Video Plus Series DICOM Conformance Statement
|
|
- Kathlyn Lucas
- 5 years ago
- Views:
Transcription
1 Video Plus Series DICO Conformance Statement Covering the following products: Video Plus Cardiac Video Plus Cardiac/A Video Plus Angio Video Plus RF Video Plus RF/A Camtronics, Ltd. Version 1.1 Revision C Document. No Copyright Camtronics, Ltd. 1997, 1998
2 Table of Contents 0 Introduction Implementation odel Application Data Flow Diagram Functional Definition of Application Entities Sequencing of Real-World Activities Application Entity Specifications VP-DICO Application Entity Specification Association Establishment Policies General Number of Associations Asynchronous Nature Implementation Identifying Information Time-outs Association Initiation Policy Real-World Activity User selects DICO Send option Associated Real-World Activity Proposed Presentation Contexts X-Ray Angiographic Image Storage SOP Class Conformance X-Ray Radiofluoroscopic Image Storage SOP Class Conformance Secondary Capture Image Storage SOP Class Conformance Real-World Activity User selects Archive Patient option Associated Real-World Activity Proposed Presentation Contexts Storage Commitment Push odel SOP Class Conformance Real-World Activity User selects DICO Echo option Associated Real-World Activity Proposed Presentation Contexts Verification SOP Class Conformance Association Acceptance Policy Real-world Activity Associated Real-world Activity Accepted Presentation Contexts Verification SOP Class Conformance Presentation Context Acceptance Criterion Transfer Syntax Selection Policies Communication Profiles TCP/IP Stack API...19 Camtronics, Ltd. Video Plus Series DICO Conformance Statement Revision C Version 1.1 Document. No Page i Copyright Camtronics, Ltd. 1997, 1998
3 3.1.2 Physical edia Support Extensions/Specializations/Privatizations Standard Extended/Specialized/Private SOP s Private Transfer Syntaxes Configuration AE Title/Presentation Address apping Configurable Parameters Support of Extended Character Sets...20 Camtronics, Ltd. Video Plus Series DICO Conformance Statement Revision C Version 1.1 Document. No Page ii Copyright Camtronics, Ltd. 1997, 1998
4 Preface The intent of any DICO conformance statement is to provide a knowledgeable user with the information required to determine whether and to what extent independent DICO implementations may be able to inter-operate. However, the information contained in a DICO conformance statement is not sufficient to ensure independent implementations will, in fact, be able to inter-operate. The user or system integrator must be aware of the following potential issues related to interoperation: Using only the information provided by this Conformance Statement does not guarantee interoperability of Camtronics equipment with non-camtronics equipment. It is the user s (or system integrator s) responsibility to analyze thoroughly the application requirements and objectives to determine if they can be met by the connection of Camtronics equipment to non-camtronics equipment. Camtronics equipment has been tested to assure that the actual implementation of the DICO interface corresponds with this Conformance statement. It is the responsibility of the user (or system integrator) to specify and carry out additional validation testing, which covers a broad spectrum of potential interactions between the independent implementations. Camtronics reserves the right to make changes to its products or to discontinue their delivery. Therefore, the user (or system integrator) should ensure that any future versions of Camtronics or non-camtronics equipment are regression tested to verify that new software releases have not adversely impacted the ability to inter-operate. Camtronics provides this documentation as is without warranty of any kind, either expressed or implied, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose. Some states or countries do not allow disclaimers of express or implied warranties in certain transactions; therefore, this statement may not apply to you. No part of this publication may be reproduced, transmitted, transcribed, recorded, or translated into any language in any form or by any means, electronic, magnetic, optical, chemical, physical or otherwise. Camtronics reserve the right to correct this publication and to make changes to its contents without the obligation of informing anyone of these revisions or changes. Camtronics, Ltd. Video Plus Series DICO Conformance Statement Revision C Version 1.1 Document. No Page iii Copyright Camtronics, Ltd. 1997, 1998
5 0 Introduction This document provides the conformance statement for the Camtronics edical Systems Video Plus Series. Details on the supported DICO service classes and roles are contained within. All of the DICO features mentioned in this document are optional and may not be available on certain products within the Video Plus Series. 1 Implementation odel 1.1 Application Data Flow Diagram The Application Data Flow Diagram for the Video Plus is shown in Figure 1. Local Real-World Activity Local Application Entity Remote Real-World Activity User selects DICO Send option User selects Archive Patient option VP-DICO AE Image stored remotely Remote takes ownership of images Remote responds to verification User selects DICO Echo option Figure 1 Implementation odel DICO Standard Interface Remote requests verification The Video Plus implementation consists of a single Application Entity called VP-DICO. The VP-DICO AE provides all the DICO services for the Video Plus. Document. No Page 1 of 20 Copyright Camtronics, Ltd. 1997, 1998
6 When the user selects the DICO Send option, the VP-DICO AE will initiate an association for the purpose of having an image or study stored remotely. When the user selects the Archive Patient option, the VP-DICO AE will initiate an association for the purpose of requesting that the remote accept responsibility for safekeeping of images. The VP-DICO AE will then expect to receive an association when the remote system returns the result of the storage commitment operation. When the user selects the DICO Echo option, the VP-DICO AE will initiate an association for the purpose of verifying that a remote DICO application is responding to requests. The VP-DICO AE will expect to receive an association when a remote system requests verification but will cause no local real world activity. 1.2 Functional Definition of Application Entities The VP-DICO AE provides all of the necessary functionality to implement the network sending and ownership transfer of the DICO images. This includes, but is not limited to, association management, image and demographics encoding, and response interpretation. The images in the Video Plus are acquired by digitizing a video input signal. Demographic information is provided by the user via the keyboard. 1.3 Sequencing of Real-World Activities Not applicable. 2 Application Entity Specifications 2.1 VP-DICO Application Entity Specification The VP-DICO AE provides Standard Conformance to the following DICO V3.0 SOP Classes as a Service Class User (SCU): SOP Class Name SOP Class UID X-Ray Angiographic Image Storage X-Ray Radiofluoroscopic Image Storage Secondary Capture Image Storage Storage Commitment Push odel Verification The VP-DICO AE also provides Standard Conformance to the following DICO V3.0 SOP Classes as a Service Class Provider (SCP): Document. No Page 2 of 20 Copyright Camtronics, Ltd. 1997, 1998
