WIRELESS RF Implantable Stimulation User s Manual

Similar documents
Stimulation User s Manual. Version 2.0

UserGuide_TempSensor_with_Alarms Issue 4/

DYB2200 INEXPENSIVE, RELIABLE USB PRODUCTS.

HS-18 User Manual. A preamplifier headstage with 18 channels of unity gain including 16 electrode channels and 2 references.

ADPT-HS36-N2T-32A User Manual

HOBO State Data Logger (UX90-001x) Manual

Thermal Transient Test Installation and Operating Manual

Installing LE History Record Reader program software.

HH Channel Data Logging Headstage Manual. Version 6

HOBO State Data Logger (UX90-001x) Manual

DATA LOGGER (Version V1.3)

Quick Start Guide. Data Logger TrueLog100

Remora Acoustic Recorder

Quick Start Guide. January EAZ0081L02A Rev. A

Pico Installation Guide

DIABLOSPORT PREDATOR REVISION UPDATE INSTRUCTIONS

Prizmatix Pulser. USB-TTL Interface. User Manual

CRIBS Installation and User Guide

contents Page 1 of 6 Field Programming Software User Guide Rev. A 10/16

Cheetah16/32/64 Setup Guide

CarScope VISO Getting started manual DITEX. CarScope VISO. Getting started manual

Plus-X IIU / ICRU Installation Guide

MODEL DL-1 and TRAC 4 Software

Freeduino USB 1.0. Arduino Compatible Development Board Starter Guide. 1. Overview

DataNet Installation Guide + Quick Start Guide Updated: August 2008

YourBell INEXPENSIVE, RELIABLE USB PRODUCTS.

Help Volume Agilent Technologies. All rights reserved. Agilent E2485A Memory Expansion Interface

CORD-XL Dual-Channel Electronic Chart Recorder User s Manual

UsbScope.eu Manual. UsbScope.eu Manual. Version channel version

1. Place your CD driver in your PC drive for software installation 2. Follow the instruction on the screen to finish the software installation

ViZion DR + Wireless Install Guide

COMPASS PLM-4-BT USER MANUAL. Bluetooth. Rev. 1.2

HS-27 Manual 5/2/2013. Neuralynx, Inc. 105 Commercial Drive, Bozeman, MT Phone Fax

TSR 3DXP and TSR 6DXP User s Manual May 2011

SONOFLOW Monitor Short Instructions. For SONOFLOW Sensors of the Series CO.55 V2.0

SPCPro USER MANUAL.

Introduction. Chapter 2

IP RECEIVER GRx 8-32 User's Guide

Thermal Transient Test

Swift Responses with Accurate Measurement

User Guide. Version 2.

THREAT-FIRE MANUAL SYSTEM SPECS AND OPERATION INSTRUCTIONS. Version

VWR Protocole Builder Software User manual Summary

Trouble shooting the DeskCNC controller:

HS-27-Mini. Headstage preamplifier providing 27 channels of unity gain amplification.

M950 Thermometer Instructions Connectors: Probe Connector Please Attach Probe Temperature Probe Error Please Try Another Probe

SensorWATCH Basic RH - Wireless Setup Users Help Guide Part Number: A Revision: 1.1.0

Using PATCHMASTER in Standard Patch-Clamp Experiments

Danfoss commercial compressors Secop compressors produced for Danfoss Bluetooth Gateway

User Guide for Digidata 1440A

Raspberry Pi Relay Board v1.0

Embedded Navigation Solutions VN 100, VN 200 & VN 300 Development Board User Manual

Quick Start Instructions for Mitsar Installation

RST INSTRUMENTS LTD.

512 channels WIFI-DMX Interface

IkaScope WS MAR.2018 PRODUCT DATA SHEET. IKALOGIC S.A.S. 19 Rue Columbia Limoges FRANCE

VP- X Pro & VP- X Sport

GDC75 ARINC 429 BUS READER. User's Guide. Rev IR

Lab Rat User Guide Updated: 2/9/2018

DPL4000 Portable Low Range Dew Point Analyzer

KWJ-WSN & USB Gateway. Quick Start Guide

INSTRUCTION MANUAL C WA2 WELD ANALYZER

PowerTrac USB and Link Users Guide Software Installation & Operation Manual

Bluetooth Keyless Entry System - Lite

Lab 4: Introduction to ELVIS II+ Introduction to ELVIS II+

Technical data. General specifications. factory setting: 125 ms Memory Non-volatile memory Write cycles Indicators/operating means

