315 ph/ion. Installation. Specifications. Installing the Batteries a595. Input Connectors

Size: px
Start display at page:

Download "315 ph/ion. Installation. Specifications. Installing the Batteries a595. Input Connectors"

Transcription

1 315 /Ion Installation Input Connectors socket (automatic temperature compensation) socket data socket reference socket If you ordered an electrode with the 315, connect it as follows: Corning Incorporated Science Products Division Corning, New York USA. Tel: / For Technical Assistance: US and Canada call / Specifications Measurement Ranges to mv ± Temp to C Conc to Resolution mv 0.1 Temp. 0.1 Conc. ± 1 LSD Relative Accuracy mv 0.2 mv Temp. 0.3 C Conc. 0.5% Memory 50 memories Isopotential Point Conc. User selectable Calibration Buffers 3 sets of 4 buffers Calibration Points 3 of 3 selectable Conc. 5 of 5 selectable Temp. Compensation -5.0 to C, automatic and manual Display LCD Outputs Recorder Serial Auto Off 10 minutes from last keypress Operating Conditions Temperature 5 to 40 C Humidity 5 to 80% (non condensing) Size/Weight 3 1/4 x 7 3/4 x 1 3/4 inches (85 x 200 x 45 mm) 1lb (0.45kg) Batteries 4 x AA/LR6 alkaline 1.5V d.c. Regulatory Compliance Complies with Part 15 of FCC Rules (Class A computing device). Waterproof to IP67* *when used as directed with electrodes with waterproof covers a595 socket O rings connector alignment cover 1. Make sure connector cap is loosened. 2. Connect the electrode BNC connector to the socket. 3. Lightly grease the O rings with the grease supplied, then slide the waterproof cover over the socket. Make sure the cover is lined up correctly with the socket. NOTE: Lightly greasing the cables will help removal of the connector cover in future. 4. Finger tighten the connector cap. 5. Repeat steps 1-4 connecting the phono connector to the socket. CAUTION: To avoid damaging the connectors always loosen the connector cap before removing the cover. Installing the Batteries hand grip coin battery cover connector cap O ring seal 1. Remove the hand grip, and unscrew the battery cover using a coin. 2. Insert the batteries following the illustration. 3. Lightly grease the O ring seal, and replace the battery cover and hand grip.

2 Installation (continued) Fitting the Wrist Strap Controls Press & Release Press & Hold for 2 Seconds Program Menu The Program Menu allows you to set a calibration reminder interval; date and time; manual temperature compensation; calibration buffers; concentration standards; ISE (ion selective electrode) slope value; isopotential point and concentration units. You can only enter the Program Menu if the current measurement has endpointed. Press and hold the mode key for 2 seconds to access the Program Menu - appears. Press mode to scroll through the options. Press read to exit the Program Menu at any time. If you press read when a value is flashing that value will not be entered. Display low batt. indicator shown here electrode condition shown here starts/stops measurement stability/auto endpoint indicator M 25.0 c 50 temperature, electrode offset/slope shown here turns auto endpoint on/off starts calibration recalls last cal data for current mode selects /mv/ enters Program Menu conc/off. Displays mv and temperature of buffer or standard after calibration. stores result turns continuous in memory measurement (up to 50) on/off recalls result turns continuous from memory data transfer on/off memory, cal and error number shown here conc. units shown here prog. menu, cont. read and data transfer icons shown here MTC The calibration reminder interval can be set, in hours, between 0 and 9 (where 0 = no reminder). When the selected time has elapsed the calibration reminder symbol will appear. The 315 has a preset reminder interval of 1. Change the interval using and. Press mode to enter the value and move on. Time and date are displayed during calibration recall, and will be sent via the serial output. Time is shown on the main display, the date is shown on the lower part of the display. Set the time and date using and. Press mode to enter the value and move on. You can enter temperature manually between 5.0 and C. (An probe will override manual compensation.) The 315 has a preset temperature of 25 C. Use and to change temperature. Press mode to enter the value and move on. If you entered the Program Menu from mode (or mv mode), continue with the mode instructions. If you entered from concentration mode follow the concentration mode instructions mode You can select three of four calibration buffers. The buffers are grouped in sets (b = 1, b = 2, b = 3); select your required set first using and : set 1 = 7.000, 4.000, , set 2 = 7.000, 4.010, 9.210, set 3 = 6.860, 4.010, 9.180, Press mode to enter the set and move on.

3 Program Menu (continued) cal 1 cal 2 cal 3 c = 1 ISo.Pt. cal 1 to cal 5 Choose three buffers (cal 1, cal 2, cal 3) from the set for calibration. For ease of use, set the buffers in the order that you will use them. For example: Set 1 (factory settings) cal 1=7.000, cal 2=4.000 and cal 3= not selected. These can be rearranged in any order using and. Press mode to enter the value and move on. Press read to exit the Program Menu. Concentration mode You can select one of four preset slope values for 1 point calibrations (c = 1, c = 2, c = 3, c = 4). If you are not sure of your electrode type, or wish to do 2- (3, 4, 5) point calibrations, do not select a slope as the 315 will calculate the slope and use this value rather than the factory setting. c = mv/decade, e.g. for BF 4, Br, Cl, ClO 4, F, I, NO 2, NO 3, SCN 2 c = mv/decade, e.g. for CO 3, S 2 c = mv/decade, e.g. for Ag +, H +, K +, Na + +, NH 4 c = mv/decade, e.g. for Ca 2+, Cd 2+, Cu 2+, Pb 2+, WH 2+ Select your required slope using and, press mode to enter the value and move on. You can enter the isopotential point (in mv) of your electrode. The range is 99 to +99 mv. The 315 has a preset isopotential point of 0 mv. Change the isopotential point using and, press mode to enter the value and move on. You can set five concentration standards to values of your choice, the range is to For ease of use, set the standards in the order that you will use them. Set the values using or, press mode to enter the value and move on. For example, to set 500 Press or ; the display shows 10000, the 10 will flash. Press or to change the value to 0000 (no thousands), then press mode to confirm. The value changes to 000.0, with the first 0 ( hundreds ) flashing. Program Menu (continued) conc Press or to change the hundreds digit until the display shows 500.0, then press mode to confirm. The tens digit will flash. Press mode until the display reads with the tenths digit flashing, then press mode again to enter value. For example, to set 0.5 Press or ; the display shows 10000, the 10 will flash. Press or to change the value to 0000 (no thousands), then press mode to confirm. The value changes to 000.0, with the first 0 ( hundreds ) flashing. Press mode until the display reads 0.000, with the tenths digit flashing. Press or to change the tenths digit until the display shows 0.500, then press mode to confirm. The hundredths digit will flash. Press mode again, the display reads with the thousandths digit flashing, press mode again to enter value. Select concentration units from the following list: blank (no units), ppm, mg/g, mol/l, mmol/l, g/l, mg/l, eq/l, meq/l, %, ppt. The 315 is preset to blank (no units). Use or to change units, press mode to enter the units and move on Press read to exit the Program Menu. Buffer Tables The 315 automatically corrects buffers for temperature using the values shown in the table C C C C C C C C C C

