (AACMMs) have spread out in the manufacturing industry thanks to their flexibility and reduced cost. Nevertheless, their performance has been barely studied unlike traditional coordinate measuring machines. Therefore, a lack of traceability and reliability have been found AACMM field. AACMM performance is affected by many factors which are partially studied and compensated. Among them, human factor is one of the main factors with a significant impact on AACMM performance, however it is not considered in current evaluation or calibration methodologies. In this work, a new methodology is presented in order to calculate operator contribution to AACMM errors. Furthermore, operator behavior changes according to the measured feature and type of tolerance. For this reason, a features-based gauge capable of materialize a wide range of tolerances types has been designed and built for using with this methodology. Test results provide with a reasonable basis for performing both operator and AACMM qualification. 2014 The Authors. Published by Elsevier Ltd. 2014 The Authors. Published by Elsevier Ltd. Open access under CC BY-NC-ND license. Selection and peer-review under responsibility of DAAAM International Vienna. Selection and peer-review under responsibility of DAAAM International Vienna Keywords: Coordinate Measuring Arms; CMA; AACMM; operator qualification; coordinate metrology; dimensional gauge 1. Introduction AACMMs have broadened inspection tasks into a more flexible activity of manufacturing process. They not only avoid meticulous measurement planning but also work on quite different environments and tasks such as, production * Corresponding author. Tel.: +034-985-182-136. E-mail address: ecuesta@uniovi.es 1877-7058 2014 The Authors. Published by Elsevier Ltd. Open access under CC BY-NC-ND license. Selection and peer-review under responsibility of DAAAM International Vienna doi: 10.1016/j.proeng.2014.03.010

Daniel González-Madruga et al. / Procedia Engineering 69 ( 2014 ) 442 448 443 lines, workstation, part assembly places, reverse engineering, etc. Their flexibility and portability bring great advantages; nevertheless, they also bring additional error sources that have a deeper impact in AACMM performance than in traditional CMMs. Operator factor reveals as an important error source that contributes to AACMM uncertainty. However, this error source is not considered in current evaluation or calibration methodologies. Vrhovec et al. [1] listed the main error sources and their approximated contribution to AACMM performance. Kinematic and encoder errors are pointed as two of the main error sources followed by thermal errors and elastics deformations. While kinematic errors are dealt by new calibration methodologies in order to improve the kinematic model, manufacturers develop lighter structures and high resolution encoders for reducing encoder errors. Santolaria et al. [2] improved AACMM performance in terms of distances and point repeatability by measuring a ball bar. Furutani et al. [3] calibrated their AACMM with several ball bar combinations. Kovac et al. [4] proposed a linear gauge with an interferometer for AACMM calibration. Gao et al. [5, 6] applied neural networks to their calibration process with kinematic seats. These calibration methodologies involve an overall error optimization, that is, kinematic errors and encoder errors. Also, elastic and thermal errors than happen during calibration are included in the calibration results. Thermal errors were also controlled by Santolaria et al. [7] who carried out their calibration methodology at several temperatures. Consequently, a set of kinematic parameters is calculated according to the environment temperature. Regarding to elastic deformation, weight of AACMM components and contact force remain as non-controlled error sources. Vrhovec et al. [1] also proposed a laser sensor that detects deflection on AACMM links. In a previous work, Cuesta et al. [8] proved a greater contribution of the operator factor in AACMM performance that the one stated by Vrhovec previously. They also identified a considerable impact of other parameter such as part geometry or probe type. In fact, operator factor has a different behavior depending on the measured feature [9, 10] and it should be evaluated. International standards [11, 12] define several tests capable of quantifying AACMM performance at some extent and, therefore, they establish reference values for evaluation over time and comparison with other AACMMs. Both standards provide with similar performance tests which consist of measuring a ball bar within the AACMM workspace. In other tests, a calibration sphere or a kinematic seat are used in order to evaluate point repeatability or the sphere