SPECIAL PROJECTS 3800 Hwy 271 North Powderly, TX Phone: Fax: Aqua 20/20 OWNER S MANUAL

Spectrometer Visible Light Spectrometer V4.4

Digital Surveillance Camera

Model BS-150. Instruction Manual. Video Inspection Camera. reedinstruments. www. com

BushBaby DVR: Retrieving Stored Video / Pictures

User s Manual. SLP-F7M Smart Loader Package for the F7M Micro Flow Rate Liquid Flow Meter. No. CP-SP-1423E

Chapter 1 Introducing the OM-USB Functional block diagram... 5

Features. IDS : Inclinometer Display System with RS232 Output

A variety of ECONseries modules provide economical yet flexible solutions. Waveform Generation

DVR 520 / 1.8 Digital Video Recorder

BF ENTRON LTD. Monitoring equipment. Weld Analyser WA2

Instruction Manual. ROYTRONIC EXCEL Series AD9 Programming Guide Electronic Metering Pumps

Digital Surveillance Camera INT SPYP. IR-B model. User s manual.

BESANTEK USB Data Logger Instructional Manual (BST-DL10, BST-DL13, BST-DL14)

ATS Automatic transfer switches Advanced version

Bluetooth Light Gun. Please also see the Aimtrak Setup Guide at : Setup Guide.pdf

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB

Model BS-150. Instruction Manual. Video Inspection Camera. reedinstruments www.

USER MANUAL S6000 / S7000 S8000 / S9000. Associated Research, Inc., West Laurel Drive Lake Forest, Illinois, U.S.A.

midi LOGGER GL240 Quick Start Guide GL240-UM-852

IPM650 Intelligent Panel-Mount Display

DaqBoard/3000USB Series USB 1-MHz, 16-Bit Data Acquisition Boards

Wavy for PAT-T Ver. 5.2

Remote Display User Manual

SBS-200CT BATTERY DISCHARGE CYCLER USER MANUAL

User s Guide. LA5034 Operation Manual

Installation and Operation Back-UPS BR1000G-IN / BR1500G-IN

Multi-Function Ground Resistance Tester Model 6471

120 WATT FLEX ATX TYPE WITH ACTIVE PFC

Compliance test method and detailed spec for - USB2.0. Tektronix Korea YJ.PARK

Vercise DBS Programming Manual

Wavy for PAS&PWR Ver. 5.0

User s Manual. USB Bluetooth Dongle

ION Demo Kit. Quick Start Guide PERFORMANCE MOTION DEVICES

Transcription:

WIRELESS RF Implantable Stimulation User s Manual Triangle BioSystems International 2224 Page Rd. Suite 108 Durham, NC 27703 Phone: (919) 361-2663 Fax:(919) 544-3061 www.trianglebiosystems.com

Table of Contents Document Overview... - 2 - Shipped Items List... - 2 - Headstage Pinout... - 3 - Hardware Installation... - 4 - Software Installation... - 5 - System Requirements... - 5 - StimWare Installation... - 5 - Software Navigation... - 7 - Loading/Saving Configurations... - 7 - Programming the Stim Channels... - 8 - System Setup and Testing... - 9 - Application Notes... - 19 - Electrodes... - 19 - Troubleshooting... - 19 - Version history for this manual... - 20 - Version 1.0-1 - 9/21/2016

Document Overview This document will demonstrate how to use the TBSI Implantable STIMWARE hardware and software. The program controls a single 2 channel implantable headstage that allows researchers to generate and upload two separately customizable waveform patterns via a wireless USB dongle. The headstage is battery operated therefore only needs to be connected to the charging plug for recharging. Shipped Items List 2 Channel Stimulation Implantable Headstage StimWare Installation CD USB Dongle Inductive Power Charger Cage 2 Trigger Interface Cables Stim Signal Test Load Board Implantable Headstage Installation CD USB Dongle Inductive Power Charger Trigger Cables Load Board Figure 1: Images of Items Version 1.0-2 - 9/21/2016

Headstage Pinout Figure 2: Top View Figure 3: Side View Version 1.0-3 - 9/21/2016

Hardware Installation Figure 4: System Setup Diagram Version 1.0-4 - 9/21/2016

Software Installation System Requirements The recommended minimum PC requirements are as follows: At least 1.0 GHz Processor 2 GB of storage memory 1 GB RAM (1) USB 2.0 port Windows 7 StimWare Installation To install and run StimWare: 1. Insert the StimWare installation disc into your computer. Navigate to STIMWARE Installer folder within the installation disc and double click setup.exe,then follow the prompts until installation is complete. 2. Navigate to Silabs-CDC_Install folder within the installation disc and double click Silabs- CDCInstaller_x64.exe if Operation System is 64bit or Silabs-CDCInstaller_x86.exe if Operation System is 32bit,then follow the prompts until installation is complete. Figure 5: Select dongle driver Version 1.0-5 - 9/21/2016