7 SOP Class Name SOP Class UID Verification Association Establishment Policies General The VP-DICO AE will always propose the DICO Application Context Name (ACN), which is: Application Context Name The maximum length PDU for an association initiated by the VP-DICO AE is: aximum Length PDU 120 Kbytes Number of Associations The VP-DICO AE will initiate only one DICO association at a time to perform an image storage or storage commitment operation as an SCU. If the user requests an echo (verification service class) during an image storage or storage commitment operation, the VP-DICO AE can initiate a second association. A maximum of two DICO associations will be initiated by the VP-DICO AE at any time Asynchronous Nature ultiple outstanding transactions are not supported by the VP-DICO AE Implementation Identifying Information The VP-DICO AE will provide the following Implementation Identifying Information in the User Information Field of the A-ASSOCIATE request primitive during association establishment: Implementation Class UID The VP-DICO AE will have an Implementation Version Name which is formed by appending the software version to the end of the letter V. For example, the Implementation Version Name for software version 4.0 will be V Time-outs The time-outs for the VP-DICO AE apply for all operations and functions are configurable by service personnel. These time-outs limit the amount of time spent waiting for DISE operations to complete. There are specific time-outs for establishing a connection, writing a command request, writing a command response, reading a command request, reading a command response, writing data, reading data, and closing a connection. Document. No Page 3 of 20 Copyright Camtronics, Ltd. 1997, 1998
8 2.1.2 Association Initiation Policy The VP-DICO AE will attempt to establish an association whenever the user selects the DICO Send option, the Archive Patient option, or the DICO Echo option. All three operations will cause the initiation of an association with the same proposed presentation contexts Real-World Activity User selects DICO Send option Associated Real-World Activity When the user is viewing some part of a multi-framed image and selects the DICO Send option, that image will be queued for transfer. When the user has a patient high-lighted on the patient directory screen and selects the DICO Send option, all the images associated with that patient will be queued for transfer. In both the above scenarios, after the user selects the OK button confirming the requested operation, the VP-DICO AE will initiate a connection to the remote node. The images will be sent to the Send Destination configured at the time the operation was queued. If the Auto-archive option is enabled, all acquired images (not reference or photofile images) will be queued for transfer immediately after they are acquired. In this scenario, the user will not be prompted for confirmation of the operation and the images will be sent to the Archive Destination configured at the time the operation was queued. This option does not cause a Storage Commitment operation to occur. The Video Plus has the ability to modify the pixel matrix, bit depth and to some extent SOP class of the object for image transfer. Proper configuration of these options will allow interoperability with destinations with limitations in these areas Proposed Presentation Contexts Abstract Syntax Transfer Syntax Role Extended Name UID Name UID Negotiation X-Ray Angiographic Image Storage X-Ray Radiofluorosco pic Image Storage JPEG Lossless Process 14 (Selection Value 1) Explicit VR Little Endian Implicit VR Little Endian JPEG Lossless Process 14 (Selection Value 1) SCU None SCU None SCU None SCU None Document. No Page 4 of 20 Copyright Camtronics, Ltd. 1997, 1998
9 Abstract Syntax Transfer Syntax Role Extended Name UID Name UID Negotiation Storage Commitment Push odel Explicit VR Little Endian Implicit VR Little Endian Explicit VR Little Endian Implicit VR Little Endian SCU None SCU None SCU None SCU None Verification Implicit VR Little Endian SCU None Note: SCU/SCP role negotiation is performed during association establishment for all of the above listed presentation contexts classes X-Ray Angiographic Image Storage SOP Class Conformance When a Success or Warning (Bxxx) status is received by the VP-DICO AE, the transmission is assumed to be successful. In the case of a warning status, the exact status returned by the remote system is logged in the service log. When a Refused or Error status is received by the VP-DICO AE, the connection is closed and the transmission is retried. In this case, the exact status returned by the remote system is also logged in the service log. Extended negotiation is not supported by the VP-DICO AE for this SOP class. The following tables show the optional elements that may be included for the XA IOD. IE odule Usage Patient Patient Study General Study Series General Series Equipment General Equipment Image General Image Image Pixel Contrast/Bolus Cine ulti-frame ask C - Required if contrast media was used in this image C - Required if pixel data is multi-frame cine data C - Required if pixel data is multi-frame cine data C - Required if the image may be subtracted Document. No Page 5 of 20 Copyright Camtronics, Ltd. 1997, 1998
10 IE odule Usage Display Shutter X-Ray Image X-Ray Acquisition X-Ray Table XA Positioner SOP Common U C - Required if the image is created with table motion Patient odule Patient's Name 0010, Patient ID 0010, Patient's Birth Date 0010, Patient's Sex 0010, General Study Study Instance UID 0020, 000D 1 Study Date 0008, Study Time 0008, Referring Physician s Name 0008, Study ID 0020, Accession Number 0008, Study Description 0008, General Series odality 0008, XA Series Instance UID 0020, 000E 1 Series Number 0020, Laterality 0020, C Performing Physicians(s) Name 0008, Document. No Page 6 of 20 Copyright Camtronics, Ltd. 1997, 1998
11 Physicians(s) Name General Equipment anufacturer 0008, CATRONICS, LTD. Institution Name 0008, General Image Image Number 0020, Patient Orientation 0020, C Sent 0 length Image Date 0008, C Image Time 0008, C Image Comments 0020, Reference Image Sequence >Reference SOP Class UID >Referenced SOP Instance UID 0008, ,1150 1C , C Image Pixel Rows 0028, Columns 0028, Pixel Data 7FE0, Pixel Aspect Ratio 0028, C 1\1, 2\1 Contrast/Bolus odule Contrast/Bolus Agent 0018, Document. No Page 7 of 20 Copyright Camtronics, Ltd. 1997, 1998
12 Cine Frame Time 0018, C ulti-frame Number of Frames 0028, ask ask Subtraction Sequence 28, >ask Operation 28, AVG_SUB >ask Frame Numbers 28, C >Recommended Viewing ode 28, SUB Display Shutter Shutter Shape 0018, CIRCULAR, RECTANGULAR, RECTANGULAR\CIRCULAR Shutter Left Vertical Edge Shutter Right Vertical Edge Shutter Upper Horizontal Edge Shutter Lower Horizontal Edge Center of Circular Shutter Radius of Circular Shutter 0018, C 0018, C 0018, C 0018, C 0018, C 0018, C Document. No Page 8 of 20 Copyright Camtronics, Ltd. 1997, 1998
