Breast Tomosynthesis: Impact on IT Systems. dondennison.com

Size: px
Start display at page:

Download "Breast Tomosynthesis: Impact on IT Systems. dondennison.com"

Transcription

1 Breast Tomosynthesis: Impact on IT Systems MIIT 2016, HAMILTON

2 Don K Dennison Industry consultant with 14+ years experience in imaging IT Society for Imaging Informatics in Medicine (SIIM) Board of Directors (2011-Current) Committees: Program, Hackathon, ecommunity American College of Radiology (ACR) Committees: Connect (Chair), Informatics Industry Activities (Chair), Clinical Data Sciences Contact 2

3 Don K Dennison Faculty/Presenter Disclosure Relationships with commercial interests: Consulting Fees: Provider and Vendor clients 2

4 Today s Talk What is Breast Tomosynthesis and why is its popularity growing? What makes the exam data so large and what factors contribute to the size of each exam? OK, how big are we talking about? OMG, that big? What will this do to my imaging IT systems? What can I do to prepare for, or avoid, negative system (or IT cost) impacts?

5 Breast Tomosynthesis (Tomo) New modality; market share leader is Hologic 1 Increasing evidence as to its effectiveness over traditional Mammograms, especially for women with dense breast tissue Increasingly adopted as physicians and patients view Breast Tomo as superior to Mammograms; traditional Mammogram exam volumes are often replaced by Breast Tomo exam volumes Healthcare providers that do not offer Breast Tomo often see a loss of volume to providers that do 1) Most technical details in this talk refer to Hologic and their products

6 Visual Differences MAMMOGRAM (MG) DIGITAL BREAST TOMOSYNTHESIS (DBT) Source: American Journal of Roentgenology: Issues to Consider Before Implementing Digital Breast Tomosynthesis Into a Breast Imaging Practice

7 Breast Tomo Data Types Several different types of series are normally contained within the study C-View 2D image that visually appears similar to a Mammogram (MG) exam; average size is 1.5 MB per image (small) Projections Average size is 8.5 MB per image (medium) Mammogram 2D Traditional MG exam data (medium) Reconstructions 3D multi-frame DICOM object with a varying number of frames, based on the size of the breast; average size ranges from 70 to 200 MB, or more (large) Note: Many facilities do not archive the Projection series data

8 Breast Tomo Data in PACS RCC LCC DBT C-View Screen captures courtesy McKesson Imaging and Workflow Solutions RCC Slab

9 Breast Tomo More Details Each exam will have a different number of Views acquired, which normally range between 4 and 6 Typical lossless image compression methods can (and should) be applied U.S. MQSA regulations do not allow lossy compression to be applied Multiplying the number of Views for each series type results in large studies If a patient requires many Views to be acquired, has large breasts, and the Projection series are archived, each study can be very large (many GBs in size!) and then there are the formats

10 Breast Tomo Multiple Formats Available BTO DICOM standard defined SOP Class object Valid DICOM Transfer Syntaxes applied Many PACS and Enterprise Viewers now support display ~2.5 times larger than SCO data SCO Secondary Capture Proprietary to Hologic Easy to store in PACS/VNA, but not display Can be converted to BTO, in future, using Hologic-provided tool Viewed using Hologic workstation Additional information from SIIM 2013:

11 Deciding on a Format: Choose Wisely BTO DICOM standard, so interoperable among systems and organization enterprises Supported by many PACS, so can reduce types of clients required, desktop integrations Also, can support integrated display of Tomo, MG, MRI, US SCO Much smaller than BTO, so will cost less to store and will take less time to transmit Cannot be viewed by any affiliate that does not have a Hologic workstation SCO can be converted to BTO by Hologic Tech workstation, tool

12 Introducing the Breast Tomosynthesis IT Impact Calculator Available for no-cost download, along with instructions: Allows entry of your study size averages, volumes, etc. Calculates the impact to your long-term archive and PACS cache, along with costs Estimates file transmission times Includes some default graphs

13 So, Exactly How Large are we Talking About? The average number of Views acquired, number of frames in the Recon object, and whether or BTO or SCO data are captured all have a large and multiplicative impact on imaging IT systems Could store both BTO and SCO data, but the cost will be prohibitive

14 Impact on the PACS Cache 2 Years 1 Year PACS cache capacity is consumed much faster, reducing the amount of data available for all studies; therefore, data is deleted from cache much sooner and more priors will need to be retrieved from the archive

15 Impact on the Archive (PACS, VNA) Costs Based on US$0.70/GB The archive storage capacity consumed is significant, requiring more to be purchased, much sooner

16 Breast Tomo Compared to Other Imaging For example: Using typical values, in 10 years, Breast Tomo (BTO) data will consume more storage than 25 years of all other imaging data types combined

17 Network Requirements for Fast Display All but the fastest networks will be a bottleneck in transmitting Breast Tomo data, especially if there are multiple priors and large Recon objects Note: Client workstations also often require a 64-bit operating system and additional RAM to be able to load the large datasets

18 Closing Comments Organizations that do not prepare for the impact of Breast Tomo data in their imaging IT systems will be caught off guard and may suffer from performance issues and/or unplanned costs The choice between SCO and BTO formats is multi-faceted, involving interoperability, accessibility, performance and costs aspects: consider all of these when making you decision Due to the size of Breast Tomo imaging study data, and the IT costs to store them, shared per-study PACS cost models may need to be revised the per-study cost charged to a facility may not cover the cost of storing the study data

