Rigaku Mercury CCD on Ultima III PXRD in CSL 1011

Similar documents
Precautions The XRD uses x-ray radiation. The instrument is designed to prevent radiation exposure.

Panalytical MRD X-Ray Diffraction SOP

1. Right-click the worksheet tab you want to rename. The worksheet menu appears. 2. Select Rename.

PANALYTICAL X PERT PRO XRD

Panalytical XPert Powder X-Ray Diffraction SOP

Grazing Angle 2 Theta Phase Analysis

Screen Projector in back of room. Document Camera. Computer Monitor. Smart Classroom Quick Start Guide. Art 110. Orientation

Powder X-Ray Diffraction (PXD/XRPD)

Attenuator Correction Package Measurement

Start Up and Shutdown Procedures (Unix)

Printing Achievement Series Answer Sheets

Generic OLLI Classroom Technology Updated March 22, 2019

Retrieve Download File from CSDI Installation Instructions

Smart Classroom Quick Start Guide Wirtz 316

WHAT TO DO IF THE SERVO SYSTEM REPORTS FAILURES.

Mobile Recorder User Manual

Room Classroom Technology Updated March 4, 2019

Screen Projector in back of room. Document Camera. Computer Monitor. Smart Classroom Quick Start Guide. Art 102. Orientation

Installing VPN client by Jupiter Networks:

Room Classroom Technology Updated October 25, 2018

HomeTroller S6 Quick-Start Guide

Fiery proserver User Guide

Projector. Screen. Computer Monitor. Document Camera Smart Classroom Quick Start Guide Still Gym 302. Orientation

Room Classroom Technology Updated March 4, 2019

Room Classroom Technology Updated August 29, 2018

Screen. Projector. Document Camera. Computer Monitor. Smart Classroom Quick Start Guide Music 173. Orientation

Projector Screen. Sympodium Computer Monitor. Document Camera. Smart Classroom Quick Start Guide. Davis 116. Orientation

BASIC ipad TROUBLESHOOTING IN THE CLASSROOM

On occasion, due to a failure or water damage it may become necessary to replace an APEX, APEX LITE, or APEX JR with another controller.

Boston College McGuinn Auditorium

UNINSTALL THE BPR FOR WIN-Rev2.bprm

Maintenance Package Measurement

NetProfiler 3 Getting Started Guide

MicroBloggingMD (MBMD) Support Documentation

Table of Contents. Overview. Installing and Using Afaria with your Motorola XOOM End User Document v0.3

Smart Classroom Quick Start Guide. La Tourette 200. Orientation. Projector. Screen Control on East Wall by double doors. Document Camera.

` EVG 620 MASK ALIGNMENT SYSTEM Bay 1 STANDARD OPERATING PROCEDURE

Scanning Protocol: Cyberware Model 15

Projector. Screen Computer Monitor. Document Camera. Smart Classroom Quick Start Guide. Psych/CompSci 156. Orientation

VIV-CV-1225V v2 Digital Camera Binoculars. Instruction Manual

Information Technology Help Desk George Washington Hall B College Avenue Fredericksburg, VA 22401

AWS DeepRacer Getting Started Guide

XRD-DSC General (BB) D/teX Package Measurement. XRD-DSC General (medium resolution PB/PSA) Package Measurement

Computer Instructions

Qvis NVR Mini Driver Installation and Configuration Guide for Paxton Access Control

Smart Classroom Quick Start Guide. Wirtz 110. Orientation. Projector. Screen Document Camera. Computer Monitor

Equipment Operating Procedure. Model Number: Dage-Series-4000PA Machine Name: Shear Tester Date: 8/17/2013 Serial Number:

Product Manual. Intova Duo

ThinManager Interface

Fiery proserver User Guide

Digital 2.4GHz 3.5 LCD Monitor

Room Classroom Technology Updated March 4, 2019

Operating Procedure PerkinElmer LAMBDA 750

IT Services TRI IT... 3 AV Services... 3 TRI Credential... 4

Smart Classroom. Instruction Manual

P.E.O. STAR Scholarship Online Recommendation Instructions

Title: Karl Suss MA150 Semiconductor & Microsystems Fabrication Laboratory Revision: E Rev Date: 03/01/06

Catalogue. Assembling... 1 Unpack Charge the Battery Install the Memory Card Product Basics Buttons and Connections...

Technology Classrooms Podium at Community College of Philadelphia

Smart Door Lock. Set-Up Guide

Quick Use Guide. WorkCentre 7132

Set Password. After you complete the setup, click OK button, system pops up the following interface for you to confirm. Click OK button to exit

Smart Classroom Quick Start Guide

Quick Installation Guide Direct and Transfer Thermal Printer

B-Works 2 User s Guide

Fundamentals Exam Registration Guide

350 SMART MANOMETER OPERATING INSTRUCTIONS

This job aid will show Portal Administrators and users how to perform self-serve tasks for their account.

AFN Setup. If you have any trouble setting it up please contact the Help Desk by at Outlook 2007 and 2010

ETVR360CAMW-User Guide

Projector Screen Computer Monitor Document Camera. Smart Classroom Quick Start Guide. Wirtz 103B. Orientation

Operating Procedure for Horiba Raman Microscope

Battery Charging The tablet has a built-in rechargeable Li-polymer battery.

XPS 13 Convertible Service Manual

Getting Started Instructions for the Access USB and Access 2 USB Probers

Smart Classroom Quick Start Guide

HomeTroller SEL Getting Started Guide

OneRoof CyberCafePro Client Installation, Setup and User Manual

Computer Essentials Session 1 Lesson Plan

Wirtz 101 (Auditorium)

Polaris G50 Dash cam incident recorder.