13 X-Ray Image Frame Increment Pointer 0028,0009 1C (0018, 1063) Image Type 0008, ORIGINAL\PRIARY\SINGLE PLANE, ORIGINAL\PRIARY\BIPLANE A, ORIGINAL\PRIARY\BIPLANE B, DERIVED\SECONDARY\SINGLE PLANE Pixel Intensity Relationship 0028, DISP Samples per Pixel 0028, Photometric Interpretation 0028, ONOCHROE2 Bits Allocated 0028, , 16 Bits Stored 0028, , 10 High Bit 0028, , 9 Pixel Representation 0028, X-Ray Acquisition KVP 0018, Radiation Setting 0018, GR Exposure 0018,1152 2C Sent 0 length X-Ray Table Table otion 0018, Sent 0 length X-Ray Positioner Positioner otion 0018,1500 2C Sent 0 length Positioner Primary Angle Positioner Secondary Angle 0018, Sent 0 length 0018, Sent 0 length Document. No Page 9 of 20 Copyright Camtronics, Ltd. 1997, 1998
14 SOP Common SOP Class UID 0008, SOP Instance UID 0008, X-Ray Radiofluoroscopic Image Storage SOP Class Conformance When a Success or Warning (Bxxx) status is received by the VP-DICO AE, the transmission is assumed to be successful. In the case of a warning status, the exact status returned by the remote system is logged in the service log. When a Refused or Error status is received by the VP-DICO AE, the connection is closed and the transmission is retried. In this case, the exact status returned by the remote system is also logged in the service log. Extended negotiation is not supported by the VP-DICO AE for this SOP class. The following tables show the optional elements that may be included for the XRF IOD. IE odule Usage Patient Patient Study General Study Series General Series Equipment General Equipment Image General Image Image Pixel Contrast/Bolus Cine ulti-frame ask Display Shutter X-Ray Image X-Ray Acquisition SOP Common C - Required if contrast media was used in this image C - Required if pixel data is multi-frame cine data C - Required if pixel data is multi-frame cine data C - Required if the image may be subtracted U Patient odule Patient's Name 0010, Patient ID 0010, Document. No Page 10 of 20 Copyright Camtronics, Ltd. 1997, 1998
15 Patient's Birth Date 0010, Patient's Sex 0010, General Study Study Instance UID 0020, 000D 1 Study Date 0008, Study Time 0008, Referring Physician s Name 0008, Study ID 0020, Accession Number 0008, Study Description 0008, General Series odality 0008, RF Series Instance UID 0020, 000E 1 Series Number 0020, Laterality 0020, C Performing Physicians(s) Name 0008, General Equipment anufacturer 0008, CATRONICS, LTD. Institution Name 0008, General Image Image Number 0020, Patient Orientation 0020, C Sent 0 length Image Date 0008, C Document. No Page 11 of 20 Copyright Camtronics, Ltd. 1997, 1998
16 Image Time 0008, C Image Comments 0020, Reference Image Sequence >Reference SOP Class UID >Referenced SOP Instance UID 0008, ,1150 1C , C Image Pixel Rows 0028, Columns 0028, Pixel Data 7FE0, Pixel Aspect Ratio 0028, C 1\1, 2\1 Contrast/Bolus odule Contrast/Bolus Agent 0018, Cine Frame Time 0018, C ulti-frame Number of Frames 0028, ask ask Subtraction Sequence 28, Document. No Page 12 of 20 Copyright Camtronics, Ltd. 1997, 1998
17 >ask Operation 28, AVG_SUB >ask Frame Numbers 28, C >Recommended Viewing ode 28, SUB Display Shutter Shutter Shape 0018, CIRCULAR, RECTANGULAR, RECTANGULAR\CIRCULAR Shutter Left Vertical Edge Shutter Right Vertical Edge Shutter Upper Horizontal Edge Shutter Lower Horizontal Edge Center of Circular Shutter Radius of Circular Shutter 0018, C 0018, C 0018, C 0018, C 0018, C 0018, C X-Ray Image Frame Increment Pointer 0028,0009 1C (0018, 1063) Image Type 0008, ORIGINAL\PRIARY\SINGLE PLANE, ORIGINAL\PRIARY\BIPLANE A, ORIGINAL\PRIARY\BIPLANE B Pixel Intensity Relationship 0028, DISP Samples per Pixel 0028, Photometric Interpretation 0028, ONOCHROE2 Bits Allocated 0028, , 16 Bits Stored 0028, , 10 High Bit 0028, , 9 Document. No Page 13 of 20 Copyright Camtronics, Ltd. 1997, 1998
18 Pixel Representation 0028, X-Ray Acquisition KVP 0018, Radiation Setting 0018, GR Exposure 0018,1152 2C Sent 0 length SOP Common SOP Class UID 0008, SOP Instance UID 0008, Secondary Capture Image Storage SOP Class Conformance The following tables show the optional elements that may be included for the SC IOD. IE odule Usage Patient Patient Study General Study Series General Series Equipment General Equipment U SC Equipment Image General Image Image Pixel SC Image SOP Common Patient odule Patient's Name 0010, Patient ID 0010, Patient's Birth Date 0010, Document. No Page 14 of 20 Copyright Camtronics, Ltd. 1997, 1998
19 Patient's Sex 0010, General Study Study Instance UID 0020, 000D 1 Study Date 0008, Study Time 0008, Referring Physician s Name 0008, Study ID 0020, Accession Number 0008, Study Description 0008, General Series odality 0008, OT Series Instance UID 0020, 000E 1 Series Number 0020, Laterality 0020, C Performing Physicians(s) Name 0008, General Equipment anufacturer 0008, CATRONICS, LTD. Institution Name 0008, SC Equipment Conversion Type 0008, DV General Image Document. No Page 15 of 20 Copyright Camtronics, Ltd. 1997, 1998
20 Image Number 0020, Image Date 0008, C Image Time 0008, C Image Comments 0020, Image Pixel Samples per Pixel 0028, Photometric Interpretation 0028, ONOCHROE2 Rows 0028, Columns 0028, Bits Allocated 0028, Bits Stored 0028, High Bit 0028, Pixel Representation 0028, Pixel Data 7FE0, SOP Common SOP Class UID 0008, SOP Instance UID 0008, Real-World Activity User selects Archive Patient option Associated Real-World Activity When the user has a patient high-lighted on the patient directory screen and selects the Archive Patient option, a Storage Commitment operation is queued. After the user selects the OK button confirming the requested operation, the VP-DICO AE will initiate a connection to the remote node. The request will be sent to the Archive Destination configured at the time the operation was queued Proposed Presentation Contexts The presentation contexts proposed are the same as the ones specified in section Document. No Page 16 of 20 Copyright Camtronics, Ltd. 1997, 1998
21 Storage Commitment Push odel SOP Class Conformance The VP-DICO AE may request the storage commitment of the following objects. SOP Class Name SOP Class UID X-Ray Angiographic Image Storage X-Ray Radiofluoroscopic Image Storage Secondary Capture Image Storage The Referenced Study Component Sequence Attribute is not supported. The storage commitment transaction will remain valid until the user chooses to cancel the job or the Video Plus is turned off or otherwise loses power or the transaction completes successfully. The optional Storage edia File-set ID and UID Attributes are not supported. When an N-ACTION response is received with anything other than a Success status, the connection is closed and the transmission is retried. When an N-EVENT-REPORT is received indicating no failures (the event type indicates that the operation was successful), the patient study is marked as archived, and is more easily deletable by the user. When an N-EVENT-REPORT is received indicating failures exist (the event type indicates some failures exist), the failed images are automatically requeued along with another storage commit operation Real-World Activity User selects DICO Echo option Associated Real-World Activity When the user is viewing the I/O Status screen, selects the DICO enu option and then selects the DICO Echo option, a screen appears listing all the configured DICO destinations. After a destination is selected and the operation initiated, the VP-DICO AE will initiate a connection to the selected remote node Proposed Presentation Contexts The presentation contexts proposed are the same as the ones specified in section Verification SOP Class Conformance The VP-DICO AE provides standard conformance to the Verification SOP class Association Acceptance Policy The VP-DICO AE is always ready to accept associations for the purpose of receiving a verification request or a storage commitment response. Document. No Page 17 of 20 Copyright Camtronics, Ltd. 1997, 1998
