Comparison of Scatter Correction Methods for CBCT. Author(s): Suri, Roland E.; Virshup, Gary; Kaissl, Wolfgang; Zurkirchen, Luis

Similar documents
DUE to beam polychromacity in CT and the energy dependence

Scatter Correction Methods in Dimensional CT

TEP Hounsfield units. Related topics Attenuation coefficient, Hounsfield units

SYSTEM LINEARITY LAB MANUAL: 2 Modifications for P551 Fall 2013 Medical Physics Laboratory

Spectral analysis of non-stationary CT noise

HIGH RESOLUTION COMPUTED TOMOGRAPHY FOR METROLOGY

Acknowledgments. High Performance Cone-Beam CT of Acute Traumatic Brain Injury

Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography

A study on image quality provided by a kilovoltage cone-beam computed tomography

Radiology. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

CT vs. VolumeScope: image quality and dose comparison

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

Artifact Mitigation in High Energy CT via Monte Carlo Simulation

Applying Hounsfield unit density calibration in SkyScan CT-analyser

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Novel evaluation method of low contrast resolution performance of dimensional X-ray CT

Quality control phantoms and protocol for a tomography system

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University

Impact of X-ray Scatter When Using CT-based Attenuation Correction in PET: A Monte Carlo Investigation

Evaluation of Spectrum Mismatching using Spectrum Binning Approach for Statistical Polychromatic Reconstruction in CT

Beam Attenuation Grid Based Scatter Correction Algorithm for. Cone Beam Volume CT

DUAL energy CT (DECT) is a modality where one and. Empirical Dual Energy Calibration (EDEC) for Cone-Beam Computed Tomography

Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Basics of treatment planning II

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

Computed Tomography & 3D Metrology Application of the VDI/VDE Directive 2630 and Optimization of the CT system

S. Guru Prasad, Ph.D., DABR

Clarify behavioral factor of X-ray scatter in MDCT scanners based on evaluation data by a wide variety of MDCT scanners

Refinement of Imaging Processing of Scatter Correction and Beam Hardening Tools for Industrial Radiography and Cone Beam CT

Ch. 4 Physical Principles of CT

Frequency-based method to optimize the number of projections for industrial computed tomography

CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Effects of the difference in tube voltage of the CT scanner on. dose calculation

COMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Image Acquisition Systems

Simulation of Beam Hardening in Industrial CT with x-ray and Monoenergetic Source by Monte Carlo Code

AN X-ray system with a large area detector, such as is used

Metal Artifact Reduction CT Techniques. Tobias Dietrich University Hospital Balgrist University of Zurich Switzerland

BME I5000: Biomedical Imaging

Introduction to Biomedical Imaging

Fast iterative beam hardening correction based on frequency splitting in computed tomography

Automated Quality Assurance for Image-Guided Radiation Therapy

LAB DEMONSTRATION COMPUTED TOMOGRAPHY USING DESKCAT Lab Manual: 0

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

Recognition and Measurement of Small Defects in ICT Testing

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

Reduction of Metal Artifacts in Computed Tomographies for the Planning and Simulation of Radiation Therapy

X-ray tomographic study of wetting bentonite

X-ray Tomography. A superficial introduction, but sufficient enough to get us started in surgical navigation.

HIGH-SPEED THEE-DIMENSIONAL TOMOGRAPHIC IMAGING OF FRAGMENTS AND PRECISE STATISTICS FROM AN AUTOMATED ANALYSIS

Evaluation of Computed Tomography Techniques for Material Identification in Low Level and Intermediate Level Waste

Corso di laurea in Fisica A.A Fisica Medica 4 TC

Brilliance CT Big Bore.

Digital Laminography and Computed Tomography with 600 kv for Aerospace Applications

CIVA Computed Tomography Modeling

[PDR03] RECOMMENDED CT-SCAN PROTOCOLS

Medical Image Reconstruction Term II 2012 Topic 6: Tomography

Using the ACR CT accreditation phantom for routine image quality assurance on both CT and CBCT imaging systems in a radiotherapy environment

PROVIDE A UNIFORM DOSE IN THE SMALL ANIMAL.