4 Calibrating 1. Select or conc. mode Measuring Samples 1. Select, mv or concentration mode 2. 1-point calibration Example buffer or standard 2. Measure sample Example cal1 cal buffer/standard reminder Auto endpoint or 2 seconds c cal cal1 E0 mv sample Auto endpoint or c c 3. Rinse electrode and blot dry H 2 O to return to sample measurement, or continue with step 4 for further cal points (3 or 3, 4, 5 concentration) point calibration, repeat steps 2 and 3 using other buffer/standard auto or manual endpoint only 97.2 % slope cal % % % electrode in electrode electrode good condition needs cleaning needs conditioning 3. Rinse, blot and store electrode H 2 O If you press mode to select, mv or concentration the 315 begins measuring immediately, and you will not need to press read to start the sample measurement. The decimal point flashes during sample measurement. When the electrode output has stabilized the stability indicator appears. Manual endpoint - press read to endpoint Auto endpoint - the 315 automatically endpoints. When you press cal the buffer or concentration standard you selected for cal 1 (Program Menu) is displayed for 2 seconds, and then the 315 starts measuring. If you want to use another buffer (choice of 3) or conc. standard (choice of 5) press cal again. Press cal repeatedly to exit calibration routine. The decimal point flashes during cal measurement. When the electrode output has stabilized the stability indicator appears. Manual endpoint - press read to endpoint Auto endpoint - the 315 automatically endpoints. When a calibration has endpointed you can press mode to display the absolute mv value and temperature of the buffer or standard; the display reverts to the buffer or standard and E 0 ( only) values, or the buffer or standard and slope values for a 2-(3, 4, 5) point calibration after 2 seconds.

5 Maintenance General The 315 requires very little maintenance. Occasionally wipe the meter with a damp cloth. The casework is made of ABS/PC which is known to be affected by some organic solvents, including toluene, xylene and methyl-ethyl-ketone. It is good practice to wipe away any spillages as soon as they occur. Every six months inspect and lightly grease the socket seals, and connector and battery O rings with the grease supplied. Replace the O rings if damaged. CAUTION: To prevent permanent damage care should be taken to prevent liquid entering the 315 when fitting electrodes or replacing batteries. If it does, remove the batteries and allow the meter to dry before using. Replacing the Batteries Replace batteries as soon as the low battery indicator appears. To prevent data loss always turn the 315 off using the mode key before replacing the batteries. See the Installation section for details on installing batteries. Do not mix old and new batteries. NOTE: Check the Program Menu settings after changing the batteries. Electrode Maintenance Refer to the electrode product insert for full details on maintaining your electrode. Spares and Accessories Epoxy, 3-in-1 Polymer Gel Electrode (4' cable) Epoxy, 3-in-1 Refillable Electrode (4' cable) Replaceable Junctions 3/pk Rubber Meter Holder and Neck Strap Mini Electrode Stand Hard Carrying Case RS232 Interface Kit Recorder Cable Wrist Strap Instrument Sealing Kit Neck Strap Batteries, AA, 4/pk Guide to Measurement Guide to ISE Measurement Buffer Sachets, 30/case Buffer Sachets, 30/case Buffer Sachets, 30/case Rinsing Solution Sachets, 30/case Buffer Sachet, Multi-Pack, 32/case Buffer Red, 500 ml, 2/case Buffer Yellow, 500 ml, 2/case Buffer Blue, 500 ml, 2/case Rainbow Buffer Pack, 500 ml (2 each of 4.00, 7.00, 10.01) 6/case Problem Solving Err 1 - offset value (E 0 ) out of range Check correct buffer is used. Check mv reading for 7 buffer is 0 ± 30 mv. If it is not, clean or replace the electrode. Err 2 - slope out of range in mode Slope less than 85.0%, or not + Err 2 calculable. Electrode needs cleaning, conditioning or replacing. Check correct buffers are used. + Err 2 Slope more than 105.0%. Check calibration buffers. NOTE: Where 0 mv/ unit = 0%, and mv/ unit = 100% Err 2 - slope out of range in concentration mode Check that for a decade change in concentration, the mv reading is between +10 to +65 mv, or -10 to 65 mv, depending on the valency and charge of the species being measured. If it is not, check the concentration standards and clean, condition or replace the electrode. Err 3 - buffer outside temperature limits buffers must be between 5 C and 50 C for accurate temperature compensation. Displayed (out of range) Check electrode is connected and immersed in sample. Check wetting cap is removed. Data Entry Errors Date flashes and reverts to previous setting - invalid date entered. Entered temperature value changes to 5.0 or the 315 will not accept values outside this range. Dashes displayed instead of buffer or standard value during calibration buffer or standard set up - that buffer or standard has already been selected for a cal point. Cal points must have different values. Using the Memory Entering a Reading into Memory The 315 can store up to 50 endpointed results. Press M when the measurement has endpointed. M 1 is displayed (or M 2 to M 50 if readings have already been stored). M 50 indicates the memory is full.

