1 BROY LOGMASTER LM750 TABLE OF CONTENTS 1.0 INTRODUCTION Data Presentation Software LM750 OPERATION Description of Recording Methods Range and Power Switches Installing and Removing the LM Using the Display Using the Computer Interface Adapter Transferring and Processing LM750 Data FIELD INITIALIZATION Summary Initialization Procedure Setting Recording Interval and Mode Settin g Elapsed Time Setting Data Mode Clearing Data Memory Checking Paramet er Settings BATTERY MAINTENANCE TROUBLESHOOTING Hi gh or Low Readings Ra ndom Readings Technical Assistance LM750 SPECIFICATIONS...7 Instruction Manual 10/30/02 2

2 1.0 INTRODUCTION LM750 RECORDING VOLTMETER W ITH DISPLAY The Broy Model LM750 is an electronic battery-powered instrument used to measure and record voltages up to 750 Vac. The voltage is digitized and recorded by the LM750. The recorded data is downloaded into a computer and may then be processed, viewed and plotted for purposes such as feeder regulation studies, planning, monitoring, or troubleshooting. Contacts: Logmaste r@broy.com Toll free Fax Web LM750 OPERATION 2.1 Description of Recording Methods 1.1 Energraf EG2000V Data Presentation Software EG2000V is available FREE on request to users of the LM750 recording voltmeter. Please or fax your name, location, and shipping address to us, and we will promptly send a free copy of EG2000V on a CD-ROM. The LM750 samples the voltage applied to its inputs once per second as timed by an internal clock. These samples are used to calculate the average, maximum, and minimum voltage as required over the duration of a user-selected time period. One of three RECORDING INTERVAL time periods can be selected: 1, 5 or 15 minutes. The information that is stored in memory at the end of each interval depends on which Data Mode is selected: AVERAGE or AVERAGE/MIN/MAX. In AVERAGE Data Mode the stored data is the average of the one-second samples taken during the interval. In AVERAGE/MIN/MAX Data Mode the maximum and the minimum of the samples as well as the average is stored. The data is stored in memory in one of two mann ers: CONTINUOUS or STOP. In CONTINUOUS Memory Mode when the memory is filled newly recorded data displaces e oldest. When the data is transferred to the computer, the recorder's entire memory contents are transferred. In STOP Memory Mode the LM750 stores data until the ime in the recorder is preserved. The time, date and data remain until the memory contents are transferred to a computer and the recorder is reinitialized. Only the data that has been recorded since the previous initialization will be transferred resulting in shorter data transfers and smaller data files than for Cont inuous Mode recording. The computer clock will determine the time and date to be used during initialization. Memory mode, recording interval duration, and data modes may be selected when the LM750 is initialized from the computer software, or later in the field, by using the front panel switches (see 2.4). It is important to clear the memory before using the recorder; this is done automatically during computer Initialization, or manually at turn-on in the field, by using the Program and DISPLAY buttons. The operational modes may be checked at any time using the DISPLAY button (see 2.4). 2.2 Range and Power Switches These switches are located under the small plastic plug and operated by sliding the small white tabs left or right to set. To use the LM750, set the power switch ON, select the range wanted, connect the LMI-1 Interface Adapter (sect 2.5), and run the Initiali zation step from the computer software which you are using. When the instrument is first turned on, the memory will contain random data (garbage); this must be cleared before using (section 3.2.5). When swi tched ON the instrument defaults to the following Modes which may be displayed by pressing the DISPLAY button. Default Mode s: E00 Elapsed time zero P15 Recording interval fifteen minute, continuous HL Data mode Min/Max/Avg Memory status not cleared 2.3 Installing and Removing the LM750 Only qualified persons should install and remove the LM750 on electrical circuits. When working on live equipment normal safety precautions are to be taken that are appropriate for the voltage level involved. Two sets of test leads with safety shrouded connectors are supplied. One set is a clip-on high voltage type, with in-line High Rupturing Capacity (HRC) fuses. The other consists of a standard 120V line cord. Repla cement leads are available from Broy Engineering. ALWAYS INSTALL THE LEADS IN THE LM750 BEFO RE CONNECTING TO THE LOAD VOLTAGE As part of the installation and removal procedure, the location and serial number of the LM750 as well as the installation and removal times should be noted since these may be useful during data processing. date and emory is full. When all the memory is filled no more data is stored and the 2.4 Using the Display The LM750 has a 4-digit display which can be activated by pressing the membrane switch area on the front panel la belled "DISPLAY". If the DISPLAY button has not 1 10/30/02 2

3 previously been pressed for at least 16 seconds, the recorder display will step through the present setup in the following order and format: ELAPSED TIME AND MEMORY STATUS Enn. - nn minutes into programmed interval memory full Enn. " memory partial RECORDING INTERVAL AND RECORDING MODE Pmm. - mm minutes recording interval (01,05,15) stop mode Pmm. " continuous mode DATA MODE -- mode Average HL " High/ Low/aver age After these three consecutive indications the LM750 will display the present voltage input. 2.5 Using the LMI-1 Desktop Computer Interface The LMI-1 Interface Adapter has a 9-socket pin (Female) D-type connecter, which should be conne cted to serial port COM1 or COM2 of a DOS-type computer. If necessary, a 9 pin to 25 pin adapter may be used. Power is supplied to the interface adapter from the power adapter supplied with the interface. Connection to the LM750 instrument is through the 6-pin circular connector at the end of the case by removing the large red plastic plug and inserting the interface connector. The FUNCTION switch on the interface should be at "STANDBY" when t he conne ction is made. The LO BATTERY indicator light will turn OFF if the battery in the LM750 is satisfactory for further use. LMI-2 Portable Battery Powered Computer Interface The LMI-2 is powered by an internal 9V battery. Connecting the round connector to your LM750 will activate data transfer and provide the recorder battery check. The battery LED will light if the reorders battery is ok for further use. Your software may refer to a Standby/Transfer switch, the LMI-2 does not have one. Disconnected, the LMI-2 is in Standby, upon connection, data transfer will begin and the transfer LED will light. 2.6 Transferring and Processing LM750 Data Data may be output fr om the LM750 is down loaded to a comput er by means of the LMI- 1 adapter. The output, coded in ASCII, consists of a header containing the serial number (the number stamped on the nameplate) and the recorder's time and date followed by the recorded voltage data with newest data first. Data fil es are stored on the hard drive under file names chosen by the user; the default file extension is.dat. 3.0 FIELD INITIALIZATION 3.1 Summary The LM750 is initialized at the start of a recording session by either the membrane switches on the front panel or by using the computer. The LM750 has four parameters that can be s et in the field as wel l as the capa bility of zeroing the da ta memory : * memory mode * recording interval * data mode * el aps ed t ime (pa nel swi tches only) These parameters should not be changed until the recorded data has been transferred to the computer. Available are two memory Modes, Stop or Continuous, three Recording Intervals of 1, 5, or 15 minutes, and two Data Modes, Average or Average/Minimum/Maximum Initialization Procedure THE "PROGRAM" BUTTON MUST BE HELD DOWN THROUGHOUT THESE PROCEDURES. If it is released, the procedure will be terminated. To stop the programming process after any step, simply release the PROGRAM button. A description of all displayed symbols may be found in Section Setting Recording Interval and Mode To program the recording interval and mode, hold the PROGRAM button, press and release the DISPLAY button once. The display will show "888.8" indefinitely. Press and hold the DISPLAY button a second time; the present setting will be displayed for one second. After one second, the recorder will cycle through the available modes, (P01. P05. P15. P01.- P05.- P15.-). To select a mode, release the DISP LAY button when the desired mode is shown. If necessary, continue t o hold the DISPLAY button; the LM750 continues to cycle through the possible modes until DISPLAY is released. The selected mode will be displayed as long as the PROGRAM button is held. LMI-1 Computer Interface LMI-2 Computer Interface Setting Elapsed Time When the instrument is switched ON the internal clock will start up with a time of 00:00. In the one minute interval mode, Elapsed Time is not applicable and the value will be zero; the setti ng option will not be available. When using 5 or 15 minute interval mode, 3 10/30/02 4