19 Closing Comments Organizations will need to examine their policies, procedures, and tools regarding imaging record retention it may be time to start purging old imaging data to offset the new, higher costs caused by Breast Tomo The impact on PACS cache may require the Breast Tomo data to be segregated to its own cache, or have advanced lifecycle management policies applied to maintain expected availability of all imaging

Digital Breast Tomosynthesis PACS/Workstation Vendor Perspective. Ron Muscosky Product Line Manager for Carestream Health

Digital Breast Tomosynthesis PACS/Workstation Vendor Perspective. Ron Muscosky Product Line Manager for Carestream Health Digital Breast Tomosynthesis PACS/Workstation Vendor Perspective Ron Muscosky Product Line Manager for Carestream Health Digital Breast Tomosynthesis Digital Breast Tomosynthesis (DBT) is an important

More information

Lossless Compression of Breast Tomosynthesis Objects Maximize DICOM Transmission Speed and Review Performance and Minimize Storage Space

Lossless Compression of Breast Tomosynthesis Objects Maximize DICOM Transmission Speed and Review Performance and Minimize Storage Space Lossless Compression of Breast Tomosynthesis Objects Maximize DICOM Transmission Speed and Review Performance and Minimize Storage Space David A. Clunie PixelMed Publishing What is Breast Tomosynthesis?

More information

Informatics I - The digital breast image interoperability and management

Informatics I - The digital breast image interoperability and management Informatics I - The digital breast image interoperability and management Alain Gauvin Montebello, 2017-02-01 DICOM HL7 VARIABLE Orders, demography, (HL7 ORM, ADT) Images + metadata (dicom storage) Modality

More information

A Closer Look at SERVER-SIDE RENDERING. Technology Overview

A Closer Look at SERVER-SIDE RENDERING. Technology Overview A Closer Look at SERVER-SIDE RENDERING Technology Overview Driven by server-based rendering, Synapse 5 is the fastest PACS in the medical industry, offering subsecond image delivery and diagnostic quality.

More information

10/9/2018. Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT

10/9/2018. Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT 1 Mammography Five FFDM approved for Tomo Hologic, Ge Senoclair and GE Pristina, Siemens, and Fujifilm 2 Why 2D Digital

More information

Context Management and Tag Morphing in the Real World Wayne T. DeJarnette, Ph.D. President, DeJarnette Research Systems, Inc.

Context Management and Tag Morphing in the Real World Wayne T. DeJarnette, Ph.D. President, DeJarnette Research Systems, Inc. White Paper Series Context Management and Tag Morphing in the Real World Wayne T. DeJarnette, Ph.D. President, DeJarnette Research Systems, Inc. January 4, 2010 DeJarnette Research Systems, Inc., 2010

More information

This document contains confidential information that is proprietary to SonoSite. Neither the document nor the information contained therein should be

This document contains confidential information that is proprietary to SonoSite. Neither the document nor the information contained therein should be This document contains confidential information that is proprietary to SonoSite. Neither the document nor the information contained therein should be disclosed or reproduced in whole or in part, without

More information

DICOM Conformance Statement for Advanced Workflow Manager Software Version 1.8 Part Number MAN Revision 001

DICOM Conformance Statement for Advanced Workflow Manager Software Version 1.8 Part Number MAN Revision 001 DICOM Conformance Statement for Advanced Workflow Manager Software Version 1.8 Part Number MAN-04090 Revision 001 July 2014 Table of Contents TABLE OF CONTENTS... 3 1.0 OVERVIEW... 4 2.0 INTRODUCTION...

More information

4/19/2016. Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT

4/19/2016. Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT 1 Three manufacturers approved for Tomo Hologic and GE, and Siemens Why 2D Digital Mammography 2D FFDM it appears to

More information

DICOM V3.0 Conformance Statement. SenoIris 1SP2

DICOM V3.0 Conformance Statement. SenoIris 1SP2 DICOM V3.0 Conformance Statement SenoIris 1SP2 Image Diagnost International Reproduction of this document without prior written approval is prohibited. Image Diagnost International reserves the right to

More information

Hologic Physician s Viewer 5.0 DICOM Conformance Statement

Hologic Physician s Viewer 5.0 DICOM Conformance Statement Hologic Physician s Viewer 5.0 DICOM Conformance Statement MAN-00239 REV 001 Page 1 of 18 1 INTRODUCTION... 4 1.1 Purpose of the Document...4 1.2 References...4 1.3 Definitions...4 2 IMPLEMENTATION MODEL...

More information

Selenia Dimensions 3Dimensions Advanced Workflow Manager. DICOM Conformance Statement For Software Versions 1.9 and 2.0

Selenia Dimensions 3Dimensions Advanced Workflow Manager. DICOM Conformance Statement For Software Versions 1.9 and 2.0 Selenia Dimensions 3Dimensions Advanced Workflow Manager DICOM Conformance Statement For Software Versions 1.9 and 2.0 Part Number MAN-05339 Revision 001 July 2017 2014-2017 Hologic, Inc. Printed in the

More information

SenoClaire DICOM CONFORMANCE STATEMENT. Senographe Essential Version:

SenoClaire DICOM CONFORMANCE STATEMENT. Senographe Essential Version: SenoClaire Senographe Essential Version: ADS 56.10, 56.11, 56.12 Recon 01.10, 01.10.1, 01.10.2, 01.10.3 DICOM CONFORMANCE STATEMENT 5415902-3-8EN 2013-2014 by General Electric Company All Rights Reserved.