22 Real-world Activity Associated Real-world Activity No real-world activity occurs when a verification request is received from a remote node. The real-world activity that occurs when a storage commitment response is received from a remote system is discussed is section Accepted Presentation Contexts Abstract Syntax Transfer Syntax Role Extended Name UID Name UID Negotiation Verification Implicit VR Little Endian SCP None Storage Commitment Push odel Explicit VR Little Endian Implicit VR Little Endian SCU None SCU None Note: SCU/SCP role negotiation is performed during association establishment for all of the above listed presentation contexts classes Verification SOP Class Conformance The VP-DICO AE provides standard conformance to the Verification SOP class Presentation Context Acceptance Criterion Not applicable Transfer Syntax Selection Policies The Explicit VR Little Endian transfer syntax will be chosen over the Implicit VR Little Endian transfer syntax if the remote system proposes both. 3 Communication Profiles The only communication stack supported by the Video Plus Series is the TCP/IP stack as specified in PS3.8 (DICO Part 8). Neither the OSI nor point-to-point stacks are supported with this product. Document. No Page 18 of 20 Copyright Camtronics, Ltd. 1997, 1998
23 3.1 TCP/IP Stack The TCP/IP stack is Fusion version 3.6 by Pacific Softworks which has been ported to the Video Plus environment API Not applicable Physical edia Support The supported physical medium for the Video Plus Series are: IEEE BASE-FX via II IEEE BASE-TX via II IEEE BASE-T via II 4 Extensions/Specializations/Privatizations 4.1 Standard Extended/Specialized/Private SOP s None. 4.2 Private Transfer Syntaxes None. 5 Configuration The Video Plus maintains several configurable parameters which may be modified by a service person. 5.1 AE Title/Presentation Address apping The AE Title, IP address and TCP port for each destination is configurable on the Video Plus by Camtronics authorized service personnel. For each destination, the following information can be entered: Description AE Title IP address Port Transfer type Document. No Page 19 of 20 Copyright Camtronics, Ltd. 1997, 1998
24 512 x 512, XA and SC images, 8 bit depth 512 x 512, XA images only, 8 bit depth (Basic Cardiac Application Profile) Up to 1024 x 1024, XA images only, 8 bit depth Up to 1024 x 1024, XA and SC images, 8 bit depth Up to 1024 x 1024, XA and SC images, acquired bit depth Up to 1024 x 1024, XRF and SC images, 8 bit depth Up to 1024 x 1024, XRF and SC images, acquired bit depth 5.2 Configurable Parameters The following parameters are configurable by Camtronics authorized service personnel: Local AE title Local IP address TCP listen port Institution Name (Hospital Name) Time-outs Connection establishment (this is the ARTI timer) Command request write Command response write Command request read Command response read Data write Data read Connection close 6 Support of Extended Character Sets The Video Plus Series does not support the use of extended character sets. Document. No Page 20 of 20 Copyright Camtronics, Ltd. 1997, 1998
DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL ADR-1000A /R2 (MIIXR0003EAE)
DICOM CONFORMANCE STATEMENT FOR TOSHIBA DIGITAL RADIOGRAPHY SYSTEM MODEL ADR-1000A /R2 (MIIXR0003EAE) 1998-2004 IMPORTANT! (1) No part of this manual may be copied or reprinted, in whole or in part, without
More informationDICOM CONFORMANCE STATEMENT STORAGE SCU FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A. with XIDF-053A or XIDF-056A (MIIXR0001EAB)
DICOM CONFORMANCE STATEMENT STORAGE SCU FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A with XIDF-053A or XIDF-056A (MIIXR0001EAB) 2000 IMPORTANT! (1) No part of this manual may be copied or reprinted,
More informationThe CT Synergy (V3.1/V3.2/V3.3/V3.4/V3.5) Conformance Statement
The CT Synergy (V3.1/V3.2/V3.3/V3.4/V3.5) Conformance Statement for DICO V3.0 Direction 2210079 Revision 1.0 do not duplicate Copyright 2000 by General Electric Company GE EDICAL SYSTES STATEENT DIR 2210079
More informationTechnical Publications
g GE Medical Systems Technical Publications Direction 2206867 Revision 1.0 DICOM Software V2.0 for DICOM V3.0 Copyright ª 1997 By General Electric Co. Do not duplicate 1 g GE Medical Systems THIS PAGE
More informationVilla Sistemi Medicali DRF 4343
Digitec S.r.l. Via Caduti Lecchesi a Fossoli, 17 23900 LECCO Italy digitec@digiteclecco.com Villa Sistemi Medicali DRF 4343 Network and Media Interchange DICOM Conformance Statement August 13, 2009 Draft
More informationDICOM 3.0 Conformance Statement for PlatinumOne Systems
DICOM 3.0 Conformance Statement for PlatinumOne Systems 0.0 Revision History: Revision ECN Date Description A 3032 09/12/2002 New Release 1.0 Purpose: Define the DICOM Conformance statement associated
More informationSonoWin Picture Archiving System
Page 1 of 13 SonoWin Picture Archiving System DICOM SSCU Module Version 4.1 DICOM Conformance Statement Page 2 of 13 1 DOCUMENT HISTORY...4 2 INTRODUCTION...4 2.1 SOURCES OF INFORMATION...4 2.2 ACRONYMS
More informationDICOM Conformance Statement. DSI Release 4.4. Document Number June Copyright Philips Medical Systems Nederland B.V.
Philips Medical Systems DICOM Conformance Statement DSI Release 4.4 Document Number 4522 220 84061 6 June 1997 Copyright Philips Medical Systems Nederland B.V. 1997 Philips Medical Systems apple PHILIPS
More informationDAR-8000 Series DICOM Conformance Statement
Series DICOM Conformance Statement for 03.01.xx No. : S517-1070 Rev. : D [NO TEXT] 2 0.0 Revision History: Revision Date Description First Edition 2005/08/03 New Release A 2005/11/09 Added Worklist B 2006/03/02
More informationDICOM Conformance Statement
DICOM Conformance Statement DICOM Pro Plus as an SCU Array Corporation Index 0. INTRODUCTION...3 1. IMPLEMENTATION MODEL...4 1.1 APPLICATION DATA FLOW DIAGRAM...4 1.2 FUNCTIONAL DEFINITION OF APPLICATION
More informationTechnical Publications
g GE Medical Systems Technical Publications Direction 2216638-100 Revision 0.0 DSX DICOM Software V2.1 for DICOM V3.0 Copyright ª 1997 By General Electric Co. Do not duplicate 1 g GE Medical Systems THIS
More informationCardiac Workstation Last Updated January 10,
. Camtronics Ltd. P.O. Box 950 900 Walnut Ridge Drive Hartland, WI 53029 Phone: 262-367-0700 Cardiac Workstation Last Updated January 10, 2001......... DICOM Conformance Statement AUTHOR: DATE: APPROVED
More informationCMT MEDICAL TECHNOLOGIES
CMT MEDICAL TECHNOLOGIES DICOM CONFORMANCE STATEMENT ADR / Product Lines Document Number: 81AR800000 CMT MEDICAL TECHNOLOGIES CMT MEDICAL TECHNOLOGIES, 2007 ALL RIGHTS RESERVED IMPORTANT! 1. The information
More informationDICOM Conformance Statement for IMAGEnet 5
DICOM Conformance Statement for IMAGEnet 5 Ver.1.0 2013/6/3 TOPCON Corporation ALL RIGHTS RESERVED Conformance Statement for IMAGEnet 5 1 Contents 1. Foreword... 4 1.1. References... 4 1.2. Abbreviations...
