Thermo-Anemometer Data Logger Model 1227

Similar documents
QUICK START USER GUIDE. Data Logger Model L452

Simple Logger II Series

Micro-Ohmmeter Model 6292

THERMO-ANEMOMETER DATA LOGGER

Clamp-on Ground Tester Model 6417

Megohmmeter Models 6550 & 6555

POWER & ENERGY LOGGER MODEL PEL 102 & PEL 103

OX 5022-CK OX 5042-CK

POCKET MULTIMETER 5115 ENGLISH User Manual

THERMO-HYGROMETER DATA LOGGER

200A MICRO-OHMMETER ENGLISH. User Manual

PowerPad QUICK START 8335 USER GUIDE IMPORTANT WARRANTY NOTE:

PowerPad III Model 8336

CL600 SIMPLE LOGGER RMS CLAMP-ON MODULE E N G L I S H. User Manual

L610 L620 L630 SIMPLE LOGGER THERMOCOUPLE MODULE E N G L I S H. User Manual

CA846 THERMO-HYGROMETER ENGLISH. User Manual. 99 Washington Street Melrose, MA Phone Toll Free

AC/DC Current Probe Model SL201

AC/DC Current Probe Model SL206 USER MANUAL

L605 SIMPLE LOGGER TEMPERATURE MODULE E N G L I S H. User Manual

LIGHTMETERS CA811 CA813. User Manual

SIMPLE LOGGER II TRMS VOLTAGE MODEL L261. User Manual

SL261 AC/DC CURRENT OSCILLOSCOPE PROBE. User Manual ZERO. 100 mv/a 10 mv/a OFF

LIGHTMETERS CA811 CA813. User Manual

SENTINEL VOLTAGE DATA LOGGERS DATA LOGGERS. User Manual

CLAMP MULTIMETER METRACLIP 85 ENGLISH. Start Guide

CLAMP-ON METER ENGLISH. User Manual

User Manual. Data Logging Anemometer. Model with NIST-Traceable Calibration THE STANDARD IN PRECISION MEASUREMENT

MICRO-OHMMETER ENGLISH. User Manual

WIRE MAPPER PRO LAN CABLE TESTER CA7028. User Manual

Instruction Manual Extreme Temperature (XT) Data Logger

SIMPLE LOGGER DC MODULE L320 L410 L430. User Manual

Large Vane CFM/CMM Thermo-Anemometer

374 FC/375 FC/376 FC. Clamp Meter. Safety Information

Temperature & Humidity Datalogger

MEASUREMENT OF INDOOR AIR QUALITY CO 2 AMBIENT TEMPERATURE RELATIVE HUMIDITY C.A Shop for AEMC products online at:

Simple Logger II Data Loggers

Large Vane CFM/CMM Anemometer / Psychrometer

CFM/CMM Hot-Wire Thermoanemometer

User's Guide. Phase Sequence and Motor Rotation Tester Model

MICRO-OHMMETER. User Manual

Micro-Ohmmeter. Digital 200A. Model Measure low resistance with high accuracy. Technical Hotline: (800)

MICRO-OHMMETER. User Manual

User Manual. Heavy Duty Hot Wire CFM Thermo-Anemometer. Model

User Manual. Dual-Input Data Logging Thermometers. with NIST-Traceable Calibration. Thermocouple Meter, Type J/K/T Input Model

Digital Vane Thermo-Anemometer Model

Model P4017 Single Channel USB Oscilloscope. Quick Start Guide

Micro-Ohmmeter Model 6290

User's Guide. Cup Thermo-Anemometer. Model AN400

Energy Logger. User's Guide. Model EM Washington Street Melrose, MA Phone Toll Free

Digital Multimeter Safety Tips

ACT-1B Series Panel Tachometer

Humidity & Temperature Datalogger

Tinytag Thermohygrometer User Guide

C.A 6165 Appliance multitester

12-Channel Thermocouple Datalogger

C.A A microhmmeter

User Manual. Stoltzen SHSP14 HDMI Splitter 1x4. 4K HDMI Splitter 1x4. All Rights Reserved. Version: SHSP14_2016V2

Operating Manual. English. Digital-Analog Converter DACON8-UMB.