4 the clock time may be adjusted so that so that data recording takes place on "even" times of 5 or 15 minutes in real time; for example, 10:05 or 13:45. This is done by telling the clock the number of minutes passed since the last standard recording time by entering what is called Elapsed Time. For example, while using a recordin g interval of 15 minutes, with an actual startup time of 10:06, 6 minutes have passed since the last standard recording time and the Elapsed Time will thus be 6 minutes. It should be noted that only the minutes will be affected. The Elapsed Time value may be set using the buttons on the front panel as follows. Press and hold the PROGRAM button; at the same time press and release the DISPLAY button; the display will show "888.8". Now press and release the DISPLAY button a second time, the selected recording interval and mode will be displayed. Press and hold the DISPLAY button again and the existing elapsed time setting will be displayed; the display will then begin cycling through all the possible values of E00 through E05 or E15, depending on the recording interval. When the desired value is shown, release the DISP LAY button and the selected Elapsed Time will be set to the last value shown. To use the default setting of zero minutes do not hold the DISPLAY button; release it immediately Setting Data Mode To program DATA MODE; hold the PROGRAM button, press and release the DISPLAY button three times. Now press and hold the DISPLAY button a fourth time, the display will show t he current setting either "--." or "HL.". The display will then begin to alternate between the two modes. Release the DISPLAY button when the des ired mode is shown, the se lected mode w ill then remai n on the di spl ay Clearing Data Memory To zero the data memory; hold the PROGRAM button, press the DISPLAY button four times. Now press and hold the DISPLAY button a fifth time, the display will show "---.- ". Continue holding the DISPLAY button until the display shows "222.2". The memory has now been zeroed. The recorder will then continue into a parameter check/display mode Checking Parameter Settings At lower temperatures, battery life is reduced. The useful life at -20C will be about 7 weeks, with a reserve of about 2 weeks. Extremely low temperatures approaching -40C will have a life of about 4 to 5 weeks and it is recommended that new batteries be installed whenever recordings are to be made under these conditions. To install or change the batteries in the recorder, remove t he cover from the recorder and insert the new batteries into the holders provided. Replace the cover carefully so as not to pinch the wires. Following battery replacement or removal the recorders' date must be reset using the initialization procedure. 5.0 TROUBLESHO OTING 5.1 High or Low Readings Voltage readings of five times the expected amount. This may be corrected by making sure the range is set to 0-200V, pressing and holding the DISPLAY button until "00.0" is displayed, and repeating the data transfer (Sec 2.6). Voltage readings of one fifth the expected amount. This may be corrected by making sure the range is set to 0-750V, pressing and holding the DISPLAY button until "00." is displayed, and repeating the download. 5.2 Random Readings Downloads containing large numbers of random values, which may be very high in value, beyond the normal maximum voltage range. Probable cause is failing to clear the memory after swi tching the un it ON. Soft war e initialization cl ears the me mory aut omatically; field start-up requires manual clearing using the front panel switch. 5.3 Technical Assistance Technical assistance is available via telephone, , and our website: Logmaster@broy.com or To check the chosen parameter settings, press and release the DISPLAY button five times. The recorder will then step through the selected settings in the order: Pnn,Emm,HL, or Pnn,Emm,-- etc. as long as the PROGRAM button is held. 4.0 BATTERY MAINTENANCE The LM750 operates from two 9V alkaline batteries. At a temperature of 20C, new batteries should operate about 16 weeks before the Computer Interface LED indicates "lo bat tery" by lighting dur ing download. At this indication, the b atteries have ab out 21 d ays of reserve remaining. It is not necessary for the interface to be connected to the computer to test the batteries. Once "l o battery" is indicated, the ba tteries ha ve less than 21 d ays of life left, and should be replaced. Both batteries should be replaced together. 5 10/30/02 6

5 6.0 LM750 SPECIF ICATIONS Fre que ncy: Voltage Ranges: Hi: V Accuracy: Hi: (1% of rdg. +1) V Resolution: Hi: 1 V Maximum Voltage: Stored Data: Average Mode - Average/Minimum/Maximum Mode - Data Output: Data ports: Data Download time: Memory Modes: Stop - Records until memory full. Continuous Hz. Lo: V Lo: (1% of rdg. +0.2) V Lo: 0.2 V 750 Vac Average of all samples during the specified period based on one sample per second. 3 readings stored per period; average as above, the minimum and maximum voltage values which occurred during this period. Serial RS232, 9600 baud using the LMI-1 or 2 computer interface. COM1, COM2. 50 seconds or less. When memory filled, newer data displaces old. Dat a Ca pacity: Rec Int. Data M ode Days Storage 15 Avg. 66 * 5 " 22 1 " Avg/Min/Max 22 5 " " 1.5 * may be limited by battery life at low temperatures Batteries: (2) Standard 9V Alkaline. Battery Life: At 20C temperature, 16 weeks. Shorter life at lower temperatures. Timing: Quartz regulated; Maximum drift 1 minute over 2 weeks. Temperature Range: -40 C to +55 C Weight: 1.7 lb (0.8 Kg) Dimensions: 19 x 8.1 x 5.5 cm. Printer Ports: LPT1, LPT2. 7