diameter obtained with the AACMM. Piratelli et al. [13, 14] proposed a similar to ASME standard by using a virtual sphere gauge, bar and plate. In addition, in a previous work [15] it was developed a similar methodology with virtual circles instead of virtual spheres. In any case, a positive result of the evaluation would guarantee the validity of subsequent measurement, assuming that the rest of features (cylinders, planes ) are included into the evaluation. However, all these tests are carried out by only one operator, thus the impact into the evaluation could lead to a miscalculation of the AACMM performance. In addition, if the evaluation is carried out by another operator, it would probably bring different results according to his experience, training or measurement ability that could lead to a false acceptance of subsequent measurements. In this paper, a new evaluation methodology capable of revealing operator contribution to the measurement process uncertainty is proposed as well as a true value for AACMM performance. This methodology is based on the considered feature, so evaluation will show the capability of AACMM to measure different features, the mos common in actual manufacturing parts. Table 1. AACMM ROMER Omega technical specifications. Technical specification Model Range Repeatability, Sphere test Point repeatability, Cone test Length Accuracy, Volumetric accuracy test Value Romer Omega (6DOF) 1800 [mm] 0.010 [mm] 0.018 [mm] 0.025 [mm]

444 Daniel González-Madruga et al. / Procedia Engineering 69 ( 2014 ) 442 448 2. Materials and test setup A new methodology capable of evaluating operator effect as well as AACMM performance for several part features has been developed. This methodology is based on repeatability and reproducibility studies. This survey was performed with an AACMM ROMER, Omega series, 2018 model with a measurement range of 1800 mm (Fig. 1a). A "hard probe" of 50 mm long with a ruby sphere 6 mm in diameter was used. At the beginning, the AACMM and the probe were qualified with the manufacturer's software (Table 1). In order to evaluate AACMM and operator performance for each type of tolerance, a features-based gauge was designed and manufactured, Fig. 1b. It includes most common geometrical characteristics: planes, spheres, inner and outer cylinders which can materialize different geometric and dimensional tolerances. This gauge consists of a main length of 1000 mm with T-inverted section. Along this length, 12 parallel planes are determined by 6 crenels. Each plane has enough surface area (about 2 cm 2 ) and accessibility for a proper manual measurement with AACMM. Furthermore, the gauge includes 5 spheres on crenels top. In each crenel 6 inner cylinders have been machined (drilled and bored). In addition, 4 cylindrical parts have been machined and then assembled in the main structure of the gauge. These parts contain 4 outer cylinders and 4 inner cylinders. The features-based gauge was made of aluminum alloy with hard anodized treatment to increase surface hardness. Fig. 1. a) Geometric features-based gauge; b) portable measuring arm used. Geometric features and their combinations provide a wide range of geometric and dimensional tolerances. Table 2 shows tolerance types measured for this study and the associated dimension. The features-based gauge was previously measured with a DEA Global Image 091508 CMM, whose expanded uncertainty is given by MPE E = 2.2+3L/1000 [μm]. These measurements are used as reference values for this work since its repeatability level is far better than AACMM s repeatability (MPE p < 2.2 m). Table 2.Geometric features and tolerances included in the study. Geometric features Number of elements Geometric tolerance Dimensional tolerance Spheres 2 Sphericity Diameter (20 mm) Distance between spheres (960 mm) Outer cylinders 2 Cylindricity Diameter (50 mm) Distance between outer cylinders (700 mm) Parallelism between outer cylinders Inner cylinders 2 Cylindricity Diameter (30 mm) Distance between inner cylinders (900 mm) Parallelism between inner cylinders Parallel planes 2 pair Flatness Distance between planes (1000 mm) Parallelism between planes Perpendicular planes 1 pair Flatness Perpendicularity between planes Inclined planes 2 pairs Flatness Angularity between inclined planes

