Norland Centroc1000. Operating Manual Ver

Size: px
Start display at page:

Download "Norland Centroc1000. Operating Manual Ver"

Transcription

1 1 Norland Centroc1000 Operating Manual Ver Ferrule Concentricity/Eccentricity Measurement Ferrule Cylindricity Measurement Insertion Loss Assessment Ferrule Tuning Means Visual Examination Norland Products Incorporated Phone: Fax: Route 130 Building 100, Cranbury NJ USA

2 2!! CAUTION: DON T touch the ferrules with bare hands at any times. To load a ferrule into the machine, you MUST either wear silicon gloves, or utilize a pair of clear tweezers (supplied).!! CAUTION: DON T RUN Centroc1000 if you haven t placed a ferrule into the front loading fixture of the machine. Clicking ALIGN FERRUEL or RUN TEST without the physical presence of an actual testing ferrule, or a dummy ferrule, may damage the driving belt of Centroc1000!! 2

3 3 INDEX Section No. Section Name Page 1. Introduction 4 2. General Description 4 3. Equipment Setup 5 4. Software Setup 6 5. Centroc1000 Main Screen Menus File Menu Tools Menu Display Menu Options Menu About Menu: Measuring Eccentricity Visual Examination Changing Mounting Fixtures Changing Driving Belt Service Information Specifications Appendixes Tuning Criteria for reference connectors Insertion Loss Assessment 38 3

4 4 1. Introduction The Centroc1000 unit, for noncontact measuring of ferrule endface geometry, is part of Norland Product s range of measuring equipment for Optical Cable Assembly shops. The unit complies with: IEC ISO 2538 IEC IEC IEC General description The Centroc1000 is a special designed auto-focusing microscope with a precision chuck for mm or for mm fiber optic ferrules. In Order to achieve maximum accuracy the magnification in model 1000 is 900 X on a 17 monitor. 4

5 5 3. Equipment Setup Connect the "Centroc1000" and the PC using the cable supplied (9-pin female to Centroc1000 and 9-pin male to the frame grabber card interface at the PC). The Centroc1000 uses the USB interface for motor and lighting control. Connect the USB cable to Centroc1000 and insert the other end into one of the USB ports. If plugging into that port for the first time, the following setup procedure may be required. Connect monitor, mouse, printer and keyboard to the PC. Connect the Printer and the Personal Computer to AC- power mains. Verify the computer monitor Screen Area is set to 1024x768 with Colors set to True Color (24 bit) 5

6 6 4. Software Setup The Centroc1000 system comes with any required software already installed onto its associated computer. The Centroc1000 application is designed to operate under Windows 2000 Pro or Windows XP Pro. The minimum requirement for PC: Intel P III 800 MHz. In the event the Centroc1000 software needs to be reinstalled, it must be installed in the following order: 4.1 Euresys drivers for the Picolo frame-grabber. 4.2 Centroc1000 measurement application. If updating just the Centroc1000 application, only step 2 need be performed. Check for any new versions on the Norland Products web-site Euresys drivers for Picolo frame-grabber Installation from the Euresys CD: a. Place the CD, and select INSTALL to install the software 6

7 7 b. Select Run-time and select evision and Multicam for Picolo and Click Next c. Click NEXT on the following screen 7

8 8 d. Select Easy Grab, Drivers, ActiveX and Click NEXT on the following screen. e. Leave the default and click NEXT on the following screen 8

9 9 f. After successful installation the click OK on the following message and click QUIT from the initial installation screen to close the screen. 4.2 Centroc1000 Software Installation Installation from the Centroc1000 CD: a. Insert the CD to install the software. Click Next to the following display 9

10 10 b. Choose I accept and click Next c. Enter the Customer Information (or accept the default) and click Next. 10

11 11 d. Click Next to start the installation e. Click Finish to complete the installation Configuring the Centroc1000 USB Interface: a. Plug in the USB cable: Plug in one end of the USB cable to the Centroc1000 instrument and its other end into one of the computer s USB ports. Once the cable is plugged in, the following message will be appear at the bottom of the screen. 11

12 12 b. Install the new USB driver ELEKTOR: The following screen prompts the user to install the new drivers. Choose Install the software automatically and click Next. System is now searching for the pertinent drivers: Click Next after it has found the files. 12

13 13 Click Continue Anyway Click Finish to complete. Display of the following message shall indicate that the installation has been successful. 13

14 14 5. Centroc1000 Main Screen Upon activating the application, the following main screen will show up. Menubar Image: Menu bar: Align Ferrule: Live image from the camera. Collection of pull-down menus of various program functions and options.!! CAUTION: DON T touch the ferrules with bare hands at any times. To load a ferrule into the machine, you MUST either wear silicon gloves, or utilize a pair of clear tweezers (supplied).!! CAUTION: STOP!! DON T move forward to running the machine if you haven t actually placed a ferrule into the unit. Clicking ALIGN FERRUEL or RUN TEST without the physical presence of a real ferrule, at least a dummy ferrule, may damage the driving belt of Centroc1000!! After inserting a ferrule, click this button to bring its end face fully into the fixture before starting a measurement. 14

15 15!! CAUTION: To ensure every ferrule is well aligned and brought into focus in the machine, click Align Ferrule at least once before RUN TEST on each new ferrule!! Run Test: Eject: Focus: Click this button to calculate the eccentricity or concentricity for the ferrule. Ferrule rotates automatically. While rotating, several measurements are taken at each rotation point. The points along with eccentricity values are displayed on a separate Results window. Before clicking Run test make sure to click the Align Ferrule button first. Use this option to remove the ferrule from the mount. Click to activate focusing. CAUTION: DON T perform focusing on a ferrule until it has been aligned in the unit. See Align Ferrule above. NOTE: A refocusing is normally NOT required unless (a) a new batch of samples are to be tested; or (b) a new work shift has just started and the machine has been just boot up. Tuning On/Off: Back light: Status of Tuning the ferrule for marking is displayed. Tuning On for a measurement means rotate the ferrule, so the end ferrule eccentricity or concentricity is aligned at 12 o clock position at the end of the measurement. Tuning On is required for the ferrules, which need marking. Tuning Off for normal measurement. To select Tuning On or Off, use the Perform tuning at the end of measurement option from Options/Menu from tool bar. Click + or - buttons to adjust the level of illumination. 15

16 16 6. Menus 6.1. File Menu Exit: To close the application Tools Menu Change Mount On/Off This option, when selected, will activate an Image circle template and onscreen instructions that assist the user in changing the mounting fixture. (See Section 7.Changing Mount Fixtures) Inspection Mode On/Off Choose this option to increase illumination on the connector endface. When selected, use additional set of buttons ( +/- ) to make smaller increases or decreases to the illumination level as needed. 6.3 Display Menu 16

17 Statistics When selected, a separate form with statistical report of the results history is displayed (see figure below). At top left in the above figure, a table lists the results from all stored measurements. Statistical displays from a subset of those measurements are at top center as well as the two bottom displays. The amount of measurements used in the statistical displays can be selected by the operator (by entering how many of the last measurements into the top center box). From the graph at top right, an estimation of a particular ferrule s worst-case insertion loss can be made using the its eccentricity result. See the Theory Section in Appendix A for more information. a. File Menu File Menu from the Statistics display contains choices of Data, Print, & Exit. i. Data: On choosing Data the following options are shown Clear List: To clear results from the form. Restore Last list: To restore last cleared results. 17

18 18 ii. Print: On choosing Print the following options are shown Print Form: Print the current form. Print all records: Print all the stored results from the database. iii. Print (sort): This option will show Data Record Selection to print only data from a particular date range or with a particular Product ID. Print: After choosing the Date or Product ID click Print to print the data. Exit: Click to Exit from this form. 18

