Ultralow temperature freezers Platilab & Irilab Series
ANGELANTONI LIFE SCIENCE propose a wide range of ultralow temperature freezers, from 110 to 800 liters, chests and uprights kind. This brochure includes all ULT freezers available at ALS for storage of general non-flammable materials, for laboratories use, pharmaceutical, hospital and for any other kind of storage when such a low temperature is required. PLATILAB models have the classic insulation with high density injected CFC-free PU foam of minimum 10 mm thickness IRILAB models have a mix of vacuum insulated panels (V.I.P.) and injected PU foam for total 70 mm thickness. This solution increases the internal capacity by consuming less footprint area. NEXT versions of both PLATILAB and IRILAB have the intelligent controller Cold Brain
Ultralow temperature freezers Platilab & Irilab Series Chest and upright freezers operating at, designed for long term storage of general non-flammable materials Stainless steel interiors Racks for classic fibercard boxes for vials storage are available for all models (optional) CO2 / LN2 back systems (optional) Remote alarms standard Auto recharging battery backup to feed the electronics up to 6 h in case of main power off Doors key lock or push button or by personal card opening system (see option BIOGUARD ) Inner doors are standard in all models Two kind of microprocessor controllers, both battery buffered: The standard controller has two independent sections, one for the cooling system and one for the alarms The ColdBrain controller, installed with all NEXT versions, includes a full monitoring system of the freezer performance. ColdBrain has an internal temperature graph recorder, alarms and touch key buttons too. Besides the Cold Brain is ready for Ethernet connection and is supplied with a SD memory card for local data storage (requires CRYOLOG software to read SD memory data: see optional accessories for computer communication) 550 L 800 L 520 L 0 L 70 L 110 L
Ultralow temperature freezers Platilab & Irilab Series Operating temperature range: -50 C to Technical features PLATILAB : 10 mm insulation of HD Injected PU Foam IRILAB : 70 mm insulation with vacuum panels and PU foam Standard porthole with cap (diam. 2 mm) included with all models Heavy frame with: Exterior in galvanized steel epoxy coated powder, Stainless steel interior with rounded corners Inner Doors Adjustable Shelves (upright models only) High resistance castors Main door: With magnetic double tight gasket Easy one hand opening system PLATILAB: key lock as standard or key card opening system (see options BIOGUARD) IRILAB: push button as standard or key card opening system (see options BIOGUARD) NEXT versions: push button or Bioguard (see options BIOGUARD) Vent valve to reduce the waiting time from a door opening to another Cooling system: Static with two high performance compressors (cascade system) CFC free gases Low noise level Removable front filter for easy cleaning (upright models only)
Equipment /Controllers Standard controller for PLATILAB and IRILAB series. The standard controller is made of two independent parts: one for the cooling system control and another one for all local and remote alarms. It is equipped with rechargeable battery back-up which can provide energy to the electronics up to 8h even in case of power black-outs. The standard controller can be equipped with an optional Ethernet board, for connection and data handling to the CRYOLOG software (See optional accessories). Cooling system: Temperature setting Temperature display Cooling system ON/OFF Alarms section: Alarms setting for Hi-Low temperature Door ajar Main power-off Acoustic and visual alarms with LEDs Free contacts for remote alarms Intelligent controller for PLATILAB Next and IRILAB Next The ColdBrain microcontroller uses a back lighted LCD graphic display of 20 x 128 pixels with a touch keyboard and independent electronics: one for power & temperature management, one for the alarms, and one for the display, communication and data logging. A self-diagnostic routine it s full time active to double check both electronic sections and to avoid data loss. ColdBrain Power & Temperature management section: Temperature setting Cooling system ON/OFF Power diagnostic view by two Pt100 and two NTC probes Alarms section: Hi-Low temperature Door ajar Main power-off or out of range Free contacts for remote alarms Compressors failure Alarms diagnostic view by two Pt100 and two NTC probes Probes failure Battery back-up handling, and battery low LCO2/LN2 back-up system handling Display, communication and data logging section: Temperature and alarms display Acoustic & visual alarms with LEDs Temperature graph on display for last twelve weeks Password protected settings Password protected memory card (SD) for graphs, alarms and data Ethernet port with configurable TCP/IP address Data communication failure Connection with CRYOLOG Data Management Software BIOGUARD handling, recording of accesses and failure detection (if available) 5