More information

Certificate Extension II 8/1/2016. Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role

Certificate Extension II 8/1/2016. Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role Kish Chakrabarti, Ph.D., FAAPM Division of Mammography Quality Standards Center for Devices and Radiological

More information

DICOM Conformance Statement for PenFetch

DICOM Conformance Statement for PenFetch Penrad Technologies, Inc. DICOM Conformance Statement for PenFetch 10580 Wayzata Blvd. Suite 200 Minnetonka, MN 55305 www.penrad.com Copyright PenRad Technologies, Inc. May 2007 Page 1 of 9 1 Introduction...3

More information

No. MIIUS0015EA DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM APLIO MODEL SSA-770A TOSHIBA CORPORATION 2001 ALL RIGHTS RESERVED

No. MIIUS0015EA DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM APLIO MODEL SSA-770A TOSHIBA CORPORATION 2001 ALL RIGHTS RESERVED No. MIIUS0015EA DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM APLIO MODEL SSA-770A TOSHIBA CORPORATION 2001 ALL RIGHTS RESERVED IMPORTANT! (1) No part of this manual may be copied or reprinted,

More information

Mississippi State University RFP Enterprise Imaging Informatics Solutions Questions and Answers September 13, 2017

Mississippi State University RFP Enterprise Imaging Informatics Solutions Questions and Answers September 13, 2017 Mississippi State University RFP 17-73 Enterprise Imaging Informatics Solutions Questions and Answers September 13, 2017 The list of questions below were received. Please use this information when preparing

More information

available technologies in remote diagnostics THE WORLD IN YOUR HANDS

available technologies in remote diagnostics THE WORLD IN YOUR HANDS TELERADIOLOGY available technologies in remote diagnostics THE WORLD IN YOUR HANDS AVAILABLE TECHNOLOGIES access your medical images anytime and anywhere Selecting the right solution for teleradiology

More information

DICOM Conformance Statement

DICOM Conformance Statement DICOM Conformance Statement TOPCON Corporation TOPCON SPECULAR MICROSCOPE SP-1P Version 1.10 Document Version: 1 Date: 2014-02-17 1 CONFORMANCE STATEMENT OVERVIEW This document declares conformance to

More information

DICOM Conformance Statement for Selenia Dimensions Acquisition Workstation Software Version 1.8 Part Number MAN Revision 002

DICOM Conformance Statement for Selenia Dimensions Acquisition Workstation Software Version 1.8 Part Number MAN Revision 002 DICOM Conformance Statement for Selenia Dimensions Acquisition Workstation Software Version 1.8 Part Number MAN-04089 Revision 002 April 2015 Table of Contents Selenia Dimensions 1.8 DICOM Conformance

More information

FusionXD 2.0 DICOM Conformance Statement

FusionXD 2.0 DICOM Conformance Statement FusionXD 2.0 DICOM Conformance Statement 1 Introduction 1.1 Revision History See the Revision History in the beginning of this document 1.2 References Reference REF1 Document Digital Imaging and Communications

More information

DAR Revision Date: July,

DAR Revision Date: July, S517-E053B DAR-9000 Revision Date: July, 25. 2005 DICOM service classes in this manual are provided as an optional item of DAR-9000. For your suitable system configuration, please consult with SHIMADZU

More information

Media Writer DICOM Conformance Statement

Media Writer DICOM Conformance Statement Media Writer DICOM Conformance Statement July 12, 2016 Page 1 of 16 Lexmark Enterprise Software LLC - Proprietary 2016 Lexmark. Lexmark and the Lexmark logo are trademarks of Lexmark International, Inc.,

More information

IHE Integration profiles

IHE Integration profiles Integrating the Healthcare Enterprise Access to Radiology Information Cor Loef Co-chair IHE Radiology Technical Committee 1 IHE Integration profiles Scheduled Workflow of Grouped Procedures Patient Information

More information

Medical Imaging Consultants, Inc.

Medical Imaging Consultants, Inc. IA2000 Conformance Statement 1 Medical Imaging Consultants, Inc. Conformance Statement IA2000 Document: MICI-210-01 Revision: 6.0 Document Status: Approved When printed, this is NOT a controlled copy Reviewers

More information

CADstream DICOM Conformance Statement

CADstream DICOM Conformance Statement CADstream 5.2.9 DICOM Conformance Statement CADstream 5.2.9 DICOM Conformance Statement This document has been prepared by Cedara Software Corp. doing business as Merge OEM (hereinafter referred to as

More information

Healthcare IT A Monitoring Primer

Healthcare IT A Monitoring Primer Healthcare IT A Monitoring Primer Published: February 2019 PAGE 1 OF 13 Contents Introduction... 3 The Healthcare IT Environment.... 4 Traditional IT... 4 Healthcare Systems.... 4 Healthcare Data Format

More information

HEALTHCARE DICOM Conformance Statement

HEALTHCARE DICOM Conformance Statement MED/PL/000406 Page 1 of 15 HEALTHCARE DICOM Conformance Statement IMPAX Web1000 (Release 2.0) Status: Released Page 2 of 15 MED/PL/000406 26 September, 2002 HealthCare Document Information Author Patricia

More information

Web Access of DICOM Objects (WADO)

Web Access of DICOM Objects (WADO) Web Access of DICOM Objects (WADO) Engineer Amer khraisat, Engineer Mahmoud Al Ikour, Mohammad Nour, Engineer AhmadAlkouz, AbdAlazizAlqisy, Ahmad Elamaireh. The Institute of biomedical technology, Jordan.