Multi-Channel RS-232 Current Loop Interface Converter (CLIC)

Multi-Channel RS-232 Current Loop Interface Converter (CLIC) DECEMBER 1996 CL204A CL208A CL212A CL216A Multi-Channel RS-232 Current Loop Interface Converter (CLIC) CLIC 1/2A,SLO BLO,115VAC 1/4,SLO BLO,230VAC CUSTOMER SUPPORT INFORMATION Order toll-free in the U.S.:

More information

Universal AC Power Source + AC Power Analyzer

Universal AC Power Source + AC Power Analyzer User's Guide Universal AC Power Source + AC Power Analyzer Model 380820 Introduction Congratulations on your purchase of the Extech Model 380820. This Universal AC Power Source and AC Power Analyzer can

More information

CONTENTS 1. SAFETY INFORMATION SPECIFICATION PARTS & CONTROLS General Specification Electrical Specification...

CONTENTS 1. SAFETY INFORMATION SPECIFICATION PARTS & CONTROLS General Specification Electrical Specification... Title CONTENTS page 1. SAFETY INFORMATION... 1 2. SPECIFICATION... 2 2-1 General Specification... 2 2-2 Electrical Specification... 5 3. PARTS & CONTROLS... 6 3-1 The Name of Each Parts and Positions...

More information

Visit us at Back to the Extech Product Page

Visit us at   Back to the Extech Product Page 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech 407355 Product Page User's Guide Noise Dosimeter with PC Interface

More information

TS-ADA-029. Wireless Tone Generator. User's Manual / Installation Guide. Version 1.03

TS-ADA-029. Wireless Tone Generator. User's Manual / Installation Guide. Version 1.03 TS-ADA-029 Wireless Tone Generator User's Manual / Installation Guide Version 1.03 Visiplex, Inc. 2012 TS-ADA-029 Wireless Tone Generator Copyright The product described in this manual includes copyrighted

More information

VBT-75 S2 VACUUM BOTTLE TESTER

VBT-75 S2 VACUUM BOTTLE TESTER VBT-75 S2 VACUUM BOTTLE TESTER USER S MANUAL Vanguard Instruments Company, Inc. 1520 S. Hellman Ave. Ontario, California 91761, USA TEL: (909) 923-9390 FAX: (909) 923-9391 July 7, 2017 Revision 1 SAFETY

More information

Always make sure ALL power to the equipment you are working with is turned OFF before starting any maintenance.

Always make sure ALL power to the equipment you are working with is turned OFF before starting any maintenance. Thank you for purchasing the enclosed PMP Serial Communications Module. This module is designed and manufactured by PMP as a direct replacement for the Veeder-Root TLS-350 Serial RS-232 Module 847490-301

More information

RS Stock No Instruction Manual RS Input Data Logging Thermometer

RS Stock No Instruction Manual RS Input Data Logging Thermometer RS Stock No. 730-0458 Instruction Manual RS-1384 4 Input Data Logging Thermometer EN FR IT DE ES TABLE OF CONTENTS / EN TITLE TABLE OF CONTENTS PAGE 1. INTRODUCTION FEATURE... 1 2. SPECIFICATIONS... 2

More information

AELP PALLET BEAMS SCALES

AELP PALLET BEAMS SCALES Adam Equipment AELP PALLET BEAMS SCALES (P.N. 4327, Rev. A7, Mar 2016) Adam Equipment Company 2016 Adam Equipment Company 2016 CONTENTS 1.0 INTRODUCTION... 2 2.0 TECHNICAL SPECIFICATIONS... 3 3.0 UNPACKING

More information

HAND HELD PROGRAMMER QUICK START GUIDE

HAND HELD PROGRAMMER QUICK START GUIDE HAND HELD PROGRAMMER QUICK START GUIDE IMPORTANT INFORMATION 1) Do not leave the programmer connected to the PC adapter or a target system, as this will drain the battery. LOADING FROM THE PC 1) Connect

More information

WARNING: Do not use the thermometer/data logger before you read the users manual and the following instructions.

WARNING: Do not use the thermometer/data logger before you read the users manual and the following instructions. 55 This unit passes the following tests EN 61326-1:2006 (CISPR11,IEC/EN 61000-3-2:2006, IEC/EN 61000-3-3: 1995+A1 :2001+A2:2005 IEC/EN 61000-4-2/-3/-5/-6/-11) WARNING: Do not use the thermometer/data logger

More information

User Guide CPSMP VAC Power Supply Module: PointSystem CPSMC Accessory CPSMC Accessory. Contents.

User Guide CPSMP VAC Power Supply Module: PointSystem CPSMC Accessory CPSMC Accessory. Contents. User Guide CPSMP-205 110 240 VAC Power Supply Module: PointSystem CPSMC1800-200 Accessory CPSMC1900-100 Accessory Contents Contents...1 Description...1 Cautions and Warnings...2 Definitions...2 Power supply

More information

CORD-XL Dual-Channel Electronic Chart Recorder User s Manual

CORD-XL Dual-Channel Electronic Chart Recorder User s Manual CORD-XL Dual-Channel Electronic Chart Recorder User s Manual Rohrback Cosasco Systems Inc. 11841 E. Smith Ave Santa Fe Springs, CA 90670 Tel: (562) 949-0123 Fax: (562) 949-3065 P/N 720701-Manual Rev E

More information

Communi-Cell Communicator Nokia Version

Communi-Cell Communicator Nokia Version Communi-Cell Communicator Nokia Version Jan 2007 Instruction Manual 1 ABOUT THE COMMUNI-CELL COMMUNICATOR SYSTEM The Communi-Cell Communicator system is based on GSM SMS technology. It uses a standard

More information

CyberComm Pro Data Acquisition Software Installation & User Guide

CyberComm Pro Data Acquisition Software Installation & User Guide CyberComm Pro 2.2.3 Data Acquisition Software Installation & User Guide ph 1100 and ph 2100 Bench ph and Bench ph/ion Meter Technology Made Easy... 68X090822 rev 1 Aug 2002 2 PREFACE Thank you for selecting

More information

DPM-3221 Digital DC Power Meter with data logging capability.

