Agilent MSD Productivity ChemStation G1701 & G1710. Performance Report Parameters

Similar documents
Agilent W2100 Antenna Modeling Design System

Agilent OpenLAB Data Store Administration. Guide for Administrators

Agilent SureCall 3.5. Installation Guide. Agilent Technologies. For Research Use Only. Not for use in diagnostic procedures.

Agilent CytoGenomics 2.5

EZChrom Elite Compact. What s New in EZChrom Elite Compact Service Pack 2

Keysight E7515A UXM Firmware Update Tool

BenchCel Workstations Software

Agilent EZChrom SI. Startup Guide

Agilent G2721AA/G2733AA

Agilent 34826A BenchLink Data Logger for 34980A. Getting Started Guide. Agilent Technologies

Agilent 1120 Compact LC Firmware

Agilent ChemStation Plus

Agilent Genomic Workbench 7.0

Agilent ChemStation Plus

Agilent 1260 Infinity Purification Solution

Agilent CSV Export Utility

Agilent 85130C NMD 3.5 mm to Type-N Adapter Kit

Agilent OpenLAB. Data Store. Backup and Restore Guide

Agilent MassHunter Drug Analysis Mode Using Quantitative Analysis Workflow Guide

Agilent 5977 Series MSD System

Microscan Barcode Reader

Nimsoft Monitor. proxy Guide. v3.1 series

Agilent MassHunter Metabolite ID Software. Installation and Getting Started Guide

Agilent OpenLAB Chromatography Data System (CDS)

Agilent G1978B Multimode Source for 6410 Triple Quad LC/MS

OpenLAB CDS Report Template Editor

Keysight GUI Data Logger Software for U1731C, U1732C, and U1733C Handheld LCR Meter

Agilent Series Logic Analysis System

Agilent N2739A 1000 Series Oscilloscope Rack Mount Kit

Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS

Agilent OpenLAB Chromatography Data System (CDS)

Agilent Genomic Workbench Lite Edition 6.5

Agilent E4418/19A Power Meter Hardware Upgrade Kit

Agilent Genomic Workbench 6.5

Agilent Genomic Workbench 7.0

Agilent OpenLAB Chromatography Data System (CDS)

Keysight 11636B APC-3.5 mm Power Divider DC to 26.5 GHz

Automated Plate Labeling

Agilent OpenLAB. Data Store. Maintenance Guide

Errata. Agilent References in this manual. About this manual. Support for your product

Agilent Technologies E5385A 100-Pin Probe

Agilent OpenLAB Chromatography Data System

Agilent ChemStation for UV-visible Spectroscopy

Keysight N5990A DisplayPort Extended Tests Embedded DisplayPort

Keysight N5990A Test Automation Software Platform

Agilent Technologies E5346A 38-Pin Probe and E5351A 38-Pin Adapter Cable

Agilent N2916B 6000 and 5000 Series Oscilloscope Rack Mount Kit

Agilent OpenLAB Data Analysis Upload Download Tool. User s Guide

G5584A Labware MiniHub

Keysight 11878A 50 Ohm 3.5 mm Adapter Kit

Retention Time Locking with the MSD Productivity ChemStation. Technical Overview. Introduction. When Should I Lock My Methods?

Agilent OpenLAB ECM Intelligent Reporter

Automated Plate Centrifugation

Agilent Genomic Workbench 6.0

Agilent Triple Quadrupole LC/MS Peptide Quantitation with Skyline

MSD SECTOR Reader. Device Driver User Guide. Agilent Technologies

Agilent EZChrom Elite. PDA Analysis

Keysight 85130F NMD 2.4 mm to 3.5 mm Adapter Kit

Operations and Installation Manual. Keysight 85132E/F 3.5 mm to 7 mm Test Port Return Cables