6 Using the Memory (continued) Recalling Memory You can only recall stored memories if the current measurement has endpointed. Press R - the last stored memory is displayed. Press or to scroll through memories. RM 1 to RM 50 indicates which memory is being displayed. M 0 indicates no memories stored. Clearing the Memories Press R then press or until M C is displayed. Press mode to clear the memories, M 0 indicates the memories have been cleared (press read to exit without clearing the memories). Using Continuous Measurement Mode In normal mode, the 315 will auto-off 10 minutes after the last key press, whether the unit is measuring or has endpointed. If you select continuous measurement mode, the 315 will not auto-off. To select continuous measurement mode press and hold M for 2 seconds. Using Continuous Data Transfer Mode In continuous data transfer mode readings are sent to the data output approximately every second. If the measurement endpoints (manually, or using auto endpoint) data transfer stops. If you recall memories all stored readings are output, (from the first to the last stored). To select continuous data transfer mode press and hold R for 2 seconds. If you want to use this mode to monitor continually, you also need to select continuous measurement mode. To maintain continuity of sample readings calibration data is not output in continuous data transfer mode. Interfacing Serial Output The 315 can interface with computers, printers and other RS232 compatible devices. The RS232 Interface Kit (Cat ) converts data from the 6 pin socket to RS232 format. Type of communication - uni directional, baud rate 1200, data format - 1 start bit : 8 data bits : 1 stop bit : no parity Data is sent to the output socket at measurement endpoint (manual or automatic), calibration recall and memory recall. Interfacing (continued) To use the output to monitor continually, you need to select continuous measurement mode and continuous data transfer mode. Recorder Output An output is available, via the Recorder Cable (Cat ), for all modes. The polarity of the output is the same as the polarity of the input signal. Approximate Recorder Outputs - Output follows display and provides approximately 60 mv output/ unit. This value varies with changes in electrode slope. mv - Output follows display, i.e. output varies from to mv. Concentration - Output depends on valency and charge of ionic species being measured, and is typically approximately ± 60 or ± 30 mv/decade. Recorder output is not available for temperature. 315 Data Socket CAUTION: The 315 does not meet IP67 when the waterproof output socket seal has been removed. Operating Hints protective ground (0V) recorder output Remove the wetting cap from the end of the electrode and the rubber cap from the fill hole (if fitted) before using the electrode. Calibrate using buffers or standards with values that bracket that of the sample. With a new electrode, or after maintenance, we recommend you use a buffer close to 7 for the first calibration point. For greatest accuracy, buffers or standards and samples should be at the same temperature. Do not use solutions after the expiration date. When transferring the electrode from one solution to another, rinse it with distilled water and blot dry with tissue paper - do not wipe the electrode as this may cause polarization and slow response. When you select the calibration buffers/ standards (Program Menu), set the buffer or standard you will use most frequently for 1-point calibrations as cal 1, the one you will use most frequently for the second calibration point as cal 2, and so on. Calibrate in concentration mode using the units you want for sample measurement. Changing the units does not change the reading; e.g. the 315 does not convert from ppm to mmol.

Operating Instructions. SenseLine ph meter F410

Operating Instructions. SenseLine ph meter F410 Operating Instructions SenseLine ph meter F410 Contents Contents 1. Introduction 3 2. Safety measures 3 3. Installation 5 3.1 Installing the batteries 5 3.2 Fitting the wrist strap 5 4. Operating the F410

More information

450 ph/ion Meter. Instruction Manual

450 ph/ion Meter. Instruction Manual 450 ph/ion Meter Instruction Manual This manual contains complete instructions for setting up and using the 450 ph/ion meter. Applications information is also available. The information contained in this

More information

Thermo Scientific Orion Star A324 Portable ph/ise Meter. Instruction Sheet

Thermo Scientific Orion Star A324 Portable ph/ise Meter. Instruction Sheet English ph / ISE Thermo Scientific Orion Star A324 Portable ph/ise Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the appropriate wall socket plug plate.

More information

Thermo Scientific Orion Star A214 Benchtop ph/ise Meter. Instruction Sheet

Thermo Scientific Orion Star A214 Benchtop ph/ise Meter. Instruction Sheet English ph / ISE Thermo Scientific Orion Star A214 Benchtop ph/ise Meter Instruction Sheet Preparation Power Source 1. Power adapter (included with meter) a. Select the appropriate wall socket plug plate.

More information

STARTER 3000 Bench ph Meter Instruction Manual

STARTER 3000 Bench ph Meter Instruction Manual STARTER 3000 Bench ph Meter Instruction Manual Table of Contents 1 Introduction... 2 1.1 Safety measures... 2 1.2 Display and controls... 3 2 Installation... 5 2.1 Package contents... 5 2.2 Installing

More information

USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements

USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements www.adwainstruments.com Dear Customer, Thank you for choosing an Adwa product. Please read carefully this manual before starting

More information

INSTRUCTION MANUAL. Portable ph and ph/ise Meters. Models 210Aplus 230Aplus 250Aplus 290Aplus

INSTRUCTION MANUAL. Portable ph and ph/ise Meters. Models 210Aplus 230Aplus 250Aplus 290Aplus INSTRUCTION MANUAL Portable ph and ph/ise Meters Models 210Aplus 230Aplus 250Aplus 290Aplus Table Of Contents Chapter I Introduction................................1 Chapter II General Information.........................3

More information

User Manual. Thermo Scientific Orion Star A320 Series

User Manual. Thermo Scientific Orion Star A320 Series User Manual Thermo Scientific Orion Star A320 Series Electrochemistry Portable Meters 68X000443 Revision A June 2015 Important Note Please read this user guide thoroughly before using your meter. Any use

More information

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet English ph / ISE / COND / RDO / DO Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the

More information

INSTRUCTION MANUAL ORP-METER MODEL COR-01

INSTRUCTION MANUAL ORP-METER MODEL COR-01 INSTRUCTION MANUAL ORP-METER MODEL COR-01 EL-221, MIDC Electronic Zone, Mhape, Navi Mumbai-400701. Tel: 022-61944000 CHAPTER 1 INTRODUCTION CONTECH ORP meter measures ph and mv, can used for Oxidation

More information

HI HI HI HI HI HI 98185

HI HI HI HI HI HI 98185 Instruction Manual HI 98180 HI 98181 HI 98182 HI 98183 HI 98184 HI 98185 Calibration Check Waterproof ph/mv/ise/temperature Meters www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments

More information

USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter

USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter www.adwainstruments.com 1 Dear Customer, Thank you for choosing an Adwa product. Please read carefully this manual before starting operations. This instrument

More information

Operating the meter 13. j h g. 4.1 Backside layout Pin assignments RS232 connection

Operating the meter 13. j h g. 4.1 Backside layout Pin assignments RS232 connection 4 Operating the meter 4.1 Backside layout c d e f a b j h g a b c d e f g h i BNC socket for mv/ph signal input RCA (Cinch) socket for temperature signal input Reference socket for reference electrodes

More information

User Guide. ExStik FL700. Fluoride Meter

User Guide. ExStik FL700. Fluoride Meter User Guide ExStik FL700 Fluoride Meter Introduction The model FL700 is a system specifically designed for the quick and accurate measurement of fluoride ions in drinking water and other aqueous samples.

