MODEL ATV 4/6 BATTERY POWERED DIGITAL VACUUM INSTRUMENT

Size: px
Start display at page:

Download "MODEL ATV 4/6 BATTERY POWERED DIGITAL VACUUM INSTRUMENT"

Transcription

1 MODEL ATV 4/6 BATTERY POWERED DIGITAL VACUUM INSTRUMENT Contains Operating and Programming Information

2

3 Model ATV 4/6 Battery Powered Digital Vacuum Instrument ã 1996, Teledyne Electronic Technologies, Hastings Instruments Division All rights reserved. Hampton, Virginia, U.S.A. First Printing March 1996 Document Number: Rev. B

4 Manual Print History The print history shown below lists the printing dates of all revisions and addenda created for this manual. The revision level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered sequentially. When a new revision is created, all Addenda associated with the previous Revision of the manual are incorporated into the new Revision of the manual. Each new Revision includes a revised copy of this print history page. Revision A (Document Number )... March 1996 Revision B (Document Number )... Feburary 1999 Warranty Hastings Instruments warrants this product, for a period of one year from the date of shipment, to be free from defects in material and workmanship. This warranty does not apply to defects or failures resulting from unauthorized modification, misuse or mishandling of the product. This warranty does not apply to batteries or other expendable parts, nor to damage caused by leaking batteries or any similar occurrence. This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty as to fitness for a particular use. Hastings Instruments shall not be liable for any indirect or consequential damages. Hastings Instruments will, at its option, repair, replace, or refund the selling price of the product if Hastings Instruments determines in good faith, that it is defective in materials or workmanship during the warranty period, and provided the item is returned to Hastings Instruments together with a written statement of the problem and a repair authorization number.

5 Table of Contents 1 General Information Introduction Features Vacuum Gauge Tube Accessories Specifications Receiving Inspection Power Supply Installation Battery Installation Optional Power Supply Installation Optional Gauge Installation Instrument Configuration Factory Calibration Values User Calibration Values Connection and Operation Instrument Connection to a Gauge Tube Basic Instrument Operation and Pressure Measurement Pressure Measurements using the DV-4 Gauge Tube Pressure Measurements using the DV-6 Gauge Tube Maintenance and Troubleshooting Instrument Maintenance Troubleshooting Guide Warranty Repair Non-Warranty Repair

6

7 1 General Information

8 1-2 General Information Introduction The Model ATV 4/6 is a compact, hand-held, battery powered, digital, LCD display vacuum instrument. It is designed to operate using a single 9 Volt battery (included with the unit), and either a DV-4 or DV-6 Hastings gauge tube dependant upon the range of measurement desired. The instrument comes packaged in a lightweight carrying case which can be used for the storage and protection of the instrument. Features Operates using an internal battery and/or an external AC/DC power adapter Direct reading LCD digital display. A single switch selects user parameters Automatic shut off circuitry to prolong battery life Custom two-point calibration using external instrumentation Selectable units of measure, mbar or Torr. Selectable tube, DV-4 or DV-6 Low battery voltage indicator Lightweight protective carrying case Vacuum Gauge Tube The ATV 4/6 is capable of performing pressure measurements from to 20 Torr or from to 20 mbar using a Hastings DV-4 and DV-6 vacuum gauge tube. DV-4 pressure range: 0.10 to 20 Torr or 0.13 to 20 mbar. DV-6 pressure range: to 1.0 Torr or to 1.3 mbar. Note: 1 Torr is equal to 1 mmhg or 1000 microns Hg. Hastings vacuum gauge tubes, DV-4 and DV-6, are not included with the purchase of an ATV 4/6 instrument. Accessories External Power Supply Any power supply capable of delivering VDC at 50mA with a 3.5mm diameter positive (+) polarity tip-end can be used with the ATV 4/6 instrument.

9 General Information 1-3 Specifications Pressure range to 20 Torr or 0.13 to 20 mbar (with DV-4 gauge tube) to 1 Torr or to 1.3 mbar (with DV-6 gauge tube) Operating temperature... 0 to 40 0 C ambient Storage temperature to 50 0 C ambient Current drain madc (ATV 4/6 with tube) Internal power supply...battery, 9V Internal battery usable life hours continuous (9V alkaline) Dimensions... Hand held display and control unit 4.72 x 2.36 x 1 (120 x 60 x 25 mm) Gauge tube interface module 3.00 x 1.30 dia (75mm x 33 mm dia) Weight (approx)...7 oz. (195 gm) Receiving Inspection Carefully unpack the Hastings Battery Operated Vacuum Instrument (p/n ATV 4/6). Inspect all items received at this time for any obvious damage that may have occurred during shipment. Immediately advise the shipping agent if there is any damage to any items in your shipment. Identify each item and the quantity received with those listed on the packing list. Accessories will be listed separately on the packing list.

10

11 2 Power Supply Installation

12 2-2 Power Supply Installation This section is designed to assist in placing the battery operated ATV 4/6 vacuum instrument into operation as quickly and easily as possible. Please read the following instructions thoroughly before installing and placing the instrument into service. Battery Installation Open the carrying case and remove the ATV instrument and battery. Place the instrument, display side down on a suitable flat surface. Depress the battery cover release latch, and slide the cover off of the bottom of the unit. Connect the supplied 9 Volt battery to the unit, insert it into the back of the instrument and reinstall the battery cover. (9 Volt Lithium or Nickel- Cadmium batteries may be substituted for the standard alkaline battery) Press and release the mode button. The following Display Sequence will indicate that the battery has been properly installed. -DISPLAY SEQUENCE- d--4 or d--6 down arrow plus d--4 or d--6 down arrow plus Err Instrument will power down and the display will go blank after approximately 30 seconds. Optional Power Supply Installation Hastings Instruments offers an AC/DC power supply that is suitable for use with the ATV instrument. Use of this power supply or any commercially available unit with the following specifications is recommended if making long-term continuous measurements, and the operating time of the instrument will exceed 8 to 10 hours. Class 2 transformer with output cable termination of 3.5mm with a positive (+) polarity tip-end. Input power : Nominal 120V 4 watts Output power: Rated 12VDC 50 ma Before connecting the external power supply make sure that the unit is off (display should be blank). Connect the AC/DC adapter male power plug into the female receptacle on the side of the unit and plug the adapter into a V 60Hz power source.

13 Power Supply Installation 2-3 Optional Gauge Tube Installation If you purchased a DV-4 or DV-6 gauge tube for use with your ATV instrument the following is information on the installation of these tubes. WARNING: Compression seal fittings Quick Connects are not for use in systems where the tube can be pressurized above atmosphere. All Hastings gauge tubes are shipped with a protective cap at the evacuation port to keep the thermopile elements clean and free from physical damage. Remove the protective cap prior to installation. The tube can be mounted in any orientation without affecting the calibration or performance. The recommended position is with the tube vertical and the stem down. If using Quick Connect compression fittings, installation of an isolation valve between the compression fitting and the vacuum system is recommended.

