User Guide. ExStik FL700. Fluoride Meter
|
|
- Laurence Whitehead
- 5 years ago
- Views:
Transcription
1 User Guide ExStik FL700 Fluoride Meter
2 Introduction The model FL700 is a system specifically designed for the quick and accurate measurement of fluoride ions in drinking water and other aqueous samples. Unlike other electrode based systems the FL700 consists of the sensing electrode, measuring electronics, and the display in one convenient package. This meter is shipped fully tested; with proper use, this instrument will provide years of reliable service. Features Automatic temperature compensation (±10 o C of calibration temperature) Automatic calibration Stability sensing to optimize accuracy Internal Datalogger for storing up to 25 readings Direct reading of ppm units Direct reading of relative mv units Automatic shut down after 12 minutes to preserve battery life Internal error detection 2
3 Description Meter Description 1. Battery compartment cover 2. LCD Display 3. MODE/HOLD button 4. CAL/RECALL button 5. ON/OFF button 6. Electrode Retaining Collar (ring) 7. Electrode Sensor (Note: The Electrode storage cap is not shown this diagram) Electrode Sensor Description The sensing electrode is a europium doped lanthanum fluoride single crystal that has been incorporated into a removable sensing module that houses a reference electrode and temperature measurement system. The high resistance electrode signals are impedance converted to a low resistance output in the sensing module to ensure stable and noise free performance. Reagent Tablets The FL700 allows the users to follow the ASTM and EPA standard methodology using any of the TISAB reagents and standards already in use. Also, Extech has developed tablet-form TISAB which contains all of the essential and approved chemicals that are found in the usual TISAB reagents. The dry reagent does not contribute to sample dilution. The benefits of this method are: 1. No volumetric errors 2. Independent of sample size (± 20% of nominal sample volume) 3. Easy to use in the field or laboratory 4. Can be shipped more easily than liquid reagents 5. Lower cost per test LCD Display 1. Bargraph display 2. Measurement units 3. Main display 4. Low (L) and High (H) Calibration icons 5. Data HOLD indicator 6. Low Battery indicator 7. Temperature display
4 Quick Start The following procedure describes a quick-start for the FL700 using the standard TISAB chemistry. Detailed operating instructions can be found on the pages that follow. Preparation 1. Remove the FL700, electrode module and sample cup from its box. Remove caps from module. 2. Fit the electrode module onto the end of the meter body, making sure that the slots line up correctly, and tightly turn the module retaining ring to secure the assembly. 3. Wipe the fluoride crystal and reference junction with a damp tissue Calibration 1. Prepare a 1 ppm fluoride standard and TISAB reagent, or use pre made mixed TISAB and 1 ppm standard 2. Pour mls of this standard solution into the sample cup 3. Rinse the end of the FL700 module in TISAB solution and wipe thoroughly with paper tissue 4. Place the FL700 into the 1 ppm standard 5. Switch the instrument on using the ON/OFF key. The instrument will now go through its internal calibration ppm reading will stabilizes in ~35 seconds and the instrument will enter HOLD mode 7. When in HOLD mode press the CAL key and hold until 1.0 ppm and CAL appear in the display. Release the CAL key 8. Wait until the display stops blinking; the instrument will enter the HOLD mode 9. The instrument is now calibrated and ready for use Measurement 1. Prepare unknown solution by adding TISAB reagent to the sample. Thoroughly wipe the end of the FL Place the FL700 into the prepared unknown sample 3. If the display is indicating HOLD press the HOLD key to enter the Measure mode (the HOLD display will switch off) 4. After ~ 35 seconds the instrument will display the value of the unknown concentration 5. The readings can be stored in memory by pressing the MODE/HOLD key for ~ 3 seconds Storage and maintenance 1. After use, store the electrode in an analyzed sample 2. Thoroughly wipe the sensor with paper tissue. The flat ended sensors can be wiped vigorously 3. The fluoride module can be replaced once the automatic calibration no longer sufficiently calibrates the instrument 4. Other maintenance information is provided in a later section of this guide 4