Simplified statistical image reconstruction algorithm for polyenergetic X-ray CT. y i Poisson I i (E)e R } where, b i

DUAL energy X-ray radiography [1] can be used to separate

Optimization of the geometry and speed of a moving blocker system for cone-beam computed tomography scatter correction

An investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory

Research Article An Investigation of Calibration Phantoms for CT Scanners with Tube Voltage Modulation

Digital Image Processing

A new concept for high-speed atline and inline CT for up to 100% mass production process control allowing both 3D metrology and failure analysis

Introduction to XCT and its growth

Enhanced material contrast by dual-energy microct imaging

Correlation between Model and Human Observer Performance on a Lesion Shape Discrimination Task in CT

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Proton dose calculation algorithms and configuration data

CALIBRATIONS FOR ANALYZING INDUSTRIAL SAMPLES ON MEDICAL CT SCANNERS

LABORATORY SYSTEM FOR X-RAY NANOTOMOGRAPHY

Computed Tomography. Principles, Design, Artifacts, and Recent Advances. Jiang Hsieh THIRD EDITION. SPIE PRESS Bellingham, Washington USA

MEDICAL IMAGING 2nd Part Computed Tomography

Lucy Phantom MR Grid Evaluation

Carestream s 2 nd Generation Metal Artifact Reduction Software (CMAR 2)

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

Computer-Tomography I: Principles, History, Technology

Radon Transform and Filtered Backprojection

Investigation of the kinematic system of a 450 kv CT scanner and its influence on dimensional CT metrology applications

CIVA CT, an advanced simulation platform for NDT

Identification of Shielding Material Configurations Using NMIS Imaging

Reconstruction Methods for Coplanar Translational Laminography Applications

Computed tomography (Item No.: P )

Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region

Micro-CT Methodology Hasan Alsaid, PhD

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

Metal Streak Artifacts in X-ray Computed Tomography: A Simulation Study

Coordinate Measuring Machines with Computed Tomography

Research Collection. Localisation of Acoustic Emission in Reinforced Concrete using Heterogeneous Velocity Models. Conference Paper.

An Iterative Approach to the Beam Hardening Correction in Cone Beam CT (Proceedings)

Industrial Computed Tomography Innovations

Transcription:

Research Collection Working Paper Comparison of Scatter Correction Methods for CBCT Author(s): Suri, Roland E.; Virshup, Gary; Kaissl, Wolfgang; Zurkirchen, Luis Publication Date: 2010 Permanent Link: https://doi.org/10.3929/ethz-a-007587136 Rights / License: In Copyright - Non-Commercial Use Permitted This page was generated automatically upon download from the ETH Zurich Research Collection. For more information please consult the Terms of use. ETH Library

Comparison of Scatter Correction Methods for CBCT Roland E. Suri (1), Gary Virshup (2), Wolfgang Kaissl (1) (1) Varian Medical Systems Imaging Laboratory GmbH, Täfernstrasse 7, CH-5405 Baden-Dättwil, Switzerland (2) Varian Medical Systems, Ginzton Technology Center, 2599 Garcia Ave, Mountain View, CA, USA, 94043 In contrast to the narrow fan of clinical Computed Tomography (CT) scanners, Cone Beam scanners irradiate a much larger proportion of the object, which causes several times more X-rays scattering. If this scatter is not corrected, the reconstructed images exhibit artifacts: The middle area of the object becomes darker than the outer area, as the density in the middle of the object is underestimated. We compared three methods of correcting for scatter artifacts. 1) A heuristically-estimated constant was subtracted from each projection image (Uniform Scatter Fraction). 2) A beam-hardening-type correction followed by comparing with a cylindrical norm. 3) A combination of both. 4) Using the projections, the object dimensions were estimated in order to compute the scatter with a physical model. In our preliminary results, the first method significantly reduced scatter artefacts. Method two and method three lead to similar image quality and effectively reduced scatter artifacts. Keywords: Algorithms (ALG), CT, Scatter (SCAT) Preferred presentation type: Oral presentation Description of Purpose In Cone Beam Computed Tomography (CBCT) the whole area of interest is irradiated. This causes scatter contributions to the measured projections which are typically between 40 % and 95 % of the minimual pixel intensity for object sizes ranging from 20 cm diameter to 40 cm diameter (without bowtie and without grid). The distribution of the scatter over the projection is rather uniform (about 15 % deviation from average). Since in the reconstruction algorithm the logarithm of the measured intensities is taken, the scatter fraction impacts the reconstructed image, which is large where the object is thick and small where the object is thin. Therefore, in the reconstructed images the middle area of the object appears darker than it should. This reasoning leads to a simple heuristic method for correcting scatter, which is here called the Uniform Scatter Fraction method. From each projection a uniform image is subtracted with an amplitude that depends on the projection minimum and on the object size. Even if it may be inaccurate, this method should improve the image quality if it is compared to not using a scatter correction at all. Our CBCT system is already using other methods that reduce the impact of scatter on image quality even if they were not specifically designed for this purpose: A beam-hardening correction is performed and the beam-hardening corrected projections are compared with a cylindrical norm of similar size as the object. Therefore, we were interested in the question whether adding the Uniform Scatter Fraction method to correct for scatter would improve image quality. Methods 1) The Uniform Scatter Fraction method assumes that the scatter in each projection is uniform distributed over the image. A uniform distributed scatter is always less than the minimum of the projection. The fraction of the scatter contribution to this minimum is assumed to be a linear function of the average object diameter. Further multiplicative parameters were used to model the influence of grid and bowtie. These model parameters were estimated using about 50 scatter measurements. To exclude contributions from metal markers to the projection minimum, the projection mimimum was defined as the middle of the pixel percentile with the lowest count values. 2) For the Beam Hardening Correction the projection of a cylindric phantom of known size was measured