14

15 3 Instrument Configuration

16 3-2 Instrument Configuration This information is designed to prepare the instrument for a specific setup that will result in proper operation in the selected modes d-4 or d-6 when using either DV-4 or DV-6 gauge tubes. Factory Calibration Values (factory values are for use with any DV-4 or DV-6 gauge tube) Select Unit of Measure, DV-4 Tube, and Factory Calibration Values NOTE: Do Not connect ATV to gauge tube at this time. Press and hold Mode button, either mbar or Torr will be displayed. Release mode button. If desired unit of measure is not displayed, press and release mode button until display is correct. Either d--4 for DV-4 or d--6 for DV-6 will be displayed. If d--4 is not displayed, press and release mode button until d--4 is obtained. Cal-H is displayed. Press and release mode button. P 20 is displayed. Press and release mode button. Cal.L is displayed. Press and release mode button. Down arrow is displayed. Press and release mode button. Factory values for DV-4 have been entered. Down arrow and Err is displayed Following a delay period, the ATV will power down and the display blanks. Select Unit of Measure, DV-6 Tube, and Factory Calibration Values NOTE: Do Not connect ATV to gauge tube at this time. Press and hold Mode button, either mbar or Torr will be displayed. Release mode button. If desired unit of measure is not displayed, press and release mode button until display is correct. Either d--4 for DV-4 or d--6 for DV-6 will be displayed. If d--6 is not displayed, press and release mode button until d--6 is obtained. Cal-H is displayed. Press and release mode button. P 1 is displayed. Press and release mode button. Cal.L is displayed. Press and release mode button. Down arrow is displayed. Press and release mode button. Factory values for DV-6 have been entered. Down arrow and Err is displayed Following a delay period, the ATV will power down and the display blanks.

17 Instrument Configuration 3-3 User Calibration Values (user values are for use with only one DV-4 or DV-6 gauge tube) Select Unit of Measure, DV-4 Tube, and enter User Calibration Values. Install DV-4 gauge tube and pressure standard in vacuum system but Do Not connect ATV to gauge tube at this time. Press and hold Mode button, either mbar or Torr will be displayed. Release the mode button. If desired unit of measure is not displayed, press and release mode button until display is correct. Either d--4 for DV-4 or d--6 for DV-6 will be displayed. If d--4 is not displayed, press and release mode button until d--4 is obtained. Cal.H is displayed. Press and release mode button. P 20 is displayed. Connect ATV to DV-4 gauge tube and evacuate to a pressure less than 2.0E + 01 Torr (<20.00 Torr). Raise pressure by backfilling system using nitrogen gas or high purity dry air until pressure standard in use indicates a stable value at 2.0E + 01 Torr (20.00 Torr). Press and release mode button to enter user calibration value for Cal.H into ATV memory. Cal.L is displayed. Press and release mode button, 0.10 is displayed. Evacuate DV-4 gauge tube to a pressure less than 1.0E-01 Torr (<0.1 Torr). Raise pressure by backfilling system using nitrogen gas or high purity dry air until pressure standard in use indicates a stable value at 1.0E- 01 Torr (0.100 Torr). Press and release mode button to enter user value for Cal.L into ATV memory. Following a delay period, the ATV will power down and the display blanks. Note: Making changes while in the calibration mode, other than selecting unit of measure, may result in the lose of user calibration values. Select Unit of Measure, DV-6 Tube, and enter User Calibration Values. Install DV-6 gauge tube and pressure standard in vacuum system but Do Not connect ATV to gauge tube at this time. Press and hold Mode button, either mbar or Torr will be displayed. Release the mode button. If desired unit of measure is not displayed, press and release mode button until display is correct. Either d--4 for DV-4 or d--6 for DV-6 will be displayed. If d--6 is not displayed, press and release mode button until d--6 is obtained. Cal.H is displayed. Press and release mode button. P 1 is displayed. Connect ATV to DV-6 gauge tube and evacuate to a pressure less than 1.0E + 00 Torr (<1.00 Torr). Raise pressure by backfilling system using nitrogen gas or high purity dry air until pressure standard in use indicates a stable value at 1.0E + 00 Torr (1.00 Torr). Press and release mode button to enter user calibration value for Cal.H into ATV memory. Cal.L is displayed. Press and release mode button, is displayed. Evacuate DV-6 gauge tube to a pressure less than 1.0E-03 Torr (<0.001 Torr). Raise pressure by backfilling system using nitrogen gas or high purity dry air until pressure standard in use indicates a stable value at 1.0E-03 Torr ( Torr). Press and release mode button to enter user value for Cal.L into ATV memory. Following a delay period, the ATV will power down and the display blanks. Note: Making changes while in the calibration mode, other than selecting unit of measure, may result in the lose of user calibration values.

18

19 4 Connection and Operation

20 4-2 Connection and Operation Instrument Connection to a Gauge Tube CAUTION: When disconnecting the ATV from a gauge tube, pull the interface module straight away from the tube until the two items are completely separated. Failure to do this can result in the breaking of the center post or key guide of the gauge tube. Connect the interface module of the ATV to the connector base of either a DV-4 or DV-6 gauge tube. Do not install any extension or adapter cables between the module and the gauge tube.

21 Connection and Operation 4-3 Basic Instrument Operation and Pressure Measurement There are only two basic modes of operation, Auto-Off (about 40 seconds of active display) and Continuous On (display is always active). Instrument operation is initiated by momentarily pressing the mode button. Upon doing this the instrument is now powered on and defaults to the Auto-Off mode. To change the mode of operation to a continuous mode press and hold the mode button until the up arrow symbol appears in the display. At this point release the button. Select the mode of operation by momentarily pressing the mode button. When doing this the display will switch from displaying the up arrow and the down arrow each time the button is pressed. By leaving the down arrow symbol on the display, the instrument will remain in the Auto-Off mode. Leaving the up arrow will place the instrument in the Continuous On mode. Pressure Measurements using the DV-4 Gauge Tube Connect the ATV s interface module to the DV-4 gauge tube. Momentarily press the mode button and verify that d--4 and Torr or mbar are displayed and are correct for the instruments intended use. If the displayed values are incorrect, refer back to Section 3. Press and release the mode button of the instrument, d--4 will appear on the display momentarily. The pressure reading will now be displayed for about 40 seconds if the instrument is in the auto-off mode (down arrow symbol). If continuous operation is desired, press and hold the mode button until the up arrow symbol appears on the display, then release the mode button. After completing measurements using the continuous on mode, press and hold the mode button until the down arrow symbol appears in the display. Release the mode button. The instrument will continue to display the system pressure for approximately 40 seconds before the unit powers down and the display goes blank. If the system pressure is lower than the instruments range, ur will appear on the display indicating an underrange condition. When system pressure is greater than the instruments measuring capabilities, or will be displayed indicating an overrange condition. Pressure Measurements using the DV-6 Gauge Tube Connect the ATV s interface module to the DV-6 gauge tube. Momentarily press the mode button and verify that d--6 and Torr or mbar are displayed and are correct for the instruments intended use. If the displayed values are incorrect, refer back to Section 3. Press and release the mode button of the instrument, d--6 will appear on the display momentarily. The pressure reading will now be displayed for about 40 seconds if the instrument is in the auto-off mode (down arrow symbol). If continuous operation is desired, press and hold the mode button until the up arrow symbol appears on the display, then release the mode button. After completing measurements using the continuous on mode, press and hold the mode button until the down arrow symbol appears in the display. Release the mode button. The instrument will continue to display the system pressure for approximately 40 seconds before the unit powers down and the display goes blank. If the system pressure is lower than the instruments range, ur will appear on the display indicating an underrange condition. When system pressure is greater than the instruments measuring capabilities, or will be displayed indicating an overrange condition.