More information

Operating Instructions

Operating Instructions Beckman Instructions 511290-AA Φ ph/temperature/mv/ise Meters Operating Instructions Beckman Instruments, Inc. 4300 N. Harbor Blvd., Fullerton, CA 93835 Copyright 1998 Beckman Instruments, Inc. Printed

More information

Operating Instructions Bedienungsanleitung Mode d emploi Istruzioni d uso Instrucciones de manejo 使用说明. Conductivity Meter SG3

Operating Instructions Bedienungsanleitung Mode d emploi Istruzioni d uso Instrucciones de manejo 使用说明. Conductivity Meter SG3 Operating Instructions Bedienungsanleitung Mode d emploi Istruzioni d uso Instrucciones de manejo 使用说明 Conductivity Meter SG3 中文 Español Italiano Français Deutsch English Contents Contents 1. Introduction

More information

Agilent 3200M Multi-Parameter Analyzer

Agilent 3200M Multi-Parameter Analyzer Agilent 3200M Multi-Parameter Analyzer User Guide Agilent Technologies Notices Agilent Technologies (Shanghai) Co., Ltd. 2012 No part of this manual may be reproduced in any form or by any means (including

More information

USER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe

USER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe USER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe www.adwainstruments.com Dear Customer, Thank you for choosing an Adwa product. Please read carefully this manual before starting

More information

USER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe

USER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe USER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe ADWA HUNGARY Kft. Alsókikötõ sor 11, 6726 Szeged, Hungary Tel. +36 62 317 878 Fax +36 62 550 610 www.adwainstruments.com MANAD630

More information

As a result, our corporate philosophy is focused on total customer satisfaction. The Headquarter of Adwa Instruments is located in Szeged, Hungary:

As a result, our corporate philosophy is focused on total customer satisfaction. The Headquarter of Adwa Instruments is located in Szeged, Hungary: Adwa Instruments Founded in 1998, Adwa Instruments is one of the leading manufacturers of instrumentation to measure, Conductivity, ORP, Dissolved Oxygen and erature for Research and Industry. Headquartered

More information

Thermo Scientific Orion Star A326 Portable ph/rdo/do Meter. Instruction Sheet

Thermo Scientific Orion Star A326 Portable ph/rdo/do Meter. Instruction Sheet English ph / RDO / DO Thermo Scientific Orion Star A326 Portable ph/rdo/do Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the appropriate wall socket plug

More information

Instruction Manual ph 600/610/620 Waterproof Handheld ph Meter ph/mv/ion/ºc/ºf meter Technology Made Easy...

Instruction Manual ph 600/610/620 Waterproof Handheld ph Meter ph/mv/ion/ºc/ºf meter Technology Made Easy... Instruction Manual ph 600/610/620 Waterproof Handheld ph Meter ph/mv/ion/ºc/ºf meter Technology Made Easy... 68X415301 Rev 0 JUN 06 Preface This manual serves to explain the use of the ph600/610/620 Waterproof

More information

Instruction Manual Mi 160 Bench Meter

Instruction Manual Mi 160 Bench Meter Instruction Manual Mi 160 Bench Meter GENERAL DESCRIPTION... 4 FUNCTIONAL DESCRIPTION... 2 SPECIFICATIONS... 5 OPERATIONAL GUIDE... 6 ph CALIBRATION... 9 ph BUFFER TEMPERATURE DEPENDENCE... 12 RELATIVE

More information

Research Grade Meters

Research Grade Meters HI5521 HI5522 Research Grade Meters ph/orp/ise and EC/TDS/Resistivity/Salinity and Temperature bench Conductivity/TDS Measure up to Eight Parameters HI5521 and HI5522 are research grade benchtop instruments

More information

Operations Manual EcoSense ph100a

Operations Manual EcoSense ph100a Operations Manual EcoSense ph100a Portable ph, mv and Temperature Instrument English CONTENTS GENERAL INTRODUCTION... 3 INITIAL INSPECTION... 3 THE INSTRUMENT... 3 BATTERY INSTALLATION... 3 Battery Disposal...

More information

Thermo Scientific Orion Star A325 Portable ph/conductivity Meter. Instruction Sheet

Thermo Scientific Orion Star A325 Portable ph/conductivity Meter. Instruction Sheet English ph / COND Thermo Scientific Orion Star A325 Portable ph/conductivity Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the appropriate wall socket plug

More information

Resolution selectable from 0.01 and ph. Range ph. Accuracy ± ph

Resolution selectable from 0.01 and ph. Range ph. Accuracy ± ph ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or

More information

ph/orp Meter SD Card real-time Datalogger

ph/orp Meter SD Card real-time Datalogger User's Guide ph/orp Meter SD Card real-time Datalogger Model SDL100 Introduction Congratulations on your purchase of the Extech SDL100 ph/orp Meter, an SD Logger Series meter. This meter displays and stores

More information

User Manual. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500. Additional User Manual Translations available at

User Manual. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500. Additional User Manual Translations available at User Manual ExStik EC500 ph/conductivity / TDS / Salinity / Temperature Meter EC500 Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the

More information

Introduction: Features:

Introduction: Features: Index Features:... 3 Specifications:... 4 Accessories:... 5 Functions of Keyboard:... 7 Preparation:... 8 ... 8 ... 8 Calibration:... 8 ... 8 ... 9 ...