Installation Guide. QBox-V6. Standalone/Spare V6 SDI QBox. Standalone/Spare V6 SDI QBox. Part No. A

ETHOS Auto Ranging Digital Multimeter

Net-Ritef Continuity and Wiremap Tester Trace-Ritef Tone Generator and Amplified Tone Tracer User's Guide.

english Thanks for choosing the Xbox GB Hard Drive. The Xbox GB Hard Drive lets you save games, save music and create

Keysight U1731C/ U1732C/U1733C Handheld LCR Meter. Quick Start Guide

PV Remote Unit A 1378 Instruction manual Version 2.1, Code no

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9310/9350/9510/9550/9500 QUICK START GUIDE

Heavy Duty Datalogger Module

User Manual SP2-4K. 4K HDMI Splitter 1x2. All Rights Reserved. Version: SP2-4K_2016V1.0

3-Axis G-Force Datalogger

Quick Start Guide CJB1950ALAAB

P344 4-lane MIPI DPhy Stimulus Generator Data Sheet & User Manual. January Rev 1.31

AEROTRAK HANDHELD AIRBORNE PARTICLE COUNTER MODEL 9306 QUICK START GUIDE

Mini Digital Multimeter

User Manual. Thermocouple Thermometer with Dual Type K/J Inputs and Data Logging. Model with NIST-Traceable Calibration

SafeNet Reader CT1100. For Windows 7, 8.1 and 10 User Guide

User's Guide. Hot Wire Thermo-Anemometer. Model

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at

PAT-Cal 2 TECHNICAL MANUAL

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS:

User's Guide. Vane Thermo-Anemometer Datalogger. Model Introduction

User Manual TL-FO-HD. 4K Optical Fiber Extender. All Rights Reserved. Version: TL-FO-HD_160628

1. Introduction 2. Specifications 3. Symbol Definition and Button Location 4. Operation Instructions 5. Temperature Measurement 6.

RIGOL. Quick Guide. DP800A Series Programmable Linear DC Power Supply. Aug RIGOL Technologies, Inc.

Datalogging Hygro-Thermometer

User Manual. Thermocouple Datalogging Thermometer. Model EA15

Earth Resistance and Resistivity Tester

User Manual. Data Logging Light Meter. Model with NIST-Traceable Calibration THE STANDARD IN PRECISION MEASUREMENT

Heavy Duty CFM Thermo-Anemometer Model

POCKET MULTIMETER Model No: MM18

LQBHPS Bluetooth Headphones User Manual Please read this manual before operating your device, and keep it for future reference.

INSPECTION CAMERA MODEL NO: CIC2410 OPERATION & SAFETY INSTRUCTIONS PART NO: GC0116

Network Camera. Quick Guide DC-B1203X. Powered by

RS Stock No Instruction Manual RS Input Data Logging Thermometer

USER MANUAL. WBV412U01 AC voltage transducer

Quick Start/Installation Guide

SLA Battery Capacity Analyzer

Datalogging Hygro-Thermometer

Heavy Duty Hot Wire CFM Thermo-Anemometer

Heavy Duty Datalogger Module

11 in 1 Environmental Meter

Sound Level Meter SD Card real-time Datalogger

Transcription:

Thermo-Anemometer Data Logger Model 1227 Quick Start Guide ENGLISH www.aemc.com

Statement of Compliance Chauvin Arnoux, Inc. d.b.a. AEMC Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards. We guarantee that at the time of shipping your instrument has met its published specifications. An N.I.S.T. traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge. The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com. Serial #: Catalog #: 2121.72 Model #: 1227 Please fill in the appropriate date as indicated: Date Received: Date Calibration Due: Chauvin Arnoux, Inc. d.b.a AEMC Instruments www.aemc.com 1