3. Plug the USB dongle into PC. Silicon Labs CDC Serial port driver will be installed. Figure 6: Dongle driver installation done 4. Right click on My computer and select Manage to open the Device Manager. If you do not have My computer on your desktop click the start button then right click on Computer and select Device Manage. Dongle should be indicated by Silicon Labs CDC Serial port (COM#) under COM and LTP Figure 7: Information in Device Manager Version 1.0-6 - 9/21/2016

Software Navigation Launch the StimWare software application. StimWare is in Run mode once launched. Loading/Saving Configurations Figure 8: StimWare Launch Screen Default parameters are loaded into StimWare from the configuration (.ini) file located on your computer: C:\2Ch STIMWARE Version 1.0-7 - 9/21/2016

Figure 9: Loading Configuration File Customized pattern parameters can be saved by pressing the Save As button. The new configuration file, by default, will be saved to C:\2Ch STIMWARE Figure 10: Saving Parameters Programming the Stim Channels The two stim channels can be enabled and programmed separately from each other. To enable a channel, select the checkbox next to its name. A channel that is unchecked will not output any signal, so it may be advantageous to disable an unused channel to conserve battery life. Version 1.0-8 - 9/21/2016

Figure 11: Distinguishing Separate Channels The graphics in the StimWare interface are intended as conceptual references and do not reflect the actual pattern set by the user-defined parameters. The Initial Delay (ID) will be executed once at the start of the pattern. By clicking the Balance Charge button, the software will verify and/or balance the current of the individual pulses. Checking the Invert box will change the polarity of the entire signal. To enable higher order tiers, check the box next to its name. Keep unused tiers disabled if they are not in use. Certain parametric limitations are programmed into the software to ensure signal consistency and prevent saturation of the headstage amplifiers. If a parameter has exceeded its allowed range (ex. a frequency is set too high in a higher order tier), it will automatically be reset to its highest allowed value. PI1 and PI2 can t exceed 500uA. PD1 and PD2 can t be lower than 50uS. System Setup and Testing 1) Connect USB Dongle to computer Open the StimWare program application. If the dongle is successfully recognized by your computer, the Dongle indicator will illuminate green in StimWare. Version 1.0-9 - 9/21/2016

Figure 12: Dongle Recognition 2) Search and Select Headstage Once the headstage is charged; Click the Search HS tab. Press the Search HS button on that window and you will see your headstage # under Available Headstages. Once your HS is visible, press the Select HS button and the green Headstage light at the bottom will illuminate and turn on Headstage. Version 1.0-10 - 9/21/2016

Figure 13:Searching for Headstage Figure 54:Selecting Headstage Multiple Headstage If you have multiple headstages with different address, you will be able to control each headstage with StimWare. At one time, dongle can only communicate with one headstage. Once the headstages are charged; Set up Start Address and End Address based on headstage s address range ; Click the Search HS tab; Press the Search HS button on that window and you will see your headstage listed in Available Headstages table. Once your headstages are visible, click on headstage # that you want to control with mouse and press the Select HS button. The green Headstage indicator at the bottom will illuminate and selected headstage address will be shown right inside indicator. Version 1.0-11 - 9/21/2016

Figure 15: Multiple Headstage Detection 3) Remove possible DC Offsets Click the DC Offset Zero button to set the signal level to 0V before downloading a pattern. The light will blink yellow while the DC offsets are being removed. DC offset adjust should only be performed when resistance changes. Figure 16:Removing DC Offset Version 1.0-12 - 9/21/2016

4) Generate and download pattern in StimWare Choose the stim signal parameters you wish to upload to the headstage. (Refer to the Software Navigation section of this document for more pattern programming instructions.) When finished, click the Download Pattern button to send the pattern data to the headstage. The pattern will be saved on headstage until you download new pattern. Figure 17:Downloading Pattern 4.1) Software Trigger Press Trigger to start the pattern. The Pattern light will illuminate and blink while the pattern is running and will stop blinking when pattern is over. Version 1.0-13 - 9/21/2016