More information

NumaStore 1.0 DICOM 3.0 Conformance Statement

NumaStore 1.0 DICOM 3.0 Conformance Statement External Distribution Revision 2 February 19, 2001 NumaStore 1.0 Revision 2 Date February 19, 2001 Author Bill Kilgore, George Panousis Revision History Revision Date By Comment 1 January 13, 2001 Bill

More information

Copyright 2011, TeraMedica, Inc.

Copyright 2011, TeraMedica, Inc. CHAPTER 1 Implementation Information Make certain that you understand how and who will be providing the support that you need to implement the solution. How many engineering teams are involved in the development

More information

What is unique? EasyViz adds complete mobility to any PACS and EMR

What is unique? EasyViz adds complete mobility to any PACS and EMR EasyViz is an Image Distribution System providing the market s most versatile clinical image viewer. EasyViz works with any modern PACS archive and EMR system directly, there is no need to push or replicate

More information

FusionXD 2.1 DICOM Conformance Statement

FusionXD 2.1 DICOM Conformance Statement FusionXD 2.1 DICOM Conformance Statement Contents 1 Introduction... 3 1.1 References... 3 1.2 Integration and Features... 3 1.3 Definitions, Acronyms and Abbreviations... 3 2 Services... 4 3 Networking...

More information

DICOM Conformance Statement * /02* /02 MADE IN GERMANY

DICOM Conformance Statement * /02* /02 MADE IN GERMANY EN DICOM 9000-608-109/02 *9000-608-109/02* MADE IN GERMANY Overview Vet-Exam Intra controls the digital DÜRR MEDICAL imaging products. The Vet-Exam Intra DICOM Interface translates between DICOM applications

More information

Philips Medical Systems DICOM Conformance Statement. EasyWeb 2.0. Document Number April 1999

Philips Medical Systems DICOM Conformance Statement. EasyWeb 2.0. Document Number April 1999 Philips Medical Systems DICOM Conformance Statement EasyWeb 2.0 Document Number 9896 050 60201 12 April 1999 Copyright Philips Medical Systems Nederland B.V. 1999 Philips Medical Systems apple PHILIPS

More information

FUJIFILM MEDICAL SYSTEMS

FUJIFILM MEDICAL SYSTEMS FUJIFILM MEDICAL SYSTEMS Product Data Synapse Release Version 4.2 Foundation Technologies Application Synapse server is a collection of software modules built on Microsoft Windows Server, which together

More information

Enhanced Multi-frame Images The New Core Paradigm for DICOM

Enhanced Multi-frame Images The New Core Paradigm for DICOM THE DICOM 2014 INTERNATIONAL SEMINAR August 26 Chengdu, China Enhanced Multi-frame Images The New Core Paradigm for DICOM Harry Solomon Interoperability Architect, GE Healthcare Co-chair, DICOM Standards

More information

Separating PACS Servers from VNA and then connecting them

Separating PACS Servers from VNA and then connecting them Separating PACS Servers from VNA and then connecting them By Don Dennison May 2014 Sponsored by: Hitachi Data Systems and Brocade Communication Systems 1 Contents Executive Summary... 3 First, what is

More information

Version 9 DICOM Conformance Statement. Version 3.05, September 2015

Version 9 DICOM Conformance Statement. Version 3.05, September 2015 Version 9 DICOM Conformance Statement Version 3.05, September 2015 1 Conformance Statement Overview The application described in this conformance statement, VEPRO EMR Manager PACS, is a collection of processes

More information

DICOM 3.0. ENsphere CONFORMANCE STATEMENT. Physician s Workstation

DICOM 3.0. ENsphere CONFORMANCE STATEMENT. Physician s Workstation DICOM 3.0 CONFORMANCE STATEMENT ENsphere Physician s Workstation Copyright Statement ADAC Laboratories, a Philips Medical Systems Company, has taken care to ensure the accuracy of this document. However,

More information

Truly Advanced Visualization

Truly Advanced Visualization A FEATURED EBOOK Truly Advanced Visualization Understanding the most important dimensions Truly Advanced Visualization 1 CONTENTS 3 4 5 Look Who s Talking Connecting Workflow How Do I Choose? Truly Advanced

More information

OASIS V4.0 DICOM Conformance Statement Rev. 3

OASIS V4.0 DICOM Conformance Statement Rev. 3 0BMR Imaging System OASIS V4.0 DICOM Conformance Statement Rev. 3 Tokyo, Japan E1E-BM7403-3 Copyright Hitachi Medical Corporation. 2011. All rights reserved. Revision History Revision Date Change Description

More information

STANDARDISATION IN DIGITAL PATHOLOGY. Marcial García Rojo Hospital de Jerez. Cádiz. España Vice-President Spanish Society for Health Informatics

STANDARDISATION IN DIGITAL PATHOLOGY. Marcial García Rojo Hospital de Jerez. Cádiz. España Vice-President Spanish Society for Health Informatics STANDARDISATION IN DIGITAL PATHOLOGY Marcial García Rojo Hospital de Jerez. Cádiz. España Vice-President Spanish Society for Health Informatics Multiple image sources Introduction Digital imaging in pathology

More information

DBSWIN DICOM Conformance Statement. DBSWIN DD-DICOM Interface. DICOM Conformance Statement V2.1

DBSWIN DICOM Conformance Statement. DBSWIN DD-DICOM Interface. DICOM Conformance Statement V2.1 DBSWIN DD-DICOM Interface DICOM Conformance Statement V2.1 DÜRR DENTAL Page 1 / 23 Conformance Statement Overview The DBSWIN software controls the digital Dürr Dental imaging products. The DD-DICOM Interface