5 Operation Preparation for use 1. Remove the FL700, electrode module and sample cup from its box. Remove caps from module 2. Fit the electrode module onto the end of the meter body, making sure that the slots line up correctly, and tightly turn the module retaining ring to secure the assembly 3. Wipe the fluoride crystal and reference junction with a damp tissue (cleaning techniques are addressed later in this guide) Powering the FL700 The FL700 uses four (4) CR2032 lithium ion batteries. Press the ON/OFF button to turn the meter on or off. If the batteries are weak, the BAT indicator appears on the LCD. The auto power off feature shuts the meter off automatically after approximately 12 minutes of inactivity. The auto power off feature may be temporarily disabled for convenience or for extended polarization time. Power-On Diagnostics 1. When the meter is switched ON the LCD displays SELF and CAL while the meter runs a diagnostic routine 2. During this time the meter is recalling the user calibration data, performing self diagnostics & initializing the circuitry 3. When completed, the meter proceeds to the normal measurement mode Calibration The FL700 can be calibrated between 1.0 ppm and 10 ppm or between 0.5 and 5.0 ppm Fluoride ion. The following calibration procedure assumes the normal 1.0 to 10 ppm range has been chosen. 1. Prepare a 1 ppm fluoride standard and TISAB reagent, or use pre made mixed TISAB and 1 ppm standard 2. Pour mls of this standard solution into the sample cup 3. Rinse the end of the FL700 module in TISAB solution and wipe thoroughly with paper tissue 4. Place the FL700 in the 1 ppm standard and switch the instrument ON using the ON/OFF key. The instrument will now run its self-calibration 5. The instrument will enter the HOLD mode when stabilized in the 1 ppm solution 6. Press the CAL key; CAL will appear in the display followed by 0.5ppm and 5.0 ppm. Continue holding until 1.0 ppm is shown. Release the CAL key. After the display stops blinking the instrument will enter the HOLD mode 7. The instrument is now calibrated and ready for use. The circled L and H icons on the display indicate that the low range (L) and high range (H) calibrations have been completed Calibration frequency Calibrate the FL700 prior to each new measurement batch or if more than 12 hours has elapsed since the last calibration. 5
6 Slope Adjustment 1. Slope adjustment although not a frequent requirement can be carried out by following the instructions in Calibration steps above except for the fact that a 10 ppm standard is used after calibrating with the 1 ppm standard 2. Press the Cal button until 10 ppm appears. Slope adjustment is then complete Other standards As mentioned the FL700 can also be calibrated between 0.5 and 5.0 ppm F. Follow the calibration instructions above but substitute 0.5 ppm for 1.0 ppm and 5.0 ppm for 10 ppm. Measurements 1. Prepare an unknown solution by adding TISAB reagent to the sample in the same dilution ratio as for the calibration procedure. Mix thoroughly 2. Rinse the end of the FL Place the FL700 into the prepared unknown sample. If the instrument is in the HOLD mode, press MODE/HOLD to unlock HOLD 4. After 25 seconds, the instrument will display the value of the unknown concentration and will then enter the HOLD mode Note: The readings can be stored in the memory by pressing the MODE/HOLD key for ~ 3 seconds as explained in a subsequent section of this user guide. Temperature Units (ºF / ºC) 1. With the unit OFF, press and hold the CAL/RECALL button 2. With the CAL/RECALL button depressed, momentarily press the ON/OFF button to turn the unit ON 3. The CAL/RECALL button can be released when Self Cal is shown in the display 4. To switch back to the previous unit of measure, repeat steps 1 through 3. Auto-Power OFF Feature The auto power off feature automatically shuts the meter off 12 minutes after the most recent button press. Disabling the Auto-Power OFF Feature With the unit ON, momentarily press the CAL/RECALL button, then quickly press and hold both the MODE/HOLD and ON/OFF buttons until off is displayed. To restore the Auto Power Off Feature (auto power OFF enable) simply turn the meter off and on again using the ON/OFF button. Low Battery Indication When the battery voltage falls below the operating threshold, BAT will appear on the display. Refer to the Maintenance section for battery replacement information. 6
7 Storing Readings Up to 25 readings can be stored in memory for later recall. 1. With the meter in the HOLD mode, press and hold the MODE/HOLD button for three (3) seconds to store a reading. Release the button when the memory location number appears on the lower display. 2. After approx. 30 seconds (measurement duration) the meter will return to the HOLD mode and another reading can then be stored. 3. If more than 25 readings are stored, previously stored readings (starting with reading number 1) are overwritten. Recalling Stored Readings 1. Momentarily press the CAL/RECALL button and then within 4 seconds momentarily press the MODE/HOLD. The last stored data point location will be displayed (1 to 25). Each time the MODE/HOLD button is momentarily pressed the next most recently stored data point will be displayed. 2. After the last data point is displayed, pressing the MODE/HOLD button again returns the display to the beginning of the list. 3. Pressing the CAL/RECALL button at anytime stops the data retrieval process and returns the meter to the normal measurement mode. Clearing Stored Readings 1. With the unit ON press and hold the ON/OFF button for 4 seconds 2. When clr is shown in the main display the memory is cleared. 7