Daniel González-Madruga et al. / Procedia Engineering 69 ( 2014 ) 442 448 445 PCDMIS software was used to control measurement of AACMM and CMM. Additionally, AACMM and the features-based gauge were mounted on a CMM table at fixed positions. Test measurements were carried out in the CMM laboratory where temperature is controlled within a range of 20 ± 1 ºC. 3. Methodology The evaluation test intention is to calculate operator and AACMM uncertainty when measuring a feature in a part. Because of this, several operators are necessary to carry out the test and to extend the methodology to any kind of tolerance. This way, evaluation results will qualify certain operators for the measurement with a specific AACMM. Also, AACMM capability to measure each type of tolerance is finally determined, which assures AACMM reliability and traceability. Three operators measured the features-based gauge 5 times, with 10 repetitions each time. Given the importance of operator factor, the test analysis use statistical techniques similar to repeatability & reproducibility studies which allow us to evaluate operators. Two ways ANOVA (analysis of variance) compare the mean of several measurement samples, classified into two categories by their variance study. As a result, this technique obtains if any factor is significant and, therefore, affects AACMM measurement. On the way, variance of factors are calculated and used to set the operator and AACMM uncertainty. Measurement results were used to calculate the sum of squares for each factor or group of measurement, eq. 1-5. Sum of squares of operator (OSS) and Sum of squares of gauge (GSS) consider the variability associated to each operator or set of gauge measurements against the total mean. Interaction between operator and gauge factors (ISS) measures the variability of their combined effects. This value is used for comparison against the variability of each individual factor and the total mean. Error sum of squares (ESS) represents the variability which is non-explained by operator or gauge factor. At last, total sum of squares (TSS) shows the variability of the whole test measurements. Letters a, b and r are the number of operators (3 operators), the sets of measurements of the gauge (5 sets) and the repetitions (10), respectively. (1) (2) (3) (4) (5) These values of sum of squares are divided by the degree of freedom of each group in order to obtain the mean squares. The ratio for each mean square and the Error mean squares provides the F parameter. This parameter is introduced in the Fisher distribution to obtain the significance of the factors. The level of confidence is given by p values: p value lower than 0.05 (95% level of confidence) means that the factor is statistically significant. These significant p factors explain the measurement variability and prove that there are differences between operators performance. Table 3 shows the ANOVA methodology for the test analysis. Table 3. ANOVA methodology for the test analysis. Sources Sum of squares Degree of freedom Mean squares F p Operator OSS a-1 OMS=OSS/(a-1) OMS/EMS 0.05 Gauge GSS b-1 GMS=GSS/(b-1) GMS/EMS 0.05 Interaction ISS (a-1)(b-1) IMS=ISS/(a-1)(b-1) IMS/EMS 0.05 Error ESS ab(r-1) EMS=ESS/ab(r-1) Total TSS abr-1 TMS=TSS/( abr-1)

446 Daniel González-Madruga et al. / Procedia Engineering 69 ( 2014 ) 442 448 The non-explained variability due to operator or gauge is assumed as the AACMM contribution to the measurement uncertainty. The total variability is the sum of the variability of each factor. In addition the standard deviation was calculated for each factor so the true contribution of operator and AACMM to the measurement test is obtained. 4. Results Two kinds of results were obtained: direct comparison among operator s measurements and contribution of operator and AACMM to measurement performance in terms of standard deviation. As example, Fig. 2 shows the test results for sphere diameter which includes diameter deviation and frequency of each diameter value. It is observed that each operator is characterized by a different behavior; whilst operator 1 and operator 2 achieve similar curves with different dispersion, operator 3 performance is considerably worse. These results show the most suitable operator for measuring sphere diameters and his measurement dispersion is determined. Frequency 60 50 40 30 20 10 0-0,060-0,040-0,020 0,000 0,020 0,040 0,060 0,080 0,100 Diameter error [mm] Mean: 19.984 mm Std. Dev.: 0.0248 mm Total Operator 1 Operator 2 Operator 3 Fig. 2. Operators performance comparison. Another operator comparison is carried out for each gauge measurement set. Fig. 3 shows the mean value for each gauge measurement set and the standard deviation associated to each operator. The mean diameter of the spheres is in a range of 0.040 mm with a maximum error of 0.021 mm with respect to the mean of the measurements. Regarding dispersion, operators 1 and 2 obtain similar results, about 0.020 mm, and operator 3 reaches up to 0.030 mm. Mean diameter (mm) 20,020 20,000 19,980 19,960 19,940 1 2 3 4 5 Gauge measurement set index Operator 1 Operator 2 Operator 3 Fig. 3. Mean diameter and standard deviation for the 5 sets of measurements of a sphere.