19 Last Results This display shows a graphical representation of the concentricity or eccentricity measurements, as well as other measured or calculated quantities from the most recent scan. Note: Either Concentricity or Eccentricity values will be displayed depending on the Measurement Type, as configured by the Operator in Options/Menu before the measurements. There are three main parameters that have been displayed from the last measurement. They are, Concentricity /Fitted (or: Eccentricity/Fitted): The mean Concentricity, or Eccentricity, value of the ferrule obtained by means of a least-squared circular fitting to all rotation steps of the scan (µm). Ferrule Cylindricity: Assessed cylindricity of the ferrule. Ferrule Cylindricity is a measure of the deviation of the ferrule outside diameter (OD) from a perfect round cylinder (µm). End Angle: Angle of Concentricity, or Eccentricity, measured at the end point of the ferrule rotation, as illustrated by the RED dot on the screen. Note: When TUNING ON is selected in the option menu, this End Angle is always aligned to the 12:00 o clock position (or, zero degree position.) This Tuning option provides the means for the operator to add a mark on the ferrule at the orientation of the measured Concentricity/Eccentricity. Other less significant parameters are also displayed, for reference purpose. They are, Concentricity /Start: The first measured concentricity during scan. Concentricity /End: The last measured concentricity during scan. Maximum Concentricity: Maximum measured concentricity during scan. Minimum Concentricity: Minimum measured concentricity during scan. Calculated Insertion Loss: Assessed Insertion Loss based on fitted Concentricity/Eccentricity. Ferrule Type: Type of the ferrule as entered from the Options Menu. Product ID: ID of the sample as entered before the measurement. Operator: Operator s Name as entered in the Options Menu. Centroc 1000 model sn: Serial Number of the Centroc1000 machine. 19

20 20 Press [Print] to print the current screen. Press [Next] to return to the start screen to measure the next product. Press [Stats] to see statistical presentation of results. Press [Excel Viewer] to view the saved scan results in Microsoft Excel. The text file viewed can be selected or changed in the Options/Menu under Save Results to section. Excel Viewing the Text File 20

21 Options Menu Adjust the different parameters as required per connector specification and as desired by the operator. Nomenclature for the Options Menu: SerialNumber Picolo: Picolo frame grabber serial number from the Board. SerialNumber Centroc: Centroc1000 serial number 21

22 22 Calibration Factor: The magnification calibration factor entered at the factory for the unit. It should match the value marked on its rear panel. If there is any question about the correctness of this value or need to recalibrate, contact Norland Products Technical Support for assistance. Ferrule Type: Choose the appropriate selection for the sample being tested. Pass /Fail Criteria: Enter desired Min and Max limits for the average Eccentricity ( Ecc/Fitted ) or Concentricity depending on Measurement Type choice. At measurement completion the result will be compared against these limits and displayed on Red (Fail) or Green (Pass) background. Perform Tuning at the end of measurement: Select this option to rotate the ferrule at the end of the measurement, so the ferrule end Eccentricity or Concentricity is aligned at 12 o clock position. The status of Tuning is displayed on the main screen below the live image. Measurement Type: Depending on the measurement type chosen the scan results will be displayed on Results screen and Statistics screen as shown below. For Eccentricity Display: 22

23 23 For Concentricity Display: Show Maximum/Minimum: Maximum and Minimum Eccentricity values will only be shown on the Results screen if this option is checked. 23

24 24 Product ID: Information for Product ID a. Prompt For Input Select this option if it is desired to be prompted for a Product ID before each new measurement. At the time of measurement, the following prompt will be presented. After entering an ID (or accepting the value already there), click OK to continue with the measurement or Cancel to exit from the measurement. b. Auto Increment: Check this option if the ID number is to be incremented automatically for the next measurement. Save Results To: Choose/enter the ASCII text filename and location where measurement data will be stored. This file can be later imported into other programs like Microsoft Excel for additional processing and can be viewed through Results screen. Operator: Select this option if it is desired to include an Operator s Name among the displayed and stored results of a measurement. Defaults: Load the factory defined defaults. Note: Different defaults will be loaded depending on choice for Ferrule Type. Cancel: Apply: Close the form without saving any changes. Save any changes and close the form. 24

25 25 Note: It is possible to select and then save (using this Apply button) two completely different configurations of options; one for the SM and another for MM ferrule type choices. Then it is easy to switch from one configuration to the other simply by changing Ferrule Type selection. 25

26 About Menu: Click Help/About to show the software product information. 26

27 27 7. Measuring Eccentricity/Concentricity Go to the measurement screen (as pictured above). If the current display is the Statistics screen, click EXIT. If, instead, it is the Results screen, click NEXT. Set the required options in Options screen. Open the mounting fixture by lifting the lever to the left of the fixture. Place the ferrule into the mounting fixture s V-groove, let the lever down to hold the ferrule in place.!! CAUTION: DON T touch the ferrules with bare hands at any times. To load a ferrule into the machine, you MUST either wear silicon gloves, or utilize a pair of clear tweezers (supplied).!! CAUTION: STOP!! DON T proceed from this step forward if you haven t actually placed a ferrule into the machine. Clicking ALIGN FERRUEL or RUN TEST without the physical presence of a real ferrule, at least a dummy ferrule, may damage the driving belt of Centroc1000!! Click Align Ferrule before measuring.!! CAUTION: To ensure every ferrule is well aligned and brought into focus in the machine, click Align Ferrule at least once before RUN TEST on each new ferrule!! 27

28 28 Click Run Test. The ferrule will rotate automatically. While rotating, several measurements are taken at each rotation point. The points along with eccentricity values are displayed on a separate Results window. See Last Results in section

29 29 8. Visual Examination Example of ferrule image on a Centroc1000 Length Measurement Planar dimensions on the ferrule such as ferrule diameter, adhesive ring thickness, and scratch width can be checked using the live Image and a mouse. By clicking the left mouse button at two points on the Image, the distance between those two points can be measured. Additional Tools A variety of additional tools can be found by clicking the right mouse button while the cursor is on the live Image. Features such as zooming, panning, and Image copying can be selected and used as part of the visual examination process. 29

30 30 9. Changing Mounting Fixtures This procedure should be followed when the mounting fixture needs to be changed for the testing of different ferrule geometries, for example, from 1.25mm diameter to 2.50mm diameter ferrules. Pull down Change Mount menu. Remove front main-panel. Remove holding-down spring from driving-lever assembly. Using one hand, lift left-end handle of driving-lever assembly upwards till it stops, and hold the handle at that position. Using M3 Hex Key, replace v-groove mount by new mount. Release driving-lever assembly. Replace holding-down spring on driving-lever assembly. Insert a testing ferrule of appropriate size into v-groove mount, and click Align Ferrule. If needed, perform manual focusing of ferrule image by turning front-bottom knob. If needed, perform manual centering of ferrule image by turning micrometers at both sides behind v-groove mount. Replace front main-panel, and press Close to resume measurements. 30

31 31 Mounting heads and Driving Belts 31

32 Changing Driving Belt CAUTION: Belt has to be changed with a pair of clean tweezers (supplied). Touching of the belts by bare hands is prohibited at all times. This procedure should be followed when the synthetic rubber driving belt at the front end of the machine is worn out and needs to be changed. NOTE: Observations of skidding or otherwise jumping around of the ferrule or the belt itself are normally indications of a worn out belt. Remove front main-panel. Remove holding-down spring from driving-lever assembly. Using one hand, lift left-end handle of driving-lever assembly upwards till it stops, and hold the handle at that position. Replace driving belt by a new belt. Release driving-lever assembly. Replace holding-down spring on driving-lever assembly. Insert a testing ferrule of appropriate size into v-groove mount, and click Align Ferrule several times. (NOTE: Click Align Ferrule 3 ~ 4 times shall help aligning the new belt well inside the grooves of the driving pulleys.) If needed, perform manual focusing of ferrule image by turning front-bottom knob. If needed, perform manual centering of ferrule image by turning micrometers at both sides behind v-groove mount. Replace front main-panel, and resume measurements. 32