DPM-3221 Digital DC Power Meter with data logging capability. Introduction DPM-3221 Digital DC Power Meter with data logging capability. User Manual As a digital DC power meter, it measures the real time DC Voltage (5-60V), DC Current (0-200 A with the correct shunt),

More information

Installation and Operation Back-UPS BR1000G-IN / BR1500G-IN

Installation and Operation Back-UPS BR1000G-IN / BR1500G-IN Installation and Operation Back-UPS BR1000G-IN / BR1500G-IN Important Safety Information Read the instructions carefully to become familiar with the equipment before trying to install, operate, service

More information

AVR Intermediate Development Board. Product Manual. Contents. 1) Overview 2) Features 3) Using the board 4) Troubleshooting and getting help

AVR Intermediate Development Board. Product Manual. Contents. 1) Overview 2) Features 3) Using the board 4) Troubleshooting and getting help AVR Intermediate Development Board Product Manual Contents 1) Overview 2) Features 3) Using the board 4) Troubleshooting and getting help 1. Overview 2. Features The board is built on a high quality FR-4(1.6

More information

CyberComm Pro 2.4 Data Acquisition Software Installation & User Guide. CyberScan DO 1500

CyberComm Pro 2.4 Data Acquisition Software Installation & User Guide. CyberScan DO 1500 CyberComm Pro 2.4 Data Acquisition Software Installation & User Guide CyberScan DO 1500 Bench Dissolved Oxygen Meter Technology Made Easy... 68X292341 Rev.0 01/04 PREFACE Thank you for selecting the CyberScan

More information

Copyright December 2017 Nidec Issue: D

Copyright December 2017 Nidec Issue: D General Information The manufacturer accepts no liability for any consequences resulting from inappropriate, negligent or incorrect installation or adjustment of the optional parameters of the equipment

More information

RS Stock No Instruction Manual RS Input Data Logging Thermometer

RS Stock No Instruction Manual RS Input Data Logging Thermometer RS Stock No. 730-0458 Instruction Manual RS-1384 4 Input Data Logging Thermometer EN FR IT DE ES TABLE OF CONTENTS / EN TITLE TABLE OF CONTENTS PAGE 1. INTRODUCTION FEATURE... 1 2. SPECIFICATIONS... 2

More information

EA-GSM-Interface-GB.odt

EA-GSM-Interface-GB.odt EA-GSM-Interface (1.8/1.82) 4/7/213 EA-GSM-Interface GSM-Gateway +41 ()56 648 4 4 / www.leitronic.ch 1/12 EA-GSM-Interface (1.8/1.82) 4/7/213 Contents 1 Overview... 3 2 Specification... 4 2.1 Connectors...

More information

ME6008M Batch Monitor Printer Driver Unit

ME6008M Batch Monitor Printer Driver Unit ME6008M Batch Monitor Printer Driver Unit FEATURES Prints QA batch totals automatically (programmable delay), or manually. Internally logs 00 batch events, downloadable to your laptop/pc via optional front-access

More information

T3 Interface Converters

T3 Interface Converters MT1103A JULY 2003 MT1103A-DC T3 Interface Converters CUSTOMER SUPPORT INFORMATION Order toll-free in the U.S.: Call 877-877-BBOX (outside U.S. call 724-746-5500) FREE technical support 24 hours a day,

More information

STYLITIS-POWER OPERATOR S GUIDE

STYLITIS-POWER OPERATOR S GUIDE STYLITIS-POWER OPERATOR S GUIDE Copyright 2008-2016, SYMMETRON Electronic Applications. Second edition in English. April 2016. No part of this publication may be reproduced, stored in retrieval system,

More information

POWER AVAILABILITY 2U POD USER MANUAL. Power Output Distribution 230 Volt

POWER AVAILABILITY 2U POD USER MANUAL. Power Output Distribution 230 Volt POWER AVAILABILITY 2U POD USER MANUAL Power Output Distribution 230 Volt TABLE OF CONTENTS IMPORTANT SAFETY INSTRUCTIONS.......................................... 1 GLOSSARY OF SYMBOLS..................................................

More information

SY021 Portable Load/Force Meter User instructions

SY021 Portable Load/Force Meter User instructions SY021 Portable Load/Force Meter User instructions Relates to firmware version 5.2 INTRODUCTION The SY021 is a portable load meter, which can indicate the load present on any attached cell. A dual channel

More information

Dual channel temperature logger with two binary inputs and LCD display Instruction Manual

Dual channel temperature logger with two binary inputs and LCD display Instruction Manual LOGGER S0841 Dual channel temperature logger with two binary inputs and LCD display Instruction Manual Instruction Manual for use of S0841 logger Instrument is designed for measurement and record of temperature

More information

Instruction Manual. CCP Supply Type 12AL SOUND & VIBRATION. Skovlytoften 33, 2840 Holte, Denmark

Instruction Manual. CCP Supply Type 12AL SOUND & VIBRATION. Skovlytoften 33, 2840 Holte, Denmark Instruction Manual CCP Supply Type 12AL G. R. A. S. SOUND & VIBRATION Skovlytoften 33, 2840 Holte, Denmark www.gras.dk gras@gras.dk CCP Supply Type 12AL Acknowledgement: ICP is a registered trademark of

More information

5000V Digital/Analog Megohmmeter Model 5060

5000V Digital/Analog Megohmmeter Model 5060 5000V Digital/Analog Megohmmeter Model 5060 The Model 5050 and 5060 are the latest design in 5000V Megohmmeters, the most innovative products in their class on the market today. The features and functions

More information

Model R5005. Instruction Manual. True RMS Industrial Multimeter. reedinstruments. www. com

Model R5005. Instruction Manual. True RMS Industrial Multimeter. reedinstruments. www. com Model R5005 True RMS Industrial Multimeter Instruction Manual reedinstruments com Table of Contents Safety... 4 Features... 5 Specifications...5-6 Instrument Description...7-8 Operating Instructions...9-13

More information

Sound Analyser Real Time 1/1 & 1/3 Octave band Analysis

Sound Analyser Real Time 1/1 & 1/3 Octave band Analysis ISO-TECH Sound Analyser Real Time 1/1 & 1/3 Octave band Analysis SLA-1358 INSTRUCTION MANUAL L CONTENTS Title Page I. SAFETY INFORMATION...1 II. GENERAL INFORMATION & FEATURES...2 III. SPECIFICATIONS...2