Figure 68: Triggering Pattern 4.2) External Trigger Two external triggers are supported. As Figure1 shows, one end of trigger cable can be plugged in dongle and the other end are two bare wires that can be connected with external system in which external trigger signals are generated. Red wire is external trigger signal whose voltage is up to 5V. Black wire is ground and should be connected with the ground of external system. All external trigger settings are located in Set up tab of StimWare as Figure16 shows. Either rising edge or falling edge can be selected as active trigger edge. The channel should be disabled to avoid false trigger when you finish experiment or when you don t use external trigger. Trigger setting won t take effective until you click Set Up External Trigger button. Version 1.0-14 - 9/21/2016

Figure 19: External Trigger Setting When valid external trigger is detected, it will trigger the pattern loaded in headstage. Trigger Cable# will show up in log window and Pattern indicator starts blinking. Figure 20: External Triggering Version 1.0-15 - 9/21/2016

Trigger Signal Rising edge active If Rising Edge is selected in Ch# Edge Selection, the rising edge of external trigger signal will trigger pattern. If the positive pulse T1 is more than 4ms, it s regarded as valid trigger. The pattern will start after delay T2. If the positive pulse T1 is less than 4ms, it s regarded as false trigger and ignored and the pattern won t start. T3 is interval between the end of pattern and the next coming trigger edge. Users should set up the reasonable pattern duration by considering T4, which is interval between the two consecutive trigger pulses. VH T1 VL Pattern Start Pattern Stop Rising edge T2 T3 T4 Figure 21: External Trigger Rising Edge Active Falling edge active If Falling Edge is selected in Ch# Edge Selection, the falling edge of external trigger signal will trigger pattern. If the negative pulse T1 is more than 4ms, it s regarded as valid trigger. The pattern will start after delay T2. If the negative pulse T1 is less than 4ms, it s regarded as false trigger and ignored and the pattern won t start. T3 is interval between the end of pattern and the next coming trigger edge. User should set up the reasonable pattern duration by considering T4,which is interval between the two consecutive trigger pulses. Version 1.0-16 - 9/21/2016

T1 Trigger Signal VH VL Pattern Start Pattern Stop Falling edge T2 T3 T4 Figure 22: External Trigger Falling Edge Active Parameter Description Min. Max. Unit VH External Trigger High 2.1 5 V VL External Trigger Low 0 0.95 V T1 Trigger Pulse width 4 ms T2 trigger edge to pattern start time 9 20 ms T3 pattern end to next trigger edge time 5 ms T4 Trigger interval 50 ms 20Hz Figure 23 External Trigger Table 5) View stim pattern Confirm the signal shape coming out of the headstage with an oscilloscope or similar device. Either pressing Trigger button in StimWare or using external trigger signal connected with Dongle will start the pattern that is already loaded on the headstage. The entire pattern will execute once for each valid trigger, and then the headstage will go into standby mode. Version 1.0-17 - 9/21/2016

Test Setup Stimware USB Dongle Oscilloscope Head Stage on Test Board Figure 24: Test Setup Hardware Correlations 6) Turn off Headstage Press the Turn Off HS button in StimWare to turn off headstage when experiment is over. The Headstage indicator will go gray when headstage is off. Version 1.0-18 - 9/21/2016

Figure 25:Turning Off Headstage Application Notes Charging the Headstage Lithium Ion Battery The wireless stimulator headstage includes an integrated rechargeable battery which should be recharged when not in use. Recharging is accomplished by the following procedure: 1. Connect supplied charging plug to the charge port on the headstage. 2. The LED on the charging plug will turn green when the battery is charged. Electrodes Microprobes Inc., NeuroNexus, Neuralynx and CD Neural Technologies offer a variety of electrode arrays that connect to all of our headstages. Troubleshooting Problem: The headstage won t turn on. Version 1.0-19 - 9/21/2016

Suggestion: Charge the battery via the charging plug provided with the system. Refer to the section titled Application Notes for charging instructions. Problem: One or more output signals are missing or incorrect. Suggestion: Download the stim signal parameters to the headstage with StimWare, making sure to follow all instructions within this manual. Check to make sure that the channels are enabled within the software prior to downloading. Check all equipment used to monitor the signal to ensure continuity. Call TBSI customer support at (919) 361-2663 if there are any problems with system setup and function. You can also submit a Technical Issue report on our website here. Version history for this manual 1.0 Initial release with Specifications 1.1 Removed Specifications, Added cover page 1.2 Updated output pinout, added mating connector part number 1.3 New functionality added 11/2013 1.4 Updated pinout, images 5/2014 1.5 Overhauled Version 1.0-20 - 9/21/2016