22

23 5 Maintenance and Trouble shooting

24 5-2 Maintenance and Troubleshooting Instrument Maintenance The only authorized maintenance at the user level for the ATV 4/6 is the replacement of the 9V battery. Consult the troubleshooting guide for additional information. Reference Tube Calibration The low end of the vacuum range can be set with a Hastings reference tube, such as the DB-16D, which is used with the d 4 range, or the DB-20 which is used with the d 6 range. The DB-16D must have a marked calibration value between 0.10 Torr and 0.20 Torr. The DB-20 must have a marked calibration value between Torr and Torr. Connect the ATV coupler to the Reference Tube. Press and hold down the MODE switch until the unit of measure appears in the upper right hand corner of the display. Release the mode switch. If the unit is mbar, press and release the MODE switch to get Torr. In about 5 seconds the display will indicate the tube type, either d-4 or d-6. If the Reference Tube is a DB-16D the tube type should be d 4, if the Reference Tube is a DB-20, the tube type should be d 6. If it is correct, do nothing. If you wish to change it, press and release the MODE switch. In about 5 seconds the display will indicate CAL.H. Do not press the MODE switch again until the display indicates CAL.L. CAUTION: Pressing the MODE switch while in the CAL.H mode will cause the ATV to accept the Reference Tube output as the high calibration value rather than the low calibration value. In about 5 seconds, the display will indicate CAL.L. Press and release the MODE switch. The display will indicate 0.10 Torr if d 4 was selected or Torr if d 6 was selected. Press and hold down the MODE switch until the indicated pressure matches that marked on the Reference Tube. NOTE: Some DB-16D Reference Tubes may be marked in mmhg and some DB-20 Reference Tubes may be marked in microns Hg. To convert: 1 mmh = 1 Torr 1 micron Hg = Torr This completes the Reference Tube calibration. In about 25 seconds the ATV will turn off. Note: Making changes while in the calibration mode, other than selecting unit of measure, may result in the lose of user calibration values. Troubleshooting Guide Symptom: Display reads Err instead of the system pressure. Cause: Gauge tube disconnected, or defective gauge tube Action: Check connection between the instruments interface module and the tube. If problem still persists connect instrument to a new gauge tube. If unit is still not functioning properly, consult factory. Symptom: Display reads batt. Cause: Battery voltage is too low. Action: Replace the 9V battery.

25 Connection and Operation 5-3 Symptom: Abnormal pressure readings. Cause: Incorrect gauge tube type connected to instrument. Action: Verify that the tube is of the correct type that the instrument is setup and calibrated for. If the problem still persists, consult the factory. Symptom: Blank display, mode switch is inoperative. Cause: Battery voltage is too low. Action: Replace the 9 V battery. If problem still persists, consult factory. Symptom: Inoperative gauge, power input is normal. Cause: Defective instrument. Action: Return to factory for repair or replacement. Warranty Repair Hastings Instruments warrants this product, for a period of one year from the date of shipment, to be free from defects in material and workmanship. This warranty does not apply to defects or failures resulting from unauthorized modification, misuse or mishandling of the product. This warranty does not apply to batteries or other expendable parts, nor to damage caused by leaking batteries or any similar occurrence. This warranty does not apply to any instrument which has had a tamper seal removed or broken. This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty as to fitness for a particular use. Hastings Instruments shall not be liable for any indirect or consequential damages. Hastings Instruments will, at its option, repair, replace, or refund the selling price of the product if Hastings Instruments determines in good faith, that it is defective in materials or workmanship during the warranty period. Defective instruments should be returned to Hastings Instruments together with a written statement of the problem and a Return Material Authorization (RMA) number. Please consult the factory for your RMA number before returning any product for repair. Non-Warranty Repair Any product returned for a non-warranty repair must be accompanied by a purchase order, RMA form and a written description of the problem with the instrument. If the repair cost is higher, you will be contacted for authorization before we proceed with any repairs. If you then choose not to have the product repaired, a minimum will be charged to cover the processing and inspection. Please consult the factory for your RMA number before returning any product for repair. TELEDYNE ELECTRONIC TECHNOLOGIES Hastings Instruments 804 Newcombe Avenue Hampton, Virginia U.S.A. ATTN: Repair Department Telephone FAX

TELEDYNE HASTINGS INSTRUMENTS

TELEDYNE HASTINGS INSTRUMENTS TELEDYNE HASTINGS INSTRUMENTS INSTRUCTION MANUAL 760 PLUS VACUUM GAUGE ISO 9001 C E R T I F I E D Manual Print History The print history shown below lists the printing dates of all revisions and addenda

More information

TELEDYNE HASTINGS INSTRUMENTS

TELEDYNE HASTINGS INSTRUMENTS TELEDYNE HASTINGS INSTRUMENTS INSTRUCTION MANUAL MASS FLOWMETER CONTROLLER READOUT POWERPOD-100 Manual Print History The print history shown below lists the printing dates of all revisions and addenda

More information

HASTINGS INSTRUCTION MANUAL HASTINGS POWER SUPPLY MODEL A-PAGE1

HASTINGS INSTRUCTION MANUAL HASTINGS POWER SUPPLY MODEL A-PAGE1 HASTINGS HASTINGS POWER SUPPLY MODEL 400 INSTRUCTION MANUAL 113.A-PAGE1 113.A-PAGE2 113.A-PAGE3 113.A-PAGE4 1.0 GENERAL INFORMATION The Hastings Model 400 Power Supply is a combination power supply and

More information

Vacuum Meter Instruction Manual

Vacuum Meter Instruction Manual Vacuum Meter 840064 Instruction Manual TABLE OF CONTENTS 1. INTRODUCTION...2 2. METER DESCRIPTION...3 3. OPERATING INSTRUCTIONS 3-A PRECAUTIONS - READ BEFORE FIRST USE...4 3-B MEASUREMENT PROCEDURE...4

More information

Model 8702 DP-CALC Micromanometer

Model 8702 DP-CALC Micromanometer Ventilation Testing/Balancing Model 8702 DP-CALC Micromanometer Operation and Service Manual 1980260, Revision F July 2006 Model 8702 DP-CALC Micromanometer Operation and Service Manual 1980260, Revision

More information

Manometers 2PSI, 5PSI, 15PSI, 30PSI, 100PSI , , ,

Manometers 2PSI, 5PSI, 15PSI, 30PSI, 100PSI , , , Manometers 2PSI, 5PSI, 15PSI, 30PSI, 100PSI 840080, 840081, 840082 840083, 840085 Manometers 840080, 840081, 840082, 840083, 840085 Copyright 2015 by Sper Scientific ALL RIGHTS RESERVED Printed in the

More information

Wide Range Pressure Meter Instruction Manual

Wide Range Pressure Meter Instruction Manual Wide Pressure Meter 840065 Instruction Manual Wide Pressure Meter - 840065 Copyright 2015 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or

More information

Exposure Monitoring. Model 8560 INSPECTAIR CO 2. Meter. Operation and Service Manual , Revision B June 2006

Exposure Monitoring. Model 8560 INSPECTAIR CO 2. Meter. Operation and Service Manual , Revision B June 2006 Exposure Monitoring Model 8560 INSPECTAIR CO 2 Meter Operation and Service Manual 1980265, Revision B June 2006 Model 8560 INSPECTAIR CO 2 Meter Operation and Service Manual June 2006 P/N 1980265 Rev.