Keysight N5191A/93A UXG X-Series Agile Signal Generator

Agilent OpenLAB Chromatography Data System

PlateLoc ActiveX v3.1

Unified Infrastructure Management Compatibility Matrix June 26, 2015

Nimsoft Monitor. sysstat Guide. v1.1 series

Setting Up Probes. Online Help

Nimsoft Unified Management Portal

Agilent U2941A Parametric Test Fixture

Firmware Guide. Keysight PXIe Chassis Family

Setting up a LAN Instrument Network. Installation Guide

Agilent 4155C Semiconductor Parameter Analyzer Agilent 4156C Precision Semiconductor Parameter Analyzer

Keysight M9393A PXIe Vector Signal Analyzer

Agilent Dissolution Tester Driver

Unified Infrastructure Management Compatibility Matrix September 05, 2017

Tisio CE Release Notes

Agilent 89600B VSA. Software Installation Guide

Agilent ChemStation. ECM Interface Guide. Agilent Technologies

Agilent InfiniiMax II 1168A/1169A Probes

Upgrading BankLink Books

Nimsoft Monitor. cluster Guide. v2.6 series

Agilent M9502A and M9505A AXIe Chassis Firmware Revision. Firmware Update Guide. Agilent Technologies

Agilent Genomic Workbench 7.0

Automated Serial Dilution

Agilent E2943A/E2944A ATCA Probes for Advanced Switching Interconnect

Nimsoft Monitor. xendesktop Release Notes. All series

Keysight 85130D NMD 3.5 mm to 3.5 mm Adapter Kit

New Features of Deconvolution Reporting Software Revision A.02 Technical Note

BIT-3000 Dynamic Sequencing Generator and Analyzer. Datasheet 1.11

Agilent OpenLAB Chromatography Data System (CDS)

Agilent OpenLAB Chromatography Data System (CDS)

Agilent Lab Advisor. IT Administrator's Guide. Agilent Technologies

Errata. Agilent References in this manual. About this manual. Support for your product

Nimsoft Monitor. ntp_response Guide. v1.2 series

PHERAstar. Device Driver User Guide. Agilent Technologies

Creating EMPro Projects. EMPro 2012 October 2012 Creating EMPro Projects

Nimsoft Monitor. db2mon Guide. v2.2 series

Nimsoft Monitor. ocs_monitor Guide. v1.3 series

Cavro Pump. Device Driver User Guide. Agilent Technologies

Agilent OpenLab CDS ChemStation Edition

Nimsoft Service Desk. Single Sign-On Configuration Guide. [assign the version number for your book]

Nimsoft Monitor. reboot Guide. v1.4 series

Transcription:

Agilent MSD Productivity ChemStation G1701 & G1710 Performance Report Parameters

Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number G1701-90101 Edition First Edition March 30, 2012 Printed in USA Agilent Technologies, Inc. 5301 Stevens Creek Blvd. Santa Clara, CA 95051 Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with such license. Restricted Rights Legend U.S. Government Restricted Rights. Software and technical data rights granted to the federal government include only those rights customarily provided to end user cus-tomers. Agilent provides this customary commercial license in Software and technical data pursuant to FAR 12.211 (Technical Data) and 12.212 (Computer Soft-ware) and, for the Department of Defense, DFARS 252.227-7015 (Technical Data -Com-mercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Computer Software Documentation).

Contents Introduction 4 Report Layout 5 Heading... 5 RTE Integration Parameters... 5 Graphics Area... 5 Noise characteristics... 5 Performance Report Results... 6 G1701 MSD Productivity ChemStation Performance Report Parameters 3

Introduction 1. Introduction This document describes the calculations used by the Performance Report in the Agilent MSD Productivity ChemStation. These calculations are dependent on the integrator selected for the Data Analysis. The MSD Productivity ChemStation allows use of the Chemstation integrator and the RTE integrator. The report is created by selecting Performance Report from the Chromatogram menu. The user is then presented with a dialog and asked to enter a retention time range where the noise will be calculated. A Total Ion Chromatogram (TIC) of this specified noise range is displayed in a new window and the report shown below is printed: G1701 MSD Productivity ChemStation Performance Report Parameters 4