33 Service Information Item Description Accessories Std Fixture for 1.25mm ferrules (LC and MU connectors) Std Fixture for 1.60mm ferrules (supports M29504/4 and M29504/5 connectors) Std Fixture for 2.00mm ferrules (supports M29504/14 and M29504/15 connectors) Std Fixture for 2.50mm ferrules (supports SC and FC connectors)

34 Specifications Power Requirements Supported Fiber Types Temperature Humidity Wavelength (LED) Measurement Uncertainty (STD) Minimum PC Requirements Operating system Standards/Reference IEC IEC IEC IEC ISO V DC (supplied by the PC via the Frame Grabber connector) Suitable for most SM and MM fiber connector ferrules (Supports Maximum fiber diameter up to 140µm) C Operating 5-55 C Storage RH Maximum 90%; no condensation λ 475 nm 0.1 µm (Eccentricity), 0.2 µm (Concentricity) Intel P4 or AMD Athlon Preferably +2 GHz. 1 free PCI slot for Picolo frame-grabber. Windows 2000 or XP Pro 34

35 Appendixes 13.1 Tuning The worst insertion loss in a production population resulting from the core-to-core misalignment between any two mating patchcords, that have been randomly selected from the population, can be reduced significantly via a tuning process of all patchcords in the population. When a single mode fiber is operating at 1310nm wavelength, the insertion loss of a connector measured against an ideally concentric reference connector (i.e.; with zero eccentricity) is illustrated in the figure below. (NOTE: More insertion loss charts of other fiber types can be found in Appendix 13.3.) INSERTION LOSS OF SM Insertion Loss (db)) Eccentricity (um) CASE I: Un-tuned Connectors When a connector is not tuned, its angle of eccentricity can be anywhere within +/- 180 of its key orientation. The worst case core-to-core offset distance between any two randomly picked patches from a large connector population can be estimated by adding the two highest eccentricities in the population (See Graph Illustration Below, CASE 1: UNTUNED CONNECTORS). For example, when the maximum eccentricity (Ecc) of an untuned population is 0.90um, the maximum possible core-to-core offset distance between two random connectors is: OFFSET = =1.80 µm 35

36 36 Then, the maximum possible insertion loss between two randomly chosen connectors from the given untuned population, in accordance with the Insertion Loss (IL) chart above, is: IL = 0.66 db CASE II: Tuned Connectors When a group of connectors are all tuned, the eccentricity orientation of all the connectors are all aligned within a small fan area within the vicinity of the key position. For example, when the maximum eccentricity in a population is 0.90um and the connectors are all tuned to within +/-30 around the key position, the maximum possible core-to-core offset distance between any two randomly selected connectors in the tuned population, is (See Graph Illustrations Below, CASE 2: TUNED CONNECTORS): OFFSET = 0.90 µm (FACT: THIS VALUE IS 1/2 OF THAT IN CASE I). Subsequently, the maximum possible insertion loss between the two randomly selected connectors in the tuned population, in accordance with the Insertion Loss (IL) chart above, is: IL = 0.17 db (FACT: THIS VALUE IS 1/4 OF THAT IN CASE I). FACT: The Maximum Insertion Loss between any two randomly selected fiber connectors in a tuned population has been reduced to one-quarter (1/4) of the Maximum Insertion Loss in an corresponding untuned population. 36

37 Criteria for reference connectors A connector is suitable for reference purposes if the requirements for reference products as presented in the relevant specification or related IEC documents are met. For the case of a user having their own purchasing specification, use the criteria for reference components from this document. If no specification for reference is given, a reference can be selected from the table below. The maximum allowable eccentricity for a batch of connectors is dependent not only on the connector under test, but also on the reference connector used. Example: If the Loss Specification is 0.20 db according to IEC method 7 (first column in table), then the maximum allowed (total) eccentricity as 0.95 µm. This is the sum of the maximum eccentricity of the connector under test and the maximum eccentricity of the reference connector (from the table). This example shows that a reference connector should be selected with an eccentricity value as low as possible. Careful selection of a reference connector with the lowest possible value for eccentricity results in best yield (more connectors that will be considered to have acceptable eccentricity) in the final loss measurement. Loss against concentric reference connector. (IEC method 7) Maximum loss in the field 2% Single Mode Connectors. Maximum Eccentricity Reference connector Maximum eccentricity of connector under test 0.10 db 0.40 db 0.10 µm 0.58 µm 0.20 db 0.77 db 0.20 µm 0.75 µm 0.30 db 1.16 db 0.20 µm 0.98 µm 0.40 db 1.75 db 0.25 µm 1.11 µm 0.50 db 2.10 db 0.30 µm 1.24 µm 37

38 Insertion Loss Assessment In a fiber optic network, the center-to-center alignment between the outside diameters of two mating ferrules are normally very finely maintained at the joint section between two fiber optic connectors, by means of the well established precision coupling mechanism of a spring-loaded split sleeve (see Figure below). However, perfect eccentricity between two mating ferrules DOES NOT always promise perfect concentric mating between two transmitting fiber optic cores, due to the often non-perfect eccentric offset from the center of the ferrule Outside Diameter (OD) to the center of the core for each individual patchcord. In the field, the power Insertion Loss (IL), while the optical signal is transmitted from one fiber core to another fiber, is heavily dependent upon the precision of the ferrule-to-core eccentricities of both individual patchcords obtained from the manufacturing process. Alignment sleeve Ferrules In this Manual, Eccentricity of a patchcord is defined as the offset distance from the center of the ferrule outside diameter to the center of the fiber optics core, and is measured in micrometers (µm). And Concentricity, also measured in micrometers, is defined as two times the Eccentricity. For simplicity, we shall exploit letter E to denote the Eccentricity of the fibers in all formulas involved in this section. The insertion loss (IL) formulas provided in this section are all measured between a testing fiber, whose ferrule-to-core eccentricity is E(um), and an ideal reference fiber, whose eccentricity is defined as, ideally, 0(um). When Should I Apply the Maximum Eccentricity or the Maximum Concentricity to Calculate the Maximum Insertion Loss? In field applications, the worst case insertion loss obtained between two randomly selected fibers from a tuned connector population can be very different than that from a non-tuned connector population (see section 12.1 for detailed merit of the tuning). We shall utilize symbol E max to delineate the maximum value of the measured eccentricities in a given population. In a tuned connector population, the eccentricities of all connectors are tuned to within +/-30 angular range (or smaller) from the key position. The resulting 38

39 39 maximum core-to-core offset between any two mating fibers in the population is always less than, or equal to, E max. Thus, the maximum possible insertion loss between any two randomly selected connectors from the population can be computed by substituting directly E max in places of symbol E in the formulas. In a non-tuned connector population, however, the worst case core-to-core mismatch between two randomly selected patches can be as large as two times of the actually measured maximum eccentricity in the population. Therefore, in computing the maximum insertion loss between two randomly selected connectors from the population, one has to substitute two-times of the Maximum Eccentricity (2E max ), or referred to as Maximum Concentricity, in places of symbol E in the formulas INSERTION LOSS BETWEEN TWO SINGLE-MODE FIBERS: When a single-mode fiber is used for 1310nm wavelength transmission, the insertion loss (IL) formula is (ModeFieldDiameter=9.2um; Based on a Gaussian Beam Profile Model): IL (db) = *E² When a single-mode fiber is used for 1550nm wavelength transmission, the insertion loss (IL) formula is (ModeFieldDiameter=10.4um; Based on a Gaussian Beam Profile Model): IL (db) = *E² INSERTION LOSS OF SM FIBERS Insertion Loss (db)) Eccentricity (um) SM@1310nm SM@1550nm 39