More information

Copyright 2014 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future

Copyright 2014 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future Copyright 214 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material

More information

C-View 1.0. DICOM Conformance Statement. MAN Rev 001

C-View 1.0. DICOM Conformance Statement. MAN Rev 001 C-View 1.0 DICOM Conformance Statement MAN-02865 Rev 001 Technical Support For support in North America contact: Toll Free: +1.866.243.2533 (+1.866.CHECKED) Email: scsupport@hologic.com Hours: Monday Friday,

More information

DICOM Conformance Statement

DICOM Conformance Statement DIMEX.HYBRID Schäf Systemtechnik GmbH Falkenstrasse 22 D-91580 Petersaurach Germany Phone: +49 (0) 9872 2002 Fax: +49 (0) 9872 2823 E-Mail: support@sstmed.com WWW: http://www.sstmed.com Page 1 TABLE OF

More information

Topcon Medical Systems

Topcon Medical Systems Topcon Medical Systems EZ Lite 2 Version 1.2 DICOM Conformance Statement November 1, 2013 Rev 1.1 Topcon Medical Systems, Inc. 111 Bauer Drive, Oakland, NJ 07436, USA Phone: (201) 599-5100 Fax: (201) 599-5250

More information

DICOM Conformance Statement. EasyWeb 3.0. Document Number XZR January Copyright Philips Medical Systems Nederland B.V.

DICOM Conformance Statement. EasyWeb 3.0. Document Number XZR January Copyright Philips Medical Systems Nederland B.V. Philips Medical Systems DICOM Conformance Statement EasyWeb 3.0 Document Number XZR 951-000010.00 20 January 2000 Copyright Philips Medical Systems Nederland B.V. 2000 Philips Medical Systems apple PHILIPS

More information

Artwork consists of sixty-five (65) 8 ½ inch x 11 inch pages.

Artwork consists of sixty-five (65) 8 ½ inch x 11 inch pages. Artwork consists of sixty-five (65) 8 ½ inch x 11 inch pages. REV AUTHORED BY DATE P.KUPOVICH 2/14/08 REV DRAFTED BY DATE G CORRIDORI 2/27/08 PROPRIETARY: This document contains proprietary data of Hologic,

More information

A Web PACS Architecture Based on WADO Service of DICOM Standard

A Web PACS Architecture Based on WADO Service of DICOM Standard A Web PACS Architecture Based on WADO Service of DICOM Standard G. KOUTELAKIS, G. TRIANTAFYLLOU, G. MANDELLOS, M. KOUKIAS, D. LYMPEROPOULOS Wire Communication Lab., Electrical & Computer Engineering Dept.,

More information

ROUTER. taking teleradiology to the next level

ROUTER. taking teleradiology to the next level ROUTER Image compression, teleradiology and workflow management taking teleradiology to the next level VERSiON 2.1 iq-router iq-router accelerates the transmission of any medical image through any network

More information

ModLink DICOM Conformance Statement

ModLink DICOM Conformance Statement ModLink DICOM Conformance Statement May 18 th 2015 LM-ENG-ML-DCS-REVA Page 1 of 12 - Proprietary Copyright 2015 - All rights reserved. Any comments or questions regarding the contents of this document

More information

ACR Accreditation Image Upload Instructions

ACR Accreditation Image Upload Instructions ACR Accreditation Image Upload Instructions Contents I. Introduction... 2 II. Your Application... 2 III. Your Testing Package... 3 A. ACRedit Web Client... 4 B. TRIAD Windows Client... 7 Step 1: Select

More information

HCP DICOM Net DICOM Conformance Statement

HCP DICOM Net DICOM Conformance Statement HCP DICOM Net DICOM Conformance Statement Version 4.00.00 32-bit version for Windows NT/2000/XP Copyright - 1997-2004 SoftLink International Pvt. Ltd. Microsoft, Windows NT, 2000,XP are trademarks of Microsoft

More information

Running and maintaining a secure Unity RIS/CVIS/PACS

Running and maintaining a secure Unity RIS/CVIS/PACS TechNote: Unity System Backend for System Administrators Running and maintaining a secure Unity RIS/CVIS/PACS Unity Releases 9, 10, and 11 April 3, 2015 For system administrators Contents Getting started...

More information

DICOM Conformance Statement

DICOM Conformance Statement Group of Companies ifa systems AG ifa united i-tech, Inc. Augustinusstr. 11b 1850 SE 17 th Street Suite C Suite 107 50226 Frechen Fort Lauderdale, FL 33316 Germany USA Phone: +49 2234 93367-0 Phone: +1

More information

Digital Imaging COmmunications in Medicine. DICOM in a Nutshell

Digital Imaging COmmunications in Medicine. DICOM in a Nutshell Digital Imaging COmmunications in Medicine DICOM in a Nutshell Ben Gorissen Philips Medical Systems Nederland B.V. ICS - ARC 1 /Ben Gorissen/ICS-ARC/March 1997/XPB080-970021.01 Presentation Overview Scope

More information

Kodak Point of Care CR systems. DICOM Conformance Statement

Kodak Point of Care CR systems. DICOM Conformance Statement Kodak Point of Care CR systems DICOM Conformance Statement Kodak QC Software Version 2.1.x Aug 21, 2005 Document #AT001008-00 Revision 1.0 Table of Contents REVISION HISTORY 4 ACRONYMS, ABBREVIATIONS AND

More information

ACR Accreditation User Instructions for Electronic Submission of Images

ACR Accreditation User Instructions for Electronic Submission of Images ACR Accreditation User Instructions for Electronic Submission of Images Page 1 of 25 Contents I. Introduction... 3 II. Your Application... 3 III. Your Testing Package... 5 A. ACRedit Web Client... 7 B.