More information

OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide

OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide OLS Series Light Sources, OPM Series Optical Power Meters, and Related Test Kits User s Guide 2004-2009, AFL Telecommunications, all rights reserved. COM4-00-1001 Revision E, 2009-06-16 Specifications

More information

GARNET. SEELEVEL II TM Tank Monitor INSTRUMENTS LTD. Model 709 Owner s Manual IMPORTANT OPERATOR INFORMATION DATE INSTALLED: SERIAL NUMBER:

GARNET. SEELEVEL II TM Tank Monitor INSTRUMENTS LTD. Model 709 Owner s Manual IMPORTANT OPERATOR INFORMATION DATE INSTALLED: SERIAL NUMBER: GARNET INSTRUMENTS LTD. SEELEVEL II TM Tank Monitor Model 709 Owner s Manual IMPORTANT OPERATOR INFORMATION DATE INSTALLED: SERIAL NUMBER: Signal Level Sender Height Black Water Tank Grey Water Tank Fresh

More information

Model M140 Mini Solar Power Meter

Model M140 Mini Solar Power Meter Model M140 Mini Solar Power Meter 2 Table of Contents 1 INTRODUCTION... 3 1.1 The M140 Features:... 3 1.2 The M140 Applications:... 3 2 SAFETY SUMMARY... 4 3 COMPLIANCE STATEMENTS... 5 4 PRODUCT CONTENTS

More information

SoniChek MC. 99 Washington Street Melrose, MA Phone Toll Free TRIPLETT. Mini Sound Level Meter C Weighted

SoniChek MC. 99 Washington Street Melrose, MA Phone Toll Free TRIPLETT. Mini Sound Level Meter C Weighted 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 TRIPLETT SoniChek MC Mini Sound Level Meter C Weighted Instruction Manual 84-888 4-10 Visit us at www. I. Safety information

More information

User Guide. Heavy Duty Differential Pressure Manometer. Model HD750

User Guide. Heavy Duty Differential Pressure Manometer. Model HD750 User Guide Heavy Duty Differential Pressure Manometer Model HD750 Warranty EXTECH INSTRUMENTS CORPORATION (A FLIR COMPANY) warrants this instrument to be free of defects in parts and workmanship for three

More information

V-9939B MICROPHONE ADAPTER

V-9939B MICROPHONE ADAPTER Issue 10 INTRODUCTION These instructions provide identification, installation, connection, operation and maintenance information for the Microphone Adapter. The is a Microphone Adapter designed to be used

More information

4-20mA Display. Installation and Operation Manual. Rev 4/3/2019 Part #

4-20mA Display. Installation and Operation Manual. Rev 4/3/2019 Part # 4-20mA Display Installation and Operation Manual Rev 4/3/2019 Part # 12050769 Table of Contents DOCUMENTATION CONVENTIONS... 2 Section 1: System Description... 3 Function and Theory... 3 Section 2: System

More information

User Guide. ExStik FL700. Fluoride Meter

User Guide. ExStik FL700. Fluoride Meter User Guide ExStik FL700 Fluoride Meter Introduction The model FL700 is a system specifically designed for the quick and accurate measurement of fluoride ions in drinking water and other aqueous samples.

More information

Platinum RTD Thermometer MODEL NO

Platinum RTD Thermometer MODEL NO Platinum RTD Thermometer MODEL NO. 93410-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com

More information

VELOCICALC Air Velocity Meter

VELOCICALC Air Velocity Meter ENERGY AND COMFORT Ventilation Testing VELOCICALC Air Velocity Meter Model 9525 Operation and Service Manual Copyright TSI Incorporated / May 2007 / All rights reserved. Address TSI Incorporated / 500

More information

Secured Series: Hub Plus Kit Single Door Controller Package Installation Manual

Secured Series: Hub Plus Kit Single Door Controller Package Installation Manual Secured Series: Hub Plus Kit Single Door Controller Package Installation Manual This package is designed to simplify the connections to our Secured Series Hub Plus Controller. This will translate into

More information

CR87B TEMPERATURE RECORDER

CR87B TEMPERATURE RECORDER CR87B TEMPERATURE RECORDER Sealed Unit Parts Co., inc. PO Box 21, 2230 Landmark Place, Allenwood, NJ 08720 USA Phone: 732-223-6644 Fax: 732-223-1617 www.supco.com info@supco.com 19204 INTRODUCTION The

More information

CONTENTS Safety Information Description Features Specifications Nomenclature And Function Operating Preparation Calibration Procedure

CONTENTS Safety Information Description Features Specifications Nomenclature And Function Operating Preparation Calibration Procedure CONTENTS Title Page 1. Safety Information.. 1 Environment Conditions.1 Maintenance & Clearing.1 Safety symbols 1 2. Description.. 2 3. Features.....2 4. Specifications...... 2 Output sound pressure levels...2

More information

WalkLAB ph mv/orptemperature

WalkLAB ph mv/orptemperature 16,1,16,1,2,15,2,15,14,3,14,3,4,13,4,13,12,5,12,5,6,11,6,11,10,7,10,7,8,9,8,9 WalkLAB ph mv/orptemperature meter HP9000 Operations Manual TABLE OF CONTENTS PAGE 1. INTRODUCTION 1 1.1 Features 1 1.2 Technical

More information

IAQ-CALC INDOOR AIR QUALITY METER MODEL 7525

IAQ-CALC INDOOR AIR QUALITY METER MODEL 7525 IAQ-CALC INDOOR AIR QUALITY METER MODEL 7525 OPERATION AND SERVICE MANUAL P/N 1980572, REVISION E FEBRUARY 2016 Copyright TSI Incorporated / 2007 2016 / All rights reserved. Address TSI Incorporated /

More information

Model SRV945-BT 24 X 36 Platform Scale App-Enabled Operating and Service Manual Part No.: MAN945-BT_ Page 1 of 16.

Model SRV945-BT 24 X 36 Platform Scale App-Enabled Operating and Service Manual Part No.: MAN945-BT_ Page 1 of 16. Part No.: MAN945-BT_180417 Page 1 of 16 S by S Instruments, Inc. Model SRV945-BT Portable Platform Scale App Enabled Wireless Operating and Service Manual Part No.: MAN945-BT_180417 Page 2 of 16 TABLE

More information

CSS1 Contractor Series Light Sources and CSM1 Contractor Series Optical Power Meters User s Guide

CSS1 Contractor Series Light Sources and CSM1 Contractor Series Optical Power Meters User s Guide CSS1 Contractor Series Light Sources and CSM1 Contractor Series Optical Power Meters User s Guide www.aflglobal.com or (800) 321-5298, (603) 528-7780 Limited Warranty All NOYES test equipment products

More information

SERVICE MANUAL MODEL SSP-363-E (FORMERLY SSP-365-E)

SERVICE MANUAL MODEL SSP-363-E (FORMERLY SSP-365-E) SSP-363-E-(SSP-365-E)-ADT1.03-ISSUE4.0 SERVICE MANUAL FOR MODEL SSP-363-E (FORMERLY SSP-365-E) STAINLESS STEEL PANEL TELEPHONE WITH 12 BUTTON AUTOMATIC DIALER EQUIPPED WITH ADT1.03 FIRMWARE Serving the

More information

Model OI-6900-LEL Sensor Assembly

Model OI-6900-LEL Sensor Assembly Model OI-6900-LEL Sensor Assembly Operation Manual Revision 2.4w Product Overview The Otis Instruments, Inc. GenII WireFree OI-6900-LEL ambient air gas sensor assembly is a WireFree gas sensor assembly

More information

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389 User s Guide 600A True RMS AC/DC Clamp Meter Model 38389 Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further

More information

Global Water Instrumentation, Inc.