40 INSERTION LOSS BETWEEN TWO 50um/125um MULTI-MODE FIBERS: When the 50/125 fibers are SI (Step Index), the insertion loss (IL) formula is (Based on a Cylindrical Uniformly Distributed Beam Profile Model): IL (db) = ABS{10* LOG 10 ( *E)} (NOTE: LOG 10 denotes base-10 Logarithm Operation; same below.) When the 50/125 fibers are GI (Graded Index), the insertion loss (IL) formula is (Based on a Downward-Opening Parabolic Beam Profile Model): IL (db) = *E² INSERTION LOSS OF MM-50/125 FIBERS Insertion Loss (db)) Eccentricity (um) MM/SI 50/125 MM/GI 50/125 40

41 INSERTION LOSS BETWEEN TWO 62.5um/125um MULTI-MODE FIBERS: When the 62.5/125 fibers are SI (Step Index), the insertion loss (IL) formula is (Based on a Cylindrical Uniformly Distributed Beam Profile Model): IL (db) = ABS{10* LOG 10 ( *E)} When the 62.5/125 fibers are GI (Graded Index), the insertion loss (IL) formula is (Based on a Downward-Opening Parabolic Beam Profile Model): IL (db) = *E² INSERTION LOSS OF MM-62.5/125 FIBERS Insertion Loss (db)) Eccentricity (um) MM/SI 62.5/125 MM/GI 62.5/125 41

42 INSERTION LOSS BETWEEN TWO 100um/140um MULTI-MODE FIBERS: When the 100/140 fibers are SI (Step Index), the insertion loss (IL) formula is (Based on a Cylindrical Uniformly Distributed Beam Profile Model): IL (db) = ABS{10* LOG 10 ( *E)} When the 100/140 fibers are GI (Graded Index), the insertion loss (IL) formula is (Based on a Downward-Opening Parabolic Beam Profile Model): IL (db) = *E² INSERTION LOSS OF MM-100/140 FIBERS Insertion Loss (db)) Eccentricity (um) MM/SI 100/140 MM/GI 100/140 ************ We continuously upgrade software and documentation. Check on for the latest versions. 42

Norland CentrocGP/GPM. Operating Manual Ver

Norland CentrocGP/GPM. Operating Manual Ver Norland CentrocGP/GPM Operating Manual Ver 5.63.32 Connector Eccentricity/Concentricity Measurement Ferrule Cylindricity Assessment Insertion Loss Estimation Tuning of Connectors PM Fiber Alignment and

More information

ZX-1 Array Plus Zoom Interferometer

ZX-1 Array Plus Zoom Interferometer ZX-1 Array Plus Zoom Interferometer Key Benefits: Single unit can measure multi-fibre MT, Mini-MT, MT-RJ, MPO, MPX and MTP array connectors AND single-fibre PC, APC connectors, ferrules and ferrule blanks

More information

Table of Contents. Section Description Page

Table of Contents. Section Description Page Table of Contents Section Description Page 1.0 Getting the System Up and Running. 4 ZX-1 mini PMS Cable Connections. 4 Making a Measurement on a PC Connector. 9 Making a Measurement on an APC Connector.

More information

LS-785 High Throughput Lens Spectrograph User Manual

LS-785 High Throughput Lens Spectrograph User Manual 15 Discovery Way, Acton, MA 01720 Phone: (978)263-3584, Fax: (978)263-5086 Web Site: www.piacton.com LS-785 High Throughput Lens Spectrograph User Manual Ver. 1.2 Table of Contents Chapter 1 -- General...5

More information

CONNECTORVIEW. Operation Guide. VIP-45 Software V

CONNECTORVIEW. Operation Guide. VIP-45 Software V CONNECTORVIEW VIP-45 Software Operation Guide V.6.23.16 Table of Contents 1. VIP-45...3 2. Install ConnectorView...5 3. General settings...8 3.1 Setting Options...9 3.2 Main window controls...12 3.3 Image

More information

NORLAND PRODUCTS INC 2540 Route 130, Bldg 100 P.0. BOX 637 Cranbury, NJ TEL , FAX

NORLAND PRODUCTS INC 2540 Route 130, Bldg 100 P.0. BOX 637 Cranbury, NJ TEL , FAX NORLAND PRODUCTS INC 2540 Route 130, Bldg 100 P.0. BOX 637 Cranbury, NJ 08512 TEL. 609-395-1966, FAX 609-395-9006 ACCISNT RELEASE NOTES Changes for ACCISNT v6.02: Changed SETUP menu to allow selection

More information

B & C Grade Tuned Connectors

B & C Grade Tuned Connectors B & C Grade Tuned Connectors White paper White Paper B & C Grade Tuned Connectors v1.0 1 Introduction Modern systems are exponentially increasing the amounts of data being transmitted, requiring ultrahigh

More information

Cable Assembly Test System. MS12001 System

Cable Assembly Test System. MS12001 System Cable Assembly Test System KEY FEATURES MS10P is a powerful controller unit with 10 slots MS10E Expansion Chassis for up to 10 extra slots Remote control via Ethernet/GPIB card Turn-key software Local

More information

LEAD Fiber Optics PRODUCT CATALOGUE

LEAD Fiber Optics PRODUCT CATALOGUE LEAD Fiber Optics PRODUCT CATALOGUE OPTICAL TERMINATOR TEL:+886-2-8672-2371 FAX:+886-2-8672-3275 Web:http:// Optical Terminator Optical Terminators are used to terminate unused fiber connector ports in

More information

Complete Barrel Measuring and Inspection System. PS Series. User s manual

Complete Barrel Measuring and Inspection System. PS Series. User s manual Complete Barrel Measuring and Inspection System PS Series User s manual SAFETY PRECAUTIONS... 3 ELECTROMAGNETIC COMPATIBILITY... 3 LASER SAFETY... 3 GENERAL INFORMATION... 3 BASIC DATA AND PERFORMANCE

More information

Handheld Video Microscope

Handheld Video Microscope Handheld Video Microscope SAFETY SUMMARY Personnel concerned with the operation of this instrument must thoroughly understand and follow the safety guidelines below. The manufacturer assumes no liability

More information

RIFOCS Corp. Component Services Division

RIFOCS Corp. Component Services Division RIFOCS Corp. Component Services Division Overview The Components Division is one of four business units within RIFOCS Corporation. The other three divisions are Instruments, Harsh Environments, and International.

More information

OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide

OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide 2004-2009, AFL Telecommunications, all rights reserved. COM4-00-1001 Revision E, 2009-06-16 Specifications

More information

ConnectorMax Analysis Software DELIVERING FAST PASS/FAIL ASSESSMENT OF CONNECTOR ENDFACES

ConnectorMax Analysis Software DELIVERING FAST PASS/FAIL ASSESSMENT OF CONNECTOR ENDFACES DELIVERING FAST PASS/FAIL ASSESSMENT OF CONNECTOR ENDFACES 2011 GLOBAL PORTABLE FIBER OPTIC TEST EQUIPMENT MARKET SHARE LEADERSHIP AWARD ConnectorMax delivers clear-cut pass/fail verdicts, eliminating

More information

Contents. - i - Ver.:2

Contents. - i - Ver.:2 Contents 1 Accessories of D-Point 2... 2 2 Connecting Projector to PC... 4 3 Pen Action... 4 4 Driver Installation... 5 5 Configuration Setting... 8 6 D-Point 2 Pull-up Menu... 12 7 DT02 Multi-touch Interactive

More information

SpectraTec X. Multi- λ Laser Light Source BLUE SKY RESEARCH WAVELENGTHS

SpectraTec X. Multi- λ Laser Light Source BLUE SKY RESEARCH WAVELENGTHS Multi- λ Laser Light Source BLUE SKY RESEARCH Up to four individual laser wavelengths One single mode PM fiber output Ultra-stable power output Compact : 165mm x 125mm x 34mm Integrated control and drive

More information

LEAD Fiber Optics PRODUCT CATALOGUE

LEAD Fiber Optics PRODUCT CATALOGUE LEAD Fiber Optics PRODUCT CATALOGUE FIBER OPTIC ATTENUATOR Fixed plug-in type TEL:+886-2-8672-2371 FAX:+886-2-8672-3275 Web:http://www.twfiberoptic.com Fixed plug in type Attenuator Fixed plug in fiber

More information

Instruction Manual R&M Connectors. R&M Fiber Optic Components. System security for future requirements. Version 2.4

Instruction Manual R&M Connectors. R&M Fiber Optic Components. System security for future requirements. Version 2.4 Instruction Manual R&M Connectors R&M Fiber Optic Components System security for future requirements Version 2.4 1/13 R&M Fiber Optic Components... 1 System security for future requirements... 1 1. Foreword...