More information

Test Content Outline. Approved Date: April 2016 Implementation Date: March 2017

Test Content Outline. Approved Date: April 2016 Implementation Date: March 2017 Test Content Outline Approved Date: April 2016 Implementation Date: March 2017 A. Procurement 5% B. Project Management 5% C. Operations 10% D. Communications 10% E. Training and Education 5% F. Image Management

More information

SecurView RT/TechMate Workstation DiCom Conformance statement

SecurView RT/TechMate Workstation DiCom Conformance statement breast imaging solutions DICOM Standard Interface SecurView RT/TechMate Workstation DiCom Conformance statement Software Version 8.4 man-03818 revision 002 SecurView RT/TechMate 8.4 Workstation DICOM

More information

Media Writer. Release Notes LX-DOC-MW5.1.9-RN-EN-REVB. Version 5.1.9

Media Writer. Release Notes LX-DOC-MW5.1.9-RN-EN-REVB. Version 5.1.9 Media Writer Release Notes Version 5.1.9 Regulations and Compliance Tel: 1-844-535-1404 Email: TS_PACSGEAR@hyland.com 2018 Hyland. Hyland and the Hyland logo are trademarks of Hyland LLC, registered in

More information

Artwork consists of twenty three (23) 8½ x 11 inch pages.

Artwork consists of twenty three (23) 8½ x 11 inch pages. Artwork consists of twenty three (23) 8½ x 11 inch pages. REV AUTHORED BY DATE P KUPOVICH 9/17/04 REV DRAFTED BY DATE G CORRIDORI 9/17/04 PROPRIETARY: This document contains proprietary data of Hologic,

More information

Summary DICONDE Opportunities, Advantages & Benefits

Summary DICONDE Opportunities, Advantages & Benefits Summary DICONDE Opportunities, Advantages & Benefits 1 BACKGROUND During the digital revolution in the medical industry, a single imaging standard, DICOM, emerged from the community. Digital Imaging and

More information

DICOMReader READING PORTABLE DICOM MEDIA INTO ANY PACS FASTER... Working TimeS. VERSiON 5.0

DICOMReader READING PORTABLE DICOM MEDIA INTO ANY PACS FASTER... Working TimeS. VERSiON 5.0 DICOMReader READING PORTABLE DICOM MEDIA INTO ANY PACS Working TimeS FAsteR 9than V4.0 fast... FASTER... DICOMReader VERSiON 5.0 DICOMReader Achieving THE highest compatibility DICOMReader can automatically

More information

SIEMENS. DICOM Conformance Statement

SIEMENS. DICOM Conformance Statement SIEMENS Siemens Medical Solutions USA, SY CARD, Ann Arbor DICOM Conformance Statement Revision: 9.0 11-12-10 Table of Contents 1. Introduction... 4 1.1 Purpose of this Document... 4 1.2 Sources for this

More information

TOSHIBA AMERICA MEDICAL SYSTEMS

TOSHIBA AMERICA MEDICAL SYSTEMS TOSHIBA AMERICA MEDICAL SYSTEMS ULTRASOUND DICOM IMAGE MANAGER Conformance Statement Revision: 1.6 Revision: 1.6 Date: 03/08/99 Reviewed by: Desmond Hirson Approved by: Thomas Little Revision History Revision

More information

DE32-DCM DentalEye 3.2. DICOM conformance statement

DE32-DCM DentalEye 3.2. DICOM conformance statement DentalEye 3.2 Table of contents 1. INTRODUCTION... 4 1.1. DentalEye Enterprise Application... 4 1.2. About this Document... 4 1.3. Integration and Validation... 4 1.4. Future Evolution... 5 1.5. Sources

More information

Philips Medical Systems DICOM Conformance Statement USIT 1.5 L3

Philips Medical Systems DICOM Conformance Statement USIT 1.5 L3 Philips Medical Systems DICOM Conformance Statement USIT 1.5 L3 Document Number 4512 131 81001 19 February 2004 Copyright Philips Medical Systems Nederland B.V. 2004 Page ii DICOM Conformance Statement

More information

PACS QUICK REFERENCE GUIDE

PACS QUICK REFERENCE GUIDE Version: 7.6.2.22 QUICK REFERENCE ADMIN/RADIOLOGIST PACS QUICK REFERENCE GUIDE Critical Findings The purpose of this document is to instruct Radiologists how to mark images with any critical finding or

More information

CHILI CHILI PACS. Digital Radiology. DICOM Conformance Statement CHILI GmbH, Heidelberg

CHILI CHILI PACS. Digital Radiology. DICOM Conformance Statement CHILI GmbH, Heidelberg CHILI Digital Radiology CHILI PACS DICOM Conformance Statement 2010-10 2013 CHILI GmbH, Heidelberg Manufacturer CHILI GmbH Friedrich-Ebert-Str. 2 D-69221 Dossenheim/Heidelberg Tel. (+49) 6221-18079-10

More information

IHE Endoscopy Technical Framework Supplement. Endoscopy Image Archiving (EIA) Rev. 1.1 Trial Implementation

IHE Endoscopy Technical Framework Supplement. Endoscopy Image Archiving (EIA) Rev. 1.1 Trial Implementation Integrating the Healthcare Enterprise 5 IHE Endoscopy Technical Framework Supplement 10 Endoscopy Image Archiving (EIA) 15 Rev. 1.1 Trial Implementation 20 Date: November 28, 2018 Author: IHE Endoscopy