More informationCONFORMANCE STATEMENT
Philips Medical Systems CONFORMANCE STATEMENT BV Family XA Document Number XPR 080-020061 July 15, 2002 Copyright Philips Medical Systems Nederland B.V. 2002 All rights reserved DICOM Conformance Statement
More informationSonoWin Picture Archiving System
Page 1 of 12 SonoWin Picture Archiving System DICOM SSCP Module Version 4.5 DICOM Conformance Statement Page 2 of 12 Table of Contents 1 DOCUMENT HISTORY... 4 2 INTRODUCTION... 4 2.1 SOURCES OF INFORMATION...
More informationMed X Change DRSHD Digital Recording System High Definition DICOM Conformance Statement Document Version: 1.5
Med X Change DRSHD Digital Recording System High Definition DICOM Conformance Statement Document Version: 1.5 All rights reserved Med X Change 2008 Page 1of 31 1.OVERVIEW This document provides the DICOM
More informationDEJARNETTE DICOM/FFP RESEARCH SYSTEMS. DICOM Conformance Statement
DEJARNETTE RESEARCH SYSTEMS DICOM/FFP DICOM Conformance Statement DICOM/FFP Conformance Statement Document Number 090-00-0001 A 7 October 1996 (090-00-0001 A) DICOM_FFP DICOM Conformance Statement.doc
More informationDICOM Conformance Statement
SafeCT TM Product Line DICOM Conformance Statement Medic Vision Imaging Solutions Ltd. February 2010 All rights reserved. Tel: +972.73.7262226 Fax: +972.73.7262262 E-mail: contact@medicvision.com 1 Contents
More informationOASIS V4.0 DICOM Conformance Statement Rev. 3
0BMR Imaging System OASIS V4.0 DICOM Conformance Statement Rev. 3 Tokyo, Japan E1E-BM7403-3 Copyright Hitachi Medical Corporation. 2011. All rights reserved. Revision History Revision Date Change Description
More informationDICOM CONFORMANCE STATEMENT
DICOM CONFORMANCE STATEMENT VERICIS Release 2.2 AUTHOR: DATE: APPROVED BY: DATE: Camtronics, Ltd. Doc. Version 2.0 Document. No. 09610-0045 Copyright Camtronics, Ltd. 2001 Camtronics, Ltd. Doc. Version
More informationTechnical Publication. DICOM Conformance Statement. Trauma 3.0. Document Revision 2. October 5 th, 2010
Technical Publication DICOM Conformance Statement 3.0 Document Revision 2 October 5 th, 2010 2010 Copyright BrainLAB AG October 6, 2010 Document Revision 1 Page 1 of 23 1 Conformance Statement Overview
More informationPhilips Medical Systems DICOM Conformance Statement. Inturis Suite R2.2
Philips Medical Systems DICOM Conformance Statement Inturis Suite R2.2 Document Number XZR 080-010044 8 February 2002 Copyright Philips Medical Systems Nederland B.V. 2002 Page ii DICOM Conformance Statement
More informationVersion 7 DICOM Conformance Statement. Document Version 3.02, June 2009
Version 7 DICOM Conformance Statement Document Version 3.02, June 2009 1 Conformance Statement Overview The application described in this Conformance Statement VEPRO EMR Manager PACS is a collection of
More informationTechnical Publications
Technical Publications Direction LU42818 Revision 1 4.0 Express Insight Copyright 2006 By General Electric Co. Do not duplicate g GE Medical Systems REVISION HISTORY REV DATE REASON FOR CHANGE 1 June 30,
More informationPunctual Dicom Workstation
DICOM Conformance Statement Punctual Dicom Workstation Company Name Product Name Product Version 2.0 Document number 2008-1001 Date June 1, 2009 Punctual Software Inc. Punctual Dicom Workstation TABLE
More informationDICOM Conformance Statement
DICOM Conformance Statement TOPCON Corporation TOPCON SPECULAR MICROSCOPE SP-1P Version 1.10 Document Version: 1 Date: 2014-02-17 1 CONFORMANCE STATEMENT OVERVIEW This document declares conformance to
More informationTechnical Publications
Technical Publications Direction 2009076-070 Revision A Centricity Cardiology Data Management System 2.0 CONFORMANCE STATEMENT for DICOM Copyright 2002 By General Electric Co. Do not duplicate THIS PAGE
More informationDICOM Conformance Statement
DICOM Conformance Statement SCENARIA Tokyo, Japan E1E-HC0002-01 Hitachi, Ltd. 2016. All rights reserved. History Revision Description Date Rev. 1.0 Initial. 2016/04/01 ii E1E-HC0002 Contents History...
More informationFusionXD 2.1 DICOM Conformance Statement
FusionXD 2.1 DICOM Conformance Statement Contents 1 Introduction... 3 1.1 References... 3 1.2 Integration and Features... 3 1.3 Definitions, Acronyms and Abbreviations... 3 2 Services... 4 3 Networking...