More information

DO Meter Kit. Instruction Manual

DO Meter Kit. Instruction Manual DO Meter Kit 850048 Instruction Manual DO Meter Kit 850048 Copyright 2013 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or transmitted in any

More information

Electrodes ph EC DO /14 dge 2 P/L e

Electrodes ph EC DO /14 dge 2 P/L e ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand Manual log on stability Interval logging Temperature readout ( C or

More information

Research Grade Meters. ph, ISE, EC and DO. Laboratory bench meters with a color, graphic display and capacitive touch keypad

Research Grade Meters. ph, ISE, EC and DO. Laboratory bench meters with a color, graphic display and capacitive touch keypad Research Grade Meters ph, ISE, EC and DO Laboratory bench meters with a color, graphic display and capacitive touch keypad HI5000 Series Research Grade Meters Overview HI5000 Series Research Grade Meters

More information

HI 3220 HI 3221 HI 3222

HI 3220 HI 3221 HI 3222 Instruction Manual HI 3220 HI 3221 HI 3222 Calibration Check ph/mv/ise/temperature Bench Meters WARRANTY The HI 322x series are guaranteed for two years against defects in workmanship and materials when

More information

Handheld Water Quality Meters 100 Series

Handheld Water Quality Meters 100 Series ph ORP Ion Conductivity Resistivity TDS DO Salinity Handheld Water Quality Meters 100 Series METERS 2 YEAR W A R R A N T Y M A D E I N J A P A N www.horiba-laqua.com In the lab, in the field or anywhere

More information

WK7200. Water Quality Test Kit USER S MANUAL

WK7200. Water Quality Test Kit USER S MANUAL WK7200 USER S MANUAL Water Quality Test Kit INTRODUCTION: Thank you for selecting model WK7200 microprocessor-based waterproof Water Quality Test Kit. It is used to measure a wide range of ph, ORP, Conductivity,

More information

Integrated Instruction Manual WP 600 Series Meters

Integrated Instruction Manual WP 600 Series Meters Integrated Instruction Manual WP 600 Series Meters ph 600/ 610/ 620, COND 600/610, DO 600, PC 650, PD 650, CD 650, PCD 650 68X415307 Rev. 1 Oct 07 Part of Thermo Fisher Scientific Technology Made Easy...

More information

ph 300 ph 301 ph 302 Microprocessor GLP Bench-top ph/mv/ise/ C Meters Instruction Manual

ph 300 ph 301 ph 302 Microprocessor GLP Bench-top ph/mv/ise/ C Meters Instruction Manual Instruction Manual ph 300 ph 30 ph 302 Microprocessor GLP Bench-top ph/mv/ise/ Meters Rel mv ppm ph DATE TIME MEM mv ppm ph INTV WRONG m BUF 2 s LOG 3 ON OFF PAPER PRINT ph302 ph METER RANGE / TIME PGM

More information

TDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT. Pool Professional s Meter

TDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT. Pool Professional s Meter TDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT Pool Professional s Meter TRACER EC/TDS/SAL POCKETESTER TM CODE 1749-KIT TABLE OF CONTENTS Introduction... 4 Specifications... 4 Contents... 5 Parts &

More information

Indoor Air Quality Meter/Datalogger

Indoor Air Quality Meter/Datalogger User Manual Indoor Air Quality Meter/Datalogger Model EA80 CC Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the Extech EA80 Indoor Air

More information

Table of Contents. Chapter 1 Introduction General Information... 1 Meter Packing List... 2

Table of Contents. Chapter 1 Introduction General Information... 1 Meter Packing List... 2 Table of Contents Chapter 1 Introduction General Information.................................................. 1 Meter Packing List................................................... 2 Chapter 2 Meter

More information

Operations Manual EcoSense ph100

Operations Manual EcoSense ph100 Operations Manual EcoSense ph100 Portable ph, mv and Temperature Instrument English Français Español Deutsch Italiano WARRANTY The EcoSense ph100 Instrument is warranted for one year from date of purchase

More information

Single Parameter Meters Lightweight and versatile ph, EC and DO meters that can be used in portable, wall-mount and benchtop configurations

Single Parameter Meters Lightweight and versatile ph, EC and DO meters that can be used in portable, wall-mount and benchtop configurations Single Parameter Meters Lightweight and versatile ph, EC and DO meters that can be used in portable, wall-mount and benchtop configurations Digital electrodes edge performs measurements through its unique

More information

User's Guide. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500

User's Guide. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500 User's Guide ExStik EC500 ph/conductivity / TDS / Salinity / Temperature Meter EC500 Introduction Congratulations on your purchase of the ExStik EC500 ph/conductivity/total Dissolved Solids (TDS) / Salinity

More information

CLEAN PH200/ PH500 ph/ mv/ ORP Ion Versatile Water Analysis Meter www.cleaninst.com E-mail: CS@cleaninst.com CLEAN INSTRUMENTS CO. LTD 2F No.140, Zhongxiao St., Zhonghe Dist., New Taipei City,Taiwan ph/

More information

ph, ORP, and TDS ORP Continuous Monitor OPERATING INSTRUCTIONS 8. Warranty 9. Return of Goods

ph, ORP, and TDS ORP Continuous Monitor OPERATING INSTRUCTIONS 8. Warranty 9. Return of Goods ENTER 8. Warranty Oakton Instruments warrants this product to be free from significant deviations in material and workmanship for a period of three year from the date of purchase (excludes electrode).

More information

User Guide. Indoor Air Quality Meter/Datalogger. Model EA80

User Guide. Indoor Air Quality Meter/Datalogger. Model EA80 User Guide Indoor Air Quality Meter/Datalogger Model EA80 Introduction Congratulations on your purchase of the Extech EA80 Indoor Air Quality Meter. This meter measures Carbon Dioxide (CO 2, ppm) levels,

More information

Instruction Manual. ph 11 & ph 110. Hand-held ph / mv / Temperature / RS 232 Meter 68X Rev. 2-11/03. Technology Made Easy...