More information

Caseaccess 1.0 DICOM Conformance Statement

Caseaccess 1.0 DICOM Conformance Statement Caseaccess 1.0 DICOM Conformance Statement 1 Introduction 1.1 Integration and Features The integration of medical devices may require functions that are beyond the scope of the DICOM standard. Consequently,

More information

WorkstationOne. - a diagnostic breast imaging workstation. User Training Presentation. doc #24 (v2.0) Let MammoOne Assist You in Digital Mammography

WorkstationOne. - a diagnostic breast imaging workstation. User Training Presentation. doc #24 (v2.0) Let MammoOne Assist You in Digital Mammography WorkstationOne - a diagnostic breast imaging workstation User Training Presentation WorkstationOne Unique streamlined workflow for efficient digital mammography reading Let MammoOne Let MammoOne Assist

More information

digital imaging A world of dedicated to animal healthcare Fast, affordable, easy-to-use solutions that offer high-quality digital imaging

digital imaging A world of dedicated to animal healthcare Fast, affordable, easy-to-use solutions that offer high-quality digital imaging Digital radiography small animals A world of digital imaging dedicated to animal healthcare Fast, affordable, easy-to-use solutions that offer high-quality digital imaging 2 Digital radiography small animals

More information

Matrix Server and Client DICOM Conformance Statement

Matrix Server and Client DICOM Conformance Statement Matrix Server and Client DICOM Conformance Statement Revision: 3.0 October 2005 All information contained in this document is accurate to the best knowledge of Merge emed as of date of publication and

More information

NumaStore Preclinical FAQ

NumaStore Preclinical FAQ NumaStore Preclinical FAQ 1. What is NumaStore Preclinical? 2. How does NumaStore Preclinical work with Inveon? 3. What data types does NumaStore Preclinical support? 4. How much storage space do I need

More information

nstream Version 3.1 DICOM Conformance Statement

nstream Version 3.1 DICOM Conformance Statement Image Stream Medical nstream Version 3.1 DICOM Conformance Statement Revision History: Revision Date Author Notes A 03/02/2006 TMT/EP Initial Control A 03/05/2006 TMT Minor cosmetic changes A 03/07/2006

More information

Philips Medical Systems DICOM Conformance Statement USIT 1.5

Philips Medical Systems DICOM Conformance Statement USIT 1.5 Philips Medical Systems DICOM Conformance Statement USIT 1.5 Document Number 4512 131 81001 19 April 2001 Copyright Philips Medical Systems Nederland B.V. 2001 Page ii DICOM Conformance Statement 4512

More information

DICOMReader READING PORTABLE DICOM MEDIA INTO ANY PACS FASTER... Working TimeS. VERSiON 5.0

DICOMReader READING PORTABLE DICOM MEDIA INTO ANY PACS FASTER... Working TimeS. VERSiON 5.0 DICOMReader READING PORTABLE DICOM MEDIA INTO ANY PACS Working TimeS FAsteR 9than V4.0 fast... FASTER... DICOMReader VERSiON 5.0 DICOMReader Achieving THE highest compatibility DICOMReader can automatically

More information

Uscan. DICOM Conformance Statement

Uscan. DICOM Conformance Statement DICOM Conformance Statement - Uscan Uscan DICOM Conformance Statement Software Version: 4.1.1 Date: 3 August 2018 Signostics, Inc., a subsidiary of EchoNous, Inc. 8310 154th Ave NE, Suite 200 Redmond,

More information

X-Porte DICOM Conformance Statement

X-Porte DICOM Conformance Statement 21919 30th Dr. SE, Bothell, WA 98021-3904 USA Telephone 1.425.951.1200 Facsimile 1.425.951.1201 www.sonosite.com Document Number: Revision: Title: D10711 D X-Porte DICOM Conformance Statement CHANGE HISTORY:

More information

Management of Acquisition Protocols (MAP) IHE Radiology Domain Kevin O Donnell, Toshiba Medical

Management of Acquisition Protocols (MAP) IHE Radiology Domain Kevin O Donnell, Toshiba Medical Management of Acquisition Protocols (MAP) IHE Radiology Domain Kevin O Donnell, Toshiba Medical Motivation One Scanner can have 100 or more protocols Quality of Care Good, consistent image quality depends

More information

SecurView DX Workstation DICOM Conformance Statement

SecurView DX Workstation DICOM Conformance Statement breast imaging solutions DICOM Standard Interface SecurView DX Workstation DICOM Conformance Statement Software Version 8.2 MAN-03404 Revision 001 SecurView DX 8.2 Workstation DICOM Conformance Statement

More information

New Enhanced Multi-frame DICOM CT and MR Objects to Enhance Performance and Image Processing on PACS and Workstations

New Enhanced Multi-frame DICOM CT and MR Objects to Enhance Performance and Image Processing on PACS and Workstations New Enhanced Multi-frame DICOM CT and MR Objects to Enhance Performance and Image Processing on PACS and Workstations SCAR 2004 Hot Topics - 22 May 2004 David Clunie, RadPharm Charles Parisot,, GE Healthcare

More information

PACS: Image Distribution

PACS: Image Distribution PACS: Image Distribution Henri Rik Primo Director, Marketing Siemens Medical Solutions USA, Inc. Image Distribution: Why So..the Radiology Department is ready to become filmless, except maybe for mammography

More information

Philips Medical Systems. Xcelera R1.1L1. Document Number XPR July 2003

Philips Medical Systems. Xcelera R1.1L1. Document Number XPR July 2003 Philips Medical Systems DICOM CONFORMANCE STATEMENT Xcelera R1.1L1 Document Number XPR 080-030073 11 July 2003 Copyright Philips Medical Systems Nederland B.V. 2003 All rights reserved 3 DICOM Conformance

More information

Note: The abbreviation e.g. is used in this document to indicate examples of the topics covered, but the examples are not a complete list.