8 Maintenance Battery Replacement 1. Twist off the battery compartment cover 2. Holding the battery housing in place with a finger, pull out the battery carrier using the two small tabs 3. Replace the four (4) CR2032 batteries observing proper polarity 4. Replace the battery carrier, reattach the battery compartment cap and tighten securely Electrode Replacement 1. To remove the electrode, first turn the instrument OFF and then unscrew and remove the electrode retaining collar. (turn the collar counter-clockwise to remove) 2. Gently rock the electrode from side to side, pulling it away from the meter until it disconnects 3. To attach an electrode, align the positioning keys on the electrode and the main body housing and then carefully push the electrode into the meter socket until it is fully seated 4. Tighten the electrode retaining collar firmly enough to seal the electrode with the meter Electrode Storage 1. The module and can be stored wet or dry. If stored dry it will be necessary to allow approximately 15 minutes of soaking in a fluoride solution before the specified performance can be achieved. It is recommended that the electrode be stored wet in the last test solution used by the instrument (fluoride ion plus TISAB reagent). 2. The instrument will give an error code when the electrode can no longer be calibrated 3. If the instrument will not calibrate, clean the fluoride electrode surface with the Extech cleaning paste and recalibrate the instrument. If the meter still does not calibrate, replace the electrode. Extech cleaning paste Extech offers a unique cleaning procedure for the lanthanum fluoride crystal that removes any coating from the surface of the crystal and dramatically increases response time. This is available from Extech and a sample of this paste is shipped with the product. 8
9 Specifications Range Accuracy Resolution Display Electrode Electrode life Response Time Operating Temp. Range ATC Range Measurement Storage Battery Power Low Battery Indication Auto Power Off Dimensions/Weight 0.10 to 9.99ppm (mg/l) ± 3% of reading or ± 0.1ppm (whichever is greater) 0.1ppm 2000 count, Dual function 3 ½ digit LCD with Bargraph, Display size: 24 mm x 20 mm Europium doped lanthanum fluoride single crystal 6 months minimum 90% of change in less than 30 seconds (typical) 32 to 140 F (0 to 60 C) 32 to 140 F (0 to 60 C) 25 tagged (numbered) data sets with recall Four (4) CR2032 button batteries BAT appears on the LCD After 12 minutes of inactivity 1.4 x 6.8 x 1.6 (36 x 173 x 41mm); 7.4 oz (210g) 9
10 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a six month limited warranty applies to sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department at (781) ext. 210 for authorization or visit our website for contact information. A Return Authorization (RA) number must be issued before any product is returned to Extech. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied. Calibration and Repair Services Extech offers repair and calibration services for the products we sell. Extech also provides NIST certification for most products. Call the Customer Care Department for information on calibration services available for this product. Extech recommends that annual calibrations be performed to verify meter performance and accuracy. Support line (781) Technical support: Extension 200; support@extech.com Repair & Returns: Extension 210; repair@extech.com Product specifications subject to change without notice For the latest version of this User s Guide, Software updates, and other up-to-the-minute product information, visit our website: Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA Copyright 2007 Extech Instruments Corporation. All rights reserved including the right of reproduction in whole or in part in any form. 10
User Guide. ExStik TM Model RE300 Waterproof ORP Meter. Patent Pending RE300 ORP
0 User Guide ExStik TM Model RE300 Waterproof ORP Meter Patent Pending 500 MV 1000 TM RE300 ORP ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE button 4. CAL
More informationUser's Guide. Cup Thermo-Anemometer. Model AN400
User's Guide Cup Thermo-Anemometer Model AN400 Introduction Congratulations on your purchase of the Extech Cup Thermo-Anemometer. The AN400 measures air velocity in five units of measure: feet per minute
More informationUser Guide. Moisture Meter. Model MO250
User Guide Moisture Meter Model MO250 Introduction Congratulations on your purchase of the Extech MO250 Moisture Meter. The MO250 detects moisture in wood and other building materials such as brick, wall
More informationVisit us at Back to the Extech DO600 Product Page
User's Guide II ExStik DO600 Dissolved Oxygen Meter RECALL Dissolved Oxygen DO600 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com