More information

Optical CWDM Power Meter

Optical CWDM Power Meter User Manual Optical CWDM Power Meter TheFibers Inc. A-1109 Keumkang Penterium IT Tower, 282 Hagui-ro, Dongan-gu, Anyang-City, Korea Tel: +82-31-381-6108 Fax: +82-31-381-6109 Email: sales@thefibers.com

More information

ViewConn Pro VC-8200

ViewConn Pro VC-8200 ViewConn Pro VC-8200 1. Introducing ViewConn 2. Using ViewConn Pro 3. Optional Integrated Optical Power Meter 4. Specifications Safety Fiber optic lasers use light wavelengths not visible to the human

More information

OPTICAL CONNECTOR CONTAMINATION

OPTICAL CONNECTOR CONTAMINATION OPTICAL CONNECTOR CONTAMINATION 6 What do the standards say? The test procedures specified by 568-C require tools and procedures for ensuring connector quality. End faces on cords and trunks shall be in

More information

INSPECTION TOOLS EYL10X. 10X Eye Loupe

INSPECTION TOOLS EYL10X. 10X Eye Loupe INSPECTION TOOLS Fiber Scope with 200X Magnification 6X and 10X Eye Loupes Head Magnifier and Inspection Mirror SDPK Scratch-Dig Paddle 10125HG SMA Height Gauge FS200-LC LC Adapter for FS201 Scope MAG200K

More information

User Guide. Video Borescope. Models BR200 and BR250

User Guide. Video Borescope. Models BR200 and BR250 User Guide Video Borescope Models BR200 and BR250 Introduction Congratulations on your purchase of this Extech BR200 (17mm camera head) or BR250 (9mm camera head) Video Borescope. This instrument was designed

More information

VISIBLE FIBER OPTIC FAULT LOCATOR

VISIBLE FIBER OPTIC FAULT LOCATOR VISIBLE FIBER OPTIC FAULT LOCATOR Features: High visibility (up to 6 km with a 1 mw, 635 nm source) Higher output up to 30 mw, non-contact style power versions available Continuous light or pulse modulation

More information

Autofocus Video Inspection Probe G0382A. Product Introduction

Autofocus Video Inspection Probe G0382A. Product Introduction focus Video Inspection Probe G0382A Product Introduction What is Video Inspection Probe? Scratches and stains to optic fiber ferrule endfaces are often said to have a negative impact on transmission quality.

More information

QUALITY ASSURANCE OF MULTIFIBER CONNECTORS

QUALITY ASSURANCE OF MULTIFIBER CONNECTORS QUALITY ASSURANCE OF MULTIFIBER CONNECTORS Eric A. Norland/ Daniel Beerbohm Norland Products Inc. Cranbury, NJ 08512 ABSTRACT Industry requirements for compact, high fiber count connectors in both telecom

More information

BENCHMIKE 283 SERIES. Off-Line ID/OD/Wall measurement systems for pipe and tube samples

BENCHMIKE 283 SERIES. Off-Line ID/OD/Wall measurement systems for pipe and tube samples BENCHMIKE 283 SERIES Off-Line ID/OD/Wall measurement systems for pipe and tube samples Measure manufactured parts or cut samples fast and with the highest accuracy Perform reliable, non-contact measurements

More information

General SC F728 & FC F729

General SC F728 & FC F729 General SC F728 & FC F729 Radiall, thanks to its new range of SC and FC connectors has greatly improved these traditional fiber optic connectors in terms of technology and reliability. One-piece connector

More information

1 : LC/PC 0 : None S : Simplex 09 : 0.9mm G : Green W : White. A : LC/APC S : SM D : Duplex 20 : 2.0mm L : Blue. M : MM 30 : 3.

1 : LC/PC 0 : None S : Simplex 09 : 0.9mm G : Green W : White. A : LC/APC S : SM D : Duplex 20 : 2.0mm L : Blue. M : MM 30 : 3. LC Connector The LC connector was the first small form factor connector on the market. The use of a high performance 1.25 mm ferrule gives the connector extremely high performance characteristics. The

More information

PixelPro 3.0 User Manual

PixelPro 3.0 User Manual PixelPro 3.0 User Manual To ensure proper use of this instrument as well as to avoid injury while operating instrument, understanding this manual completely before use is highly recommended. Table of Contents

More information

2.2. Facilities Requirements

2.2. Facilities Requirements 2.2. Facilities Requirements Facilities requirements for the alpha-se system are listed in Table 2-1 and the system dimensions are given in Figure 2-1. As shown in Figure 2-2, the preferred clear work

More information

TL Laser Systems for Tool Measurement

TL Laser Systems for Tool Measurement TL Laser Systems for Tool Measurement Tool monitoring with a TL laser system is a very flexible solution. The contact-free optical measurement enables you to check even the smallest tools rapidly, reliably

More information

User's Guide. Video Borescope. Models BR200 and BR250

User's Guide. Video Borescope. Models BR200 and BR250 User's Guide Video Borescope Models BR200 and BR250 Introduction Congratulations on your purchase of this Extech BR200 (17mm camera head) or BR250 (9mm camera head) Video Borescope. This instrument was

More information

OLP-55 Sm a r t Optical Power Meter A Sm a r t, Future-Proof Optical Power Meter

OLP-55 Sm a r t Optical Power Meter A Sm a r t, Future-Proof Optical Power Meter COMMUNICATIONS TEST & MEASUREMENT SOLUTIONS OLP-55 Sm a r t Optical Power Meter A Sm a r t, Future-Proof Optical Power Meter Key features Industry s first auto-zeroing function provides outstanding accuracy

More information

ABM's High Resolution Mask Aligner Features:

ABM's High Resolution Mask Aligner Features: ABM's High Resolution Mask Aligner is a very versatile instrument with interchangeable light sources which allow Near-UV (405-365 nm) as well as Mid- and Deep-UV (254 nm, 220 nm) exposures in proximity

More information

BENCHMIKE PRO NEW. The Industry s leading off-line diameter & ovality measurement system

BENCHMIKE PRO NEW. The Industry s leading off-line diameter & ovality measurement system BENCHMIKE PRO NEW The Industry s leading off-line diameter & ovality measurement system Industry s leading sample inspection system Measure manufactured cut samples fast and with the highest accuracy in

More information

501, , 1052, , 1602, 1604 EXCEL EXCEL 1602UC EXCEL 1052UC EXCEL 501HC. Micro-Vu Corporation. Precision Measurement Systems

501, , 1052, , 1602, 1604 EXCEL EXCEL 1602UC EXCEL 1052UC EXCEL 501HC. Micro-Vu Corporation. Precision Measurement Systems 501, 502 1051, 1052, 1054 1601, 1602, 1604 1602UC 1052UC 501HC Precision Measurement Systems 501, 502 1051, 1052, 1054 1601, 1602, 1604 Excel 501 HM/HC Excel 502 HM/HC Excel 501 Excel 502 Scale Resolution

More information

DigiPot Software Configuration Menu

DigiPot Software Configuration Menu C Option Centers the beam axially with the laser module s housing Available for the following modules: PM, PMT, SPM, & SPMT D Option Settings stored in nonvolatile memory Works with customer-supplied computer

More information

BENCHMIKE 283 SERIES. BETA LaserMike. Off-Line ID/OD/Wall measurement systems for pipe and tube samples

BENCHMIKE 283 SERIES. BETA LaserMike. Off-Line ID/OD/Wall measurement systems for pipe and tube samples BENCHMIKE 283 SERIES Off-Line ID/OD/Wall measurement systems for pipe and tube samples Measure manufactured parts or cut samples fast and with the highest accuracy Perform reliable, non-contact measurements

More information

Contents. Text Notations. Copyright. Using the digitizer pen (Optional) 10. Contents 3. Basic Items 4. Using the Slate PC Dock (Optional) 11

Contents. Text Notations. Copyright. Using the digitizer pen (Optional) 10. Contents 3. Basic Items 4. Using the Slate PC Dock (Optional) 11 Contents Contents 3 Basic Items 4 Overview 5 Setting up the computer 8 Using the digitizer pen (Optional) 10 Using the Slate PC Dock (Optional) 11 Using the Bluetooth Wireless Keyboard (Optional) 14 Text

More information

F-3000 connector family SFF reliable optic connections.