More informationFusionXD 2.0 DICOM Conformance Statement
FusionXD 2.0 DICOM Conformance Statement 1 Introduction 1.1 Revision History See the Revision History in the beginning of this document 1.2 References Reference REF1 Document Digital Imaging and Communications
More informationRaffaello Conformance Statement for DICOM V3.0
Raffaello Conformance Statement for DICOM V3.0 Copyright 2003-2006 by I.M.S. s.r.l. DOCUMENT VERSIONS Version Date Author Changes 1.00 04-Oct-03 IMS s.r.l. First Version 1.01 27-Feb-04 IMS s.r.l. Updated
More informationDICOM Conformance Statement August 23, 2010 Version 1.0 ImageWorks Internal document number:
DICOM Conformance Statement August 23, 2010 Version 1.0 ImageWorks Internal document number: 0000062023 ImageWorks 250 Clearbrook Road, Suite 240 Elmsford, NY 10523 (914) 592-6100 1 Revision History Version
More informationDICOM Conformance Statement
DICOM Conformance Statement McKesson Cardiology Release 13.2.1 June 2016 Doc ID DCM-10018 Doc Revision 1.0 Produced in Cork, Ireland Revision History Revision Revision Date Summary of Changes Rev. 1.0
More informationDAR-9400f / DAR-9500f DICOM Conformance Statement Revision J
S517-E057J Apr. 2016 DAR-9400f / DAR-9500f DICOM Conformance Statement Revision J Revision Date: April, 2016 Revision History Date Rev. Comment 2005.10.4 First First Revision 2006.11.10 A Correct the
More informationSCORE 3D Workstation DICOM CONFORMANCE STATEMENT
Document No.: S517-E105 Revision No.: A DIGITAL ANGIOGRAPHY SYSTEM DAR-9500f SCORE 3D Workstation DICOM CONFORMANCE STATEMENT Revision History Rev. Date Content of Change First 2013.09 Newly issued. A
More informationPhilips Medical Systems DICOM Conformance Statement
Philips Medical Systems DICOM Conformance Statement DSI Pro 2.0 Document Number 4522 164 09301 4 February 2002 Copyright Philips Medical Systems Nederland B.V. 2002 Page ii DICOM Conformance Statement
More informationDICOM Conformance Statement
DICOM Conformance Statement ECLOS Tokyo, Japan (SN-C357E) Copyright Hitachi Medical Corporation. 2007, 2008. All rights reserved. History Revision Description Date Rev. 1.0 Initial. 2007/02/01 Rev 2.0
More informationNo. : S Rev. : A. DICOM Conformance Statement For Side Station Rev02.00
No. : S517-1076 Rev. : A DICOM Conformance Statement For Side Station Rev02.00 Overview: Side Station DICOM3.0 Conformance Statement This conformance statement details the Side Station s compliance to
More informationDICOM CONFORMANCE STATEMENT MEDIA STORAGE FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A. with XIDF-037A and XIDF-038A or SRS-1000A
DICOM CONFORMANCE STATEMENT MEDIA STORAGE FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A with XIDF-037A and XIDF-038A or SRS-1000A (MIIXR0002EAC) 2001 IMPORTANT! (1) No part of this manual may
More informationPhilips Medical Systems. Xcelera R1.1L1. Document Number XPR July 2003
Philips Medical Systems DICOM CONFORMANCE STATEMENT Xcelera R1.1L1 Document Number XPR 080-030073 11 July 2003 Copyright Philips Medical Systems Nederland B.V. 2003 All rights reserved 3 DICOM Conformance
More informationCopyright FUJIFILM Corporation, Japan. August, th Edition 897N100760F
DICOM Conformance Statement ***** FDR-1000AWS CR-IR363AWS for DICOM Storage DICOM Storage Commitment DICOM MWM DICOM MPPS DICOM Print DICOM Query / Retrieve DICOM Media Storage (Standard) August, 2010
More informationNo. MIIMS0009EA DICOM CONFORMANCE STATEMENT FOR MODEL TFS-3000 (MIIMS0009EA) TOSHIBA CORPORATION 2001 ALL RIGHTS RESERVED
DICOM CONFORMANCE STATEMENT FOR MODEL TFS-3000 (MIIMS0009EA) TOSHIBA CORPORATION 2001 ALL RIGHTS RESERVED IMPORTANT! (1) No part of this manual may be copied or reprinted, in whole or in part, without
More informationX-Porte DICOM Conformance Statement
21919 30th Dr. SE, Bothell, WA 98021-3904 USA Telephone 1.425.951.1200 Facsimile 1.425.951.1201 www.sonosite.com Document Number: Revision: Title: D10711 D X-Porte DICOM Conformance Statement CHANGE HISTORY:
More informationDICOM Conformance Statement. CT AV Series Product Line. Document Number February 1997
Philips Medical Systems DICOM Conformance Statement CT AV Series Product Line Document Number 4522 983 65911 28 February 1997 Copyright Philips Medical Systems Nederland B.V. 1997 Philips Medical Systems
More informationDICOM Conformance Statement June 24, 2011 Version 0.4 ImageWorks Internal document number:
DICOM Conformance Statement June 24, 2011 Version 0.4 ImageWorks Internal document number: 0000062015 ImageWorks 250 Clearbrook Road, Suite 240 Elmsford, NY 10523 (914) 592-6100 www.imageworkscorporation.com
More informationPrivilege/Prestige. DICOM Conformance Statement. Version Algotec Systems Ltd.
Privilege/Prestige DICOM Conformance Statement Version 1.0.0 Algotec Systems Ltd. April 12, 1995 Elscint Ltd. DICOM Conformance Statement Contents Introduction 1 1. Implementation Model 2 1.1 Application
More informationSIEMENS. DICOM Conformance Statement
SIEMENS Siemens Medical Solutions USA, SY CARD, Ann Arbor DICOM Conformance Statement Revision: 9.0 11-12-10 Table of Contents 1. Introduction... 4 1.1 Purpose of this Document... 4 1.2 Sources for this
More informationWhole Body X-ray CT System Supria. DICOM Conformance Statement
Whole Body X-ray CT System Supria DICOM Conformance Statement Tokyo, Japan (SN-C633E) Copyright Hitachi Medical Corporation. 2013, 2014. All rights reserved. History Revision Description Date Rev. 1.0
More informationTechnical Publications
g GE Medical Systems Technical Publications Direction 2347917-100 Revision 1 Precision 500D for DICOM V3.0 Copyright 2002 By General Electric Co. Do not duplicate THIS PAGE LEFT INTENTIONALLY BLANK i TABLE
More informationNo. MIIMS0002EAG TOSHIBA MEDICAL SYSTEMS CORPORATION ALL RIGHTS RESERVED
DICOM CONFORMANCE STATEMENT FOR MODEL TWS-2000 SERIES MODEL TWS-3000 SERIES MODEL TWS-5000 SERIES MODEL STWS-005 SERIES MODEL TIS-3000 MODEL TIS-4000 (MIIMS0002EAG) TOSHIBA MEDICAL SYSTEMS CORPORATION
More informationKodak Point of Care CR systems. DICOM Conformance Statement
Kodak Point of Care CR systems DICOM Conformance Statement Kodak QC Software Version 2.1.x Aug 21, 2005 Document #AT001008-00 Revision 1.0 Table of Contents REVISION HISTORY 4 ACRONYMS, ABBREVIATIONS AND
More informationAVIA (Dx MM) DICOM 3.0 Conformance Statement
AVIA (Dx MM) DICOM 3.0 Conformance Statement AVIA (Dx MM) is a product of Medasys Digital Systems. Table of Contents 1INTRODUCTION... 4 1.1SCOPE AND AUDIENCE... 4 1.2REFERENCES... 4 1.3ACRONYMS AND ABBREVIATIONS...
More informationDICOM Conformance Statement
DICOM Conformance Statement Application Annex: AneurysmFlow R1.1 On Interventional Workspot R1.5 Koninklijke Philips N.V. 2018 All rights are reserved. Issued by: Philips Medical Systems Nederland BV,
More informationTechnical Publications
Technical Publications Direction 2193691DNT Revision 4 STENOSCOP 6000/9000 MDA and MDA Plus for DICOM V3.0 Copyright 1997 By General Electric Co. Do not duplicate GE Medical Systems 1/42 REVISION HISTORY
More informationDBSWIN DICOM Conformance Statement. DBSWIN DD-DICOM Interface. DICOM Conformance Statement V2.1
DBSWIN DD-DICOM Interface DICOM Conformance Statement V2.1 DÜRR DENTAL Page 1 / 23 Conformance Statement Overview The DBSWIN software controls the digital Dürr Dental imaging products. The DD-DICOM Interface
More informationVersion 9 DICOM Conformance Statement. Version 3.05, September 2015
Version 9 DICOM Conformance Statement Version 3.05, September 2015 1 Conformance Statement Overview The application described in this conformance statement, VEPRO EMR Manager PACS, is a collection of processes
More informationTechnical Publication. DICOM Conformance Statement. Patient Browser 2.1. Document Revision 2. April 13, Copyright BrainLAB AG
Technical Publication DICOM Conformance Statement 2.1 Document Revision 2 April 13, 2011 2011 Copyright BrainLAB AG 1 Conformance Statement Overview This is a conformance statement for the BrainLAB software.