More informationUser's Guide. ExStik EC400. Conductivity / TDS / Salinity / Temperature Meter MODE HOLD
User's Guide ExStik EC400 Conductivity / TDS / Salinity / Temperature Meter MODE HOLD Introduction Congratulations on your purchase of the ExStik EC400 Conductivity/Total Dissolved Solids (TDS) / Salinity
More informationUser's Guide. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500
User's Guide ExStik EC500 ph/conductivity / TDS / Salinity / Temperature Meter EC500 Introduction Congratulations on your purchase of the ExStik EC500 ph/conductivity/total Dissolved Solids (TDS) / Salinity
More informationUser Guide. ExStik II EC400. Conductivity / TDS / Salinity / Temperature Meter
User Guide ExStik II EC400 Conductivity / TDS / Salinity / Temperature Meter Introduction Congratulations on your purchase of the ExStik II EC400 Conductivity/Total Dissolved Solids (TDS) / Salinity /
More informationUser's Guide. Model Laser Photo Tachometer
User's Guide Model 461920 Laser Photo Tachometer Introduction Congratulations on your purchase of Extech's Mini Laser Photo Tachometer, Model 461920. This Tachometer provides non-contact RPM and Revolution
More informationINSTRUCTION MANUAL. Model True RMS AC/DC 30A Mini Clamp-on Meter. Introduction. True RMS AC Current and Voltage
INSTRUCTION MANUAL Model 380942 True RMS AC/DC 30A Mini Clamp-on Meter True RMS AC Current and Voltage Measure low current with high resolution to 0.1mA AC and 1mA DC Auto Power Off One touch DCA zero
More informationMini Photo Tachometer Model Mini Contact Tachometer Model
User's Guide Mini Photo Tachometer Model 461700 Mini Contact Tachometer Model 461750 Introduction Congratulations on your purchase of the Extech Tachometer. The Model 461700 Photo Tachometer uses precision
More informationTDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT. Pool Professional s Meter
TDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT Pool Professional s Meter TRACER EC/TDS/SAL POCKETESTER TM CODE 1749-KIT TABLE OF CONTENTS Introduction... 4 Specifications... 4 Contents... 5 Parts &
More informationUser s Manual. Model and Sound Level Calibrators
User s Manual Model 407744 and 407766 Sound Level Calibrators Introduction Congratulations on your purchase of Extech s Sound Level Calibrator. Extech Calibrators accommodate Sound Level Meters with 0.5
More informationUser Manual. ExStik EC500. ph/conductivity / TDS / Salinity / Temperature Meter EC500. Additional User Manual Translations available at
User Manual ExStik EC500 ph/conductivity / TDS / Salinity / Temperature Meter EC500 Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the
More informationUser's Guide. Phase Sequence and Motor Rotation Tester Model
User's Guide Phase Sequence and Motor Rotation Tester Model 480403 Introduction Congratulations on your purchase of the Extech Model 408403 Motor and Phase Rotation Indicator. This handheld instrument
More informationEC/TDS/SAL TRACERTM POCKETESTER CODE Conductivity/TDS/Salinity/Temperature/Meter
EC/TDS/SAL TRACERTM POCKETESTER CODE 1749 Conductivity/TDS/Salinity/Temperature/Meter WARNING! This set contains chemicals that may be harmful if misused. Read cautions on individual containers carefully.
More informationUser's Guide. Model High Precision Quad Output DC Power Supply
User's Guide Model 382270 High Precision Quad Output DC Power Supply Introduction Congratulations on your purchase of the Extech 382270 DC Power Supply. The Model 382270 can be used for many applications
More informationSoilStik ph Meter PRODUCT MANUAL
SoilStik ph Meter PRODUCT MANUAL Item # 2105 2 Contents General Overview...3 LCD Display...4 LCD Display Messages...5 Meter Components...6 Calibration...7 Changing the Temperature Units...8 Taking Liquid/Soil
More informationUser's Guide. Datalogging Light Meter with PC Interface. Model
User's Guide Datalogging Light Meter with PC Interface Model 401036 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date
More informationTrue RMS AC Voltage/Current Datalogger
User's Guide True RMS AC Voltage/Current Datalogger Model DL150 Introduction Congratulations on your purchase of this Voltage or Current datalogger. With this meter, you can monitor and log data over long
More informationHeavy Duty Datalogging Light Meter
User's Guide Heavy Duty Datalogging Light Meter with PC Interface Model HD450 Introduction Congratulations on your purchase of the Extech HD450 Digital Light Meter. The HD450 measures illuminance in Lux
More informationUser's Guide. Video Borescope. Model BR100
User's Guide Video Borescope Model BR100 Introduction Congratulations on your purchase of this Extech BR100 Video Borescope. This instrument was designed for use as an inspection device. It can be used
More informationUser s Guide. 600A AC Clamp Meter. Model 38387
User s Guide 600A AC Clamp Meter Model 38387 Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.