Daniel González-Madruga et al. / Procedia Engineering 69 ( 2014 ) 442 448 447 As an example of the methodology, Table 4 shows the result for sphere diameter tolerance. In this case, both factors, operator and gauge, have a significant impact on the measurement results, although gauge contribution is quite lower than operators. This fact shows that the operator can lead to measurement error regardless the AACMM qualification. In fact, operators should be evaluated and qualified as an important part of the measurement system as well. Table 4. Results of the repeatability and reproducibility for sphere diameter. Main effects Sum of squares Degree of freedom Mean squares F p Operator 0.008403 2 0.004201 8.556152 0.000317 Part 0.007471 4 0.001868 3.803545 0.005803 Interaction 0.009248 8 0.001156 2.354289 0.021104 Error 0.066288 135 0.000491 Total 0.091409 149 0.000613 Furthermore, the variance and standard deviation associated to each factor were calculated. Since the sum of each factor variance is the total variance, unlike standard deviation, the contribution for each factor was also calculated as the percentage of variance, Table 5. It was noted that although gauge factor is significant (p>0.05), its contribution to the total variance is so low that this gauge is suitable for AACMM evaluation. In addition, the contribution of the operator and the AACMM to the measurement performance is quantified. This way AACMM is evaluated subtracting the influence of operator. The values in bold show the highest influences of each column factor. Note that the sum of the four factor percentages gives 100% although the percentage of interaction factor is not showed. Table 5. Contribution of operator and gauge for all measured features. Tolerance σ 2 % AACMM σ 2 % Operator σ 2 % Gauge σ Total variance Sphericity 56.85 33.05 0.00 0.0103 Cylindricity, outer cylinder 34.11 44.97 4.61 0.0165 Cylindricity, inner cylinder 50.79 16.52 9.06 0.0134 Flatness 86.80 12.67 0.53 0.0033 Parallelism between outer cylinders 55.77 0.00 0.00 0.0139 Parallelism between inner cylinders 60.54 22.44 6.78 0.0189 Parallelism between planes 45.20 51.40 3.40 0.0167 Perpendicularity between cylinders 39.50 38.24 0.00 0.0433 Perpendicularity between planes 91.63 2.83 3.65 0.0121 Diameter, sphere 76.47 9.48 3.69 0.0253 Diameter, outer cylinder 22.90 24.85 7.17 0.0446 Diameter, inner cylinder 42.46 10.41 0.00 0.0297 Distance between spheres 9.70 12.19 0.00 0.0641 Distance between outer cylinders 3.86 28.21 2.00 0.1043 Distance between inner cylinders 3.61 20.34 0.00 0.0996 Distance between planes 2.12 27.72 0.00 0.1038 Angularity 29.80 62.79 4.55 0.0590 5. Conclusion Current methodologies for AACMM evaluation are clearly based on previous MMC experience, both gauges and procedures. Usually, only distances between spheres (exceptionally with distances between planes) are measured so it is excluded the evaluation of operator performance, which controls many measurement parameters, measurement technique. In other words, evaluation methodologies do not agree to real measurements carried out in industrial environment. Throughout this paper, we propose a completely different methodology and far more ambitious. On

448 Daniel González-Madruga et al. / Procedia Engineering 69 ( 2014 ) 442 448 one side, a features-based gauge is proposed, manufactured and test it. The use of this gauge allows us to evaluate the capacity of the operator to measuring several geometric features with an AACMM. The corresponding methodology allows for assessing the agreement between AACMM and CMM for different dimensional and geometrical tolerances. In addition, this methodology was repeated with different operators in order to evaluate, by means of ANOVA, which quantity the contribution to total error of the gauge, operator and AACMM. Results of the evaluation test show that this approach is capable of quantifying the operator contribution to measurement performance as well as the AACMM real contribution without