Global Water Instrumentation, Inc. Instrumentation, Inc. 11390 Amalgam Way Gold River, CA 95670 T: 800-876-1172 Int l: (916) 638-3429, F: (916) 638-3270 Display: EZ100 11/12/04-1 - Congratulations on your purchase of the EZ100 Display.

More information

INSTRUCTION MANUAL. Digivac 760 Series. Models 100Led P / P / 29.9" 100Led 760 / SixVac 801 / 760 / 2c 801 Micro

INSTRUCTION MANUAL. Digivac 760 Series. Models 100Led P / P / 29.9 100Led 760 / SixVac 801 / 760 / 2c 801 Micro INSTRUCTION MANUAL Digivac 760 Series Models 100Led 760 100P / 760 100P / 29.9" 100Led 760 / SixVac 801 / 760 / 2c 801 Micro Ranges 0 to 760 Torr 0 to 19.99 Torr 0 to 29.9 inches of mercury The Digivac

More information

User's Guide. Heavy Duty Differential Pressure Manometer. Model

User's Guide. Heavy Duty Differential Pressure Manometer. Model User's Guide Heavy Duty Differential Pressure Manometer Model 407910 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date

More information

DIGITAL SOUND LEVEL METER. Instruction Manual

DIGITAL SOUND LEVEL METER. Instruction Manual DIGITAL SOUND LEVEL METER 840029 Instruction Manual CONTENTS I. Introduction... 3 II. Panel Description... 3 III. Operating Instructions A. Measuring Procedure... 4 B. Settings for General Applications...

More information

Dual Channel Active Direct Box. Artcessories. User's Manual

Dual Channel Active Direct Box. Artcessories. User's Manual Dual Channel Active Direct Box Artcessories User's Manual IMPORTANT SAFETY INSTRUCTION READ FIRST This symbol, whenever it appears, alerts you to the presence of uninsulated dangerous voltage inside enclosure-voltage

More information

350 SMART MANOMETER OPERATING INSTRUCTIONS

350 SMART MANOMETER OPERATING INSTRUCTIONS 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com 350 SMART MANOMETER OPERATING INSTRUCTIONS Meriam Instrument s 350 Smart Manometer

More information

ARA FTS Flow Calibrator. Operation Manual August 1, 2016

ARA FTS Flow Calibrator. Operation Manual August 1, 2016 ARA FTS Flow Calibrator Operation Manual August 1, 2016 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 1 2. GETTING STARTED 1 2.1. Navigation 1 2.2. Charge Battery 1 2.3. Set Date and Time 2 2.4. Plug-In

More information

DP-CALC MICROMANOMETER MODEL 5825

DP-CALC MICROMANOMETER MODEL 5825 DP-CALC MICROMANOMETER MODEL 5825 OPERATION AND SERVICE MANUAL P/N 1980568, REVISION F FEBRUARY 2016 Copyright TSI Incorporated / 2007 2016 / All rights reserved. Address TSI Incorporated / 500 Cardigan

More information

4" AMPLIFIED RECESSED CEILING SPEAKERS 8" AMPLIFIED RECESSED CEILING SPEAKERS

4 AMPLIFIED RECESSED CEILING SPEAKERS 8 AMPLIFIED RECESSED CEILING SPEAKERS VSP-V-1011/V-1012 Issue 5 4" AMPLIFIED RECESSED CEILING SPEAKERS 8" AMPLIFIED RECESSED CEILING SPEAKERS INTRODUCTION These instructions contain the specifications and guidelines necessary to install, operate

More information

Series F75 Fail-Safe Module Installation, Operation and Maintenance Instructions

Series F75 Fail-Safe Module Installation, Operation and Maintenance Instructions WCAIM2029 (Part 16022) Series F75 Fail-Safe Module Installation, Operation and Maintenance Instructions CAUTION: Flowserve recommends that all products that must be stored prior to installation be stored

More information

CLEVELAND CONTROLS MODEL 6650 DIGITAL MANOMETER INSTRUCTION MANUAL DM

CLEVELAND CONTROLS MODEL 6650 DIGITAL MANOMETER INSTRUCTION MANUAL DM CLEVELAND CONTROLS MODEL 6650 DIGITAL MANOMETER INSTRUCTION MANUAL DM-6650.02 1 Instruction Manual DM-6650.02 TABLE OF CONTENTS I.0 INTRODUCTION 3 1.1 General Description... 3 1.2 Specifications... 3 1.3

More information

Platinum RTD Thermometer MODEL NO

Platinum RTD Thermometer MODEL NO Platinum RTD Thermometer MODEL NO. 93410-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com

More information

Fiber Optic Prism Switch USER S MANUAL

Fiber Optic Prism Switch USER S MANUAL Fiber Optic Prism Switch USER S MANUAL Warranty Newport Corporation warrants this product to be free from defects in material and workmanship for a period of one year from the date of shipment. If found

More information

VIS Video Inspection Scope

VIS Video Inspection Scope User s Guide VIS 300 - Video Inspection Scope VIS1-00-2000, Rev A, 6-09-10 Specifications subject to change without notice Contents Section 1 - General Overview A. Contact ODM for customer service B. Unpacking

More information

3000 Series Instructions Read Instructions Carefully!

3000 Series Instructions Read Instructions Carefully! Series Instructions Read Instructions Carefully! With buttons Without buttons CURTIS INSTRUMENTS, INC. Kisco Avenue, Mt. Kisco, NY 9 Tel. (9) 9 www.curtisinst.com. Technical Specifications. Electrical

More information

V-9941A ONE ZONE HANDSFREE TALKBACK CONTROL UNIT

V-9941A ONE ZONE HANDSFREE TALKBACK CONTROL UNIT VSP- Issue 4 ONE ZONE HANDSFREE TALKBACK CONTROL UNIT INTRODUCTION The is a One Zone Handsfree Page Unit for use on an Electronic Key system line key position, a PABX loop start trunk port, a dedicated

More information

STATIC METERS Models 211 & 212. SURVEYORSTAT Models 211XL & 212XL

STATIC METERS Models 211 & 212. SURVEYORSTAT Models 211XL & 212XL STATIC METERS Models 211 & 212 SURVEYORSTAT Models 211XL & 212XL Operating Manual 5/09 1.0 GENERAL DESCRIPTION The Model 211 and 212 Static Meters, shown in Figures 1.0-1a and b, are accurate, compact

More information

LBO-H2 Series DIRECT PLUGGABLE LINKBRIDE TM FIBER OPTIC HDMI 2.0 TRANSMISSION SYSTEM

LBO-H2 Series DIRECT PLUGGABLE LINKBRIDE TM FIBER OPTIC HDMI 2.0 TRANSMISSION SYSTEM LBO-H2 Series DIRECT PLUGGABLE LINKBRIDE TM FIBER OPTIC HDMI 2.0 TRANSMISSION SYSTEM BCI reserves the right to make changes to the products described herein without prior notice or consent. No liability