More informationSuperHeat Psychrometer Models RH350 and RH355 (kit)
User's Guide SuperHeat Psychrometer Models RH350 and RH355 (kit) Introduction Congratulations on your purchase of the Extech RH350 SuperHeat Psychrometer. This device measures Differential Temperature
More informationHeavy Duty Datalogger Module
User's Guide Heavy Duty Datalogger Module Model 380340 Introduction Congratulations on your purchase of Extech s 380340 Datalogger Module. The Datalogger connects to and records data from Extech Heavy
More informationOyster Meter ph-conductivity-tds-salinity-orp(mv)
User's Guide Oyster Meter ph-conductivity-tds-salinity-orp(mv) Model 341350A INTRODUCTION Congratulations on your purchase of the Extech's Oyster Series meter. This device measures ph, Conductivity, TDS,
More informationUser's Guide. Digital Sound Level Meter. Model
User's Guide Digital Sound Level Meter Model 407730 Introduction Congratulations on your purchase of the Extech 407730 Digital Sound Level Meter. The 407730 measures and displays sound pressure levels
More informationUser's Guide Video Borescope Model BR200
User's Guide Video Borescope Model BR200 Introduction Congratulations on your purchase of this Extech BR200 Video Borescope. This instrument was designed for use as a remote inspection device. It can be
More informationUser's Manual. Model Heavy Duty Datalogging Module with Windows Software
User's Manual Model 380340 Heavy Duty Datalogging Module with Windows Software Stores data for later recall and analysis Can be used with any Extech Heavy Duty meter Selectable recording interval Battery
More informationUser's Guide. 1/16 th DIN ph Controller. Model 48PH2. Introduction
User's Guide 1/16 th DIN ph Controller Model 48PH2 Introduction Thank you for selecting the Extech Model 48PH2 controller. This 1/16 th DIN-size controller offers dual display, high resolution measurements,
More informationUser's Guide. Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer. Patented
User's Guide 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer Patented Introduction Congratulations on your
More informationLaser Photo/ Contact Tachometer. m/min. ft/min
User's Manual Model 461995 Laser Photo / Contact Tachometer Laser Photo/ Contact Tachometer 461995 MEMORY rpm PHOTO CONTACT ft/min m/min Introduction Congratulations on your purchase of Extech's Laser
More informationIntegrating Sound Level Datalogger Model
User's Guide Integrating Sound Level Datalogger Model 407780 Introduction Congratulations on your purchase of the Extech 407780 Integrating Sound Level Meter. The 407780 with programmable integrating time
More informationUser s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389
User s Guide 600A True RMS AC/DC Clamp Meter Model 38389 Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further
More informationUser's Guide. Power Analyzer Model Power Analyzer Datalogger Model Introduction
User's Guide Power Analyzer Model 380801 Power Analyzer Datalogger Model 380803 Introduction Congratulations on your purchase of the Extech 380801 or 380803 Power Analyzer Datalogger. This device offers
More informationTracer TM Salt/EC/TDS PockeTester Code 1749
Tracer TM Salt/EC/TDS PockeTester Code 1749 Pool Professional's Meter! 2 WARNING! This set contains chemicals that may be harmful if misued. Read cautions on individual containers carefully. Not to be
More informationDigital Vane Thermo-Anemometer Model
User's Guide Digital Vane Thermo-Anemometer Model 451104 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment
More informationUser's Guide. Extech AM A AC Analog Clamp Meter
User's Guide Extech AM300 300A AC Analog Clamp Meter Introduction Congratulations on your purchase of the Extech AM300 Analog Clamp Meter. This device measure AC Voltage and Current, DC Voltage, and Resistance.
More informationHeavy Duty Psychrometer + IR Thermometer Model HD500
User Guide Heavy Duty Psychrometer + IR Thermometer Model HD500 Introduction Congratulations on your purchase of the Extech HD500 Psychrometer. This handheld meter measures and displays Air Temperature,
More information3-Axis G-Force Datalogger
User's Guide 3-Axis G-Force Datalogger Model VB300 Introduction Congratulations on your purchase of the VB300 G-Force Datalogger. The Model VB300 can measure and record shock and vibration (acceleration)
More informationUser's Guide. Hot Wire Thermo-Anemometer Model
User's Guide Hot Wire Thermo-Anemometer Model 407123 Introduction Congratulations on your purchase of the Extech Hot Wire Anemometer. This instrument measures air flow and temperature by placing the sensor
More informationHeavy Duty Mini Vane CFM Thermo Anemometer
User's Manual Heavy Duty Mini Vane CFM Thermo Anemometer Model 407117 WARRANTY EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for three years from
More informationUser's Guide. Heavy Duty Vane Thermo-Anemometer. Model
User's Guide Heavy Duty Vane Thermo-Anemometer Model 407112 407112 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date
More informationUser Guide. Heavy Duty Differential Pressure Manometer. Model HD750
User Guide Heavy Duty Differential Pressure Manometer Model HD750 Warranty EXTECH INSTRUMENTS CORPORATION (A FLIR COMPANY) warrants this instrument to be free of defects in parts and workmanship for three
More informationUser's Guide. MiniTec TM Series Model MN25 MultiMeter
User's Guide MiniTec TM Series Model MN25 MultiMeter Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a
More informationUser's Guide. Vibration Meter and Laser Combination Tachometer Model
User's Guide Vibration Meter and Laser Combination Tachometer Model 461880 Introduction Congratulations on your purchase of the Extech 461880 Vibration Meter and Combination Laser Tachometer. The 461880
More informationHeavy Duty Datalogger Module
User's Guide Heavy Duty Datalogger Module Model 380340 Introduction Congratulations on your purchase of Extech s 380340 Datalogger Module. The Datalogger connects to and records data from Extech Heavy
More informationUser's Guide. Heavy Duty Differential Pressure Manometer. Model
User's Guide Heavy Duty Differential Pressure Manometer Model 407910 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date
More informationMini Digital Multimeter
User's Guide Mini Digital Multimeter Model MN15 Introduction Congratulations on your purchase of the Extech MN15 MultiMeter. The MN15 offers AC/DC Voltage, AC/DC Current, Resistance, Diode, and Continuity
More informationTracer ph/tds/salt PockeTester TM Code 1766
Tracer ph/tds/salt PockeTester TM Code 1766 Do Not allow ph sensor to dry out. Store Upright Before first use, tap meter against palm to move electrolyte to tip. See pages 10 and 19. ph/conductivity/tds/salinity
More informationHeavy Duty Vibration Meter
User Guide Heavy Duty Vibration Meter Model 407860 Introduction Congratulations on your purchase of the Extech 407860 Vibration Meter. The Model 407860 measures vibration levels in industrial machinery.