Thank you for purchasing the Thermo-Anemometer Data Logger Model 1227. For best results from your instrument: read these operating instructions carefully comply with the precautions for use WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears. Information or useful tip. Battery. Magnet. The product has been declared recyclable after analysis of its life cycle in accordance with the ISO14040 standard. AEMC has adopted an Eco-Design approach in order to design this appliance. Analysis of the complete lifecycle has enabled us to control and optimize the effects of the product on the environment. In particular this appliance exceeds regulation requirements with respect to recycling and reuse. Indicates conformity with European directives and with regulations covering EMC. Indicates that, in the European Union, the instrument must undergo selective disposal in compliance with Directive WEEE 2002/96/EC. This instrument must not be treated as household waste. The Model 1227 is designed to operate with system voltages less than 75V DC. Precautions This instrument is compliant with safety standard IEC 61010-2-030, for voltages up to 5V with respect to ground. Failure to observe the following safety instructions may result in electric shock, fire, explosion, and damage to the instrument and/or the installation in which it is located. The operator and/or the responsible authority must carefully read and clearly understand all precautions to be taken prior to using the instrument. Thorough knowledge and awareness of electrical hazards are essential when using this instrument. Observe all conditions of use, including temperature, relative humidity, altitude, pollution degree, and location of use. Do not use instrument if it appears damaged, incomplete, or improperly closed. Before each use, check condition of housing and accessories. Any item on which the insulation is deteriorated must be set aside for repair or disposal. Always hold the sensor handle, and keep your fingers away from the propeller. Troubleshooting and metrological checks must be done by accredited personnel. 2

Installing Batteries INITIAL SETUP 1. Press the tab of the battery compartment cover and lift it clear. 2. Remove the battery compartment cover. 3. Insert the new batteries, ensuring correct polarity. 4. Close the battery compartment cover; ensuring it is completely and correctly closed. Connecting to a Computer The Model 1227 must be connected to DataView before it can be configured. (For detailed setup instructions, see the User Manual in the USB drive that comes with the instrument.) To connect the Model 1227 to your computer: 1. Install the DataView software, making sure to select the Data Logger Control Panel as an option (it is selected by default). De-select any Control Panels you do not need. 2. If prompted, restart the computer after installation is complete. 3. Connect the instrument to the computer using a USB cable or pair with Bluetooth. 4. Wait for the drivers to install. The drivers are installed the first time the instrument is connected to the computer. The Windows operating system will display messages indicating when the installation is complete. 5. Start the Data Logger Control Panel by double-clicking the Data Logger shortcut icon in the DataView folder placed on the desktop during the installation. 6. Click Instrument in the menu bar, and select Add an Instrument. 7. The Add an Instrument Wizard dialog box opens. This is the first of a series of screens that lead you through the instrument connection process. The first screen prompts you to select the connection type (USB or Bluetooth). Choose the connection type and click Next. 8. If the instrument is identified, click Finish. The instrument is now communicating with the Control Panel. 9. When you are finished, the instrument will appear in the Data Logger Network branch in the Navigation frame, with a green check mark indicating a successful connection. Setting the Instrument s Clock To ensure an accurate time stamp of measurements recorded in the instrument, set the instrument s clock as follows: 1. Select the instrument in the Data Logger Network. 2. In the menu bar, select Instrument. In the drop-down menu that appears, click Set Clock. 3. The Date/Time dialog box appears. Complete the fields in this dialog box. If you need assistance, press F1. 4. When you are finished setting the date and time, click OK to save your changes to the instrument. 3

INSTRUMENT CONFIGURATION In addition to setting the instrument s clock, other basic setup tasks include: Enabling Bluetooth (can be done on the instrument or via DataView) Setting measurement units to F or C (can be done on the instrument or via DataView) Changing the Auto OFF interval (requires DataView) Detailed information for configuring the instrument via the DataView Data Logger Control Panel is available by pressing the Help button. Enabling Bluetooth Long press (>2 seconds) the button to enable/disable Bluetooth. Selecting Temperature Units Press to toggle between C and F. OPERATION Making Measurements Air Speed: 1. Connect the sensor to the instrument. 2. If the instrument is OFF, press and hold down the button until it turns ON. 3. Press and hold down the button. While holding this button down, shortpress the button to cycle through the measurement options until VELOC appears on the LCD. 4. While continuing to hold down, short-press to select the air speed units of measure. Options are M/S (meters per second), F/M (feet per minute), KM/H (kilometers per hour), and M/H (miles per hour). When the desired measurement is displayed, release. 5. To change temperature units, short-press. 6. Hold the sensor in the air flow. The arrow on the inside of the propeller must point in the direction of the flow. This ensures the temperature sensor faces the incoming air, allowing it to display an accurate value more quickly. 7. The air speed and temperature measurements are displayed on the LCD. (Wait for the measurements to stabilize before accepting the reading.) 4