Instruction Manual. ph 11 & ph 110. Hand-held ph / mv / Temperature / RS 232 Meter 68X Rev. 2-11/03. Technology Made Easy... Instruction Manual 11 & 110 Hand-held / mv / Temperature / RS 232 Meter 68X361301 Technology Made Easy... Rev. 2-11/03 Preface This manual serves to explain the use of the hand-held series meters. The

More information

INSTRUCTION MANUAL CONDUCTIVITY-METER MODEL CC-01

INSTRUCTION MANUAL CONDUCTIVITY-METER MODEL CC-01 INSTRUCTION MANUAL CONDUCTIVITY-METER MODEL CC-01 27618366 / 74 EL-221, MIDC Electronic Zone, Mhape, Navi Mumbai-400701. Tel: CHAPTER 1 INTRODUCTION Contech CC-01 measures Conductivity, Total Dissolved

More information

Operation Manual. Hand-held ph/mv/temperature Meter. ph6810

Operation Manual. Hand-held ph/mv/temperature Meter. ph6810 Operation Manual Hand-held ph/mv/temperature Meter ph6810 CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......2 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS FUNCTIONS..... 4

More information

Resolution selectable from 0.01 and ph. Range to ph. Accuracy ± ph

Resolution selectable from 0.01 and ph. Range to ph. Accuracy ± ph ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 to 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand Manual log on stability Interval logging Temperature readout ( C

More information

Operating Instructions

Operating Instructions Operating Instructions Seven2Go ph Meter S2 Cal METTLER TOLEDO Seven2Go Table of Contents 1 Introduction 5 2 Safety Measures 6 2.1 Definition of signal warnings and symbols 6 2.2 Product specific safety

More information

Research Grade Meters. ph, ISE, EC and DO. Laboratory bench meters with a color, graphic display and capacitive touch keypad

Research Grade Meters. ph, ISE, EC and DO. Laboratory bench meters with a color, graphic display and capacitive touch keypad Research Grade Meters ph, ISE, EC and DO Laboratory bench meters with a color, graphic display and capacitive touch keypad HI5000 Series Research Grade Meters Overview HI5000 Series Research Grade Meters

More information

Water Quality Meter. Water Quality Meter Kit. Instruction Manual

Water Quality Meter. Water Quality Meter Kit. Instruction Manual Water Quality Meter 850081 Water Quality Meter Kit 850086 Instruction Manual Contact : EASTERN ENERGY CO.,LTD. (HEAD OFFICE) 40/4 Vitoondumri Rd., Banbueng, Banbueng, Chonburi 20170 Tel: 0-3844-6117, Fax:

More information

Thermo Scientific Orion Star A223 and Star A323 Portable RDO/DO Meter. Instruction Sheet

Thermo Scientific Orion Star A223 and Star A323 Portable RDO/DO Meter. Instruction Sheet English RDO / DO Thermo Scientific Orion Star A223 and Star A323 Portable RDO/DO Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the appropriate wall socket

More information

Operation Manual. MODEL ph/ec80. Microcomputer Based ph/condctivity/tds/ Sanility & Temperature Pocket Meter. ph/ec80

Operation Manual. MODEL ph/ec80. Microcomputer Based ph/condctivity/tds/ Sanility & Temperature Pocket Meter. ph/ec80 Operation Manual MODEL ph/ec80 Microcomputer Based ph/condctivity/tds/ Sanility & Temperature Pocket Meter ph/ec80 CONTENTS INITIAL INSPECTION AND ASSEMBLY...2 BEFORE YOUR FIRST USE......2 A. Replace the

More information

Resolution selectable from 0.01 and ph. Range ph. Accuracy ± ph

Resolution selectable from 0.01 and ph. Range ph. Accuracy ± ph ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or F) Automatic

More information

Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter

Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter English Cond Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter Instruction Sheet Preparation Power Source 1. Power adapter (sold separately) a. Select the appropriate wall socket

More information

Instruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy...

Instruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy... Instruction Manual ph 5+ ph/ C ph 6+ ph/ C/mV Ion 6+ ph/ C/mV/Ion Technology Made Easy... 68X243633 Rev 1-1 3/2011 Part of Thermo Fisher Scientific Preface This instruction manual serves to explain the

More information

Tracer ph/tds/salt PockeTester TM Code 1766

Tracer ph/tds/salt PockeTester TM Code 1766 Tracer ph/tds/salt PockeTester TM Code 1766 Do Not allow ph sensor to dry out. Store Upright Before first use, tap meter against palm to move electrolyte to tip. See pages 10 and 19. ph/conductivity/tds/salinity

More information

GeneMate Benchtop ph Meters Operation Manual

GeneMate Benchtop ph Meters Operation Manual GeneMate Benchtop ph Meters Operation Manual TABLE OF CONTENTS SAFETY PRECAUTIONS.. 1 GETTING STARTED.... 1 INTRODUCTION..... 2 METER OVERVIEW..... 3 ph, mv & OXYGEN REDUCTION POTENTIAL (ORP) MEASUREMENT....

More information

USER GUIDE. Salinity Meter Pen Style Water Quality Meter. Model EC170

USER GUIDE. Salinity Meter Pen Style Water Quality Meter. Model EC170 USER GUIDE Salinity Meter Pen Style Water Quality Meter Model EC170 Introduction Congratulations on your purchase of the Extech Pen Style Water Quality instrument; the Model EC170 measures Salinity and

More information

Visit us at Back to the Extech DO600 Product Page

Visit us at   Back to the Extech DO600 Product Page User's Guide II ExStik DO600 Dissolved Oxygen Meter RECALL Dissolved Oxygen DO600 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com

More information

HI98190, HI Calibration Check Waterproof ph/mv/ise/temperature Meters INSTRUCTION MANUAL

HI98190, HI Calibration Check Waterproof ph/mv/ise/temperature Meters INSTRUCTION MANUAL HI98190, HI98191 Calibration Check Waterproof ph/mv/ise/temperature Meters INSTRUCTION MANUAL Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully

More information

Bluelab ph Meter TM Instruction Manual

Bluelab ph Meter TM Instruction Manual TM Instruction Manual www.getbluelab.com Table of Contents 1.0 Introduction to Bluelab ph Meter 2 2.0 Preparing the Bluelab ph Meter for use 3 3.0 Calibrating Meter 4 4.0 Measure ph Value 5 5.0 Cleaning