More informationUser's Guide. Hot Wire Thermo-Anemometer Model Introduction
User's Guide Hot Wire Thermo-Anemometer Model 407123 Introduction Congratulations on your purchase of the Extech Hot Wire Anemometer. This meter measures air flow and temperature by placing the sensor
More informationHeavy Duty CFM Thermo-Anemometer Model
User's Guide Heavy Duty CFM Thermo-Anemometer Model 407113 3 Introduction Congratulations on your purchase of the Extech 407113 CFM meter. This handheld meter can display Air Flow (CFM) with Area or Air
More informationUser's Guide. Temperature / Humidity Datalogger. Model 42270
User's Guide Temperature / Humidity Datalogger Model 42270 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment
More informationUser's Manual Sound Level Alert Model SL130
User's Manual Sound Level Alert Model SL130 Meets ANSI and IEC Type 2 Sound Level Meter Standards Settable High Limit with large bright High/Low indication and Alarm Output Wall or Desk mountable Extra
More informationUser Guide. Model Temperature Datalogger Kit Model Temperature and Humidity Datalogger Kit Model SW276 Datalogging Software SW276
User Guide Model 42265 Temperature Datalogger Kit Model 42275 Temperature and Humidity Datalogger Kit Model SW276 Datalogging Software SW276 Introduction Congratulations on your purchase of Extech Instrument
More informationINSTRUCTION MANUAL. Model Dual Input RTD Thermometer. Measures two temperatures simultaneously. Dual RTD probe inputs
INSTRUCTION MANUAL Model 421504 Dual Input RTD Thermometer Measures two temperatures simultaneously Dual RTD probe inputs Clock and Elapsed Timer functions Special functions include Data Hold, MIN/MAX/AVG,
More informationUser's Guide. Programmable DC Power Supply 200 Watt (40 Volts / 5 Amps) Model Introduction
User's Guide Programmable DC Power Supply 200 Watt (40 Volts / 5 Amps) Model 382280 382280 Introduction Congratulations on your purchase of the Extech 382280 Programmable DC Power Supply. This 200 watt
More informationDigital Sound Level Meter
User's Guide Digital Sound Level Meter Model 407732 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 FAX 781.665.0780 - TestEquipmentDepot.com information. A Return Authorization
More informationUser's Guide. Model Noise Dosimeter with PC Interface. Introduction
User's Guide Noise Dosimeter with PC Interface Model 407355 Introduction Congratulations on your purchase of the Extech 407355 Personal Noise Dosimeter. The 407355 is designed to test noise exposure in
More informationTracer ph/tds/salt PockeTester TM Code 1766
Tracer ph/tds/salt PockeTester TM Code 1766 Do Not allow ph sensor to dry out. Store Upright Before first use, tap meter against palm to move electrolyte to tip. See pages 10 and 19. ph/conductivity/tds/salinity
More informationUser's Guide. Vane Thermo-Anemometer Datalogger. Model Introduction
User's Guide Vane Thermo-Anemometer Datalogger Model 451126 Introduction Congratulations on your purchase of Extech's Thermo-Anemometer Datalogger. This Vane-type Anemometer can indicate Air Velocity in
More informationPower Analyzer Datalogger Model
User's Guide Power Analyzer Model 380801 Power Analyzer Datalogger Model 380803 Extech 380803 Appliance Tester/Power Analyzer Extech 380803 True RMS Power Analyzer Data Logger Extech 380803-NIST True RMS
More informationHygro Thermo-Anemometer Model
User's Manual Hygro Thermo-Anemometer Model 407412 WARRANTY EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a
More informationUser's Guide. Mini Microscope. Model MC108
User's Guide Mini Microscope Model MC108 Introduction Congratulations on your purchase of this Extech Meter. This digital microscope with 7 to 27X optical magnification, 4X digital zoom, image capture
More informationWK7200. Water Quality Test Kit USER S MANUAL
WK7200 USER S MANUAL Water Quality Test Kit INTRODUCTION: Thank you for selecting model WK7200 microprocessor-based waterproof Water Quality Test Kit. It is used to measure a wide range of ph, ORP, Conductivity,
More informationAutoranging Multimeter plus IR Thermometer Extech 450 Patented
User's Guide Autoranging Multimeter plus IR Thermometer Extech 450 Patented Introduction Congratulations on your purchase of the Extech 450 (part number EX450) Autoranging Multimeter plus IR Thermometer.