Air Flow: 1. Press and hold down the button. While holding this button down, short-press the button to cycle through the measurement options until the desired FLOW setting appears on the LCD. Choices are: In - measures incoming (blowing) air flow with a cone. out - measures outgoing (suction) air flow with a cone. CUSt - measures air flow without a cone. If you select this option, you must enter the area of the ventilation register into the DataView Data Logger Control Panel, as explained in step 4 below. The instrument can then calculate air flow using the speed measurement and ventilation area setting. 2. While continuing to hold down, short-press to select the air flow units of measure. Options are M3/S (cubic meters per second), M3/H (cubic meters per hour), L/S (liters per second), and CF/M (cubic feet per minute). When the desired measurement is displayed, release. 3. To change temperature units, short-press. 4. If you are using an air flow cone: Skip the remainder of this step and proceed with step 5 below. If you are not using a cone: Connect the instrument to a computer using DataView. Select the instrument in the Data Logger Network and click Instrument -> Configure. Then go to the Anemometer tab, select Custom, and enter the area of the ventilation register. (Consult the DataView Data Logger Help system for assistance.) When finished, place the sensor in front of the ventilation register, and go to step 7 below. 5. Place the air flow measurement cone on the sensor. Depending on the measurement you want to make (flow in or flow out), place the sensor with the arrow pointing towards the cone (outgoing flow) or away from it (incoming flow). 6. Place the cone on the ventilation register. The arrow inside the propeller must point in the direction of the air flow. This will be towards the outside of the cone for a blowing measurement, or towards the inside of the cone for a suction measurement. 7. The air flow and temperature measurements are displayed on the LCD. (Wait for the measurements to stabilize before accepting the reading.). Recording Measurements You can start and stop a recording session on the instrument. Recorded data is stored in the instrument s memory, and can be downloaded and viewed on a computer running the DataView Data Logger Control Panel. You can record data by pressing the button: 5

A short press (MEM) records the current measurement(s) and date. A long press (REC) starts the recording session. While the recording is in progress, the symbol REC appears at the top of the display. A second long press of stops the recording session. Note that while the instrument is recording, a short press of has no effect. To schedule recording sessions, and download and view recorded data, see the Data Logger Control Panel Help. Repair and Calibration To ensure that your instrument meets factory specifications, we recommend that it be sent back to our factory Service Center at one-year intervals for recalibration, or as required by other standards or internal procedures. For instrument repair and calibration: You must contact our Service Center for a Customer Service Authorization Number (CSA#). This will ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you want a standard calibration; or a calibration traceable to N.I.S.T. (includes calibration certificate plus recorded calibration data). Ship To: Chauvin Arnoux, Inc. d.b.a. AEMC Instruments 15 Faraday Drive Dover, NH 03820 USA Phone: (800) 945-2362 (Ext. 360) (603) 749-6434 (Ext. 360) Fax: (603) 742-2346 or (603) 749-6309 E-mail: repair@aemc.com (Or contact your authorized distributor.) Cost for repair, standard calibration, and calibration traceable to N.I.S.T. are available. NOTE: You must obtain a CSA# before returning any instrument. Technical and Sales Assistance If you are experiencing any technical problems, or require any assistance with the proper operation or application of your instrument, please call, fax, or e-mail our technical support team: Contact: Chauvin Arnoux, Inc. d.b.a. AEMC Instruments Phone: (800) 945-2362 (Ext. 351) (603) 749-6434 (Ext. 351) Fax: (603) 742-2346 E-mail: techsupport@aemc.com 6

01/18 99-MAN 100450 v1 Chauvin Arnoux, Inc. d.b.a. AEMC Instruments 15 Faraday Drive Dover, NH 03820 USA Phone: (603) 749-6434 Fax: (603) 742-2346 www.aemc.com