More information

Simple Logger II Series

Simple Logger II Series Simple Logger II Series Quick Start Guide ENGLISH www.aemc.com CHAUVIN ARNOUX GROUP Statement of Compliance Chauvin Arnoux, Inc. d.b.a. AEMC Instruments certifies that this instrument has been calibrated

More information

Operation Manual. Imara S100 & S60 DMX. Imara S100 DMX Yoke Mount. Imara S100 DMX Pole-Op

Operation Manual. Imara S100 & S60 DMX. Imara S100 DMX Yoke Mount. Imara S100 DMX Pole-Op Operation Manual Imara S100 & S60 DMX Imara S100 DMX Yoke Mount Imara S100 DMX Pole-Op Part No. 3100083 Rev A 03-02-2015 Imara DMX Yoke Mount Imara Fixtures IMR-S100-120U Imara S100 DMX Yoke Mount, Univ

More information

Multi-IQ. Firmware: RS232-Data Manager for UPS 4 x RS232 / 1 x LAN-Contacts flash upgradeable. User manual. (Serial Number / Seriennummer)

Multi-IQ. Firmware: RS232-Data Manager for UPS 4 x RS232 / 1 x LAN-Contacts flash upgradeable. User manual. (Serial Number / Seriennummer) Version: 2016-08-18 Multi-IQ Firmware: (Serial Number / Seriennummer) RS232-Data Manager for UPS 4 x RS232 / 1 x LAN-Contacts flash upgradeable User manual MANUAL ENGLISH... 2 GENERAL PROPERTIES:... 2

More information

SWC1090 configuration software for G.M. International Isolators series D1000 and E1000. Installation and use manual

SWC1090 configuration software for G.M. International Isolators series D1000 and E1000. Installation and use manual SWC1090 configuration software for G.M. International Isolators series D1000 and E1000 Installation and use manual ISM0084-2 Installation and use of the SWC1090 calibration software Page 1 of 12 Index

More information

8051 Intermidiate Development Board. Product Manual. Contents. 1) Overview 2) Features 3) Using the board 4) Troubleshooting and getting help

8051 Intermidiate Development Board. Product Manual. Contents. 1) Overview 2) Features 3) Using the board 4) Troubleshooting and getting help 8051 Intermidiate Development Board Product Manual Contents 1) Overview 2) Features 3) Using the board 4) Troubleshooting and getting help 1. Overview 2. Features The board is built on a high quality FR-4(1.6

More information

OVEN INDUSTRIES, INC.

OVEN INDUSTRIES, INC. OVEN INDUSTRIES, INC. OPERATING MANUAL MODEL 5C7-195 THERMOELECTRIC MODULE TEMPERATURE CONTROLLER TABLE OF CONTENTS Page Features...1 Description...2 Block Diagram...2 Mechanical Package Drawing...3 RS232

More information

Getting started with your DGH module:

Getting started with your DGH module: Getting started with your DGH module: This document contains step-by-step instructions to quickly connect and communicate with your new DGH modules. For most applications, new modules require a one-time

More information

AutoMax Network Communication Option Board for use with GV3000/SE AC Drives M/N 2AX3000

AutoMax Network Communication Option Board for use with GV3000/SE AC Drives M/N 2AX3000 AutoMax Network Communication Option Board for use with GV3000/SE AC Drives M/N 2AX3000 Instruction Manual D2-3308-7 The information in this manual is subject to change without notice. Throughout this

More information

User's Guide. Model W 3-in-1 Switching DC Power Supply

User's Guide. Model W 3-in-1 Switching DC Power Supply User's Guide Model 382260 80W 3-in-1 Switching DC Power Supply Introduction Congratulations on your purchase of the Extech 80W 3-in-1 Switching DC Power Supply. The Model 382260 has three output ranges

More information

DEFECTOMETER High detection sensitivity to surface flaws (even underneath paint) Simple, microprocessor-controlled operation

DEFECTOMETER High detection sensitivity to surface flaws (even underneath paint) Simple, microprocessor-controlled operation DEFECTOMETER 2.837 High detection sensitivity to surface flaws (even underneath paint) Simple, microprocessor-controlled operation Automatic lift-off and zero compensation Several alternatives for signal

More information

Multi-Function Smart Indicator User Manual

Multi-Function Smart Indicator User Manual Multi-Function Smart Indicator User Manual EC30006 Content 1 Notice... 2 2 Specification... 2 3 Main parameter... 3 4 How to fix... 3 5 Port... 4 6 On/Off... 4 7 Charging... 4 8 Display... 5 9 Screen keypad...

More information

AppleCD 300e Plus/600e

AppleCD 300e Plus/600e K Service Source AppleCD 300e Plus/600e AppleCD 300e Plus AppleCD 600e K Service Source Specifications AppleCD 300e Plus/600e Specifications Characteristics - 1 Characteristics 300e Access Time (Including

More information

I. SAFETY INFORMATION II. FEATURES III. GENERAL SPECIFICATIONS

I. SAFETY INFORMATION II. FEATURES III. GENERAL SPECIFICATIONS CONTENT Page I. SAFETY INFORMATION ------------------- 2 II. FEATURES ------------------------------------- 2 III. GENERAL SPECIFICATIONS ----------- 3 IV. CONTROLS ------------------------------------

More information

5000V Digital/Analog Megohmmeter. Models 5050 & 5060

5000V Digital/Analog Megohmmeter. Models 5050 & 5060 5000V Digital/Analog Megohmmeter The Model 5050 and 5060 are the latest design in 5000V Megohmmeters, the most innovative products in their class on the market today. The features and functions incorporated

More information

Talk-N-Tracef Communication and Tracing Set. User's Guide.

Talk-N-Tracef Communication and Tracing Set. User's Guide. Talk-N-Tracef Communication and Tracing Set User's Guide www.jdsu.com/know ON / RING BUTTON When unit is off, pressing this button turns on power to talk circuitry. When unit on, pressing this button

More information

OFFICE UPS MULTI-DEVICE PROTECTION UPS 500S/600S/750S

OFFICE UPS MULTI-DEVICE PROTECTION UPS 500S/600S/750S OFFICE UPS MULTI-DEVICE PROTECTION UPS 500S/600S/750S USER S MANUAL 1.Safety instructions Thank you for selecting this uninterrupted power source. It provides you with better protection for connected equipment.