Note: The abbreviation e.g. is used in this document to indicate examples of the topics covered, but the examples are not a complete list. Test Content Outline Approved Date: June 2018 Implementation Date: March 2019 Number of Scored Questions 1 A. Procurement 6 B. Project Management 7 C. Operations 13 D. Communications 10 E. Training and

More information

AGFA HEALTHCARE DICOM Conformance Statement

AGFA HEALTHCARE DICOM Conformance Statement HE/001512 Page 1 of 124 AGFA HEALTHCARE DICOM Conformance Statement Agfa HealthCare Enterprise Imaging 8.0.x Document No. 001512 Revision: 13 When printed, this is NOT a controlled copy. HE/001512 Page

More information

StellarPACS DICOM Conformance Statement. Version 1.3, August 2008 SoftTeam Solutions

StellarPACS DICOM Conformance Statement. Version 1.3, August 2008 SoftTeam Solutions StellarPACS DICOM Conformance Statement Version 1.3, August 2008 SoftTeam Solutions www.softteam.com Table of Contents 1 INTRODUCTION... 3 1.1 REFERENCE... 3 1.2 DEFINITIONS... 3 1.3 ACRONYMS, ABBREVIATIONS

More information

Technical Publications

Technical Publications g GE Healthcare Technical Publications Direction DOC1629000 Revision 2 LOGIQ P9/P7 R2.x Copyright 2016 by General Electric Co. g GE Healthcare Do not duplicate THIS PAGE LEFT INTENTIONALLY BLANK CONFORMANCE

More information

Application Launcher 2.2 DICOM Conformance Statement

Application Launcher 2.2 DICOM Conformance Statement Contents mm761-0 Application Launcher 2.2 DICOM Conformance Statement 1 Introduction... 3 1.1 Integration and Features... 3 1.2 Definitions... 3 2 NETWORKING... 4 2.1 Implementation Model... 4 2.1.1 Application

More information

DIRECT RADIOGRAPHY. flexible. Delivering. imaging excellence. DX-D Direct Radiography from Agfa HealthCare

DIRECT RADIOGRAPHY. flexible. Delivering. imaging excellence. DX-D Direct Radiography from Agfa HealthCare DIRECT RADIOGRAPHY flexible Delivering imaging excellence DX-D 300 - Direct Radiography from Agfa HealthCare versatile a and compact solution for all General Radiographic exams Maximum flexibility. The

More information

Using Pinnacle 16 Deformable Image registration in a re-treat scenario

Using Pinnacle 16 Deformable Image registration in a re-treat scenario Introduction Using Pinnacle 16 Deformable Image registration in a re-treat scenario This short Hands On exercise will introduce how the Deformable Image Registration (DIR) tools in Pinnacle can be used

More information

DICOM Conformance Statement. cvi 42. Version 5.6. Workstation for Cardiovascular MR and CT

DICOM Conformance Statement. cvi 42. Version 5.6. Workstation for Cardiovascular MR and CT DICOM Conformance Statement cvi 42 Version 5.6 Workstation for Cardiovascular MR and CT 2017-02-24 CIRCLE Cardiovascular Imaging Inc Panarctic Plaza Suite 250, 815 8 th Avenue SW Calgary, Alberta, Canada

More information

Review of WG17 Work Item 1 Output: DICOM Encapsulated STL. Allan Noordvyk & Justin Ryan Co-Chairs of WG17: 3D Manufacturing

Review of WG17 Work Item 1 Output: DICOM Encapsulated STL. Allan Noordvyk & Justin Ryan Co-Chairs of WG17: 3D Manufacturing Review of WG17 Work Item 1 Output: DICOM Encapsulated STL Allan Noordvyk & Justin Ryan Co-Chairs of WG17: 3D Manufacturing Agenda WG17 Mandate & Work Item Specifics of Proposal New IOD / SOP Other related

More information

7/30/2018. QC Considerations for Mobile Devices. Outline. No disclosures.

7/30/2018. QC Considerations for Mobile Devices. Outline. No disclosures. QC Considerations for Mobile Devices Alisa Walz-Flannigan, PhD (DABR) Mayo Clinic, Rochester, Minnesota AAPM Annual Meeting July 30, 2018 2017 MFMER slide-1 No disclosures. Any commercial products referenced

More information

Version 7 DICOM Conformance Statement. Document Version 3.02, June 2009

Version 7 DICOM Conformance Statement. Document Version 3.02, June 2009 Version 7 DICOM Conformance Statement Document Version 3.02, June 2009 1 Conformance Statement Overview The application described in this Conformance Statement VEPRO EMR Manager PACS is a collection of

More information

Philips Medical Systems DICOM Conformance Statement. Inturis Suite R2.2

Philips Medical Systems DICOM Conformance Statement. Inturis Suite R2.2 Philips Medical Systems DICOM Conformance Statement Inturis Suite R2.2 Document Number XZR 080-010044 8 February 2002 Copyright Philips Medical Systems Nederland B.V. 2002 Page ii DICOM Conformance Statement

More information