More information

ph/water Quality Meter F-7X High-spec Command Reference

ph/water Quality Meter F-7X High-spec Command Reference CODE:GZ0000438936 February, 2016 ph/water Quality Meter F-7X High-spec Command Reference 2016 HORIBA, Ltd. Preface This manual describes the communication command list of the ph/water Quality Meters with

More information

WalkLAB HC Professional Conductivity-TDS-Salinity Meter. Operations Manual 15,0,11,4,1,14,5,10,13,2,9,6,3,12,7,8

WalkLAB HC Professional Conductivity-TDS-Salinity Meter. Operations Manual 15,0,11,4,1,14,5,10,13,2,9,6,3,12,7,8 15,0,11,4,1,14,5,10,13,2,9,6,3,12,7,8 15,0,15,0,1,14,1,14,13,2,13,2,3,12,3,12,11,4,11,4,5,10,5,10,9,6,9,6,7,8,7,8 WalkLAB Professional Conductivity-TDS-Salinity Meter HC 9021 Operations Manual 0 INTRODUCTION

More information

Bante 221 Portable ph/orp Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

Bante 221 Portable ph/orp Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD Bante 221 Portable ph/orp Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante 221 Portable ph/orp Meter 1 Introduction Thank you for selecting the Bante 221 portable ph/orp meter. This manual provides

More information

AquaShock ph Meter K

AquaShock ph Meter K AquaShock ph Meter 850057K 7720 E. Redfield Rd. Suite #7, Scottsdale, AZ 85260 Tel: (480) 948-4448 Fax: (480) 967-8736 Web: www.sperscientific.com AquaShock ph Meter 850057K Copyright 2011 by Sper Scientific

More information

Handheld Water Quality Meters 100 Series

Handheld Water Quality Meters 100 Series ph ORP Ion Conductivity Resistivity TDS DO Salinity Handheld Water Quality Meters 100 Series METERS 2 YEAR W A R R A N T Y M A D E I N J A P A N www.horiba-laqua.com In the lab, in the field or anywhere

More information

Integrated Instruction Manual WP 600 Series Meters

Integrated Instruction Manual WP 600 Series Meters Integrated Instruction Manual WP 600 Series Meters ph 600 / 610 / 620, COND 600 / 610, DO 600, PC 650, PD 650, CD 650, PCD 650 68X415307 Rev. 4 June 2010 Technology Made Easy... Part of Thermo Fisher Scientific

More information

OPERATION MANUAL. separate probe, RS-232 SOUND LEVEL METER. Model : SL-4013

OPERATION MANUAL. separate probe, RS-232 SOUND LEVEL METER. Model : SL-4013 separate probe, RS-232 SOUND LEVEL METER Model : SL-4013 Your purchase of this SOUND LEVEL METER marks a step forward for you into the field of precision measurement. Although this SOUND LEVEL METER is

More information

Colorado Series. CR-10 Portable ph / mv / Temperature Meter Operation Manual

Colorado Series. CR-10 Portable ph / mv / Temperature Meter Operation Manual Colorado Series CR-10 Portable ph / mv / Temperature Meter Operation Manual 2 Contents Page General Introduction and Initial Inspection.............................. 4 Safety Infomation..................................................

More information

OPERATION MANUAL. ph/mv SD card real time data logger ph METER. Model : PH-230SD

OPERATION MANUAL. ph/mv SD card real time data logger ph METER. Model : PH-230SD ph/mv SD card real time data logger ph METER Model : PH-230SD Your purchase of this ph METER with SD CARD DATA LOGGER marks a step forward for you into the field of precision measurement. Although this

More information

Operation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M

Operation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M Operation Manual Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......2 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS

More information

Quality Electrochemical Analysis Professional Meters Basic Meters Sensors

Quality Electrochemical Analysis Professional Meters Basic Meters Sensors Quality Electrochemical Analysis Professional Meters Basic Meters Sensors Sartorius Electrochemistry: ph Conductivity DO Meters Quality and Convenience Sartorius approach is to offer a complete range of

More information

Quality Electrochemical Analysis Professional Meters Basic Meters Sensors

Quality Electrochemical Analysis Professional Meters Basic Meters Sensors Quality Electrochemical Analysis Professional Meters Basic Meters Sensors Sartorius Electrochemistry: ph Conductivity DO Meters Quality and Convenience Sartorius approach is to offer a complete range of

More information

Operation Manual. MODEL ph6175. Microcomputer Based ph/orp/temperature Benchtop Meter

Operation Manual. MODEL ph6175. Microcomputer Based ph/orp/temperature Benchtop Meter Operation Manual MODEL ph6175 Microcomputer Based ph/orp/temperature Benchtop Meter ph6175 0 CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 POWER INPUT.......2 INSTALLING THE BATTERIES.....3

More information

Professional Benchtop Conductivity meter BC3020

Professional Benchtop Conductivity meter BC3020 15,0,15,0,1,14,1,14,13,2,13,2,3,12,3,12,11,4,11,4,5,10,5,10,9,6,9,6,7,8,7,8 Professional Benchtop Conductivity meter BC3020 Operation Manual 0 INTRODUCTION Your purchase of this professional benchtop Conductivity

More information

Pen type, with separate electrode PH METER Model : PH-220 OPERATION MANUAL

Pen type, with separate electrode PH METER Model : PH-220 OPERATION MANUAL Pen type, with separate electrode PH METER Model : PH-220 Your purchase of this PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and delicate

More information

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS OPERATION MANUAL JENCO MODEL 6230M/6230N/6231M/6231N MICROCOMPUTER BASED ph/mv/temp PORTABLE METER JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS Model 6230M 6230N 6231M 6231N Storage Yes

More information

Benchtop Multi-parameter Water Quality Meter

Benchtop Multi-parameter Water Quality Meter LMPWM-A Series Benchtop multi-parameter water quality meter LMPWM-A series is comes with different measurement modes for quality and testing of water on different parameters. It measures ph, Conductivity,

More information

PH SD Card Datalogger. Instruction Manual

PH SD Card Datalogger. Instruction Manual PH SD Card Datalogger 850060 Instruction Manual PH SD Card Datalogger 850060 Copyright 2010 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or