F-3000 connector family SFF reliable optic connections. F-3000 connector family SFF reliable optic connections. Outline Standard product and versions The F-3000 r and rs Available technologies and standards The F-3000 fusion 2 The connector 3 Features and benefits

More information

Volition. Multimode VF-45 Socket and Patch Cords. 3Innovation

Volition. Multimode VF-45 Socket and Patch Cords. 3Innovation Volition Multimode VF-45 Socket and Patch Cords O b D s oc ol e um te en t It's as easy as plugging in a rj-45. The Volition VF-45 Socket and Patch Cords from 3M are the latest duplex, fiber optic interconnects

More information

FA-2448 SIX POSITION Filter Wheel

FA-2448 SIX POSITION Filter Wheel 15 Discovery Way, Acton, MA 01720 Phone: (978)263-3584, Fax: (978)263-5086 Web Site: www.acton-research.com Operating Instructions Acton Research Corporation FA-2448 SIX POSITION Filter Wheel Rev. 3.05.17

More information

BENCHMIKE PRO NEW. The Industry s leading off-line ID/OD/Wall measurement system

BENCHMIKE PRO NEW. The Industry s leading off-line ID/OD/Wall measurement system BENCHMIKE PRO NEW The Industry s leading off-line ID/OD/Wall measurement system Industry s leading sample inspection system Measure manufactured cut samples fast and with the highest accuracy in the industry

More information

PM-212-MPO Pocket Optical Power Meter INSTRUCTION MANUAL

PM-212-MPO Pocket Optical Power Meter INSTRUCTION MANUAL PM-212-MPO Pocket Optical Power Meter INSTRUCTION MANUAL Revision 1.2 is the registered trademark of OPTOKON, a.s. Other names and trademarks mentioned herein may be the trademarks of their respective

More information

Manual. LC-16 system. LC-16 Inkjet Printer 1

Manual. LC-16 system. LC-16 Inkjet Printer 1 Manual LC-16 system LC-16 Inkjet Printer 1 Index ENVIRONMENT. 3 OPERATOR S SAFETY 3 OPERATION SAFETY 3 PART 1 INSTALLATION AND PARAMETER SETTING 4 1) Preparing 4 2) Installation 4 3) Priming 4 4) Parameter

More information

User Manual. Optical DWDM Power Meter ODPM-48

User Manual. Optical DWDM Power Meter ODPM-48 User Manual Optical DWDM Power Meter TheFibers Inc. A-1109 Keumkang Penterium IT Tower, 282 Hagui-ro, Dongan-gu, Anyang-City, Korea Tel: +82-31-381-6108 Fax: +82-31-381-6109 Email: sales@thefibers.com

More information

AF-OLK5A Series Fiber Optic Test kit

AF-OLK5A Series Fiber Optic Test kit AF-OLK5A Series Fiber Optic Test kit Description The AF-OLK5A series test kits are the complete solution necessary for the installer to test, terminate and trouble shoot fiber optic systems. These test

More information

C o n n e c t o r s C h a p t e r 183

C o n n e c t o r s C h a p t e r 183 Connectors Chapter 5 183 Connectors Contents Copper RJ45 FAMILY 700A8 185 LC FAMILY LC Connectors 186 LazrSPEED 189 LC Couplings 190 SC FAMILY SC Connectors 191 SC Connector Clip 192 SC Couplings 193 ST

More information

MU Type Fiber Optic Connectors

MU Type Fiber Optic Connectors Jun.1.2018 Copyright 2018 HIROSE ELECTRIC CO., LTD. All Rights Reserved. MU Type Fiber Optic Connectors HMU Series Features 1. IEC, JIS standard compliant and intermateability test certified. Comply with

More information

Product Specifications

Product Specifications Product Specifications The ksa RateRat Pro is a turnkey, real-time, in-situ optical reflectance probe designed for deposition monitoring of semi-absorbent thin films. The RateRat Pro measures deposition

More information

Calibration of a portable interferometer for fiber optic connector endface measurements

Calibration of a portable interferometer for fiber optic connector endface measurements Calibration of a portable interferometer for fiber optic connector endface measurements E. Lindmark Ph.D Light Source Reference Mirror Beamsplitter Camera Calibrated parameters Interferometer Interferometer

More information

Volition. VF-45 Singlemode Socket and Patch Cords. 3Innovation IT'S AS EASY AS PLUGGING IN AN RJ-45.

Volition. VF-45 Singlemode Socket and Patch Cords. 3Innovation IT'S AS EASY AS PLUGGING IN AN RJ-45. Volition VF-45 Singlemode Socket and Patch Cords IT'S AS EASY AS PLUGGING IN AN RJ-45. The Volition VF-45 Singlemode Socket and Patch Cords from 3M are the latest duplex, fiber optic interconnect products

More information

MAP Insertion Loss/Return Loss Meter (morl-a1 with PCT Automation Software)

MAP Insertion Loss/Return Loss Meter (morl-a1 with PCT Automation Software) COMMUNICATIONS TEST & MEASUREMENT SOLUTIONS MAP Insertion Loss/Return Loss Meter (morl-a1 with PCT Automation Software) Key Features Mandrel-free optical return loss measurements on fiber-optic patch-cords

More information

VOLITION INTERCONNECT VF-45 STYLE

VOLITION INTERCONNECT VF-45 STYLE Volition Cabling System VOLITION INTERCONNECT VF-45 STYLE 62.5/125µm Fiber 3Innovation 1. Product Description & Requirements Volition Interconnect (VF-45 ) Socket and Patch Cords The Volition VF-45 Interconnect

More information

CHEMINSTRUMENTS HIGH SPEED RELEASE TESTER MODEL HSR-1000 OPERATING INSTRUCTIONS

CHEMINSTRUMENTS HIGH SPEED RELEASE TESTER MODEL HSR-1000 OPERATING INSTRUCTIONS CHEMINSTRUMENTS HIGH SPEED RELEASE TESTER MODEL HSR-1000 OPERATING INSTRUCTIONS PRODUCT DESCRIPTION...2 UNPACKING...3 ASSEMBLY...4 KEY COMPONENTS...5 OPERATION...6 SAMPLE PREPARATION...8 EZ LAB SOFTWARE...9

More information

Volition. Multimode VF-45 Socket and Patch Cords. 3Innovation

Volition. Multimode VF-45 Socket and Patch Cords. 3Innovation Volition Multimode VF-45 Socket and Patch Cords It's as easy as plugging in a rj-45. The Volition VF-45 Socket and Patch Cords from 3M are the latest duplex, fiber optic interconnects for affordable, high