the operator influence. For example, it can be said that, talking about geometric tolerances, AACMM error contribution is an important part of total error, while for dimensional tolerances the same factor is considerably lower. Furthermore, operator has a relevant influence in the error form of spheres and cylinders - both outer and inner. Regarding the features-based gauge, the low error contribution indicates that this gauge is suitable for AACMM evaluation. It goes without saying that this features-based gauge was not only capable to evaluate the AACMM and the operator s technique for each type of tolerances, but also allows the operator training. Acknowledgements Authors thank to the Spanish Ministry of Economy and Competitiveness for financial support of the project entitled Quality assurance and knowledge modelling applied to portable coordinate measuring systems (ref. DPI2012-36642-C02-01) and to Instituto Universitario de Tecnología Industrial de Asturias (IUTA) for the support to the project Desarrollo de sensor miniaturizado con sistema de medición de fuerzas para Brazos de Medir por Coordenadas (ref. SV-13-GIJON-1.8). References [1] M. Vrhovec, M. Munih. Improvement of coordinate measuring arm accuracy. Proceedings of the 2007 IEEE/RSJ International Conference on Intelligent Robots and Systems, San Diego, CA, USA, October 29 - Nov 2, 2007. [2] J. Santolaria, J.J. Aguilar, J.A. Yagüe, J. Pastor. Kinematic parameter estimation technique for calibration and repeatability improvement of articulated arm coordinate measuring machines. Precision Engineering, 32 (2008), pp. 251 68. [3] R. Furutani, K. Shimojima, K. Takamasu, Kinematical calibration of articulated CMM using multiple simple artifacts, XVII IMEKO World Congress (2003). [4] I. Kovac, A. Klein, Apparatus and a procedure to calibrate coordinate measuring arms. Journal of Mechanical Engineering, 48-1 (2002), pp. 1732. [5] G. Gao, W. Wang, K. Lin, Z. Chen, Structural Parameter Identification for Articulated Arm Coordinate Measuring Machines. International Conference on Measuring Technology and Mechatronics Automation, 2 (2009), pp. 128-131. [6] G. Gao, W. Wang, K. Lin, Z. Chen. Kinematic Calibration for Articulated Arm Coordinate Measuring Machines Base on Particle Swarm Optimization. Second International Conference on Intelligent Computation Technology and Automation, 1 (2009), pp. 189-192. [7] J. Santolaria, J.A. Yagüe. R. Jimenez, J.J. Aguilar, Calibration based thermal error model for articulated arm coordinate measuring machines. Precision Engineering. 33 (2009) pp. 476 485. [8] E. Cuesta, B.J. Álvarez, S. Martinez, J. Barreiro, D. González-Madruga. Evaluation of influence parameters on measurement reliability of coordinated measuring arms, AIP Conference Proceedings, 1413 (217) (2012), pp 214-224. [9] D. González-Madruga, E. Cuesta, J Barreiro, S. Martinez-Pellitero, B.J. Álvarez. Real-Time contact force measurement system for Portable Coordinate Measuring Arms DAAAM 2012. 23 (1) (2012), pp. 267-272. [10] D. González-Madruga, E. Cuesta, J Barreiro, A.I. Fernandez-Abia. Application of a forcé sensor to improve the reliability of measurement with articulated arm coordinate measuring machines. Sensors, 13 (2013), pp. 10430-10448. [11] ASME B89.4.22-2004. Methods for performance evaluation of articulated arm coordinate measuring machines, (2004). [12] VDI/VDE 2617 Part 9, Acceptance and reverification test for articulated arm coordinate measuring machines, Verein Deutscher Ingenieure, (2009). [13] A. Piratelli-Filho and G.R. Lesnau. Virtual spheres gauge for coordinate measuring arms performance test. Measurement, 43 (2009), pp. 236-244. [14] A. Piratelli-Filho, F. H. Tavares Fernandes and R. Valdés Arencibia. Application of Virtual Spheres Plate for AACMMs evaluation. Precision Engineering, 32-2 (2012), pp. 349-355. [15] D. González-Madruga, E. Cuesta, H. Patiño, J. Barreiro, S. Martínez-Pellitero. Evaluation of AACMM using the virtual circles method. Mesic 5th (2013).