More information

Dual channel temperature logger with display Instruction Manual

Dual channel temperature logger with display Instruction Manual LOGGER S0122 Dual channel temperature logger with display Instruction Manual ie-log-s0122-05 Instruction manual for use of temperature logger S0122 Logger is designed for measurement and record of ambient

More information

User's Guide. Model Noise Dosimeter with PC Interface. Introduction

User's Guide. Model Noise Dosimeter with PC Interface. Introduction User's Guide Noise Dosimeter with PC Interface Model 407355 Introduction Congratulations on your purchase of the Extech 407355 Personal Noise Dosimeter. The 407355 is designed to test noise exposure in

More information

Table 1. RS232 Serial Adapter DEBUG Connector Pin Descriptions

Table 1. RS232 Serial Adapter DEBUG Connector Pin Descriptions RS232 SERIAL ADAPTER (EC2) USER S GUIDE 1. Contents The RS232 Serial Adapter (EC2) package contains the following items: RS232 Serial Adapter (RS232 to Debug Interface) 7 Ribbon Cable 2. RS232 Serial Adapter

More information

Part No. Z , IA Nov OPERATION MANUAL. High Voltage Digitalmeter A

Part No. Z , IA Nov OPERATION MANUAL. High Voltage Digitalmeter A Part No. Z1-109-920, IA001723 Nov. 2005 OPERATION MANUAL High Voltage Digitalmeter 149-30A Use of Operation Manual Please read through and understand this Operation Manual before operating the product.

More information

TRC-190 User s Manual

TRC-190 User s Manual First Edition, November 2008 www.moxa.com/product 2008 Moxa Inc. All rights reserved. Reproduction without permission is prohibited. The software described in this manual is furnished under a license agreement

More information

Customer Maintenance Parts List

Customer Maintenance Parts List C US R MODEL SUFFIX Customer Maintenance Parts List Model 0 R0000 dot recorder Made in China RECORDER STYLE FREQUENCY NO. SUPPLY N00 0 TAG NO. Note: Parts marked with a symbol are Customer Maintenance

More information

SY061 Portable Load Meter with IEEE TEDS capability User Instructions

SY061 Portable Load Meter with IEEE TEDS capability User Instructions SY061 Portable Load Meter with IEEE1451.4 TEDS capability User Instructions Preliminary, relates to firmware version 1.2i Introduction The SY061 is a portable load meter, which can indicate the load present

More information

Application Note. Use of Palm Switches with QuadTech Hipot Testers

Application Note. Use of Palm Switches with QuadTech Hipot Testers Application Note Use of Palm Switches with QuadTech Hipot Testers Operator safety is an important aspect to consider when dealing with high voltage devices and instruments such as hipot testers. To improve

More information

U P S USER S MANUAL IMPORTANT SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS. Uninterruptible Power System. Line Interactive (Network) UPS

U P S USER S MANUAL IMPORTANT SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS. Uninterruptible Power System. Line Interactive (Network) UPS Important Safety Instructions IMPORTANT SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS U P S Uninterruptible Power System Line Interactive (Network) UPS 400VA/ 500VA/ 600VA/ 800VA USER S MANUAL WARNING (SAVE

More information

General Operating, Maintenance and Installation Manual

General Operating, Maintenance and Installation Manual General Operating, Maintenance and Installation Manual Hardware Platform for Protocol Converter Small Embedded Controller - SEC2-91056 Erlangen Telephone +49 9131 92076-0 Fax: +49 9131 92076-10 Internet:

More information

Dual channel temperature logger with two voltage inputs 0-5V Instruction Manual

Dual channel temperature logger with two voltage inputs 0-5V Instruction Manual LOGGER S0541 Dual channel temperature logger with two voltage inputs 0-5V Instruction Manual Instruction Manual for use of S0541 logger Instrument is designed for measurement and record of temperature

More information

HYAMP. Quick Start Guide

HYAMP. Quick Start Guide Quick Start Guide HYAMP Model 40 SAFETY CHECKLIST Survey the test station. Make sure it is safe & orderly. Always keep unqualified/unauthorized personnel away from the test area. Familiarize yourself with

More information

A1H Portable Humidity Temperature Indicator Instruction Manual

A1H Portable Humidity Temperature Indicator Instruction Manual A1H Portable Humidity Temperature Indicator Instruction Manual -2- CONTENTS Description...3 Operation...3 Measurement...4 Maintenance...4 Probe Installation and Removal...5 Output Connector...6 Accessories

More information

epowerswitch-8 Power Management Master Slave User's Guide

epowerswitch-8 Power Management Master Slave User's Guide epowerswitch-8 Power Management Master Slave User's Guide epowerswitch-8 epowerswitch-8 is a couple of power control units made of a Master named epowerswitch-m8 and a slave named epowerswitch-s8. epowerswitch-m8

More information

Net-Ritef Continuity and Wiremap Tester Trace-Ritef Tone Generator and Amplified Tone Tracer User's Guide.

Net-Ritef Continuity and Wiremap Tester Trace-Ritef Tone Generator and Amplified Tone Tracer User's Guide. Net-Ritef Continuity and Wiremap Tester Trace-Ritef Tone Generator and Amplified Tone Tracer User's Guide www.jdsu.com/know Voltage Probe Tracer/Power Button Volume Up/Down Indicator LED Power/Tone Type

More information

OM-DLTH DATA LOGGER INSTRUCTIONS

OM-DLTH DATA LOGGER INSTRUCTIONS OM-DLTH DATA LOGGER INSTRUCTIONS The OMEGA OM-DLTH Data Logger is a self-contained precision instrument for recording the temperature, humidity and dew point of the surrounding environment. The OM-DLTH

More information

TRC-190 User s Manual

TRC-190 User s Manual User s Manual Edition 3.2, May 2017 www.moxa.com/product 2017 Moxa Inc. All rights reserved. User s Manual The software described in this manual is furnished under a license agreement and may be used only

More information

Notes: 1. The module is not protected against ESD, avoid potential difference between yourself and the module before use.

Notes: 1. The module is not protected against ESD, avoid potential difference between yourself and the module before use. The Digital PIR USB Interface module is a link between most* Digital PIR Detectors and a Personal Computer. The microcontroller on the module reads all available information from the Digital Detector on

More information

Instruction Manual for BE-SP3 Circuit. 10/21/07

Instruction Manual for BE-SP3 Circuit. 10/21/07 Page 1 of 54 Instruction Manual for BE-SP3 Circuit. 10/21/07 Page 1 Index: Page 2 BE-SP3 Circuit Specifications. Page 3-4 Intro to the BE-SP3. Page 5 Basics of serial to parallel. Page 6-7 ASCII Code.