More informationDigital Lightbox 1.0
Technical Publication DICOM Conformance Statement 1.0 Document Revision 1 June 3, 2008 2008 Copyright BrainLAB AG 1 Conformance Statement Overview This is a conformance statement for the BrainLAB software.
More informationThis document contains confidential information that is proprietary to SonoSite. Neither the document nor the information contained therein should be
This document contains confidential information that is proprietary to SonoSite. Neither the document nor the information contained therein should be disclosed or reproduced in whole or in part, without
More informationUscan. DICOM Conformance Statement
DICOM Conformance Statement - Uscan Uscan DICOM Conformance Statement Software Version: 4.1.1 Date: 3 August 2018 Signostics, Inc., a subsidiary of EchoNous, Inc. 8310 154th Ave NE, Suite 200 Redmond,
More informationDICOM Conformance Statement RadWorks 4.0 Product Line
RadWorks 4.0 Product Line Author : Frank Wessels Revision Date : November 1998 Version : 4.0 Document Number : 1996-1998 Applicare Medical Imaging B.V. Name Signature Date Consulting Radiologist Frits
More informationMedical Imaging Consultants, Inc.
IA2000 Conformance Statement 1 Medical Imaging Consultants, Inc. Conformance Statement IA2000 Document: MICI-210-01 Revision: 6.0 Document Status: Approved When printed, this is NOT a controlled copy Reviewers
More informationDx Server for Windows NT DICOM 3.0 Conformance Statement
Dx Server for Windows NT DICOM 3.0 Conformance Statement NAME FONCTION VISA AUTHOR Alain Battistini Senior Engineer CONFORM TO ORIGINAL REVIEW Jean-François Hoh Senior Engineer CONFORM TO ORIGINAL APPROVAL
More informationDAR-9400f/DAR-9500f with Gateway Option DICOM Conformance Statement. Revision D
S517-E068D Apr. 2014 DAR-9400f/DAR-9500f with Gateway Option DICOM Conformance Statement Revision D Revision Date: April, 2014 Copyright (C) 2005-2014 Shimadzu Corporation. All rights reserved Date Rev.
More informationSurgimap. DICOM Conformance Statement. Revision 2.0
Surgimap DICOM Conformance Statement Revision 2.0 Date Name Comment 7/8/2013 Raul Macule Creation 1/12/2016 Raul Macule Wildcard policy and unique keys for higher level support Table of Contents 0. Introduction...
More informationDx Server for Windows NT DICOM 3.0 Conformance Statement
Dx Server for Windows NT DICOM 3.0 Conformance Statement NAME FONCTION VISA AUTHOR Alain Battistini Senior Engineer CONFORM TO ORIGINAL REVIEW Jean-François Hoh Senior Engineer CONFORM TO ORIGINAL APPROVAL
More informationDICOM CONFORMANCE STATEMENT
Document No.: S517-1087 Revision: B DAR-7500 DICOM CONFORMANCE STATEMENT MEDICAL SYSTEMS DIVISION Revision History History Date Revision Description 1 st Edition July 2007 - A July.2008 Entirely recreated.
More informationDICOM Conformance Statement * /02* /02 MADE IN GERMANY
EN DICOM 9000-608-109/02 *9000-608-109/02* MADE IN GERMANY Overview Vet-Exam Intra controls the digital DÜRR MEDICAL imaging products. The Vet-Exam Intra DICOM Interface translates between DICOM applications
More informationFluoroscan InSight V5.0.6 DICOM Conformance Statement
Fluoroscan InSight V5.0.6 DICOM Conformance Statement MAN-02996 Rev 002 Page 1 of 30 1. INTRODUCTION... 4 1.1. Purpose of the Document... 4 1.2. References... 4 1.3. Definitions... 4 2. IMPLEMENTATION
More information1 Introduction Scope and audience References Acronyms and abbreviations 3. 2 Implementation model... 4
NAME FONCTION VISA AUTHOR Alain Battistini Senior Engineer CONFORM TO ORIGINAL REVIEW Jean-François Hoh Senior Engineer CONFORM TO ORIGINAL APPROVAL Pascal Gilbert-Jeantet CONFORM TO ORIGINAL Revision
More informationDICOM Conformance Statement RadWorks 2.1 Product Line
RadWorks 2.1 Product Line Author : Frank Wessels Revision Date : August 1997 Version : 2.1a Document Number : 1996-1997 Applicare Medical Imaging B.V. Name Signature Date Consulting Radiologist Frits Barneveld-
More informationDAR Revision Date: July,
S517-E053B DAR-9000 Revision Date: July, 25. 2005 DICOM service classes in this manual are provided as an optional item of DAR-9000. For your suitable system configuration, please consult with SHIMADZU
More informationTechnical Publications GE MR DICOM CONFORMANCE STATEMENT. Optima MR430s Software Version DOC Revision 1. GE Healthcare.
GE Healthcare Technical Publications DOC0923546 Revision 1 GE MR DICOM CONFORMANCE STATEMENT Optima MR430s Software Version 3.54 Copyright 2011 by General Electric Co. Do not duplicate CONFORMANCE STATEMENT
More informationSep, th Edition 897N101668H
DICOM Conformance Statement Console Advance DR-ID 300CL/700CL 800CL/900CL for DICOM Storage DICOM Storage Commitment DICOM MWM DICOM MPPS DICOM Print DICOM Query / Retrieve DICOM Media Storage DICOM Dose
More informationDICOM CONFORMANCE STATEMENT FOR ZIOBASE 4.0
DICOM CONFORMANCE STATEMENT FOR ZIOBASE 4.0 DICOM Conformance Statement, Ziobase 4.0 1 Copyright 2006-2012, Ziosoft, Inc. 0 DICOM CONFORMANCE STATEMENT OVERVIEW... 4 1 IMPLEMENTATION MODEL... 5 1.1 APPLICATION
More informationTopcon Medical Systems
Topcon Medical Systems EZ Lite 2 Version 1.2 DICOM Conformance Statement November 1, 2013 Rev 1.1 Topcon Medical Systems, Inc. 111 Bauer Drive, Oakland, NJ 07436, USA Phone: (201) 599-5100 Fax: (201) 599-5250
More informationWhole Body X-ray CT System Supria. DICOM Conformance Statement
Whole Body X-ray CT System Supria DICOM Conformance Statement Tokyo, Japan E1E-HC0001-02 Hitachi, Ltd. 2016. All rights reserved. ii E1E-HC0001 History Revision Description Date Rev. 1.0 Initial. 2016/4/1
More informationREGIUS CONSOLE CS-2. Revision History. Date Version Description 02/08/2004 Ver This Conformance Statement applies to DICOM3.0 (2000 version).