More information

Paint Test Equipment. Eban 4000 Coating Thickness Meter Data Sheet

Paint Test Equipment. Eban 4000 Coating Thickness Meter Data Sheet Paint Test Equipment Eban 4000 Coating Thickness Meter Data Sheet Coating Thickness Gloss Porosity Adhesion Surface Roughness Surface Cleanliness Climatic Conditions Electrostatic Inspectors Accessories

More information

DIAMOND Fiber Optic Components

DIAMOND Fiber Optic Components DIAMOND Fiber Optic Components CABLE ASSEMBLIES AND ADAPTERS E-2000 Simplex PC/APC MULTIMODE PC Constant innovation in the field of fiber optics has created demand for products with ever increasing performance

More information

Agilent 10706B High Stability Plane Mirror Interferometer

Agilent 10706B High Stability Plane Mirror Interferometer 7D Agilent 10706B High Stability Plane Mirror Interferometer Description Description The Agilent 10706B High Stability Plane Mirror Interferometer (see Figure 7D-1) is an improved version of the Agilent

More information

MantraJet 1100 CD/DVD autoprinter Operator s manual

MantraJet 1100 CD/DVD autoprinter Operator s manual MantraJet 1100 CD/DVD autoprinter Operator s manual Rev 1.00 May 7, 2008 Table of contents 1. Specifications.. Page 2 2. Unpacking you MantraJet 1100 Page 3 3. Using your autoprinter for the first time.

More information

A Closer Look at Fiber Optic Cable Assemblies

A Closer Look at Fiber Optic Cable Assemblies A Closer Look at Fiber Optic Cable Assemblies Our thanks to Siemon for allowing us to reprint the following article Many network performance problems occurring at the physical layer are directly related

More information

Manual for MantraJet 1100 CD/DVD auto-printer

Manual for MantraJet 1100 CD/DVD auto-printer Manual for MantraJet 1100 CD/DVD auto-printer Rev 1.03 September 7, 2010 Table of contents Specifications...3 Unpacking MantraJet 1100...4 Quick installation reference MantraJet 1100...7 Installation of

More information

Polarization Microscope. Omano OM349P Series. User Guide

Polarization Microscope. Omano OM349P Series. User Guide Polarization Microscope Omano OM349P Series User Guide Table of Contents Getting Started... 3 Components... 3 Technical Specifications... 5 Setup Instructions... 6 Basic Operation... 7 Changing the Light

More information

Fiber Optic Distributed Temperature Sensors (B-DTS)

Fiber Optic Distributed Temperature Sensors (B-DTS) Fiber Optic Distributed Temperature Sensors (B-DTS) Low-cost Brillouin BOTDA scattering version For more information about our strain and temperature sensor system and related products, please visit www.ozoptics.com

More information

SPECIALIZED FIBER OPTIC ENDFACE TESTING

SPECIALIZED FIBER OPTIC ENDFACE TESTING SPECIALIZED FIBER OPTIC ENDFACE TESTING PRODUCTS // ACCESSORIES FiBO APPLICATIONS Fiber Optic Connector and Termini Testing Aerospace & military vehicles on location inspection NIST/ISO traceable testing

More information

SC Type Fiber Optic Connectors

SC Type Fiber Optic Connectors SC Type Fiber Optic Connectors HSC Series MU SC FC Features 1. JIS, IEC standards compliant and fully intermateable with NTT's SC products. JIS C 573 (F04 type fiber optic connectors) IEC 61754-4 2. Plugs

More information

PM-212-SI3 Pocket Optical Power Meter INSTRUCTION MANUAL

PM-212-SI3 Pocket Optical Power Meter INSTRUCTION MANUAL PM-212-SI3 Pocket Optical Power Meter INSTRUCTION MANUAL Revision 2.4 is the registered trademark of OPTOKON, a.s. Other names and trademarks mentioned herein may be the trademarks of their respective

More information

Product Introduction. G0306A USB 400x Video Inspection Probe

Product Introduction. G0306A USB 400x Video Inspection Probe Product Introduction G0306A USB 400x Video Inspection Probe G0306A USB 400x Video Inspection Probe Product Introduction March 2014 Anritsu Corporation Slide 1 VIP Series Support Products Mainframe MU909014x/15x

More information

WATER MAZE VIDEO TRACKING

WATER MAZE VIDEO TRACKING instrumentation and software for research SOF-845 USER S MANUAL DOC-214 Rev. 2.2 Copyright 2016 All Rights Reserved Med Associates Inc. P.O. Box 319 St. Albans, Vermont 05478 Phone: 802.527.2343 Fax: 802.527.5095

More information

SPECIALIZED FIBER OPTIC ENDFACE TESTING

SPECIALIZED FIBER OPTIC ENDFACE TESTING SPECIALIZED FIBER OPTIC ENDFACE TESTING PRODUCTS // ACCESSORIES FiBO APPLICATIONS Fiber Optic Connector and Termini Testing Aerospace & military vehicles on location inspection NIST/ISO traceable testing

More information

Table of Contents. Unpacking and Inspection Setup Loading the Media Mount the Printer on the Wall... 16

Table of Contents. Unpacking and Inspection Setup Loading the Media Mount the Printer on the Wall... 16 WPL25/WHC25 Table of Contents Unpacking and Inspection... 1 Setup... 5 Loading the Media... 6 Mount the Printer on the Wall... 16 LED and Button Functions... 17 Troubleshooting... 18 Unpacking and Inspection

More information

Standard-Area Silicon PIN Photodiodes (500µm)

Standard-Area Silicon PIN Photodiodes (500µm) Standard-Area PD-LD Inc. offers a variety of standard and custom PIN Photodiodes and APDs in fiber coupled packages. The semiconductors offered are of proven manufacture and design. Our Silicon devices

More information

Cross hairs for vertical and horizontal profiles. Profile can be linked to cross hair position, centroid, peak or sum profile can be displayed

Cross hairs for vertical and horizontal profiles. Profile can be linked to cross hair position, centroid, peak or sum profile can be displayed 3.1.2 Laser Beam Analysis Systems 3.1.2.2 BeamStar Greatest ease of use Automated operation Effortless report generation Start/stop measurement Background subtraction for improved accuracy measurements

More information

PointWrite Touch Module User Manual

PointWrite Touch Module User Manual PointWrite Touch Module User Manual Version: 1 Keep this user manual for future references. Table of contents Shipping Contents... 3 Product Overview... 4 About Touch Module...4 Dimensions...4 Preparation

More information

MarSurf. The new generation of contour measurement systems MarSurf XC 20 MarSurf XC 2

MarSurf. The new generation of contour measurement systems MarSurf XC 20 MarSurf XC 2 MarSurf The new generation of contour measurement systems MarSurf XC 20 MarSurf XC 2 The new generation of contour measurement systems Ladies and Gentlemen, There is an increasing need in industrial production

More information

3700 SERIES USER MANUAL

3700 SERIES USER MANUAL SAFETY GUIDE This manual contains the precautions necessary to ensure your personal safety as well as for protection for the products and the connected equipment. These precautions are highlighted with

More information

OTS-600 Series Optical Sources, Meters, Testers and Kits with Data Storage Capabilities

OTS-600 Series Optical Sources, Meters, Testers and Kits with Data Storage Capabilities features and benefits Large LCD screen and soft key menus Source and meter in one unit Auto wavelength switching and detection USB data ports OTS-600 Series Light Source and Power Meter Photo LAN1193 Ease

More information

MInIMAT-Ec. flexible data recording highly accurate

MInIMAT-Ec. flexible data recording highly accurate Screwdriving technology Automation Air motors Air tools Handheld Screwdriver electric MInIMAT-Ec straight handle design - torque ranges between 0.03-18 nm pistol-grip design - torque ranges between 0.06-18

More information

DIAMOND Fiber Optic Components

DIAMOND Fiber Optic Components DIAMOND Fiber Optic Components CABLE ASSEMBLIES AND ADAPTERS F-3 Simplex PC/APC MULTIMODE PC Telecommunications and Fiber Optics are developing at an incredible speed. To keep up with tomorrow s requirements