More information

Neve 8803 Dual EQ User Manual

Neve 8803 Dual EQ User Manual Neve 8803 Dual EQ User Manual 527-370 Issue 2 AMS NEVE Billington Road Burnley Lancs BB11 5UB England Phone +44 (0)1282 457011 Fax: +44 (0)1282 417282 info@ams-neve.com www.ams-neve.com www.neve.eu - Page

More information

CAMit I Camera with built in Modem

CAMit I Camera with built in Modem CAMit I Camera with built in Modem User s Manual CAMit I AP revision: 3.3 CAMit I Setup revision: 2.0.1 Manual revision: 2.0 Date: February 27, 2002 Congratulations You just acquired a fine product from

More information

Fluke Metrology Software

Fluke Metrology Software Fluke Metrology Software Version 7 MET/CAL 5500/CAL Getting Started Guide P/N 1275404 July 1999 Rev. 2, 9/04 1996-2004Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

4-20 ma Current Data Logger

4-20 ma Current Data Logger 1 4-20 ma Current Data Logger OM-DL420 DATA LOGGER INSTRUCTIONS The OMEGA OM-DL420 Data Logger is a self-contained precision instrument for recording the current in a 4 20 ma loop. The OM-DL420 can store

More information

OPERATING INSTRUCTIONS AND SERVICE MANUAL BASEBALL SCOREBOARD MODEL MP-3350 WITH MP-2002 CONTROL

OPERATING INSTRUCTIONS AND SERVICE MANUAL BASEBALL SCOREBOARD MODEL MP-3350 WITH MP-2002 CONTROL OPERATING INSTRUCTIONS AND SERVICE MANUAL BASEBALL SCOREBOARD MODEL MP-3350 WITH MP-2002 CONTROL EFFECTIVE S.N. XXXX JAN 1, 1993 1 1. General Information TABLE OF CONTENTS 1.1 Description 1.2 Identification

More information

OPERATING INSTRUCTIONS AND SERVICE MANUAL. MODEL MP-3804 BLACK FLAG DISPLAY WITH MP-2002 Control

OPERATING INSTRUCTIONS AND SERVICE MANUAL. MODEL MP-3804 BLACK FLAG DISPLAY WITH MP-2002 Control OPERATING INSTRUCTIONS AND SERVICE MANUAL MODEL MP-3804 BLACK FLAG DISPLAY WITH MP-2002 Control EFFECTIVE S.N. 18,959 May 2, 2003 TABLE OF CONTENTS 1. General Information 1.1 Description 1.2 Identification

More information

MC 11 EB-2 Power supply cabinet with external bus, AC version

MC 11 EB-2 Power supply cabinet with external bus, AC version MC 11 EB-2 Power supply cabinet with external bus, AC version USER/MAINTENANCE MANUAL 1 SLOT 0 SLOT 1 SLOT 2 SLOT 3 SLOT 4 SLOT 5 SLOT 6 SLOT 7 SLOT 8 SLOT 9 SLOT 10 SLOT 11 EB-2 (a) MC11 (b) (c) Figures

More information

8 Button RS232/IR. Control Panel. MuxLab Inc A / SE A

8 Button RS232/IR. Control Panel. MuxLab Inc A / SE A 8 Button RS232/IR Control Panel 500816 MuxLab Inc. 2016 94-000833-A / SE-000833-A SAFETY PRECAUTIONS To insure the best use from the product, please read all instructions carefully before using the device.

More information

2214 EVENT LOGGER MODULE FOR THE VX2200 SYSTEM

2214 EVENT LOGGER MODULE FOR THE VX2200 SYSTEM 2214 EVENT LOGGER MODULE FOR THE VX2200 SYSTEM CONTENTS PAGE Manual Introduction 3 Product Overview 3 Operation and Connection 3 RS232 Connection 3-4 RS485 Connection 3-4 2214 PCB layout 5 RS485 Bus Terminal

More information

1. CP430, CP470, CP474, CP770 and CP774

1. CP430, CP470, CP474, CP770 and CP774 1. CP430, CP470, CP474, CP770 and CP774 1.1 Order data CPUs CP430, CP470, CP474, CP770 and CP774 CP430, CP470, CP770 CP474, CP774 Model number 7CP430.60-1 7CP470.60-2 7CP474.60-2 7CP770.60-1 7CP774.60-1

More information

User Guide Automatic Transfer Switch (ATS)

User Guide Automatic Transfer Switch (ATS) User Guide Automatic Transfer Switch (ATS) V.1.0 Table of Contents 1. Introduction... 2 2. Product Overview... 2 3. Important Safety Warnings... 3 4. Operation Indicators & Status... 3 5. Installation...

More information

D6600 Communications Receiver/Gateway

D6600 Communications Receiver/Gateway Communications Receiver/Gateway Quick Start.0 Parts List able : System Components able : Documentation Qty. Description Receiver/Gateway AC Power Cord Battery Cables I/O Cable Rack Mount Ears 8/ Philip-Head

More information

OPERATING PROCEDURES for the UPS-S2 Universal Power Supply Series 2

OPERATING PROCEDURES for the UPS-S2 Universal Power Supply Series 2 OPERATING PROCEDURES for the Universal Power Supply Series 2 Vanguard Instruments Co., Inc. 1520 S. Hellman Ave. Ontario, California 91761 TEL: 909-923-9390 November 2013 FAX: 909-923-9391 REV. 2 SAFETY

More information

D8000 SERIES QUICK START GUIDE

D8000 SERIES QUICK START GUIDE D8000 SERIES QUICK START GUIDE Version 1.0 Overview The D8000 series modules require a DC Voltage power supply, a USB cable and an unused computer USB port for proper operation. Connecting the D8000 series

More information

Installation and Operation Back-UPS Pro 900

Installation and Operation Back-UPS Pro 900 Us er Documentation Installation and Operation Back-UPS Pro 900 Inventory bu001a User Documentation (2) Safety and General Information Inspect the package contents upon receipt. Notify the carrier and

More information

INSTRUCTION MANUAL. Model Dual Input RTD Thermometer. Measures two temperatures simultaneously. Dual RTD probe inputs

INSTRUCTION MANUAL. Model Dual Input RTD Thermometer. Measures two temperatures simultaneously. Dual RTD probe inputs INSTRUCTION MANUAL Model 421504 Dual Input RTD Thermometer Measures two temperatures simultaneously Dual RTD probe inputs Clock and Elapsed Timer functions Special functions include Data Hold, MIN/MAX/AVG,

More information

MC-12 Software Installation Instructions

MC-12 Software Installation Instructions MC-12 DOCUMENTATION CONVENTIONS This document contains software installation instructions for the MC-12/MC-12 Balanced. Refer to the MC-12 User Guide for general safety, installation, and operating instructions.