Revision History Date Version Description 02/08/2004 Ver.1.00 This Conformance Statement applies to DICOM3.0 (2000 version). REGIUS CONSOLE CS-2 DICOM 3.0 Conformance Statement Ver.1.00 02/08/2004 1 Contents
More informationMed X Change DRSHD Digital Recording System High Definition DICOM Conformance Statement Document Version: 1.4
Med X Change DRSHD Digital Recording System High Definition DICOM Conformance Statement Document Version: 1.4 All rights reserved Med X Change 2008 Page 1of 31 1.OVERVIEW This document provides the DICOM
More informationDICOM CONFORMANCE STATEMENT. BrainLAB PatXfer 5
Technical Publication DICOM CONFORMANCE STATEMENT BrainLAB PatXfer 5 Revision 1.0 BrainLAB AG last change: 22.10.2007 17:48 Page 1 of 11 Table of contents DICOM Conformance Statement 0. INTRODUCTION...3
More informationTable of contents REVISION HISTORY INTRODUCTION DICOM CONFORMANCE STATEMENT IMPLEMENTATION MODEL
JA 02 Table of contents REVISION HISTORY... 4 1 INTRODUCTION... 5 2 DICOM CONFORMANCE STATEMENT... 6 2.1 IMPLEMENTATION MODEL... 6 2.1.1 Application Data Flow Diagram... 6 2.1.2 Fundamental Definition
More informationDICOM Conformance Statement RadWorks Product Line
RadWorks Product Line Version : 5.1 Status : Final Document Number : DCM-1015-050/00-001 1996-2000 Applicare Medical Imaging B.V. Author Position Release Date Signature Jan-Pieter Diender Software Engineer
More informationDICOM Conformance Statement Product Name HCP DICOM Net Version
Title DICOM Conformance Statement Product Name HCP DICOM Net Version 5.00.00 Copyright - 2000-2013 SoftLink International Pvt. Ltd. SoftLink International Pvt. Ltd. Page 1 of 37 2, Anand Park, Aundh, Pune
More informationDICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL SSA-640A V5.0
DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM TM MODEL SSA-640A V5.0 TOSHIBA MEDICAL SYSTEMS CORPORATION 2015 ALL RIGHTS RESERVED Trademarks Viamo is a trademark of Toshiba Medical Systems
More informationInstaPACS / InstaRISPACS / InstaViewer (v Release) DICOM Conformance Statement. Meddiff Technologies Pvt. Ltd.
InstaPACS / InstaRISPACS / InstaViewer (v3.0.22 Release) DICOM Conformance Statement Meddiff Technologies Pvt. Ltd. Table of Contents 1. Introduction... 3 1.1 Scope and Field of Application:... 3 1.2 References
More informationDxMM/DxWin DICOM 3.0 Conformance Statement. Document Reference (Référence du document) : 99/ Oct30/ABA/MM103/398B
DxMM/DxWin DICOM 3.0 Conformance Statement Document Reference (Référence du document) : 99/ Oct30/ABA/MM103/398B Table of Contents (Table des Matières) 1 Introduction 6 1.1 Scope and Audience 6 1.2 References
More informationHCP DICOM Net DICOM Conformance Statement
HCP DICOM Net DICOM Conformance Statement Version 4.00.00 32-bit version for Windows NT/2000/XP Copyright - 1997-2004 SoftLink International Pvt. Ltd. Microsoft, Windows NT, 2000,XP are trademarks of Microsoft
More informationGE Healthcare. Technical Publications. ConnectR Plus Version 5.0 DICOM CONFORMANCE STATEMENT. GE Healthcare IT. Direction DOC Revision 0.
GE Healthcare Technical Publications Direction DOC0509300 Revision 0.1 ConnectR Plus Version 5.0 GE Healthcare IT Copyright 2008. General Electric Company 1 Revision History Revision # Author Description
More informationNumaStore 1.0 DICOM 3.0 Conformance Statement
External Distribution Revision 2 February 19, 2001 NumaStore 1.0 Revision 2 Date February 19, 2001 Author Bill Kilgore, George Panousis Revision History Revision Date By Comment 1 January 13, 2001 Bill
More informationTITAN DICOM Conformance Statement
TITAN DICOM Conformance Statement This document contains confidential information that is proprietary to SonoSite. Neither the document nor the information contained therein should be disclosed or reproduced
More informationTechnical Publications
Technical Publications Direction DOC0304290 Revision 1 Image Vault 5.0 DICOM Copyright 2008 by General Electric Co. Do not duplicate Page 1 of 45 THIS PAGE LEFT INTENTIONALLY BLANK Page 2 of 45 TABLE OF
More informationPOP-Net Essential DICOM Conformance Statement
POP-Net Essential DICOM Conformance Statement Revision 4.0 Software Version 4.0 and after Date: February 9, 2005 Prepared for Page 1 Copyright 2000-2005 EBM Technologies, Inc. Head office Japan office
More informationTechnical Publications
Technical Publications Direction 40472 Revision 6 TM 9.0 CONFORMANCE STATEMENT for DICOM Prodigy Advance Prodigy DPX Duo DPX Bravo DPX-NT DPX-MD+ Copyright ª 2004 By General Electric Co. Do not duplicate
More informationDAR-9400f DICOM Conformance Statement Revision A
S57-E057A DAR-9400f DICOM Conformance Statement Revision A Revision Date: November 0, 2006 Date Rev. Author Comment 2005.0.4 First Go Takayanagi First Revision 2006..0 A Go Takayanagi Correct the description
More informationTechnical Publications
Technical Publications Direction 2316173-100 Revision 8.00 LOGIQ 7 CONFORMANCE STATEMENT for DICOM Do not duplicate Copyright ª 2008 By GE Healthcare THIS PAGE LEFT INTENTIONALLY BLANK REVISION HISTORY
More informationTechnical Publications
g GE Healthcare Technical Publications DOC1987748 Revision 1 DoseWatch 2.2 Copyright 2017 by General Electric Co. Do not duplicate THIS PAGE LEFT INTENTIONALLY BLANK CONFORMANCE STATEMENT OVERVIEW DoseWatch
More informationAloka ProSound DICOM
11/18/09 Page 1 of 17 Aloka ProSound DICOM Aloka ProSound C3 and C3cv Ultrasound Imaging System ProSound Versions 1.0 and 1.1 DICOM 11/18/09 Page 2 of 17 Table of Contents 1 Introduction...4 1.1 Scope...4
More informationStellarPACS DICOM Conformance Statement. Version 1.3, August 2008 SoftTeam Solutions
StellarPACS DICOM Conformance Statement Version 1.3, August 2008 SoftTeam Solutions www.softteam.com Table of Contents 1 INTRODUCTION... 3 1.1 REFERENCE... 3 1.2 DEFINITIONS... 3 1.3 ACRONYMS, ABBREVIATIONS
More informationTechnical Publication. DICOM Conformance Statement. iplan 3.0. Document Revision 1. November 17, Copyright BrainLAB AG
Technical Publication DICOM Conformance Statement Document Revision 1 November 17, 2009 2009 Copyright BrainLAB AG 1 Conformance Statement Overview This is a conformance statement for the BrainLAB software
More information