and these measured absorptions were rescaled to fit the theoretical values. The beam-hardening-corrected projections of a cylindrical norm (of similar size than the object) were divided by the beam-hardeningcorrected projections of the object. 3) To combine both methods, the scatter was subtracted from all projections before further processing: The scatter-corrected projections were used for the Beam Hardening Correction, the Single Norm Calibration and for reconstructing the scan of the object. 4) For each projection, the object thickness was estimated using the known absorptions and its position was estimated using the couch height. The scatter was then computed with a physical model. Scatter Kernels (25 cm x 25 cm field, 100 kv, water plates, only Al wedge filter) 1 Normalized relative intensity 0.8 0.6 0.4 0.2 0-25 -20-15 -10-5 0 5 10 15 20 25 location x on imager [cm] air plate thickness of 10 cm plate thickness of 20 cm plate thickness of 30 cm Figure 1. Scatter Kernel for a squared field of 25 cm times 25 cm and different Plexiglas plates. These kernels were estimated by fitting a physical scatter model by scatter measurements on a Varian Acuity system (data by Waldemar Ulmer, Varian, Baden). Results The same projection data were reconstructed with and without using the Uniform Scatter Fraction method for scatter correction. Without scatter correction, and by using neither a beam hardening correction nor a single norm image, a cupping of about 60 HU values occurred (Figure 2).

(A) No Scatter Correction (B) With Scatter Correction Figure 2. Two reconstructions using the same projections of a head phantom (no Beam Hardening Correction, no Single Norm, no filter, no bowtie, no grid, 125 kv, 80 ma, 10 ms). The reconstruction used a modified Blackman filter, 512x512 resolution, and 2.5 mm slices. Reconstruced by the Varian CBCT Application using a constant Air Norm Image. (A) Without scatter correction a cupping of about 60 HU values occurred (see bottom graph that shows average of the area between the green horizontal lines of the green rectangle). (B) Scatter correction with Uniform Scatter Fraction method eliminated the cupping. A Scatter Fraction of 7/12 was used that was detemined by exploration. The Varian CBCT Application offers to correct for the beam hardening and to normalize the projections by a cylindric norm. Both may reduce scatter artifacts. Therefore, the scatter-corrected projections were used for the beam hardening, to normalize the projections, and to reconstruct the scan. This combined method (method three) did not lead to a visible quality improvement as compared to using only beam hardening and norm corrections (method two) (Figure 3).

No Scatter Correction With Scatter Correction Figure 3. Comparison between without (left) and with (right) scatter correction for the same body phantom. No obvious difference was found in the quality of both images. Both methods similarly reduced cupping, but did not correct for fluctuations of the Hounsfiled Unit (HU) values over the slice. The HU values of the main phantom material had an inaccuracy of about 40 HU over the image (see bottom graph that shows average of the area between the green horizontal lines of the green rectangle for a window of 360 HU). The scatter correction was performed taking the pixel value with the 2000th lowest counts and multiplying this value by 0.75. The scans were taken with 125 kv, 80 ma, 13 ms, half fan, bowtie, and grid. Conclusions Reducing scatter artifacts in projections that were not corrected by any other method (method 1) was quite simple as compared to reducing scatter artifacts if the projection had already been corrected by other correction methods (method 3). The accuracy and stability of the used heuristic scatter correction method may not be sufficient for the second purpose. We estimate that for half fan scans with bowtie and grid, our current heuristic scatter model has an average error in estimating the scatter of about 30 %, which translates to a cupping in the order of 70 HU. If this cannot be improved, it is not sufficient to correct Varian CBCT images using standard settings, as they have typically much better HU value accuracies. For this reason, we are working on a scatter correction method that uses a physical scatter model to improve the accuracy of the scatter estimate. This work has not been submitted elsewhere.