More informationUser's Manual. Digital Sound Level Meter. Model
User's Manual Digital Sound Level Meter Model 407768 Introduction Congratulations on your purchase of the Extech 407768 Sound Level Meter. This Auto Range Sound Level Meter meets ANSI and IEC Type II accuracy
More informationUSER GUIDE. Salinity Meter Pen Style Water Quality Meter. Model EC170
USER GUIDE Salinity Meter Pen Style Water Quality Meter Model EC170 Introduction Congratulations on your purchase of the Extech Pen Style Water Quality instrument; the Model EC170 measures Salinity and
More informationUSER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements
USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements www.adwainstruments.com Dear Customer, Thank you for choosing an Adwa product. Please read carefully this manual before starting
More informationOperations Manual EcoSense ph100
Operations Manual EcoSense ph100 Portable ph, mv and Temperature Instrument English Français Español Deutsch Italiano WARRANTY The EcoSense ph100 Instrument is warranted for one year from date of purchase
More informationManual No Revision MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER
Manual No. 9835 Revision 0-401 MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER Please read this instruction manual carefully before using the instrument. This manual will provide you with all
More informationAPO F T1 F T1-T2 ESC - 0 ENTER. [Limits] CLR? SET[ ] LOG READ SAVE READ SHIFT [OFS] [APO] MAX/MIN REL HOLD [TIME] T1-T2 TYPE T1/T2 TYPE T1/T2
User s Guide Dual Input, 3-Display Thermometer/Datalogger Model 421509 Accepts J, K, T, E, R, S, & N thermocouples Internal Memory stores up to 16 data sets, with 1024 maximum data capacity Backlit Electro-luminescent
More informationAutoranging Mini Multimeter Model MN16A
User's Guide Autoranging Mini Multimeter Model MN16A Introduction Congratulations on your purchase of the Extech MN16A Autoranging Multimeter. This meter measures AC/DC Voltage, AC/DC Current, Resistance,
More informationUSER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe
USER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe www.adwainstruments.com Dear Customer, Thank you for choosing an Adwa product. Please read carefully this manual before starting
More informationUSER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe
USER MANUAL AD 630 DO & Temperature Waterproof Meter with Galvanic Probe ADWA HUNGARY Kft. Alsókikötõ sor 11, 6726 Szeged, Hungary Tel. +36 62 317 878 Fax +36 62 550 610 www.adwainstruments.com MANAD630
More informationOperation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M
Operation Manual Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......2 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS
More informationUser's Guide. Heavy Duty Conductivity / Temperature. Model Introduction
User's Guide 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech 407303 Product Page Heavy Duty Conductivity / Temperature
More informationOperation Manual. Hand-held ph/mv/temperature Meter. ph6810
Operation Manual Hand-held ph/mv/temperature Meter ph6810 CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......2 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS FUNCTIONS..... 4
More informationOperation Manual. MODEL ph 630. Microcomputer Based ph&temperature Pocket Meter. ph630
Operation Manual MODEL ph 630 Microcomputer Based ph&temperature Pocket Meter ph630 CONTENTS INITIAL INSPECTION AND ASSEMBLY...2 BEFORE YOUR FIRST USE... 2 A. Replace the Batteries...2 B. Soak the Electrode......2
More informationWalkLAB ph mv/orptemperature
16,1,16,1,2,15,2,15,14,3,14,3,4,13,4,13,12,5,12,5,6,11,6,11,10,7,10,7,8,9,8,9 WalkLAB ph mv/orptemperature meter HP9000 Operations Manual TABLE OF CONTENTS PAGE 1. INTRODUCTION 1 1.1 Features 1 1.2 Technical
More informationWalkLAB HC Professional Conductivity-TDS-Salinity Meter. Operations Manual 15,0,11,4,1,14,5,10,13,2,9,6,3,12,7,8
15,0,11,4,1,14,5,10,13,2,9,6,3,12,7,8 15,0,15,0,1,14,1,14,13,2,13,2,3,12,3,12,11,4,11,4,5,10,5,10,9,6,9,6,7,8,7,8 WalkLAB Professional Conductivity-TDS-Salinity Meter HC 9021 Operations Manual 0 INTRODUCTION
More informationHI HI HI HI HI HI 98185
Instruction Manual HI 98180 HI 98181 HI 98182 HI 98183 HI 98184 HI 98185 Calibration Check Waterproof ph/mv/ise/temperature Meters www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments
More informationDO Meter Kit. Instruction Manual
DO Meter Kit 850048 Instruction Manual DO Meter Kit 850048 Copyright 2013 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or transmitted in any
More informationph, ORP, and TDS ORP Continuous Monitor OPERATING INSTRUCTIONS 8. Warranty 9. Return of Goods
ENTER 8. Warranty Oakton Instruments warrants this product to be free from significant deviations in material and workmanship for a period of three year from the date of purchase (excludes electrode).