More information

QUICK START USER GUIDE. Data Logger Model L452

QUICK START USER GUIDE. Data Logger Model L452 QUICK START USER GUIDE Data Logger Model L452 Statement of Compliance Chauvin Arnoux, Inc. d.b.a. AEMC Instruments certifies that this instrument has been calibrated using standards and instruments traceable

More information

ZENSOL CIRCUIT BREAKER PERFORMANCE ANALYZER

ZENSOL CIRCUIT BREAKER PERFORMANCE ANALYZER ZENSOL AUTOMATION INC. COMPUTERIZED TEST EQUIPMENT ZENSOL CIRCUIT BREAKER PERFORMANCE ANALYZER CBA-32P Micro CBA MANUAL 1W E OPERATOR S GUIDE Version 1.70 January 2001 www.zensol.com Man-1we.doc Rév 5

More information

RS-422 Code-Operated Switches

RS-422 Code-Operated Switches JUNE 2000 SW421A-R2 SW422A-R2 RS-422 Code-Operated Switches COS/4 TEXT TRANSPARENT GRAPHICS MODE RESET ST LO CUSTOMER SUPPORT INFORMATION Order toll-free in the U.S. 24 hours, 7 A.M. Monday to midnight

More information

Blue Point Engineering

Blue Point Engineering Blue Point Engineering Board - Pro Module (E) Instruction Pointing the Way to Solutions! Controller I Version 2.1 The Board Pro E Module provides the following features: Up to 4 minutes recording time

More information

User Manual Back-UPS BE650MC

User Manual Back-UPS BE650MC User Manual Back-UPS BE650MC Inventory Back-UPS ES 650 bu075b This unit is intended for indoor use only. Do not operate this unit in direct sunlight, in contact with fluids, or where there is excessive

More information

EZCT S2A. current transformer test set. Vanguard Instruments Company, Inc.

EZCT S2A. current transformer test set. Vanguard Instruments Company, Inc. EZCT S2A current transformer test set Vanguard Instruments Company, Inc. www.vanguard-instruments.com EZCT S2A current transformer test set The EZCT S2A is Vanguard s third-generation, microprocessor-based,

More information

Supplement for module D041 incl. ATMega8 Prozessor

Supplement for module D041 incl. ATMega8 Prozessor Supplement for module D041 incl. ATMega8 Prozessor V 1.4 16. March 2006 2006 by Peter Küsters This document is in copyright protected. It is not permitted to change any part of it. It is not permitted

More information

S&C M Series Switch Operator. Sophisticated Microprocessor Control of Overhead Distribution Switches

S&C M Series Switch Operator. Sophisticated Microprocessor Control of Overhead Distribution Switches S&C M Series Switch Operator Sophisticated Microprocessor Control of Overhead Distribution Switches S&C s M Series Switch Operator is a fully integrated, microprocessor-based operator that lets you automate

More information

Series L Model 8 Recording Volt-Ammeter Instrument Operation, Software and Modem Manual

Series L Model 8 Recording Volt-Ammeter Instrument Operation, Software and Modem Manual 2038-003 Revision J March 2001 Series L Model 8 Recording Volt-Ammeter Instrument Operation, Software and Modem Manual Copyright 2001 AVO International. All rights reserved. The information presented in

More information

User Manual MHD44. HDMI Matrix Switcher. All Rights Reserved. Version: MHD44_2014V1.0

User Manual MHD44. HDMI Matrix Switcher. All Rights Reserved. Version: MHD44_2014V1.0 User Manual MHD44 HDMI Matrix Switcher All Rights Reserved Version: MHD44_2014V1.0 HDMI 4x4 Matrix Switcher NOTICE: Please read this user manual carefully before using this product. This manual is only

More information

User's Manual. Autoranging Datalogging Insulation Tester (Megohmmeter) with PC Interface. Model

User's Manual. Autoranging Datalogging Insulation Tester (Megohmmeter) with PC Interface. Model User's Manual Autoranging Datalogging Insulation Tester (Megohmmeter) with PC Interface Model 380366 Introduction Congratulations on your purchase of the Extech 380366 Datalogging Insulation Tester. The

More information

Copyright Black Box Corporation. All rights reserved.

Copyright Black Box Corporation. All rights reserved. Copyright 1998. Black Box Corporation. All rights reserved. 1000 Park Drive Lawrence, PA 150551018 7247465500 Fax 7247460746 SAM232 Compact MARCH 1998 TS158A DB 9P cable test 1 5 6 8 7 2 3 9 4 7 6 5 4

More information

HI-POT TESTER User Manual

HI-POT TESTER User Manual HI-POT TESTER 7600 User Manual Contents 1. Safety...- 3-1.1 General...- 3 - AC Power Supply...- 4-1.2 Adjustment, Maintenance and Repair...- 5-1.3 Static Electricity...- 5-2. General:...- 6-2.1Packing

More information

805HP. Handheld Digital Weight Indicator User s Manual (v1703) Anyload Transducer Co. Ltd Website:

805HP. Handheld Digital Weight Indicator User s Manual (v1703) Anyload Transducer Co. Ltd Website: 805HP Handheld Digital Weight Indicator User s Manual (v1703) Anyload Transducer Co. Ltd Website: www.anyload.com Email: info@anyload.com TABLE OF CONTENTS 1. Introductions and Features 2 2. Safety Recommendations

More information

1. Introduction. 2. Installation INSTALLATION INSTRUCTIONS: MCD USB MODULE

1. Introduction. 2. Installation INSTALLATION INSTRUCTIONS: MCD USB MODULE Introduction INSTALLATION INSTRUCTIONS: MCD USB MODULE 1. Introduction The USB Module can be used in conjunction with WinMaster to manage Danfoss soft starters. These instructions detail the installation,

More information

7561-PSD Manual Portable Battery Powered Indicator

7561-PSD Manual Portable Battery Powered Indicator 7561-PSD Manual Portable Battery Powered Indicator Lebow Products Inc. 1728 Maplelawn Drive P.O. Box 1089 Troy, Michigan 48084-1089 (800) 803-1164 Phone: (248) 643-0220 FAX: (248) 643-0259 Visit our web

More information