Report Layout 2. Report Layout The report has several sections as described below. Calculations for noise are located in the Noise characteristics section. Calculations relating to individual peaks are shown in the table at the bottom of the report. Heading The heading displays the data file and path used by the report, the date the data was acquired, the operator s name, sample name, misc, and vial number. This is followed by the integration parameters file and the name of the integrator used. RTE Integration Parameters When the integrator is the RTE integrator, the RTE integration parameter settings are displayed next. Graphics Area The top window contains a Total Ion Chromatogram (TIC) of the data file. The lower graphic contains the noise area specified by the user. Noise characteristics Noise Start and Noise End These are the retention time values entered by the user to define the range used for calculating noise. Number of points The number of data points found in the specified noise range. Maximum noise The abundance value of the data point with the maximum abundance in the noise range. Minimum noise The abundance value of the data point with the minimum abundance in the noise range. G1701 MSD Productivity ChemStation Performance Report Parameters 5

Report Layout Peak-to-Peak noise PPNoise = Maximum noise - Minimum noise Root-mean-square noise RMS Noise = (N (x2 ) ( x) 2 ) N(N 1) Where: N = Number of points (see definition under Noise characteristics above) x = abundance of an individual data point Performance Report Results #: Peak number. RetTime: Retention time of the peak. Area: Area of the peak as determined by the selected integrator. BslPH: The peak height as measured from the corrected baseline. PWidth: Peak width for the ChemStation Integrator is the width of the peak at ½ the peak height. For the RTE Integrator, peak width is the width of the peak start minus the peak end. G1701 MSD Productivity ChemStation Performance Report Parameters 6

Report Layout Resol: ChemStation Integrator Resolution, as determined using the "ChemStation" Integrator. RT 2 RT 1 Resol = 1.18 PW hh1 + PW hh2 Where: RT 2 = retention time of the second peak; RT 1 = retention time of the first peak; PW hh1 = peak width at ½ height of the first peak; PW hh2 = peak width at ½ height of the second peak. RTE Integrator Resolution, as determined using the "RTE" Integrator. RT 2 RT 1 Resol = 1.18 PW BL1 + PW BL2 Where: RT 2 = retention time of the second peak; RT 1 = retention time of the first peak; PW BL1 = peak width at baseline of the first peak; PW BL2 = peak width at baseline of the second peak. Please observe that the calculation for Resolution when using the RTE integrator does not use the peak width at ½ height. Instead the resolution is calculated using the peak width at the baseline. This is not one of the classical methods for calculating resolution but can be used as a measure to compare resolutions on the Agilent MSD Productivity ChemStation. G1701 MSD Productivity ChemStation Performance Report Parameters 7

Report Layout Tailing: Peak Tailing Factor This is a measure of peak symmetry. For a perfectly symmetric peak, the peak tailing factor = 1. Tailing = RT b RT f 2.0 (RT RT f ) Where: RT = retention time of the peak. RT b = RT of peak back at 10% of peak height; RT f = RT of peak front at 10% of peak height; PtP S/N: Peak-to-Peak signal to noise ratio. PtP S/N = (BslPH)/(PPNoise) Where: BslPH = baseline corrected peak height; PPNoise (see definition under Noise characteristics above) RMS S/N: Root Mean Square signal to noise ratio RMS S/N = (BslPH)/(RMSNoise) Where: BslPH = baseline corrected peak height; RMSNoise (see definition under Noise characteristics above) G1701 MSD Productivity ChemStation Performance Report Parameters 8

Report Layout Plates: Number of Theoretical Plates This is a measure of the column efficiency (the ability to produce narrow peaks). The result is the number of theoretical plates per meter of column length. Plates = 5.54 RT 2 PW hh Where: RT = retention time of the peak; PW hh = width of the peak at ½ height. G1701 MSD Productivity ChemStation Performance Report Parameters 9

Agilent Technologies, Inc. 3/30/2012 G1701-90101