More information

Humidity + Temp. + Dew Point Pocket size HUMIDITY METER. Model : HT-315

Humidity + Temp. + Dew Point Pocket size HUMIDITY METER. Model : HT-315 Humidity + Temp. + Dew Point Pocket size HUMIDITY METER Model : HT-315 TABLE OF CONTENTS 1. FEATURES...1 2. SPECIFICATIONS... 2 3. FRONT PANEL DESCRIPTION...5 3-1 Display...5 3-2 Power Button...5 3-3 Hold

More information

Operation Manual. Hand-held ph/orp/temperature Meter 6010M

Operation Manual. Hand-held ph/orp/temperature Meter 6010M Operation Manual Hand-held ph/orp/temperature Meter 6010M CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......3 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS FUNCTIONS..... 4

More information

Datalogging Conductivity/ TDS Meter. Instruction Manual

Datalogging Conductivity/ TDS Meter. Instruction Manual Datalogging Conductivity/ TDS Meter 850039 Instruction Manual 1 TABLE OF CONTENTS 1. INTRODUCTION...3 2. PANEL DESCRIPTION...4 3. MEASURING PROCEDURE...5 3-A General Measurement...5 3-B Auto and Manual

More information

Portable Conductivity Meter M1410. User Manual 1.002

Portable Conductivity Meter M1410. User Manual 1.002 Portable Conductivity Meter M1410 User Manual 1.002 General information This manual describes the operation of the Portable Conductivity Meter M1410. Please read this entire manual before operating this

More information

User's Guide. ExStik EC400. Conductivity / TDS / Salinity / Temperature Meter MODE HOLD

User's Guide. ExStik EC400. Conductivity / TDS / Salinity / Temperature Meter MODE HOLD User's Guide ExStik EC400 Conductivity / TDS / Salinity / Temperature Meter MODE HOLD Introduction Congratulations on your purchase of the ExStik EC400 Conductivity/Total Dissolved Solids (TDS) / Salinity

More information

Tracer ph/tds/salt PockeTester TM Code 1766

Tracer ph/tds/salt PockeTester TM Code 1766 Tracer ph/tds/salt PockeTester TM Code 1766 Do Not allow ph sensor to dry out. Store Upright Before first use, tap meter against palm to move electrolyte to tip. See pages 10 and 19. ph/conductivity/tds/salinity

More information

CONDUCTIVITY METER OPERATION MANUAL. Model : CD-4322

CONDUCTIVITY METER OPERATION MANUAL. Model : CD-4322 1,600 point data logger, RS232, AC/DC adapter 2,000 us 20 ms, 100 ms, auto range, auto calibration CONDUCTIVITY METER Model : CD-4322 Your purchase of this CONDUCTIVITY METER marks a step forward for you

More information

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter Operation Manual MODEL 2TX 2-wire Isolated ph/orp Transmitter 0 2TX CONTENTS INITIAL INSPECTION.....2 INTRODUCTION......2 ASSEMBLY...3 PREPARATION....4 CONNECTING THE ELECTRODE...4 CONNECTING THE TEMPERATURE

More information

MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL

MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL 470 350/REV A/10-03 MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL CONTENTS Introduction 1 Specification 1-2 Installation 2 Displays

More information

Using I9131 Sodium Combination ISE on 3200I/M

Using I9131 Sodium Combination ISE on 3200I/M Overview Using I9131 Sodium Combination ISE on 3200I/M This document demonstrates how to use I9131 Sodium Combination ISE on 3200I or 3200 M to measure pna or the concentration of [Na + ]. Applied to 3200I

More information

HI 2300 Auto-ranging Microprocessor EC/TDS/NaCl/ºC Bench Meter

HI 2300 Auto-ranging Microprocessor EC/TDS/NaCl/ºC Bench Meter / LOCK C Instruction Manual HI 2300 Auto-ranging Microprocessor EC/TDS/NaCl/ºC Bench Meter T Functions Select Measurement RANGE Range ALT GLP ATC TC SETUP Start Calibration LOCK Lock Current Range Change

More information

Agilent 3200M Multi-Parameter Analyzer

Agilent 3200M Multi-Parameter Analyzer Agilent 3200M Multi-Parameter Analyzer User Guide Agilent Technologies Notices Agilent Technologies (Shanghai) Co., Ltd. 2013 No part of this manual may be reproduced in any form or by any means (including

More information

Oyster Meter ph-conductivity-tds-salinity-orp(mv)

Oyster Meter ph-conductivity-tds-salinity-orp(mv) User's Guide Oyster Meter ph-conductivity-tds-salinity-orp(mv) Model 341350A INTRODUCTION Congratulations on your purchase of the Extech's Oyster Series meter. This device measures ph, Conductivity, TDS,

More information

Instruction Manual. 24" and 48" Waterproof Electronic Digital Level Model Nos &

Instruction Manual. 24 and 48 Waterproof Electronic Digital Level Model Nos & 7640H-English_Manuals 12/11/13 4:18 PM Page 1 24" and 48" Waterproof Electronic Digital Level Model Nos. 1880-2400 & 1880-4800 Instruction Manual Congratulations on your choice of this Digital Level. We

More information

EC/TDS/SAL TRACERTM POCKETESTER CODE Conductivity/TDS/Salinity/Temperature/Meter

EC/TDS/SAL TRACERTM POCKETESTER CODE Conductivity/TDS/Salinity/Temperature/Meter EC/TDS/SAL TRACERTM POCKETESTER CODE 1749 Conductivity/TDS/Salinity/Temperature/Meter WARNING! This set contains chemicals that may be harmful if misused. Read cautions on individual containers carefully.

More information

User Guide. ExStik II EC400. Conductivity / TDS / Salinity / Temperature Meter

User Guide. ExStik II EC400. Conductivity / TDS / Salinity / Temperature Meter User Guide ExStik II EC400 Conductivity / TDS / Salinity / Temperature Meter Introduction Congratulations on your purchase of the ExStik II EC400 Conductivity/Total Dissolved Solids (TDS) / Salinity /

More information

Table of Contents accumet

Table of Contents accumet AR5 USER Manual Table of Contents accumet Introduction 4 Unpacking the Meter 5 Specifications 6 Getting Started Connectors Attaching ph and ISE electrodes 8 0 Using the Meter Touch Screen Operation Choosing

More information