More information

OLS-55/56 Sm a r t Optical Laser Source A Sm a r t, Future-Proof Optical Laser Source

OLS-55/56 Sm a r t Optical Laser Source A Sm a r t, Future-Proof Optical Laser Source COMMUNICATIONS TEST & MEASUREMENT SOLUTIONS /56 Sm a r t Optical Laser Source A Sm a r t, Future-Proof Optical Laser Source Key features New Laser source with up to four wavelengths offers flexibility

More information

FIBER. WP-TEST-FIBER-400 Optical Power Meter Owner s Manual

FIBER. WP-TEST-FIBER-400 Optical Power Meter Owner s Manual FIBER WP-TEST-FIBER-400 Optical Power Meter Owner s Manual Technical Support: p. 866.838.5052 2017 Wirepath Ver. 170419-1000 Contents Chapter 1. Standard Configuration... 2 Chapter 2. Overview... 2 Chapter

More information

FISCHER MINIMAX TM SERIES

FISCHER MINIMAX TM SERIES TECHNICAL SPECIFICATIONS FISCHER TM SERIES Fischer Connectors SA All rights reserved Web version 1.0-03.2016 Changes without prior notice FISCHER TM SERIES KEY FEATURES The Fischer MiniMax Series increases

More information

ND 1300 QUADRA-CHEK the Digital Readouts for Convenient 2-D Measurement

ND 1300 QUADRA-CHEK the Digital Readouts for Convenient 2-D Measurement ND 1300 QUADRA-CHEK the Digital Readouts for Convenient 2-D Measurement The ND 1300 QUADRA-CHEK digital readouts can support up to four axes. They function as measuring computers with 2-D acquisition of

More information

Handheld Optical Power Meter. Model SSF-TKITP-400 OPTICAL POWER METER. Manual

Handheld Optical Power Meter. Model SSF-TKITP-400 OPTICAL POWER METER. Manual Model SSF-TKITP-400 OPTICAL POWER METER Manual August 2016 Contents Chapter 1. Standard Configuration.....2 Chapter 2. Overview...........2 Chapter 3. Data Sheet.....3 Chapter 4. Function.........4 4.1

More information

General SC F728 & FC F729

General SC F728 & FC F729 General SC F728 & FC F729 Radiall, thanks to its new range of SC and FC connectors has greatly improved these traditional fiber optic connectors in terms of technology and reliability. One-piece connector

More information

Led Illuminator and Light Stimulator. Model # LSD-1. Ver

Led Illuminator and Light Stimulator. Model # LSD-1. Ver Led Illuminator and Light Stimulator Model # LSD-1 Ver. 2.0.0 5.16 ALA Scientific Instruments Inc. 60 Marine Street, Farmingdale, NY 11735 Voice: +1 631.393.6401 fax: +1 631.393.6407 www.alascience.com

More information

High Precision Fusion Splicer

High Precision Fusion Splicer OR-300 High Precision, with high performance price ratio, is the main product of Oren Electrical technology. With smart appearance, reliable quality, The is a special design for the installation, operation

More information

MInIMAT-ec. flexible data recording highly accurate

MInIMAT-ec. flexible data recording highly accurate Screwdriving technology Automation Air motors Air tools Handheld Screwdriver electric MInIMAT-ec straight handle design - torque ranges between 0.03-18 nm pistol-grip design - torque ranges between 0.06-18

More information

SMART FORCE AND TORQUE SENSORS. User s Guide

SMART FORCE AND TORQUE SENSORS. User s Guide SMART FORCE AND TORQUE SENSORS Thank you Thank you for purchasing a Mark-10 Plug & Test TM remote sensor, designed for use with a Mark-10 model 5i or 3i force/torque indicator. With proper usage, we are

More information

VIAVI. OLA-54 and OLA-55. SmartClass. Data Sheet. Optical Level Attenuator. VIAVI SmartClass optical handhelds go beyond the basics.

VIAVI. OLA-54 and OLA-55. SmartClass. Data Sheet. Optical Level Attenuator. VIAVI SmartClass optical handhelds go beyond the basics. Data Sheet VIAVI TM SmartClass Optical Level Attenuator OLA-54 and OLA-55 VIAVI Solutions VIAVI SmartClass optical handhelds go beyond the basics With more than 100,000 optical handhelds already in use,

More information

How to Measure Wedge. Purpose. Introduction. Tools Needed

How to Measure Wedge. Purpose. Introduction. Tools Needed Purpose Optical Wedge Application (OWA) is an add-on analysis tool for measurement of optical wedges in either transmission or reflection. OWA can measure a single part or many parts simultaneously (e.g.

More information

IR Varifocal IP66 Weatherproof Cameras

IR Varifocal IP66 Weatherproof Cameras IR Varifocal IP66 Weatherproof Cameras Products: CFC6067VF, CFC6067VF2 Please read this manual before using your camera, and always follow the instructions for safety and proper use. Save this manual for

More information

Thermo Scientific Spectronic 200 Visible Spectrophotometer. The perfect tool. for routine measurements

Thermo Scientific Spectronic 200 Visible Spectrophotometer. The perfect tool. for routine measurements Thermo Scientific Spectronic 200 Visible Spectrophotometer The perfect tool for routine measurements The Standard for Routine Measurements Thermo Scientific SPECTRONIC spectrophotometers have served as

More information

INSTALLATION MANUAL. For HOLO3DGRAPH I, HOLO3DGRAPH II & HD-Aux

INSTALLATION MANUAL. For HOLO3DGRAPH I, HOLO3DGRAPH II & HD-Aux INSTALLATION MANUAL For HOLO3DGRAPH I, HOLO3DGRAPH II & HD-Aux www.immersiveinc.com Phone: 716.667.2531 Fax: 360.358.2503 22 Breezewood Drive, Orchard Park, New York 14127 Table Of Contents Introduction

More information

Product information 19" 1U UniRack2 24xLC-Duplex G.652.D, PC, ceramic, C/2 R819330

Product information 19 1U UniRack2 24xLC-Duplex G.652.D, PC, ceramic, C/2 R819330 Product information 19" 1U UniRack2 24xLC-Duplex G.652.D, PC, ceramic, C/2 R819330 1 Pcs. 19" 1U slide-in unit, splice tray plate with 2 splice trays radius 40 mm, fastening material, labeling strips and

More information

Quick Scope. CNC/Manually-operated Vision Measuring Microscope

Quick Scope. CNC/Manually-operated Vision Measuring Microscope Bulletin No. 1579 Quick Scope CNC/Manually-operated Vision Measuring Microscope Quick Scope Non-contact Vision Measuring Microscopes Offer High Accuracy, Excellent Affordability and Powerful Capabilities!

More information

Max. db Channel Uniformity

Max. db Channel Uniformity General Information Applications: - TTX Deployments - GPON Networks - CATV Links - Optical Signal Distribution euroclust offers splitters based on planar lightwave circuit (PLC) technology and fusion based

More information

STEP-BY-STEP INSTRUCTIONS FOR BUILDING A FLUORESCENCE MICROSCOPE. TECHSPEC Optical Cage System

STEP-BY-STEP INSTRUCTIONS FOR BUILDING A FLUORESCENCE MICROSCOPE. TECHSPEC Optical Cage System STEP-BY-STEP INSTRUCTIONS FOR BUILDING A FLUORESCENCE MICROSCOPE TECHSPEC Optical Cage System INTRODUCTION 2 What is a Digital Fluorescence Microscope? Unlike traditional microscopes, which utilize an

More information

User Manual. Trinocular Stereo Microscope

User Manual. Trinocular Stereo Microscope User Manual Trinocular Stereo Microscope Model V436 XV436 Series MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 1 1. Component Illustration... 2 2. Installation 3 3. Operation....4

More information