Service Bulletin SB685. Date: 8/18/2017 TriPac EVOLUTION Communications Update Bulletin Location: TSA Info Central\Service Bulletins

Operating Instructions

Danalock V3 SCANDI Mounting guide.

Streaming Smart Stick. Quick Start Guide

Evans Field House 128

NEO 4.5. User Manual

Prepulse Inhibition (PPI) Boxes

UNIVERSAL SOFTWARE. Universal Software. Data Sheet

XRD Training Notebook

CESecure Quick Start Guide

1. Introduction P Package Contents 1.

Cut-True 29A & 29H Guillotine Cutters

2MASS Observer s Guide. Steward Observatory 61 Kuiper Telescope

ipad and iphone Basics

Nicolet RAMAN Procedure

Getting Started with VPN

Table of contents. 6. Search & Filter functionality 7. Including truck data to tech support request 8. Troubleshooting 9. Truck Tool Legacy Trucks

Power Xpert Meter 2000 Gateway Card Kit

PANalytical X-Ray Diffractometer with PW3830 Generator. Standard Operating Procedure. Revised on 02/15/2007

Transcription:

Rigaku Mercury CCD@ FSU for Ultima III PXRD [CSL 1011] Quick Shut-down and start-up procedure for Rigaku MercuryCCD on Ultima III XRD [Updated by Thayumanasamy Somasundaram, V.: Apr 18, 2014; Original by B. vande Burgt] Shut-down and Start-up Procedure for frozen Mercury CCD Rigaku Mercury CCD on Ultima III PXRD in CSL 1011 Note: After a general power-failure please follow the Rigaku Ultima III Standard Operating Procedure and NOT the procedure described here. The CCD camera should not freeze during power failures. If the CCD camera freezes during operation, stop the AreaMAX (icon with red RGKExpand) and Exit. Left Click on RGKExpand on the left side of System Tray, then select Stop AreaMax This closes the AreaMax window (better way to close it)

Turn the control computer (Control PC) off by closing all Windows programs and executing a Windows 2000 shutdown. Switch the monitor to the frame grabber computer (the left PC) by pressing <Scroll Lock> twice and execute a Windows 2000 shutdown. Press <Scroll Lock> twice to switch to the Control Computer. Now, in the Ultima III Generator s Left Top Panel, Under X-RAYS o Press the Grey OFF button In the Ultima III Generator s Left Top Panel, under the SAFETY RELEASE panel, press the DOOR button. While the DOOR button is blinking, open the sliding door all the way to left and right. Inside the x-ray enclosure, locate the DALSA power supply (usually located on the right hand side and sits on the base plate) for the CCD detector. Power it down by switching it OFF (the power switch is located in the front of the unit, however, only the unit s back is visible) by locating the switch on the far right hand side.

Now, in the Ultima III Generator s Left Top Panel, Under POWER Press Red OFF button Then, in the Ultima III Generator s Left Bottom Panel, open the door and locate the SERVO AMP UNIT, Turn it OFF by pressing the black power switch to off position. Press the reset button located on the vertical card second from the right (white button next to the label CPU2-TF ).

Now open the Left Hand Side Bottom panel on the generator Press the third switch from top on the leftmost PC card once This should make the control panel flash once Turn the power ON at the SERVO AMP UNIT and close the front door Now, in the Ultima III Generator s Left Top Panel, Under POWER o Press Green ON button o Press the X-Rays ON button (after the READY light is lit) o Now the RED LED light for X-RAYS will come on Turn the control computer (Control PC) on by pressing the power button on the front of the PC. You should see the normal boot up windows (if not press <Scroll Lock> twice on the keyboard, the monitor will switch to Control PC s display). Log in using the username administrator and password (ask PXRD Personnel). Leave the frame grabber PC OFF (for now)

Wait for the Right window to finish two initializations Open the RINT 2200 Right System (located in the Rigaku Control Panel) Change the Slit to Auto Variable Slit (Micro Area/Ultima III) Click OK Then Yes to update Initialization Step Open Manual Measurement and click Initialize. Watch for the rotation of camera beam The initialization is finished when the Right window stops updating Open the RINT2200 Right System (located in the Rigaku Control Panel folder Change the Slit to Auto Variable slit (Mercury CCD (General/UI)) Click OK Then Yes to update it Check the Alignment Settings by going to the Manual Measurement window and selecting View, Setting Results and comparing the values listed there to the values obtained during the last calibration (this values can be found on a sheet of paper in the upper left hand drawer of the PC desk) If the values are good, click OK. If not change

the values to those listed in the calibration paper and reinitialize. (Initialization Step) Close the x-ray shutter, open the door and turn on Dalsa Camera Interface. Close the door. Switch the PC Monitor to Frame Grabber PC (the one of the left) by pressing, <Scroll Lock> twice. Power up the Frame Grabber PC and login using the following credentials Username: Administrator Password: *************** [Ask PXRD Person Responsible] Press <Scroll Lock> twice to switch to Host PC If the x-ray tube has been off more than an hour, the tube needs to be Aged by opening XG Operation control button and executing Aging Icon. Aging the tube properly extends its lifetime. This procedure will take 45 minutes. Otherwise proceed to next step If the Julabo chiller is at or below -4 C, the CCD detector is ready Open Standard Measurement Multi-thread Listening Client connection

Open RAX Video The system in now ready! V. 20140418/TS