Technical specifications 12086 1208 1208 1070 1071 111 Model with standard controller PLATILAB 110 H PLATILAB 70 H PLATILAB 550 H PLATILAB 0 V--STD PLATILAB 500 V--STD IRILAB 800 V-5-STD Version and insulation kind Chest Chest Chest V.I.P. Operating range Inner doors 1 2 5 Shelves none none none Max vials capacity 2 ml 600 21000 000 2000 0000 60000 Volume liters 110 70 550 0 500 800 Interior 50x560x50 1050x520x670 1590x520x670 50x50x150 700x50x150 906x609x172 Exterior 70x850x1080 1780x855x110 250x855x110 765x900x1985 1015x900x1985 1086x890x1995 Package 800x1020x1270 1850x970x100 280x970x100 800x950x2150 1080x1000x2120 1151x990x2120 Net weight *with pack. 180 *205 Kg 260 *299 Kg 12 *60 Kg 2 *262 Kg 00 *29 Kg 6 *89 Kg Electrical 50Hz 9.0FLA 50Hz 9.0FLA Model with Coldbrain controller Version and insulation kind Operating range Inner doors Shelves Max vials capacity 2 ml Volume liters Interior Exterior Package Net weight *with pack. Electrical 1259 10 11 1 18 1 PLATILAB NEXT 70 H PLATILAB NEXT 550 H PLATILAB NEXT 0 V--STD PLATILAB NEXT 500 V--STD IRILAB NEXT 520V--STD IRILAB NEXT 800V-5-STD Chest Chest V.I.P. V.I.P. 2 5 none none 21000 000 2000 0000 800 60000 70 550 0 500 520 800 1050x520x670 1590x520x670 50x50x150 700x50x150 586x61x170 906x609x172 1780x855x110 250x855x110 765x900x1985 1015x900x1985 70x85x1980 1086x890x1995 1850x970x100 280x970x100 800x950x2150 1080x1000x2120 800x950x2150 1151x990x2120 260 *299 Kg 12 *60 Kg 26 *262 Kg 01 *29 Kg 260 *290 Kg 6 *89 Kg 50Hz 9.0FLA 50Hz 9.0FLA 50Hz 9.0FLA NOTE: All models are available upon request for 20 or 120 VAC, 60Hz Optional Accessories for computer communication 1212 180 Description Cryolog monitoring software for Windows computers (see details on last page too) Ethernet connection board necessary only for STD versions (NEXT versions include as standard the Ethernet board) IMPORTANT NOTES: 1. To connect a freezer to the computer where Cryolog is installed, please provide by yourself for two Ethernet cables of proper length and one Ethernet Switch: one cable is for freezer Switch connection and one is for Switch computer connection 2. If you have more than one freezer to put on line, please buy as many Ethernet cables of proper length as the number of freezers you have and Ethernet Switch with sufficient numbers of ports equal or greater than the number of freezers plus one (last one free to be used for computer connection). Example of computer interfaces of CRYOLOG software - see specific section: CRYOLOG data management software 6
Cryogenic back-up systems 1190 12278 15 15 12107 155 Description CO2 back-up system for PLATILAB 0V, 500V and IRILAB 520, 800V CO2 back-up system for PLATILAB 110H, 70H and 550H CO2 back-up system for PLATILAB NEXT 0V, 500V, IRILAB NEXT 520V and 800V CO2 back-up system for PLATILAB NEXT 70H and 550H LN2 back-up system with hose for all models except IRILAB NEXT 520V and 800V LN2 back-up system with hose only for IRILAB NEXT 520V and 800V LCO2 / LN2 cryogenic back-up systems to automatically supply liquid nitrogen or carbon dioxide into the freezers. In case of chamber warm up beyond - 65 C, the system automatically injects the liquid gas, to keep the internal temperature around -65 C. These back-up systems include the probes, the electronics and the cryogenic valves. The cryogenic cylinders with LN2 or LCO2 are not included and must be locally provided through a qualified supplier, as well as the connection piping. The LCO2 / LN2 are emergency devices only. In none case, they can replace the job of the cooling systems. The task of such emergency systems, it is to ensure few hours extra at a reasonably low temperature, even in case of compressors failure or mains power black-out. The few hours ensured by the back-up systems represent the time necessary to move the freezer contents into another working unit until the electric power is back. This time may differ significantly depending on several factors: kind and quantity of cryogen gas availability, freezer filling volume and external ambient temperature. -86 C FREEZER LCO2 - example of connection diagram of a LCO2 complete back-up system General accessories for PLATILAB and IRILAB versions 1152 1229 11976 12 Description 7-days temperature chart recorder (pen + diagrams for 1 year) Kit for chart recorder (100 discs and 2 pens) Additional user badge for BIOGUARD system Additional PT 100 temperature probe class A 1212 180 Additional certified PT 100 temperature probe, class A with calibration points Ethernet board (only for PLATILAB and IRILAB STD, request Cryolog swr for computer connection) 12680 BIOGUARD - RFID card door opening system, fits any models except PLATILAB 110 H BIOGUARD system uses RFID cards to open the door without key. Each unit equipped with the Bioguard is supplied complete with three opening cads but in addition to that, You may add some more cards (optional). The BIOGUARD option increase the safety of your valuable stored products. In case of card damaging or electronic failure, there is an emergency rescue key to open the door manually. BIOGUARD 7
Racks for vial storage Storage racks for chest freezers, for boxes with sizes of 129x129x 50 H mm PLATILAB HORIZONTAL Description Dimension mm Storage capacity 110 L 70 L 550 L 11685 Vertical without boxes W10xD15xH91 7 9 11687 Vertical without boxes W18xD18xH60 10 21 10 Storage racks for chest freezers, for boxes with sizes of 129x129x 75 H mm PLATILAB HORIZONTAL Description Dimension mm Storage capacity 110 L 70 L 550 L 11686 Vertical without boxes W10xD15xH91 5 9 11688 Vertical without boxes W18xD18xH60 7 21 7 Storage racks for upright freezers, for boxes with sizes of 129x129x 50 H mm PLATILAB VERT. IRILAB VERT. Description Dimension mm Storage capacity 0 L 500 L 520 L 800 L 165 Horizontal without boxes W10xD5xH281 20 12 20 20 121 Horizontal without boxes W10xD5xH19 2 16 1228 Horizontal without boxes W10xD5xH266 20 0 Storage racks for upright freezers, for boxes with sizes of 129x129x 75 H mm PLATILAB VERT. IRILAB VERT. Description Dimension mm Storage capacity 0 L 500 L 520 L 800 L 1196 Horizontal without boxes W10xD5xH29 12 12 12 20 12 0 16 Horizontal without boxes W10xD5xH18 16 16 Fiberboard boxes for all racks 188 189 Description 10 boxes 129x129x50H mm without dividers 10 boxes 129x129x75H mm without dividers Internal dividers for all boxes 180 181 182 Description 25 internal dividers 9 cells, diameter each 15 mm 25 internal dividers 100 cells, diameter each 12 mm 25 internal dividers 81 cells, diameter each 1 mm 8
Equipment /CRYOLOG data management software Main features of CRYOLOG Acquisition and monitoring of all data parameters Ethernet communication with programmable TCP/IP address Creation of a complete data base on computer hard disk Log storage and display of all events such as alarms, failures, door opening Fully programmable temperature graphs view (curve,zoom, color, grid selection...) Possibility of multiple view of different temperature curves by sliding cursor Reading and decrypting to.csv file, of SD memory card data used by ColdBrain. Minimum computer requirements Before installing the CRYOLOG application, please be sure that your computer includes at least the following equipment: Intel Pentium processor or higher version 1 GB RAM 200MB of free space on the Hard-Disk 1280 x 720 minimum resolution, 16 million colors One Ethernet card Mouse & keyboard Windows Vista / 7 / 8 To connect any device to the computer where CRYOLOG is installed, two Ethernet cables of proper length and one Ethernet Switch are necessary: one cable is for the device and another one for computer connection. If you have more than one device to connect with CRYOLOG, you need as many Ethernet cables as the number of devices you wish to connect to the computer, plus one extra cable, for the connection between the Switch and the computer. FREEZER n... FREEZER 1 FREEZER 2 FREEZER (Ethernet Switch) CRYOLOG SWR (PC) (Ethernet cables) Schematic connection of -86 C freezers to computer with CRYOLOG 9
Biomedical refrigerators and freezers (+ /), chambers for stability tests and LN2 robotic storage system for vials (-180 C) Refrigeration systems for industrial processes Angelantoni Infection Control, steam sterilizers, washers and biohazard waste treatment Biosafety laminar flow cabinets Copyright 2015 Angelantoni Life Science S.r.l. / Rev. 01 - Illustration, description and technical data may change without notice.