More information

Torque Meter. Instruction Manual

Torque Meter. Instruction Manual Torque Meter 840062 Instruction Manual Torque Meter - 840062 Copyright 2013 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or transmitted in

More information

MODEL 245 / 345 High Range

MODEL 245 / 345 High Range MODEL 245 / 345 High Range PRESSURE TRANSMITTER INSTALLATION DATA MANUAL 199 Fire Tower Drive Tonawanda, NY 14150 Toll Free: 1-800-688-0030 International: 716-629-3800 Fax: 716-693-9162 www.viatran.com

More information

Datalogging Conductivity/ TDS Meter. Instruction Manual

Datalogging Conductivity/ TDS Meter. Instruction Manual Datalogging Conductivity/ TDS Meter 850039 Instruction Manual 1 TABLE OF CONTENTS 1. INTRODUCTION...3 2. PANEL DESCRIPTION...4 3. MEASURING PROCEDURE...5 3-A General Measurement...5 3-B Auto and Manual

More information

Series STA DIGITAL TORQUE TOOL TESTERS. User s Guide

Series STA DIGITAL TORQUE TOOL TESTERS. User s Guide Series STA DIGITAL TORQUE TOOL TESTERS User s Guide Series STA Torque Tool Testers Thank you! Thank you for purchasing a Mark-10 Series STA Digital Torque Tool Tester. We are confident that you will get

More information

GVIS Video Inspection Scope

GVIS Video Inspection Scope User s Guide GVIS 300 - Video Inspection Scope GVIS1-00-2000, Rev A, 6-09-20 Specifications subject to change without notice Contents Section 1 - General Overview A. Contact Tempo for customer service

More information

ma Input Module. Installation and Operation Guide

ma Input Module. Installation and Operation Guide 780 4-20 ma Input Module Installation and Operation Guide Part #60-9003-064 Copyright 1995. All rights reserved, Isco, Inc. Revision E, April 20, 2011 Foreword This instruction manual is designed to help

More information

DCS200/DCS200-09/DCS DCS300/DCS300-09/DCS355

DCS200/DCS200-09/DCS DCS300/DCS300-09/DCS355 THE SEEKER 200 SERIES & THE SEEKER 300 SERIES VIDEO INSPECTION SYSTEMS DCS200 (shown) DCS300 (shown) DCS200/DCS200-09/DCS200-05 DCS300/DCS300-09/DCS355 USER S MANUAL Please read this manual carefully and

More information

IAQ-CALC TM Indoor Air Quality Meter

IAQ-CALC TM Indoor Air Quality Meter ENERGY AND COMFORT Indoor Air Quality IAQ-CALC TM Indoor Air Quality Meter Model 7535 Operation and Service Manual Copyright TSI Incorporated / May 2007 / All rights reserved. Address TSI Incorporated

More information

User's Guide. Video Borescope. Model BR100

User's Guide. Video Borescope. Model BR100 User's Guide Video Borescope Model BR100 Introduction Congratulations on your purchase of this Extech BR100 Video Borescope. This instrument was designed for use as an inspection device. It can be used

More information

SP-400 Portable Water Analyzer Operation Manual. Rev. A Firmware version 1.02

SP-400 Portable Water Analyzer Operation Manual. Rev. A Firmware version 1.02 SP-400 Portable Water Analyzer Operation Manual Rev. A Firmware version 1.02 Pyxis Lab, Inc 120 Jeffrey Ave Holliston, MA 01746 USA www.pyxis-lab.com service@pyxis-lab.com 1 Contents 1 General Description...

More information

Rev. F January 13, 1999 Part Number: OFL Operations Manual for the OFL 100 Mini-OTDR

Rev. F January 13, 1999 Part Number: OFL Operations Manual for the OFL 100 Mini-OTDR Rev. F January 13, 1999 Part Number: OFL1-00-1000 Operations Manual for the OFL 100 Mini-OTDR Contents Page 1.0 General Information... 1 1.1 Unpacking and Inspection... 1 1.2 Precautions... 1 1.3 Recommended

More information

Series ST. User s Guide DIGITAL TORQUE TOOL TESTERS. .com. Series ST Torque Tool Testers. User s Guide

Series ST. User s Guide DIGITAL TORQUE TOOL TESTERS. .com. Series ST Torque Tool Testers. User s Guide Series ST DIGITAL TORQUE TOOL TESTERS.com Mark-10 Corporation has been an innovator in the force and torque measurement fields since 1979. We strive to achieve 100% customer satisfaction through excellence

More information

IVC-222HPII Precision Voltage/Current Calibrator

IVC-222HPII Precision Voltage/Current Calibrator IVC-222HPII Precision Voltage/Current Calibrator Table of Contents 1. Introduction 1.1 Customer Service................2 1.2 Unpacking.....................2 2. Operating Procedure...............3 3. Operating

More information

AirPro Surveyor 2 Manual

AirPro Surveyor 2 Manual AirPro Surveyor 2 Manual AirPro Surveyor Specifications Table of Contents Size 3/8 x 7 1/2 x 4 3/8 Weight 4.6 lbs. 2094 g Dynamic Range 1-1000 ml/min. total flow/constant flow Flow Capacity (8 Hrs.) 1000

More information

Light Meter with Bar Graph

Light Meter with Bar Graph Light Meter with Bar Graph 840020 Instruction Manual SPER SCIENTIFIC LTD. TABLE OF CONTENTS INTRODUCTION... 2 PANEL DESCRIPTION... 3 MEASUREMENT PROCEDURES... 3-4 ZERO ADJUSTMENT... 4 BATTERY REPLACEMENT...

More information

MiniSim 330 EEG Simulator

MiniSim 330 EEG Simulator INSTRUCTION MANUAL MiniSim 330 EEG Simulator Netech Corporation 110 Toledo Street Farmingdale, NY 11735 USA www.gonetech.com Dear User: We appreciate your purchase of the MiniSim 330 EEG Simulator. Properly

More information

V-9908 MESSAGE/PAGE PANEL

V-9908 MESSAGE/PAGE PANEL Issue 4 V-9908 MESSAGE/PAGE PANEL Introduction These instructions contain the specifications and guidelines necessary to install, operate, and maintain the V-9908, /Page Panel. The V-9908 /Page Panel provides

More information

VIP-812A DUAL NETWORKED STATION PORT

VIP-812A DUAL NETWORKED STATION PORT ISSUE 1 VIP-812A DUAL NETWORKED STATION PORT INTRODUCTION The VIP-812A Dual Networked Station Port allows most loop start terminal devices to be connected to a managed IP-based LAN/WAN. SPECIFICATIONS

More information

INSTRUCTION MANUAL. Model True RMS AC/DC 30A Mini Clamp-on Meter. Introduction. True RMS AC Current and Voltage

INSTRUCTION MANUAL. Model True RMS AC/DC 30A Mini Clamp-on Meter. Introduction. True RMS AC Current and Voltage INSTRUCTION MANUAL Model 380942 True RMS AC/DC 30A Mini Clamp-on Meter True RMS AC Current and Voltage Measure low current with high resolution to 0.1mA AC and 1mA DC Auto Power Off One touch DCA zero

More information

AIR PATHWAYS AP340 OWNERS MANUAL

AIR PATHWAYS AP340 OWNERS MANUAL AIR PATHWAYS VII WARRANTY/ SERVICE AIR PATHWAYS warrants this product to be free from defect in materials and workmanship for a period of two years from the date of installation to the original purchaser

More information

User Guide. ExStik TM Model RE300 Waterproof ORP Meter. Patent Pending RE300 ORP

User Guide. ExStik TM Model RE300 Waterproof ORP Meter. Patent Pending RE300 ORP 0 User Guide ExStik TM Model RE300 Waterproof ORP Meter Patent Pending 500 MV 1000 TM RE300 ORP ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE button 4. CAL

More information

XBDM. 1015LV, 1020LV, 1030LV, 1020HV Models USER & INSTALLATION MANUAL BYPASS DISTRIBUTION MODULE

XBDM. 1015LV, 1020LV, 1030LV, 1020HV Models USER & INSTALLATION MANUAL BYPASS DISTRIBUTION MODULE XBDM 1015LV, 1020LV, 1030LV, 1020HV Models USER & INSTALLATION MANUAL www.xpcc.com 2013 Xtreme Power Conversion Corporation. All rights reserved. Table of Contents IMPORTANT SAFETY INSTRUCTIONS:... 4 INTRODUCTION...