More informationOperation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter
Operation Manual MODEL 2TX 2-wire Isolated ph/orp Transmitter 0 2TX CONTENTS INITIAL INSPECTION.....2 INTRODUCTION......2 ASSEMBLY...3 PREPARATION....4 CONNECTING THE ELECTRODE...4 CONNECTING THE TEMPERATURE
More information315 ph/ion. Installation. Specifications. Installing the Batteries a595. Input Connectors
315 /Ion Installation Input Connectors socket (automatic temperature compensation) socket data socket reference socket If you ordered an electrode with the 315, connect it as follows: Corning Incorporated
More informationPROGRAMMABLE DIGITAL REFRACTOMETER INSTRUCTION MANUAL
PROGRAMMABLE DIGITAL REFRACTOMETER 300037 INSTRUCTION MANUAL 1 CONTENTS INTRODUCTION... 2 PANEL DESCRIPTION... 3 FACEPLATE & LCD DISPLAY...3 OPERATING INSTRUCTIONS General Measurement...4 Calibration Procedures...4
More informationOperation Manual. MODEL ph/ec80. Microcomputer Based ph/condctivity/tds/ Sanility & Temperature Pocket Meter. ph/ec80
Operation Manual MODEL ph/ec80 Microcomputer Based ph/condctivity/tds/ Sanility & Temperature Pocket Meter ph/ec80 CONTENTS INITIAL INSPECTION AND ASSEMBLY...2 BEFORE YOUR FIRST USE......2 A. Replace the
More informationSP-400 Portable Water Analyzer Operation Manual. Rev. A Firmware version 1.02
SP-400 Portable Water Analyzer Operation Manual Rev. A Firmware version 1.02 Pyxis Lab, Inc 120 Jeffrey Ave Holliston, MA 01746 USA www.pyxis-lab.com service@pyxis-lab.com 1 Contents 1 General Description...
More informationOperation Manual. MODEL ph6175. Microcomputer Based ph/orp/temperature Benchtop Meter
Operation Manual MODEL ph6175 Microcomputer Based ph/orp/temperature Benchtop Meter ph6175 0 CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 POWER INPUT.......2 INSTALLING THE BATTERIES.....3
More informationPRO Owner's Manual
PRO 3600 Owner's Manual Feature Overview The Pro 3600 operates normally in a standard reference mode where level is displayed as 0.00. However, a new reference point for 0.00 can easily be established
More informationDual Input, 3-Display Thermometer/Datalogger Model
USER GUIDE Dual Input, 3-Display Thermometer/Datalogger Model 421509 Accepts J, K, T, E, R, S, & N thermocouples Internal Memory stores up to 16 data sets, with 1024 maximum data capacity Backlit Electro-luminescent
More informationDatalogging Conductivity/ TDS Meter. Instruction Manual
Datalogging Conductivity/ TDS Meter 850039 Instruction Manual 1 TABLE OF CONTENTS 1. INTRODUCTION...3 2. PANEL DESCRIPTION...4 3. MEASURING PROCEDURE...5 3-A General Measurement...5 3-B Auto and Manual
More informationPRO 360. Owner's Manual
One-Year Limited Warranty If, within one year from the date of original purchase, the Pro 360 Digital Protractor fails to function because of defects in materials or workmanship, the manufacturer will,
More informationUSER MANUAL AD 3000 EC/TDS/Temperature Bench Meter
USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter www.adwainstruments.com 1 Dear Customer, Thank you for choosing an Adwa product. Please read carefully this manual before starting operations. This instrument
More informationOperations Manual EcoSense ph100a
Operations Manual EcoSense ph100a Portable ph, mv and Temperature Instrument English CONTENTS GENERAL INTRODUCTION... 3 INITIAL INSPECTION... 3 THE INSTRUMENT... 3 BATTERY INSTALLATION... 3 Battery Disposal...
More informationPIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics
PIECAL 322 Automated Thermocouple Calibrator Operating Instructions Product Description Easy to use With the PIECAL 322-1 you can check & calibrate all your thermocouple instruments and measure thermocouple
More informationOperation Manual. Hand-held ph/orp/temperature Meter 6010M
Operation Manual Hand-held ph/orp/temperature Meter 6010M CONTENTS GENERAL INTRODUCTION...2 INITIAL INSPECTION......2 WATER PROOF.......3 INSTALLING THE BATTERIES.....3 DISPLAY & KEYS FUNCTIONS..... 4
More informationPRO 360. Owner's Manual
PRO 360 Owner's Manual Introduction The Pro 360 Digital Protractor is a revolutionary measuring tool that provides an immediate, digital reading of all angles in a 360 circle. The machined aluminum frame
More informationBante 221 Portable ph/orp Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD
Bante 221 Portable ph/orp Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante 221 Portable ph/orp Meter 1 Introduction Thank you for selecting the Bante 221 portable ph/orp meter. This manual provides
More information