More information

POWER SERIES Plus. 3 in 1 AC Voltage. Digital Switchboard Meter. User s Manual IM2493VVV-2

POWER SERIES Plus. 3 in 1 AC Voltage. Digital Switchboard Meter. User s Manual IM2493VVV-2 POWER SERIES Plus 3 in 1 AC Voltage Digital Switchboard Meter User s Manual General Description The POWER SERIES Plus digital switchboard meters incorporate the latest DSP microprocessor technology. Careful

More information

CO2 Controller Operating Instructions Models: RAD-0501, RAD-0501A, RAD-0501E 1. Product Description

CO2 Controller Operating Instructions Models: RAD-0501, RAD-0501A, RAD-0501E 1. Product Description CO2 Controller Operating Instructions Models: RAD-0501, RAD-0501A, RAD-0501E 1. Product Description RAD-0501 Greenhouse Mode: Controls CO2 generator or regulator to increase CO2 levels during daylight

More information

User s Manual. Digital Baby Scales. Hopkins Featherweight Baby Scale # Hopkins Featherweight XL Baby Scale # # #688340

User s Manual. Digital Baby Scales. Hopkins Featherweight Baby Scale # Hopkins Featherweight XL Baby Scale # # #688340 Max=44s/2kg d=.1s/5g RECALL Digital Baby Scales User s Manual Hopkins Featherweight Baby Scale # Hopkins Featherweight XL Baby Scale #68834 Max=44s/2kg d=.1s/5g # ON/TARE LB/KG #68834 Table of Contents

More information

IAQ-CALC TM Indoor Air Quality Meter

IAQ-CALC TM Indoor Air Quality Meter ENERGY AND COMFORT Indoor Air Quality IAQ-CALC TM Indoor Air Quality Meter Model 7525 Operation and Service Manual Copyright TSI Incorporated / August 2008 / All rights reserved. Address TSI Incorporated

More information

QuickTouch (QT4) Owner s Manual

QuickTouch (QT4) Owner s Manual QuickTouch (QT4) Owner s Manual 4-Function Hand-Held Wireless Remote Control IMPORTANT SAFETY INSTRUCTIONS READ AND FOLLOW ALL INSTRUCTIONS SAVE THESE INSTRUCTIONS Table of Contents SECTION I. APPLICATION...

More information

Datalogging Manometer

Datalogging Manometer Datalogging Manometer 840086, 840098 and 840099 Instruction Manual TABLE OF CONTENTS Introduction.. 3 Front Panel Descriptions.. 4 Set-up.. 5-6 Calibration.. 7 Measurement Procedures.. 8-14 Error Codes..

More information

Residential/Light Commercial Remote Control System

Residential/Light Commercial Remote Control System MODULAR CONTROLLER REMOTE CONTROL Residential/Light Commercial Remote Control System OWNER S MANUAL AND INSTALLATION INSTRUCTIONS CONTENTS INTRODUCTION 2 SYSTEM COMPONENTS - REMOTE 3 SYSTEM COMPONENTS

More information

DCS-E 1kW Series, DLM-E 3kW & 4kW Power Supplies

DCS-E 1kW Series, DLM-E 3kW & 4kW Power Supplies DCS-E 1kW Series, DLM-E 3kW & 4kW Power Supplies M51A Option: Isolated Analog Programming Manual Power Supplies Elgar Electronics Corporation 9250 Brown Deer Road San Diego, CA 92121-2294 1-800-73ELGAR

More information

User's Guide. MiniTec TM Series Model MN25 MultiMeter

User's Guide. MiniTec TM Series Model MN25 MultiMeter User's Guide MiniTec TM Series Model MN25 MultiMeter Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a

More information

Model CR103J. DC Current and Voltage Calibrator/Standard ±0.01µA to ± mA ±0.1µV to ± V. Operating Manual

Model CR103J. DC Current and Voltage Calibrator/Standard ±0.01µA to ± mA ±0.1µV to ± V. Operating Manual Model CR103J DC Current and Voltage Calibrator/Standard ±0.01µA to ±111.1110mA ±0.1µV to ±11.11110V Operating Manual This page intentionally left blank. CR103J OPERATORS MANUAL Serial No. This page intentionally

More information

TC100 Precision Thermocouple Calibrator

TC100 Precision Thermocouple Calibrator TC100 Precision Thermocouple Calibrator Table of Contents 1. Introduction.....................1 2. Accessories....................5 3. Set-Up Basic/Advanced...........5 4. Operating Procedure.............6

More information

SERVICE MANUAL MODEL SSW-521-D2-ADA

SERVICE MANUAL MODEL SSW-521-D2-ADA SSW-521-D2-ADA-SPK1.07UNVLr3-ADA-ISSUE4.0 SERVICE MANUAL FOR MODEL SSW-521-D2-ADA STAINLESS STEEL HANDS FREE WALL TELEPHONE WITH OPTIONAL ADA FEATURE AND OPTIONAL SECOND-NUMBER AUTO-DIAL FEATURE EQUIPPED

More information

Mini-DVI-WP Series WALL PLATE MINIATURE MULTIMODE FIBER OPTIC DVI TRANSMISSION SYSTEM

Mini-DVI-WP Series WALL PLATE MINIATURE MULTIMODE FIBER OPTIC DVI TRANSMISSION SYSTEM Mini-DVI-WP Series WALL PLATE MINIATURE MULTIMODE FIBER OPTIC DVI TRANSMISSION SYSTEM BCI reserves the right to make changes to the products described herein without prior notice or consent. No liability

More information

MD9 MULTIDROP INTERFACE INSTRUCTION MANUAL

MD9 MULTIDROP INTERFACE INSTRUCTION MANUAL MD9 MULTIDROP INTERFACE INSTRUCTION MANUAL REVISION: 01/05/03 COPYRIGHT (c) 1987-2003 CAMPBELL SCIENTIFIC, INC. This is a blank page. WARRANTY AND ASSISTANCE The MD9 MULTIDROP INTERFACE is warranted by

More information

eforce 150 Keyless Entry Owner s manual & User s guide For Model 3090

eforce 150 Keyless Entry Owner s manual & User s guide For Model 3090 eforce 150 Keyless Entry Owner s manual & User s guide For Model 3090 This manual contains important operation, maintenance & warranty information. Save this manual for future reference TABLE OF CONTENTS

More information

FA-2448 SIX POSITION Filter Wheel

FA-2448 SIX POSITION Filter Wheel 15 Discovery Way, Acton, MA 01720 Phone: (978)263-3584, Fax: (978)263-5086 Web Site: www.acton-research.com Operating Instructions Acton Research Corporation FA-2448 SIX POSITION Filter Wheel Rev. 3.05.17

More information

353 PRECISION ALTIMETER TESTER OPERATING INSTRUCTIONS

353 PRECISION ALTIMETER TESTER OPERATING INSTRUCTIONS 353 PRECISION ALTIMETER TESTER OPERATING INSTRUCTIONS The Model 353 provides Altitude, Rate of Climb (R of C), and Leak testing at the stroke of a key. It is microprocessor based and its accuracy is ±

More information

SmartBoom PRO PHS-SB100-4F, PHS-SB100-5M, PHS-SB100-U, PHS-SB200-4F, PHS-SB200-5M, PHS-SB200-U. Operating Manual

SmartBoom PRO PHS-SB100-4F, PHS-SB100-5M, PHS-SB100-U, PHS-SB200-4F, PHS-SB200-5M, PHS-SB200-U. Operating Manual SmartBoom Operating Manual PRO PHS-SB100-4F, PHS-SB100-5M, PHS-SB100-U, PHS-SB200-4F, PHS-SB200-5M, PHS-SB200-U Thank You We at Pliant Technologies, LLC want to thank you for purchasing the SmartBoom PRO

More information

CTI 2551-A 8-CHANNEL ISOLATED THERMOCOUPLE INPUT MODULE INSTALLATION AND OPERATION GUIDE. Version 1.3 CTI Part # * *

CTI 2551-A 8-CHANNEL ISOLATED THERMOCOUPLE INPUT MODULE INSTALLATION AND OPERATION GUIDE. Version 1.3 CTI Part # * * CTI 2551-A 8-CHANNEL ISOLATED THERMOCOUPLE INPUT MODULE INSTALLATION AND OPERATION GUIDE Version 1.3 CTI Part #062-00305-013 *062-00305-011* 2551-AIOG 110106 $25 Copyright 2006 Control Technology Inc.

More information

Telex Operating Instructions

Telex Operating Instructions Telex Operating Instructions Echelon ANR TM 150 Headset Figure 1 Note: See page 7 for available replacement parts. General Description The Echelon ANR 150 is a medium-weight aircraft communications headset.

More information

Operations Manual EcoSense ph100

Operations Manual EcoSense ph100 Operations Manual EcoSense ph100 Portable ph, mv and Temperature Instrument English Français Español Deutsch Italiano WARRANTY The EcoSense ph100 Instrument is warranted for one year from date of purchase

More information

DICKSON FH320/325 & FT300/325 DICKSON. Temperature/Humidity Recorder. Specifications. Applications & Product. Product. Contents: Getting Started

DICKSON FH320/325 & FT300/325 DICKSON. Temperature/Humidity Recorder. Specifications. Applications & Product. Product. Contents: Getting Started FH320/325 & FT300/325 Temperature/Humidity Recorder Contents: and Instrument Anatomy Accessories Warranty / Accessories, & Calibration FH325 Remote Probe Temperature & Humidity FT325 Remote Probe 2 Channel

More information

Hybrid AC Driver [GCNC-1110]

Hybrid AC Driver [GCNC-1110] Page 1 Installation Manual and Datasheet Page 2 Key Features Smooth and quiet operation at all speeds and extremely low motor heating Industrial grade performance for an alternating current servo motor

More information

VELOCICALC AIR VELOCITY METER MODEL 9545/9545-A

VELOCICALC AIR VELOCITY METER MODEL 9545/9545-A VELOCICALC AIR VELOCITY METER MODEL 9545/9545-A OPERATION AND SERVICE MANUAL P/N 1980564, REVISION C SEPTEMBER 2013 Copyright TSI Incorporated / 2007-2013 / All rights reserved. Address TSI Incorporated

More information

VE8014A/VE8014AR QUAD NETWORKED STATION PORT

VE8014A/VE8014AR QUAD NETWORKED STATION PORT ISSUE 2 VE8014A/VE8014AR QUAD NETWORKED STATION PORT INTRODUCTION The VE8014A/VE8014AR Quad Networked Station Port allows most loop start terminal devices to be connected to a managed IP-based LAN/WAN.

More information

SERVICE MANUAL MODEL SSW-521-F (FORMERLY SSW-521-FS)

SERVICE MANUAL MODEL SSW-521-F (FORMERLY SSW-521-FS) SSW-521-F-SPK1.07UNVL-ADA-ISSUE4.0 SERVICE MANUAL FOR MODEL SSW-521-F (FORMERLY SSW-521-FS) STAINLESS STEEL HANDS FREE WALL TELEPHONE OPTIONAL ADA FEATURE AVAILABLE EQUIPPED WITH SPK1.07UNVL FIRMWARE Serving

More information

INSTALLATION AND USER GUIDE 2800MWB SINGLE LINE BASIC FEATURE TELEPHONE

INSTALLATION AND USER GUIDE 2800MWB SINGLE LINE BASIC FEATURE TELEPHONE INSTALLATION AND USER GUIDE 2800MWB SINGLE LINE BASIC FEATURE TELEPHONE TeleMatrix Copyright 2005 COMPLIANCE AND SAFETY As specified by FCC regulation, we are required to inform you of specific governmental

More information

Series RMA Hot Runner Controller Setup and Operation Manual

Series RMA Hot Runner Controller Setup and Operation Manual Series RMA Hot Runner Controller Setup and Operation Manual Littlefuse is a registered trademark of Littlefuse, Inc. G is a registered trademark of the DME Corporation. SafeChange is a trademark of Athena

More information

SM-10 SM-20-A. Sound Level Meters. Users Manual. For detailed specifications and ordering info go to

SM-10 SM-20-A. Sound Level Meters. Users Manual. For detailed specifications and ordering info go to SM-10 SM-20-A Sound Level Meters Users Manual For detailed specifications and ordering info go to www.testequipmentdepot.com SM-10 / SM-20-A Sound Level Meters English Users Manual SM10_Rev002 2009 Amprobe

More information

DICKSON TC200/TH300 DICKSON. Handheld Temperature and Temperature / RH Indicators. Useful Features. Applications & Product. Specifications.

DICKSON TC200/TH300 DICKSON. Handheld Temperature and Temperature / RH Indicators. Useful Features. Applications & Product. Specifications. / Handheld Temperature and Temperature / RH Indicators Contents: Applications and Accessories Warranty / This versatile temperature indicator accepts J, K and T thermocouple probes with mini-connectors.

More information

HI ma Simulator

HI ma Simulator Instruction Manual HI 931002 4-20 ma Simulator http://www.hannainst.com These Instruments are in Compliance with the CE Directives Dear Customer, Thank you for choosing a Hanna Instruments Product. Please

More information

SERVICE MANUAL MODEL SSW-520-F

SERVICE MANUAL MODEL SSW-520-F SSW-520-F-ISSUE4.0 SERVICE MANUAL FOR MODEL SSW-520-F HANDS FREE STAINLESS STEEL TELEPHONE EQUIPPED WITH LOW POWER SPEAKER BOARD Serving the Telephone Industry Since 1930 Communication Equipment 519 West

More information